Professional Documents
Culture Documents
The Basics of Lc-Msms Troubleshooting: Tools, Cases, Strategy
The Basics of Lc-Msms Troubleshooting: Tools, Cases, Strategy
LC-MSMS
TROUBLESHOOTING:
TOOLS, CASES,
STRATEGY
JUDY STONE, MT(ASCP), PH.D., DABCC
UNIVERSITY OF CALIFORNIA, SAN DIEGO
HEALTH SYSTEM, MASS
SPECTROMETRY/TOXICOLOGY LABORATORY
THANKS TO:
• Larry McAndrew, AB Sciex
• Judy Chang, Camilla Otten,
Kathy Chen at TPMG Kaiser
Regional Laboratories
• Rob Fitzgerald, Heather
Hochrein, Josh Akin, Krista
Pratico at UC San Diego
OUTLINE
• Troubleshooting tools
• Cases
• Strategy and ground-
work, the infrastructure
necessary for better
troubleshooting (Appendix)
TROUBLESHOOTING
(DIAGNOSTIC) TOOLS
1. Chromatographic images
2. Pressure traces
3. System Suitability Test (SST)
results and Maintenance
Calendar annotation
4. Acquisition data file records
5. Divide and Conquer
1. CHROMATOGRAPHIC
IMAGES
• Peak Review vs XIC Overlay
Qualifier
MRM
I.S. Quantifier
MRM
I.S. Qualifier
MRM
Courtesy of Grace vanderGugten (St. Paul’s Hospital, Vancouver, BC
1.TIC OR XIC OVERLAY
• Unprocessed data (no
integration, no assignment
of MRM to analyte or I.S.)
• Multi-analyte, qualitative
view
Electronic record –
what parameters were
actually used to
acquire a file?
5. DIVIDE & CONQUER
• SST: Sample Prep vs
Instrument
• Unextracted standard post-
column infusion: LC vs
MSMS
• Maintenance calendar: Silly
human error vs instrument
5. LC DIVIDE & CONQUER
• Disconnect components
sequentially & check
flow/pressure to find leaks
OR source of overpressure
• $64,000 question – what is
normal pressure for that
segment?
CASES
• Peak shape
• Peak area/Peak Height
• Peak Retention Time (Rt)
PEAK SHAPE CASE 1
Quantifier
MRM
Qualifier
Normal
MRM
I.S.
MRM
PEAK SHAPE CASE 2
Morphine
Overlay
peakofafter
two Low QC injections
fix
for an opiates method
Hydromorphone
Oxymorphone
Morphine
peak
before fix
POLLING QUESTION #2
Would you change the guard
column or column to solve
this problem (without any
additional information)?
PEAK SHAPE CASE 3
THC-COOH Normal
Quantifier
MRM
Qualifier
MRM
I.S.
MRM
WHAT WOULD CAUSE
ABNORMAL PEAK SHAPES ?
(ALL SAMPLES)
Stainless steel
ferrule shape,
length, size and
compression
screw shape,
thread length,
size varies by LC
vendor
http://www.sepscience.com/Techniques/LC/Articles/265-/HPLC-
Solutions-55-PEEK-Fitting-Slippage
MAKING A GOOD LC CONNECTION
REQUIRES TRAINING
1. Slide PEEK fitting quite far back from end of tubing
2. Bottom the tubing out in the mating port
3. Advance the fitting into the mating port while holding
the tubing firmly in the bottomed out position
4. Finger tighten, then + a ¼ turn, don’t over tighten
5. Tug to check that tubing doesn’t move
6. PEEK wears out – replace compression screws w
ferrule often
7. Cut PEEK tubing with perfectly flat ends, exact same
length as the old piece
of tubing (an ART)
POLLING QUESTION #3
Have you ever received any
training about the importance
of extra-column dead volume
and how to minimize it by
making good LC connections
& optimizing LC plumbing?
WHY - COLUMN OVERLOAD?
30μL
10 μLinjection
injectionof
or
low
high%%organic
organic
LC 2.1 mm
inner
column diameter
0.13 mm i.d. inlet
PEEK tubing
OR
Same
vial
after
new
ferrule
THC-COOH Qualifier
After re-extraction MRM
I.S.
MRM
POLLING QUESTION #4
Abnormal
appearance of
appearance
today’s Qualifier of THC-
THC-COOH MRM COOH
LLOQ calibrator
LLOQ
calibrator
Internal
Standard
MRM
LOW S/N – CASE 2
Today’s High
Benzodiazepine calibrator
Why does
the Fixed TIC w Fixed Y scale – 1.4 e8
Y scale 12%
matter?
100%
Without no additional
information – can we
conclude that these 3 low S/N
cases are all MSMS related
(MSMS is source of the
problem)?
HOW TO PROCEED?
Pump – Sample Prep – Autosampler – Column – Divert Valve – MSMS
(Tubing & Connections) (Unauthorized method edits)
X
X
Column Oven
ALS
MSMS
X
X
Pumps
Q3 collision Q1
Divert
Valve
cell
Waste
WHAT WOULD CAUSE LOW PEAK AREAS, W
NORMAL RTS, NORMAL PEAK SHAPES?
1. Sample preparation error
2. Too little volume injected from autosampler
a. Programming error
b. Autosampler problem (leak, partially
plugged needle or needle seal, ran out of
needle wash)
3. Post-column leak (why only post column?)
4. MSMS needs cleaning or calibration or
source component replacement or detector
voltage too low or M.P. contaminant = ion
suppression Wrong method – unauthorized method edit
HOW TO FIND THE SOURCE OF
LOW PEAK AREAS WITH NORMAL
CHROMATOGRAPHY?
1. Look at SST results, today & previous days
2. Inject stable extract from previous batch
6. Check for method edits
3. Autosampler problem 7. Check for solvent lot
change
a. Check vials & volume injected, check that
vial/plate caps were pierced
b. Inject extracts from other assays with similar
injection volume
4. Look for post-column leak, check LC pressure
5. Check MSMS signal, calibration, Rs - infusion
LOW S/N CASE 1
1. No SST done for THC-COOH, but Vit D
SST passed (so is the LC-MSMS OK?)
2. Autosampler OK? Inject THC-COOH
LLOQ calibrator from previous batch –
normal (autosampler [LC] & MSMS OK)
3. No leak from column to MSMS found,
normal starting pressure & pressure
trace Instrument looks OK –
problem with Sample Prep?
4. Ask the analyst who extracted the batch
– anything abnormal happen?
POLLING QUESTION #6
Abnormal
appearance of
appearance
today’s Qualifier of THC-
THC-COOH MRM COOH
LLOQ calibrator
LLOQ
calibrator
Degradation
Decreased ionization
WHY MASS CALIBRATION
– MASS RESOLUTION?
Q1 Mass Resolution & Mass Calibration result table
Target Found Mass Pass? Peak Pass?
Mass at (Da) Shift Width
(Da) (+/- 0.1) (0.6-0.8)
59.05 59.04
58.03 -0.01
-1.02 YES
NO 0.72
0.54 YES
NO
175.13 175.15
174.32 +0.02
-0.81 YES
NO 0.68
0.58 YES
NO
500.38 500.37
499.76 -0.01
-0.62 YES
NO 0.76
0.57 YES
NO
616.46 616.46
616.03 +0.003
-0.43 YES
NO 0.69
0.60 YES
906.67 906.65
906.47 -0.02
-0.20 YES
NO 0.65
0.63 YES
1254.93 1254.95
1254.84 +0.02
-0.09 YES 0.68 YES
POLLING QUESTION #8
With quadrupole mass
analyzers, higher mass
resolution (narrower MS peak
widths) = lower signal
abundance
True = YES
False or don’t know = NO
WHY DETECTOR VOLTAGE?
• The detector counts ions & amplifies
signal *Not a substitute for cleaning the interface,
increases noise as well as signal!
With clean
interface
components
With clean
interface
components
RT SHIFT CASE
WHAT MIGHT CAUSE A
PROGRESSIVELY SHORTER RT?
• Typically think leak = late Rt
• Failing A pump or leak before
mixer might = early Rt
• Check the pressure trace
OVER-PRESSURE HISTORY
• Rarely a problem until last few
weeks, then…..
• Several Rt shifts per week on both
instruments for protein-crash 25-OH
Vitamin D method, sometimes
associated with system shutdown
for over-pressure
• What changed?
INVESTIGATION
• Protein crash in ALS vials,
multi-mixer, centrifuge
• Checked multi-mixer &
centrifuge speeds – normal
• No extraction reagent
change or error found
• Checked ALS needle heights
• Raised needle 2 mm, no
further overpressure
POLLING QUESTION #9
With the same column, flow rate and
column temperature – will a 50:50
(vol:vol) H2O:Methanol mobile phase
yield a higher column backpressure
than a 20:80 (vol:vol) H2O:Acetonitrile
mobile phase?
True = YES
False or don’t know = NO
THE APPROACH TO
SUCCESSFUL
TROUBLESHOOTING
1. Know your instrument