Professional Documents
Culture Documents
WS-3811 Strain Gauge Signal Acquisition System: 1. Features
WS-3811 Strain Gauge Signal Acquisition System: 1. Features
In India)
3. Specification:
3.1 Measuring method: Static and dynamic strain measurement, voltage measurement;
3.2 Channel options: 2, 4, 6, 8, 10, 12, 14, 16 channels. Multi-channel may consist of
multiple instruments through Ethernet network.
3.3 Computer Interface: USB or Ethernet port with 10/100BaseT.
3.4 Bridge excitation voltage
① Programmed bridge voltage: 2,5,10 and 15V,
② Bridge voltage accuracy error: <0.2%,
…………………………………………………………………………………………………………………
Address Huda Sector23A Gurugram India 122017 www.sensor.in
Beijing Wavespectrum S&T Co., Ltd (Mechanics Lab. In India)
③ Common mode rejection ratio (CMRR): not less than 110 dB;
④ Input protection: when the input voltage is more than ±10V, amplifier adopts full
protection
3.8 Excursion
① Time excursion ≤2/2h
② Temperature drift ≤1/°C (in working temperature)
③ Nonlinear error ≤0.05%/h(FS)
④ Offset drift ≤2/2h
3.9 Digital Filter (Program Control)
① Filter Mode: 8 levels ellipse lowpass filter; Strain Gauge Box
② Cutoff frequency from 2Hz to 50kHz programmable;
③ Undulation of smoothness of the filter is less than 0.1dB during f<0.5fo;
④ Attenuation of the filter is -80dB/Oct;
3.10 Programmable frequency range is from DC(0kHz) to 50kHz;
3.11 Noise (When input with short circuit, max gain and max bandwidth reduced to input)
① Dynamic measuring noise<5Vrms
② Static measuring noise<1Vrms
3.12 Strain Gauge
① Gauge resistance is 120Ω. It may make of 1/4 gauge, half gauge and full gauge
② Strain sensor resistance: 120Ω
③ Strain sensor sensitivity K=2
④ Standard strain gauge’s cable length is 2m
⑤ Gauge resistor resistance range from 50Ω to 10000Ω (Customized)
3.13 Working Temperature Range: -10°C50°C
3.14 Humidity: ≤85%RH
3.15 Dimension: W370xH165xD365mm
3.16 Power Supply: AC220V/50Hz
3.17 Weight ① For 2 to 8 channels about 5.4kg
② For 10 to 16 channels about 8.2kg
3.18 Software: The software’s function is static and dynamic strain measurement,
oscilloscope, compensation of sensitivity factor (K), stress calculation, curve display
and plot, data structure convention for Word, MatLAB and Excel.
…………………………………………………………………………………………………………………
Address Huda Sector23A Gurugram India 122017 www.sensor.in
Beijing Wavespectrum S&T Co., Ltd (Mechanics Lab. In India)
Extension Cable:
Order Type Length(m) Order Qty
WS-3811/C5 5
WS-3811/C10 10
WS-3811/C20 20
WS-3811/C50 50
5. BX Series Strain Gauges (Optional)
Gage Size Base Size
Type Model
Length mm Width mm Length mm Width mm
BX120-5AA 5.0 3.0 9.5 5.3
BX120-6AA 6.0 4.0 10.0 4.5
BX120-7AA 7.0 4.0 12.0 7.0
BX120-8AA 8.0 3.0 13.0 6.0
BX120-2CA 2.0 1.0 7.2 7.2
BX120-3CA 3.0 2.0 11.5 11.5
BX120-5CA 5.0 3.0 16.5 16.5
BX120-10CA 10.0 2.0 19.6 19.6
Sensitivity Coefficient 2.06-2.12
6. Strain Acquisition and Calculation Software
6.1 Strain Signal Acquisition Software Panel
…………………………………………………………………………………………………………………
Address Huda Sector23A Gurugram India 122017 www.sensor.in
Beijing Wavespectrum S&T Co., Ltd (Mechanics Lab. In India)
The software may calculate cable resistance compensation of 1/4 bridge and two mode of half
bridge.
…………………………………………………………………………………………………………………
Address Huda Sector23A Gurugram India 122017 www.sensor.in
Beijing Wavespectrum S&T Co., Ltd (Mechanics Lab. In India)
…………………………………………………………………………………………………………………
Address Huda Sector23A Gurugram India 122017 www.sensor.in