Professional Documents
Culture Documents
Efecto Del Autoclave y Energía
Efecto Del Autoclave y Energía
Efecto Del Autoclave y Energía
Group 1, intensity
values were obtained
after each autoclave
cycle for the first
seven cycles and
then every fifth cycle
from cycles 10 to 25.
We measured the
light intensity at the
same intervals for
the tips in Group 2,
but we polished
these tips only once
(immediately before
the first autoclaving
cycle) using a com-
mercially available
cleaning kit (Optical
Maintenance Kit,
Demetron Research
Corp.). The tips in
the third group were
cleaned after every
fifth autoclaving
cycle. We measured
NO. OF AUTOCLAVE CYCLES the light intensity
values of this group
after every fifth au-
toclaving cycle, as
well as after polish-
Figure 1. The graph illustrates the influence of repeated autoclaving on transmitted ing at similar cycle
energy in light-curing tips. Boxed points indicate no tip polishing (Group 1). stages.
Diamond points indicate one initial polishing only (Group 2). The T-bar represents
- I standard deviation. To test the influ-
ence of autoclaving
on light intensity,
Demetron Research Corp.). We Corp.). We tested three as-re- we performed a two-factor, re-
established a calibration scale ceived tips for the intensity val- peated measures analysis of
between the dial-face intensity ues before and after repeated variance on the intensity values
readings from the analog meter autoclaving procedures. The recorded for specimens from
and the measured millivoltage curing light tips were auto- Group 1 and Group 2. Post-hoc
drop values, which enabled us claved at a temperature of means comparison was per-
to convert the measured milli- 270 F for four minutes followed formed using Tukey's Honestly
voltage readings into intensity by a 14-min drying cycle in a Significant Difference Test.
values. Using the control tip, we standard autoclave (no. 3013, The ability of the polishing
monitored the light source be- AMSCO Int.) and were pack- tip to recover the level of pre-
fore and after each intensity aged in autoclavable envelopes autoclaved light intensity after
reading to ensure that it provid- before sterilization. a series of autoclave cycles with
ed the desired control value. In addition to establishing a subsequent polishing was tested
The light tips used were con- control group, we designed the using multiple regression anal-
ventional 8-millimeter-diameter study to include three test ysis. The difference between in-
glass bundles with a 30-degree groups, each with three light tensity values of autoclaved and
curve at the end of the tip (no. tips. All of the light tips were polished treatment values was
20941 N, Demetron Research autoclaved for 25 cycles. In studied as a function of the
number of auto-
clave cycles.
The data were
logarithmically
transformed to
provide a linear
relationship. All
statistical test-
ing was per-
formed at the
significance
level of P = .05.
RESULTS
As Figure 1 il-
lustrates, the
transmitted
light energy
quickly decreas-
es after only a
few sterilization
treatments
when the tips
are not pol-
ished. After
only one cycle,
the intensity
dropped drasti- Figure 2. The effect of repeated autoclaving and subsequent tip polishing on transmitted
light intensity. The red bar indicates pre-polishing intensity. The blue bar indicates in-
cally from the tensity reading after polishing. The T-bar represents 2 I standard deviation.
pre-autoclaved
value of 100 per-
cent to 74 percent. The intensi- similar numbers of cycle treat- a significant influence on the dif-
ty continued to decrease but at ments. ference in intensity values
a slower rate after each succes- The data from Group 3 speci- (P = .0003), while no particular
sive sterilization cycle. After mens indicated that after the tip acted differently from the
only three cycles, the transmit- initial five autoclave cycles, the others (each P > .3). All the re-
ted intensity decreased to 50 light intensity dropped to 50 per- covered, polished tips produced
percent of its pre-autoclaved cent of its original reading. After intensity values similar to 100
value. The ANOVA indicated the tips were polished, the light percent of the as-received value.
that the autoclaving treatment intensity returned to its original Magnification of the as-re-
itself had a significant effect on level (Figure 2). After five more ceived tip x50 showed no gross
the light intensity (P = .0008), cycles, the intensity dropped contamination (Figure 3). After
as did repetition of the auto- again, but to only 80 percent of five autoclave cycles, magnifica-
claving treatment (P = .017). its original value. The original tion at the same level revealed
Figure 1 also shows that pol- value again was completely re- a dark, noncontinuous film over
ishing the tips only once before covered after polishing. The in- the surface (Figure 4). After pol-
autoclaving (Group 2) decreased tensity value after each subse- ishing, the appearance of the
the rate at which light intensity quent set of five autoclaving autoclaved curing tip was iden-
declined with repeated auto- cycles interrupted by polishing tical to that of the as-received
claving. This decrease is demon- remained near 95 percent of the state (Figure 3).
strated by the relative differ- maximum transmitted value.
ence in transmission values The regression analysis indicat- DISCUSSION
between the two graphs' having ed that the number of cycles had Our data show that the light in-