Professional Documents
Culture Documents
Radiotherapy Planning (Isodose Distributions) : "Isodose Chart" For A Given Beam
Radiotherapy Planning (Isodose Distributions) : "Isodose Chart" For A Given Beam
Radiotherapy Planning
(Isodose Distributions)
Aim:-
Theory:
7
Figure-2
For ortho-voltage (about 400 kVp) and low energy x-rays, the reference depth
(at which maximum build up occurs) is usually the surface (do = 0). For higher
energies, the maximum build up is below the surface of the patient. This is called skin
sparing effect, where the patient's skin is saved from burning.
A number of parameters affect the central axis depth dose distribution as follows:-
2. Depth
The percentage depth dose decreases with depth beyond the depth of
maximum dose. However, there is an initial build-up of dose which becomes
more and more pronounced as the energy is increased. The region between the
surface and the point of maximum dose is called the dose build-up region. The
dose build-up effect of the higher energy beams gives rise to what is clinically
known as the skin-sparing effect.
3. Field size
Percentage depth dose increases with increasing the field size.
8
that we are dealing with a primary beam without scatter. In a given clinical
situation, collimation or other scattering materials in the beam may cause
deviation from the inverse square law. So, the percent depth dose increases with
SSD because of the effects of the inverse square law.
Figure-3
9
Figure-4
2. Multiple fields
One of the most important objectives of treatment planning is to deliver the
maximum dose to the tumour and minimum dose to the surrounding tissues. In
addition, dose uniformity within the tumour volume and sparing critical
organs are important considerations in judging a plan. To achieve these goals,
we should do the following:
10
The basic dose distribution data are obtained under standard conditions, which
include homogeneous unit density phantom, perpendicular beam incidence, and flat
surface. During treatment, however, the beam may be obliquely incident with respect
to the surface and in addition, the surface may be curved or irregular in shape. Under
such conditions, the standard dose distributions cannot be applied without proper
modifications or corrections. Contour corrections may be avoided by using a bolus or
a compensator but under some circumstances it is permissible to determine the actual
dose distribution by calculation. We can correct for the contour irregularities by three
methods:
The first two methods are useful for making individual point dose calculations. In this
experiment, we will focus only on the third method.
For manual treatment planning, it is convenient to correct the entire isodose chart
for contour irregularities. This can be done by an empirical method, known as the
isodose shift method. Suppose that S - S in figure 5,is the patient contour drawn
on a transparent paper, and S`- S` is a flat surface line passing through the point of
intersection of the central axis with the contour. This method can be summarized
as follows;
1. From the line S`-S` draw vertical grid lines, parallel to the central axis and
spaced about 1 cm apart, to cover the full field width.
11
2. Place the standard isodose chart underneath this paper and align the central
line of the chart with that of the grid.
Figure-5
4. For each grid line, slide the isodose chart up or down, depending on whet-
her there is tissue excess or deficit along that line, by an amount k x h,
where “k” is a factor less than 1.
N.B. Factor “k” depends on the radiation quality (as shown in the table-1), field
size, depth of interest, and SSD.
5. Mark the isodose values at points of intersection of the given grid line and the
shifted isodose curves.
6. After all the isodose positions along all the grid lines have been marked,
newisodose curves are drawn by joining the marked points having the
sameisodose values.
12
N.B: In case of tissue excess, the shift is upward k x h, where h is the distance between
S`-S` and the contour S-S. In case of air-gap, the shift
is downward – (k x h`), where h` is the distance between S`-S` and S-S.
Table-1
Important Definitions
Target volume
This is the volume which needs to be irradiated to a specified absorbed dose.
Hot spot
It is an area outside the target volume, which receives a higher dose than the
specified target dose.
Procedure:-
3. Correct the isodose chart for contour irregularities (considering air-gaps and
excess tissue) by using the “isodose shift method”.
References:
F. M. Khan 1994 (2nd edition). The Physics of Radiation Therapy. pp. 226 – 314
13