Subject: The New software and hardware of BC-20s is released
RECORD IDENTIFIER
CONTENT
Old V01.11.00.2302
New V01.12.00.2403
2.The new version of the software modifies the timing sequence for solving the
problem of the Rotation or Elevator Motor Error
3.The new version of the software adds ‘Drain Fluidcs’ and ‘Drain Preheat Bath’ functions
4. The new version of the software adds a 【save】button in the system configuration
interface.
2.Hardware change
2.1 Change the inlet of the diluent to a metal joint to avoid interference
Step2: Connect the Dilent tube with the new metal connector, ensure the tightly
connection.
2.2 Solve the screen blinking problem by changing the screen line
Tips :New wire is compatible with old wire, the part number does not change
Part Number:
009-004471-00 Small LCD Signal Wire