SERVICE MANUAL

PHYSIOMED-Expert 2
PHYSIOMED–IF–Expert 2

Revision: 09 / 2008
Valid from Instrument No.: PME 2 - 0300211
And PIE 2 - xx001xx 17900
Table of Contents
1 TECHNICAL DATA .................................................................................................3

2 HOUSING CONSTRUCTION...................................................................................4

3 GENERAL NOTES ..................................................................................................4

4 NOTE .....................................................................................................................4

5 ADDITIONAL EQUIPMENT REQUIRED..................................................................4

6 TEST MENU ...........................................................................................................5

6.1 Test Menu Stimulation Current ................................................................................................5

7 BALANCING PROCEDURE .....................................................................................6

7.1 Setting up ...................................................................................................................................6

7.2 Checking the Power Supply; PC Board No. 9902-2..................................................................7


7.2.1 Not Stabilized.......................................................................................................................7
7.2.2 Stabilized .............................................................................................................................7

7.3 Balancing PC-Board No.: 9902-2 (Stimulation Current)...........................................................7


7.3.1 Balancing I: Offset Set Up of Output Transformer.................................................................7
7.3.2 Balancing II: Offset Set Up of Pulse Generation ....................................................................7
7.3.3 Adjustment of Output Current .............................................................................................8

8 CHECKING THE INSTRUMENT ..............................................................................9

8.1 Checking Connection .................................................................................................................9

8.2 Checking Display and Display Elements ..................................................................................9

8.3 Checking the Therapy Timer ...................................................................................................10

8.4 Checking the Current Types....................................................................................................10

8.5 Checking Function of the Selection Fields .............................................................................10

8.6 Display Wire of Air Gap and Leakage Distance.....................................................................10

9 ERROR CODES.....................................................................................................11

9.1 Important Service Note ...........................................................................................................12


Technical Data

1 Technical Data

Protection class I acc. To VDE 0750 / IEC 601


Typ: BF

CE characterization acc. to Council Directive concerning medical devices


(93/42 EEC)

Class acc. EEC 93/42 IIa

Power supply 230 VAC ± 10% or 115 VAC ± 10%


Note Settings indicated at the mains module!

Power line frequency 50 … 60 Hz

Current consumption 0.3 A (at 230 V) or 0.6 A (at 115 V)

Power consumption 68 VA

Line fuses at 230 V: T1A


at 115 V: T2A

Output data 75 mAs (maximum) at 500 Ohm


STIMULATION

Ambient temperature + 10°C ... + 40°C

Dimensions 34.5 x 13.3 x 34.8 cm


(W x H x D)

Weight 5.1 kg

PHYSIOMED-Expert 2 - Service Manual 3


Housing Construction

2 Housing Construction
PHYSIOMED-Expert 2 / PHYSIOMED-IF-Expert 2 is a single shell construction in a plastic
housing.
Always disconnect the power lead before opening the instrument.

3 General Notes
PHYSIOMED-Expert 2 / PHYSIOMED-IF-Expert 2 is a two-channel stimulation current therapy
unit. You can apply the whole range of current modes from low- to medium-frequency
including the classic interferential current. The instrument also offers you programs for
diagnosis by stimulation current. The unit can also be used for simultaneous therapy
(stimulation current and ultrasound in ONE treatment if you combine it with a suitable
ultrasound therapy unit (e.g. PHYSIOSON-Expert).

4 Note
The structure of the following service instructions allows you to do a basic balancing.
As a rule, for later balancing (e.g. repair), only parts of the balancing procedure are
necessary, which can all be adjusted by the trimming potentiometer (provided that
grave errors concerning the balancing procedure are not apparent or have been
cleared).

5 Additional Equipment Required


State of calibration *
Oscilloscope (with probe) B
Digital multimeter AC and DC B
Frequency meter B

* State of calibration B = precision 5%

The following documents provide information necessary for the balancing


procedure, such as positions of the trimmers or test points:
• Wiring Scheme
• Parts List
• Operating Instructions

4 PHYSIOMED-Expert 2 - Service Manual


Test Menu

6 Test Menu
The test menu is used to select and display service relevant parameters of the
instrument. In order to get into the test menu you have to set short-circuit plug JP2 at
Pc board 9902-2 on position 1+2 and then turn on the instrument.

6.1 Test Menu Stimulation Current

Selection Channel 1

Selection Channel 2

Display Circuit: Set Value I

Value measured output I


Value measured output U

PHYSIOMED-Expert 2 - Service Manual 5


Balancing Procedure
You will find a similar layout of the display for circuit II (ch2). By using the following
overview table you can get information for possible adjustable parameters and displays:

Selection Field Circuit I Available Options


( ) + Positive signal / Negative signal
( ) CH1 NF NF/ MF / MC basic signals
( ) monph Monophasic signal / Biphasic signal
( ) kb_trap1 Signal waveform
( ) T Impulse duration of the signal
( ) R Pause duration of the signal
Selection Field Circuit II Available Options
( ) + Positive signal / Negative signal
( ) monph (unused; CH1 also switches CH2)
( ) CH2 f0 f0 / AMF / NF / NF_HV / NF_2 basic signals
( ) f0 4 KHz Adjusting of basic frequency
( ) Δ f 100Hz Adjusting of modulation frequency
Lower Selection Field Available Options
( ) DA1_255 Adjust circuit I by means of the DA converter
( ) DA2_255 Adjust circuit II by means of the DA converter
[ ] EEPRCL EPROM clear (e.g. clearing all individual programs)
[ ] CV1_ON Activate constant voltage
[ ] CV2_ON Activate constant voltage
[ ] R_ext Switch signal to external lead resistor
[ ] Amp2_ON Activate output transistor in circuit 2
[ ] Pl Ù Vak Switching between plate electrodes – vacuum
electrodes
[ ] US-TestMen Selection of the ultrasound test menu (unused)
[ ] Test END Exit test menu

7 Balancing Procedure

7.1 Setting up
• Check line fuses; refer to technical data and reference labels.
• Check fuses F1-F4 of secondary circuit; refer to parts list of power supply and to
reference labels.
• Connect power cord

6 PHYSIOMED-Expert 2 - Service Manual


Balancing Procedure

7.2 Checking the Power Supply; PC Board No. 9902-2


7.2.1 Not Stabilized
Measure +7.8 V at TP5 Nominal Value: +6.6 V...+9.0 V
Measure voltage towards GND at TP6.
Measure +34 V at TP4 Nominal Value: +30 V…+38 V
Measure voltage towards –15 V at TP3.
Measure +125 V at R119 Nominal Value: +110 V ... +140 V
Measure – 125 V at R114 Nominal Value: - 110 V ... – 140 V
Measure voltage towards GND at TP10.

7.2.2 Stabilized
Measure + 5 V at TP1, Nominal Value: + 4.9 V...+ 5.2 V
Measure +15 V at TP2, Nominal Value: +14.8 V...+15.2 V
Measure – 15 V at TP3,Nominal Value: -14.8 V...- 15.2 V
Measure voltage towards GND at TP10.

7.3 Balancing PC-Board No.: 9902-2 (Stimulation Current)


7.3.1 Balancing I: Offset Set Up of Output Transformer
Preparation:
• JP2 to position 1+2 : (Test Mode)
• Turn intensity to zero
• Adjust earth connection of the oscilloscope at TP10 (GND)
• Switch on the instrument
• Measure offset-voltage at housing of V33
• Do following adjustments in the stimulation current test menu
• kb_trap1 adjust signal waveform
• R_ext select external resistance

Adjust offset-voltage to zero by using pad-trimmer R104


When selecting CV1_ON in the stimulation current test menu, the offset voltage
should not change. If it does, use pad-trimmer R104 to adjust.

7.3.2 Balancing II: Offset Set Up of Pulse Generation


Preparation:
• CV1_ON not activated
• Turn intensity to maximum
• Adjust earth connection of the oscilloscope at TP10 (0 V)
• Measure offset-voltage at housing of V33

Adjust offset-voltage to zero by using pad trimmer R53

PHYSIOMED-Expert 2 - Service Manual 7


Balancing Procedure

7.3.3 Adjustment of Output Current


Preparation :
• CV1_ON : not activated
• kb_trap1 : adjust signal waveform
• R_ext . external resistance activated
Plug in the external test resistance R = 500 Ohm

Channel 1:
Setting up low frequency output current
CH 1 NF is set in the menu channel 1
Turn intensity to maximum
The low frequency output current of circuit 1 is to be adjusted by trimmer R60

The value appearing to the right of the intensity display in circuit 1 must be 74 … 75
mA. The maximum value of the measured current must be between
US,min = 33.8 V and US,max = 37.5 V.

Channel 2:
Setting up interference output current
CH 2 fo is set in the menu channel 2
Turn intensity to maximum
The Interference output current of circuit 2 is to be adjusted by trimmer R72.

The value appearing to right of the intensity display in circuit 2 must be 74 … 75 mA


The maximum value of the measured current must be between
US,min =33.8 V and US,max =37.5 V.

Setting up low frequency output current


CH 2 NF is set in the menu channel 2
Turn intensity to maximum
The low frequency output current of circuit 2 is to be adjusted by trimmer R252.
The value appearing to right of the intensity display in circuit 2 must be 74 … 75 mA
The maximum value of the measured current must be between
US,min =33.8 V and US,max =37.5 V.

The value appearing right to intensity-display in circuit 2 has to be 74 … 75mA


Maximum value of the measured currency has to be between US,min = 33,8V and US,max
= 37,5V

Setting up frequency of special current modes (e.g. KOTS)


CH 2 NF_2 is set in the menu channel 2
Turn intensity to maximum
The special current mode of circuit 2 is to be adjusted by trimmer R251.
The value appearing to the right of the intensity display in circuit 2 must be 67 mA
To return to regular operation mode, set JP2 to position 2+3.

8 PHYSIOMED-Expert 2 - Service Manual


Checking the Instrument
Instrument self-test:
(in the selection stimulation current test menu)
Select the [ ] Test END
The instrument will carry out an automatic self-test routine

9902-2
CH1 NF/ IF R 60

CH2 NF R252
CH2 NF2 R 251 ( +40/-40 )
CH2 f0 R 72

Tolerance of selftest values

Stimulation current 10 + 2 50 +5 70 +5 80 78 - 85
mA:

The values displayed here are important assessment criteria for our service department.
The assessment of the parameters enables us to make a diagnosis of the technical state
of the device.
If the measured values are not within tolerance, the automatic self-test is interrupted
and an error code is displayed. It can be restarted by pressing the selection-button. If
settings are done, the self-test routine will end with the message:
CH1+CH2 HWT OK
To return to THERAPY MODE, set JP2 back to position 2 + 3.

8 Checking the Instrument

8.1 Checking Connection


Check the whether line voltage of the device (see “red window“ in the mains module)
and the rated voltage correspond.

8.2 Checking Display and Display Elements


Turn both intensity controls to 0.
Switch off the device and switch it on after 5 sec.
The instrument will carry out an automatic self-test routine.
• Display is tested
• All functions and output values of the stimulation current unit are checked
• Pulse indicators are illuminated
• Patient current and output indicators are illuminated
• An acoustic signal is audible
Either the start menu or the display of the last selected current type appears.

PHYSIOMED-Expert 2 - Service Manual 9


Checking the Instrument

8.3 Checking the Therapy Timer


• Connect oscilloscope to the output
• Select any current type
By using the Therapy timer you can adjust time from 0-30 min in 1 min. intervals.

• Adjust a therapy time of 30:00 min.


• Turn on intensity
Timer is started, therapy time decreases.

Turn therapy time to 0.


An acoustic signal is audible, intensity is decreasing and Intens Å appears in the lower
status line.

8.4 Checking the Current Types


Connect oscilloscope to output.
Check all current types, pulse and rest times according to operating instructions.

8.5 Checking Function of the Selection Fields


Check all selection fields according to operating instructions.

8.6 Display Wire of Air Gap and Leakage Distance


Check air gap and leakage distance (2.5 mm / 4 mm); you should also pay attention
to the position of the display wire.

10 PHYSIOMED-Expert 2 - Service Manual


Error Codes

9 Error Codes
Self Test

ERROR 0mA CH1 Operating point difference current circuit 1 at 0 mA


Reason: control power supply and F1 ( 160mA),
flat ribbon cable main-/frontboard, V25/24
control function of K4
ERROR 10mA CH1 Operating point difference current circuit 1 at 10 mA
Reason: control power supply and F1 ( 160mA),
flat ribbon cable main-/frontboard , V25/24, R113
ERROR 50mA CH1 Operating point difference current circuit 1 at 50 mA
Reason: control power supply ,
V 48 and CC/ CV function of N9 (81C55)
flat ribbon cable main-/frontboard
ERROR 70mA CH1 Operating point difference circuit 1 at 70 mA
Reason: control power supply, flat ribbon cable main-
/frontboard, V36,V44/45
I80_HARD_ERROR Current circuit 1 > 80 mA Hardware triggers
Reason: control power supply ( +15DCV , out of tolerance) V10,
N3, flat ribbon cable main-/front board
I80_SOFT_ERROR Current circuit 1 > 80 mA Software triggered
Reason: control power supply ( +15DCV , out of tolerance) V10,
N3, flat ribbon cable main-/front board
ERROR 0mA CH2 Operating point difference current circuit 2 at 0 mA
Reason: control power supply
flat ribbon cable main-/front board
ERROR 10mA CH2 Operating point difference current circuit 2 at 10 mA
Reason: control power supply
flat ribbon cable main-/front board
ERROR 50mA CH2 Operating point difference current circuit 2 at 50 mA
Reason: control power supply
flat ribbon cable main-/front board
ERROR 70mA CH2 Operating point difference circuit 2 at 70 mA
Reason: control power supply
flat ribbon cable main-/front board
I80_HARD_ERROR CH2 Current circuit 2 > 80 mA Hardware triggers
Reason: control power supply ( +15DCV , out of tolerance) V10,
N3, flat ribbon cable main-/front board
I80_SOFT_ERROR CH2 Current circuit 2 > 80 mA Software triggered
Reason: control power supply ( +15DCV , out of tolerance) V10,
N3, flat ribbon cable main-/front board

PHYSIOMED-Expert 2 - Service Manual 11


Error Codes

Therapy

ERROR uP Watchdog Watchdog-error


Reason: replace μP Processor
ERROR T2 Watchdog Timer2 Watchdog-error
Reason: replace μP Processor
ERROR T/R Impulse-/resting-time error
Reason: flat ribbon cable main-/front board
KFG Eprom N11,
ERROR CH1 I 20% Operating point difference circuit 1
Reason: flat ribbon cable main-/front board, dose potentiometer
mechanically damaged, faulty balance
ERROR CH2 I 20% Operating point difference circuit 2
Reason: flat ribbon cable main-/front board, dose potentiometer
mechanically damaged, faulty balance
ERROR CH1 Current circuit 1 > 0, although control 1 is at zero position
INTENS.ZERO Reason: flat ribbon cable main-/front board ,faulty offset
balance, dose potentiometer mechanically damaged
ERROR CH2 Current circuit 2 > 0, although control 2 is at zero position
INTENS.ZERO Reason: flat ribbon cable main-/front board, faulty offset
balance, dose potentiometer mechanically damaged
I80_HARD_ERROR Current circuit 1 > 80 mA Hardware triggers
Reason: control power supply ( +15DCV, out of tolerance ) V10,
N3
I80_SOFT_ERROR Current circuit 1 > 80 mA Software triggered
Reason: control power supply ( +15DCV , out of tolerance ) V10,
N3
I80_HARD_ERROR CH2 Current circuit 2 > 80 mA Hardware triggers
Reason: - control power supply ( +15DCV , out of tolerance)
V10, N3
I80_SOFT_ERROR CH2 Current circuit 2 > 80 mA Software triggered
Reason: - control power supply ( +15DCV , out of tolerance)
V10, N3
CH2 NF ERROR NF input voltage for circuit 2 is too high (R72)

9.1 Important Service Note

In case of service and updating the system please note:

Adapter PCB 0607-1 2CH


is the substitute of IC N23 ML2035 and N24 ML2035

Software Versions:

ISE2 / PME2 xx.x1 only to use for N23 ML2035 and N24 ML2035
(picture 1) at 9902-2 Rev. 1.10

12 PHYSIOMED-Expert 2 - Service Manual


Error Codes

ISE2 / PME2 xx.x only to use for adapter PCB 0607-1


(picture 2) at 9902-2 Rev. 1.10

This is valid from the SW Ver. 12.3 from May 2006.

Picture 1

Picture 2

PHYSIOMED-Expert 2 - Service Manual 13


Spare Parts List - PHYSIOMED ELEKTROMEDIZIN AG Date: 30.11.2005 12 Page: 1

part number / O-No description / ND step/cost perpos.no. item notice quantity

p.ID: 00091 var./sw: 1/0-0-0 desc 1/2:PHYSIOMED-Expert Reizstromgerät


O#: 00091 desc 3/4:
ND: PHYSIOMED EXPERT desc 5/6:
Ersteller:Legl desc 7/8:

50076 [¶] acrylic plastic front plate 1 1 1 1,0000 Stk


PHYSIOMED-Expert
50078 [»] 9805-2 cpl. 1 1 2 1,0000 Stk
front pc-board
51311 [¶] 9902-2-Exp full set a. checked 1 1 3 1,0000 Stck
main pc-board Rev 1.10
50046 [»] 9806-1 full set 1 1 4 1,0000 Stk
inverter pc-board
51200 [¶] Display 1 1 5 1,0000 Stck
SP14N001
51199 [¶] foil cable 1 1000 6 1,0000 Stck
FFC1.00A26/0240L5.0-5.0-10.0-1
50011 [¶] Control button/ transmitter 1 1 7 1,0000 Stk
CESI 16 / LS-PM
50015 [¶] fuse 1 100 8 2,0000 Stk
T1A
50020 [¶] select buttom small 24mm 1 1 9 2,0000 Stk
A1024069
50022 [¶] lable scale RAL 5021 1 100 10 2,0000 Stk
A 1105011
50023 [¶] select buttom large RAL 5021 1 1 11 1,0000 Stk
102001
50021 [¶] button top cover d=24mm 1 1 12 1,0000 Stk
666215
50560 [¶] flat ribbon cable 1 100 13 1,0000 Stk
Typ 7-2 "F" cpl.
50535 [¶] Net input module 1 1 14 1,0000 Stk
PSOS XD H3A
operating unit monitoring hardware

measurement and
control unit

interface
sono unit stimulation current
amplifier circuit 1

interface
interface
vacuum unit monitoring hardware stimulation current
electrodes

measurement and
control unit

pulse
stimulation current
interface generation
amplifier circuit 2
manual release key

interface
power supply unit
power supply microprocessor

Datum Name

Bearb. 27.02.06 Nie Block Diagram


Gepr. PHYSIOMED - Expert

PHYSIOMED Platine PC-Board


ELEKTROMEDIZIN AG
Hutweide 10
Blatt
91220 Schnaittach /
Laipersdorf /Germany 1
Rev Aenderung Datum Name File bd-gb-pme2
99-000045-2 REV 1.10
C46

1/1

be9902-2_rev_1_10
Blatt
C100 C97 R77 C181 C182 C187 C188 R253 R254
R252
X4 C98 R80 C61
C62
C86 C50
X1
N53 R136 N54 N55
C45 R247 N13
C189 R250 R251 R74 R35
C183 C186 R249
N33 N32

Component Placement

PHYSIOMED-Expert 2
C96 R65 C82 C185 R45 R48

9902-2 Rev. 1.10


TP9 R72
C87
R81 C146 R170 C139 R164 R219 C190

IONOSON-Expert 2
R66 R71 R69
R70
X9 X13 V15 C99 R78 R85
R172 R171 R166 R165 C88

Bestückungsplan
N29

R79 R55 C66 N56 R82


R229 R221 R54
C128 R91 C94
N15

Platine PC-Board
R230
R150

Hauptplatine
N44 N43 C191
R259 R256 R86
N30

Main Board
R231 R139 C52
C65 C92 C85
R232 R142
V45 V44
C148 C142 R56 N14 C93 C89 N11
R149
R233 R141 R174 R173 R222 R169 C140 R57 C53 R255 C73 R73
C194
R234 R124 V43 R175 C147 R168 R167 R220 R49 C78 R90 C91 R75
R258 R225
R191

File
R235 R147 C77 C57
N40

R140 R135 R260 R226 R228 R76 R43 R44 R42 R46 C51
V57 N16 C75
R236 C131 R185 R60 R257
V59 C125 N27 C42 R47
R237 R137 C84 C90
C130 C150
N8

R184 V42 C72


R182 R53 N26

ELEKTROMEDIZIN
R238 V48 R138
C129 R188
C151 C149

Name
R239
C74 R227
R146 V56 R134 R50 C76
R240 R145 V46
1

R59

Nie
V34 V35 V39 C59
V55 C58

Laipersdorf /Germany
R241 R144 R183 R133 C126 R67 R68 R223
C153 N45 R58

91220 Schnaittach /
V47
N22

R143 V60 V38 C81 N17 R224


V66
C157 R187 V53 R126 R129
C171
N9

02.09.03
R261
R264
R265

Datum
R190 N39 V40 C44
R176 C80
R179 R180 R127

Hutweide 10
R198 C154 V65 V54 C83
R201

Physiomed
R178 C152 R181 R128 C124 C60
R199 C155 V64 C63 R51 N25
C169 R125
N21

R177 V36 C127 R132


R197 C168 C79 R52
N41

R155

Bearb.
C54

Gepr.
N46 C170 V37 C64 R64 C172
V63 R156 R130 R131
V61
1

V62 R154 R63


R159 R242 N18

Name
N5

T1
4N35
R206
R207

R192 R193 C156 R195 N12


R152 C49 C48
10

X5 R194 R196 C55 C68

Grö
1

R189 R200 C47


V88
2 1 N10 C40
R269 R271 R41 V13
R153 R40
ML2035
N24
N19

C192 6
C193 V90 R268 V71

Datum
L7 R215
C163

R273 AU-04
C134 R83
V58 N20 V72 R262 -5V C56
R274 3 C159
R214
L30 V70
N28
R204
C160 R205
R87
C71 C69
V49 V87 R160 5 OPA548T C133 C158 C67
R162
N47

R216

ML2035
N23
R163 C137 V68 C135
C162 V14

R61
R212
6
5

V83 R161 C136 R210

02.09.2003
R211 C161 N6
V84 V52 V51 V67 V69 R213 R88
C184 C70
R151 TP7
R208
R209
R248 R267 K4 C37 C41
K5

N48
K2 V12 R33 C112 C110 JP2
R203

C132 R263
R39 R38 R31
TP8
N7 C43
R186
R32

Ä
6 OK1 C109 C39 C36 C111

g
ru
se
d
n
C108
C106
C38 R34 C113 C32 F2
1 R115

1.10
C107
V33 R62

Rev
V92
R272
C104 C33
R123
R122

C105

E
V91
K1

L5
C102 C13
C31

L4
C103 F3
R158 V74 F4 F1
R157 R217 V73 C180 N4 R84
C30 BR1 BR3 R10 C21
R148
V50 V79 V41 V6

R9
N2 +

N31
C164
1

V9
BR2 BR4
V85 V80 V81 V32 C14 C16
MC34063AP1
C178 C35 C34 R37 R36 C179 C1
X12 C95
C7 C15
V7

B
R12 C22
C176

L6

N51 N50
KK3 C5 R13 R11 +

+
IL420 IL420 C8 C6
L2
V89
V82

C17

R112
C141 C173
V78 Q1 V8
R244 R245 R3

E
C175 14 N57 R7 R6

R14
V4 C9
V30

V11
R202 1
C24 C3
R118
C177
R246 R243 C18
X6 V2
R111 R109 V1

C120
C138 N1
C174
V28 V29 C23
L1

R116
R117
IL420 IL420 C145 R121 R107
TP10
MC34063AP1 V3 C25
L13 L12
L10 N52 N49 L8 C143 V77 R105 C26 R18

4N35
N36
N3

R1
R4 R5 C10
V23
C167

C166

L9 N38

R8
C165 C121 V31 V21 R99 V19 R96
L3 R17 C20

-
C2

V5
MC34063AP1
R218 C118 C117 C144 KK2
R97 C11
C114 R98
L11
V24 R30 R27 R23 R19 V10 C27
R108
V26 V18 C19

4N35
N37
V75 C123 V25 JP1 R28 R25 R21

N35
R102 C116 V27 R101 V20 C12 R15
V17
C119 R29 R89 R26 R22 C29
1

C122 R16
C115 R100 R95
R104 V22 R24 R20

R114
R120
R113
V76 C28
R103 R94
1

TP4

TP3

0V

TP1

TP2
R92

4N35
X8

R119
R106
R110

N34
1 X7 X2 1 R93
1 R2
1 X11 X3 V16 C101 C4
X10 K3 TP6 TP5
43,2 C6
C5

C1
N1
TL082
3H C2

R9

R2
R8
R7
V3

R1
R6
C7

2H
1H

C3

R25
R24
R21
R5
R4

V2
V1

N2

V6

R10
R17

R11
R18

R20
TL082

R19
5H 4H
N4

V5
V4
R3
C8

R15
R16
R14
R13
R12

N3
X4 X3 1
C4
X1 X2

130,8

Soldered side:
Lötseite:

H1, H2, H3, H4, H5


X3, X4

Datum Name
Component Placement
Bearb. 13.11.03 Nie Front Board

Soldered side
Lötseite
Component side
Bestückungsseite
Gepr. Bestückungsplan
Frontplatine
Physiomed ELEKTROMEDIZIN
Platine PC-Board

X3, X4 Hutweide 10
91220 Schnaittach / 9805-2 Rev 1.10 Blatt
Laipersdorf /Germany 1
Rev s
g
ru
e
d
n
Ä Datum Name File Be9805-2_rev_1_10
X1
1
C1
C2

F1
V1

Soldered side
C3 Lötseite

E1811

Datum Name Component Placement


09.01.06 Nie Inverter Board
Bestückungsplan
Inverterplatine
Physiomed Elektromedizin Platine
Hutweide 10
91220 Schnaittach /
9806-3 1
Rev Datum Name Laipersdorf / Germany File Be9806-3.brd

You might also like