BI Series

Weighing Indicator

OWNER’S MANUAL
CONTENTS
INTRODUCTION 4
CAUTIONS 4
FEATURES & MAIN FUNCTIONS 4
SPECIFICATIONS 5
PART NAME AND FUNCTION 5
SYSTEM MODE 7
GENERAL FUNCTION & DESCRIPTION 13
CONVERSION MODE 15
RS-232C COMMUNICATION 15
INSTALLING 18
USING & KEEPING 19
BATTERY 19
TEST FUNCTION 21
CALIBRATION MODE 23
SEALING & OPTIONAL DRAWING 26
ERROR MESSAGE DESCRIPTION
AND MANAGEMENT 27
INTRODUCTION
We greatly appreciate for you to purchase BI-SERIES of CAS. This
goods has the excellent performance and splendid properties through strict
test under severe quality management.
It is recommended to read this manual in full before using BI-SERIES
for good function application.

CAUTIONS!
Do not press the key strongly.
Do not use the inflammables at cleaning.
Do not install in the place with sudden temperature change.
Do not install in the place with high voltage and terrible electrical noises.
Do not use in the place exposed directly to the sun and including much
dust.
Do not use in the place with excessive electrical noises and shaking.
Charge a battery in full to use after negligence for a long time.
Install the platform on a flat place.

FEATURES
Simple outdoor using in use of battery
Using general (Mn, Alkaline) battery or rechargeable battery
Easy operation and various functions
SPAC TM Single pass automatic span calibration
Waterproof(IP65)
High precise resolution
Liquid crystal display of 5 place weight (LCD)
Using long-time battery
Various options (Wall mount, Fixed bracket, RS232)

Main functions
Weight averaging function(Hold function)
COUNT function (Unit weight setting, unit weight setting by sample)
Low/High weight comparing
Auto power OFF

4
SPECIFICATION
MODEL BI-SERIES
ITEM BI-100D BI-100R BI-100DB BI-100RB
DC1.2V(6EA)
DC1.2V(6EA) DC1.5V(6EA) Recharging
DC1.5V(6EA) Recharging Mn battery battery
OPTIONAL
Mn battery battery DC12V adapter
DC12V adapter
BACK LIGHT
LOADCELL PERMISSION
5V DC
VOLTAGE
TEMPERATURE RANGE 10 C 40 C
ZERO CALIBRATION RANGE 0.05mV 5mV
REZERO RANGE 2% of Max. Capacity
INITIAL ZERO RANGE 20% of Max. Capacity
INPUT SENSITIVITY 1 V/D more
A/D INTERNAL RESOLUTION 1/100,000 (MAX.)
A/D EXTERNAL RESOLUTION 1/10,000 (MAX.)
MAX. TARE RANGE Maximum capacity
WEIGHT DISPLAY TYPE 5 place LCD(Char. size 23.5mm)

1.Stable ( ) 2. High, Normal, Low (HI, OK, LO)


CHARACTER DISPLAY 3.Battery discharge (BAT) 4. quantity unit (PCS)
5.Weight unit (kg, lb)
CONSUMPTION Approx. 0.15W

Part name and function


Part name
Adapter connector
Lcd Display Loadcell

Contact output, RS232C

key Top cover Bottom Cover

5
Display part

HI
pcs
POWER
OK lb
kg
CHARGE
LO BAT

ZERO GROSS NET TARE HOLD

ZERO lamp : ON when the current weight is°∞0°±.


TARE lamp : ON when the TARE weight is saved.
GROSS weight lamp : ON at displaying the total weight.
NETHI weight lamp : ON at displaying the pure weight.
pcs
Hold
HI lamp : ON at weighing the moving object.
HI pcs
lb
OK pcs
BATHI lamp : In case of rechargeable battery,
pcs lb
kg display the battery
OK lb
recharging
OK
LO BATtime. lb kg
OK kg
In LO
case of general Mn/Alkaline battery,
LO
BAT
BAT kgdisplay the battery
replacing
LO
time.
BAT

CHARGE lamp : OFF when the charging is completed at charging


in use of adapter.
HI
POWER lamp : ON at connecting DC 12V adapter.
HI
lamp : ON when the current weight
pcs exceeds normal weight.
HI
HI

OK
OK lamp : ON when the current weightlb is normal.
OK
OK lamp : ON when the current weightkg is 80-90% of normal
weight
lamp: ON when the current weight is 50-80% of normal
weight
LO lamp: ON when the current weight is below 50% of normal
weight.

KEY PART
KEY FUNCTIONS
ON/OFF KEY:Use to power ON/OFF.

ZERO
MODE KEY:Use in case of rezero(within 2% of max. capacity)
or to enter the system mode by pressing for 2 sec.
TARE KEY:Use at weighing in use of TARE.

KEY:Use to turn ON/OFF BACK LIGHT.(Option)

6
Key functions in Simple weighing mode
& in Low/High comparing mode.
NET
GROSS KEY: Use to display NET weight or GROSS weight.

HOLD KEY: Use to weigh the moving object.


(Manual, Automatic)
Key functions in COUNT mode

U.W. KEY: Use to display the unit weight.(for approx. 5 sec.)

W. KEY: Use to display the current weight.


(for approx. 5 sec.)

SYSTEM MODE
Simple weighing mode : Display the weight after weighing.
High/Low comparing mode : Determine that the weight is high,
normal or low value and perform the display.
Count mode : Display the quantity by setting the unit or sample
weight.
Quantity comparing mode : Perform the display after discrimination
by setting High/Low quantity.

MODE CONVERSION
Press ZERO
MODE
key for 2 sec. to display the following message.

LO
kg

Execute depending on the below mode conversion and setting


method.

DIGIT INPUT METHOD


Input the digit in use of the below keys.

KEY: Use to set the inputted value to 0 .

KEY: Use to increase the first place of value by one.

7
KEY : Use to move the inputted value to the left by one place.
ENTER KEY: Use to input a decimal point.
MODE KEY: Use to complete the input

Ex) For 20.5 input


DISPLAY KEY DESCRIPTION

Press once
Step 1
(Set to 0 )

Press twice
Step 2 (Input 2 as the
first place value)

Press once
Step 3 (Move to the left
by one place)

Press once
Step 4 ENTER
LO (Decimal point input)

Press five times


Step 5 (Input 5 as the
first place value)

Step 6 MODE Input completion

SIMPLE WEIGHING MODE CONVERSION


DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press once

Press once.
Step 2 ENTER Empty
kg Generally operated

8
High/Low comparing mode conversion
In case of no resetting the High/Low value
DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press twice

Press once
Step 2 ENTER Empty
LO
kg Generally operated

In case of resetting the High/Low value


LO

DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press twice

ZERO Press once


Step 2 MODE kg (Low weight input)

Refer to digit input


Step 3
kg kg in page 8.

pcs
Step 4 High value input
kg

Refer to digit input


Step 5
kg kg in page 8.

Step 6 pcs
Empty Generally operated
LO
kg

9
COUNT MODE CONVERSION
In case of no setting the unit/sample weight
DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press three times

pcs
Press once.
Step 2 ENTER Empty
Generally operated

In case of setting the unit weight


(Min. range of unit weight : Min. Division)
DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press three times

kg
Step 2 MODE Press once

kg
Display the set
Step 3 MODE
kg weight
pcs

Set digit
Step 4 input in Input the unit weight
kg
page 8.

pcs
Press ‘Mode’ key to
Step 5 MODE round and display
Empty
unit weight.
Generally operated.

Step 5. Ex) Min. division is 0.1


Input the unit weight as 0.532 and round off the second place of decimal to 0.5.

10
In case of setting the unit weight by sample
(Samples : 10-200, 10 unit)
DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press three times

Step 2 MODE Press once

Step 3 Press once

pcs
Step 4 MODE Press once

pcs Press five times.


Step 5 Empty Simple count
input. (50pcs)

Step 6 ENTER Empty Press once

Press once with


Step 7 ENTER Sample platform loaded.

Unit weight display.


Step 8 ENTER
kg Press once.

Step 9

pcs
Step10 Empty Generally operated

Note : For setting the unit weight by sample count, the min. unit weight shall
be more than min. division.

11
COUNT COMPARING MODE CONVERSION
In case of no setting the High/Low quantity
DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press four times

pcs
Press once.
Step 2 ENTER
Generally operated.

pcs
In case of setting the High/Low quantity
DISPLAY KEY PLATFORM DESCRIPTION

Step 1 Press four times


kg

pcs
Press once.
Step 2 MODE
kg
kg Low value setting.

pcs Set digit


pcs
Step 3 kg input in Low value input.
kg
LO page 8.

pcs
Press once.
Step 4 MODE
kg High value setting.

Set digit
pcs Press once.
Step 5 input in kg
High value input.
page 8.

pcs
Press once.
Step 6 MODE Empty kg
LO
Generally operated.

12
GENERAL FUNCTION & DESCRIPTION
REZERO FUNCTION(Use at change zero point)
Rezero band range : Within 2% of Maximum capacity

kg kg

Zero point is changed Press Zero key.


Zero lamp is ON.

TARE FUNCTION(Use at weighing in use of TARE)


Maximum TARE range : Maximum capacity
Note : The weight including TARE weight can’t exceed the maximum capacity.

kg kg kg

Put the TARE on platform. Press TARE key. Put the object on platform.
(TARE weight: 10kg) (The TARE weight is saved) (object weight: 20kg)

In case to want know the Gross weight

kg Press NET/GROSS key.


(object TARE weight display)

In case to want know the Net weight


Press NET/GROSS weight key.
(object weight display)
kg Remove the TARE and object from platform to display
the saved TARE weight.

In case to remove the Tare weight

kg Remove the TARE and object from


loadcell and press TARE key.

13
HOLD function (Use at weighing the moving object)
Auto hold function(Hold function operated whenever weighing)

kg kg kg kg kg

Empty the platform. Press Hold key. Hold lamp is ON.


kg Auto hold message is
displayed for 1 sec.

kg kg

When the weight change is The weight is averaged for The hold weight is displayed
stable after putting the approx. 3-5 sec
object(20kg) on platform,
Hold message is displayed.
kg

To release the hold weight Empty the platform or press Hold key
and perform the normal operation.

Auto hold function release

kg kg

Empty the platform. Press Hold key.


Auto hold release message
is shown for 1 sec.
and then normally operated.
Hold lamp is OFF.

Manual hold function (Executed at pressing Hold key)

kg kg kg

Put the object(10kg) on Press Hold key. The weight is averaged for
platform. kg Hold message is displayed 3-5 sec.
for 1 sec.

The hold weight is displayed.


Empty the platform or press Hold key
to release the hold weight and perform
kg
the normal operation.
Hold lamp is OFF.

14
CONVERSION MODE
Enter method
If you press the * key for three seconds to normal operation state,
after display SET message for one second and then enter
F-1 menu.

Key using
: Increase the set point.
: Move to the next menu.
ENTER : Return to the normal operation at completion of setting.

Conversion menu (F01-F02)


F-1: Auto power OFF setting
F-2: RS-232C communication setting

F-1: Auto power OFF setting


Auto power OFF function is used to spare the battery by means
of power off when this scale is not used.
Set point Meaning
F-1 0 Auto power OFF is not used.
Automatically power off when the weight is not changed or no key is
F-1 1
used for 5 min.

F-2: RS-232C communication setting(option)


Set point Meaning
F-2 0 No transmit.
F-2 1 Transmit at both stable and unstable weight.
F-2 2 Transmit at stable weight.
F-3 3 Transmit at data request. (Data request signal D )

RS-232C communication (option)


Port connection method
Connect the computer serial port with RS-232C port in the back panel of
indicator as follows to communicate with computer.

15
RXD 1 2 Transmit Data
TXD 2 3 Receive Data
GND 5 7 Signal Ground
INDICATOR RS-232C port 4 Request To Send
5 Clear To Send
6 Data Set Ready
8 Carrier Detect
20 Data Terminal Ready
Computer serial 25 pin port

RXD 1 3 Transmit Data


TXD 2 2 Receive Data
GND 5 5 Chassis Ground
INDICATOR RS-232C port 1 Carrier Detect
4 Data Terminal Ready
6 Data Set Ready
7 Request To Send
8 Clear To Send
Computer serial 9 pin port

RS-232C Output data format


Baudrate : 9600bps
Data bit: 8, Stop bit: 1, Parity bit: None
Code: ASCII

Transmit data format


Total 22 bytes
, , , data(8bytes) K G CR LF
L B

US(Unstable) GS(GROSS W.) Empty


ST(Stable) NT(NET W.) Empty
OL(Overload) Lamp status byte Weight unit

Lamp status byte: Display the current state(ON/OFF) of indicator


lamp.
bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
GROSS
1 Stable High Low Hold TARE ZERO
W.

Data(8bytes) : Weight data including a decimal point and “-”.


That is, for -13.5kg, : Transmit ASCII code 8 bytes corresponding
to “-”,“ ”,“ ”,“ ”,“1”,“3”,“.”,“5”

16
Data request mode
If F-2 is set to 3, the indicator outputs the specified data when
ASCII code corresponding to "D" is sent to it.

*** Simple transmit/receive program(Language: BASIC)


10 OPEN "COM1: 9600, N, 8, 1" AS #1
20 IF LOC(1) = 0 THEN 60
30 A$ = INPUT$(1, 1)
40 PRINT A$; " " ;
50 GOTO 20
60 B$ = INKEY$ : IF B$ = " " THEN 20
70 PRINT B$ ; " ";
80 PRINT #1, B$
90 GOTO 20

*** Simple transmit/receive program( C-Language)


#include <bios.h>
#include <conio.h>

#define COM1 0
#define DATA_READY Ox100
#define TRUE 1
#define FALSE 0

#define SETTING ( 0x00 | 0xE0 | 0x00 | 0x03 )


int main(void)
{
int in, out, status, DONE = FALSE;
bioscom(0, SETTING, COM1);
cprintf(". . . BIOSCOM [ESC] to exit . . .\n);
while(!DONE)
{
status = bioscom(3, 0, COM1);
if (status & DATA_READY)
if (( out = bioscom(2, 0, COM1) & 0x7F) ! = 0)
putch(out);
if ( kbhit())
{
if ((in =getch()) == ' \x1B ')
DONE = TRUE;
bioscom(1, in, COM1);
}
}
return 0;
}

17
INSTALLING
Loadcell connection
Loadcell connector connect with the loadcell port of indicator back
panel.

Loadcell and connector connection Resolution to loadcell output rate.


method
5V impression to loadcell Recommended
Load Cell Max. loadcell output resolution
Connector
EX+ 2 mV 1/2,000 (Max.)
5 mV 1/5,000 (Max.)
SIG+ SIG-

10 mV 1/10,000 (Max.)
EX-

Reference : please refer to following loadcell color


because the wire color may be different by loadcell
production company and model.

Loadcell color table by company


Connector No. 1 No. 2 No. 3 No. 4 No. 5
Company (EX+) (EX-) (SIG+) (SIG-) (SHIELD)
CAS RED WHITE GREEN BLUE EXTERNAL
BONGSIN RED WHITE GREEN BLUE YELLOW
JEANGWOO RED WHITE GREEN BLUE YELLOW
KYOWA RED BLACK GREEN WHITE EXTERNAL
INTERFACE RED BLACK GREEN WHITE EXTERNAL
P.T RED BLACK GREEN WHITE EXTERNAL
BLS GREEN BLACK WHITE RED YELLOW
SHOWA RED BLUE WHITE BLACK EXTERNAL
SHINKOH RED WHITE GREEN WHITE EXTERNAL
TMI RED WHITE GREEN BLUE YELLOW
TML RED BLACK WHITE GREEN EXTERNAL
TFAC RED BLUE WHITE BLACK YELLOW
HUNTLEIGH GREEN BLACK RED WHITE EXTERNAL

18
USING
Replace the general(Mn, Alkaline)battery or charge the rechargeable
battery in full at using after long-time storage.
Open the top cover and check that the power switch in the left bottom
is ON as follows.
Power on by pressing ON/OFF key in the front panel.
Power switch and connector position

Battery select S/W


Battery Conn.
Ground
RECH DRY

RELAY CN3
GND BATT
CN4
+

EX
EX
IN
NO + +
CN2
RS232

Adapter Conn. L/C Conn.


Main power S/W RS232C Conn.

KEEPING
Power off by pressing ON/OFF key.
Open the top cover and power off for a long-time storage.

BATTERY
Battery replacement or charging time
Power is off in a specific time after BAT signal flickers in the left
bottom of display part on using
Replace the general battery or charge the rechargeable battery.

General (Mn, Alkaline) battery using and replacing


Open the top cover and make as follows in the left bottom of
cover at using the general (Mn, Alkaline) battery.
Insert the battery into battery case after checking the battery
polarity.

19
RECH DRY
Battery
BATT
CN4
+

RECH DRY + C

RELAY CN3
GND BATT
CN4
+

EX
EX
IN
NO + +

CN2
RS232

Rechargeable battery using and recharging


Open the top cover and make as follows in the left bottom of cover at
using the rechargeable battery.
Open the jack protecting cap in the back panel and connect the
adapter as follows.
When the adapter is connected with indicator, RED(Power) lamp is
ON.
When the battery is charging, RED lamp is ON at charge lamp part.
When the charging is completed, GREEN lamp is ON at charge
lamp part.
It takes approximately 4-5 hours to charge in full.
Adapter

RECH DRY

BATT
CN4
+

RECH DRY + C

RELAY CN3
GND BATT
CN4
+
EX
EX
IN

NO + +
CN2
RS232

Battery using period


Conditions Using period

Recharging battery : Approx. 100 hours


BACK LIGHT(E.L) is
Mn battery : Approx. 150 hours
not used.
Alkali battery : Approx. 300 hours

Recharging battery : Approx. 30 hours


BACK LIGHT(E.L) is Mn battery : Approx. 45 hours
always used Alkali battery : Approx. 90 hours

20
TEST FUNCTION
Moving method
Open the top cover of indicator and power on with pressing CAL switch in
the left top. At that time, LCD screen shows “ZERO = TEST TARE = CAL”
message and pressing ‘Zero’ key moves to Test 1.

Test menu (tESt 1 ~ tESt 4)


TEST 1: Liquid crystal display test
TEST 2: A/D conversion test
TEST 3: Key test
TEST 4: RS-232C communication test

TEST 1(Enter ‘TEST’ mode to start TEST 1.)


Function : LCD TEST
Used key LCD screen Description

ENTER key :
Press once: Test running Display TEST 1 state.
Press one more time: HI

OK
pcs
lb
Next menu LO BAT
kg

TEST 2
Function : A/D conversion test
Used key LCD screen Description

ENTER key : Display TEST 2 state.


Press once: Test running
Press one more time: Display the digital value
Next menu corresponding to current weight.

TEST 3
Function : Key test
Used key LCD screen Description

ENTER key : Display TEST 3 state.


Press once: Test running
Press one more time: Pressing the key to test shows
Next menu the key code.

21
Key code
Key name code Key name code
1 ZERO 2
TARE 3 GROSS/NET 4
HOLD 5

TEST 4
Function : RS-232C communication test (Computer link test)
Used key LCD screen Description

ENTER key : Display TEST 4 state.


Press once: Test running
Press one more time: Receive : “1” Transmit : “2”
Move to initial menu

Reference 1. Connect the computer COM port with RS232C connector of indicator
before running this test.(Possible only at installing RS-232C option)
Reference 2. Press “1” in computer keyboard and check that “1” is shown in LCD
screen of indicator and check that the next ASCII code value, “2”
is inputted in computer.

ASCII code table


Char Code Char Code Char Code Char Code Char Code Char Code
32 0 48 64 P 80 96 p 112
33 1 49 A 65 Q 81 a 97 q 113
34 2 50 B 66 R 82 b 98 r 114
35 3 51 C 67 S 83 c 99 s 115
36 4 52 D 68 T 84 d 100 t 116
37 5 53 E 69 U 85 e 101 u 117
38 6 54 F 70 V 86 f 102 v 118
39 7 55 G 71 W 87 g 103 w 119
40 8 56 H 72 X 88 h 104 x 120
41 9 57 I 73 Y 89 i 105 y 121
42 58 J 74 Z 90 j 106 z 122
43 59 K 75 91 k 107 123
44 60 L 76 92 l 108 124
45 61 M 77 93 m 109 125
46 62 N 78 94 n 110 126
47 63 O 79 95 o 111

22
Loadcell test method
Power off before testing the loadcell output voltage and check that the
loadcell output voltage(SIG+, SIG-) is same as the specified output voltage
with ON/OFF key in the front panel pressed.

CALIBRATION MODE
Moving method
Open the top cover of indicator and power on with CAL switch pressed. At
that time, LCD screen shows “ZERO = TEST TARE = CAL” message and
pressing ‘Tare’ key moves into CAL 1.

Calibration menu (CAL 1 ~ CAL 5)


CAL 1: Maximum capacity setting
CAL 2: Minimum division setting
CAL 3: Span weight setting
CAL 4: Zero calibration
CAL 5: Span calibration

Digit input method


: Increase the first place value by 1.

: Move to the left by 1 place.


ZERO : Set the inputted value to “0”.
ENTER : Move to the next menu.

Ex) 150 input


ZERO key : Press once (LCD screen:0)

key : Press once (LCD screen:1)

key : Press once (LCD screen:10)

key : Press once (LCD screen:15)

key : Press once (LCD screen:150)

23
CAL 1(Enter ‘Calibration’ mode to start CAl 1.)
Function : Maximum Capacity setting
Setting range from 1kg to 99999kg
Used key LCD screen Description

Input by digit input 5kg


method.
ENTER key : Move to 200kg
the next menu.

CAL 2
Function : Minimum Division setting
Setting range from 0.001kg to 50kg
Used key LCD screen Description
key : Input the next
division 0.001kg
key : Input the
previous division
0.5kg
ENTER key : Move to
the next menu.

Min. Min. Min. Min. Min.


division division division division division
1 0.001kg 4 0.01kg 7 0.1kg 10 1kg 13 10kg
2 0.002kg 5 0.02kg 8 0.2kg 11 2kg 14 20kg
3 0.005kg 6 0.05kg 9 0.5kg 12 5kg 15 50kg

CAL 3
Function : Span weight setting
Setting range from 1kg to 99999kg
Used key LCD screen Description

Input by digit input Set the span weight


to 5kg
method.
ENTER key : Move to Set the span weight
the next menu. to 200kg

24
Reference 1. The span weight shall be within 10-100% of maximum capacity.
The initial span weight is 100% of max. capacity but if the current span
weight is different from it, input the desired weight again.
Reference 2. The span weight shall be above 10% of maximum capacity.
If not so, an error message(ERR22) is shown.
Reference 3. Never install the span weight higher than maximum capacity because an
error message(ERR23) is shown.

CAL 4(Auto zero calibration takes about 30 second)


Function : Zero Calibration
Used key LCD screen Description

Empty the platform and when LCD


ENTER key : Move into screen shows “ULOAD”, press
initial menu after span ‘ENTER’ key. Then move into span
calibration calibration after counting down
from 5.

Reference 1. Zero calibration is performed automatically by program.


Loadcell output is automatically adjusted within 0.05mV and 5mV

CAL 5(Auto span calibration takes about 30 sec.)


Function : Span Calibration
Used key LCD screen Description
Put the weight set in CAL3 on
platform and press ‘ENTER’ key.
ENTER key :
Move into initial Span calibration is terminated.
menu after span
Error occurrence.
calibration
Move into initial menu
automatically.

Reference 1. Display “END” message and move into the initial menu at completion
of span calibration
Reference 2. At occurrence of ERR24, reset by lowering the resolution.

25
SEALING
Perform the sealing as shown in the below figure after the calibration is
completed.
Open the top cover.
Cover CAL switch in the left top of cover with sealing plate and
screw 2 sealing bolt.
Connect the sealing wire as follows and compress the sealing lead.

Sealing bolt

Sealing wire Sealing plate


Sealing lead

OPTIONAL DRAWING

WALL MOUNT FIXED BRACKET


INDICATOR GUDE BRACKET

60
5 2.5

4.5
50
3
61

50
89

94

FIXED BRACKET

26
ERROR MESSAGE DESCRIPTION AND MANAGEMENT
Errors in weighing mode
Err 01
Cause
The indicator can' t be initialized due to platform shaking.
Management
Put the platform on a flat place without shaking and power on.
Err 02
Cause
The loadcell is connected wrongly or A/D conversion part has a fault.
Management
Check that the loadcell is well connected with main unit.
Err 03
Cause
The calibration isn' t executed or the data of internal memory is lost
due to a electrical shock.
Management
Perform the required specification again in setting mode.
Err 06
Cause
The printer isn' t properly connected.
Management
Check that the printer connector has no fault.
If the message is displayed in spite of no fault in printer and printer
connector, call to A/S part of CAS CO., LTD.
Err 09
Cause
The current weight exceeds zero range.
Err 10
Cause
The specified TARE weight exceeds the maximum capacity.
Management
Set the TARE weight lower than max. capacity.
Or set the maximum capacity higher than the desired TARE weight
in calibration menu. and perform the calibration again.

27
Err 13
Cause
The initial zero range exceeds above 10% of maximum capacity.
Management
Check the loadcell state.
Over
Cause
Too heavy object is put on platform so that it may exceed the
maximum capacity.
Management
Never put the weight exceeding the maximum capacity
If the loadcell is damaged, replace the loadcell.

Errors in calibration mode


Err 22
Cause
The span weight for span calibration is set below 10% of max.
capacity.
Management
Set the span weight for span calibration more than 10% of max.
capacity in CAL 3 of calibration menu.
Err 23
Cause
The span weight for span calibration exceeds 100% of max. capacity.
Management
Set the span weight for span calibration within allowable capacity
range in CAL 3 of calibration menu.
Err 24
Cause
The resolution is set too high in comparison with loadcell output.
Management
Set the high minimum division in CAL 2.
Err 26
Cause
Zero value is too high.
Management
Check that the platform is loaded on.

28
Err 27
Cause
Zero value is too low.
Management
Check that the platform is empty.
Err 28
Cause
The weight trembles.
Management
Check that the loadcell connector is properly connected.

Errors in count mode


Err 61
Cause
The count unit weight and sample weight are set too high.
Management
Set the count unit weight low or the sample quantity high.
Err 62
Cause
The count unit weight and sample weight are set too low.
Management
Set the count unit weight high or the count sample quantity low.

29