Evaluation of Calibration Measurement Capability of Multifunction Calibrator 10 V At: Lab

You might also like

Download as xlsx, pdf, or txt
Download as xlsx, pdf, or txt
You are on page 1of 3

EVALUATION OF CALIBRATION MEASUREMENT CAPABILITY OF MULTIFUNCTION CALIBRATOR

Details Of UUC:- Instrument Name: Multifunction CalibraID No:


Make: FLUKE Range: 0 ~ 10 V Sr No:
Model: L.C: 0.1 Type: Digital 2
STANDARDS / EQUIPMENT USED FOR CALIBRATION
1. Instrument Name: Multifunction Calibrator 8. Model:
2. Range: 0 ~ 1000 V 9. Sr. No:
3. Resolution / L.C: 0.01 10. ID. No:
4. Unc in Measurement: 0.035 V 11. RDG %+D:
5. Drift / Accuracy : 0.05 V ( RDG %+D)
6. Validity / Tracebility:
7. Certificate No:

Environmental Conditions: Avg. Temp: 25.3 °C Avg. Humidity: 54 % RH

TYPE:A Maximum Range: 1000 Reading Taken on: 5 Type A Results are in
Step:1
Test Readings X1 X2 X3 X4 X5 n 5
5.4 5 5.3 5.2 5.1 n-1 4

Mean Ẍ 5.2 V
Step:2 Xi Mean Xi -Mean (Xi -Mean)2
Standard Deviation 5.4 5.2 0.2 0.040000
5 5.2 -0.2 0.040000
S.D = 0.158114 V 5.3 5.2 0.1 0.010000
5.2 5.2 0 0.000000
5.1 5.2 -0.1 0.010000
Total 0.100000
Step:3

Type A= 0.070711 V
Step:4
Degree of Freedom for Type A
DOF= 4
TYPE:B
Step:5
Uncertainity from Certificate (Master) b1= 0.0175 V
Uncertainity due to Accuracy of master(Master) b2= 0.028868 V
Uncertainity due to Test Resolution (Test) b3= 0.028868 V
Step:6
Combined Uncertainty Uc = 0.083504
Step:7
Effective Degree of freedom Veff= 7.779451
Step:8
Expanded Uncertainity
U= 0.19757 V
CTION CALIBRATOR 10 V AT: LAB
Date: ###
Location: LAB

1+2D V

Air Pressure: --

e A Results are in V

0.05 Digital 50% 2


Analog 20% 5

2.366