From the main screen, Long-press on the button will make the oximeter display e No flammable substance, overtop

No flammable substance, overtop or lower temperature and humidity existed in


Settings Pages as shown in Figure 3.3.1 or Figure 3.3.2. Selecting "AIm Setup" or operation conditions.
"Sounds Setup" and long press will make this two pages display alternately.
4.5 Troubleshooting

Table 4.5 Troubleshooting


Settings Settings
Am Setup Sounds Setup Problems I Possible Reason Resolutions
Sp02 AIm Hi 99
AIm on Oxyhemoglobin or , 1. Finger is not plugged correctty. 1. Retry by plugging the finger.
Sp02 Alm Lo 94
Beep off PR AIm Hi 130 heart rate can not 2. Patient's perfusion is too low to be 2. Try sorne more times, if you can
Restore OK PRAlm lo 50
l- be shown normally. measured. make sure about no problem existing
Exit Exit
in the product, Please go to a

Figure 3.3.1 Figure 3.3.2 hospital timely for exact diagnosis.

The ltems in "AIm Setup" are: Oxyhemoglobin or 11. Finger might not be plugged deep 1. Retry by plugging the finger.
heart rate is shown enough. 2. Ty not to move, Let the patient
ltems Default Opti ons Oesc ripti on
unstably. 2.Finger is trembling or patient's body [k eep calm.
AIm on on, off Turn on ar off the sounds when the value if out of
is in movement status.
the limit set in Figure 3.3.2.
The oximeter can't 11. Power of batteries might be 1. Please replace batteries.
Beep off on, off Set Pulse Beep on or off. go into the working inadequate or not be there at all. 2. Please reinstall the batteries.
Restore OK OK,Blank Reset al settings as default value. state. 2. Battelies might be instaUed incorrectly.13. Please contact with local
Exit N/A NIA Retum to the Main Scre en 3. The Oximeter might be damnaged. customer service center.
The Items in "Sounds Setup" are: The screen are 1. The product is automatically standby 11. Normal.
ltems De fault Opti ons Desc ription suddenly off. or sleep when no signal is detectad 2. Replace the batteries.
Sp02Alm Hi 99 51-99 longar than 1 O seconds.
The upper limit of Sp02.
Sp02 AIm Lo 2. Power quantity of the batteries is
94 50-98 The lower limit of Sp02.
exhausted.
pR AIm Hi 130 50-250 The upper limit of PR.
PR AIm Lo 4.6 Disposal
50 50-245 The lower limit of PR
+l- + ,- To avoid contaminating or infecting personnel, the environment or other equipment,
Increase or decrease the number.
make sure you disinfect or decontaminate the device appropriately before disposing of it
Exit NIA NIA Retum to the Main Screen.
in accordance with your country's law fer equipment containing electrical and electronic
The upper limit cannot be lower than the lower limit.
parts.
3.4 Turn off the Oximeter
Sectio n 5 Specifi catio n
The oximeter will turn off automatically after 10 seconds after removing the finger out
of the device. Physical Characteristics
Machine Dimensions: 58mm (L) x 30mm (W)x 30 mm (D)
ectio n 4 Mainten an ce
Machine Weight -approx: 28g (not including battery)
4.1 Cleaning Classification
Switch off the power and take out the batteries before cleaning, Cleaning exterior Anti-electric Shock Type: lnternally powered equipment
surface (OLED display screen included) of the unit with a dry and soft cloth. Use 75% Anti-electric Shock Degree: Type BF equipment
density of medical alcohol to clean the surface and use dry fabric with little alcohol to EMC: Group 1 Class B
avoid alcohol permeates into the device.
Mode of operation: Continuous Operation
4.2 Disinfection
Enclosure Degree of ingress protection: IP22
Disinfectig the machine after using by the palien! if multiple palien! use the machine 3 1P22 means shell of this product can with stand the water dropping to the surface
in the hospital. Use 75% density of medical alcohol to clean the surface that contacting when the shell deviate 15 degree from horizontal surface.
with the patient. Power
CAUTION: Don't use strong solvent. Fer example, acetone. Intern al: 2xAAA1.5V battery
Power Consumption: 30mA (Normal)
CAUTION: Never use an abrasive such as steel wool or metal polish. Environmental
Operating Tempera tura: 5C to 40C
CAUTION: Do not allow any liquid into the product, and do not immerse any parts of Storage Temperature. - 20 C to 55°C
the device into any liquids. Relative Humidity . 15% to 85% non-condensing
Electronics Parameters
CAUTION: Avoid pouring liquids on the device while cleaning. Para meter Value
Sp02 (Oxygen saturation) 35- 100%
CAUTION: Don'! remain any cleaning solution on the surface of the device. PR (Pulse Rate) 25- 250 bpm
4.3. Warranty PI (Perfusion Index) 0- 30%
The host product' design lite is 2 years, and 1 years warranty. Under normal Sp02 (Oxygen saturation} 1%
circumstances, the malfunction of the product during the warranty period (from the date Resolution PR (Pulse Rate) 1 bpm
of purchase) should be sent back to the company fer maintenance, and our company is PI (Perfusion Index) 0.1%
responsible fer ali maintenance costs (users should cover the freight themselves). 2% (80% - 100% )
Outside the warranty period, our company shall charge sorne maintenance fee (users Sp02 (0xygen saturation) 1 3% (70% - 80% )
should cover the freight themselves)nerodoctna
s brok i
endon ana issent back tor Measure Accuracy.
maintenance. The battery is beyond the scope of the warranty. lf you have the purchase
and sale contract, the costs of the maintenance shall be in accordance with the PI (Perfusion lndex) 1%0 - 2 0%)
purchase and sale contrae! execution. Besides, it is recommended that users should
Unspecified (20% - 30% )
use them no more than five years. And over the using lite, the using risks may increase
dueto the equipment' aging.
• Am., accuracy is a statistical calculation of the difference between device
measurements and reference measurements. Approximately two-thirds of the device
4.4 Maintenance measurements fell within +/- Arms of the reference measurements in a controlled study.
• Replace the batteries timely when battery indication is low. Clean surface of the
Pulse Oximeter befere it is used in diagnosis fer patients.
• Remove the batteries inside the battery cassette if the Oximeter will not be
operated fer a long time.
• lt is better to preserve the product in a place where ambient temperature is -20-
55C and humidity is 15%-85%
• Regular inspection to make sure that no obvious damage existed to affect the
safety and performance of device.
The information you
FINGER PULSE OXIMETER Caution:
should know to protect The important information you
Note:
the equipment from should know

USER'S MANUAL V1.3 possible damage

Section 1 Safety Section 2 lntroduction

1.1 Instructions for the Safe Operation and Use of the Pulse Oxim eter 2.1 Display Param eter and Brief Device Description

• Do not attempt to service the pulse oximeter Only qualified service personnel SpO2(Functional oxygen saturation), is the amout of oxyhemoglobin expressed as
should attempt any needed internal servicing. a percentage of the Functioning Hemoglobin. Functioning Hemoglobin is capable of

• Do not use the oximeter in situations where alarms are required. carrying oxygen and includes oxygenated hemoglobin (HbO2) and deoxygenated
• Prolongad use or the patient's condition may require changing the sensor site hemoglobin (Hb).

periodically. Change sensor site and check skin integrity, circulatory status PR(Pulse Rate), measured in beats per minute(bpm), is the frequency of heart
and corree! alignment every hour. beats.

• SpO2 measurements may be adversely affected in the presence of high PI(Perfusion Index) is the ratio of the pulsatile blood flow to the non-pulsatile static
ambient light. Shield the sensor area (with a surgical towel, or direct sunlight, blood in peripheral perfusion. Perfusion index is an indication of the pulse strength at

for example) if necessary. the sensor site.


wilcauseinterference.
Tiefoilowin@reasons Plethysmogram The amount ofl
ight absorbed by the varying quantities of arterial
High-frequency electrosurgical. blood changes with the pulse beats, the waveform is named Plethysmogram. This

■ Placement of a sensor on an extremity with a blood pressure cuff arterial waveform and its variation is used for assigning signal integrity, physiological and
catheter, or intravascular line. artifactual changes such as perfusion changes, dysrhythmia, motion artifact, and
■ The patient has hypotension severe vasoconstriction severe anemia or electrical interference.
hypothermia. Bargraph, Indicate the beat of pulse, displays the amplitude changes with each
■ The patient is in cardiac arrest or is in shock. pulse.

■ Fingernail polish or false fingernails may cause inaccurate SpO2 readings. The Pulse Oximeter, based on all digital technology, is intended for noninvasive
spot-check measurement of functional oxygen saturation of arterial hemoglobin
1.2 Warnings
(SpO2). Advanced DSP algorithm can reduce the influence of motion artifact and
WARNING: EXPLOSION HAZARD Do not use the oximeter in a flammable improve measurement accuracy of low perfusion.
atmosphere where concentrations of flammable anesthetics or other materials may 2.2 Intended Use
occur. This product is suitable tor the clinic, oxygen bar, sports health(using it betore or after
sports, not advise using them during the movement), and community health care, etc.
WARNING: Do not throw batteries in fire as this may causes them to explode.
2.3 Contraindication
lt is not for intensive care and person whose finger is injured.
WARNING: Do not use the pulse oximeter in an MRI or CT environment.
Section 3 lnstallation, Setup, and Operation
CAUTtON: Keep the operating environment free of dust, vibrations, corrosive, or 3.1 Description
flammable materials, and extremes of temperature and humidity.
)2

CAUTION: Do not operate the unit if it is damp or wet because of condensation or spills. k.------ Perfusion lndex

Avoid using the equipment immediately after moving it from a cold environment to
warm, humid location.
ªI ~~ rla
.
Battery Indicator
Pulse Rate Button

WARNING: Do not attempt to recharge normal dry-cell batteries, they may leak. And
may cause a fire or even explode.

CAUTION: Neve, use sharp or pointed objects to operate the front-panel switches.
Sp0% a PR bpm uudq84 D 0ds 11111111111 a
CAUTION: The battery must be taken out from the battery compartment if the device 98.s 99'86 ó
2g$
fo%
will not be used to, a long time.
Figure 3.1.1 OLED display
Note: when battery power is at lowest level, the battery capacity indicates symbol
CAUTION: The device shall only be used if the battery cover is closed.
of "0 "in OLED, remind users of replacement of battery.
3.2 Operation
CAUTION: The battery must be proper disposed according to local regulation after
their use. 3.2.1 Install battery
lnstalling two AAA batteries into battery cassette in correct polarities and cover it
WARNING:The oximeter is not intentad to use in ICU, because the device's alarm does (as Figure 3.2.1.1)
not meet requirements of EN 60601-1-8.
WARNING: Do not attempt to
recharge normal alkaline batteries,
1.3 Definitions and Symbols they may leak and may cause a fire
or even explode.
Symbol Description Sym bol Description

#l Type BF Equipment
el Information of manufacture,
including name and address
Figure 3.2.1.1
3.2.2 Turn the Pulse Oximeter on
Put one of fingers into rubber hole of the oximeter (it is best to put the finge,
thoroughly) with nail surface upward (as Figure 3.2.2.1 ), then releasing the clamp.
Refer to the instruction 1
[sN] ] Serial NO Press the button, oximeter will go into the working state.
1 manual /booklet
~

When the end-user


wishes to discard this The information you should
product, it must be sent know to protect patients and Figure 3.2.2.1
Warning:
to separata collection medical staff from possible 3.2.3 Read corresponden! data from display screen(as Figure 3.1.1 ).
facilities for recovery injury 3.3 Parameter setting
and recycling There are two ways to operate the button according to the pressing time:
long-press is longer than half a second and short-press is shorter than half a second.
Short-press is used to select a item by moving a • to the line of this item, and long-press
is used to change the item's value, status, open a new page or make it take effect.