Load Calibration Procedure For TUC-6 - WF
Load Calibration Procedure For TUC-6 - WF
Load Calibration Procedure For TUC-6 - WF
X = 00,0 t/h Press & to enter the OFF LINE menu from the main
S1 = 6356,12 t program.
Password Enter 5 digit password and press to access the OFF LINE
Enter password : menu
X X X X X
Analog Input
Press to proceed the calibration. If the calibration menu in
Calibration bottom
CALIBRATION LOAD LHS counts (auto) w.r.t load cell mV & RHS counts determined
Keep the test
0000 => 0000 d the calibration percentage. LHS counts will increase w.r.t to
weight on the knife
increment of the mV from Tare mV. After study state of
edge.
counts enter the “d” counts in RHS w.r.t to calibration
CALIBRATION LOAD percentage. Press to complete the calibration & close the
2756 => 15000 d calibration menu.
Note: 1% calibration = 20 d counts
SYSTEM DATA Press to run the main program, while main program
MAINPROGRAM menu in bottom line.
ISSUE: 1.0 - 10/03/17 Please go through below course of action for repeatability check & zero correction.
REPEATIBILITY CHECK IN STOP CONDITION
BB = 75 % Keep the test weight & ensure the “BB” value in line with
Ta = 0000 Hz calibration percentage. (Exp: here showing 75%)
BB = 00 % Remove the test weight & ensure the “BB” value Is 0%.
Ta = 0000 Hz
ZERO CORRECTION TUC display the zero correction error. It is auto accepted if
ERROR = -1.5% the error is less than the zero limit settings in configuration.
Password
Enter five digit password and press
Enter password :
X X X X X
Dl COUNT : 03567
Note down this values for future reference.
SPAN COUNT: 06538