CISA S.p.A.
AND PRESS..
REFERENCE TABLE OF STEAM
TEMPERATURE
PresSure
RESSURE
Temperature
C kPa
100.0 93.1
101.0 98 2
102.0 102.7
103.0 107 2
104 O 111.8
105.0 116.4
106 0 121 2
107.0 126.0
108.0 131 0
109-0 13 0
110.0 141 1
111.0 146.4
112.0 151.7
113.0 157 1
114.0 1&2.7
115.0 168.3
116 0 174 1
117.0 180.0
118.0 186.0
119.0 192.1
120.0 198.4
121.0 204 8
122.0 211 3
123.0 218.0
124.0 224 9
125.0 231.8
126.0 231.0
127.0 246 3
128.0 253.8
129.0 261.4
130.0 269 2
131.0 277.2
132.0 285.4
133.0 293.9
134 0 302.5
135.0 311.3
136 0 320.4
137.0 329.7
138.0 339.2
139.0 349-0
1400 359.1
C I S a )
6-
SAFETYDEVICES
empe
erature probes.
if
Device preventing the door openingtoduring the cycles for liquids in open and closed containers,
is above or equal 90C (in the autoclaves for the sterilizations of the liquids in
the termperature
bottles)
OPERATING DEVICES
17-SET-POINT OF SAFETY AND
and operating components and their settings.
The following tables show safety
Safety
Set-point
Component
SS3 3 Bar
Generator safety valve
See attached hydraulic scheme
300°c
Heating elements thermostat TC3
See attached hydraulic scheme
Pag.16
CSA Autoclaves
CISA S.p.A.
- User manual Vere:
ersion 2.1 /08
(CIsa)
ill
cycle will
be signalledaand
1.8 A L A R M S
the
functioning
possible i printed.d.The
evel cotds e
during be
intervenes and it will
The alarmsthat
canbe alam type intervened
instructions:
the sequent
phase and follow the
showtheor reset the alarms
VISualize
22/10/08
:MENU: 08:36:49
Cycle Nr:00100
Values
->Alarms (?)
confim with
arrows, and
select "Alarms (?)" by the
From this
screen
: : : : : A L A R M : : : : : :
->OPEN RESET
Press to exit.
: ::CODE: : : : : :
***
OK
By pressing the two keys and select the code and confirm with
The under score will shift
under the next "*", until the 4 number are selected. Then o
CISA Autoclaves Pag.17
(CIsa) CISA S.p.A. - User manual Version 2.1/08
The 'RESET function is available only after the end of the cycle, after the foliow display
CYCLE COMPLETED
ATTENTION
HIGH TEMPERATURE
Canceled cycle
This alarm activates when the cycle is canceled by using the procedure described on the touchn
Screen.
Door opening
This alarm activates when the stop position doesn't indicate the corect closing of the door while the
cycle is in progress.
Maximum temperature
This alarm activates during the sterlization phase, when the chamber temperature rises above the
set point (sterilization temperature + 3°C).
Emergency temperature
CISA Autoclaves Pag.18
CISA S.p.A. -
User manual
CIsa
x c e e d s 140°c
ture eexceeds
temperatun
140°C.
Version21/08
activates
when the
This alarm
the
the maximum
maxir
Maximum pressure
when the
th pressure
exceeds
value for the
This alarm
activates
p r e s s u r e (3,5
bar, 450 kPa). chamber
Communication error
when there
Is non
communication between
between dient.
display and plc
activates
This alarm
Unsaturated steam
This alarm activates when the presSSure doesnt correspond to the
e temperature
temperature values
(shown in the reference table).