 Operation introduction

1. Key Functions
There are three buttons on the front panel, the instrument can be carried out by these three
buttons
Debugging. Display of measured values on the LCD screen after commissioning.

2, after the electricity meter display, press the Set button ((SET) to enter the menu.
3, the height of the probe entered in the "Reference Zero," "reference zero" bit in the menu See
Schedule III set menu structure. (Probe height of the probe into the weir tank emission surface
water port
Distance)
4, calibration "4mA flow values" and "20mA flow values"
4mA flow value: when the instantaneous flow rate equal to the value of the output 4mA.
20mA flow value: when the instantaneous flow rate equal to the value of the output 20mA.
"4mA flow values" and "20mA flow values" position in the menu, see Schedule II menu knot
Composition.
5. Choose a weir tank, to consider the size of the flow within the channel, the water flow within the
channel 。State whether the formation of the free stream. .The maximum flow rate of less than 40
liters / sec (144 t / h) recommend the use of a right angle triangle weir; Greater than 40 liters / sec
is recommended to use Parshall flume; shorter upstream channels, the maximum flow rate and
large .To 40 liters / sec is recommended to use a rectangular weir. When using the first zero
calibration reference instrument measuring .Point of reference for the probe zero zero distance to
the weir tank water level. (The instrument default choice Parshall flume)
① triangular weir
Triangular weir, can in the menu "9 weir tank type" → "1 triangular weir" → "1 worker
As a state "choice" open "," 2 triangular weir angle "select instrument on actual angle
Can be automatically calculated based on the flow rate of water corresponding to the water
level.
② rectangular weir
Rectangular weir, the menu can be "9 weir tank type" → "2 rectangular weir" → "1 . Work Status
"select" On ", and" the "Select" 2 standard channels 0.25 . M,,,, 0.75 m 0.50 m 1.00 m,,
non-standard channel ", the instrument can be based on . Automatically calculate the level
corresponding to the level of traffic.
③ trapezoidal weir
Use trapezoidal weir, in the menu "9 weir tank type" → "3 trapezoidal weir" → "1 Work status
"choice" open "and in the" Enter the actual two weir sill width B "in .The actual channel width weir
sill, the instrument can be automatically calculated based on water level corresponding flow The
amount.
④ Parshall flume
Use Parshall slot, you can in the menu "9 weir tank type" → "4 Parshall flume" → n .According
to the throat width "b", from the "Schedule II
Parshall flume water - flow formula "in isolated Repair work coefficient c and exponent n, input to
the menu "9 weir tank type" → "4 Parshall flume "→" 2 repair work coefficient c "and" 3 index n ".
Meter can be based on water Automatic water level corresponding to the calculated flow rate.

 First.Product introduction

Ultrasonic flow meter with a weir tank with the use of measuring the flow of water in the open
channel. Mainly used to measure the waste water treatment plant, sewage outfall enterprises, the
city sewer and water conservancy in the flow channels.
As the Company through the air using ultrasonic instruments to non-contact measurement
method. Therefore, under dirt, corrosive liquid condition, than contact type instrument, with higher
reliability and durability.

 Second.Work principle

This series of instruments to directly measure the height of the channel or the liquid inside the
tank. When used for open channel flow measurement in open channel installation weir tank. Weir
tank the size of transfer within the open channel flow into high and low level. Instrument
measuring the amount of water weir tank, then the corresponding weir tank water level -
discharge relation deduced traffic.
1.An ultrasonic liquid level measuring principle
Emit ultrasonic transducer emits ultrasonic pulses propagating through the medium of
propagation to the measured surface, the reflection and then return to the receiving transducer
through an acoustic medium, the measured ultrasonic pulse propagation from the transmitter to
the receiver in the acoustic medium time. Then according to the acoustic medium speed of sound,
it can be regarded as the transducer to the surface from a distance. Thereby determining the level.
Therefore, we can calculate the distance from the probe to the reflective surface D = C * t / 2
(divided by 2 is because the sound waves from the transmitter to the receiver is actually a round,
D is the distance, C is the speed of sound, t is the time). Can be obtained through subtraction
level value.
2, the measured flow weir tank principle
Smooth circulation within the open channel flow, the higher level; flow rate, the lower level
(Figure 2.1). By measuring the water level can calculate the flow rate. Correspondence between
open channel flow and water level within the ordinary between the affected channel slope ratio
and surface roughness influence. In the channel installation weir tank, resulting in throttling effect,
so open channel flow and level within a fixed correspondence between this relationship depends
on the amount of water the size of the weir tank constructed to reduce the impact as much as
possible channels small.
Commonly used weir tank there, right triangular weirs, rectangular weirs and Parshall flume
(Figure 2.1)

Figure 2.1 common weir tank and the water level and flow changes Relations

Use ultrasonic flowmeter, you must know the amount of water when installing with weir tank -
Flow correspondence.
Weir tank of water - flow relationship can be from national verification procedures "open
channel weir tank .Meter "JJG711-90 in found. Extract a part of this specification (sixth, the
amount of water Weir tank). Parshall flume know throat width b, you can use the corresponding
formula calculates water - Flow correspondence.Weir is also a right-angle triangle with the
appropriate formula to calculate the water - flow correspondence.

Rectangular weir has a corresponding formula. But also with the channel size installation,
determine the Water - flow relationship, rectangular weir and channel width B, the opening width b,
upstream of the weir sill height p Related.

If you calculate the amount of water weir trough level - traffic are not familiar, you can use the
weir tank ginseng .The number of notifications meter production plant. Plant help calculate. It
should be noted at the same time provide the above and determined Water - parameters related
to the flow relationship.

 Third. Technical parameter

Function integrated Split-type


Measurement range type
0.1 L/second ~ 99999.99 m3/H
Cumulative flow Max: 4,290,000,000.00 m3
Level maximum range 3m
Level Measurement 0.5%
Accuracy
Resolution 3mm or 0.1%(take the bigger one)
Flow measurement Standard weir tank is 1 to 5% (weir tank and channels
meet the national standard requirement)
accuracy Non-standard weir tank isload
10 to 30 percent.
Analog Output 4-wire 4 ~ 20mA / 750Ω
Relay output (Option) Group 2 AC 220V / 8A or DC 24V / 5A,
Power supply Standard 24VDC 100mA; optional 220V AC + 15%
50Hz。
Power supply (Optional item) 12VDC, battery-powered,
Operating ambient Display Meter -20 ~ + 60 ℃, the probe -20 ~ + 80 ℃
temperature
Workplace stress Standard atmospheric pressure
Operating humidity ≤90% RH, non-condensing
Process temperature -20~80℃;
Process pressure Standard atmospheric pressure
Communication Optional 485,232 communication, MODBUS protocol
Protection class Display IP67, Display meter IP65, probe IP68
probe IP68
Probe cable no Standard 10 meters and the maximum is 100 m
Probe mounting M48 × 2mm screw + supporting nuts
dimensions
Probe material ABS is standard, and in a corrosive environment to use PTFE
materials.

 Fourth. Installation

1.Installation of open channel flow meter

Integrated open channel flow meter

Split-type open channel flow meter


2 , install the probe
Ultrasonic flow meter probe can be mounted directly above the weir tank water level observation
points . Emission surface of the probe surface to be aligned and perpendicular with the surface.
Probe cover can be placed on the horizontal ruler , cover school level by aligning the probe
surface . parshall flume level observation point from 0.1 to 0.5 m upstream of the slot position ;
triangular weir on the upstream side of the maximum depth of water over the weir from 3 to 4
times the slice rectangular weir .

installation

22 , install the probe


Ultrasonic flow meter probe can be mounted directly above the weir tank water level observation
points . Emission surface of the probe surface to be aligned and perpendicular with the surface.
Probe cover can be placed on the horizontal ruler , cover school level by aligning the probe
surface . Parshall flume level observation point from 0.1 to 0.5 m upstream of the slot position ;
triangular weir on the upstream side of the maximum depth of water over the weir from 3 to 4
times the slice rectangular weir .
The bottom of the threaded mounting

① at the top of the object mounted a flange ②put a same inner diameter gasket on flange

③ The transducer alignment flange hole ④ The transducer into the flange hole

⑤ See the situation from the bottom flange ⑥ Flange down a piece of inside diameter of the same gasket

⑦ Tighten the nut fixed transducer ⑧ Installed the transducer


Top threaded mounting - - - - lifting method to install

① At the top of the probe with a nut screwed ② Probe into the line protection tube

Mounted on the jar , the pool, on the cover, install the same way with the above bracket

After the probe is installed, the probe will expose the cover or emission surface waveguide . Not
shrink in the cover or inside the waveguide .

3. 3, site material flow weir trough installation drawing


◆ Triangular weir installation

◆ Rectangular weir installation


◆ Parshall Flume installation

4.Installation of measuring weir flume

(1). Measuring weir trough the centerline to the channel center line overlap, the flow doesn't appear drift into the
measuring weir flume.
(2). After measuring weir trough drainage systems, the flow of water to flow freely.
Triangle weir, rectangular weir downstream water level is lower than weir candy (as shown in figure 4.4.1);Slot
is smaller than the submergence degree Parshall Flume "groove " Parshall Flume critical submergence
degree (figure 4.4.1).

(3). Minimum upstream of measuring weir trough to be greater than 5 times the channel width of flat section,
measuring weir flow smoothly into the slot, the standard is "there is no waves on the water surface.That it has
no drift, and channel slope formation of momentum.

(4). Measuring weir flume is installed on the channel to be strong.And channel wall, canal bottom link to close,
can't leak.All water flows through measuring parts of measuring weir tank.Are measuring weir plate measuring
weir crest;Measurement part of the flume is groove in throat.
4.4.1

5.Electrical wiring

◆Terminals real figure and schemes of split type open channel flow meter
Wiring methods :
Transducer : red : connect Trans1 transducer Blue Line : Temp 1 + temperature sensor +
Black line : GND Ground

Current output: " Current + " connect mA1 +; " Current - " connect mA- / GND
Relay : RLlnA and RLnB normally open

If you want to relay default state is " open" , the wiring is connected RLlnA and RLnB.
RLnA and RLnC normally closed

If you want to relay default state is " normally closed " wiring to connect RLlnA and RLnC.
Power cord: the power is AC : Connect L, N
Power is DC : 24V + connect 24VDC +, GND connect 24V DC-

Terminals schemes of integrated type open channel flow meter


integrated open channel flow meter 24 VDC terminal diagram

integrated open channel flow meter 220V AC terminal diagram

Current flow, current +connect mA + , current -connect mA - -.


Relay: connect the RLn + with RLn - terminal, the default status to normally open.
n= 1 or 2 , refers to the first one or the second relay
The power cord: if it is 220 v alternating current (ac), firewire connect L terminal, zero N terminal.
If it is a direct current (dc) : power supply 24 v + connect 24 v + terminals, power supply 24 v -
connect 24 v -
Terminals.
 Fifth .Set

1 , operating mode interface Introduction


This series of ultrasonic flow meter has two modes of operation and setup , after the device is
powered on and complete the initialization process , the meter will automatically enter Run mode
and start measuring data , record data.

3 , the menu instructions


⑴ Key Description :
① instrument has three buttons: on Up, Under Down, set the SET.
②Set key is used to : Select OK to enter or exit.
③ up and down keys are used to : move the cursor up and down , in the selection box, select the
option in the input box, select or modify the value.
④ Check echo state diagram : Hold down the "▲" link, then hold down the "SET" button for 2
seconds , you can go to the echo state diagram .

⑤ quit echo state diagram : Hold down the " " hold , then press and hold the "SET" button for
2 seconds , you can exit the echo state diagram

⑵ menu interface and operating instructions :


① in run mode screen, press t Set button to enter a menu interface :
◆ parameter is not locked a menu interface :
Parameter lock a menu interface :

② a menu the description :


◆ "0 0 0 0 End Settings"
When you select this option, press the Set button to return to the operating mode interface
◆ "0 0 0 0 not locked"
Menu is not locked , allowing others to change.

◆ "1 1 1 1 Global Lock "

Menu locked , when you set up the parameters , you do not want others to tamper with , the menu
lock , so we must enter a password to unlock the menu. The initial password of the flowmeter 25 ,
the user can modify the initial password arbitrarily set your own password ( please remember the
password remind yourself set , should forget should contact the manufacturer ) .
★ When the parameter is not locked , press t Set to enter the parameter setting interface locked :
Unlocked : not locked, it will all menus can be freely modified .
Global Lock : After the Global Lock , you must enter a password to modify .

★ When the parameter is locked, press t Set

◆ "2 Range Settings"


1 Reference 0:00 : Set flowmeter reference zero , this is mainly level measurement when it makes
sense ; factory set default values ​ ​ corresponding to the maximum range .
Low range .
2 : Set 4mA corresponding to the output level meter measurements ; lower limit can be set as the
flow value, i.e., when the level is less than the set value, the flow rate is zero. The factory default
is 0 .

3 High Point Range : Set the output level corresponding to the measurement count is set 20mA ;
upper limit can be set as the value of the flow rate , i.e. the flow rate to maintain the set value of
the flow when the liquid level exceeds the set value . For example, to indicate when the liquid
level exceeds 0.5m 0.5m 0.5m is to maintain the flow of traffic. The factory default setting for
maximum range .
4 Display Unit: : There are m, cm, mm three units can be selected , m: meters display ; cm:
centimeters display ; mm: millimeters shows the default factory settings for m.

◆ "3 measurement modes "


1 Mode Select : distance measurement and level measurement two to choose from. Distance
measurements: the horizontal distance measured value display to probe ; level measurement :
Show value of zero to the reference surface distance. The factory default setting is level
measurement .
2 Response time: There slow, medium , fast three to choose from. Slow : Slow response rate ,
high accuracy, is not susceptible to interference ; speed : between slow and fast ; Fast: response
rate, the measurement accuracy is low, susceptible to interference . The factory default setting for
the speed .
3 Security Level : holding , minimum, maximum , setting four to choose from. Hold: throw wave
after the system displays the value last measured value , the corresponding current value ;
Minimum: throw wave after the display system is 4mA, current is 4mA; Max: throw wave after the
display system is 20mA, current of 20mA ; setting: throw wave after the system displays the value
last measured value , the current output is set current settings. The factory default setting is
maintained.
4 set the current: Set throw wave after the specified output current greater than 3.6mA, less than
22mA, and then select to Hold / Max / Min invalid . The factory default setting is 3.6mA.

4 Probe Set " ( ( ( ( Please do not modify this , is factory set with ) ) ) )
1 probe options: Do not modify, need to be modified under the guidance of professional and
technical personnel.
2 blind setting : Do not modify, need to be modified under the guidance of professional and
technical personnel.
3 Short Sensitivity : Do not modify, need to be modified under the guidance of professional and
technical personnel.
4 Short threshold : Do not modify, need to be modified under the guidance of professional and
technical personnel.
5 Long Sensitivity : Do not modify, need to be modified under the guidance of professional and
technical personnel.
6 Long threshold : Do not modify, need to be modified under the guidance of professional and
technical personnel.

5 algorithm to select " ( ( ( ( Please do not modify this , is factory set with ) ) ) ) algorithm
selection : There are special circumstances a special environment two , three special
circumstances , special environment 4 ,
Special environment 5 , 6 special circumstances , special environment 7 , a total of seven to
choose from. The factory default setting for special environments 7 .

1 level alarm 1111 Mode


Relay 1 for water level alarm .
According to the water level is different, there are close , low alarm , high alarm three to choose
from.
OFF: Relay 1 does not act ; Low alarm: relay a low alarm ; high alarm :
Relay 1a high level alarm . The factory setting is off by default

2 level alarm 1111 value


In m , factory- set default is 0.00m.
3 1111 level alarm hysteresis in m , reset the alarm after alarm is triggered the alarm value
measurement requires only a valid + alarm hysteresis . The factory setting defaults to 0 .
Alarm hysteresis can also make a relay to control the pump from the low water level to high level
of the entire work process .

1 For example, for drainage : Requires the pool to one meter below the water level , the pump
stops and drainage ;
The water level to rise 5 meters , pumps started out drainage. Specific settings are as follows :
Alarm 1 mode: high alarm . Alarm 1 Value : 5.00m; Alarm 1 hysteresis : 4.00m.
2 For example, for water : Requires the pool water level to one meter below the water pump
starts ;
The water level rose to 5 meters , the water pump stopped . Specific settings are as follows :
Alarm 1 mode: low alarm . Alarm 1 Value : 1.00m; Alarm 1 hysteresis : 4.00m.
4444 2222 level alarm mode
Relay 2 is used for water level alarm .
According to the water level is different, there are close , low alarm , high alarm three to choose
from.
OFF: Relay 1 does not act ; Low alarm: relay a low alarm ; high alarm :
Relay a high level alarm . The factory setting is off by default .
5555 level alarm 2222 value
In m , factory- set default is 0.00m.
6666 level alarm 2222 Backlash
In m , reset the alarm after alarm is triggered measurements need to alarm the alarm value +
hysteresis
If valid. The factory setting defaults to 0 .

7777 instantaneous flow alarm mode :


Instantaneous flow for the upper and lower limit alarm . There are close , low alarm , high alarm
three
You can choose . OFF: Relay 3 does not act ; Low Alarm : Relay 3 low alarm ;
High Alarm : Relay 3 high alarm . The factory setting is off by default .
8888 instantaneous flow alarm value ::: :
With t / h as a unit, the factory set default is 0 .

9 instantaneous flow alarm hysteresis :


With t / h as a unit, reset the alarm after alarm is triggered the alarm value measurements needed
+ alarm back
Only effective difference . The factory setting defaults to 0 .
010 cumulative flow proportional output :
There are closed , open two to choose from. OFF: Relay 4 does not act ; open : Relay
4 alarm . The factory setting is off by default .
111 preset cumulative flow ::: :
With t as a unit, the factory set default is 0 .
◆ "7 7 7 7 parameter correction " ( ( ( ( Please do not modify this , is factory set with ) ) ) )
Make span calibration , sound velocity correction , the current output calibration , the reference
level correction operation.
Range Correction: : Enter the actual value, the system automatically span calibration . The factory
set default values ​ ​ for the measurement .
Sound Velocity Correction: Enter the actual value, the system automatically correcting speed of
sound , is not used in the gas composition of air time .

4mA correction : Modify the value until the actual output current of 4mA so far. The factory default
setting is 3100 . ( Using z this , according ▲ increase a value , will output 4ma value. )
20mA correction : Modify the value until the actual output current of 20mA so far. The factory
default setting is 7200 .
Reference Level : Enter the appropriate voltage test points measured . The factory default setting
is 5.00 .

◆ "8 8 8 8 Communication Settings"


1111 Address : Select the communication address , the default value is 1.
2222 Baud Rate : Select the communication frequency , there 2400,4800,9600,19200 optional ,
default is 9600 .

3 3 3 three work modes: automatic query report way and method."Automatic" report way is
ultrasonic open channel flow meter automatically send data to the PC, don't need a PC to send
inquiry instruction.
(to be automatic report mode only manufacturer agreement have, MODBUS protocol reported no
automatic mode)
"Query" is given a query need the upper machine instructions, open channel flow meter only
answer once

◆ "9 9 9 9 weir trough type "


1111 triangular weir :
Work Status: Closed : Description not chosen triangular weir ; default is " off ."
ON: Description chose triangular weir .
Triangular weir angle : 90 ° (30 ° / 60 ° / 45 °), corresponds to the angle of the triangular weir , the
user simply selects a different angle , the machine automatically calculated flow value based on
water level .

2222 rectangular weir ::: :


Work Status: Closed : Description not choose rectangular weir ; default is " off ."
ON: Description chose rectangular weir .
Standard Channel : 0.25 m ( 0.50 m /0.75 m /1.00 m / non-standard channels ) : These values
​ ​ represent a rectangular weir throat width, as indicated throat width 0.25 m 0.25 m rectangular
weir , the user simply selects different throat width , the machine automatically calculated flow
value based on water level . If you choose the "non -standard channels " : hide the menu is
displayed :
3333 weir width bbbb; 4 4 4 4 upstream channel width BBBB; 5 5 5 5 weir wall height PPPP, the
user can enter the actual channel size .

3333 trapezoidal weir ::: :


Work Status: Closed : Description not choose trapezoidal weir ; default is " off ."
ON: Description chose trapezoidal weir .
Weir sill width BBBB: user based on the actual size of the input channels , the unit automatically
calculated flow value based on water level .

4444 Parshall flume :


Work Status: Closed : Description not choose Parshall flume ; open : Description chose Parshall
flume , the default value is "on" is the default choice Parshall slot.
Repair work coefficient cccc: Set C value based on supporting the weir tank . The factory default
setting is 0.01 .
Index nnnn: Set n values ​ ​ based on supporting the weir tank . The factory default setting is
0.01 .

Special Note: At this point requires the user to input coefficient cccc repair workers and index
nnnn, the user can find the appropriate repair work coefficient cccc and index nnnn two
parameters depending on the specifications of the groove.

◆ "0 10 Other Parameters"


A 20mA flow values ​ ​ : 20mA flow rate setting value , said output at 20mA corresponding
instantaneous flow
Value. The factory default setting for maximum flow .
A 4mA flow values :: 4mA flow set value that indicates when the output 4mA corresponding
instantaneous flow values.
The factory setting defaults to 0 .
Cumulative flow: ::: copy accumulate when the water meter replacement values . The factory
setting defaults to 0 .
Water Cleared : The cumulative value of water can be cleared .

Flow Unit : : You can change the flow rate display units , "t / h" for ( m ) / h t , "l / s"
For l / s , "t / s" to t ( m ) / sec . The factory default setting is "t / h".
Interface switching: the interface can be displayed in flow and level switch the display between
each other interfaces.
Current output :: 4 ~ 20mA resistive load capacity of 750Ω. For flow meter , 4 ~ 20mA
According to the size of the output instantaneous flow ; timing for level by level low output. The
factory default setting for flow output .
◆ "1 11 benchmark select"
Benchmark selection : Yes : Use benchmarking to correct the measurement accuracy. No : Do not
use a benchmark to calibrate the measurement accuracy. No default factory settings .
Benchmarking calibration : When selecting a benchmark to calibrate the measurement accuracy
is required to conduct benchmark calibration, the probe onto the underside surface of the
benchmark measurements can enter this menu .

◆ "2 12 Reset Choose "


Factory Reset : Yes: to restore to factory settings just state . No : exit. No default factory settings .
System reset : Yes : Restore system settings. No : exit. No default factory settings . ( ( ( ( Please
do not modify this , is factory set with ) ) ) )

 Sixth .Main function

1 , the measurement level .


2 , open channel flow measurement .
Suitable for right-angle triangle weir , rectangular weir , trapezoidal weirs and Parshall flume .
3,4 ~ 20mA current output 4 ~ 20mA remote resistive load capacity of 750Ω. For flowmeter , 4 ~
20mA output by traffic ; timing for level by level output. 4mA corresponds to the level or flow of
zero value ; 20mA corresponds to the value set in "Setting the constant " menu item in the "20mA
current value" menu.
4 SPDT relays
Use AC 250V / 8A or DC 30V / 5A relays, programmable state . Will be one of the relay 's size limit
for instantaneous flow , in addition to a relay output for the cumulative flow ratio , that every
cumulative flow preset closing time .
5 , large storage capacity
The maximum cumulative amount of water meter can reach 12 digits ( including two decimal
places ) , when accumulated over 4290000000.00t, automatically cleared , re- cumulative . The
maximum instantaneous flow rate can reach 99999.99t / h.

 Seventh .The volume of water would sink

Choose quantity of water would sink to the type, taking into account channel flow the size
of the channels within the water. Internally, the possibility of formation of free. Under most flows
are different, we can choose not to With the Weir tank.
① The largest flows of less than 40 litres/second recommended the use of a right angle
triangular Weir; and
② Greater than 40 l/s recommended to use the Marshall sink; and
③Upstream channels is relatively short, maximum flow, and greater than 40 litres/second is
recommended for use rectangular Weir with.
Conditions permitting, the best choice of the Marshall tanks. The Marshall sink, the
water-flow relationship byLab objectives set out, and on the upstream of the channel sink, the
conditions required to be weak. Triangular Weir And rectangular Weir the water level-flow
relationship originated in the theoretical calculations, it is easy due to the neglect of some use
Conditions that bring in additional errors.
Use of glass, steel production volume of the water Weir or tanks. Triangular Weir, the
rectangular Weir Weir population size to be accurate, Towards the water side of the surface to
smooth; the Marshall sink throat some of the size to be accurate, sink, within the To smooth.

1, the right-angle triangular Weir


Figure 7.1.1 is a right-angle triangular Weir processing. Use of the above-mentioned
right-angle triangular Weir, In the menu“9 Weir trough type”→“1 right-angle triangular Weir” →“1
working condition” Option“open”, the meter can be under the water level automatically calculate
the water should flow.
Figure 7.1.1 a right angle triangular Weir structure

Triangular Weir installed in the channels, as seen in figure 7.1.2. Weir plate to vertical, in
the channels of the Central axis. Processing triangular Weir, so that the top corners become
rounded corners, in determining the water level is equal to zero position when it is to be noted that
triangular Weir, the water levels at the zero point should be in a triangular Weir on the side of the
extension line at that point. Meter probe to install the upstream distance Weir plate by 0.5 ~ 1 m
position.

Figure 7.1.2 right-angle triangular Weir in a channel on the installation and triangular Weir,
the water levels at zero.

2, the rectangular Weir


Rectangular Weir, according to figure 7.2.1 processing. Rectangular Weir the water
level-flow relationship depends primarily on Weir population-wide“b”. Also with the upstream
channel width“B”and Weir Campbell high“p”. As the use of figure 6.2.1 rectangular Weir, in the
menu“9 Weir trough type”→“2 rectangular Weir”→“1 work status”option to“turn”, and“2 standard
channels”choose“to 0.25 m, 0.50 m, 0.75 m, 1.00 m, non-target channels”, the meter can be
under the water level automatically calculate the water should flow.
At the actual site, the site of the rectangular Weir Weir mouth width exceeding 1.00 m,
which is the use of non-standard rectangular Weir to measure. This instrument has been
equipped with this function, based on the position measurement of the non-standard rectangular
Weir b, B, P value of the input, then we can measure.

Figure 7.2.1 rectangular Weir's anatomy

Material: fiberglass , PVC or stainless steel


1. The surface is smooth , flat, distortion-free
2. The top edge of the rectangular weir must be straight and smooth.
3.j: when installing the embedded portion of the sidewall and bottom , determined according
to the situation.
4.b = weir width , B = upstream channel width , h = liquid height p = height of the weir wall
Figure b, B, h, p dimensions are as follows (unit : mm ) :
Rectangular Weir installed in the channels, as seen in figure 7.2.2. Weir plate to vertical,
installed in the channels of the Central axis. The dashboard camera installed in the Weir plate
upstream by 0.5 ~ 1 m position.

Figure 7.2.2 rectangular Weir installation

3, the ladder-shaped Weir


Use of ladders Weir, in the menu“9 Weir trough type”→“3 trapezoid Weir”→“1 work
status”option to“turn”, and“2 Weir threshold width B”enter the actual channel Weir threshold width,
the meter can be under the water level automatically calculate the water should flow. The
installation with rectangular weir trapezoidal weir.
4, the .Parshall Flume

The Marshall sink structure as figure 7.3.1 it. The Marshall sink, the marking of the size of
the throat. width“b”. First, according to the application needs of the largest flows, from
the“schedule II of the Marshall sink, the water-flow formula”to find out suitable the Marshall trough
the throat.“b”. Again, from“the first schedule to the Marshall sink structure size”response throat.
width is equal to“b”.the rest sink, the other size. As“L1”, the La”, And L”, the L2”, etc. These
size enter the chart 7.3.1 the right side of the column, as in figure 7.3.1 processing shape,
installed in the channels, as seen in figure 7.2.3.

Parshall flume water - flow relationship is generally shaped like : “Q = Chan “formula. According to
the throat width "b", from - find out repair work coefficient “c “and exponent “n " Schedule II
Parshall flume water flow formula ", the input to the menu "9 weir tank type " → "4 Parshall flume
" → "2 repair work coefficients” c" and "3 index “n", the instrument can automatically calculate
the flow rate value corresponding to the water level .

Material: fiberglass , PVC or stainless steel


1. The size should be accurate within
2. To the inner surface is smooth, smooth.
3. The wall thickness is greater than 8mm
The probe holder if the span is too large, try to increase the intensity
5. The stent is greater than the distance from the highest level 400mm
6.j size in the channel installation concerned , determined according to the situation.
 Eighth .Error phenomenon and processing

phenomenon reason solution


Flowmeter no display 1. The power supply 1. Check the power
connection is not good cord, the voltage is
2. LCD display poor measured with a
contact multimeter.
3. The damage of LCD 2. To plug the LCD panel
screen and its wiring
3. Returned to the
factory to replace the
LCD panel
Into the liquid level display 1. The distance beyond the 1. Consider replacing a
Interface, if found Small flowmeter range to be larger range of flowmeters
screen display Beep tested 2. Wait measured media to
symbol does not 2. Strong disturbance, restore calm , the device
waves measured medium, will automatically return to
change ,Figure " " is a mixing, etc. normal measurement. Or
system into the wave 3. With frequency converter, use the diameter of the
state. motor and other strong wave guide 150 to install
interference sources flowmeters.
4. The probe measured 3. flowmeter grounding ,
plane misalignment power to do isolation,
5. Measured space with output signal lines do
obstacles, such as feed isolation.
opening support bar, and so 4. To re- calibrate the
on probe , the probe is
6. Level into the blind area measured to keep the
page vertically .
5. Re- select the
appropriate installation
location to avoid
interference occur
6. raise the probe
mounting height , the
probe emitting surface
mounting height > highest
page + blind.
Description: The repair work coefficient "C" with the index "N", with the number 1 ( the 1st slot ), for
example , repair work coefficient C is : 60.4
Index N is : 1.55