Product Data

Remote Controlled RF table for Flat Panel Detector

Movements
Tilting Motorized, -90°/+90°.
Two user selectable speeds: 4.5 or 6.5°/sec (speed can be
changed at installation).
Automatic stop in horizontal position
Elevating tabletop Motorized, max run 800 mm (31,5”)
Speed 25 mm/sec (1 inch/sec) (can be changed at installation)
Transversal tabletop movement Motorized, 320 mm = ± 160 mm (12,6” = ± 6,3”), with soft
start/stop for maximum patient comfort.
Speed 50 mm/sec (2 inches/sec).
Automatic centering
Longitudinal tapletop movement The complete patient coverage is guaranteed by the large
travel of the tube / detector assembly, not requiring any
longitudinal tabletop movement for the maximum patient
comfort and safety
Tube column - detector assembly Motorized, 1600 mm (63”).
movement Progressive variable speed: the movement starts slowly for
accurate adjustments and is progressively increased for quick
run over long distances.
Maximum speed 200 mm/sec (7,9 inches/sec)
Patient exploration 2030 mm (80”) thanks to the movement of the tube/detector
only, without patient repositioning
Tube column angulation Motorized, ± 40°.
Speed 11.2°/sec (can be customized at installation)
Automatic centering of target organ during oblique projections
in fluoroscopy.
Possibility to perform oblique exposures at both edges of the
tabletop.
Automatic centering
SID (Source to Image Distance) Motorized, 1100-1800 mm (43” – 72”) with continuous
adjustment
Speed max 41 mm/sec (1 inch/sec)
Tube rotation Manual, -90°/+180°.
Stop: -90°/-50°/-40°/0°/+40°/+50°/+90°/+180°
Focal spot to floor distance (with 515-2115 mm (20,27”- 83,27”) without column tilting
table in vertical position)

Product data Apollo DRF – Rev.8 Page 1 of 10


Curve Surface Tabletop
Maximum patient weight 284 kg (626 lbs), without any movement limitations
Dimensions 2379x750 mm (93,7’’x29,5’’)
Radio-transparent area 2214x650 mm (87,2’’x25,6’’)
Tabletop height from floor 600 - 1400 mm (23.6” - 55,1”)
(measured at tabletop center with
standard under-floor installation
plate)
Surface Concave, smooth, without gaps and ridges
Depth from tabletop center to 68 mm (2,7’’)
tabletop border
Materials Plastic (standard), Carbon fiber (option)
X-ray attenuation Plastic: ≤0.7 mm Al @100 kVp, HVL=2.7mm Al
Carbon fiber: ≤0.3 mm Al @100 kVp, HVL=2.7mm Al
Minimum distance between central 390 mm (15,4”) both sides
X-ray beam and tabletop edges
Minimum distance between 175 mm (6,9”)
detector and tabletop edge
Skin to Detector distance 72 mm (2.8”) (central, minimum)

Flat Surface Tabletop


Maximum patient weight Up to 284 kg (626 lbs), without any movement limitations
Dimensions 2356x738 mm (92,7”x29”)
Radio-transparent area 2214x554 mm (87,2”x21,8”)
Tabletop height from floor 618 - 1418 mm (24.3” - 55.8”)
(measured at tabletop center with
standard under-floor installation
plate)
Surface Flat
Material Plastic (standard)
Carbon fiber (option)
X-ray attenuation Plastic: ≤0.7 mm Al @100 kVp, HVL=2.7mm Al
Carbon fiber: ≤0.3 mm Al @100 kVp, HVL=2.7mm Al
Minimum distance between central 390 mm (15,4”) both sides
X-ray beam and tabletop edges
Minimum distance between 175 mm (6,9”)
detector and tabletop edge
Skin to Detector distance 91 mm (3.6”) minimum

Product data Apollo DRF – Rev.8 Page 2 of 10


Flat Surface “open” Tabletop
Maximum patient weight Up to 230 kg (507 lbs) distributed along the tabletop
length, without any movement limitations.
Dimensions 2356x738 mm (92,7”x29”)
Radio-transparent area 2214x554 mm (87,2”x21,8”)
Tabletop height from floor(measured 614 - 1414 mm (24.2” - 55.7”)
at tabletop center with standard
under-floor installation plate)
Surface Flat
Material Carbon fiber
X-ray attenuation ≤0.3 mm Al @100 kVp, HVL=2.7mm Al
Minimum distance between central X- 390 mm (15,4”) both sides
ray beam and tabletop edges
Minimum distance between detector 175 mm (6,9”)
and tabletop edge
Skin to Detector distance 91 mm (3.6”) minimum

Detector housing
AEC measuring chamber Predisposed
Grid Stationary, optimized for use with flat panel, delivered as
standard. Possibility to configure it as oscillating.
Cover material: carbon fiber
Ratio: 12:1, 80 lines/cm (203 lines/inch)
Multi-Grid system Motorized insertion of two grids optimized for different examination
protocols:
• Grid 1: Focal 120cm (43”) for general rad exams
• Grid 2: Focal 180cm (72”) for chest exams
• No Grid: for pediatric and extremity exams
Grid movement Depending exclusively on the anatomic procedure in function of the
patient size, without required input by the user.

Compressor
Compression Motorized, remote controlled
Compression force 3 kg (6,6 lbs) to 15 kg (33,1 lbs) step 0.5 kg (1,1 lbs)
Useful distance between 138 mm ÷ 388 mm (5,4” ÷ 15,2”),
compressor and tabletop 142 mm ÷ 392 mm (5,5” ÷ 15,7”) for Open version
Compression cone Removable, made of radiotransparent plastic
Compressor parking When not in use, the compressor is automatically parked behind
the tube for maximum patient safety
Securities Both software and hardware securities. Quick unhook in case of
blackout.

Product data Apollo DRF – Rev.8 Page 3 of 10


Collimator
Collimation Square and rectangular
Number of beam shutters 4 pairs of shutters, including near-focus shutters
Shutters material Fe + Pb
Adjustment Automatic with SID compensation, microprocessor controlled
“Hold” function The position of the diaphragms set during fluoro can be frozen
when switching to exposure
Field coverage 43x43 cm @ SID=100 cm (17”x17” @ SID=39,37”)
Maximum collimator filtration 0.5 mm Al eq @ 100 kV, HVL = 2.7 mm Al
Stray radiation ≤45mR/hr @ 150kVp, 350W
Light source LED source
Targeting Laser centering device (only for collimator with automatic filter
selection)
Additional filtration Optional, automatic and manual filtration.
Values of added filtration: 1mm Al + (1 mm Al or 0.1 mm Cu or
0.2 mm Cu)
Total filtration ≥ 2.7mm Al eq. @ 100kV
(tube+housing+collimator)

Tomography
Tomographic technique Linear tomography with Arc-Plane movement, fully electronic
(without connection bar)
Table positions for tomography Tomography can be performed in every table position
Angles 7°, 20°,30°, 45°
Speeds 4 speed, from 11.2 to 22.4°/s (can be changed at installation)
Direction Bi-directional, left-right or viceversa, user selected
Layer height Electronic adjustment
0 ÷ 350 mm, 1mm step (0 ÷ 13,8”, 0,04” step)
Automatic sequences Sequences of tomo images with automatic layer position increase
and bi-directional movement, without stopping the movement
between exposures
Source to Image Distance 1140 mm (44,9”)
Tomography speed (in sec.) Angle 1st speed 2nd speed 3th speed 4th speed
7° 0,6 0,5 0,4 0,3
20° 1,8 1,3 1,1 0,8
30° 2,5 1,9 1,7 1,2
45° 3,7 2,8 2,5 1,7

Stitching (option)
Stitching function The stitching function allows to automatically reconstruct one
full image from a set of single images obtained from a
patient scanning movement. This function is typically used
for spine and extremity reconstruction
Number of steps 2,3, or 4 based on the anatomic program
Step length 30 cm
Reconstructed image dimension 2 steps: 43x60 cm, 3 steps: 43x90 cm, 4 steps: 43x120 cm
Directions Right-left /left-right
Focus distance Changeable at installation
Number of programs 60 (10 programs for each direction/n° of steps combination)

Product data Apollo DRF – Rev.8 Page 4 of 10


Functions
Movement orientation selection It is possible for the user to select the direction of the
movements associated to the main joystick (column
scanning and transversal tabletop) according to table view
or monitor view
Controls on Touch Screen console Table movements, Collimator, Tomography, Compressor,
Fluoroscopy parameters
Tableside controls Table tilting, tabletop transversal and lift, column scanning
and tilting, SID, collimator lamp
Controls on collimator Pushbuttons for shutters open/close
X-ray control Double footpedal for fluoro and exposure.
Pushbutton on generator console for exposure.
Two-steps pushbutton for exposure (option)
Automatic positioning According to the predefined examination procedure, the
system automatically sets: table tilt, column angulation,
longitudinal detector position, transversal tabletop position,
source-detector distance, collimated area, grid type,
additional filtration (if present)
Applications • Fluoroscopy
• Radiographic exposure
• Tomography
• Angiography
Projections Perpendicular, oblique, on gurneys or wheelchairs, on chest
stand (option)
Tomoscopy The organ remains centered when taking oblique
projections during fluoro
Fault indication and memory Fault conditions are visualized with codes and text
messages on the display.
An internal memory stores the history of faults and
equipment conditions

Product data Apollo DRF – Rev.8 Page 5 of 10


Touch Screen console
Touch Screen Color graphic LCD TFT display
Joysticks Five joysticks for control of Apollo DRF’s functions and
movements
Function keys Emergency red pushbutton.
PREP and RAD pushbuttons for the access to the
generator via the remote-controlled table, or to directly
access the generator during the service
Size 12”
Display resolution 800 x 600 pixels
Brightness > 300 cd/m2
Contrast (optimum viewing angle) 450 : 1
Colours 65.536
Viewing angle ↑↓← → >35° / >55° />60° />60°
Aspect ratio 4:3

Joysticks on the Touch Screen console


For Apollo DRF version, the joystick functions on the Touch Screen console are arranged as follows:

Key Functions
A Collimator regulation
B • left/right: column rotation
• up/down: compressor movement
C • left/right: -90°/+90° tabletop tilting
• up/down: SID regulation
D • left/right: auto centering/ special function selection (tomography,
stitching)
• up/down: tabletop vertical movement
E • left/right: tube horizontal movement
• up/down: tabletop transversal movement
(can be inverted according to the selected orientation)
Note: A, C and E joysticks have 8-way movement.

Product data Apollo DRF – Rev.8 Page 6 of 10


Accessories
2 handgrips Standard. They can be fixed in any position along the
tabletop
Footrest Standard. It can be fixed along the table top with stops
every 50 mm (2”) for curved tabletop or 100mm (4”) for
flat tabletop
Head and shoulder rest Option. It can be fixed in any position along the tabletop
Couple of urological/gynecological leg Option. They can be fixed in any position along the tabletop
supports
Compression band Option. It can be fixed in any position along the tabletop
Lateral cassette support Option (only with optional overhead tube support).
It can be fixed in any position along the tabletop
In-room console Option. Complete in-room console on mobile trolley, used to
control table movements. With digital display.
Multi-functional in-room console Option. Additional multi-functional console on mobile trol-
ley. It can be used to control the table movements and the
entire digital processing from within the examination room.
It is composed by: high brightness monochrome monitor
(single or dual), Apollo console with full functionalities, key-
board, mouse, 10 meters cable and fluoro/exposure foot-
pedal.

Safeties
Collision All movement are software controlled to avoid collision of any part of the
equipment with room floor, ceiling or walls. Room size can be set by software
Single fault A dedicated microprocessor checks all the operating conditions of the
equipment in real time and stops the operation in case of a single failure that
might cause unwanted or excessive movements or radiation
High voltage cables HV cables are fully integrated in the column and are completely invisible and
protected with covers
Compressor When not in use, the compressor is automatically parked behind the tube for
parking maximum safety

Electrical features
Power supply voltage Three phase 380-400 Vac ±10%
Three phase 415-480 Vac ±10% (for USA version)
Frequency 50/60 Hz
Absorbed current Approx. 7A @ 380-400 Vac
Approx. 6A @ 415-480 Vac
Absorbed power 6 kVA
Operational heat dissipation 1570 BTU/h (460W) assuming a 20% duty cycle
Equipment type and classification Class I with type B applied parts
according to IEC 60601-1
Degree of protection according to IEC IP00
60529
Operating mode Continuous

Product data Apollo DRF – Rev.8 Page 7 of 10


Standards and regulations
0051
CE symbol grants the product compliance to the European Directive for Medical Devices 93/42/EEC
and its revised versions as a class IIB device

Additional standards for units installed in North America:


CFR 21
Code federal regulation. Sub chapter J
CMDR
Canadian Medical Device Regulation

Environmental conditions
Operating conditions Temperature: from +10° to + 40° Celsius (50° F to 104° F)
Pressure: from 700 to 1060 hPa
Humidity: from 30% to 75%
Conditions for transport and storage Temperature: from -20° to +70° Celsius (-4° F to 158° F)
Pressure: >630 hPa
Humidity: up to 95% non condensing

Mechanical features
Size (L x P x H) Weight
Table 2420 x 1920 x 2075 mm 1100 kg (2420 lb)
- See drawings for further details - (95,3” x 75,6” x 81,7”) (Table only, without
tube, accessories, ca-
bles, floor plate)
Electronics Cabinet 520 x 550 x 1950 mm 145 kg (320 lb)
(20,5” x 21,6” x 76,8”)
Touch Screen console 357 x 356 x 123 mm <5,2 kg (11,5 lb)
(14,1”x14”x4,8”)
Minimum ceiling height for 90/90° tilting 2600mm (8’ 4.5”)

Product data Apollo DRF – Rev.8 Page 8 of 10


Dimensions (mm & Inches)

NOTE: Above measures are valid for installation with standard under-floor plate

Product data Apollo DRF – Rev.8 Page 9 of 10


Dimensions for Open Version (mm & Inches)

NOTE: Above measures are valid for installation with standard under-floor plate

Note: Products are continuously under review in the light of technical improvements. The actual specification may
therefore be subject to improvement or modification without notice.