medifa 8000

medifa imaging operating table


800200

Index
I. Safety Instructions ................................... 2 XI. Replacement of Hydraulic Cylinder ............... 23-25
II. Normal Use ............................................. 3 -Trendelenburg Shifting ........................... 23
III. Removing the Table Top ........................ 4 -Lateral Shifting ....................................... 24-25
IV. Removing the Crusting Protector ............ 5 XII. Replacement of the Control Panel ................. 26
V. Removing the Column Cover .................... 6-9 XIII. Replacement of the Hydraulic Power Unit ..... 27
VI. Loosening of the Expansion Bellows ...... 10 XIV. Replacement of Accumulator......................... 28
VII. Oil Level Check ....................................... 11 XV. Replacement of Movable Parts ..................... 29-32
- Refill Oil ........................................... 11 - Replacement of the Castors(front) ........... 29
VIII. Replacement of Hydraulic Cylinder 12-15 - Replacement of the Castors (rear) ............ 30-32
Blocks XVI. Replacement of the External Battery Charger .. 33
- Longitudinal Shifting of table top..... 12-13 XVII. Wiring Diagram Breaker Box (black) ................. 34
- Transverse Shifting of table top ...... 14-15
XVIII. Electrical Wiring diagram ............................... 35
IX. Mounting of the Upper Base ................... 16
XIX. Hydraulic Diagram ......................................... 36-37
X. Replacement of the Way measurement 17-22
XX. List of Spare Parts & Drawings ...................... 38-39
System for ...........................................
XXI. Service/Maintenance Schedule ..................... 40-44
- Longitudinal Shifting ....................... 17
- Service Report ....................................... 40-41
- Transverse Shifting ......................... 18
- Maintenance Schedule .......................... 42-44
- Trendelenburg Shifting ................... 19-20
- Lateral Tilt 21-22
Deutsch
Service Manual - medifa 8000

I. Safety Instructions

The Radioscopy Operating Table medifa 8000 has been manufactured according to the state of the art as well
as to the valid safety specifications and requirements.

Nevertheless, improper handling and/or damages on the radioscopy operating table can lead to safety hazards
for the user and/or other persons and tangible assets.

Only start running the radioscopy operating table when in perfect condition. This table has been solely desig-
ned for normal use. Operators should keep in mind to handle the table in compliance with the safety rules and
recommendations as listed in this manual in order to avoid any hazards which may occur from improper hand-
ling. Any defaults which may affect the safety have to be repaired immediately!

The table comes with conductive double castors, conductive upholstery and an equipotential bonding.The ra-
dioscopy operation table can be used in a Zone M - given the area is equipped with a conductive floor.

Always keep this manual close at hand. Please notice as well all legal and further guidelines and regulations
for accident prevention and pollution control.

Do not replace, change or repair any devices of the above mentioned table without the authorization of the ma-
nufacturer. All repairs and maintenance services need to be carried out by the manufacturer or by authorized
personnel.
Spare parts need to be in accordance with the requirements of the manufacturer. Original spare parts always
guarantee the proper running of the radioscopy operation table.
Keep the mandatory deadlines for recurring controls as stated in this service manual!

Think of an environnement friendly disposal of equipment and faulty parts!

Think of a barrier free surrounding before car-


rying out a downward or tilting motion!

DANGER OF TILTING!

Avoid collisions between parts of the table top and the


column, when carrying out the following adjustments:

 Height adjustment (downwards)


 Tilt adjustment

2
II. Normal Use
In compliance with the fullfilled regulations and rules of the international standard IEC 60601-1 and IEC
60601-2-46 this table is a Class I device, following the classification of the law for medical products
(Medizinproduktegesetz)(MPG).
It has to be listed in a respective index. All maintenance works, new installations, adjustments and mountings
should be reported in that index.
Alway keep in mind, that all of the above mentioned works should be solely carried out by Medifa Service
Technicians to guarantee a long, safe and proper function of the product.

This is an important requirement for all kind of warranty claims!

Operation tables, manufactured by Medifa, have been exclusively designed for human medicine and
should only be operated by authorized personnel in areas equipped according to the known standards.

Authorized personnel by definition are persons who took part in a training for the correct use of the device
by specially skilled staff.

This training will be documented and confirmed in a respective report.

Only use tables in a perfect condition and maintenanced according to the given schedule!

Attention:

This manuel cannot replace the absolutely necessary training for the operation
and maintenance of our products!

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 3
Service Manual - medifa 8000

III. Removing The Table Top

Before starting any repair or maintenance work on the radioscopy operation table it might be necessa-
ry to take off the table plate, to prevent damages. In some cases this will become absolutely necessary
in order to reach special parts.
Adjust the table in the highest position and carry on as follows:

Adjust the table at the headboard with the help of the sliding
mechanism. You will now see the guide rails for the sliding
mechanism under the table top.
unscrew the eight screws.

fig. 1: removing the table top

Then drive the table plate to the end of the foot board. You
will now see the guide rails for the sliding mechanism.
unscrew the six screws.

The cylinder piston for the hydraulic sliding function has to


be taken off. To do so, put off the screw at the end of the
piston.

Note:
To put off the screw, push against the screw with a screwdri-
ver or similar. The piston itself is freely movable.

fig. 2+3: removing the table top

Finally move the table plate to the left (1 x) and to the right
(1x) in order to put off the last two screws through the notch
in the base plate.

Now the plate can be easily removed.

Attention:
before removing the last screws, a second person
should secure the table top to prevent the table top from
falling down.
In such a case serious injuries or damages of the table
top could be the result.
When remounting the table top, all screws should be
fastened first and then should be loosened by a half
turn.
After that, move the table top in all directions before fi-
fig. 4+5: removing the table top nally fastening the screws tightly.

4
IV. Removing The Crusting Protective Cover

After removing the table top, you can start to remove the in-
tegrated crusting protector,
Move the table top with the help of the cross-sliding function-
to the complete right, ( from the patients side of view).
You will find the connection for the crusting protector of the
sliding mechanism.
By loosening the screws you can pull the cables from the
socket. fig. 6: remove crusting protector

The cables can now be pulled off easily from the cable ducts
on the surface of the base plate.
If necessary, move the runners for the longitudinal shifting.
.

Attention:
Do not pull the slides out of the guide rail.This will de-
stroy the guide rail and will make it useless!
Therefore we recommend to protect the guide rail with a
cable tie against the complete pull out. ( Pull the cable tie
through the hole in the middle of the guide rail, while it is
in the middle position!) fig. 7: remove crusting protector

Now loosen the cover panel of the crusting protector. Remo-


ve all screws on the front and on the back.
Now the cover panel can be bend and removed.

Attention:
Before removing the last screw, a second person should
secure the cover panel to prevent it from falling down!

fig. 8+9: remove crusting protector

Now you will see a second cover panel, as well kept together
with two screws, but, in addition, secured by eight more
screws, screwed from below.
Loosen all screws and remove the cover panel by bending it
up!

Attention:
Before removing the last screw, a second person should
secure the cover panel to prevent it from falling down!
fig. 10: remove crusting protector

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 5
Service Manual - medifa 8000

V. Removing The Column Cover

For various repairs the column cover needs to be removed.


Do as follows:

Use the hand switch or the control panel and put the table into the top position. Remove the supply
cable, if necessary. Turn off the table at the base.

Loosen the screws at the back of the middle part of the co-
ver.

fig. 11: open the middle part of the cover

After removing both screws, the cover can be unhooked from


the upper cover by pushing it to the inside and pulling it down
at the same time. Repeat this procedure on the other side
and you will be able to completely uncover the middle part.

Attention:
Be careful to prevent the cover from falling down!

fig. 12+13: unhook the middle part of the


cover

When unhooked, the cover can be bend open around the


lower cover and can be dismantled. .

fig. 14: dismantle the middle part of the


column

6
Remove the screws on the left and on the right to dismantle
the lower cover.

fig. 15: loosening the lower cover

Loose the screws in the back.

fig. 16: loosening the lower cover

Dismantle the cover by bending it up around the column.

fig. 17: dismantle the lower cover

To dismantle the upper cover, the expansion


bellows need to be detached.

Remove all four expanding rivets to dismant-


le the expansion bellows from the upper co-
ver. Use a small slot screw driver or flat
pliers.
fig. 18: loosening the expansion bellows

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 7
Service Manual - medifa 8000

Now the expansion bellows can be pulled over the frame. A

fig. 19: removing the expansion bellows

After removing the expansion bellows from the upper cover,


do the same with the control panel.
Do as follows: Loosen the screws (4 screws) from the inside
of the upper cover.
Now move the control panel towards the headboard, pull the
screw heads through the notch and remove the control panel.
Inside the upper cover

fig. 20: removing the control panel

Put the visible data cable off the jack.

The control panel is now dismantled and can be put aside.

fig. 21: removing the data cable of the


control panel

Now, remove the cables from the illuminated push button,


from the inside.

Just pull each plug off the illuminated push button.

Note:
Keep in mind the plug-in location and order as well as the
respective colours, to prevent a mix up.

fig. 22+23: remove data cable from the


control panel

8
To remove the upper cover the holders of the expansion bel-
lows on the front and in the back have to be dismantled. .

Now the fastening screws on both sides can be removed as


well.

Attention:
Before removing the fastening screws on both sides, a
second person should secure the upper cover to prevent
it from falling down and causing secondary failures!

fig. 24+25+26: dismantling of expansion


bellows

Loose the screws on the back to remove the upper cover.

fig. 28: loosening the upper cover

Bend the cover around the column, like you did with the mi-
ddle and lower covers,in order to remove it.

fig. 29: removing the upper cover

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 9
Service Manual - medifa 8000

VI. Loosening The Expansion Bellows

Before being able to loose the expansion bellows, it is necessary to dismantle the crusting protector of the sli-
ding function.
Please see Chapter IV „Removing the Crusting Protective Cover“.
Do as follows:

Stand in front of the table.


Loosen the front holders of the expansion bellows by remo-
ving the two hexagon bolts on the upper side of the lower
base plate.

fig. 30: loosening the holders of the ex-


pansion bellows

Now move to the back of table and remove the two other he-
xagon bolts on the upper side of the lower base plate.

fig. 31 : loosening the holders of the


expansion bellows

After loosening all fastening screws, the expansion bellows


and holders can be left on the remaining holders on the co-
lumn.

The table construction is now accessible.

fig. 32: : loosening the holders of the


expansion bellows

10
VII. Oil Level Check

In order to carry out an oil level check, expose the table column as stated in Chapter VI „Exposing the taTable
Column“.
Please follow the instructions:

For an oil level check of the hydraulic system do as follow:

method 1:
1. Drive the table into the lowest position.
2. Additionally, tilt the table into the lowest position of the
head rest, as far as possible without the danger of colli-
sion with the floor.
3. Now, tilt the table to the left (from the patients view) as
far as possible without the danger of collision with the fig. 33: oil level check
floor.

All mounted cylinders should be in its lowest position, which


means, the highest possible oil level (max in) in the oil coll-
ecting pan has to be visible!

Attention:
Avoid collisions between the carbon table plate and the
floor!
If all above mentioned functions are in its end position,
serious damage can be the result.
fig. 34: oil level check

Method 2:
1. Drive the table into the highest position.
2. Additionally, tilt the table into the lowest position of the
foot rest.
3. Now, tilt the table to the right (from the patients view).

All mounted cylinders should be in its highest position, which


means, the lowest possible oil level (max out) in the oil collec-
ting pan has to be visible

Note:
The oil level „min out“ indicates the absolute minimum level fig. 35: oil level check
while the table is in the mentioned position.

For refilling, put off the plug and refill oil with a funnel!

oiL type

ARAL VITAM DE 32
Type: HLP-D

fig. 36: oil level check

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 11
Service Manual - medifa 8000

VIII. Replacing the Hydraulic Cylinder (Longitudinal Shifting)

To replace the hydraulic cylinder for the longitudinal shifting, first take off the table top as stated in Chapter III
“Removing The Table Top“.
Then remove the crusting protector and loosen the expansion bellows. For further information please see
Chapter IV „Removing The Crusting protector“, Chapter V“ Removing The Column Cover“ and Chapter
VI “Loosening The Expansion Bellows“.
Do as follows:

Loosen the threaded joint at the end of the hydraulic cylinder.

Note:
While loosening the screw, the piston has to be supported
with a - e.g.- screwdriver in order to disconnect the threaded
joint.

Now remove the fastening screws of one of the two guide


rails of the hydraulic cylinder block.
As soon as the guide rails are taken off the cylinder block can
be removed.

fig. 37+38: removing the hydraulic cylin-


der

In case the hydraulic cylinder is faulty it now can be replaced


by removing the screws of the tube connection.
After removing the screws, the tube can be easily removed
from the block.

Attention:
Should the cylinders or the tubes still be pressurised, oil
might penetrate.
We recommend to have a collecting pan for waste oil at
hand.

fig. 39+40: removing the hydraulic cylin-


der

12
Should the tubes be faulty, they have to be retraced to the
tube connection (quick acting coupling).
You will find those connections in the center part of the table
column.
To separate the couplings just pull the spring of the coupling.

Note:
There will be no penetration of oil while separating the coup-
lings, as they are equipped with a return valve.

Attention:
while working with tubes of the shifting function it is ab-
solutely necessary to install them exactly as provided by
the manufacturer, to prevent damages on the shifting
function.

Note: fig. 41+42: loosening quick acting coup-


We recommend to replace the tubes one by one to make
sure they will be properly installed. Separate the tube from
the coupling and connect the new tube right away. Install the
new tube while dismounting the faulty tube.
This procedure will guarantee the proper installation of the
tube.
Always take care of cable ties and band clamps.

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 13
Service Manual - medifa 8000

VIII. Replacing the Hydraulic Cylinder (Transverse Shifting)

Before replacing the hydraulic cylinder for the transverse shifting function, remove the table top. For further
help, please see Chapter III “Removing the Table Top“.
Also remove the crusting protector and loosen the expansion bellows, as stated in Chapter IV “Removing the
Crusting protector“, Chapter V “Removing the Column Cover“ and Chapter VI „ Loosening the expan-
sion bellows“.
Do as follows:

In order to approach to the cylinder block of the transverse


shifting function, the upper base plate has to be dismantled.

First loosen the connection between the cylinder block and


the upper base plate by loosening the two screws of the
connection of the cylinder from underneath the upper base
plate.
The connection is placed on the left.
Now the upper base plate is freely movable by hand.
fig. 43: loosen driving pin

Attention:
Take care as not to pull the base plate with its rails out of
the guide rails. That would destroy the guide rail system
and render it useless!

Move the base plate by hand to the left side until you will be
able to reach the screws, which connect the guide rails to
upper base plate, from underneath. (3x screws/ rails).

As soon as this procedure is carried out, do the same on the


opposite side.

Attention:
Before removing the last screw, a second person should
secure the base plate, as it is no longer connected to the
frame and therefore prevent it from falling down.

When all fastening screws are removed, the upper base plate
can be detached from the frame.

fig. 44+45: loosening base plate Note:


There are still connections, e.g. cables and hydraulic tubes
attached to base plate.
Put the plate on a place of deposit, prepared in advance and
in the same height as the table frame.i

Attention:
While removing the base plate do not damage any
electrical wires or tubes!

14
After removing the plate, you can start dismounting the cylin-
der block for the transverse shifting.
First remove the two fastening screws at the connection of
the cylinder piston.

fig. 46+47: remove hydraulic cylinder

Now remove one of the two guide rails of the hydraulic cylin-
der block. To do so, loosen the four fastening screws. When
you removed the guide rails, the whole cylinder block can be
taken out.

Should the hydraulic cylinder block be faulty, it can now be


replaced by taking off the screws at the tube connection. Af-
ter removing the screws the tubes can be easily pulled off the
block.

Attention: fig. 48+49: remove hydraulic cylinder


Should the cylinder and/or the tubes still be pressurised,
oil might penetrate.
We recommend to have a collecting pan for waste oil at
hand.

Should the tubes be faulty, they have to be traced back to the


connection point ( quick acting coupling).
To separate the couplings, just pull the spring of the coup-
lings. (please see: Removing the Hydraulic Cylinder, longitu-
dinal shifting).

Note:
While seperating couplings, no oil will pentrate as they are
equipped with a return valve.

Attention:
while working with tubes of the shifting function, it is
absolutely necessary to install them exactly as provided
by the manufacturer, to prevent damages of the shifting
function.
fig. 50+51: remove hydraulic cylinder
.

Note:
We recommend to replace the tubes one by one to make
sure they will be installed properly. Separate the tube from
the coupling and connect the new tube right away. Install the
new tube while dismounting the faulty tube.
This procedure will guarantee the proper installation of the
tube.
Always take care of cable ties and band clamps.

fig. 52: remove hydraulic cylinder

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 15
Service Manual - medifa 8000

IX. Mounting of the Upper base plate

Do as follows:

First, screw back the cylinder connection to the lower base


plate.

fig. 53: fastening hydraulic cylinder


connection

Properly place the base plate onto the guide rails. Now move
to foot position of the table.
Take care that the position screw, which you will see on the
lower base plate,exactly fits into the connection (signal gene-
rator) of the way measurement system for the transverse shif-
ting.
Improper handling might lead to serious problems of the shif-
ting functions.

fig. 54+55: signal generator ( way measu-


rement )

As soon as the upper base plate fits properly, the plate can
be fixed to the guide rails by screws.
In order to reach all screws move the plate by hand.

Attention:
Do not pull out the base plate completely out of the gui-
de rails. That would destroy the guide rail system and
render it useless!

Finally, fix the other cylinder connection to the back of the


upper base plate.

fig. 56+57:

16
X. Replacing the Way Measurement System (Longitudinal Shifting)

Before replacing the way measurement system for the longitudinal shifting the table top has to be taken off.
For further help please see Chapter III. „Removing the Table Top“.
Do as follows:

After taking off the table top, you will be able to look at the
way measurement system and dismantle it.

fig. 58: dismantle way measurement sys-


tem

To replace the way measurement system just remove the


screw at the end.

fig. 59 dismantle way measurement sys-


tem

Now retrace the cable to the connection of the expansion


bellows holder.
There, the cables can be removed by opening the cable
clamps.

fig. 60: dismantle way measurement sys-


tem

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 17
Service Manual - medifa 8000

X. Replacing the Way Measurement System (Transverse Shifting)

Before replacing the way measurement system for the transverse shifting, the table top has to be taken off. For
further help please see Chapter III „Removing the Table Top“.

Also remove the crusting protector and loosen the expansion bellows, as stated in Chapter IV “Removing the
Crusting protector“ and Chapter VI „ Loosening the expansion bellows“.
Do as follows:

First, remove the two fastening screws of the way measure-


ment system on the upper base plate.

Now take off the upper base plate!


For further help please see Chapter VIII „ Replacing the Hyd-
raulic Cylinder (Transverse Shifting).

fig. 61:

The way measurement system is now hanging on the back


of the upper base plate.
Retrace the cable to the connection of the expansion bel-
lows holder.
By opening the cable clamps, the cables can be easily remo-
ved.

Note:
Be sure to install the cables as provided by the manufacturer
to prevent problems with the shifting devices.

Note:
Before remounting the upper base plate the way measure-
ment system has to be fixed to the back of the upper base
plate.
fig. 62+63: Please see Chapter X „Installation of the Upper Base Plate“.

18
X. Replacing the Way Measurement System (Trendelenburg Shifting)

Before replacing the way measurement system for the Trendelenburg shifting, the table top has to be taken
off. For further help please see Chapter III „Removing the Table Top“.

Also remove the crusting protector and loosen the expansion bellows, as stated in Chapter IV “Removing the
Crusting protector“ and Chapter VI „ Loosening the expansion bellows“.

Additionally put off the column cover. For further help, please
see Chapter V „ Removing the Column Cover“.
Do as follows:

Before removing the way measurement system, the respecti-


ve driving pin has to be loosened from the cylinder.
Later on, the position of the driving pin on the cylinder will
indicate the preset „Zero Point Adjustment“. It is absolutely
necessary to place the driving pin correctly in order to gua-
rantee the right position of the „zero point adjustment“.
Therefore, the right position of the driving pin should be
marked in advance by drawing the contour with a pencil.
The tube clamp, which fixes the driving pin to the cylinder,
can now be opened.

Loosen the screws of the way measurement system with the


holder at the piston rod of the cylinder.

The system can be easily removed.

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 19
Service Manual - medifa 8000

Finally, sepeaate the wire of the way measurement system


( Trendelenburg shifting) .

Remove the two screws and open the way measurement


cap. You will now see the plug. Simply put it off.

fig. 66+67:

After separating the way measurement system from the cylin-


der and cutting off the connection, simply screw off the holder
of the driving pin.
Remove the two fastening screws and mount the holder di-
rectly to the driving pin of the new system.

Note:
Should the cable be faulty, too, retrace it back to the connec-
tion and replace it.
See the diagram for the correct placement of the cables in
fig. 68: looseningholder of driving pin the breaker box.

20
X. Replacement of the Way Measurement System (Lateral Shifting)

Before replacing the way measurement system for the lateral shifting, the table top has to be taken off. For
further help please see Chapter III „Removing the Table Top“.
Also remove the crusting protector and loosen the expansion bellows, as stated in Chapter IV “Removing the
Crusting protector“ and Chapter VI „ Loosening the expansion bellows“.
Additionally put off the column cover. For further help, please
see Chapter V „ Removing the Column Cover“.
Do as follows:

Before removing the way measurement system, the respecti-


ve driving pin has to be loosened from the cylinder.
Later on, the position of the driving pin on the cylinder will
indicate the preset „Zero Point Adjustment“. It is absolutely
necessary to place the driving pin correctly in order to gua-
rantee the right position of the „zero point adjustment“.
Therefore, the right position of the driving pin should be
marked in advance by drawing the contour with a pencil.
The tube clamp, which fixes the driving pin to the cylinder,
can now be opened.

Loosen the screws of the way measurement system with the


the holder at the piston rod of the cylinder.

The system can be easily removed.

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 21
Service Manual - medifa 8000

Finally, separate the wire of the way measurement system


( lateral shifting) .

Remove the two screws and open the way measurement


cap. You will now see the plug. Simply put it off.

fig. 71+72: disconnect way measurement


system

After separating the way measurement system from the cylin-


der and cutting off the connection, simply screw off the holder
of the driving pin.
Remove the two fastening screws and mount the holder di-
rectly to the driving pin of the new system.

Note:
Should the cable be faulty,too, retrace it back to the connec-
tion and replace it.
fig. 73: loosen holder of driving pin See the diagram for the correct placement of the cables in
the breaker box.

22
XI. Replacement of the Hydraulic Cylinder (Trendelenburg shifting)

Before replacing the hydraulic cylinder for the Trendelenburg shifting the table top has to be taken off. For
further help please see Chapter III „Removing the Table Top“.

Also remove the crusting protector and loosen the expansion bellows, as stated in Chapter IV “Removing the
Crusting protector“ and Chapter VI „ Loosening the expansion bellows“.

Additionally put off the column cover and separate the way-
measurement system from the cylinder. For further help,
please see Chapter V. „ Removing the Column Cover“ and
Chapter X. „Replacement of the Way Measurement Sys-
tem“.
Do as follows:

Unscrew the lower suspension of the cylinder. Remove the


snap ring on the opposite side of the hydraulic power unit.
Now remove the bolt by pushing it through to the other side.

Attention:
As soon as the bolt has been removed from the lower
suspension, the upper structure of the table is freely mo-
vable on the Trendelenburg axle.
To prevent injuries and damages, a second person
should secure the the upper structure.

fig. 74+75: dismantle hydraulic cylinder

Now unscrew the upper suspension.


Remove the screw which fixes the cylinder to the frame.
When unscrewed, it can be removed and behind it you will
find a hollow bolt. Remove it as well.
Interlock the screw in the threads of the hollow bolt und pull it
out of the suspension.

Attention:
Before pulling the bolt out of the connection, a second
person should secure the cylinder to prevent it from
falling down.

Retrace the hydraulic wires to the magnet valve block und


unscrew them to completely dismantle the cylinder.

Attention:
While dismantling the cylinder, oil might penetrate. The
cylinder should not be pressurized when dismantled!

fig. 76+77: dismantle hydraulic cylinder

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 23
Service Manual - medifa 8000

XI. Replacement of the Hydraulic Cylinder (Lateral Shifting)

Before replacing the hydraulic cylinder for the lateral shifting, the table top has to be taken off. For further help
please see Chapter III. „Removing the Table Top“.
Also remove the crusting protector and loosen the expansion bellows, as stated in Chapter IV. “Removing the
Crusting protector“ and Chapter VI. „ Loosening the Expansion Bellows“.
Additionally, put off the column cover and separate the way-
measurement system from the cylinder. For further help, plea-
se see Chapter V .„ Removing the Column Cover“ and Chap-
ter X. „Replacement of the Way Measurement System“.
Do as follows

Remove the four screws which fix the base plate to the arti-
culated frame of the lift column.

Attention:
The plate, e.g. the whole upper structure, is lying on the
articulated frame of the column, only fixed to the base
frame by tubes and cables.

fig. 78+79: dismantle hydraulic cylinder

As soon as the plate is unscrewed it can be slightly moved (a


few centimetres) with care, so that the screws of the suspen-
sion of the lateral cylinder will be visible.
Unscrew the two fastening screws which connects the outer
cheek of the upper suspension of the lateral cylinder to the
articulated frame.
Remove the snap ring at the end of the bolt and pull the
cheek from the bolt.

fig. 80: dismantle hydraulic cylinder

Take care of the inserts of the eye screw, which fixes the cy-
linder to the bolt.
It might lead to an unwanted instability of the table, should
they get lost.

Note:
Take care of the little bolt of the pulled off cheek of the sus-
pension. This little bolt should fit into the groove of the insert
of the bolt, to hold it in position!

fig. 81: dismantle hydraulic cylinder

24
now detach the lower suspension of the cylinder.
Remove one of the snap rings, which fix the bolt of the lower
suspension.
Remove the snap ring from the side of the magnet valve
block.
This will guarantee,that the bolt can be removed on the other
side, by driving it through, without hindrance.

Attention:
Before completely detaching the bolt from the suspensi- fig. 82: dismantle hydraulic cylinder
on, a second person should secure the upper structure.
After removing the bolt, the upper structure is freely mo-
vable on the lateral axle.
Please note: The upper plate is not connected to the
base frame, therefore, please secure directly at the arti-
culated frame.

After removing the bolt of the lower suspension, the cylinder


can be pulled off the bolt of the upper suspension.

Finally unscrew the hydraulic wires from the magnet valve


block, in order to dismantle the cylinder.

Attention:
While dismantling the cylinder, oil might penetrate. The
cylinder should not be pressurized when dismantled.

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 25
Service Manual - medifa 8000

XII. Replacement of the Control Panel

Before replacing the control, the colum cover has to be removed. Please see Chapter V. „Replacement of the
Column Cover“.
Do as follows:

Pull off all plugs from the now visible control.

Attention:
Some of the plugs might be fixed by screws.

After removing all cables, remove the three fastening screws


of the control.
Now you can detach the control from the table.

Note:
For further information on the cable connection to the new
control (pin allocation), please see the electrical wiring dia-
gram.

fig. 83+84+85: dismantle control

26
XIII. Replacement of the Hydraulic Power Unit

To replace the hydraulic power unit ,first take off the column cover. Please see Chapter V. „ Removing the
Column Cover“.
The control has to be removed as well, for further help please see Chapter XII. „Replacement of the Control
Panel“.
Do as follows:

Before starting the dismantling of the power unit, disconnect


the current supply.
Disconnect the respective cables on the front of the power
unit.

Abb. 86: dismantle the hydraulic power unit

Remove both screws on the back of the power unit in order


to dismantle it.

Attention:
Before entirely removing the screws, a second person
shoud secure the power unit to prevent it from falling
down!

Abb. 87: dismantle hydraulic power unit

Before unscrewing the hydraulic tubes from the power unit


the remaining oil should be filled, through the already remo-
ved depressurisation screw, into a separate pan or into the
oil tank of the new power unit.
As soon as the oil tank is empty the tubes can be unscrewed
and connected to the new power unit.

Note:
When unscrewing the hydraulic tubes, oil might penetrate.

Finally fix the new power unit to the holder and reconnect the Abb. 88: dismantle hydraulic power unit
current supply.

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 27
Service Manual - medifa 8000

XII. Replacement of the Accumulators

Before replacing the accumulators or controlling its condition, it is necessary to remove the column cover. For
further help please see Chapter V. „Replacement of the Column Cover“.
Do as follows:

First jack up the table by placing two ,e.g., squared timber


underneath.

Now separate the cable connection between the accumulator


and the control.

Open the insulating screw joint, which is fixed inside the co-
lumn at the column base plate.

Attention:
The contacts of the accumulator are now exposed and
could cause a short circuit.
We recommend to insulate the heads (insulation tape),
as soon as they are removed from the insulating screw
joint.

fig. 89+90:dismantle accumulators

Now, you will see the battery compartment,from the lower


side of the column, (at the head rest from the patients side of
view).

To open the compartment just remove the two visible screws


and remove the protective cover.

fig. 91: dismantle accumulators

Behind the protective cover you will find the accumulators,


which are hold by an elastic band. With the help of this band
you can easily pull the accumulators out of the column.

Note:
After the control and/or replacement of the accumulators, put
back the above mentioned elastic band, to make future main-
tenance easy. .

Attention:
fig. 92: dismantle accumulators Reconnect all cables to the control!

28
XIV. Replacement of the Castors (Fixed Roller, front)

To replace the castors, jack up the table.


Place two squared timber underneath the column.
Do as follows:

To replace the front castors, simply unscrew the fastening


screw for the axle.

fig. 93: dismantle castors

When unscrewed pull the screw out of the case.

Now remove the axis of rotation. It can be simply pulled out.


Only by doing this, it will be possible to take out the castors.

Information:
Do not lose the plastic disc on the inside of the twin castors.
fig. 94: dismantle castors

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 29
Service Manual - medifa 8000

XIV. Replacement of the Castors(Fixed Roller, Back)

To replace the castors jack up the table.


Place two squared timber underneath the column.
Do as follows :

The gear shift linkage of the radioscopy operation table can be dismantled by unscrewing the six screws.

fig. 95:dismantle the driving mechanism

As soon as these screws are loosened the whole gear shift linkage can be folded.
Do as follows:

First push the fixing brake pedal directly to the hub.

fig. 96: dismantle the driving mechanism

Then push the connection pipe to the pedal.

Attention: When you push the connection pipe to the pedal, it could happen that the short
piece from the gear shift linkage fall close!

fig. 97+98 dismantle the driving mechanism

Following you have to remove the short piece of the hexagonal bar. After that you can remove the whole gear
shift linkage bit by bit.

30
After dismantling the brake rods, the castors can be dis-
mantled as well.
To do, so loosen the fastening screw at the hub (intake) from
the inside of the column
The castor can be removed by pulling down.

fig. 99: dismantling castor

fig. 100: dismantling castors

The radioscopy operation table castors are equipped with three gears, but only two are used.

While using the castors please keep the below mentioned instructions in mind.

 The red marks at the intake cylinder of the castor must always
point to right!

 The gear must be brought into the middle position:

 Castor unbraked
(upper position)

 Castor unbraked
(middle position)

 Castor unbraked
(lower position)

fig. 101: mounting castors

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 31
Service Manual - medifa 8000

As soon as all requirements for the mounting of the castors


are met, they can be put back and fastened to the castor
case on the radioscopy operation table.
To do this, fix the fastening screw from the inside of the co-
lumn to the connection.

fig. 102: mounting the castors

For a newinstallation of the driving mechanism do as follows:

Take the long end of the hexagon gear linkage and put on
the pedal for the fixing brake, but do not fasten it.

Now place this part into the hexagon connection of the hub,
on the right side of base plate.

Now take the connection pipe.

Push it as well onto the hexagon pipe and pedal inside the
base plate.

Now take the short end of the hexagon gear linkage und put
it into the connection of the left hub.

Connect both parts by pulling the connection pipe over the


space in between and fasten it with four screws.

The hexagon rods have to be placed properly inside the cas-


tor, as not to hinder the lock mechanism.

Finally, just drive the fixing pedal into the upper position and
fasten it. .

fig. 103+104+105+106+107+108+109:
mount the driving mechanism

32
XIV. Replacement of the External Battery Charger

To replace or to test the external battery charger, remove the column cover. For further help, please see
Chapter V. „ Removing the Column Cover“ .
Do as follows:

First, loosen the wires of the battery charger. Retrace the red and
the black cable to the connection and loosen it.
Keep the respective colours and plug-in locations in mind,
to prevent a mix up.

fig. 110: loosen wires

Now take the cable off the battery charger.

fig. 111: loosen cable

In order to remove the external battery charger from the base


plate, loosen the handle for the cables next to it. To do so,
remove the two fastening screws.
Now, take the battery charger off the base plate. To prevent
the cables from getting locked inside, shift the battery charger
to the side.

To install the new charger, follow the above mentioned steps fig. 112: loosen handle for cable duct
in reverse order.

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 33
Service Manual - medifa 8000

XV. Wiring Diagram Breaker Box (black)

X12
X11

X1—way measurement trendelenburg (Wegmesser Trendelenburg)


X1.1 +5V DC (gn)
X1.2 signal (br)
X1.3 GND (ws)
X1.4 PE (rt)

X2—way measurement lateral (Wegmesser Lateral)


X2.1 +5V DC (gn)
X2.2 signal (br)
X2.3 GND (ws)
X2.4 PE (rt)

X3—way measurement transversal shift (Wegmesser Querverschiebbarkeit)


X3.1 +5V DC (gn)
X3.2 signal (br)
X3.3 GND (ws)
X3.4 PE (rt)

X4—way measurement longitidunal shift (Wegmesser Längsverschiebbarkeit)


X4.1 +5V DC (gn)
X4.2 signal (br)
X4.3 GND (ws)
X4.4 PE (rt)

X5— way measurement height adjustmend (Wegmesser Höhenverstellung)


X5.1 +5V DC (gn)
X5.2 signal in (br)
X5.3 GND (ws)
110 ohms resistor between signal in and GND must be removed (110 Ohm-Widerstand zwischen GND und Signaleingang
entfernen) (rt)

X6—joystick (Joystick)
X6.1 +5V DC (ws)
X6.2 joystick y-axis (Y-Achse Joystick) (gn)
X6.3 joystick x-axis (X-Achse Joystick) (ge)
X6.4 GND (br)
X6.5 Schirmung Joystickkabel

X7—X10 to sub-d connectors (zu den Sub-D-Einbausteckern/Buchsen)


X7 control box (Steuerung)
X8 handswitch (Handschalter)
X9 column keyboard (Säulentastatur)
X10 diagnostic (Diagnose)
X12 Erdungsanschluss

34
XVI. Electrical Wiring Diagram

No... Questionaire value

1 Voltage accumulator 25—27 V


2 Current accumulator >1A

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 35
Service Manual - medifa 8000

XVII. Hydraulic Diagram

36
Tube
Pos.-Nr. Tube connections length Notes
[mm]

560 01 ring connection M10x1 —ring connection M10x1 450

560 02 ring connection M10rx1 — ring connectionM10x1 420

560 03 ring connection M10x1 — ring connectionM8x1 140

560 04 ring connetion M10x1 — ring connection M8x1 270

560 05 ring connection M10x1 — ring connection M8x1 290

560 06 ring connection M10x1 — ring connection M8x1 400

560 07 ring connectionM10x1 — ring connection M8x1 HMF

560 08 ring connectionM10x1 — ring connectionM8x1 HMF

ring connectionM10x1 — quick acting coupling


560 09 800
M10x1

560 10 ring connection M8x1 — pipe socket Ø3 90° 400

ring connection M10x1 —quick acting coupling


560 11 800
M10x1

560 12 ring connectionM8x1 — pipe socket Ø3 90° 400

ring connectionM10x1 — quick acting coupling


560 13 350
M10x1

560 14 ring connection M8x1 — pipe socket Ø3 90° 1000

ring connectionM10x1 — quick acting coupling


560 15 350
M10x1

560 16 ring connectionM8x1 — pipe socket Ø3 90° 1100

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 37
Service Manual - medifa 8000

XVIII. List of Spare Parts/ Mounting Drawings

28

26

27
25

22 24 23

21 20

19

18
11
17
12

10 16

14
13 15

7
8 6
1 2
4
3

9
5

38
Spare Parts for the medifa 800200

Pos. Order No. Name Pos. Order No. Name

1 B5600B1-04 base Operational controls

2 4.TK00.DRW0 Twin castors, d=125mm B5600M10/1 Hand switch

3 TR65 2101 Trestle roller d=80mm B5600M10/2 Column control panel

4 7111300 accumulator (1 piece) B5600M8 Joystick

5 MA-2040-2420-KR External Battery Charger Way measurement systems

Equipotential bonding conductor Way measurement system-


6 POAG-ID/6 B601120M5
connection Trendelenburg-

7 MA-2040-2420-KR plug connection, black B601120M4 Way measurement system -Lateral-

8 Way measurement system


41F1594 connection -build- in plug B601120M5
transverse shifting

9 100-201-0289 connection Way measurement system


B601120M6
Longitudinal shifting

10 MXT6000 Control panel

11 B560320M4 distribution panel More spare parts on demand!

12 610-8011-0 Hydraulic power unit

13 131-8017-0 Trendelenburg cylinder

14 132-7111-0 Lateral cylinder

15 T02366/3P-04 Column lower cover

16 T02366/2P-04 Column middle cover

17 T02366/1P-04 Column upper cover

18 5600S1 Column keyboard

19 124010-0424 Expansion bellows

20 086812 illuminated push button

Hydraulic cylinder block transverse


21 190-7007-0
shifting
Guide rail transverse shifting
22 5.17254
(1 piece)

23 T03511/P-04 Lower crusting protector cover

24 T03510/P-04 Upper crusting protector cover

Guide rail longitudinal shifting


25 5.18812
(1 piece)
Hydraulic cylinder block longitudinal
26 190-7006-0
shifting
Carbon table top (w=500mm)
27 T03089/P
for model 560320
Carbon table top (w=600mm)
28 T03089/2P
for model 560330

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 39
Service Manual - medifa 8000

XIX. Service- / Maintenance Schedule Sheet No. 1/5

Maintenance Diagramm / Work Report

Date: Work ( carried out): Instructed Personnel: Carried out by:

40
XIX. Service- / Maintenance Diagramm Sheet No. 2/5

Maintenance Diagram / Work Report

Date: Work (carried out): Instructed personnel: Carried out by::

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 41
Service Manual - medifa 8000

XIX. Service Report / Maintenance Schedule Sheet No..3/5

Test Me- Actual Data/


Test Description Targeting Data
thod Test Results

Test item: Hand switch


Visual con-
1 Case hand switch ok? No cracks or damages
trol
Visual con- No squeezes or dama-
2 Cable of hand switch ok?
trol ges
Visual con-
3 Socket hand switch ok? Proper placement
trol
Hand switch icons cleary and vi- Visual con-
4 Clearly visible
sible trol
Visual con-
5 Hand switch cover ok? No cracks or damages
trol
Test of all shifting functions by Performance
6 All functions ok.!
hand switch check!

Test item: Column control panel + Joystick


Visual con-
7 Case of column control panel ok? No cracks or damages
trol
Column control panel– icons Visual con-
8 Clearly visible
clearly and visible trol
Visual con-
9 Cover ok? No cracks or damages
trol
Performance
10 Performance check control panel All functions ok !
check
Writing and case on/of Joystick Visual con- Clearly visible, no dama-
11
ok? trol ges
Functionalty test shifting functions Performance
12 All functions ok!
Joystick check!
Test item: Table top
Visual con-
13 Guide rail check Proper placement
trol
Visual con-
14 Surface Check No visible cracks
trol

Visual con-
15 Fixation of upholstery Properly taped
trol

Visible con-
trol/ perfor-
Way measurement system mance No malfunction/no noi-
16
check—all functions check/ ses
noise level
check

Test item: Driving mechanism

Performance
17 Control brake/lock function Table stable
check
Performance
18 Control driving mechanism Table easy to move
check

42
XIX. Service Report / Maintenance Schedule Sheet No. 4/5

Test Me- Actual Data


Test Description Targeting Data
thod Test Results

Test Item: Socket connections

19 Check connection socket Visual control No damages

Visual con-
No squeezes or other
trol / Per-
20 Check connecting cables damages/ table supp-
fomance
lied with electricity
check

Visual con-
trol / Per- No damages/ table can
21 Check on/off switch
fomance be turned on/off
check

22 Check oil level Visual control Enough oil available


Check equipotential bonding
23 Visual control No damages
conductor connection
No squeezes or other
24 Check equipotential cable Visual control
damages

Test Item:

25 expansion bellows Visual control No damages

Check of all cables and connec-


No squeezes or other
26 tions behind expansion bellows, Visual control
damages
according to wiring diagram
Check of all tubes and cables be-
No squeezes or other
27 hind expansion bellows according Visual control
damages
to hydraulic wiring diagram
Check of all connections of the Cable correctly
28 Visual control
accumulator connected&placed
Energy of accumulatur
29 Check of accumulators measuring
ok
Table column is not
30 Correct connection of the column shaking
shaking
Check of the right placement of all
31 Visual control All screws fastened
screws
Cover placed correctly,
32 Check of the column cover Visual control all carriers in right posi-
tioj

Test Item:
No potential diffe-
rences between
Check for earthing of cover via Measuring/
33 contacts 1/2 and cover
equipotential bonding conductor Voltmeter
and all metal parts of
the upper frame

medifa GmbH & Co. KG – Industriestraße 5 – D-57413 Finnentrop, Tel.: +49 (0) 2721-7177-0 – Fax: +49 (0) 2721-7177-77 43
Service Manual - medifa 8000

XIX. Service Report / Maintenance Schedule Sheet No. 5/5

Test Me-
Test Description Targeting Data Actual Data/ Test Results
thod

Test item: Safety check according to IEC 601

Check of earthwire, fuses, warning Visual con-


34 No damages
notices and connections trol
Diagnostic
35 Resistance control of the earthwire < 0,2 Ω
device
Resistance control of the primary Diagnostic
36 > 2 MΩ
insulation device
Check discharged current
Diagnostic
37 (earthed) AC < 1 mA
device
(Z interrupted)
Check discharged current (patient)
Diagnostic
38 P AC < 50 A
device
(Z interrupted)
Check discharged current (normal
Diagnostic
39 use) AC < 0,5 mA
device
Check discharged current (cover)
Diagnostic
40 ( normal use) AC < 100 A
device

Check discharged current (patient Diagnostic AC < 10 A


41
(normal use) device DC < 10 A

Check discharged current (cover) Diagnostic


42 AC < 0,5 mA
(earth wire interrupted) device
Check discharged current (patient) Diagnostic AC < 50 A
43
(earth wire interrupted) device DC < 50 A
Check earthed discharged current Diagnostic
44 AC < 0,5 mA
(normal use, polarity changed) device

Check discharged current (cover) Diagnostic


45 AC < 0,1 mA
(normal use, polarity changed) device

Check discharged current (patient) Diagnostic AC < 10 A


46
(normal use, polarity changed) device DC < 10 A

Check discharged current (cover)


Diagnostic
47 (earth wire interrupted, polarity AC < 0,5 mA
device
changed)
Prüfung des Patientenableitstroms For suitable meters please see:
Diagnostic AC < 50 A
48 (Schutzleiter unterbrochen, polari-
http://www.gossenmetrawatt.com/deutsch/produkte/secutest-s2nplus.htm
device DC < 50 A
tät getauscht)
Model No.:

tester: _______________________________________________
(block letters)

Date & Signature: _______________________________________

44