Professional Documents
Culture Documents
Autoclave 545
Autoclave 545
Table of contents:
• Autoclave Definition
• Autoclave Parts/ Components
• a. Pressure Chamber
• o b. Lid/ Door
• o c.Steamgenerator
• o d. Vacuum generator (if applicable)
• o e. Wastewater cooler
• Autoclave Principle/ Working o
Introduction
Pressure gauge
• A pressure gauge is present on the lid of the autoclave to
indicate the pressure created in the autoclave during
sterilization.
• The pressure gauge is essential as it assures the safety
of the autoclave and the working condition of the
operation.
Pressure releasing unit/ Whistle
• A whistle is present on the lid of the autoclave is the same
as that of the pressure cooker.
• The whistle controls the pressure inside the chamber by
releasing a certain amount of vapor by lifting itself.
Safety valve
• A safety valve is present on the lid of autoclave, which is
crucial in cases where the autoclave fails to perform its
action or the pressure inside increases uncontrollably.
• The valve has a thin layer of rubber that bursts itself to
release the pressure and to avoid the danger of
explosion.
c. Steam generator/ Electrical heater
• An electrical steam generator or boiler is present
underneath the chamber that uses an electric heating
system to heat the water and generate steam in the inner
and the outer chamber.
• The level of water present in the inner chamber is vital as
if the water is not sufficient; there are chances of the
burning of the heating system.
• Similarly, if the water is more than necessary, it might
interfere with the trays and other components present
inside the chamber.
d. Vacuum generator (if applicable)
• In some types of autoclaves, a separate vacuum
generator is present which pulls out the air from the inside
of the chamber to create a vacuum inside the chamber.
• The presence of some air pockets inside the chamber
might support the growth of different microorganisms. This
is why the vacuum chamber is an important component of
an autoclave.
e. Wastewater cooler
• Many autoclaves are provided with a system to cool the
effluent before it enters the draining pipes.
• This system prevents any damage to the drainage pipe
due to the boiling water being sent out of the autoclave.
Working:
• The autoclave works on the principle of moist heat
sterilization where steam under pressure is used to
sterilize the material present inside the chamber.
• The high pressure increases the boiling point of water and
thus helps achieve a higher temperature for sterilization.
• Water usually boils at 100°C under normal atmospheric
pressure (760 mm of Hg); however, the boiling point of
water increases if the pressure is to be increased.
• Similarly, the high pressure also facilitates the rapid
penetration of heat into deeper parts of the material, and
moisture present in the steam causes the coagulation of
proteins causing an irreversible loss of function and
activity of microbes.
• This principle is employed in an autoclave where the
water boils at 121°C at the pressure of 15 psi or 775 mm
of Hg.
• When this steam comes in contact on the surface, it kills
the microbes by giving off latent heat.
• The condensed liquid ensures the moist killing of the
microbes.
• Once the sterilization phase is completed (which depends
on the level of contamination of material inside), the
pressure is released from the inside of the chamber
through the whistle.
Qualification protocol
Equipment
Description
Calibration Equipment Equipment
-Equipment
requirement control operation
Identification
function
-Utility Requirements
-Component
Specifications
-Component Material
- Lubricants
TABLE OF CONTENTS
1.0 Objective
2.0 Scope
3.0 Responsibility
4.0 User requirement specification
1.0 Objective
User requirements specifications Covers all the needs of the manufacturing department.
2.0 scope
The scope of this qualification document is limited to the design qualification of Autoclave
The equipment shall be used for Sterilization of Media, glasswares, product etc
3.0Responsibility
Roll no: 545
Name
: Pradnesh sandikar
Class: Fympharm(QAT)
Transducer
21 Snitary piping Sanitary TRI‐
CLAMP fittings
22 Piping connections Screwed
Sry. No Specifications Description
30 Utility
Power supply
Material Construction
Parts Material
Chamber Stainless steel 316 L
Lid stainless steel CF8.
Trays Stainless steel 316.
Handle Thermo-insulated.
Safety valve Rubber
Piping system AISI 316L stainless steel.
Insulation fiberglass insulation
D.Y.PATIL. COLLEGE OF PHARMACY, PUNE
DESIGN QUALIFICATION REPORT
Instrument: Autoclave (Instrument No. )
REPORT NO. DYP/QAT/PV/DQ/RP001/2016-17 EFFECTIVE DATE
REF.PROTOCOL
DYP/QAT/PV/DQ/PC001/2016-17 PAGE NO. 1of 9
NO.
INSTRUMENT DETAILS
Name of Equipment Autoclave
Location of the Equipment Pimpri
Equipment Identification No. ADP1166
Model no. 1704
Make
Date of Purchase
TABLE OF CONTENTS
1.0 Objective
2.0 Scope
3.0 Responsibility
4.0 FAT Procedure
5.0 Design qualification approval
6.0 Reference documents
1.0 Objective:
To perform design qualification of Autoclave and assure that the machine is manufactured as
per the URS and it complies with the scope of supply.
2.0 Scope:
The scope of this qualification document is limited to the design qualification of Autoclave The
equipment shall be used for Sterilization of Media, glasswares, product etc.
3.0 Responsibility:
Roll No. 545
Name: Pradnesh P Sandikar
Class and department: FYMpharm (QAT department)
Prepared By Checked By Approved BY
Name & Pradnesh
Designation sandikar,
After the completion of the erection work of the machine, client shall be informed to perform the FAT.
Client shall perform the FAT at the manufacturer site and record all the data in the prescribed FAT document as
per details given below
1. Test criteria
2. Design verification check list
3. Deficiency and corrective action report
4. Pre installation requirement
5. Final report
5.0 Design qualification approval: According to the data collected as a result of this study,
equipment has been properly designed & qualified in accordance with standards and is now available
for process qualification.
Suppliers name Position Signature Date
APIS Technologies Manufacturing
D.Y.PATIL. COLLEGE OF PHARMACY, PUNE
DESIGN QUALIFICATION REPORT
Instrument: Autoclave (Instrument No. )
REPORT NO. DYP/QAT/PV/DQ/RP001/2016-17 EFFECTIVE DATE
REF.PROTOCOL
DYP/QAT/PV/DQ/PC001/2016-17 PAGE NO. 3of 9
NO.
Customer:
Student name Roll No. Signature Date
Pradnesh P 545 3june 2021
sandikar
INSTRUMENT DETAILS
Name of Equipment Autoclave
Location of the Equipment Pimpri
Equipment Identification No. ADP1166
Model no. 1704
Make
Date of Purchase
TABLE OF CONTENTS
1.0 Pre- Approval
2.0 Purpose
3.0 Scope
4.0 Responsibility
5.0 Requalification
6.0 System Description
7.0 Installation Qualification Procedure 8.0
Autoclave Major Component Results.
9.0 Conclusion
10.0 Appendix
Signing of this approval page of Installation Qualification Protocol No. DYPIPSR/ MQ-F / IQ/01,
indicates agreement with the installation qualification approach described in this document. If
modifications to the installation qualification become necessary, an addendum will be prepared and
approved.
2.0 PURPOSE :
2.1 To provide record of the key features of the equipment and components as currently
installed.
2.2 To ensure that there is sufficient information available to enable the equipment to be
operated and maintained safely, effectively and consistently.
2.3 To verify the installation attributes of the equipment name critical to serve the intended
purpose of the equipment.
2.4 To establish the suitability of the draft SOP prepared for the operations of equipments.
2.5 To document the observation for future reference.
3.0 SCOPE:
The scope of this qualification protocol is limited to the installation qualification of
equipment name. This qualification protocol is part of a validation activity.
4.0 RESPONSIBILITY:
Roll No. 542
Name: komal padule
Class and department: Fympharm (QAT department)
Sr.No. Specification
1 Pressure chamber
2 Lid
3 Pressure gauze
4 Pressure releasing unit
5 Safety valve
6 Steam generator
7 Vaccum Generator
LOCATION : pimpri
9. Prepare a deviation report including the justification of acceptance and impact on the
function.
10. Prepare an IQ report.
9.0 CONCLUSION :
The equipment name is qualifying the installation qualification test as per the guidelines
described in this Protocol No. DYPIPSR/ MQ-F / IQ/01. The operation qualification can be
followed now.
POST APPROVAL :
NAME SIGNATURE DATE
Written By : 3 June 2021
Arati ombase
Komal padule
Checked By:
Approved By:
10.0 APPENDIX:
ABBREVIATIONS AN DEFINITIONS:
2. IQ : Installation Qualification
3. QA : Quality Assurance
5. Acceptance criteria : The product , equipment, and / or process specifications nad limits,
such as acceptance quality level and unacceptable quality level, that are necessary for
making a decision to accept or reject.
6. Installation Qualification : Documented verification that the system or subsystem is
installed satisfactorily as per the recommendations of the manufacturers.
7. Validation : Establishing documented evidence that is system does what it purports to do.
8. Revalidation : Repetition of the validation process or a specific portion of it .
INSTRUMENT DETAILS
TABLE OF CONTENTS
1.0 Pre- Approval
2.0 Purpose
3.0 Responsibility
4.0 Requalification
5.0 System Description
6.0 Installation Qualification Procedure
6.1 Training
6.2 Key Functionality
6.3 Sop Verification
6.4 Deficiency (If Any) And Corrective Action Report
7.0 Acceptance Criteria
8.0 Conclusion
Prepared By Checked By Approved BY
2.0 PURPOSE
3.0 RESPONSIBILITY
Roll No.
Name:
Class and department:
4.0 REQUALIFICATION:
Operation qualification to be repeated in case of:
PARTS OF AUTOCLAVE:
8.0 TRAINING:
Title: Date:
A. Purpose:
The purpose of this procedure is to demonstrate that the control panel and other manual operations
of Autoclave.
Chamber steam
Increase the chamber pressure by
will blow off
15% of the working pressure
through safety
6 Working of safety valves
Valve
Increase the jacket pressure by Jacket steam will
blow off through
15% of the working pressure
safety valve
1.
Key functionality Complies
2. Complies
Safety features
3. Operator interface, display, Complies
automation and control requirements
4. Alarms and warnings Complies
Title: Autoclave
Operate the equipment as per the draft SOP and record the details given below:
Checked By:
The operating personnel understand and follow the SOP description (Yes / No):
If yes, Review
No.
Verified By:
Signature:
Date:
D.Y.PATIL. COLLEGE OF PHARMACY, PUNE
OPERATIONAL QUALIFICATION REPORT
Instrument: Autoclave (Instrument No. )
REPORT NO. DYP/QAT/PV/IQ/RP001/2016-17 EFFECTIVE DATE
REF.PROTOCOL DYP/QAT/PV/IQ/PC001/2016-17 PAGE NO.
NO.
10of 9
Date : Date:
Verified by :
Signature:
Date:
Operational qualification should be considered acceptable when all the conditions specified in
various data sheets under section 6.0 have been met.
Any deviation from the acceptance criteria of the specific check point should be reported and
decision should be taken for rejection, replacement or rectification of equipment/ component.
13.0 CONCLUSION:
The Autoclave is qualifying the operational qualification test as per the guidelines described in this
Protocol No. DYPIPSR/ MQ-F / OQ/01. The Autoclave can be tested for its Performance
Qualification as per the Protocol No. DYPIPSR/ MQ-F / PQ/01.
INSTRUMENT DETAILS
TABLE OF CONTENTS
1.0 Pre- Approval
2.0 Objective
3.0 Performance Qualification Test: Procedure
4.0 Performance Qualification Test Results
5.0 Requalification
6.0 Responsibility
7.0 Conclusion
Signing of this approval page of Performance Qualification Protocol No. DYPIPSR/ MQ-F /
2.0 OBJECTIVE:
The objective of this protocol is to collect sufficient data to establish that Autoclave
performs to meet the desired product quality in the consistent manner, when operated as
per standard operating procedure. Performance qualification protocol shall provide the
methodology of qualification studies, formats for recording the observation, criteria of
qualification and a guideline for documentation of the study.
3.0 RESPONSIBILITY
Roll No.
authorized protocol.
➢ The cycle for “Vacuum leak test” in PLC was selected and performed.
➢ The vacuum leak test was performed before and after the validation execution to ensure
➢ The drop in the vacuum shall not be more than 0.013 bar after 10 minutes of hold period.
➢ The Bowie and Dick test cycle in PLC was selected and the test was performed as per the
following parameters.
After completion of the cycle, the test sheet was retrieved and the compliance against the
Acceptance Criteria:
▪ The indicator color shall change from light brown to dark brown / black in the Bowie-Dick test
4.1.3 Heat distribution and heat penetration studies with loaded chamber.
➢ Multi point data logger was used with T- type RTD for heat penetration studies.
➢ Number of temperature sensors were used to check heat distribution in loaded chamber
(including one at drain point), and rest 12 probes were be used to check heat penetration
➢ The temperature sensors were placed with identification number inside the sterilizer
chamber and inside the sterilization load where the steam penetration could be difficult.
The overall rationale for probe placement would be combination of two or more of the
➢ The temperature sensors were placed with identification number inside the sterilizer
chamber and inside the sterilization load as per the temperature sensor location
diagram.
the sterilization load near each temperature sensor kept to check heat penetration, and
also at the locations where sensor cannot be placed but the location is defined as difficult
for steam penetration. ➢ The sterilization cycle was selected and cycle parameters are
verified.
➢ The heat penetration and distribution studies were carried out by using the Multi point
➢ After completion of sterilization cycle the data logger was stopped and the nonsterile
➢ The Biological indicators were taken out from the load and sent to
Microbiological Lab.
➢ The Temperature profile from the Data logger and temperature recorder was taken out.
➢ The F0 value for each temperature sensor of data logger was checked. Calculation
of F0 value is based on the given formula. Check the results against the
➢ After getting the reports of biological indicator check it for compliance against
➢ Temperature sensors which are used are be calibrated after completion of activity (post
calibration) to ensure that the temperature measurement system is accurate and precise.
Acceptance complies/
Sr. Instrumen
Sr. Instrument Acceptance criteria doesn’t
No t
No criteria complies
Data
1 Logger + 0.20C complies
1 Data 0
+ 0.2 C
Logger
2 RTD +0.25% complies
sensors
2 RTD +0.25%
sensors
Sr. Run
no No. Test Observations Acceptance criteria Inference
Uniform dark
1. 1 black color Complies
observed
Uniform dark Bowie-Dick test kit should show
2. 2 black color uniform dark brown/ black color complies
observed development
Uniform dark
3. 3 black color complies
observed
D.Y.PATIL. COLLEGE OF PHARMACY, PUNE
PERFORMANCE QUALIFICATION REPORT
Instrument: (Instrument No. )
REPORT NO. DYP/QAT/PV/PQ/RP001/2016-17 EFFECTIVE DATE
REF.PROTOCOL
DYP/QAT/PV/PQ/PC001/2016-17 PAGE NO. 5of 9
NO.
Observations
Steril Max. Max.
izer Min. Max. Min. Equilibr Max. fluctuation difference
Run temp temp F0 ium deviation of temp on a between Result
No. value time from average RTD sensors two RTD
temp sensors
122. 32. complies complies complies
1 121 30 complies
9 4
complies complies complies
2 121 123 32 30 complies
3 121 122.9 32.6 30 complies complies complies complies
6.0 REQUALIFICATION:
6.0 CONCLUSION: Finally, conclusion should be drawn based on the results of above test
and documented.
Prepared By Checked By Approved BY
Name & Pradnesh p Sandikar
Designation