Professional Documents
Culture Documents
User Man SP Eng
User Man SP Eng
Hand-Held Spirometer
Manuale Utente
User Manual
Manuel d’utilisation
Benutzerhandbuch
Installation .............................................................................................................................................................................. 23
Before starting ...................................................................................................................................................................................................24
Checking the packing contents ....................................................................................................................................................................24
Optional modules .............................................................................................................................. Errore. Il segnalibro non è definito.
2
Options/Accessories .................................................................................................................................................................................... 24
Preparing Spiropalm.......................................................................................................................................................................................... 25
The connectors............................................................................................................................................................................................ 25
Upper part connectors................................................................................................................................................................................. 25
Lower part connectors ................................................................................................................................................................................ 25
Battery charger ........................................................................................................................................................................................... 25
Connect the oxymeter (option) to the Spiropalm ........................................................................................................................................ 26
Assembling the flowmeter to the handle (spirometry, Spiropalm 6MWT only) ........................................................................................... 26
Connect the Spiropalm to the PC ................................................................................................................................................................. 26
Connect the Spiropalm to an external printer .............................................................................................................................................. 26
How to contact COSMED .................................................................................................................................................................................... 27
Complaints, feedback and suggestions ....................................................................................................................................................... 27
Spiropalm Operation................................................................................................................................................................. 29
Main functions of the Spiropalm ........................................................................................................................................................................ 30
Turning on/off the Spiropalm...................................................................................................................................................................... 30
Menus......................................................................................................................................................................................................... 30
How to enter the menus ............................................................................................................................................................................. 30
Keyboard .................................................................................................................................................................................................... 30
Functioning principles of the keyboard ....................................................................................................................................................... 31
Calibration......................................................................................................................................................................................................... 33
Calibrating the flowmeter ........................................................................................................................................................................... 33
Patient database management ......................................................................................................................................................................... 34
Create a new patient ................................................................................................................................................................................... 34
Search a patient in archive .......................................................................................................................................................................... 34
Deleting a patient from the archive............................................................................................................................................................. 34
Performing the spirometric tests ....................................................................................................................................................................... 35
Warnings for spirometry tests ..................................................................................................................................................................... 35
Patient preparation..................................................................................................................................................................................... 35
Using the turbine ........................................................................................................................................................................................ 35
Using the PNT ............................................................................................................................................................................................. 35
Other operations ......................................................................................................................................................................................... 35
Forced Vital Capacity (pre) .......................................................................................................................................................................... 35
Test encouragement ................................................................................................................................................................................... 36
Slow Vital Capacity...................................................................................................................................................................................... 36
Maximum Voluntary Ventilation ................................................................................................................................................................. 36
Bronchial Provocation Test .......................................................................................................................................................................... 36
Bronchodilators test.................................................................................................................................................................................... 37
Bronchoconstrictor test ............................................................................................................................................................................... 37
Performing the oximetry test (option) ............................................................................................................................................................... 38
Warnings for oxymetry tests ....................................................................................................................................................................... 38
Patient preparation..................................................................................................................................................................................... 38
Performing the test ..................................................................................................................................................................................... 38
Performing the Six minute walk test (Spiropalm 6MWT only)............................................................................................................................ 40
Safety warnings .......................................................................................................................................................................................... 40
Required equipment ................................................................................................................................................................................... 40
3
Patient’s preparation ...................................................................................................................................................................................40
Connecting the mask and the turbine ..........................................................................................................................................................41
Placing the mask on the patient’s face.........................................................................................................................................................41
Wear the Spiropalm.....................................................................................................................................................................................41
Performing the test .....................................................................................................................................................................................42
Test results display ......................................................................................................................................................................................45
Printing the test ..........................................................................................................................................................................................45
Sources of variability ...................................................................................................................................................................................48
Factors reducing the 6MWD.........................................................................................................................................................................48
Factors increasing the 6MWD ......................................................................................................................................................................48
Viewing and editing results................................................................................................................................................................................49
View a test in archive...................................................................................................................................................................................49
Edit a diagnosis ...........................................................................................................................................................................................49
Deleting a test .............................................................................................................................................................................................49
Printing results by means of an external printer .................................................................................................................................................50
Options ..............................................................................................................................................................................................................51
General settings ..........................................................................................................................................................................................51
Spirometry options ......................................................................................................................................................................................51
Bronchoprovocation options ........................................................................................................................................................................52
Printout options ..........................................................................................................................................................................................52
Advanced options ........................................................................................................................................................................................52
Environmental data .....................................................................................................................................................................................53
Oxymetry options ........................................................................................................................................................................................53
Restore of initial settings .............................................................................................................................................................................54
Other functions of Spiropalm .............................................................................................................................................................................55
Calibration...................................................................................................................................................................................................55
Control panel ...............................................................................................................................................................................................55
LCD contrast ................................................................................................................................................................................................55
Memory deletion .........................................................................................................................................................................................55
Firmware version information .....................................................................................................................................................................56
Reset..................................................................................................................................................................................................................57
Reset of the unit ..........................................................................................................................................................................................57
Formatting the unit .....................................................................................................................................................................................57
4
Cleaning ..................................................................................................................................................................................................... 62
Cleaning Agents/supplies............................................................................................................................................................................ 62
Standard cleaning procedure ...................................................................................................................................................................... 62
Disinfection................................................................................................................................................................................................. 63
Preparing the disinfecting solution ............................................................................................................................................................. 63
The turbine flowmeter ................................................................................................................................................................................ 63
Cleaning the turbine ................................................................................................................................................................................... 63
Disinfecting the turbine for spirometry tests (Spiropalm only) .................................................................................................................... 63
Disinfecting the turbine (Spiropalm 6MWT only) ........................................................................................................................................ 64
Precautions to take when cleaning, disinfecting and drying the turbine ..................................................................................................... 64
PNT X9 ........................................................................................................................................................................................................ 64
VO2max mask and mixing chamber mask.................................................................................................................................................... 64
Disassembling the mask.............................................................................................................................................................................. 64
Cleaning the mask ...................................................................................................................................................................................... 64
Disinfecting the mask ................................................................................................................................................................................. 64
Sterilizing the mask .................................................................................................................................................................................... 65
Reassembling the mask and components ................................................................................................................................................... 65
Inspections ........................................................................................................................................................................................................ 66
Appendix.................................................................................................................................................................................. 67
Declaration of conformity .................................................................................................................................................................................. 68
Service - Warranty ............................................................................................................................................................................................. 69
Warranty and limitation of liability ............................................................................................................................................................. 69
Return goods policy for warranty or non warranty repair ............................................................................................................................ 69
Repair Service Policy ................................................................................................................................................................................... 69
Privacy Information ........................................................................................................................................................................................... 71
Personal data treatment and purposes ....................................................................................................................................................... 71
How your personal data is treated............................................................................................................................................................... 71
Consent....................................................................................................................................................................................................... 71
Holder of the personal data ......................................................................................................................................................................... 71
Customer rights .......................................................................................................................................................................................... 71
Disposing of electrical equipment...................................................................................................................................................................... 72
Safety and conformity ....................................................................................................................................................................................... 73
Safety ......................................................................................................................................................................................................... 73
EMC ............................................................................................................................................................................................................ 73
Quality Assurance ....................................................................................................................................................................................... 73
Medical Device Directive (CE mark) ............................................................................................................................................................. 73
Technical features.............................................................................................................................................................................................. 74
Predicted values ................................................................................................................................................................................................ 75
Automatic diagnosis (algorithm) ................................................................................................................................................................ 76
Quality Control Messages ............................................................................................................................................................................ 76
References ......................................................................................................................................................................................................... 78
Spirometry .................................................................................................................................................................................................. 78
Oximeter ..................................................................................................................................................................................................... 78
6MWT ......................................................................................................................................................................................................... 78
General ....................................................................................................................................................................................................... 78
5
6
Getting started
Important notices
Intended use
Spiropalm / Spiropalm Plus / Spiropalm 6MWT is an electrical medical device designed to perform pulmonary function tests. It is to be used by
physicians or by trained personnel on a physician responsibility.
Caution: Federal law restricts this device to be sold by the order of a physician.
This equipment is intended to be used for the following applications:
• Formulating of a lung pathology diagnosis.
• Assisting with human physiology studies.
• Contributing to sports medicine applications.
COSMED Srl is not responsible for incidents which occur due to improper use of this device. Examples include:
• Operation of the device by unqualifiedindividuals.
• Use of the device not indicated by this manual.
• Not complying with the precautions and instructions described in this manual.
Warnings
In this manual, the name Spiropalm refers to models Spiropalm, Spiropalm Plus and Spiropalm 6MWT. Spiropalm Only, Spiropalm Plus Only and
Spiropalm 6MWT Only refer respectively to the Spiropalm device, to the Spiropalm Plus device and to the Spiropalm 6MWT device.
The device, program algorithms and presentation of the measured data has been developed in accordance with the specifications outlined by the
ATS (American Thoracic Society) and ERS (European Respiratory Society). Additional international references have also been applied where
applicable. All bibliography references are reported in the Appendix.
This User Manual has been developed in accordance with the Class IIa European Medical Device Directive requirements.
The precautions listed below should be noted before operating the device to ensure the safety of the user.
1. This User Manual should always be available as a reference when testing.
2. The following standards should be applied to ensure the accuracy of individual test results:
• Accessories should only be used as described in this manual. The manufacturer does not warranty any non-authorized accessories used by
the end user. The manufacturer may offer suggestions while using such accessories and the complications they could cause;
• Repairs or modifications of the device should ONLY be carried out by qualified and trained personnel;
• Environmental and electrical conditions in which the device operates should be in compliance with the specifications of this manual. In
particular grounding reliability and leakage current suppression can only be assured when the device three-wire receptacle is connected
to a yellow-green return connected to earth ground. Attempting to defeat the proper connection of the ground wire is dangerous for
users and equipment.
• Equipment maintenance, inspections, disinfection and cleaning should be as described in this manual.
3. Before powering on the system, the power cords and plugs should be inspected. Damaged electrical parts must be replaced immediately by
authorized personnel.
4. Residue and other contaminants in the breathing circuit pose a safety risk to the patient during testing procedures. Aspiration of contaminants
can be potentially life-threatening. If the recommended disposable anti-bacterial filters are not used, you must disinfect each part coming into
contact with the patient and patient’s breath prior to each test.
5. The cleaning procedures and inspections in the System Maintenance section should be performed prior to each test.
6. This device should not be used in the presence of flammable anaesthetics. This is not an AP or APG device (according to the EN 60 601-1
definitions).
7. The device should not come near any heat or flame sources, flammable or inflammable liquids or gases and explosive properties.
8. The device should not be used in conjunction with any other medical device unless that device is recommended by the manufacturer.
9. The device should be used with a computer with electromagnetic compatibility, CE marking and low radiation emission displays.
10. The PC connected to the device must be compliant with EN 60601-1 by means of an isolation transformer.
OFF
ON
Alternating current
Performing forced expiratory manoeuvres involved in spirometry testing may be contraindicated in certain conditions.
Relative contraindications
For FVC tests:
• Spontaneous post-pneumothorax.
• Arterial-venous aneurysm.
• Severe arterial hypertension.
• Pregnant with complications in the 3rd month.
For MVV tests:
• Hyperventilation syndrome.
Absolute contraindications
• Severe bronchial obstruction (decreased FEV1 in adults).
• Recent myocardium infarct.
• Recent cerebral vascular accident.
• Known arterial aneurysm.
• Incapacity for understanding the provocation test procedures and its implications.
Relative contraindications
• Bronchial obstruction caused by performing respiratory manoeuvres.
• Moderate or serious bronchial obstruction (FEV1 < 1.51 in men and FEV1 < 1.21 in women).
• Recent respiratory infection.
• Recent Asthma exacerbation.
• Hypertension
• Pregnancy
• Epilepsy
Relative contraindications
• resting heart rate of more than 120
• systolic blood pressure of more than 180 mmHg
• diastolic blood pressure of more than 100 mm Hg
Note: Refer to the ATS recommendation in Appendix for a detailed description of the test and its interpretation, of the
contraindications, of the risk factors.
Electrical fast ±2 kV for power ±2 kV for power Mains power quality should be that of a typical commercial or
transient/burst supply lines supply lines hospital environment.
IEC 61000-4-4 ±1 kV for ±1 kV for
input/output lines input/output lines
Surge ±1 kV differential ±1 kV differential Mains power quality should be that of a typical commercial or
IEC 61000-4-5 mode mode hospital environment.
±2 kV common ±2 kV common mode
mode
Voltage dips, short <5% UT <5% UT Mains power quality should be that of a typical commercial or
interruptions and voltage (>95% dip in UT) for (>95% dip in UT) for hospital environment. If the user of the device requires
variations on power supply 0.5 cycles 0.5 cycles continued operation during power mains interruptions, it is
input lines 40% UT 40% UT recommended that the device be powered from an
IEC 61000-4-11 (60% dip in UT) for 5 (60% dip in UT) for 5 uninterruptible power supply or a battery.
cycles cycles
70% UT 70% UT
(30% dip in UT) for 25 (30% dip in UT) for 25
cycles cycles
<5% UT <5% UT
(>95% dip in UT) for (>95% dip in UT) for
5 sec 5 sec
Power frequency (50/60 3 A/m 3 A/m Power frequency magnetic fields should be at levels
Hz) magnetic field characteristic of a typical location in a typical commercial or
IEC 61000-4-8 hospital environment.
Nota: UT is the a.c. mains voltage prior to application of the test level.
Notes:
(1) At 80 MHz, the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
Spiropalm is a device designed for lung function screening; the core of the system is the “intelligent” flowmeter, connected to the main unit, with
graphical colour display. It can be considered a complete portable spirometric laboratory.
Spiropalm can be connected to a PC in order to transfer and store the performed tests, to view the tests and, if it is used the medical grade battery
charger, to perform the tests and display them on the PC monitor.
The system is composed by the flowmeter (a turbine for the Spiropalm / Spiropalm 6MWT, a pneumotachograph for the Spiropalm Plus), the
measurement and data elaboration device, the battery charger and the Software pack.
Spiropalm unit
The main elements a user can detect in the Spiropalm unit are the following:
• A display
• A keyboard OK
• Some connectors on the bottom side of the device.
The display
The display interfaces the device with the user, allowing the user to access to the functions of the device and to
display the performed tests. abc def
1 2 3
The keyboard
Spiropalm has a keyboard (see picture), which allows the user to interact with the device. ghi
5
jkl mno
4 6
In the upper group there are the OK key (at the centre) and four direction keys.
In the lower group there are 10 alphanumeric keys, the cancel and ON/OFF keys. pqrs tuv wxyz
7 8 9
A more detailed description of the keys functions, of the keyboard logics and of the menus, see the chapter
Spiropalm Operation. on/off
X 0
The connectors
On the bottom side of the device there are some connectors, described in the chapter Installation.
Battery charger
The flowmeter
Turbine flowmeter (Spiropalm only)
The turbine flowmeter is made of a handle (the reader) with a hole, in which it is placed the turbine.
The air passing through the helical conveyors, takes a spiral motion which causes the rotation of the turbine rotor. The rolling blade interrupts the
infrared light beamed by the two diodes of the reader. Every interruption represents 1/4 turn of the rotor, this allows to measure the number of
turn in the time.
For hygienic reasons, we strongly recommend the use of a bacterial filter, to be connected as in the picture by side.
Note: While inserting the turbine, be sure to push the turbine up to touch the end of the reader.
VO2m
ax
The turbine flowmeter assembly consists of a bidirectional turbine and an optoelectronic reader. The reader measures infrared light interruptions
caused by the spinning blade inside the turbine. The device may be used to measure a wide flow range and is not affected by ambient conditions
(pressure, humidity, room temperature, exhaled gas composition). Daily calibration of the turbine is not necessary, but calibrations should be
performed regularly to assure accurate measurements.
The PNT flow measurements are obtained by measuring the pressure differential between the two sides of a polyester net. The device may be used
to measure a wide flow range and is not affected by ambient conditions (pressure, humidity, room temperature, exhaled gas composition). Daily
calibration of the PNT is not necessary, but calibrations should be performed regularly to assure accurate measurements.
The PNT should be used for spirometry testing only.
Paper mouthpieces
Paper mouthpieces are available for basic spirometry tests (FVC, VC, and MVV). The mouthpieces should not be used for any other testing.
Antibacterial filters
Nose clips
Nose clips should be used during spirometry testing to prevent respiration through the nasal passage while performing testing manoeuvres.
The oximeter
The oximeter probe should be placed on the patient’s finger to measure oxygen saturation at rest or during exercise.
The 6MWT masks are made of silicone and may be reused after proper disinfection (see the chapter Maintenance).
These blue masks are available in different sizes and should be assembled the included head cap as shown in the 6MWT section.
Before operating the Spiropalm you should inspect the equipment and complete the product registration.
Options/Accessories
The following options are available with the Spiropalm system:
Code Quantity Description
C02390-02-05 1 Oxymeter
C00600-01-11 1 3 litres calibration syringe
C03611-01-10 1 Mask VO2max Small
C03613-01-10 1 Mask VO2max Large
C03103-01-10 1 Paediatric mask Large for VO2max
The connectors
Upper part connectors
1 2 3 4
1 2 3
Battery charger
Spiropalm is powered by a battery pack, rechargeable by means of a battery charger.
Connect the battery charger to the Spiropalm unit through the connector with the symbol .
Near the keyboard, a led indicates the battery status:
Green led In charge
Orange led Full charged
Warning: Spiropalm is not conceived to work connected to the mains. The device must be used with the battery
charger disconnected from mains.
Note: The batteries must be replaced when they do not maintain their charge for enough time. Please contact the
technical support.
1. Handle
2. Turbine ID28
You may contact the manufacturer directly at the following address for information:
COSMED S.r.l.
Via dei Piani di Monte Savello, 37
00041 - Pavona di Albano
Rome - ITALY
Voice: +39 (06) 931.5492
Fax: +39 (06) 931.4580
email: customersupport@cosmed.it
Internet: http: //www.cosmed.com
All the Spiropalm’s functions are controlled by means of the keyboard (see the following picture).
OK
abc def
1 2 3
ghi jkl mno
4 5 6
pqrs tuv wxyz
7 8 9
on/off
X 0
In order to turn on or off the device, hold the key pressed for few seconds.
Spiropalm has two automatic turning-off functions:
• of the display. If you don’t use the unit for at least the time set in configuration (see later, Options), the display turns off automatically. In
order to avoid this, press any key.
• of the device. If you don’t use the unit for at least 5 minutes, the unit turns off automatically. In order to turn on the device, press
on/off
the key.
Menus
Menus are displayed in the screen or, in some cases, in the top part of it.
Keyboard
Hereunder is a brief description of the functionality of each single button of the keyboard.
Button Description
Navigator tool: to scroll up a menu and/or get back to the previous field when
entering data
Navigator tool: to scroll down a menu and/or pass to the next field when
entering data
Navigator tool: to move right through the menu. To space right when
entering data and/or to select an entry in case of multi options fields
Navigator tool: to move left through the menu. To cancel the last digit when
entering the data and/or to select an entry in case of multi options fields
abc
2 To enter the number 2 or the letters A B C
def
3 To enter the number 3 or the letters D E F
ghi
4 To enter the number 4 or the letters G H I
jkl
5 To enter the number 5 or the letters J K L
mno
6 To enter the number 6 or the letters M N O
pqrs
7
To enter the number 7 or the letters P Q R S
tuv
8
To enter the number 8 or the letters T U V
wxyz
9
To enter the number 9 or the letters W X Y Z
X Cancel
The order of the sequence of each single button is always letters and figures (figure and symbols for the button).
• Pressing the Left key one or more times, the cursor will move to left deleting one or more digits.
For the fields that must be mandatory filled in, the left key allows to scroll the entry options from the list.
Please refer to the software manual for the set up and a calibration overview.
In the following you will find some useful notes for proper data entry.
• ID is an alphanumeric sequence, which identifies the patient. It is assigned automatically by the unit, but can be modified as you want.
The maximum length is 16 chars.
• The birth date must be entered as requested. The year must be four-digit format.
X OK
• Cancel with , confirm with .
OK
Scroll the list by means of the arrows and confirm the selected patient with the key .
X
If necessary, cancel the operation with the key .
Patient preparation
Using the turbine
Calibrate the turbine as indicated in the chapter Calibration (ATS recommends a daily calibration).
Other operations
Instruct the patient to apply the nose clips.
The use of antibacterial filters is strongly recommended.
Test encouragement
During FVC manoeuvre you might experience some lack of collaboration with kids or with other patients. In this case you may find a good help in
using the encouragement tool.
To start a test with the encouragement function, select 3.Options/2.Spirometry and enable this option.
Bronchoconstrictor test
The manoeuvre is identical to the FVC test.
1. Select the menu item 1.Test/3.FVC Post BC.
2. A window with bronchoconstrictor name, quantity and measurement unit, as set in 3.Options/3.FVC Post, will be displayed.
3. If you accept the protocol, confirm, otherwise modify the values and confirm.
4. Wait until the program is ready for the test.
5. After having performed the test, press 2.Stop or wait for the automatic end (5 seconds without flow), so that the device displays the V/T
graph, the main parameters, and the predicteds values.
6. Press 0.Abort in order to abort the test and discard the results. Press 1.Start in order to restart the test.
7. Repeat the test until it is correctly performed (ATS recommends 3 times), by pressing 1.Start.
8. The graph of the best test is shown, as the results of the three best tests. The best and the last test performed are highlighted.
9. Press 3.Curve to view a different test performed in the current session.
10. Press 2.Save & Exit in order to exit saving a test. You will be asked to select the test to store: press the digit corresponding to the test to be
stored.
11. Otherwise, press 0.Exit without saving in order to exit the test mode without saving the tests.
12. Repeat all the FVC Post until the FEV1 falls down by more than 20% of the FVC Pre value.
The oximetry test measures the haemoglobin saturation, i.e. the percentage of the blood haemoglobin bearing oxygen. The test can be performed
at rest or during/after a light exercise phase (cycling, jogging).
The test is completely automatic.
The measured parameters are:
Symbol UM Parameter
SpO2 % Haemoglobin saturation
HR bpm Heart rate
Patient preparation
Connect the sensor to the Spiropalm unit through the RS232/SpO2 port.
Connect the probe to patient’s index finger and affix with adhesive tape if necessary.
The Six minute Walk Test (6MWT) is commonly used into the pneumology departments for assessing the exercise capacity in COPD patients.
The procedure states that the patient walks as far as possible for 6 minutes. The distance is measured by counting how many laps of a circuit the
patient does.
As recommended by the ATS guidelines, there are also useful information at the beginning and at the end of the test, other than the covered
distance.
Note: Refer to the ATS guidelines reported in the Appendix for further information about the patient preparation, the
test execution, the sources of variability and the results interpretation.
Note: Cellular phones should be turned off to eliminate potential electrical interferences.
Safety warnings
The following warnings are taken from the ATS guidelines reported in Appendix.
1. Testing should be performed in a location where a rapid, appropriate response to an emergency is possible. The appropriate location of a crash
cart should be determined by the physician supervising the facility.
2. Supplies that must be available include oxygen, sublingual nitroglycerine, aspirin, and albuterol (metered dose inhaler or nebulizer). A
telephone or other means should be in place to enable a call for help.
3. The technician should be certified in cardiopulmonary resuscitation with a minimum of Basic Life Support by an American Health Association–
approved cardiopulmonary resuscitation course. Advanced cardiac life support certification is desirable. Training, experience, and certification
in related health care fields (registered nurse, registered respiratory therapist, certified pulmonary function technician, etc.) are also desirable.
A certified individual should be readily available to respond if needed.
4. If a patient is on chronic oxygen therapy, oxygen should be given at their standard rate or as directed by a physician or a protocol.
Reasons for immediately stopping a 6MWT include the following: chest pain, intolerable dyspnea, leg cramps, staggering, diaphoresis, and pale or
ashen appearance.
Technicians must be trained to recognize these problems and the appropriate responses. If a test is stopped for any of these reasons, the patient
should sit or lie supine as appropriate depending on the severity or the event and the technician’s assessment of the severity of the event and the
risk of syncope. The following should be obtained based on the judgment of the technician: blood pressure, pulse rate, oxygen saturation, and a
physician evaluation. Oxygen should be administered as appropriate.
Required equipment
Other than Spiropalm 6MWT, with oximeter and accessories, for the test is needed:
1. Mechanical lap counter
2. Two small cones to mark the turnaround points
3. A chair that can be easily moved along the walking course
4. Worksheets on a clipboard
5. A source of oxygen
6. Sphygmomanometer
7. Telephone
8. Automated electronic defibrillator
Patient’s preparation
1. Comfortable clothing should be worn.
2. Appropriate shoes for walking should be worn.
3. Patients should use their usual walking aids during the test (cane, walker, etc.).
40 - Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)
4. The patient’s usual medical regimen should be continued.
5. A light meal is acceptable before early morning or early afternoon tests.
6. Patients should not have exercised vigorously within 2 hours of beginning the test.
Connect the turbine and the oximeter to the Spiropalm through the proper connectors (see Installation chapter).
Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)- 41
Performing the test
1. Repeat testing should be performed about the same time of day to minimize intraday variability.
2. A “warm-up” period before the test should not be performed.
3. Select on the Spiropalm the patient’s name (2.View/Search… or 1.New to enter a new patient) and select 1.Test/7.Walking Test.
4. The patient should sit at rest in a chair, located near the starting position, for at least 10 minutes before the test starts. During this time, check
for contraindications, measure pulse and blood pressure, and make sure that clothing and shoes are appropriate. Enter data into the input
mask on the Spiropalm.
The display shows the test options.
42 - Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)
0 Nothing
0.5 Very, very slight
1 Very, slight
2 Slight
3 Moderate
4 Somewhat, severe
5 Severe
7 Very severe
10 Very, very severe (maximal)
• SBp and DBp shows the systolic and diastolic pressure.
• Suppl. oxygen shows if oxygen is delivered to the patient.
• In case, O2 Flow shows the delivered oxygen flow.
• Test duration is set at 6 minutes, as recommended by ATS. This value can be changed by the user in the range 2-30 minutes. In any case,
the test can be stopped by pressing 5.End.
5. Confirm by pressing OK and have the patient stand. In this first phase the physician must be sure that the readings are steady and the pulse
oximeter signal artifacts are minimized. The SpO2 should not be used for constant monitoring during the exercise. The technician must not walk
with the patient to observe the SpO2. The oximeter must be held in place (perhaps by a “fanny pack”) so that the patient does not have to hold
or stabilize it and so that stride is not affected.
The display shows, among the other parameters, the time elapsed from the beginning of this phase.
6. Press 1.Baseline to store data during the preliminary phase.
7. Instruct the patient with the following information:
• Object of the test: to walk as far as possible for 6 minutes. Running or jogging is not allowed.
• Rest periods during the test: since 6 minutes is a long time to walk, the patient will be exerting him/herself. If he/she gets out of breath or
become exhausted, it is permitted to slow down, to stop, and to rest as necessary. He/she may lean against the wall while resting, but
he/she must resume walking as soon as he/she is able.
• Route to walk (with practical example by the physician or other trained personnel)
8. During this phase, an inspiratory capacity can be performed. To do that, select 4.IC and perform this maneuver.
Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)- 43
The measured IC is reported in the real time data.
9. Position the patient at the starting line, stand near the starting line during the test. The physician/technician cannot walk with the patient.
Invite the patient to start walking as soon as he/she is ready.
10. When the patient starts walking, press 2.Start Walk to start the test and reset the timer.
11. Do not talk to anyone else but the patient during the walk. Use an even tone of voice when using the standard phrases of encouragement.
Watch the patient. Do not get distracted and lose count of the laps. Each time the participant returns to the starting line, click the lap counter
once (or mark the lap on the worksheet). Let the participant see you do it. Exaggerate the click using body language, like using a stopwatch at a
race.
During the test, it is possible to store a marker in the Spiropalm 6MWT by pressing 3.Mrk. Markers will be reported in the printout.
After the first minute, tell the patient the following (in even tones): “You are doing well. You have 5 minutes to go.”
When the timer shows 2 minutes, tell the patient the following: “Keep up the good work. You have 4 minutes to go.”
When the timer shows 3 minutes, tell the patient the following: “You are doing well. You are halfway done.”
When the timer shows 4 minutes, tell the patient the following: “Keep up the good work. You have only 2 minutes left.”
When the timer shows 5 minutes, tell the patient: “You are doing well. You have only 1 minute to go.”
Do not use other words of encouragement (or body language to speed up).
If the patient stops walking during the test and needs a rest, say this: “You can lean against the wall if you would like; then continue walking
whenever you feel able.” Do not stop the timer. If the patient stops before the 6 minutes are up and refuses to continue (or you decide
that they should not continue), wheel the chair over for the patient to sit on, discontinue the walk, stop the test (5.End) and note the
distance and the reason for stopping prematurely.
Warning: An acoustic and visual alarm warns the user if the SpO2 falls below 88% or if the HR exceeds the maximum of the predicted value
(220-age). Another warning warns the user if the oximeter signal is lost.
When the timer is 15 seconds from completion, say this: “In a moment I’m going to tell you to stop. When I do, just stop right where you are
and I will come to you.” When the time is ended, say: “Stop!”. An acoustic signal informs that the test is ended, and the data will be
44 - Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)
stored automatically. Walk over to the patient. Consider taking the chair if they look exhausted. Mark the spot where they stopped by
placing a bean bag or a piece of tape on the floor.
12. At the end of the test it is possible to perform another slow vital capacity by pressing 4.IC.
13. Exit from the test by pressing 0.Exit.
14. Enter the distance covered, the Borg dyspnea and fatigue levels and any other symptoms (dizziness, angina, hip leg or calf pain).
15. Congratulate the patient on good effort and offer a drink of water.
Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)- 45
2. Walk Test Summary
Among the data, the following ones deserve to be explained:
• 6MWW is the Six minute Walking Work, i.e. the product of distance and patient’s weight
• Pred 6MWD is the distance predicted value, according to the selected predicted set (see further, Advanced Options). This value is
reported only if the patient’s age is within the predicted range.
• Min 6MWD is the predicted inferior limit for the distance.
3. SPO2 & Cardiovascular
Among the data, the following ones deserve to be explained:
• Avg SpO2 is the average value of the oxygen saturation during the test
• T(<=88%) is the total time with oxygen saturation <= 88%.
• T(dSpO2>=4%) is the total time with the difference between baseline oxygen saturation and oxygen saturation during the test >= 4%.
4. Ventilatory Response (only if Measure Patient’s Ventilation was enabled before the test)
Among the data, the following ones deserve to be explained:
• Baseline FEV1 is the FEV1 measured during the pre-test spirometry
• BR (Breathing reserve) is calculated as (MVV-VE)/MVV
The Baseline IC and the Final IC (if measured) can help the analysis of the dynamic hyperinflation.
5. Upper graph
This graph shows the SpO2 and the HR during the test, and the markers stored during the test.
Two horizontal lines show the 88% of the SpO2 and the baseline SpO2 - 4%.
6. Lower graph (only if Measure Patient’s Ventilation was enabled before the test)
This graph shows the VE and the RF during the test.
Two horizontal lines show the MVV pred and the 30% of the breathing reserve.
7. Footer
A space for the comments and the operator’s signature.
46 - Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)
1
Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)- 47
Sources of variability
Factors reducing the 6MWD
Shorter height
Older age
Higher body weight
Female sex
Impaired cognition
A shorter corridor (more turns)
Pulmonary disease (COPD, asthma, cystic fibrosis, interstitial lung disease)
Cardiovascular disease (angina, MI, CHF, stroke, TIA, PVD, AAI)
Musculoskeletal disorders (arthritis, ankle, knee, or hip injuries, muscle wasting, etc.)
48 - Spiropalm Operation - Performing the Six minute walk test (Spiropalm 6MWT only)
Viewing and editing results
It is possible to pass from data to graph and vice versa by means of the Left and Right arrows.
Edit a diagnosis
To edit the diagnosis, select 3.Diagnosis.
Deleting a test
By pressing the 5 key, the selected test will be deleted.
Press OK to confirm.
When an external printer is connected through an USB connection to the Spiropalm unit, the test will be printed on it.
1. View the test to be printed.
2. Select 2.Print.
3. Select 1.Selected Test in order to print only the selected test or 2.Report in order to print a report for the selected patient.
Note: If the printer connected to the Spiropalm is placed within the patient area (according to the IEC 60601-1-1
norm), it has to be made compliant with the IEC 60601-1-1 norm by means of an isolation transformer.
Note: The printer must support the PCL language and have an USB port for data transmission.
The Spiropalm unit allows the configuration of some options, by means of the 3.Options menu.
General settings
Select the menu item 3.Options/1.Settings.
Spirometry options
Select the menu item 3.Options/2.Spirometry.
Printout options
Select the menu item 3.Options/4.Printout.
Advanced options
Select the menu item 3.Options/5.Advanced.
Environmental data
Select the menu item 3.Options/6.Environment.
Oxymetry options
Select the menu item 3.Options/7.SPO2.
Calibration
The calibration procedure allows exact measurements of flows, volumes and so on, correcting errors due to climatic changes, use of the device,
wear, etc.
Select the menu item 4.Utilities/1.Flowmeter Calibration.
For further details, see the chapter Calibration.
Control panel
The control panel allows to monitor the device status.
Select the menu item 4.Utilities/2.Control panel.
LCD contrast
In order to modify the contrast of the Spiropalm display:
1. Select the menu item 4.Utilities/3.LCD Contrast.
2. Adjust the contrast by means of the right and left arrow keys.
OK
3. Press the key to confirm.
Memory deletion
It erases all the memory contents.
1. Select the menu item 4.Utilities/4.Erase Memory.
OK
2. Confirm by pressing the key.
All service operations which are not specified in this user manual should be performed by qualified personnel in accordance with the service
handbook (to be required to the manufacturer).
All materials used in the construction of the Spiropalm are non toxic and pose no safety risk to the patient or operator.
Prior to the device cleaning, disinfection and inspection it is necessary to switch off the device and to disconnect adapters from the supply mains.
In order to guarantee the highest accuracy of measurements we recommend you to disinfect the turbine periodically.
Use disposable anti-bacterial filters or disinfect each part in contact with the patient before each test (only Spiropalm).
The pneumotacograph (Spiropalm Plus) is a disposable device. In this case it is not necessary to use antibacterial filters.
The goal of infection control is to prevent the transmission of infection to patients/subjects and staff during pulmonary function testing.
Cleaning and disinfecting instructions should be strictly followed to control infections and assure the safety of the patient. Aspiration of residue,
particles and/or contaminated agents could be life threatening.
The recommendations in the following section are retrieved from Miller MR, Crapo R, Hankinson J, et al.: General considerations for lung function
testing. Eur Respir J 2005; 26:153–162.
Cross-contamination
To avoid cross-contamination, reusable mouthpieces, breathing tubes, valves and manifolds should be disinfected regularly. Mouthpieces, nose
clips and any other equipment that comes into direct contact with mucosal surfaces should be disinfected, or, if disposable, discarded after each
use.
Only the portion of the circuit through which rebreathing occurs must be decontaminated between patients, or, if disposable, discarded after each
use. Disposable sensors, when appropriately used, avoid the need for decontamination of sensors and mouthpieces.
Tuberculosis
In settings where tubercolosis or other diseases that are spread by droplet nuclei are likely to be encountered, proper attention to environmental
engineering controls, such as ventilation, air filtration or ultraviolet decontamination of air, should be used to prevent disease transmission.
Introduction
Decontamination is a multi-step process that includes preparation at point of use, thorough cleaning and rinsing and a microbicidal process.
Thorough cleaning and rinsing are the first and most important steps in the reprocessing of any reusable medical device. Without thorough
cleaning and rinsing it might not be possible to achieve high level disinfection or sterilization of the device. The purpose of cleaning and rinsing is to
remove all adherent visible soil, to reduce the number of particulates and microorganisms, and to reduce the amount of pyrogenic and antigenic
material. Any organic material, lubricants, or residual cleaning agents remaining on a device can inactivate liquid chemical disinfectants/sterilants
as well as protect microorganisms from destruction.
The second step in decontamination is the microbicidal process which is defined as a process to provide a particular level of microbial lethality (kill).
COSMED components are classified as "semi-critical" items which are devices that come into contact with intact mucous membranes. Semi-critical
devices at a minimum require a high-level disinfection procedure. Sterilization is not absolutely essential.
COSMED components require complete or partial disassembly for cleaning and disinfection. It is the responsibility of the user (health care
personnel) for ensuring that: the cleaning methods recommended can be duplicated in their environment, that appropriate tools, and replacement
parts are available and that instructions are followed correctly.
Cleaning
Note: Please refer to additional, specific cleaning instructions for the turbine assembly below.
Cleaning Agents/supplies
Mild detergents with a neutral pH (7) are recommended for cleaning. Use warm water (22°-43°c) with the mild detergent. To be effective, cleaning
agents must assist in the removal of residual organic soil without damaging the device. cleaning agents should be used in the correct
dilution/concentration and at the correct temperature in accordance with the cleaning agents manufacturer's directions.
Cleaning supplies are very basic, usually consisting of a surgical scrub brush, chenille pipe cleaners, cotton or foam tipped applicators, soft brushes,
and soft cloths. Cleaning supplies should be cleaned and disinfected or sterilized daily.
Water Quality: tap water is acceptable for use in cleaning COSMED components.
COSMED components should be soaked and rinsed in tap water at 22°-43°C to prevent the coagulation of solid substances onto the device and thus
facilitate the removal of debris.
Enzymatic detergents with a neutral pH (7) are recommended when processing difficult-to-clean items with dried-on matter. Soaking mask and
valve components in an enzymatic detergent solution can effectively remove visible debris except for lubricants thus providing an acceptable
alternative to manual cleaning. Rinsing is necessary to remove all traces of detergent and extraneous debris.
Disinfection
The recommendations in this sections have been retrieved from:
William A. Rutala, Ph.D., M.P.H., David J. Weber, M.D., M.P.H., and the Healthcare Infection Control Practices Advisory Committee (HICPAC):
Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008
(http://www.cdc.gov/hicpac/pdf/guidelines/Disinfection_Nov_2008.pdf)
High-level disinfection is the recommended decontamination procedure for semi-critical devices.
Devices are classified semi-critical when they touches mucous membranes or broken skin. Examples of semi-critical devices are flexible endoscopes,
laryngoscopes, endotracheal tubes, respiratory therapy and anesthesia equipment, diaphragm fitting rings, and other similar devices.
3. Rinse the turbine in a vessel filled with water and shake gently to remove the disinfectant (do not place the turbine under running water or
move the turbine while submerged).
4. In order to let it dry, please shake it in air and/or connect it to the calibration syringe through the antibacterial filter and perform some strokes.
5. After cleaning the turbine, always calibrate prior to subsequent testing.
PNT X9
For cleaning and disinfecting the PNT X9 follow the same instructions as for the turbine, being careful to observe the following:
• Use distilled water for preparing the disinfectant solution, otherwise the calcium deposit could damage the flowmeter net
• Do not dry the PNT X9 with hot air which may damage the flowmeter net.
The equipment requires easy inspections to be carried out in order to assure a proper electrical and mechanical safety level in the years.
These inspections are highly recommended after a rough use of the equipment or after a period of storage in unfavourable environmental
conditions.
Referring to the electrical safety, it is important to check the conditions of insulation materials of cables, plugs and any other visible part by means
of simple inspection, when the equipment is switched off and adapters (or electrical feeders) are disconnected from the supply mains.
Extract the turbine from the unit and verify, by inspection, that the turbine axis fits correctly its seats and the blade is strongly fastened on the axis
itself (it can be useful to shake slightly the turbine in order to note any anomalous movement).
Check if there are any torn or broken components in the breathing circuits: remember that they can create safety risk to patients during tests.
USA contact:
COSMED USA Inc
2211 North Elston, Suite 305
Chicago IL 60614 USA
Phone: +1 (773) 645-8113
Fax: +1 (773) 645-8116
email: usa.sales@cosmed.it
To ensure that you receive efficient technical service, please specify the nature of the problem as indicated on the assistance information form.
You should save the original packaging in case the need to ship the unit to a technical assistance centre should arise.
Dear Customer,
We would like to inform you that your personal data is gathered and will be used by Cosmed Srl in conformity with the requirements of the Italian
privacy law (Decreto Legislativo 196/2003). We believe it is important for you to acknowledge how your personal data is handled.
Consent
The consent to treat your personal data is optional, but if denied COSMED cannot supply the appropriate services.
Customer rights
In accordance with Art.7, you may:
a) Obtain confirmation of the existence and sharing of your personal data.
b) Obtain information on the:
• updating, correction or integration of your data;
• deletion or transformation of your personal data;
c) Deny your consent to treatment of your personal data;
These rights can be exercised by a request in writing to the holder responsible for your personal data.
The device cannot be disposed as unsorted municipal waste. Electronic equipment must be collected separately according to the European Directive
2002/96/EEC. Otherwise it can cause dangerous consequences for the environment and human health.
The crossed-out wheeled bin means that the product must be taken to a separate collection when you wish to dispose of it.
Safety
IEC 60601-1/EN 60601-1;
The complete classification of the device is as follows:
• Internally powered equipment with applied parts type BF device (used stand alone), class II with applied parts type BF device (used
connected to mains)
• Protection against water penetration: IPX1
• Non sterile device
• Device not suitable in the presence of flammable anaesthetics
• Continuous functioning equipment
EMC
The system meets the Standard IEC 60601-1-2.
Quality Assurance
UNI EN ISO 9001:2008 (Registration n° 387-A Cermet)
UNI EN ISO 13485:2003 (Registration n° 387-M Cermet)
ERS93
Standardized Lung Function Testing: Official Statement of the European Respiratory Society, The European Respiratory Journal Volume 6,
Supplement 16, March 1993.
Compilation of reference values for lung function measurements in children: Ph. H. Quanjer, J. Stocks, G.Polgar, M. Wise, J. Karlberg, G. Borsboom;
ERJ 1989, 2, Supp.4,184s-261s.
KNUDSON 83
Changes in the Normal Maximal Expiratory Flow-Volume Curve with Growth and Anging: J. Knudson, D. Lebowitz, J. Holdberg, B. Burrows; ARRD
1983; 127:725-734
ITS
Intermountain Thoracic Society: Clinical Pulmonary Function Testing, second edition (1984) pp 101, 144
LAM
A survey of ventilatory capacity in Chinese subjects in Hong Kong: Lam Kwok-Kwong, Pang Shing et Al. Annals of Human Biology, 1982, vol. 9, No.
5, 459-472.
Multicéntrico de Barcelona
Spirometric reference values from a Mediterranean population: J. Roca, J. Sanchis, A. Agusti-Vidal, F. Segarra, D. Navajas. R. Rodriguez-Roisin, P.
Casan, S. Sans. Bull. Eur. Physiopathol. Respir. 1986, 22, 217-224.
Nhanes III
Spirometric reference values from a sample of the general US population: John L. Hankinson, John. R. Odencrantz and Kathleen B. Fedan. Am J
Respir Critr Care Med 1999, 159, 1798-187.
Pneumobil (Brazil)
Valores extraìdos do Programa Pneumobil/Brasil para a Tese de Doutoramento do Dr. Carlos Alberto de Castro Pereira. (Boehringer).
Gutierrez (Chile)
Gutierrez et Al. Reference values for Chile population
Knudson, Morris and Bass
The maximal Expiratory Flow-Volume curve: Knudson et al. ARRD Vol. 123, p. 659-664, 1981
Spirometric Standard for healthy non-smoking adults: ARRD Vol. 10-3, p. 57-67, 1971
Pereira (Brazil)
Pereira CAC; Barreto SP; Simões JG; Pereira FWL; Gerstler JG; Nakatani J. Valores de Referência para Espirometria em uma amostra da população
brasileira adulta. Jornal de Pneumologia 1992; 18: 10-22.
Mallozi MC. Valores de referência para espirometria em crianças e adolescentes, calculados a partir de uma amostra da cidade de São Paulo. Valores
finais publicados em : Pereira CAC; Lemle A; Algranti E; Jansen JM; Valença LM; Nery LE; Mallozi M; Gerbasi M; Dias RM; Zim W. I Consenso Brasileiro
sobre Espirometria. Jornal de Pneumologia 1996; 22:105-164.
Scalambrini Costa F, Scueiri CEB, Silva Jr WC, Pereira CAC, Nakatani J. Valores de referência para espirometria em uma amostra da população
brasileira adulta da raça negra. J Pneumologia 1996;22: 165-170.
Neder JA; Andreoni S; Castelo-Filho A; Nery LE. Reference values for lung function tests. I. Static Volumes. Brazilian Journal Medical and Biological
Research 1999; 32:703-17.
Neder JA, Andreoni S, Lerario MC, Nery LE. Reference values for lung function tests. II. Maximal respiratory pressures and voluntary ventilation. Braz
J Med Biol Res 1999 ;32:719-27
Thai
Wanchai Dejsomritrutai; Khun Nanta Maranetra; Kittipong Maneechotesuwan; Nitipatana Chierakul; Jamsk Tscheikuna; Tasneeya Suthamsmai;
Arth Nana; Benjamas Chuaychoo; Phunsup Wongsurakiat; Suchai Charoenratanakul; Wilawan Juengprasert; Chana Naruman: Reference Spirometric
Values for Healthy Lifetime Nonsmokers in Thailand, J. Med. Assoc. May 2000 (83: 457-466)
DLCO
Standardized Lung Function Testing: Official Statement of the European Respiratory Society, The European Respiratory Journal Volume 6,
Supplement 16, March 1993.
Spirometry
ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION TESTING, Edited by V. Brusasco, R. Crapo and G. Viegi: Standardisation of spirometry,
Eur Respir J 2005; 26: 319–338
Lung function", J.E. Cotes, Blackwell scientific publications
"Guidelines for Clinical Exercises Testing Laboratories", I.L. Pina, G.J. Balady, P. Hanson, A.J. Labovitz, D.W. Madonna, J. Myers. American Heart
Association. 1995; 91, 912.
Oximeter
National Lung Health Education Program (NLHEP) - Guide to prescribing Home Oxygen. By Thomas L. Petty.
ERJ 2004, 23: 932-646 - ATS/ERS Task force, B. R. Celli, W. MacNee, committee members - Standard for the diagnosis and treatment of patients
with COPD: A summary of the ATS/ERS position paper.
6MWT
ATS Statement: Guidelines for the Six-Minute Walk Test American Journal of Respiratory and Critical Care Medicine Vol 166. pp. 111-117, (2002).
Reference equations for the six-minute walk in healthy adults. Enright PL, Sherrill DL. AM J RESPIR CRIT CARE MED 1998;158:1384–1387.
Six minute walking distance in healthy elderly subjects. T. Troosters, R. Gosselink, M. Decramer. Eur Respir J 1999; 14: 270±274.
General
ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION TESTING, Edited by V. Brusasco, R. Crapo and G. Viegi: General considerations for lung
function testing, Eur Respir J 2005; 26: 153-161
ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION TESTING, Edited by V. Brusasco, R. Crapo and G. Viegi: Interpretative strategies for
lung function test, Eur Respir J 2005; 26: 720-735.
78 - Appendix - References
www.cosmed.com
COSMED srl
Via dei Piani di Monte Savello 37,
Albano Laziale - Rome
00041 ITALY
O +39 06 931-5492
F +39 06 931-4580
info@cosmed.com
cosmed.com