Professional Documents
Culture Documents
TUME-ENG - LTD Tixel2 User Manual 09022020 - Compressed
TUME-ENG - LTD Tixel2 User Manual 09022020 - Compressed
TUME-ENG - LTD Tixel2 User Manual 09022020 - Compressed
/PO&6 QMBDFTUJDLFSXIFOBQQMJDBCMFPOMZ
User Manual TXLD0012
TABLE OF CONTENTS
Scope .......................................................................................................................................................... 5
Introduction ................................................................................................................................................ 5
This Manual ............................................................................................................................................... 5
Specifications of This Manual ................................................................................................................... 5
Glossary, Conventions, Abbreviations, Signs and Symbols ...................................................................... 6
Glossary ..................................................................................................................................................... 6
Abbreviations ............................................................................................................................................. 7
Signs and Symbols ..................................................................................................................................... 8
Operating State, Location and Position ...................................................................................................... 9
Electrical Safety Information ..................................................................................................................... 9
Electrical Medical Equipment Classification ............................................................................................. 9
Environmental Protection ........................................................................................................................ 11
Applicable Standards ............................................................................................................................... 12
The Tixel2 device – Overview ................................................................................................................. 12
System Layout and Components .............................................................................................................. 13
Principle of Operation and System Components ..................................................................................... 15
The Novoxel® Tip ................................................................................................................................... 15
The Standard Tip ...................................................................................................................................... 15
The Periorbital Tip ................................................................................................................................... 16
The Tixel2 Console .................................................................................................................................. 18
The Tixel2 Handpieces ............................................................................................................................ 18
Handpiece/s Receiving Inspection – Content and Functionality Verification ......................................... 21
Installation ................................................................................................................................................ 22
Turning on the Tixel2 device & Operating the Touchscreen ................................................................... 26
The Home Screen ..................................................................................................................................... 26
The Operating Screen ............................................................................................................................... 28
Pulse Credit & the Pulse Credit Screen ................................................................................................... 31
Purchasing of Pulse Credit ....................................................................................................................... 31
Entry of Credit Codes via the device’s Pulse Credit Screen - via the Home Screen ............................... 32
The Novoxel Application ......................................................................................................................... 33
System Shutdown ..................................................................................................................................... 34
Tip Cleaning ............................................................................................................................................. 37
Prepare the Following .............................................................................................................................. 37
Tip Cleaning Screen ................................................................................................................................. 37
Tip Sterilization ........................................................................................................................................ 42
Intended Use ............................................................................................................................................ 43
Indication for Use ..................................................................................................................................... 43
Contraindications for Use ........................................................................................................................ 43
Intended User ........................................................................................................................................... 44
Intended Treatment Areas ........................................................................................................................ 44
Pre-Treatment Patient Preparation ........................................................................................................... 44
Patient Pre-Treatment Preparation and Guidance .................................................................................... 44
In-Clinic Pre-Treatment Patient Preparation ............................................................................................ 45
TABLE OF CONTENTS
Determination of Treatment Parameters .................................................................................................. 45
First Treatment ......................................................................................................................................... 45
Single Pulse Treatment Characteristics for Caucasian Fitzpatrick I-IV Skin Phototypes ....................... 47
Comfort Level during Treatment ............................................................................................................. 47
Post-Treatment Heat Sensation ................................................................................................................ 47
Protrusion Level ....................................................................................................................................... 47
Device Preparation ................................................................................................................................... 48
Routine Treatment .................................................................................................................................... 48
Criss-Cross Mode ..................................................................................................................................... 49
Frame Segments ....................................................................................................................................... 50
Specific Skin Irregularities: Moles/Open Sores ....................................................................................... 52
Post-Treatment ......................................................................................................................................... 52
Typical Patient Response to Treatment ................................................................................................... 53
Example # 1 ............................................................................................................................................. 53
Example # 2 ............................................................................................................................................. 55
Post-Treatment Patient Guidance ............................................................................................................. 56
Safety Guidelines ..................................................................................................................................... 56
General Safety: Supervising Physician .................................................................................................... 56
Warnings and Precautions: Patient & Operator Routine Operation Safety ............................................. 56
Expected Possible Complications and Adverse Effects ........................................................................... 57
Maintenance ............................................................................................................................................. 59
Product Lifetime ...................................................................................................................................... 59
Technical Device Service ......................................................................................................................... 59
Cleaning ................................................................................................................................................... 59
Cleaning the Tixel2 Handpiece Distance Gauge (the DG) ...................................................................... 59
Storage and Packaging ............................................................................................................................. 60
Device Storage ......................................................................................................................................... 60
Disposal .................................................................................................................................................... 60
Periodic Inspection & Service Instructions .............................................................................................. 60
Annual Inspection by Certified Local Representative ............................................................................. 61
Essential Performance .............................................................................................................................. 63
Electromagnetic Emissions & Immunity ................................................................................................. 63
Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the Device ....................................................................................................................... 66
BLE Module with Chip Antenna Manufactured by Laird, Model BL652-SA ........................................ 66
The Responsibility for use of the Tixel2 device lies solemnly on its operator. By paying proper attention
to all safety information, you will be able to prevent injuries and product damage. It is within your
responsibility as the operator to use the product appropriately. You are also responsible if the product is
used for any intention other than its designation purpose or if the instructions in this document or others
are disregarded. In such case, Novoxel will assume no responsibility for such use of the product. All
procedures must be performed only after following Novoxel’s training policy, fully reading this manual
and proper familiarization with all safety and operation guidelines of the device, regulations, including all
applicable local or national safety regulations and rules for the prevention of accidents which must be
observed in all work places. The operator of the device must be familiar with the practical principles of
skin rejuvenation principles in general and fractional skin treatment in particular. The operator must
comply with all the guidelines, warnings and precautions portrayed in this document. Full compliance with
all local laws and regulations is obligatory for use of the device. It is fully within the operator’s
responsibility to verify that the use of the Tixel2 is permitted under local law. Using the product requires
technical skills and basic knowledge of English. It is therefore essential that only skilled, specialized and
trained staff or with the required skills be allowed to use the product. It is imperative to stop the operation
of the device immediately in case of detection of any malfunction.
NOTE: The approved CE medical indication by Novoxel’s NB is Actinic Keratosis (refer to page 43).
Chapter 1: Introduction
Scope
This document is the formal user manual for the Tixel2 device (P/N TXLD0012) which is a part of the
Tixel® Product Family. It provides important information regarding device safety, installation and setup,
operation, maintenance and service.
Introduction
This Manual
This manual contains important information regarding the Tixel2 device. In order to ensure the safety of
both operator and patient, as well as achieving optimal treatment results, please read this manual prior to
setup and operation. Verify that you are familiar with all safety instructions, guidelines, warnings and
precautions as well as the device’s operating principles.
Glossary
Term Definition
A metallic element, permanently mounted on the Handpiece of the
device, which performs the interaction with the treated tissue.
The Novoxel® Tip, or “Tip”
Several configurations/sizes exist, see description further down
this document.
A handheld apparatus, which is used by the operator to perform
The Tixel2 Handpiece
treatments. The Handpiece contains the Novoxel® Tip.
A back and forth motion of the Tip extending it towards the tissue
to be treated. A pulse is performed by pressing the Handpiece
trigger. A treatment is typically comprised of multiple pulses on
Pulse and Pulse Credit adjacent areas. The length of the pulse, or the “pulse duration”, is
controlled by the Tixel2 device’s Graphic User Interface (or GUI).
The Tixel2 arrives with a complimentary pulse credit. Once
exhausted, more pulse credit may be purchased by the customer.
The Tixel2 device may be operated in either single or double pulse
mode, which means that pressing the trigger of the Tixel2
Single/ Double Pulse Mode
Handpiece may result in either one or two pulses, depending on
user selection.
Overlapping Performing more than one pulse on the skin at the same location.
Performing two pulses on the same location but at a different
“Criss-Cross” Handpiece and Tip orientation. Further details, are available in the
operation segment of this manual.
The relative distance which the Tip’s vertices protrude beyond the
distance gauge’s distal end of the Handpiece towards the treated
Protrusion
tissue when a pulse is performed. The protrusion distance is
controlled by the Tixel2 device’s Graphic User Interface (or GUI).
Glossary
Abbreviations
Abbreviation Definition
The technology platform powering the Tixel2 device is named
Thermo-Mechanical Action (or TMA). The principle of operation
TMA
is described further down this manual. For further information
regarding this technology, refer to our website: www.Novoxel.com
HP The Tixel2 Handpieces or any other applicable Handpiece.
The Handpiece’s Distance Gauge, see description further down
DG
this document.
Abbreviations
Symbol Meaning
Manufacturer
Manufacturing Date
SN Serial Number
Batch Code
CE mark (Medical)
XXXX
The product may be operated only in full compliance with the operating
conditions and in the positions specified by Novoxel®.
Do not place the product on surfaces or tables that (for reasons of weight or stability, etc.) are
unsuitable for this purpose.
Do not place the product on a heat-generating surface or near heat emitting devices such as heaters.
Verify that there is sufficient clearance between the product and any other system that may exhaust
heat.
The product’s operating ambient range is temperature of 10 to 30°C (50 to 86 °F) and relative humidity of
20-80%. Novoxel® recommends that an ambient temperature of 20 to 25°C (68 to 77°F) and relative
humidity of 40-60% be maintained during normal operation, as this will result in better performance and
in longer MTBF of the equipment. The temperature must not exceed the maximum temperature specified
above.
Maximum operating altitude: 1000 m ASL (~3,300 ft.).
The following information on electrical safety must be observed, as failing to follow these instructions
may result in electric shock, fire and/or in serious personal injury or death.
Prior to switching the product on, always ensure that the nominal voltage settings of the product match
the nominal voltage of the AC supply network.
This product should be connected to the power mains using a 3-wire (two conductors and ground)
power cable and plug, complying with local regulations. Use this power cable with a properly
grounded electrical outlet to avoid electrical shock.
If extension cords or connector strips are implemented, they must be checked on a regular basis to
ensure that they are intact, positioned safely and safe to use.
The plug of the connecting cable is regarded as the disconnecting device. In such cases, always ensure
that the power plug is easily reachable and accessible at all times (corresponding to the length of
connecting cable, approx. 2 m).
Never use the product if the power cable is damaged. Check the power cable on a regular basis to
ensure that it is in proper operating condition. By taking appropriate safety measures and carefully
laying the power cable, you can ensure that the cable will not be damaged and that no one can be hurt
by, for example, tripping over the cable or suffering an electric shock.
Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into
the socket. Otherwise, sparks that result in fire and/or injuries may occur.
Do not overload any sockets, extension cords or connector strips; doing so can cause fire or electric
shocks.
Never remove the cover or any part of the housing. Doing so will expose circuits and components and
can lead to injuries, fire or damage to the product.
Use suitable over-voltage protection to ensure that no over-voltage (such as that caused by a bolt of
lightning) can reach the product. Otherwise, the person operating the product will be exposed to the
danger of electric shock.
The product is not liquid-proof; therefore, the equipment must be protected against penetration of
liquids. If the necessary precautions are not taken, the user may suffer electric shock or the product
itself may be damaged, which can also lead to personal injury.
Never use the product under conditions in which condensation has formed or can form in or on the
product, e.g. if the product has been moved from a cold to a warm environment. Penetration by water
or other types of liquids increases the risk of electric shock.
Prior to cleaning the product, disconnect it completely from the power supply. Use a soft, non-lint
cloth to clean the product, as instructed.
Prior to service procedures or replacement of any parts of the device, disconnect it completely from
the power supply.
Service procedures or replacement of any parts of the device should be performed only by qualified
service personnel.
Do not push any objects or fingers into the openings designated for Tip replacement or air ventilation.
Environmental Protection1
Waste electrical products should not be disposed of with household waste. Please recycle
where facilities exist. Check with your local authority for recycling advice or contact your
Novoxel product retailer for more information. The disposal of the Tixel system and/or
components should be performed according to the WEEE directive 2012/19/EU.
Applicable Standards
The European Council Medical Device Directive 93/42/EEC.
IEC 60601
The EMC Directive 2004/108/EC, Category IV Standards, harmonized under article 6(2).
IEC 60601-1-1-9
3.1 2 1 3.2
4.1 4.2
5.1 5.2
6.1 6.2
7
The Tixel2 Components
The main components of the Tixel2 are specified in the following table:
1 Tixel2 Console
2 Console touchscreen
5.1, 5.2 Handpiece umbilical cable connectors (left and right, respectively).
The HP moves the Tip through the distance gauge towards the skin (protrusion distance is modifiable).
The Tip comes in brief contact with the skin for a precisely controlled pulse duration (selected by the
user from a predefined number of options, depending on patient characteristics and desired effect).
The Tip is retracted back to its baseline position, safely secured and remote from the patient’s skin and
the operator, within the Handpiece’s distance gauge.
The pulse can also be modified by the user to single-pulse or double-pulse mode.
Two Tip configurations exist, the standard and the periorbital, which differ only by the size of the treatment
surface. All safety precautions and instructions for use apply in the same manner for both configurations.
Tip€31
s Treatment
Surface
A New Tip is pre-mounted on the Handpiece. It is clean and requires only minor
processing. For instructions, refer to the Tip cleaning instructions in Chapter 7 of
this document.
All information provided regarding the Tip is relevant for both Tip configurations.
Following high temperature heating, a thin oxidation layer is formed on the Titanium.
The layer acquires some mild colors such as blue, dark purple, and brown. This is a
normal phenomenon in heat-treated titanium. The tip is pre-conditioned by heat during
manufacturing, as it is important to create the oxidation layer for prevention of tissue
adhesion. It is biocompatible.
The HP is a handheld component of the Tixel2 device, used to perform treatment by producing mechanical
motion of the Novoxel® Tip, which is mounted on its distal end. Each of the Handpieces are plugged into
the console by a 1.5m (~5ft.) cable, which supplies power, control and enables cooling of the HP by
drawing air through its tubing by the Console’s blower.
Two Handpiece configurations exist: Standard and Periorbital. These are in fact identical Handpiece
designs, which differ only by: Tip configuration and Distance Gauge configuration (see further
information below regarding the Distance Gauge). Each HP also automatically sterilizes the Tip before
each treatment session.
Both Handpieces are shown below:
Distance Gauge
Trigger
The Distance Gauge is attached to the Handpiece on its distal end by magnets. The tip is pushed forward
and backwards through it during the treatment pulse. The DG is placed on the patient’s skin at the desired
location for treatment prior to a pulse. The Distance Gauge also functions as a safety precaution by
securing the operator and the patient from exposure to the hot Tip and to the heating unit. The DG can be
manually removed for cleaning of the Tip and for cleaning of the DG itself.
The Tixel2 system is supplied in rigid packaging. Open the box and place all components on a dry, stable,
clean surface. The Tixel2 device is provided with a Tip, mounted on the Handpiece.
The contents of the package is described in the following table:
The packed product may be heavy and should be handled with care. The system
installer may require suitable means of lifting and/or moving the product to avoid
physical injury.
The handles on the product packaging are designed exclusively to enable personnel to
transport the product. It is therefore not permissible to use the handles to fasten the
product to or on transport equipment or vehicle. The user is responsible for securely
fastening the products to the means of transport or lifting or on any of them.
Visually inspect (bare eye examination) the Novoxel® Tips on any Handpiece prior to first operation to
verify that:
Installation
The Handpiece cradles are packed separately from the Console. Insert each cradle’s legs (marked by the
red square) fully into the designated holes on each side of the Console. You may pull the cradle slightly
out for convenience, but always make sure it is firmly held in the Console.
Place each Handpiece in the Console’s Handpiece cradle as shown in the following illustration:
Ensure a safe path for the Handpiece’s cord, to prevent any undesired kinks, twists
or pinches.
The Handpiece connector inlet and the Handpiece connector are depicted in the following image set.
Note the Handpiece’s connector designated insertion notch and the matching console’s inlet insertion slot.
Plug the Tixel2 Handpiece connector into the Tixel2 Console inlet. Fully insert until a ‘Click’ is heard.
Follow the instructions in the following figures:
The Handpiece Connector Inlets in the Console (Left) and the Handpiece Cable Connector (Right)
Also, note the black insertion indication, which should be adjacent to the inlet’s edge.
Connecting the Handpiece to the Console [The Handpiece Connector in the image may differ from the one that you have]
Home Screen
Term Description
Navigation to the Operate Screen by Handpiece selection. By pressing on the icon, the
Operate Screen will be displayed after heating and sterilization of the selected Handpiece.
1.1 The other Handpiece will be heated to a hibernation temperature of 200°C during the
process.
1.2 The pulse count of each Handpiece is displayed below the icon, together with a color
visualization of the remaining life of the Handpiece (see Handpiece pulse count color
index table below).
2 Initiate the Tip cleaning process.
3 Decorative blue LED control (On/Off).
4 Demo Mode
5.1 Temperature indication, cord to the next color index:
Ready to work: Red 385-400°; Orange 200-384°; Yellow 80-199°; Blue 0-79°
5.2 Note: the accuracy of the temperature measurement is ±5° Celsius.
6 The Shutdown icon.
Home Screen Features
1 - 200,000
201,000 - 240,000
240,001 - 245,000
When the Handpiece expires, after 250,000 pulses, the following notification screen will be displayed until
a new Handpiece is connected:
Operating Screen
Item # Description
The Handpiece which was pre-selected in the Home screen, will be heated to
working temperature (400°C) and sterilized. When heated the icon will become
orange and blink.
The selected Handpiece shall be indicated by the red dot in the following icon:
When Handpiece temperature have reached 385°C and sterilized, the icon will
become green, and blink until treatment parameters are selected (see indents 4, 5 &
6 further below in this table).
After the treatment parameters have been selected, the Handpiece will become
active; The green icon will be displayed continuously and treatment can be
performed.
Note: The other Handpiece will remain at hibernation temperature (200°C) indicated
by an orange icon.
The icon will blink until treatment parameters are selected (see indents 4, 5 & 6
further below in this table). After the treatment parameters have been selected, the
Handpiece will become active; The green icon will be displayed continuously and
treatment can be performed.
Note: When both Handpieces are active (indicated by a green icon), switching
between them may be performed instantly. The non-active Handpiece will be
indicated by a blinking green icon.
Any activation after end of session will require re-sterilization, unless the
Handpieces remain in working temperature.
2 Indication of the selected Handpiece.
3 End of session and return to the Home screen.
Selection/De-selection of the “Double Pulse” option. This option will be displayed
4 for pulse durations of 8-16 milliseconds only. When selected, the icon will become
orange.
Pulse duration. Varies between 5-18 milliseconds. The increase/decrease is by
intervals of 2 milliseconds, except for a 1 millisecond gap between the 5-6 options.
5 Treat only after establishing adequate parameters. Not every
parameter setting is suitable for every patient. See clinical
information and guidelines for further information.
6 Protrusion. Varies between 100-1,000 microns, in 100 micron intervals.
Temperature indication, cord to the next color index:
Ready to work: Red 385-400°; Orange 200-384°; Yellow 80-199°; Blue 0-79°.
7.1
Handpiece heating will be indicated by a small upward arrow icon (˄) located above
7.2
the Handpiece temperature indication. Handpiece cooling will be indicated by a
small downward arrow (˅).
Pulse Credit counter. The counter is device specific and does not take into account
which Handpiece is used. The remaining credit is also indicated by color: Green for
up to remaining 2,500 pulses; Yellow for remaining 1,000-2,499 pulses, Blinking
Red for less than 1,000 remaining pulses.
8
For further details on pulse credit, go to the section named Pulse Credit (below) and
the Pulse Credit screen.
9 Current session pulse counter.
In order to purchase a new code, contact your local distributor. Please make sure you have your user name
(or request this information from your vendor), together with the serial number of the device for which you
would like to purchase the credit (stated on the label of the device). Once the purchase has been complete,
you will receive a message with the designated code.
Entry of Credit Codes via the device’s Pulse Credit Screen - via the Home Screen.
In order to feed a new code to the Tixel2 device, return to the home screen.
Press and hold on the letter “e” of the logo of the Tixel2 for four seconds.
The following screen will be displayed:
In case the pulse credit of the system has expired, the pulse credit screen will be
displayed automatically, and access to the Operate screen will be disabled.
Entry of a wrong code will be indicated by the letter “E” in the entry field.
Pulse credit may also be renewed via Novoxel’s designated applications, available for Android and iOS
users. The application requires an active Bluetooth connection to the Tixel2 device.
When additional pulse credit is purchased, a message is sent to the customer, with a link to download the
software in addition to the code itself and operation instructions.
System Shutdown
Cool-down of the Handpieces will be performed, followed by the following screen:
Shutdown Screen
After the Shutdown screen has been displayed, switch the system off via the power switch on the back
panel of the Console.
Chapter 5: Troubleshooting
The following Table describes all notifications/error screens which may be displayed during the operation
of the Tixel2 device:
Item # Description
The Handpiece which was pre-selected in the Home screen, will be heated to working
temperature (400°C) and sterilized. When heated the icon will become orange and blink.
The selected Handpiece shall be indicated by the red dot in the following icon:
When Handpiece temperature have reached 385°C and sterilized, the icon will become
green, and blink until treatment parameters are selected (see indents 4, 5 & 6 further below
in this table).
After the treatment parameters have been selected, the Handpiece will become active; The
green icon will be displayed continuously and treatment can be performed.
Note: The other Handpiece will remain at hibernation temperature (200°C) indicated by
an orange icon.
1.1
Preparing the Second Handpiece for Operation
1.2 Press on the hibernated Handpiece icon. The second Handpiece will be heated to working
temperature and sterilized for 1:00 minute. During the heating process, the icon will
become orange and blink.
After the sterilization process has been completed, the second Handpiece will have a
blinking green icon but without the red dot.
The icon will blink until treatment parameters are selected (see indents 4, 5 & 6 further
below in this table). After the treatment parameters have been selected, the Handpiece will
become active; The green icon will be displayed continuously and treatment can be performed.
“X” stands for either 1) signifying the left Handpiece or 2) signifying the right Handpiece.
P/N T2-UM-E-ENG Rev 10 Novoxel.com Page | 35
User Manual TXLD0012
E-01X1 Handpiece Motion failure, please shut down, restart and try
again. If error persists call service.
E-02X1 Handpiece heating failure, please shut down, restart and try
again. If error persists call service.
Tip Cleaning
A receptacle with Analytical Grade IPA (Isopropyl Alcohol; Propan-2-ol AR, 91% or higher).
Follow all safety instructions and storage for this material and its use, provided by its
supplier.
For the Distance Gauge cleaning only 70% Pharmacopoeia (medical) grade alcohol (70% Ethanol).
Cleaning should be performed using disposable gloves.
Press on the “Tip Cleaning” Button in the left upper corner of the Home Screen.
Temperature adjustment - The system will prepare the Tip for cleaning by either heating up to at least 80˚C
or cooling down below 200˚C. The following screen will be displayed:
When the desired temperature has been reached, the following screen will be displayed:
Select the Handpiece you would like to clean by pressing on its icon. After your selection, the following
screen will be displayed, instructing you to remove the Handpiece’s DG.
After removal of the Distance Gauge, the exposed Tip may be hot. The operator and/
or any other person/s standing nearby may be injured if contact with the distal end of
the Handpiece occurs. Always perform Distance Gauge removal with both hands
holding the Handpiece and facing the Tixel2 Console.
The Cleaning Process: Once the DG has been removed, the following screen will be displayed, instructing
you to commence cleaning:
Lightly soak the brush in the IPA fluid. Hold the Handpiece with the Tip pointing downwards.
Activate the brush and brush the Tip for 30 seconds at least, until visibly clean, as demonstrated
below:
After tip cleaning, clean the toothbrush using the medical grade Alcohol.
After cleaning the Tip, place the Handpiece in its cradle, without the DG. Soak a fiber-less wipe or
gauze in the medical grade Alcohol (70%) in order to clean the Distance Gauge thoroughly. Cleaning
should be performed using disposable gloves.
Heat cleaning of the selected Handpiece will now automatically commence. The process is 6:00
minutes long.
You may now select to perform cleaning of the second Handpiece as well. A countdown of 30 seconds
will be displayed, in which you may add the second Handpiece to the cleaning process by pressing on
its icon. If selected, you will then be requested to repeat the cleaning steps. The system will complete
the cleaning procedure by heating the Tip/s for the selected Handpiece/s.
If the second Handpiece was not selected, the heat cleaning process will continue to complete the
process.
After 6:00 minutes, the cleaning process has been concluded, the Operate screen will be displayed.
Tip Sterilization
Sterilization of the Novoxel® Tip shall be automatically performed before the system enables operation.
The Tixel2 device performs automated dry-heat sterilization. The process is performed after selection of
working parameters by the user and their confirmation in the “Operate” screen, which has been described
previously in this document. The process is 1:00 minute long. The Tixel2 device will notify when the
sterilization process has been completed and the device is ready.
The Cleaning process described above should be performed prior to your navigation to the “Operate”
screen. Sterilization will be performed as part of the “Operate” screen process.
When performing sterilization, the following screen will be displayed:
Intended Use
Dermatological procedures requiring ablative and/or non-ablative treatment of the skin. It is indicated for
use on human skin treatments.
Any patient who has undergone tanning during the 4 weeks prior to any treatment session and/or any
patient who plans to undergo tanning during the 4 weeks following any treatment session [patients
who may be exposed to the sun for short periods of time occasionally are not contra-indicated as long
as they apply a high SPF sunscreen (>50)].
Herpes Simplex.
Current skin cancer, malignant sites and/or advanced pre-malignant lesions or moles in the treatment
area.
Any female who is pregnant or plans to become pregnant during the treatment period, or is currently
breastfeeding.
An impaired immune system condition or use of immunosuppressive medication.
Previous invasive/ablative procedures in the areas to be treated within 3 months prior to initial
treatment or plans for such treatment during the course of the Tixel2 device treatment, or before
complete healing of such treatments has occurred.
Any patient who takes or has taken any medications (including via topical application), herbal
treatment (oral or topic), food supplements or vitamins, which may cause fragile skin or impaired skin
healing during the last 3 months.
Any patient who has used oral Isotretinoin (Accutane® or Roaccutane®) within 3 months prior to
treatment or less.
Any patient who has tattoos or permanent makeup in the treated area.
Any patient who has burned skin, blistered skin, irritated skin, or sensitive skin in any of the areas to
be treated.
Any patient who underwent thread lifting of the area to be treated in the last 3 months.
Intended User
The operation of the Tixel2 device, including any of its components, is restricted to trained, qualified and
professionals and must be under the supervision of a physician.
Any caregiver contact with the patient should be performed using disposable gloves.
It is recommended that a physician examine the patient prior to every treatment session.
Instruct the patient not to tan for 4 weeks prior to treatment. In case the patient has to be exposed to
sun for short periods, she/he must apply a high SPF sunscreen, of at least 50.
Patients with a history of Herpes Simplex may be prescribed prophylactic drugs cord to the
physician’s discretion.
Perform pre-treatment examination to verify the patient has not recently become contraindicated for
treatment.
The treatment may commence only after the patient’s skin has been thoroughly cleaned, cord to the
following instructions:
It is recommended to clean the treatment areas with soap and commercial grade distilled water/
thermal water. Remove all makeup, mascara clumps and/or any other materials, which may be on the
patient’s skin.
If the patient’s eyelashes and/or eyebrows are covered with mascara, either remove it or brush the
covered area in order to avoid mascara clumps during treatment.
Spray commercial grade distilled water/ thermal water on the treatment areas. Tap water is not
recommended.
Disinfect the patient’s skin, using an antiseptic lotion (such as Septeal by Pierre Fabre).
First Treatment
Tolerability Test: It is advised to perform a tolerability test prior to commencement of the first
treatment. Select a soft, non-frame, hairless segment and perform three non-overlapping pulses with
a protrusion level of 200 microns. Evaluate the patient’s response prior to treatment.
The first treatment should be performed with relatively low density of pulses, and mild treatment
intensity parameters (in terms of pulse duration and protrusion levels), lower than the maximum
parameters described in the parameter table further below this document. It is also recommended to
avoid double pulses, multiple passes and criss-cross pulses during the first treatment session. In
following treatment sessions, higher intensity may be reached gradually cord to the patient’s response,
treatment tolerance and physician’s discretion.
Single vs. Double Pulse Mode: The maximum single pulse duration recommended for Caucasian
skin treatment is 16 milliseconds. For double pulse duration, the maximum recommended pulse
duration is 10 milliseconds. Do not apply double pulses during a patient’s first treatment session.
Double pulses should be performed only on patients with thick skin, with relatively
low protrusion levels. Do not perform double pulses on frame skin segments (see
below).
Protrusion: The Protrusion level of the Tip should normally not exceed 500 microns. However,
increase of protrusion can be done cord to the physician’s discretion.
It is recommended that treatment parameters such as pulse duration and level of protrusion be
determined by the supervising physician, cord to the following treatment intensity recommendations
table:
Suggested Treatment Parameters
Single Pulse Treatment Characteristics for Caucasian Fitzpatrick I-IV Skin Phototypes
* Analgesic creams are not required. Comfort level is typically not dependent on skin phototype.
Protrusion Level
When performing ablative treatment, it is recommended to apply low parameters in the beginning in
order to examine the patient’s response, in terms of comfort level and heat sensation. Intensification
of the treatment’s parameters should be performed gradually.
Applying higher intensity parameters than denoted in the aforementioned segment should be
performed with caution and in accordance with the physician’s discretion, for example in cases of
thick skin.
Device Preparation
Clean the distance DG with 70% medical grade alcohol before and after every treatment.
Routine Treatment
The caregiver performing the treatment should always wear disposable gloves during
treatment.
The treatments should be at least 4 weeks apart. In case of thin/sensitive skin, consider longer healing
periods between treatment sessions.
During treatment, manual pulse overlapping of pulses is not prohibited, but should be
pursued with caution. It is advised to perform the treatment in a pre-determined,
systematic, directional manner.
Criss-Cross Mode
Consider working in Criss-Cross mode in deeply wrinkled segments. In this technique, the pulses (single
pulse mode only) are performed in the same location but not in the same orientation, as depicted in the
following figure:
It is recommended that the treatment be performed with the patient lying down.
In addition, do not overstretch the patient’s skin during treatment. In case of treatment of loose skin
areas, or areas which are not flat (such as back of hands), gently hold the skin straight in order to
flatten it. When working below the eyes (always maintain safe distance from the eyes), apply minimal
stretching by hand in order to avoid skin folding.
Position the Handpiece perpendicularly to the skin prior to activation, as demonstrated in the
following figure:
In case the perpendicular positioning of the Handpiece is not possible, do not treat the specific
location.
Do not apply excessive Handpiece pressure on the skin. The Handpiece should be positioned on the
skin but not pressed against it.
Stop applying pressure between pulses, in order to allow proper ventilation. The distance gauge
should not glide on the skin between pulses, but contact and then detach from the surface every time.
In order to prevent any sudden movements, it is recommended to provide hand support to the patient
(the hand that does not hold the Tixel2 Handpiece).
Frame Segments:
The Tixel2’s principle of operation relies on forwarding the Tip towards the skin for a controlled
pulse. Due to the fact that areas such as the jawline, forehead, back of hands, etc., are more sensitive,
it is recommended to apply lower intensity parameters when treating such segments. The facial
segments are demonstrated in the following figure:
Boney Facial
Treatment Areas
During facial treatments, the patient’s eyes should be closed, especially when working near the eyes.
The periorbital Handpiece is designated especially for use in the periorbital area (and/or any other
hard-to-reach areas). All precautions should be followed, particularly when working close to the eyes.
Décolleté (Chest): Flatten the treated segment when treating patients with loose skin. Move the skin
slightly when treating bony segments.
Do not treat.
Post-Treatment
If treated with a pulse duration of more than 10 milliseconds or a double pulse, do not apply makeup
for at least 48 hours after treatment. It is generally advised not to apply any makeup prior to soothing
of the face and crust disappearance. It is preferable to apply non-perfumed makeup.
Apply the ointment gently and avoid rubbing. Do not cover or cool the patient’s skin after applying
the lotion. No steroidal cream may be applied during the healing period.
Example # 1
Crusting, 3 Days After Treatment; on the Right - An Enlarged View of the Left Cheek
Example # 2
Crusting, 5 Days After Treatment (Left) / One Month After Treatment (Right)
The patient should be notified that the treated skin segments might become red for a few days, and
feel hot, even up to a burning sensation, for a few hours. Tightening of the skin may also be
experienced. It is recommended to apply any of the ointments mentioned above (see segment 7.10 –
Post Treatment), a few times daily for up to 8 days.
The patient should not tan for at least 4 weeks before and after treatment. In case the patient has to be
exposed to the sun for short periods, she/he must apply a high SPF sunscreen, of at least 50. Crusted
skin should not be exposed to sun at all.
In case back of hands treatment has been performed, the patient should refrain from working with hot
water/ cleaning materials/ other materials affecting the skin of the hands without the use of gloves,
especially when the skin is sensitive and/or crusting.
The patient should refrain for at least a week, after each treatment, from the following:
Any sport, especially swimming and/or any other activity involving sweating.
Safety Guidelines
It is advisable to inform the patient regarding the nature of the treatment. It is possible to demonstrate
the operation of the device by using the Demo Mode.
The operator should read this manual and its clinical guidelines in particular.
The operator should always be mindful of the fact that The Tip is hot during treatment.
As in all fractional skin treatment devices, treatment with the Tixel2 device may
pose a risk of dermal injury. The risk increases as treatment parameters are
intensified: exposure time (pulse duration) and pulse repetition (double pulse
mode in comparison to single pulse mode). Follow all clinical guidelines in this
document in order to ensure selection of the most suitable treatment parameters
for the patient.
Visually inspect the Tip prior to use. Do not use defective Tip and/or any other device components.
Damage to any of the device’s components, especially the Tip, may lead to system malfunction. In
case of such damage, please contact your vendor.
The Novoxel® Tip may be used only as a component of the Tixel2 device, in the specific fashion
portrayed in this document.
The Tixel2 device and the Novoxel® Tip do not contain or incorporate any medicinal substances and/
or any materials of biological origin.
Activate the handpiece when not touching the skin to verify proper operation.
Temporary swelling longer than a few hours, which does not recover without treatment.
Pain or burning sensation longer than a few hours and/or is not tolerable.
Forming of large scabs or scabs lasting more than 15 days.
Hyper-pigmentation
Hypo-pigmentation
Local burns.
Scars
Herpes Simplex
In case of any of the occurrence of any side-complication or adverse effect, it is imperative to consult
a physician within a short period.
Maintenance
The system does not require any specific maintenance except for a monthly visual inspection to verify that
the Console, the other components and the cables, are not damaged in any way.
In case of any detected damage, stop treatment immediately and contact your vendor.
Damaged and/or dirty equipment should not be used at all.
Product Lifetime
Product lifetime shall be 3 years after starting using the product but not more than 3.5 years after product
delivery to the customer. The Handpieces and the Tips shall be used for 250,000 pulses (assuming
approximately 1,250 treatments during a period of 3 years).
Cleaning
Clean the Tixel2 device and the Tixel2 Handpieces with 70% medical grade alcohol prior to each
treatment day. Cleaning should always be done using disposable gloves.
Prior to cleaning the product, shut it down and disconnect it from the power supply.
The device and the Handpieces and/or cables, should be cleaned after treatment using
a wipe soaked lightly with Ethanol 70% (see cleaning instructions below) and the Tip.
Remove the DG for cleaning purposes only, after the Tip has cooled down below
200°C.
Place the Handpiece in the console’s cradle.
Clean the distance DG with 70% medical grade alcohol before and after every treatment.
Cleaning the Novoxel® Tip - see Chapter 6: Caring For the Tips.
Device Storage
The device may remain plugged in to an in-wall power source while not in use. Prior to routine
storage, make sure the Tixel2 device is turned off. The Tixel2 device should be placed on a stable, dry
and clean surface. The Tixel2 Handpiece may remain plugged in as well.
For long-term storage or for shipment purposes, follow the next step prior to returning the Tixel2
device to its original packaging:
Place the Console and all components in the Tixel2 device’s designated box / bag.
For more information regarding boxing and shipment, contact your vendor.
Disposal
The disposal of the Tixel2 device and its accessories should be performed cord to WEEE directive 2012/
19/EU. In case you wish to dispose of the Handpiece and/or the Console, ship them back to Novoxel/ your
vendor.
The following instructions should be carried out by a trained operator of the device.
In general, the system does not require any special service other than cleaning and visual inspection.
However, certain inspections should be performed periodically as described herein.
Prior to activating the Handpiece, always check your surroundings to ensure safe
operation of the device. Always point the Handpiece away from yourself when
inspecting its action and integrity.
Visually inspect the system’s integrity. Verify that all the system components are well secured intact.
In case the components are damaged, the device should not be operated and delivered immediately to
a certified service provider for servicing.
Inspect the Tip for cleanliness. In case the Tip is not clean, perform the cleaning procedure as
described in this document.
Set system parameters to 6ms × 600 µm. Activate the Handpiece by pressing the trigger to produce a
pulse and visually inspect the Tip’s movement along its travel route forth and back. The movement
should be smooth without any substantial mechanical sound.
Inspect the Tip’s orientation within the Handpiece. Activate the system. Verify that the Tip moves
freely inside the lumen of the Distance Gauge during a pulse. The Tip should be centered. Slight
deviation from the DG lumen’s center is allowed.
Check the integrity of the Handpiece’s and its connection to the console.
An audible “click” must be heard upon insertion of the connector to its inlet in the console.
In any case of malfunction or any abnormality, the device and/or Handpiece should be delivered to a
certified service provider for servicing.
Carefully remove the upper cover of the handpiece to make sure that the warranty seal below is not
broken.
When turning on the device listen to the sound of the blower. If there is an abnormal noise or no sound
at all, the system should not be operated. Deliver it to a certified service provider for servicing.
Verify that the white LED in the Handpiece is lit.
Each Handpiece should be serviced by your local distributor after reaching 3 years of operation or
250,000 pulses.
Essential Performance
Temperature range shall vary between 385 and 405ºC.
Use of this device adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this device
and the other equipment should be observed to verify that they are operating
normally.
The use of accessories other than those specified for the device is prohibited. They
may result in increased electromagnetic emissions or decreased electromagnetic
immunity of the device.
The device is intended for use in the electromagnetic environment specified below. The operator of the
device should assure that the device is used in such an environment.
Electromagnetic Environment -
Immunity Test IEC 60601 Level Compliance Level
Guidance
Floors should be wood, concrete or
Electrostatic
8 kV contact 8 kV contact ceramic tile. If floors are covered with
discharge (ESD),
15 kV air 15 kV air synthetic material, the relative humidity
IEC 61000-4-2
should be at least 30%.
2 kV for power
Electrical fast
supply lines 2 kV for power
transient/burst, Mains power quality should be that of a
1 kV for SIP/SOP supply lines
IEC 61000-4-4 typical commercial or hospital
lines
environment.
Surge, 1 kV line to line 1 kV line to line
IEC 61000-4-5 2 kV line to earth 2 kV line to earth
0% UT for 0.5 0% UT for 0.5
cycle cycle Mains power quality should be that of a
typical commercial or hospital
Voltage dips and
0% UT for 1 cycle 0% UT for 1 cycle environment. If the user of the equipment
interruptions on
requires continued operation during
power supply input
70% UT for 25/30 70% UT for 25/30 power mains interruptions, it is
lines
cycles cycles recommended that the equipment be
IEC 61000-4-11
powered from an uninterruptible power
0% UT for 250/ 0% UT for 250/300 supply or a battery.
300 cycles cycles
Electromagnetic Environment -
Immunity Test IEC 60601 Level Compliance Level
Guidance
0% UT for 0.5 0% UT for 0.5
cycle cycle Mains power quality should be that of a
typical commercial or hospital
Voltage dips and
0% UT for 1 cycle 0% UT for 1 cycle environment. If the user of the equipment
interruptions on
requires continued operation during
power supply input
70% UT for 25/30 70% UT for 25/30 power mains interruptions, it is
lines
cycles cycles recommended that the equipment be
IEC 61000-4-11
powered from an uninterruptible power
0% UT for 250/ 0% UT for 250/300 supply or a battery.
300 cycles cycles
Power frequency magnetic fields should
Power frequency
be at levels characteristic of a typical
magnetic field, 30 A/m 30 A/m
location in a typical commercial or
IEC 61000-4-8
hospital environment.
Tx
Rx
/PWPYFMj-UE .EJ&VSPQB(NC)