TECHNICAL SPECIFICATIONS

Optical Emission Spectrometer

ARL iSpark Series


PMT optics
• One meter Paschen-Runge mounting vacuum spectrometer
• Special cast iron body, temperature controlled at 36 °C ±0.1 °C
• Primary optics: heated focusing lens made of MgF2, CaF2 or fused silica,
depending on application
• Ruled gratings: 1080, 1667 or 2160 grooves/mm, appropriately selected
for the analytical task
• Primary slit width: 20 μm
• Secondary slit width: 25, 37.5, 50 or 75 μm
The Thermo Scientific™
ARL iSpark™ Series is a high- • Up to 80 photomultiplier tubes (PMT) detectors
performance optical emission • PMTs: ∅ 28 mm, side-on-type, 10-stage, windows: MgF2, UV glass, borosilicate
spectrometer platform based on glass or synthetic silica appropriately selected for the specific wavelength
the famous Thermo Scientific • Wavelength range depending on grating and diffraction order, selected
one-meter vacuum PMT optics for optimum resolution and sensitivity. Wavelength range: 120-800 nm
and innovative technologies. The • Diffraction order selective filters in front of PMTs for gratings with 1080
ARL iSpark Series answers the and 1667 grooves/mm
needs of the different market
segments of the metallurgy and Spark stand
brings an optimum solution for • Stand cover with ergonomic design for easy operation with maximal
everyone. If offers high accuracy, operator security
sensitivity, precision, stability and
• Maximal sample size: 250 mm width x 250 mm depth x 150 mm height,
reliability.
maximal weight 10 kg
• Sample clamp with ergonomic design • Spark-DAT methods for determination of soluble/
• Wear-resistant stainless steel analysis table with insoluble concentration and for inclusion analysis
quick fastening system; no tools needed for
Vacuum system
assembly and dismantling
• Rotary pump for wavelength range down to 170 nm
• Design allowing argon savings and reduced maintenance with speed control for reduced electrical consumption
• Smart Argon Management (SAM) with programmable • Membrane pump combined with molecular pump for
argon flows and argon saving ECOmode operation down to 119 nm
• Argon flows can be different for each phase of the
Dimensions and weight
analytical sequence (flush, pre-burn, integration).
• Dimensions: 1750 m length x 860 m width x 1200 m
Optimized for each application, between 2 to 5 l/min
height (inches: 68.9 x 33.9 x 47.3)
(typically 3 l/min) in analysis and 0.4 /min in stand-by
requiring the minimum argon consumption • Weight: approx. 500 kg net (1100 lbs), 570 kg with
pallet (1260 lbs)
• Argon pulses after analysis for improved dust
evacuation and instrument autonomy Requirements
• Electrode: ∅ 1.4, 4 mm or pin electrode according to • Ambient temperature: 16-30°C (62-86°F); maximum
application; 3 mm analytical gap rate of change ±5°C/hour
• Relative humidity: 20-80%
Control electronics
• Instrument master control board (IMC) controlling • Voltage 230 V (+10% to -15%), single-phase with
all spectrometer functions, data flows and protective ground (2.5 kVA regulator required if
communication with OXSAS analytical software fluctuations exceed +10%)
• Current: 10A, including PC, screen and printer
Spark generator
• Frequency: 50 or 60 Hz
• intelliSource, current controlled, double current
source (i.e. a double CCS) for high shape flexibility • Grounding: < 1 Ohm
and reproducibility • Argon: > 99.998% maximum 5 ppm oxygen
• Vectored spark shape definition, up to 200 A and (maximum 2 ppm oxygen for samples with high
2500 μs single spark duration Si content). Optional argon purifier available and
recommended for low carbon analysis. For VUV lines
• Spark short-circuit (DISC) to improve spark repeatability
(carbon, nitrogen, oxygen, chlorine, hydrogen), argon
Acquisition electronics and algorithmics purifier recommended
• Programmable high voltage settings: -300 to -1000 V
Consumption
for each channel
• Electrical power: 1 kVA
• For each channel:
• Argon: typically 3 l/min. during analysis, 0.4 l/min in
–– Very low noise time gated differential integrator
stand-by and down to 0 l/min in ECOmode
allowing PMT dark current and electronic offset
subtraction Compliance to norms
–– TGA (Time Gated Acquisition), a ultrahigh • 2006/95/EC: Low Voltage Directive (LV)
precision (100 ns) TRS from 2 to 6000 μs • 2004/108/EC: Electromagnetic compability (EMC)
–– ADC bit depth: 20 bits • 2011/65/EU: Restriction of the use of certain
–– SSA (Single Spark Acquisition) and storage Hazardous Substances in Electrical and Electronic
–– FAST (Flexible Acquisition STart/stop) algorithm, Equipment (RoHS)
for usage of optimal single sparks subsets • 2012/19/EU: Waste Electrical and Electronic
• DISIRE (DIffuse Spark Intensity REmoval) algorithm Equipment (WEEE)
for improved accuracy

Find out more at thermofisher.com/ispark


©2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific. This
information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage
use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing
are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
PS41297-EN 0917