Professional Documents
Culture Documents
DTE LAB MANUAL - 2017 Regulation
DTE LAB MANUAL - 2017 Regulation
DTE LAB MANUAL - 2017 Regulation
SEM : VI
YEAR: III YEAR
SYLLABUS
BM8611 DIAGNOSTICANDTHERAPEUTICEQUIPMENTLABORATOR L T PC
0042
OBJECTIVES:
LIST OF EXPERIMENTS:
TOTAL: 60 PERIODS
OUTCOMES:
At the end of the lab, the student should be able to:
Measure different bioelectrical signals using various methods
Assess different non-electrical parameters using various methodologies
Illustrate various diagnostic and therapeutic techniques
Examine the electrical safety measurements
Analyze the different bio signals using suitable tools.
Quantity
S.No Description of Equipment
Required
GSR setup
2 1
3 2
Multi-output power supply (+15v, -15v, +30V variable, +5V, 2A)
12 Audiometer 1
15 Ventilator 1
16 Ultrasound Scanner 1
LIST OF EXPERIMENTS
10 Recording of Audiogram
13 Study of ventilators
ADDITIONAL EXPERIMENTS
Aim
Theory :
Electrical potentials that occur in the cortex after stimulation of a sense organ which can be
recorded by surface electrodes are known as Evoked Potential. Example: SEP, ABR and VEP.
Evoked potentials test and record how quickly and completely the nerve signals reach the
brain. Evoked potentials are used because they can indicate problems along nerve pathways that
are too subtle to show up during a neurologic examination or to be noticed by the patient. The
disruption may not even be visible on MRI exam. These tests can be helpful in making the
diagnosis of multiple sclerosis (MS) and other neurological disorders. Evoked potential amplitudes
tend to be low, ranging from less than a microvolt to several microvolt’s, compared to tens of
micro volts for EEG, mill volts for EMG, and often close to a volt for ECG. Signals can be
recorded from cerebral cortex, brain stem, spinal cord and peripheral nerves. Usually the term
"evoked potential" is reserved for responses involving either recording from, or stimulation of,
central nervous system structures.
Types:
Visual Evoked Potentials (VEP)
Auditory Evoked Potentials (AEP)
Sensory Evoked Potentials (SEP)
The VEP tests the function of the visual pathway from the retina to the occipital cortex. It
assesses the integrity of the visual pathways from the optic nerve, optic chiasm, and optic
radiations to the occipital cortex. Visual evoked potential (VEP) tests evaluate how the visual
system responds to light. VEP tests used to evaluate optic neuritis, optic tumors, retinal disorders,
and demyelenating diseases such as multiple sclerosis.
An evoked potential or evoked response is an electrical potential recorded from the nervous
system of a human or other animal following presentation of a stimulus. The recording of electrical
activity associated with the functioning of brain activity is known as electroencephalogram. The
10- 20 electrode system is used for placing the electrodes on the scalp. The electrodes are place at
the specific points after applying gel to reduce the contact impedance.
The terms visually evoked potential (VEP), visually evoked response (VER) and visually
evoked cortical potential (VECP) are equivalent. They refer to electrical potentials, initiated by
brief visual stimuli, which are recorded from the scalp overlying visual cortex, VEP waveforms are
extracted from the electro-encephalogram (EEG) by signal averaging. VEPs are used primarily to
measure the functional integrity of the visual pathways from retina via the optic nerves to the
visual cortex of the brain
Procedure
1. The montages are connected by 10-20 electrode system on the scalp.
2. Using DAC system connects the output of the recorder to the PC.
3. The flashes of light are exposed to the subject.
4. The EEG signals are observed and are stored.
5. The digital data is exported and analyzed using EEG software.
6. The Delta (δ), Theta ( φ ), Alpha (α ), Beta ( β ), Gamma( γ ) waveforms are identified.
RESULT:
Thus the natures of visually evoked potential were identified in the EEG waveform.
CIRCUIT DIAGRAM:
Ex. No: 2
GALVANIC SKIN RESISTANCE (GSR) MEASUREMENT
Date:
Aim
To study the measurement of skin conductivity and the operation of galvanic skin response
system
Equipment and Accessories Required
Theory
Electro-dermal activity is measured in two ways. They are:
• Basal Skin Response (BSR)
• Galvanic Skin Response (GSR)
Galvanic skin resistance (GSR) refers to the recorded electrical resistance between two
electrodes when a very weak current is steadily passed between them. The electrodes are normally
placed about an inch apart, and the resistance recorded varies in accordance with the emotional
state of the subject. This resistance is normally in the range of 100,000 to 1,000,000 ohms.
Galvanic skin potential(GSP) refers to the voltage measured between two electrodes
without any externally applied current. This is conducted by connecting the electrodes to a voltage
amplifier. This potential may range as high as 50mv. Similarly, this voltage varies with the
emotional state of
LEAD PLACEMENT:
TABULATION:
OSCILLOSCOPE:
Design features are:
•Dual screen designs (therapist and client) included
•Automatic calculation of baseline (average over first 30 seconds)
•Button option to reset baseline (to average over the last 30 seconds)
•Calculation of gain over baseline
A skin conductance meter is a device that measures the electrical conductance between 2
points, and is essentially a type of ohmmeter. Active measuring involves sending a small amount
of current through the body.
The galvanic skin response is combined with the recording of heart rate, respiratory rate,
and blood pressure because they are all autonomic dependent variables, meaning they cannot be
consciously changed.
Many biofeedback therapy devices utilize skin conductance to measure and present an
individual’s stress response with the goal of helping the user to control their anxiety.
Skin conductance measurement is also becoming popular in hypnotherapy and
psychotherapy practice where it can be used as a method of detecting depth of hypnotic trance
prior to suggestion therapy commencing. It is also used in behaviour therapy to measure
physiological
which can lead to inconsistent results. Internal factors such as medications can also change GSR
measurements. Responses have been shown to show inconsistency even when given the same
stimulus level.
GSR is also known as psychogalvanic reflex (PGR), or skin conductance level (SCL).
MODEL GRAPH:
Amplitude (V)
Time (ms)
Result: Thus GSR was measured for both dry and wet skin.
Aim
To study the working of shortwave diathermy and ultrasound diathermy for various
power settings.
Theory
The term ‘diathermy’ means ‘through heating’ or producing deep heating directly in the
tissues of the body. Externally applied sources of heat like hot towels, infrared lamps and electric
heating pads often produce discomfort and skin burns long before adequate heat has penetrated to
the deeper tissues. But with the diathermy technique, the subject’s body becomes a part of the
electrical circuit and the heat is produced within the body and not transferred through the skin.
Another advantage of diathermy is that the treatment can be controlled precisely. Careful
placement of electrodes permits localization of the heat to the region that has to be treated. The
amount of heat can be closely adjusted by means of circuit parameters. The heating of the tissues is
carried out by high frequency alternating current which generally has a frequency of 27.12 MHz
and a wavelength of 11m. The current being altering, it is possible to pass current through the
tissues of a much greater intensity to produce direct heating in the tissues similar to any other
electrical conductor.
Electrodes
Joints to be
treated
Circuit Diagram
Circuit Description
The short wave diathermy machine consists of two main circuits: an oscillating circuit,
which produces a high frequency current and a patient circuit, which is connected to the oscillating
circuit and through which the electrical energy is transferred to the patient.
Transformer T1, the primary of which can be energized from the mains supply, is a step-up
transformer for providing EHT for the anode of the triode valve. A second winding can provide
heating current for the cathode of the triode valve. The tank (resonance) circuit is formed by the
coil AB in parallel with the condenser C1. The positive feedback is generated by coil CD. There is
another coil EF and a variable condenser C2 which form the patient’s resonator circuit due to its
coupling with the oscillator coil AB.
The anode supply of such a circuit is around 4000 V. The conduction in the triode takes
place during the positive half cycle and the high frequency is generated only during this period.
More usually, the supply voltage is rectified before supplying to the anode of the oscillator valve.
In such case, the oscillations produced are continuous and more power thus becomes available. In
order to ensure that the oscillator circuit and the patient’s resonator circuit are tuned with each
other, and ammeter is placed in series with the circuit. The variable condenser C2 is adjusted to
achieve a maximum reading on the meter, the needle swinging back on either side of the tuned
position. The maximum power delivered by these machines is 500 W.
A thermal delay is normally incorporated in the anode supply which prevents the passage
of current through this circuit until the filament of the valve attains adequate temperature. The
patient circuit is then switched on followed by a steady increase of current through the patient. A
mains filter is incorporated in the primary circuit to suppress interference produced by the
diathermy unit itself.
There are several ways of regulating the intensity of current supplied to the patient from a
short-wave diathermy machine. This can be done by either (i) controlling the anode voltage, or (ii)
controlling the filament current, or (iii) adjusting the grid bias by change of grid leak resistance R 1,
the best way of finely regulating the current is by adjusting the grid bias, by putting a variable
resistance as the grid leak resistance.
Procedure
The two condenser pads is connected to the output socket. The two electrodes are
placed around the patient’s body where treatment is to be given.
Connect the mains cable is connected & the unit is switched ON. Power ON switch
must glow.
Timer is set for desired period.
BM6612 Diagnostic and Therapeutic Equipment Laboratory Department of BME
20
Tabulation
Power
Max. Applied Distance Between the Deflection at the Time of Patient
S.No Control
Current Electrode Pads Treatment Time (Sec)
Position
Ultrasound Diathermy
The heating effect is produced because of the ultrasonic energy absorption property of the
tissues. The effect of ultrasonic on the tissues is thus a high speed vibration of micro massage.
Massage as a modality in physical medicine has been used in the treatment of soft tissue lesions for
centuries. Ultrasonic energy enables this massage to be carried out, firstly to a greater depth than is
possible manually and secondly at times (in acute injuries) when pressure cannot be exerted by
hand because of intolerable pain caused to the patient. The thermal effects of ultrasound are
dependent on the amount of energy absorbed, the length of time of the ultrasound application and
the frequency of the ultrasound generator.
Ultrasonic generators are constructed on the piezo-electric effect. A high frequency
alternating current (e.g., 0.75-3.0 MHz) is applied to a crystal whose acoustic vibration causes the
mechanical vibration of the transducer head, which itself is located directly in front of the crystal.
These mechanical vibrations then pass through a metal cap and into the body tissue through a
coupling medium. The therapeutic ultrasonic intensity varies from 0.5 to 3.0 W / cm 2. Applicators
range from 70 to 130 mm in diameter.
Circuit Description
The equipment required for ultrasonic therapy is electronically very simple. The heart of
the system is a timed oscillator which produces the electrical oscillations of the required frequency.
The oscillator output is given to a power amplifier which drives the piezo electric crystal to
generate ultrasound waves. Power amplification is achieved by replacing the transistor in typical
LC tuned Colpitt oscillator by four power transistors placed in a bridge configuration.
The delivery of ultrasound power to the patient is to be done for a given time. This is
controlled by incorporating a timer to switch on the circuit. The timer can be a mechanical spring –
loaded type or an electronic one, allowing time settings from 0 to 30 minutes.
Tabulation
The dosage can be controlled by varying any one of the following variables.
Frequency of ultrasound
Intensity of ultrasound
Duration of the exposure
The output power of an ultrasonic therapy unit can be varied continuously between 0 and 3 W/cm2
Procedure
The ultrasonic transducer is connected to the power unit. The ultrasonic transducer
is placed on the patient’s body where treatment is to be given.
Switch ON the unitby connecting to the mains cable &. Power ON switch must glow.
Ultrasound selector switch is set for desired mode of operation (continuous or pulsed).
Power control is set to the desired position, generally 1 w/cm2 or 2w/cm2 is sufficient.
The treatment time is set.
Result
The working of shortwave and ultrasound diathermy are thus studied at various power settings.
Battery
Electrode
Battery
Aim
Theory
Wireless telemetry permits the examination of the physiological data of man or animal
under normal conditions and in the natural surroundings without any discomfort or obstruction
to the person or animal under investigation. The stage of a typical biotelemetry system is divided
into two functional blocks, the transmitter and receiver. In the transmitter unit the physiological
signals are obtained from the subjects by means of appropriate transducers. The signal is then
passed through a stage of amplification and processing circuits that include generation of a
subcarrier and a modulation stage for transmission. A radiofrequency (RF) carrier is a high-
frequency sinusoidal signal which when applied an appropriate transmitting antenna, is propagated
in the form of electromagnetic waves. The range of the biotelemetry system is the distance
travelled by the transmitted signal. Information to be transmitted is impressed upon the carrier by
a process known as modulation. The circuitry which generates the carrier and modulates it
constitute the transmitter. The receiver consists of a tuner to select the transmitting frequency, a
demodulator to separate the signal from the carrier wave and a mean of displaying or recording the
signal. By tuning the receiver to the frequency of the desired RF carrier, that the signal can be
selected while others are rejected.The two basic system of modulation are Amplitude Modulation
(AM) and Frequency Modulation (FM). In amplitude – modulated system, the amplitude of the
carrier is caused to vary with the information being transmitted. Amplitude modulated system are
susceptible to nature and manmade interference, since the interference generally appear as
variations in the amplitude of the
received signal. In Frequency Modulation system, the frequency
BM6612 Diagnostic and Therapeutic Equipment Laboratory Department of BME
TABULATION:
Procedure:
1. The modules are connected as per the block diagram.
2. The battery power supply is switch on.
3. The Electrodes from the subject is connected to the ECG amplifier.
4. The ECG from the output of ECG Amplifier is observed on the DSO.
5. The ECG signal on the receiver side from the output of FM Demodulator is
observed on the DSO.
RESULT:
Thus the ECG signal was transmitted and received through single channel Biotelemetry.
Aim
To study the working of a hemodialysis machine.
Theory
Hemodialysis is a treatment to filter wastes and water from the blood, as kidneys did when
they were healthy. Hemodialysis helps control blood pressure and balance important minerals,
such as potassium, sodium, and calcium, in the blood.
Hemodialysis can help the patient feel better and live longer, but it’s not a cure for kidney
failure.
During hemodialysis, blood that is drawn out of the body goes through a filter, called a dialyzer,
outside the body. A dialyzer is sometimes called an “artificial kidney.”
Hemodialysis works on two principles:
osmosis
diffusion
Osmosis is the ability of water to pass through the wall of a cell. This wall is called a semi-
permeable membrane and acts like a filter.
Diffusion is the natural movement of particles from an area with lots of particles to an area
with fewer particles.
In hemodialysis, the dialyzer (mechanical kidney) acts as the semi-permeable membrane.
Blood flows along one side of the membrane and a special dialysis solution called dialysate
flows on the other side.
The waste products removed from your child's body diffuse, or spread, from the blood (an
area with lots of waste products) into the dialysate (an area with no waste products).
The substances that the body needs cannot pass through the semi-permeable membrane. These
stay in the blood that is returned to the body.
BM6612 Diagnostic and Therapeutic Equipment Laboratory Department of BME
Hemodialysis process
Process in Dialyzer
Fistula (also known as arteriovenous fistula or A-V fistula): An artery and vein are joined
together under the skin in the arm. An A-V fistula needs 6 weeks or longer to heal before it can
be used for hemodialysis. Then, it can be used for many years.
Graft (arteriovenous graft or A-V graft): A plastic tube is used to join an artery and vein under
the skin.
Catheter (central venous catheter): A flexible tube (catheter) is put into a vein in the neck,
below your collarbone, or next to your groin. It’s only meant to be used for a short time.
Procedure
Measurement of Maximum & Minimum speed of peristaltic pump with and without load
Measurement of Time taken for transfer of 1 liter water for different pump speeds
S. No Speed Time
1. 60
2. 90
3. 100
Measurement of Time taken for transfer of 1 liter water for different pump speeds
Result
Thus the working of hemodialysis machine is studied.
BM6612 Diagnostic and Therapeutic Equipment Laboratory Department of BME
Ex. No: 6
ELECTRICAL SAFETY MEASUREMENTS
Date:
Aim
To measure the leakage currents, insulation break using electrical safety analyzer.
Theory
Test method
Normally the high voltage is applied between the two parts across the insulation being
tested, such as the primary circuit and metal enclosure of the equipment under test.
Tabulation
Input voltage:
Output voltage:
Reverse Polarity:
If the insulation between the two is adequate, then the application of a large voltage
difference between the two conductors separated by the insulator would result in the flow
of a very small current. Although this small current is acceptable, no breakdown of either the air
insulation or the solid insulation should take place. Therefore, the current of interest is the current
that is the result of a partial discharge or breakdown, rather than the current due to capacitive
coupling.
Another example would be to test the insulation between the primary and secondary
circuits of a power supply. Here, all the outputs are shorted together. The ground probe from the
tester is placed in contact with this cable bunch, and the high voltage probe is placed in contact
with L and N connectors, which are shorted together. The equipment under test does not run during
the test. It must also be noted that when applying the high voltage during the type test, the ideal
situation would require that not more than half of the prescribed voltage be applied, and then
raised gradually over a period of 10 seconds to the full value and maintained for 1 minute. Most
test equipment, however, either turns on directly to the full voltage or has an electronically
controlled ramp.
Test voltage:
The majority of safety standards allow the use of either ac or dc voltage for the test. When
using ac test voltage, the insulation in question is being stressed most when the voltage is at its
peak, i.e., either at the positive or negative peak of the sine wave. Therefore, if one decides to use
dc test voltage, one must ensure that the dc test voltage is √2 (or 1.414) times the ac test voltage, so
the value of the dc voltage is equal to the ac voltage peaks. For example, for a 15-V-ac voltage, the
equivalent dc voltage to produce the same amount of stress on the insulation would be 15 x 1.414
or
21.21 V dc.
Insulation
Resistance:
The insulation resistance test is also known as a Megger test. Its objective is to measure the
total resistance between any two points separated by insulation. The test, therefore, determines
how effective the insulation is in resisting the flow of electrical current. The voltage is typically
around 500 – 1000 V dc; hence, the current is very low. Because the current is so low, this test is
useful for checking the quality of the insulation not only when a product is first manufactured, but
also over time as the product is used.
The equipment under test is connected to the measuring instrument, and the voltage is
ramped up from zero to the final value, which in most cases is 500 V dc. Once the voltage reaches
the selected value, it is kept at that value for a brief period (typically up to 5 seconds) before the
resistance test is measured. The measured value should be very high (typically in the meg ohm
region).
This refers to the electric current through a human body or through an animal body when it
touches one or more accessible parts of installation or equipment. There is also another concept
known as “protective conductor current” and this refers to the current that flows in a protective
conductor. A protective conductor current, therefore, can never be the source of an electric shock
because, by definition, the protective conductor is connected to earth.
If touch current is excessive, and operator could receive an electric shock, which could
result in a serious injury, depending on a person’s body weight. Typically, current of more than 1.0
mA can cause and electric shock to an operator. The shock may or may not be serious, depending
on the amount of the current and the body weight.
Result
The leakage current, insulation for the given items are measured using Electrical Safety
Analyzer.
Aim
To measure the respiratory rate using displacement transducer
2 PC with Analyzer 1
3 Data Cable 1
Theory
Pulmonary function analyzers provide the means for automated clinical procedures and
analysis techniques for carrying out a complete evaluation of the lung function or the respiratory
process. The respiratory activity ensures supply of oxygen to and removal of carbon dioxide from
the tissues. This is performed by using a device called spirometer that measures the volume
displacement and amount of gas moved in a specific time. Usually this requires the patient to take
a deep breath and then exhale as rapidly and completely as possible called the forced vital
capacity, this gives an indication of how much air can be moved by the lungs and how freely this
air flows.
Spirometer
A spirometer is an apparatus for measuring the volume of air inspired and expired by the
lungs. It is a precision differential pressure transducer for the measurements of respiration flow
rates. The spirometer records the amount of air and the rate of air that is breathed in and out over a
specified time. The spirometer and attached flow head function together as a Pneumotachometer,
with an output signal proportional to airflow. The output produced by a spirometer is called a
kymograph trace. From this, vital capacity, tidal volume, breathing rate and ventilation rate (tidal
volume × breathing rate) can be measured. From the overall decline on the graph, the oxygen
uptake can also be measured.
Spirometery
Pneumotachometer
Pneumotachometer is a turbine transducer that converts the flow of air, breathed by the
patient, against a frictionless rotating vane into an electrical signal which is used to produce the
relevant plots. The arrow on the surface of the transducer indicates the direction in which it is to be
pushed inside the transducer slot in the head piece.
Respiratory
Volumes Tidal
Volume
The volume of gas inspired or expired (exchange with each breath) during normal quiet
breathing is known as tidal volume.
The volume of gas, which can be inspired from a normal end tidal volume.
The volume of gas remaining after a normal expiration less the volume remaining after a
forced expiration.
ERV = FRC – RV
Tabulation
Lung Functions
S.No. Subject Name
Volume Capacity
Respiratory Capacities
TLC = VC + RV
Vital Capacity (VC)
The greatest volume of gas that can be inspired by voluntary effort after maximum
expiration, irrespective of time.
Inspiratory Capacity (IC)
The maximum volume that can be inspired from the resting end expiratory position.
Dead space is the functional volume of the lung that does not participate in gas exchange.
Procedure
Fix the transducer into the transducer assembly. The transducer has an arrow on the surface.
Insert the transducer in the direction indicated by the arrow. The transducer will click into
position when correctly connected.
Fix the mouth piece (disposable or reusable) to the transducer assembly. The mouthpiece
will click into position when correctly connected. Connect the transducer assembly to the
computer using the USB cable.
Train thepatientin testperformance. The patient’s collaboration is essential to carry out the
maneuvers correctly. Use a nose clip to allow air to flow only through the patient’s mouth.
The patient must sit upright holding the hand piece to their mouth and throughout the
maneuver the patient should try to keep their back straight as much as possible.
The patient should hold transducer hand unit in such a way that the air passage is completely
unobstructed. Ensure the patient’s hands, fingers or clothing, etc., are not obstructing the air
flow. The area in front of the patient should also be kept clear to avoid back draft of air
entering the turbine and affecting the readouts.
Result
The respiratory rate is measured using displacement transducer and the spirogram is
recorded.
Aim
To study the medical stimulator for different types of waveforms and intensity
Equipment and Accessories Required
TNS
IG
surges per minute in most instruments. The ratio of the interval to the duration of the
surging is also adjustable so that graded exercise can be administered. This type of current
is required for the treatment of spasm and pain.
Exponentially Progressive Current: This type of current is useful for the treatment of severe
paralysis. The main advantage of this method lies in the possibility of providing selective
stimulation for the treatment of the paralyzed muscles. This means the surrounding health tissues
even in the immediate neighborhood of the diseased muscles are not stimulated.
Biphasic Stimulation: The all recovery from the effect of a stimulus can be hastened by the
passage of a lower intensity current of opposing polarity over a longer period so that the net
quantity of electricity is zero. Biphasic current is the most versatile of the stimulation therapy
waveforms. It can be used for several different purposes. The use of biphasic current for the
treating pain, strengthening muscles, re-educating muscles, increasing circulation and decreasing
swelling.
Procedure
1. Connect the output from Surged current to DSO
2. Switch on the equipment and draw the output waveform for Faradic and Surged
Faradic for different intensity and duration
3. Change the connection to TENS mode
4. Draw the output waveform for TNS and IG for different intensity and duration
Result
The waveform for the therapeutic application are studied using medical stimulator.
Aim
To study and analyze the functioning and safety aspects of surgical diathermy.
Equipment and Accessories Required
Theory
Surgical diathermy machine uses high frequency currents for surgical purposes involving
cutting and coagulation. They depend on the heating effect of electric current. When a high
frequency current flows through the sharp edge of wire loop or band loop or the point of a needle
into the tissue as shown in the figure, there is a high concentration of current at this point. The
tissue is heated to such an extent that the cells which are immediately under the electrode are torn
apart by the boiling of the cell fluid. The indifferent electrode establishes a large area contact with
the patient and the RF current is therefore dispersed so that very little heat is developed at this
electrode. This type of tissue separation forms the basis of electro-surgical cutting. Electro-surgical
coagulation of tissue is caused by the high frequency current flowing through the tissue and
heating it locally so that it coagulates from inside.
Operation
Surgical diathermy machine consists of a high frequency power oscillator. Two electrodes
are connected to the RF power generator. One electrode is active and has a very small cross
sectional area (100 cm2) with respect to the other electrode. The current flowing into the patient
cable is same as the current flowing into the active electrode. The current density is in amperes.
The current density is difference between the two electrodes. The tissue beneath the patient plate
heats up slightly, while the tissue underneath the active electrode is heated to destruction. In
bipolar mode electrode does
Tabulatio
n
Cut
1 Unipolar
Coagulation
2 Bipolar Coagulation
The generation of heat per unit volume of tissue, P/V (W/m3) is given by
P/V = ρi2,
Where ρ is the resistivity (Ωm) including the dielectric losses, and i is the current density
(A/m2).
As a means of incising tissue, electro surgery has the advantage that bleeding is diminished due to
the coagulation of the vessels. In addition some microorganisms are killed by heat.
Procedure
Result
The functioning of surgical diathermy is analyzed with various safety aspects.
Ex. No: 10
RECORDING OF AUDIOGRAM
Date:
Aim
To measure the pure tone air conduction and to study bone conduction thresholds of
hearing using audiometer
Equipment and Accessories Required
Theory
Acuity of hearing can be measured with the help of audiometer. The human ear responds to
vibrations ranging from 20-20000 Hz. Hearing is affected by anything which interferes with the
conduction of sound waves to the cochlea, such as perforated tympanic membrane ( ear drum),
disease of middle ear or disease of cochlea itself or its connection in the central nervous system.
Tabulation
Hearing Level in dB
Frequency in Hz
Left Ear Right Ear
Procedure
The head phones should be plugged into the jack socket on the side panel.
Switch on the audiometer.
The mode which can be turned to either ‘RIGHT’ or ‘LEFT’ depending upon the ear to
be tested.
The frequency dial is then adjusted to the test frequency from 1000Hz to 8000Hz step
by step.
The frequency and intensity is set so that patient hears the sound.
It is then diminished by adjusting the dB’s and note the dB
Repeat the test to verify the reliability of the patient’s response.
Then do the test below 1000Hz.
While testing each frequency, the tone should first be presented at a higher level to make
thepatient aware of the tone he is going to hear and then gradually the intensity is reduced
to histhreshold.
Pure tone Bone conduction Audiometry is conducted only for patients and the procedure is
as follows
The jack of the bone conduction vibrator is plugged into the socket on the side panel of
the audiometer.
The bone switch is positioned to BC.
Bone conduction up to 4000Hz only can be tested.
The testing procedure is the same as for air conduction.
Resul
t
The hearing of acuity is checked with audiometer and an audiogram is drawn for measured
readings.
Ex. No: 11
STUDY THE WORKING OF DEFIBRILLATOR AND
Date: PACEMAKERS
Aim
Theory
Ventricular fibrillation can be converted into more efficient rhythm (fig.1) by applying a
high energy shock to the heart. This sudden surge across the heart cause all muscle fibers to
contract simultaneously ,possibly the fibers nay then respond to normal physiological intrinsic
pulse. Restoration of normal rhythm in fibrillating heart by direct current shock. The horizontal
line after the shock shows that cardiograph was blocked or disconnected for its protection during
the period of shock.
Defibrillator
The instrument for administering the shock is called a defibrillator. So
it is used to reverse fibrillation of the heart. Electric shock by defibrillator is used to reestablish
normal activity.
Ventricular fibrillation and with electric shock
Types of waveform
Basic circuit diagram of DC Defibrillator and the typical discharge pulse of defibrillator
The shock can be delivered to the heart by means of electrode placed on chest of the
patient(External defibrillation) or the electrode may be held directly against the heart when the
chest is open(internal defibrillation). Higher voltage is required for external defibrillation than for
internal defibrillation
A Variable auto transformer forms the primary of a high voltage transformer. The output
voltage transformer is rectified by a diode rectifier and is connected to vacuum type high voltage
change over switch. In position 1, the switch is connected to one end of an oil filled micro farad
capacitor. In this position, the capacitor charge to a voltage set by the positioning of the auto
transformer. When the shock is delivered to the patient, a foot switch or a push button mounted on
the handle of the electrode is operated. The high voltage switch change over to position 2 and the
capacitor is discharged across the heart through the electrode.
The inductor in the circuit slows down the discharge from capacitor by induced counter
voltage. This gives the output pulse a physiologically favorable shape. The disadvantage of using
inductor is that any practical inductor will have its own resistance and dissipates part of the energy
during the discharge process. The shape of waveform that appears across electrodes will depend
upon the value of the capacitor and inductor used in the circuit. The discharge resistance which the
patient represent for defibrillating pulse may be regarded as purely ohmic resistance of 50-100 Ω
approximately for typical electrode size of 80 cm2. The typical discharge pulse of defibrillator is
shown in fig.
Monophasic pulse width is typically programmable from 3.0 to 12.0 msec. Biphasic
positive pulse width is typically programmable from 3.0 to 10.0 msec, while the negative pulse is
from 1.0 to 10.0 msec Studies suggest that biphasic pulses yield increased defibrillation efficacy
with respect to Monophasic pulses
Procedure
Pacemaker
A pacemaker is an electronic device that provides an electrical signal to make the heart
beat when its own, built-in pacemakers fail. The anatomical, built-in pacemakers provide what’s
called the “intrinsic” rhythm, and they can be disrupted by various conditions – ischemia for
example, or by an MI.
Cardiac arrhythmia
Abnormality in cardiac normal rhythm is called cardiac arrhythmia.There are various
types of cardiac disorder. Some of them are Bradycardia and tachycardia. In bradycardia heart
rate goes down from normal rate(i.e. below 60) and in tachycardia heart rate is higher than
normal rate.
There are several reasons for cardiac arrhythmia. One of the reason for arrhythmia is blockage in
conduction path. Heart block can occur anywhere in the specialized conduction system beginning with
the sino-atrial connections, the AV junction, the bundle branches and their fascicles, and ending in the
distal ventricular Purkinje fibers. Disorders of conduction may manifest as slowed conduction
Pacemaker classification
Pacemaker can classified in two category
1. External pacemaker
2. Implantable pacemaker
ECG Amp
pacemaker
TABULATION
FREQUENCY WAVEFORM
1. NORMAL
2. TACHYCARDIA
3. BRADYCARDIA
4. AV BLOCK
Synchronous pacemaker
In patients who have normal heart function most of the time, asynchronous pacing can be
extremely dangerous, working against their own physiological pacemaker with the danger of
stimulating in the vulnerable period of the T wave, a condition that can result in fibrillation. The
demand pacemaker consists of an ECG amplifier and a conventional pacemaker output pulse
circuit that has been modified to allow output from the ECG amplifier to inhibit the pulse
generator. This pacemaker senses R-waves and its timing and logic circuits count out an elapsed
time interval following an R- wave or previously induced pulse. If the intrinsic R-wave does not
appear before the elapsed time interval, the ventricle is stimulated. If an R-wave is received, the
counter is reset again. This type of pacemaker is used for patients with bradycardia, and it ensures
a heartbeat no slower than its set rate.
Procedure
Result
Ex. No: 12
ANALYSIS OF ECG, EEG AND EMG SIGNALS
Date:
Aim
Theory
Electrocardiograph (ECG)
Electrocardiograph (ECG) is an instrument used to record electrical activity of the heart over a
period of time, as detected by electrodes attached to the surface of the skin and recorded by a
device external to the body. The recording produced by this noninvasive procedure is
termed an electrocardiogram. An ECG is used to measure the rate and regularity of heartbeats, as
well as the size and position of the chambers, the presence of any damage to the heart, and the
effects of drugs or devices used to regulate the heart, such as a pacemaker. The conducting system
of the electrical pulses in heart is shown above. ECG provides a wide range of cardiac disorders
such as the presence of an inactive part (infarct) or an enlargement (hypertrophy) of heart muscle.
The ECG device detects and amplifies the tiny electrical changes on the skin that are caused when
the heart muscle depolarizes during each heartbeat.
The diagnostically useful frequency range is usually 0.05 to 150Hz. The interference of non-
biological noises can be handled by using modern differential amplifiers, which are capable of
providing excellent rejection capabilities. Common Mode Rejection Ratio of the order of 100-
120db with 5KΩ unbalance in the leads is a desirable feature of ECG machines. It is necessary to
use a notch filter tuned to 50Hz to reject hum due to power mains.
SHIFTING OF BASELINE:
A wandering baseline but otherwise normal ECG trace is usually due to the movement of the
patient or electrode, the baseline shift is eliminated by ensuring that the patient lies relaxed and the
electrodes are properly attached. Baseline wander is usually observed immediately after the
application of electrodes.
MUSCLE TREMOR:
Irregular trembling of ECG trace, without wandering of the baseline occurs when the patient is not
relaxed or is cold. It is generally formed in the case of older patients. Muscle tremor signals are
especially bothersome on limb leads. When a patient moves or the muscles are stretched. For
normal ECG recording, the patient must be advised to get warm and to relax so that muscle tremor
from shivering or tension is dominated. The most critical component of the ECG recorder is the
patient cable. They should be made of silicon-rubber, which provides better elasticity over long
period.
Normal Rhythms
Normal sinus rhythm - heart rhythm controlled by sinus node at a rate of 60-100 beats/min, each
P wave followed by QRS and each QRS preceded by a P wave.
For example, if there are 4 large squares between regular QRS complexes, the heart rate is
75bpm (300/4=75).
METHOD 2: The second method can be used with an irregular rhythm to estimate the rate.
Count the number of R waves in a 6 second strip and multiply by 10.
For example, if there are 7 R waves in a 6 second strip, the heart rate is 70bpm (7x10=70).
METHOD 3: When the rhythm is regular, the heart rate is 1500 divided by the number of
small squares between the QRS complexes.
For example, if there are 22 small squares between regular QRS complexes, the heart rate is
68bpm (1500/22=68).
1 small square – 0.04sec. For 60 sec the numbers of small boxes are 1500.
(1500*0.04=60sec.)
METHOD 4:
60 divided by [(number of small squares between the QRS complexes) *(0.04 sec)]
Abnormal Rhythms
Sinus tachycardia - high sinus rate of 100-180 beats/min as occurs during exercise or other
conditions that lead to increased SA nodal firing rate.
Atrial flutter - sinus rate of 250-350 beats/min.
Atrial fibrillation - uncoordinated atrial depolarization
Ventricular fibrillation - uncoordinated ventricular depolarization, leads to death if not quickly
converted to a normal rhythm or at least a rhythm compatible with life.
AV nodal blocks - a conduction block within the AV node (or occasionally in the bundle of His)
that impairs impulse conduction from the atria to the ventricles.
First-degree AV nodal block - the conduction velocity is slowed so that the P-R interval is
Electromyogram (EMG)
Generation of EMG
The EMG is generated when a motor neuron action potential from the spinal cord arrives at a
motor end plate. Its arrival causes a release of ACh (Acetylcholine) at the synaptic cleft which
causes a depolarization (Action Potential). This action potential electrically travels downward from
the surface in a transverse tubule. This in turn causes a release of Ca++, causing cross-bridge
binding and the sarcomere of the muscle to contract .
An Electromyography (EMG) is a measurement of the electrical activity in muscles as a by
product of contraction. An EMG is the summation of action potentials from the muscle fibers
under the electrodes placed on the skin. The more muscles that fire, the greater the amount of
action potentials recorded and the greater the EMG reading. The test is used to help detect
neuromuscular abnormalities. During the test, one or more small needles electrodes are inserted
through the skin into the muscle.
The contraction of the skeletal muscle results in the generation of action potentials in the
individual muscle fibres, a record of which is known as electromyogram. In this case skeletal
muscle repolarization takes place more rapidly, the action potential lasting only a few
milliseconds. Since most of EMG measurements are made to obtain an indication of the amount of
activity of a given muscle, or a group of muscles rather than an individual muscle fibre, the EMG
pattern is usually a summation of the individual action potentials from the fibres constituting the
muscles or muscles being studied. In voluntary contraction of the skeletal muscle, the muscle
potentials range from 50 µV to 5 mV and the duration from 2 to 15 ms. The values vary with the
anatomical position of the muscle and the size and location of electrode.
Result
Thus normal and abnormal ECG, EEG and EMG signals has been analyzed.
Aim
Theory
Procedure
1. Connect the flexible tubing to the air outlet on the side of the device.
2. Connect the tubing to the mask.
3. Plug the socket of the power cord into the power supply. Switch on the machine.
4. Press the wheel to highlight on/off icon and switch on the device.
5. Verify that the device beeps and the alarm and ramp LEDs light up each time therapy is started.
6. Make sure there is no leak from the mask . Adjust the mask and connect it to the test lung
or patient.
7. Observe and record the pressure readings in various therapy modes.
8. Press and hold the wheel for approximately 2 seconds to turn off therapy.
Result
Thus the working of ventilator was studied.
Ex. No: 14
STUDY OF ULTRASOUND SCANNERS
Date:
Aim
Theory
Ultrasound has become increasingly important in medicine and has taken its place along with X-
ray and nuclear medicine as a diagnostic tool. Its main attraction as an imaging modality lies in its non-
invasive character and ability to distinguish interfaces between soft tissues. In contrast, X-rays only
respond to atomic weight differences and often require the injection of a more dense contrast medium
for visualization of non-bony tissues. Similarly, nuclear medicine techniques measure the selective
uptake of radioactive isotopes in specific organs to produce information concerning organ function.
Radioactive isotopes and X-rays are, thus, clearly invasive. Ultrasound is not only non-invasive,
externally applied and non-traumatic but also apparently safe at the acoustical intensities and duty
cycles presently used in diagnostic equipment. Diagnostic ultrasound is applied for obtaining images
of almost the entire range of internal organs in the abdomen. These include the kidney, liver, spleen,
pancreas, bladder, major blood vessels and of course, the foetus during pregnancy. It has also been
usefully employed to present pictures of the thyroid gland, the eyes, the breasts and a variety of other
superficial structures. In a number of medically meaningful cases, ultrasonic diagnostics has made
possible the detection of cysts, tumours or cancer in these organs. This is possible in structures where
other diagnostic methods by themselves were found to be either inapplicable, insufficient or
unacceptably hazardous. Ultrasonic studies which do not involve image formation have also been
extensively developed to allow the dynamics of blood flow in the cardiovascular system to be
investigated with a precision not previously possible. The main limitation of ultrasound, however, is
Result
Thus the working of ultrasound scanner is studied.
Ex. No: 15
STUDY OF HEART LUNG MACHINE MODEL
Date:
Aim
Theory
The heart-lung machine (HLM) is a device used to provide blood circulation and oxygenation
while the heart is stopped. It is a means of keeping a patient alive while his/her heart is stopped or even
removed from the body. Usually called the heart-lung machine, the device also is referred to as
cardiopulmonary bypass, indicating its function as a means to substitute for the normal functions of the
heart (cardio) and lungs (pulmonary).
It is the function of the heart to provide circulation of blood at all times. It pushes blood out
into the body and through the lungs. It must function every minute of every day of life to maintain the
health of the tissues throughout the body.
To function, the heart-lung machine must be connected to the patient in a way that allows blood
to be removed, processed, and returned to the body. Therefore, it requires two hook-ups. One is to a
large artery where fresh blood can be pumped back into the body. The other is to a major vein where
used blood can be removed from the body and passed through the machine.
In fact, connections are made on the right side of the heart to the inferior and superior vena
cavae .These vessels collect blood drained from the body and head and empty into the right atrium.
They carry blood that has been circulated through the body and is in need of oxygenation. Another
connection is made by shunting into the aorta, the main artery leading from the heart to the body, or
the femoral artery, a large artery in the upper leg. Blood is removed from the vena cavae, passed into
the heart-lung machine where it is cooled to lower the patient’s body temperature, which reduces the
tissues’ need for blood. The blood receives oxygen which forces out the carbon dioxide and it is
Procedure
Result
Thus the working of heart lung machine is studied with a working model.