Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

1 Tag Number HYU-AIT -722 .

/A2R7
2 Service CO Analyzer
3 Function Monitor CO in Hydrogen Product
GENERAL 4 Enclosure Weatherproof SST Enclosure to be included. See notes.
5 Power Supply 240 VAC 1Ph 50Hz
6 Maximum Distance to Element Built in Sensor
7 N2 Purge Nitrogen Purge flow 100 l/hr at 4 barg pressure.
8 Hazardous Area Class Zone 2 Gr.IIC T4
9 Material Mfg Std
10 Mounting Pipe Stand
11 Scale Type Digital Display
CASE 12 Scale Size Mfg std
13 Range Units 0-10 ppm CO
14 Range Expansion Yes
15 Temperature Compensation Yes
16 Ambient Allowed Temperature -20 to 55 ºC
17 Length n/a
PROBE 18 Cable Length n/a
19 Model LaserGas II MultiPass
20 Instrument Output 4-20 mA for 0 - 10 ppm CO. Galvanically isolated.
21 Load Impedance 500 ohms resistive
22 Alarm Contacts yes
OUTPUT 23 Contacts Type Relay
24 Contacts Rating Mfg Std
25 Ethernet interface To be included for remote diagnostics.
26
27
28 System Accuracy +/- 1% of smallest possible span
29 Repeatability +/- 1% of reading
PERFORM 30 Drift No drift
31 Linearity <1%
32 Detection Limit 0.2 ppm CO or 4% of the measured value
33 Manufacturer NEO Monitors as
34 Model LaserGas II Multipass Monitor
PURCHASE 35 Purchase Order Number
36 Price Item Number
37 Serial Number
Notes: 1. Vendor shall provide permanently attached SS tags, stamped with complete Instrument Tag numbers.
2. Weatherproof Cabinet in Stainless Steel material to be supplied for this analyzer. The Cabinet should have proper entries for tubing and
cable connections. All connection points to be marked properly and labeled for easy identification.
3. Process Sample Gas Conditions:
Sample Flow Rate: 5 to 10 L/min
Sample Pressure: 0.1 to 1.0 kg/cm2
Sample Temperature Limit: -20 to 55°C
Sample Particulate Limit: 99.99% of 0.01 micron
Sample Dew Point Limit: - 12 °C below ambient
Sample Actual Temperature: 40°C
Sample Fluid: Hydrogen Gas Product
Process Inlet / Outlet Connection: 1/4" NPT SwageLok fitting
4. Analyser Calibration Cylinders & Regulators by Client.
Analyser CO Zero Media: (nominal) 0% CO in balance of Hydrogen ( cylinder w/ regulator )
Analyser CO Span Media: (nominal) 8 ppm CO in balance of Hydrogen ( cylinder w/ regulator )
Process Sample Connection = 1/4" NPT
7. ATEX Zone 2 Gr.IIC Approval certificate required for this project in Pakistan.
8. Nitrogen Purge flow100 l/hr required at nominal 4 Barg pressure. Vendor to make arrangement for necessary valves, fittings,pressure
reducer and flowmeter for N2 purging.
9. Sample conditioning panel with filter, pressure regulator, overpressure relief valve and sample flow meter will be provided by Hydro-Chem .
Sample pressure will be set @ 0.4 kg/cm2 with a sample flowmeter of 1.7 to 17 lit/min scale.
Sample panel will also have valves for selection between inlet sample, span and zero gas.
10.Ethernet interface for remote diagnostic is included. Vendor to provide necessary software, if any.
11.LaserGas Analyzer is IP65.

LaserGas Analyzer
3 MP MP 10/29/2014 Tag number updated
2 MP MP 9/12/2014 Revised per HEC remarks
1 MP MP 4/4/2014 Issue for Purchase Client: Attock Refinery Ltd
0 MP MP 3/11/2014 Bid issue Plant: Pakistan Code: 105 Sheet 1 of 1
Status No. By Chk App Date Description Job No.: A2R7 Dwg. No.: 117-AIT-722 Rev.: 3