Professional Documents
Culture Documents
Quality Assurance in External Beam Radiation Therapy: Goran K. Svensson
Quality Assurance in External Beam Radiation Therapy: Goran K. Svensson
Radiation Therapy
Goran K. Svensson, Ph.D.
Editor’s Note:
This is the fifth of a series of articles on Radiation Treatment Planning based on the Basic
Physics Symposium of the American Association of Physicists in Medicine that was presented at
the 73rd Annual meeting of the RSNA.
Address reprint requests to G.K. Svensson. Ph.D., Joint Center for Radiation Therapy and Department of Radiation
Therapy, Harvard Medical School, 50 Binney St., Boston, MA 02115
I 70 RadioGraphics January,
#{149} 1989 Volume
#{149} 9, Number I
Svonsson Radiation Therapy: Quality Assurance
can be facilitated by the following consider- tional intercompamisons are therefore neces-
ations (1). samy. Theme are various ways of performing
All chambers used for therapy beam cali- constancy checks. In those institutions with Co-
brations must have a calibration factor trace- 60 machines, one can make use of the inher-
able to the National Bureau of Standards ently reproducible Co beam and, as shown in
(NBS). This can be accomplished either by hay- Figure 2, one can build a fixed jig for position-
ing the hospital standard instrument calibrated ing the dosimeter in a reproducible manner.
by NBS directly, or by having the instrument
calibrated by one of the five AAPM accredited As seen in Figure 3, such a setup has pro-
calibration laboratories. The hospital standard duced interesting constancy check results,
instrument will then be used within the hospital here demonstrated on a Baldwin-Farmer ion-
for the calibration of field instruments. These ization chamber. The variations of this cham-
field instruments are used for the calibration of ben lie within ±1%. Seasonal variations of the
the therapy beams. This procedure assures chamber response can be seen, however. This
that all radiation equipment has its output nor- suggests that the reproducibility of these mea-
malized to the national calibration laboratory. surements is better than I % and that the vamia-
Specific state or NRC regulations may apply to tions are due to changing environmental con-
calibration procedures. ditions in the calibration room. Somewhat less
The accuracy of a dosimetry system is sub- precise, but still adequate in most situations, is
ject to change without obvious indications. In- a special check source containing Sm-90 (see
stitutional constancy checks or multiinstitu- Figure 4) into which the chamber is inserted.
Figure 3
Results of constancy checks versus
time, using the setup shown in Figure 2.
I I I I I I I I I I I I I
Constancy Checks
BF 2505 I 3 Serialal56
1.01 60Co Therapy Machine
U)
C 00 0
0
0.
U)
1.00 0 08 : 00 0:0
V
a)
N
D
E 0.99
0
z
0.98
0.97
O.Ji:_1:c1 , I I I ‘ I ,
Time
I 72 RadloGraphics January,
#{149} 1989 Volume
#{149} 9, Number I
Svensson Radiation Therapy: Quality Assurance
Prior to machine calibration, the dosimeter dose, i.e. 5 cm, to avoid the effect of spurious
,
Source
Figure 6
Typical calibration geometry for high energy x-ray beams
_%. . . - ., -.
.-_J :
! ..Y.’.’;’
2-, ,
0-
L #{149} I #{149} I #{149}
-4 -2 0 2 4
DAILY CHANG’E(%I CENTRAL AXIS
x
Figure 7 Figure 8
Variations of x-ray beam calibrations The Flatness and symmetry changes owing to a shift in
broaden distribution represents a situation in which the flattening filter on beam steering
the normally sealed monitor chamber was open to
atmospheric temperature and pressure variations.
I 74 RadioGraphics January,
#{149} 1989 Volume
#{149} 9, Number I
Svensson Radiation Therapy: Quality Assurance
The beam uniformity should be monitored fre- It is very important that action levels be es-
quently. There are several commercial and tablished for the Quality Assurance personnel.
home built instruments that can be used to de- Without action levels, it is likely that errors
temmine the dose at off-axis points. Beam uni- found during the Quality Assurance tasks me-
formity can then be calculated. A generic ap- main unconnected. Such levels can be deter-
proach to off-axis dose determination is illus- mined within each department. As an exam-
trated in Figure 9, which shows five detectors pie, one can specify that if the central axis
embedded in a plastic block. The detectors constancy checks deviate from a posted val-
are lined up to measure dose at off-axis points ue by less than ±5%, no action is needed by
as well as the central axis value. Detectors the technologist (the physicist will have tighter
numbers I and 5 should be at least 25 cm action levels). If the deviation is greaten than
apart. This will permit monitoring the beam ±5% but less than ±10%, a physicist on engi-
neem should be notified immediately. Treat-
ments may continue for the nest of that day,
however. If the deviation exceeds ±10%, the
treatments must stop immediately, and a
physicist on engineer must be notified.
Figure 9 AO/
U /0.
TABLE I
Constancy checks:
1 Dose/monitor
. unit,
X-rays Daily +1-3%
2. Dose/mm, Co6O Monthly +/-3%
3. Depth dose Monthly +/-2%
4. Beam uniformity Weekly +/-3%
5. Dose monitor or
timer constancy Annual Not Applic.
I 76 RadioGraphlcs January,
#{149} 1989 Volume
#{149} 9, Number I
Svensson Radiation Therapy: Quality Assurance
Figure 10 #MEASUREMENTS
Distribution of surgically implanted
clips around the mean location (X)
-ii
SET UP C8mm
BREATHING
OTHER MOTION
10
I I I
_7-::5 _3 -101 3 5 7
(X-Ajrnm
MANTLE STUDY
(I)
(I)
o\o
4
D/STANCE(rnrn)
Figure 1 1
Treatment-to-treatment variations for patients treated with a mantle
field The data were derived from daily portal nadiographs taken before
and after the treatments. The figure illustrates the percent of observations
that exceed a certain deviation from the expected value.
Cl)
0
a)
>
a)
I 78 RadloGraphics January,
#{149} 1989 Volume
#{149} 9, Number I
Svensson Radiation Therapy: Quality Assurance
PENUMBRA WIDTH, 80% to 20% gantry axis with the isocenter, (c) stability of
SINGLE FIELD,8MV X-RAYS the gantry arm under rotation, (d) couch mo-
10cm x iOcm,DEPTH=lOcm
tion, and (e) focal spot position.
0 0.5 1.0
Figure 13
Electron beam
The apparent width of a field edge versus setup pre-
cision
A B
Collimator
ALIGNMENT AIDS
Gantry Gantry
‘A
Figure 16
The double exposed film resulting from the lead block test device (12)
I 80 RadioGraphics January,
#{149} 1969 Volume
#{149} 9, Number I
Svensson Radiation Therapy: Quality Assurance
References
1. American Association of Physicists In Medicine Task surements. Determination of absorbed dose In a pa-
Group 2 1 . Physical aspects of quality assurance in ra- tient Irradiated by beams of x-ray and gamma radla-
diatlon therapy. Report 13. New York: American Insti- tion In radiotherapy procedures. Report 24. WashIng-
tute of Physics. 1984. ton, DC: International Commission on Radiation Units
2. WrIght AE, Boyer AL, eds. Advance in radiation treat- and Measurements, 1980.
ment planning. American Association of Physicists In I I . Johanson KA. Studies of different methods of ab-
Medicine Medical Physics Monograph no. 9. New York: sorbed dose determination and a doslmetrlc Inter-
American Institute of Physics. 1983: 244-268. comparison at the Nordic Radiotherapy centers. The-
3. Starkschall G. ed. Proceedings of a symposium of qual- sis. Goteborg, Sweden, 1982.
Ity assurance of radiation therapy equipment. Kansas 12. Lutz WR, Larsen RD. Bjamgard BE. Beam alignment
City, Mo: American Association of Physicists In MedI- tests for therapy accelerators. Int J Radlat Oncol Biol
cine, 1982. Phys 1981; 7:1727-1731.
4. American Association of Physicists In Medicine Task 13. DosImetry of x-ray and gamma-ray beams for radla-
Group 21 . A protocol for the determination of ab- tion therapy In the energy range of 10 key to 50 MeV.
sorbed dose from high energy photon and electron NCRP report no 69. Bethesda, Md: National council on
beams. Med Phys 1983; 10:741-77 1. Radiation Protection and Measurements, 1981.
5. American Association of Physicists In Medicine. Code 14. Rablnowltz I, Broomberg J, Goltein M, McCarthy K,
of practice for x-ray linear accelerators. Med Phys Leong J. Accuracy of radiation field alignment In dm1-
1975; 2:110-121. cal practice. nt J Radlat Oncol BIoI Phys 1985;
6. AmerIcan Association of Physicists in Medicine and BI- I 1:1857-1867.
ology Task Group 28. Radiotherapy portal Image qual- 15. Shukovsky U. Dose, time, volume relationshIps In squa-
Ity. Report 24. New York: American Institute of Physics. mous cell carcinoma of the supraglottic larynx. AJR
1987. 1970; 108:27-29.
7. Wlzenberg MJ. ed. Quality assurance In radiation ther- 16. Stewart JG, Jackson AW. The steepness of the dose
apy: A manual for technologists. Chicago. American response curve both for tumor cure and normal tissue
College of Radiology. 1982. injury. Laryngoscope 1975; 85:1107-1111.
8. CommIttee for Radiation Oncology Studies. Report to 17. Svensson GK. Quality assurance In radiation therapy:
the Director of the National Cancer Institute on criteria Physics efforts. Int J Radlat Oncol Biol Phys 1984;
for radiation oncology in multidisciplinary cancer man- 1O(suppl 1):23-29.
agement: The blue book. 1981. 18. verhay LV, Goltemn M, McNulty P. Munzenrlder JE, Suit
9. Hendrickson FR. Radiation treatment planning: The HD. Precise positioning of patients for radiation thera-
physician’s role. RadioGraphics 1988; 8:987-991. py. Int J Radiaf Oncol Blol Phys 1982; 8:289-294.
10. International Commission on Radiation Units and Mea-
I 82 RadioGraphics January,
#{149} 1989 Volume
#{149} 9, Number I