Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

1274 CHROMATOGRAPHY/High-performance Liquid Chromatography

Modified cellulose such as DEAE-cellulose can be High-performance Liquid


used for ion-exchange separations on TLC. Separ-
ation by size-exclusion chromatography can be
Chromatography
carried out with TLC plates coated with Sephadex M V Moreno-Arribas and M C Polo, Instituto de
gel but is slower and significantly more difficult Fermentaciones Industriales (CSIC), Madrid, Spain
than other forms of TLC. Polyamides such as Copyright 2003, Elsevier Science Ltd. All Rights Reserved.
polyhexamethylenediaminoadipate can be used as
the adsorbent to separate components that interact
with it by hydrogen bonding. (See Chromatography:
Principles.) Introduction
High-performance liquid chromatography (HPLC) is 0001

Applications an instrumental form of liquid chromatography that


employs stationary phases consisting of small par-
0024 The actual method employed for preparing samples
ticles, thereby achieving more efficient separations
from foodstuffs for analysis by TLC depends on the
than those used in conventional liquid chromatog-
nature of the substances being investigated. Many
raphy. Since its origin in the late 1960s, it has been
methods involve the extraction of the food with a
known by several different names, including high-
suitable solvent followed by precipitation and filtra-
pressure liquid chromatography, because of the
tion steps to remove classes of compounds that are
high pressures required to force the mobile phase or
not of interest. (See Analysis of Food.)
solvent through the stationary phase, and high-
0025 The choice of adsorbent and solvent system used in
resolution liquid chromatography, because of the
TLC is dictated by the nature of the sample to be
good resolution achieved using this technique.
analyzed. Examples of broad groups of substances
This article describes the equipment needed to 0002
that have been successfully resolved by TLC methods
carry out HPLC and summarizes some of the app-
are shown in Table 1, along with suggested solvent
lications of this technique in food analysis. Two
systems and detection reagents commonly used.
techniques related to HPLC, fast-protein liquid chro-
However, for most substances, there can be a variety
matography (FPLC) and supercritical fluid chroma-
of different solvent systems and detection reagents
tography (SFC), are also considered. The basic theory
that are suitable, depending on the precise compos-
of the chromatographic process and the factors
ition of the samples to be resolved.
that affect separation efficiency are discussed in
the section on Chromatography. (See Chromatog-
See also: Chromatography: Principles; High-
raphy: Principles; Thin-layer Chromatography;
performance Liquid Chromatography; Gas
Chromatography High-performance Liquid Chromatography; Gas
Chromatography; Supercritical Fluid Chromatog-
raphy; Combined Chromatography and Mass Spec-
trometry.)
Further Reading
Ackman RG (1982) Flame ionization detection applied to
thin-layer chromatography on coated quartz rods. Separation Modes
Methods in Enzymology 72: 205–252.
Henderson RJ and Tocher DR (1992) Thin-layer chroma- There are many different stationary and mobile 0003

tography. In: Hamilton RJ and Hamilton S (eds) Lipid phases that can be used in HPLC, and for this reason
Analysis: A Practical Approach, pp. 65–111. Oxford: there is a great variety of separation mechanisms.
Oxford University Press. The treatises on HPLC use different criteria in their
Jork J, Funk G, Fischer G and Wimmer CJ (1989) Thin- attempt to classify the modes of this technique:
layer Chromatography: Reagents and Detection type of stationary phase, predominant separation
Methods. Cambridge: VCH. mechanism, type of groups of compounds it is
Kirchner JG (1978) Thin-layer Chromatography. Tech- aimed at. These criteria sometimes overlap; that is
niques of Chemistry, vol. XIV. New York: Wiley.
to say, it is possible to work in several different chro-
Sherma J (1996) Thin layer chromatography in food
analysis. Food Testing and Analysis 2: 12–15.
matographic methods with the same stationary
Touchstone JC (1988) Instrumentation for thin-layer chro- phase, and the same group of compounds can be
matography: a review. Journal of Chromatographic separated using several different stationary phases,
Science 26: 645–649. and for this reason it is not easy to classify HPLC
Wilson ID (1996) Thin layer chromatography: a neglected techniques into specific groups. Generally speaking,
technique. Therapeutic Drug Monitoring 18: 484–492. it can be said that there are three large groups
CHROMATOGRAPHY/High-performance Liquid Chromatography 1275

that dominate the field of chromatographic separ- salt that competes with the solutes in their interaction
ations: reversed-phase chromatography, ion exchange with the charges of the stationary phase. The charges
chromatography and size exclusion chromatography. permanently bonded to the chromatographic support
There are also other types of chromatography that are compensated by counterions, that is to say, by free
are rapidly gaining in importance and cannot be in- ions with the opposite charge. Passage through the
cluded exactly in any of these three groups: ion pair column of a sample or of the mobile phase with ions
chromatography, affinity chromatography, and chiral of the same sign as the counterions of the column may
separations. result in the displacement of the counterions and the
retention of the ions of the sample or of the stationary
Reversed-Phase Chromatography phase.
0004 This is the most commonly used method for chroma- The ion exchangers can be classified as cationics 0008

tographic separations, so it could in fact be called and anionics. Within each group it is possible to
‘normal-phase’ except that it was developed later differentiate strong and weak exchangers. Strong
than the latter. In reversed-phase chromatography, cation exchangers (SCX) usually contain sulfonic
the stationary phase is of a nonpolar nature and the groups, while weak ones have carboxylic groups.
mobile phase is more polar than the stationary one. Quaternary ammonium salts are the most usual func-
Among the most widely used stationary phases are tional groups in strong anion exchangers (SAX), and
silica-based ones with cyano, phenyl, or alkyl func- primary amines are used in weak anion exchangers.
tionality and those of a polymeric nature, above all
Size Exclusion Chromatography
polystyrene-divinylbenzene (PSDVB). The mobile
phase is usually water with an organic modifier This separation method is based on the size of the 0009

(methanol, acetonitrile, or tetrahydrofuran). Ternary molecules. For many years it was known as gel
or quaternary mixtures of these components can be permeation chromatography (GPC) or gel filtration
used to improve selectivity. chromatography (GFC). The retention of the mol-
0005 There is controversy over the mechanism that ecules depends on their size: the larger molecules
controls reversed-phase separations. The three basic that do not enter the pores of the packing elute first,
models are the partitioning model, the solvophobic while the smaller molecules, whose diameter makes it
effect, and the adsorption model. It is also necessary possible for them to enter and exit the pores of the
to consider the possibility that a mixed mechanism chromatographic packing, take a longer time.
might exist.
Ion Pair Chromatography
0006 In the partitioning mechanism it is accepted that
the retention of a solute in reversed-phase liquid This chromatographic method enables the separation 0010

chromatography is due to its partitioning between of solutes that are ionized in separation conditions.
the mobile phase and the stationary phase. This On most occasions the separations are carried out on
mechanism assumes that the stationary phase behaves the same columns as those used in reversed-phase
in a similar way to a liquid, a situation that may occur separations without the formation of ion pairs. The
in certain conditions, such as when working with mobile phase is made up of a water buffer and an
monomeric phases and high percentages of water in organic modifier, to which mixture is added a coun-
the mobile phase, and the size of the hydrocarbonated terion with an opposite charge to that of the solute.
chains is large enough for them to be associated. The Quaternary ammonium salts are often used as
adsorption mechanism accepts that retention takes counterions for the analysis of anion substances,
place through the interaction of the solute to the while n-alkyl sulfate salts are among the most com-
organic molecules bonded to the surface of the monly used for the analysis of cation substances.
packing matrix. The solvophobic effect theory
accepts that the solute is expelled from the mobile Affinity Chromatography
phase towards the stationary phase. This is based on the specific interaction between a 0011

molecule, or group of molecules, and a ligand,


Ion Exchange Chromatography
which is a molecule that is attached to the station-
0007 This is used for the analysis of electrically charged ary phase to interact with those of the solute. The
substances in working conditions. Stationary phases growing importance of affinity chromatography is
with electrically charged functional groups are used due to the development of bioaffinity chromatog-
for this method of chromatography. The support can raphy in which the interactions between ligands and
be silica or a polymer. The mobile phase contains a molecules are based on recognitions of a biological
buffer which controls the loading of the solutes, and a type.
1276 CHROMATOGRAPHY/High-performance Liquid Chromatography

Chiral Separations uptake of air. To prevent bubble formation in the


detector cell with depressurization, a restrictor is
0012 These can be approached mainly in two ways; by
sometimes attached to the outlet of the detector cell.
transforming enantiomers into chromatographically
separable diasteroisomers, which does not necessitate
Pumps
the use of chiral phases, or by preparing chiral phases.
Chiral phases can be obtained by covering the pack- The pump is the system for delivering the solvent 0016

ings by making a solution pass through the column, from the solvent reservoir to the column through the
or by ionic or covalent bonding of the chiral reagent injector. Basically, two types of pumps are available:
with the column packing material. constant-pressure pumps and constant-flow-rate
pumps. The latter are more frequently used in HPLC.
Constant-pressure pumps are less expensive and 0017
Instrumental Configurations easy to operate, but the flow rate may vary with
0013 The basic equipment consists of a column packed changes in the viscosity of the mobile phase caused
with a stationary phase, a driving force to propel the by temperature fluctuations or by the accumulation
solvent through the column (pump), a system (in- of undissolved sample components in the column.
jector) for introducing the sample on to the column, These variations in flow rate affect retention times,
a system (detector) for measuring a physical property and they may also affect resolution, increasing
of the solutes being analyzed that differs from the the difficulty of both qualitative and quantitative
properties of the solvent of a property of the mobile analysis.
phase which is altered by the presence of the solute, Constant-flow-rate pumps afford the advantage of 0018

and a system for recording the detector signals and maintaining retention times irrespective of changes in
converting them into graphic traces or chromato- solvent viscosity. This type of pump includes syringe
grams. pumps, which consist of a cylinder containing the
0014 A single solvent is often used to carry out the mobile phase, which is expelled by a piston. The
separation (isocratic elution), but differing propor- piston is driven by a motor, so as to supply a constant
tions of various solvents are also often used (gradient flow devoid of pulses. This type of pump can achieve
elution), in which case a gradient device is needed. A relatively high pressures, but maintenance and chang-
variety of accessories, such as pressure controllers, ing solvents are complicated.
valves for switching solvents, valves for switching Reciprocating pumps, also a type of constant-flow- 0019

column, and ovens for heating the columns, are also rate pump, are most commonly employed. Their price
commonly employed. Today most chromatographs varies depending on their complexity of design, and
are controlled by a computer, which is also used for their main drawback is that they generate pulses that
data collection. This provides greater quality of quan- may cause noise in the detector. Single-piston pumps
titative data and enables automation of the system. are the least expensive and have a rotating eccentric
cam, which drives the plunger, discharging the liquid
Solvents through a one-way valve. Duplex pumps have two
plungers driven by a single motor by means of a
0015 The nature of the solvent will depend upon the mode
shared cam. This arrangement means that, while
of chromatographic separation employed, but a series
one of the plungers is in the discharge phase, the
of precautions that are common to all types of HPLC
other is in the intake phase, thereby superimposing
must be taken when preparing solvents. Because the
the two flow-rate profiles and considerably reducing
columns have frits at the ends to hold the packing in
pulsation. In this type of pump, delivery from the
place, the solvents must be devoid of particles and
solvent reservoir is steepled and changing solvents is
consequently must be filtered through membranes
rapid.
with a pore size of 0.5 mm or smaller, prior to use.
Bubble formation must also be avoided, since bubbles
Injectors
may cause variations in the flow rate if they reach the
pump or perturbations in the chromatogram if they Introducing the sample on to the column is one of the 0020

reach or form in the detector cell. Solvents must most critical steps in HPLC. Ideally, the sample
therefore be degassed by immersing the bottle con- should reach the column in the form of a tiny droplet
taining the solvent in an ultrasonic bath or by flushing that does not undergo diffusion, which would
the solvent with a stream of helium or nitrogen before broaden the chromatographic bandwidth and thus
delivering it to the chromatograph. A small stream of lower resolution.
helium is commonly bubbled through the solvent Several methods are employed to deliver the 0021

during the chromatographic procedure to prevent sample on to the column. In on-column injectors the
CHROMATOGRAPHY/High-performance Liquid Chromatography 1277

sample is introduced through a syringe which tra- The length of both conventional and microbore 0026

verses a septum and enables the desired quantity of columns is mainly limited by the pressure needed
sample to be deposited at the column inlet. In this to drive the solvent through the column, which is
method the mobile phase flows continuously through inversely proportional to the size of the particles
the column. The stop-flow injector is a variation of used in the packing. Minimum column diameter is
this type of injector. In this method the pump is limited by column-wall effects that cause solute mol-
stopped before the syringe is inserted in the injector, ecules flowing next to the wall to move more slowly
and injection is effected when column pressure has than those flowing through the center of the column,
dropped to atmospheric pressure. The advantage of resulting in an increase in chromatographic band-
these injectors is that they are inexpensive and of width.
simple construction, yet they do not lower efficiency. Open capillary columns similar to those employed 0027

However, reproducibility is poor, they are not suitable in gas chromatography are the most recent to have
for high working pressures, and they are complicated come into use. The column tube is made of glass and
to operate. is 20–50 mm in diameter and several meters long. The
0022 Valve injectors are most commonly used. In these stationary phase is chemically bonded to the wall of
injectors the sample is delivered on to a pressurized the tube. Capillary columns are also prepared by
column with no appreciable interruption in flow. The packing a tube with particles 5–30 mm in size and
sample is deposited by a syringe into an external loop, subsequently heating and drawing the tube to an
the valve selector is turned, and the mobile phase internal diameter of 50–125 mm.
passes through the loop on its way to the column. In addition to the column on which the separation 0028

The valve thus has two positions, a load position and is carried out, two other types of column are also uti-
an injection position. High injection reproducibility lized in HPLC to protect the analytical column. These
can be achieved with these injectors. The band- are precolumns and guard columns. Precolumns are
broadening effect is comparable to or somewhat placed between the pump and the injector to saturate
higher than that obtained using on-column syringe the mobile phase with the stationary phase and thus
injectors. The drawback afforded by these injectors prevent dissolution of the stationary phase in the
is that there is a split-second interruption in mobile- analytical column. Guard columns are placed be-
phase flow that may damage the column. To avoid tween the injector and the main column in order to
this problem, a bypass may be attached, such that retain components in the sample that might otherwise
there is always a constant flow of mobile phase become permanently adsorbed on the analytical
from the pump to the column. column, thereby affecting column efficiency and per-
0023 Automatic injectors that are capable of running meability. Both precolumns and guard columns are
analyses on up to 100 samples without operator at- normally made of a material similar to that used in
tention have been designed on the basis of the same the analytical column.
mechanism as valve injectors. There are also injectors Chromatography is customarily performed at am- 0029

equipped with a system of valves connecting them to bient temperature, but it may be necessary to regulate
several columns, which enables columns to be the column temperature or to carry out the separation
switched without stopping the flow. at a temperature other than room temperature.
In such cases thermostat-equipped compartments
Columns (ovens) and systems for heating or cooling the
0024 The columns most commonly utilized in HPLC con- columns are required.
sist of stainless-steel, plastic, or glass tubes, measuring
Detectors
15–25 cm in length and packed with small-diameter
particles (3–20 mm). Internal column diameter is In order to be suitable for use in HPLC, detectors 0030

normally between 2 and 5 mm. must meet a number of requirements. First and
0025 The use of small-internal-diameter (microbore) foremost, detector design must prevent broadening
columns has become increasingly important from of chromatographic bandwidth to insure that the
the early 1980s; they are similar to the columns de- separations achieved on the column do not deterior-
scribed above, but their internal diameter is between ate in the detector. In addition, response time must be
0.5 and 2 mm, while they range in length from 10 to short and the response must be linear over a suffi-
25 cm. Packing particle size normally ranges from 3 ciently broad range of concentrations.
to 5 mm. These columns are suitable for use when only The detectors most frequently employed are the 0031

small samples are available or when low solvent con- refractive index detector, the photometric detector,
sumption is required. and the fluorescence detector.
1278 CHROMATOGRAPHY/High-performance Liquid Chromatography

0032 Refractive index detectors measure the difference Selected Applications


between the refractive index of the mobile phase and
The use of HPLC in food analysis is growing daily, 0038
that of the column eluate. They are universal detect-
and it is now routinely applied in many laboratories.
ors that are highly sensitive to small changes in the
The many different types of columns and detectors
mobile phase and even to small variations in tempera-
that are currently available commercially make it
ture or pressure. This sensitivity means that to achieve
possible to apply HPLC in analyzing nearly all the
a suitable signal-to-noise ratio they are only capable
nonvolatile components in foods, be they present
of detecting solute concentrations in the order of
naturally or added artificially. The techniques
micromoles. In addition, they are unsuitable for
working with gradient conditions. employed for certain groups of food components are
summarized below by way of example.
0033 Photometric detectors measure absorbance in the
ultraviolet (UV) or visible regions of all the compon- Carbohydrates
ents in the column eluate. They are less universal than
refractive index detectors but by the same token are Nearly all chromatographic modes may be used in 0039

more specific. This type of detector is normally separating carbohydrates. For example, ion exchange
capable of detecting nanomoles, provided that the on strongly or weakly basic anionic resins or on
compound contains a strong chromophore. The three cationic resins, and partition chromatography on
types of photometric detector most frequently used ion exchange resins, on chemically bonded cyano,
are fixed-wavelength detectors, variable-wavelength amino, propyl-amino, or combined amino-cyano
detectors, and diode array detectors. This last type of phases, or on silica gel or gel permeation are all
detector is capable of performing complete spectral possible. Differential refractometry is the conven-
analysis of the column eluate on a continuous basis, tional detection system, although direct detection,
i.e., without stopping the flow. using short UV wavelengths or by forming derivatives
0034 Fluorimeter detectors are more specific and more detectable at longer wavelengths or fluorescent
sensitive than photometric detectors, but their linear derivatives, is also employed. (See Carbohydrates:
range is smaller. Detection limits are in the order of Determination.)
picomoles for suitably fluorescent compounds, and Acids
they are very useful in trace component analysis.
0035 Electrochemical detectors are also widely employed A variety of chromatographic modes are also applied 0040

in HPLC. These come in two types: amperometric to this group of components. Certain workers have
and conductometric detectors. Amperometric detect- employed ion exchange chromatography on strongly
ors are highly sensitive but are only applicable to acid cationic resins or strongly or weakly basic
analyses that can be oxidized or reduced; conducto- anionic resins. Reversed-phase chromatography and
metric detectors are moderately sensitive and are ion pair chromatography have also been used. Detec-
applicable for detecting anions and cations. This tion is carried out by refractometry, photometry using
type of detector is the detector most commonly used UV, or visible wavelengths, as in the case of carbohy-
in ion exchange chromatography. drates. (See Acids: Properties and Determination.)
0036 Derivatization of the components being analyzed Amino Acids and Amines
may sometimes be employed to increase the detection
limit or specificity. Most separations of amino acids and amines are per- 0041

0037 Mass spectrometry (MS) is being used more and formed using reversed-phase chromatography of dan-
more as an online detection system in HPLC. The use syl chloride, orthophthaldialdehyde (OPA), or phenyl
of HPLC-MS coupling has spread since 1980 due to dithioisocyanate derivatives. 9-Fluorenylmethyl
the improvements introduced in the different types of chloroformate (FMOC) is a suitable reagent for the
interface used. Several coupled MS-HPLC systems are analysis of secondary amino acids. Detection is
available commercially, and improvements are con- carried out by means of fluorescence or UV absorp-
tinually being introduced in existing equipment. At tion. The derivatizing, highly fluorescent, reagent
the same time, new equipment is appearing all the 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate
time. MS has been considered to be the ideal detector, (AQC) reacts rapidly and easily with both primary
since it furnishes information on component struct- and secondary amines and its use is currently spread-
ure. Microbore HPLC is particularly useful for online ing. (See Amines; Amino Acids: Determination.)
HPLC-MS, which requires low sample volumes. MS
Peptides
in combination with HPLC may become a standard
technique in a few years’ time. (See Mass Spectrom- Diverse techniques are used to separate peptides on 0042

etry: Principles and Instrumentation.) account of the broad range of molecular weights
CHROMATOGRAPHY/High-performance Liquid Chromatography 1279

of these components. Conventional reversed-phase ation of HPLC and MS analysis is commonly used for
columns are applied for peptides with molecular the sequence analysis of glycoproteins. Detection is
weights of less than 3 kDa, while reversed-phase carried out at 214 or 280 nm. (See Protein: Determin-
columns packed with particles with large pore sizes ation and Characterization.)
or size exclusion columns are used for larger peptides.
Other Compounds
Detection is performed at 214 nm. Figure 1 shows a
chromatogram of the peptides with molecular weights In addition to the major groups of food components 0044

higher and lower than 700 from a sparkling wine, as an mentioned above, HPLC has found application in
example of this type of analysis. (See Peptides.) many other areas of food analysis, and details
may be found in the relevant articles for the follow-
Proteins
ing compounds or groups of compounds: lipid
0043 Practically all known modes have been used in components, phospholipids, triglycerides, vitamins,
the separation of proteins, e.g., separations based colors, pesticides, drug residues, polycyclic aromatic
on molecule size (gel permeation chromatography), hydrocarbons, and nitrosamines. (See Colorants
on charge (ion exchange chromatography), on (Colourants): Properties and Determination of
hydrophobicity (reversed-phase chromatography Natural Pigments; Properties and Determinants
and hydrophobic–interaction chromatography), and of Synthetic Pigments; Fatty Acids: Analysis; Nitro-
even combinations of these mechanisms. The combin- samines; Pesticides and Herbicides: Types, Uses,
and Determination of Herbicides; Phospholipids:
Determination; Polycyclic Aromatic Hydrocarbons;
0.15 Triglycerides: Characterization and Determination;
Vitamins: Determination.)
Absorbance at 214 nm

0.10
Related Techniques
FPLC designates a fast chromatographic method, de- 0045

veloped by Pharmacia, which is similar to HPLC and


yields high resolution. Since FPLC needs only a rela-
0.05
tively low backpressure to drive the high flow rates at
which the separations are performed, the risk of de-
naturation caused by shearing forces is reduced.
0.00
Moreover, the mechanical components are resistant
0 10 20 30 40 50 to corrosive buffers, and there is no contamination or
(a) Time (min) inactivation or the components of interest.
Given the range of columns available in the market, 0046
0.15
a variety of separation modes can be applied using
this technique: size exclusion, hydrophobic inter-
action, chromatofocusing, ion exchange, and re-
Absorbance at 214 nm

0.10
versed-phase chromatography.
This method was developed to separate and purify 0047

biomolecules and is very useful in separating isoen-


zymes and molecular species with similar charge char-
0.05 acteristics. It is also used to distinguish between
different types of meat or grains.
SFC is another technique, related to HPLC, which 0048

uses as the mobile phase a supercritical fluid, i.e., a


0.00
0 10 20 30 40 50
fluid at a pressure and temperature above the critical
(b) Time (min) point. The properties of supercritical fluids are inter-
mediate between those of gases and those of liquid.
fig0001 Figure 1 Chromatogram of peptides with molecular weights (a) Thanks to their higher diffusivity and lower viscosity
higher and (b) lower than 700 Da from a sparkling wine. Column: as compared to liquids, high efficiencies are achiev-
Nova Pak C18. Solvents: (a) TFA/water; (b) TFA/acetonitrile.
able with shorter analysis times than those customar-
Gradient elution. Detection at 214 nm. Reproduced from Moreno-
Arribas MV, Bartolomé B, Pueyo E and Polo MC (1998) Isolation ily employed using HPLC.
and characterization of individual peptides from wine. Journal of The basic advantage of SFC with respect to gas 0049

Agriculture and Food Chemistry 46: 3422–3425, with permission. chromatography (GC) is the possibility of analyzing
1280 CHROMATOGRAPHY/Gas Chromatography

components that span a broad range of volatilities as Dabrio MV, Blanch GP, Cifuentes A et al. (2000) Cromato-
well as heat-labile components. At the same time, grafia y Electroforesis en Columna. Barcelona: Springer-
SFC is also compatible with many of the detectors Verlag Ibérica.
commonly used in GC or HPLC and SFC-MS Gonzalez-Llano D, Polo C and Ramos M (1990) Update on
HPLC and FPLC analysis of nitrogen compounds in
coupling is easy to carry out.
dairy products. Lait 70: 255–277.
0050 The critical temperature of the supercritical fluids
Gruewedel D and Whitaker JR (1987) Food Analysis. Prin-
employed as the mobile phases between 0  C and ciples and Techniques. New York: Marcel Dekker.
200  C and the critical pressure of these fluids should Henschen A, Hupe KP, Lottspeich F and Voelter W (1985)
not be too high. Carbon dioxide, nitrous oxide, High Performance Liquid Chromatography in Bio-
alkanes (such as n-pentane), and xenon, all of which chemistry. Weinheim: VCH Press.
are nonpolar, are most often used. Ammonia can be Macrae R (1987) HPLC in Food Analysis, 2nd edn.
used to elute polar solutes, but mixtures of phases, London: Academic Press.
e.g., a nonpolar mobile phase containing a small Moreno-Arribas MV, Bartolomé B, Pueyo E and Polo MC
quantity of a polar organic solvent, known as a (1998) Isolation and characterization of individual
modifier are normally employed. peptides from wine. Journal of Agriculture and Food
Chemistry 46: 3422–3425.
0051 SFC can be performed on capillary, packed, and
Oliver RWA (1989) HPLC of Macromolecules. A Practical
micropacked columns. Stationary phases should be Approach. Oxford: IRL Press.
cross-linked; otherwise the supercritical fluids, Shaw LP (1988) Handbook of Sugar Separations in Foods
which are excellent solvents for polymers, could by HPLC. Boca Raton. FL: CRC Press.
extract the stationary phase. Enantiomers can be Smith RM (1988) Supercritical Fluid Chromatography.
resolved using chiral phases. RSC Chromatography Monographs. Cambridge: Royal
0052 The equipment used in SFC is similar to that used Society of Chemistry.
in HPLC and basically consists of a high-pressure
syringe pump, an injector, and a restrictor or post-
column valve to keep the mobile phase in a supercrit-
ical condition inside the chromatographic column.
0053 Fluid density is commonly programmed to adjust
mobile-phase selectivity, in as much as the physico-
chemical properties of supercritical fluids (solvation Gas Chromatography
strength, viscosity, diffusion) are all dependent upon I Martı́nez-Castro, J Sanz and M V Dabrio,
density. Instituto de Quı́mica Orgánica General (CSIC), Madrid,
0054 This method has been applied in food analysis (oils, Spain
cheeses, coffee, etc.) Some of the most interesting Copyright 2003, Elsevier Science Ltd. All Rights Reserved.
applications include separations of acids, alcohols,
lipids, carbohydrates, vitamins, and terpenes. (See
Acids: Properties and Determination; Alcohol: Prop-
erties and Determination; Analysis of Food; Carbo- Background
hydrates: Determination; Coffee: Analysis of Coffee
Gas chromatography was born in 1951, when James 0001
Products; Vitamins: Determination.)
and Martin developed a new separation technique in
order to analyze a mixture of 17 fatty acids. This
See also: Acids: Properties and Determination; Alcohol:
technique includes different chromatographic modes
Properties and Determination; Amino Acids:
Determination; Chromatography: Principles; Coffee: that use a gas as the mobile phase: the separation
Analysis of Coffee Products; Colorants (Colourants): process takes place in a chromatographic column.
Properties and Determination of Natural Pigments; Like other separation techniques, gas chromatog-
Properties and Determinants of Synthetic Pigments; Fatty raphy (GC) can be employed on the preparative
Acids: Analysis; Nitrosamines; Peptides; scale, but more frequently, it is used as a powerful
Phospholipids: Determination; Polycyclic Aromatic analytical technique.
Hydrocarbons; Protein: Determination and
Characterization; Triglycerides: Characterization and
Determination; Vitamins: Determination Instrument
The analytical instrument consists of a mobile phase 0002
Further Reading supply (usually a gas cylinder) regulated by a con-
Charalambous G (1984) Analysis of Foods and Beverages. trol system, a sample introduction system (usually
Modern Techniques. Orlando: Academic Press. called injector), a thermostatically controlled oven

You might also like