Professional Documents
Culture Documents
Service Manual PHOENIX NCC-3300 v15.05
Service Manual PHOENIX NCC-3300 v15.05
Service Manual PHOENIX NCC-3300 v15.05
PHOENIX NCC-3300
Automated Hematology Analyzer
Service Manual
V15.05 eng
CONTENTS
COPYRIGHT .............................................................................................................................................. 3
CHAPTER 1 SYSTEM DESCRIPTION .......................................................................................................... 6
1.1 Front View ..................................................................................................................................... 6
1.2 Rear View ...................................................................................................................................... 7
1.3 Principles of Operation ................................................................................................................. 7
1.3.1 Electrical Impedance Method ................................................................................................ 8
1.3.2 WBCs Counting Principles ...................................................................................................... 8
1.4 WBC Test Principle and Differential .............................................................................................. 9
1.5 RBC Test Principles ...................................................................................................................... 10
1.5.1 RBC Total Number Test Principle ......................................................................................... 10
1.5.2 Test Principles of RBC Indexes ............................................................................................. 11
1.6 Platelet Test Principle ................................................................................................................. 11
1.7 HGB Colorimetric Method .......................................................................................................... 12
CHAPTER 2 NOTICE ............................................................................................................................... 14
2.1 External Factors .......................................................................................................................... 14
2.1.1 Voltage ................................................................................................................................. 14
2.1.2 Electromagnetic Interference .............................................................................................. 14
2.1.3 Temperature ........................................................................................................................ 14
2.2 Location Requirement ................................................................................................................. 15
2.3 Notices for Startup ...................................................................................................................... 15
2.4 Blood Samples Collection and Test ............................................................................................. 15
CHAPTER 3 CIRCUIT ............................................................................................................................... 16
3.1 Introduction ................................................................................................................................ 17
3.1.1 Electrical System .................................................................................................................. 17
3.1.2 Switched Mode Power Supply (SMPS) ................................................................................. 18
3.1.3 ARM Board ........................................................................................................................... 18
3.1.4 FPGA Board .......................................................................................................................... 19
3.1.5 Analog Amplifier Board ........................................................................................................ 19
3.1.6 Valves and Motors Drive Board ........................................................................................... 20
3.1.7 AVDD Board ......................................................................................................................... 21
page 2
3.1.9 Print Server Board ................................................................................................................ 22
3.1.10 Reorder .............................................................................................................................. 22
3.2 Electrical System Fault ................................................................................................................ 23
3.3 Parameters Detection Principle .................................................................................................. 25
CHAPTER 4 FLOW SYSTEM .................................................................................................................... 26
4.1 Dilutor ......................................................................................................................................... 26
4.2 Sample Cup ................................................................................................................................. 27
4.3 Vacuum Pump ............................................................................................................................. 29
4.4 Sampling Mechanism .................................................................................................................. 30
4.5 Solenoid Valve ............................................................................................................................. 31
4.6 Rolling Pump ............................................................................................................................... 31
4.7 Vacuum Chamber ........................................................................................................................ 32
4.8 JWLM Board ................................................................................................................................ 32
4.9 LMS Board ................................................................................................................................... 33
4.10 Function of Flow System Valves ................................................................................................ 35
4.11 Flow System Fault ..................................................................................................................... 36
CHAPTER 5 TEST .................................................................................................................................... 40
5.1 System Calibration ...................................................................................................................... 40
5.2 Gain Adjustment ......................................................................................................................... 42
5.2.1 WBC, RBC Gain Adjustment ................................................................................................. 43
5.2.2 HGB Voltage Adjustment ..................................................................................................... 43
5.2.3 PLT Gain Adjustment ............................................................................................................ 44
5.2.4 Vacuum Adjustment ............................................................................................................ 44
5.3 System Check .............................................................................................................................. 44
5.3.1 Motor Check ......................................................................................................................... 44
5.3.2 Valve Check .......................................................................................................................... 45
5.3.3 System Status Check ............................................................................................................ 46
5.4 Internal Calibration ..................................................................................................................... 47
CHAPTER 6 SOFTWARE UPGRADE AND ONLINE ................................................................................... 48
6.1 ARM Board Software Upgrade .................................................................................................... 48
6.1.1 Preparations ......................................................................................................................... 48
6.1.2 Upgrade Process .................................................................................................................. 48
6.2.2 Online Procedures ................................................................................................................ 50
CHAPTER 7 COMMON FAULTS .............................................................................................................. 50
Page 3
Copyright
Declaration
NeoMedica owns the copyright of this unpublicized issued manual, and has right to
handle as secret information. This manual just used as reference for operation,
reserved, prohibit copy and transmit any content of this manual without NeoMedica
written agreement.
NeoMedica doesn’t make any form of guarantee for this manual, including (but not
for certain purpose. NeoMedica has NO responsibility for the error included in this
Manufacturer’s Obligation:
NeoMedica is only responsible for instrument security, reliability and capability under
following condition:
WARNING:
If each hospital or institution that is responsible for using this instrument cannot
Page 4
realize a set of satisfactory service procedure, it will cause abnormal
Guarantee
Manufacture Technics and Materials
technical and material problem within one year from shipping date if it is under
Free Service
NeoMedica’s obligation under this guarantee does NOT include freight and other fee.
NeoMedica is not responsible for direct, indirect, ultimate damage and delay caused
This guarantee is not applicable for following items: improper use; damaged
analyzer or the one which has NOT been maintained; NeoMedica original S/N label
product.
If following occur, NeoMedica is not responsible for the security, reliability and run
Contact with NeoMedica sales company to get the right of return, inform S/N which
marked on outer pack. Analyzer with illegible label is NOT acceptable. Please
Version: V15.05
NeoMedica DOO
Tel: +381 (18) 573-820, +381 (18) 573-606, +381 (18) 533-935
Web: www.NeoMedica.rs
Email: info@NeoMedica.rs
3 2
1. Aspiration Probe
Aspirate samples.
2. RUN Key
Press the RUN key to startup the aspiration probe and then analyze specimen
only in the screens of main menu or Quality Control. At other screens, the RUN
key is invalid.
3. Recorder
Print the test result.
4. Touch Screen
10.4 inch LCD.
Page 7
1.2 Rear View
Figure 1-2
1.3 Principles of Operation
NCC-3300 is a multi-parameter, automated hematology analyzer. It can
Page 8
display 21 parameters and 3 histograms. Analyzer adopts electrical impedance
method for WBC, RBC and PLT test and colorimetric method for HGB test.
1.3.1 Electrical Impedance Method
Electrical impedance method is based on the non-conductivity of blood cells.
When the blood cells in Diluent pass through the ruby aperture, resistance will
change, base on that we can get the counting and volume of WBCs.
1.3.2 WBCs Counting Principles
NCC-3300 automated hematology analyzer uses two channels for
counting, which means WBCs and RBCs are in different counting chambers.
Figure 1-3
Inner and outer electrodes of constant current source are located in front
certain concentration of cell suspension, and front chamber is filled with Diluent.
Conductivity of cells is lower than that of Diluent. They are being treated as
relatively poor conductors. When a cell particle goes through the ruby aperture,
of particle that produced it. Under the effect of negative pressure, a certain
pulses.
conversion.
1.4 WBC Test Principle and Differential
Along with WBC parameters, analyzer could also give the WBC histogram
which can display the average volume of specific cells population, cells
Add a certain amount of dilution and Lyse into WBC sample cup. Lyse can
make RBC dissolved and WBC dehydrated to form "film covers core", so that
WBC, analyzer divides distribution range of WBC volume (35~450 fL) into 256
histogram (See Figure 1-4). The ordinate indicates the relative quantity of
WBC (rel.no) and the abscissa indicates the volume of WBC (fL).
Page 10
Figure 1-4
According to the volume, WBCs handled by Lyse can be subdivided into three
LYM 35—90 fL
GRAN 160~450 fL
1.5 RBC Test Principles
1.5.1 RBC Total Number Test Principle
RBC test principle is alike to WBC test principle. In sample cup which is similar
to that of WBC, with the effect of negative pressure, a certain amount of cells
Analyzer can work out total number and average volume of RBC according to
the size and height of pulse. Meanwhile, it can also get a RBC volume
volume and the percentage of cells which have the same volume.
Page 11
Figure 1-5
Under normal conditions, we can ignore WBCs due to the proportion between
1.5.2 Test Principles of RBC Indexes
HCT can be worked out by dividing the product of MCV and RBC by 10.
According to relative algorithm, analyzer can get MCH, MCHC though RBC,
MCV and HGB. Red Cell Distribution Width (RDW) can be figured out by
detecting RBC number and the difference of RBC size so as to reflect the
heterogeneity of RBC volume. RDW can reflect the degree of RBC size
1.6 Platelet Test Principle
Platelet (PLT) and RBC are being tested in the same sample cup. The
threshold (Figure 1-6). Data of platelet are being saved in 64 channels in 2~30
fL interval.
Page 12
Figure 1-6
PDW can be worked out though histogram. MPV is the arithmetic mean
volume of platelets which are shown by the curve in histogram. MPV of normal
people has a nonlinear negative correlation with platelet number. PCT is got
1.7 HGB Colorimetric Method
HGB. Add Lyse into the diluted sample, RBC will be dissolved and hemoglobin
sample cup through the monochromatic light with a wavelength of 540nm and
then compare it with the result in blank state to get the hemoglobin
concentration (blank state refers to the state that only has Diluent in sample
cup). Instruments can test automatically, then calculate and print out the result
(in g/L).
Page 13
K is a constant.
1.8 Parameters
NCC-3300 generates following 21 hematologic measurements with
Chapter 2 Notice
2.1 External Factors
2.1.1 Voltage
To ensure normal working and stability of data test, analyzer should equipped
2.1.2 Electromagnetic Interference
Because the signals gathered by analyzer are very weak, external interference
should be connected with grounding wire to make interference signals into the
2.1.3 Temperature
Analyzer working temperature is 15 ºC to 35 ºC. Low temperature will affect
reagents and cause test error. The most common thing is that the value of
WBC and HGB are too high, because the dissolving speed of Lyse is slow with
low temperature. The value of PLT is on the low side, since platelets aggregate
2. Waste container should be placed on the ground, and not on the same
2.3 Notices for Startup
1. Check tubes connectors after the power supply and reagents are
2. Check whether there are abnormal smell, sound and picture. If there are
3. Check whether the picture displayed and program initialization are normal.
2.4 Blood Samples Collection and Test
Test modes are divided into Whole Blood Mode and Pre-diluent Mode.
3. Whole Blood Mode test: Place the anticoagulant tube with blood sample
under the sample aspiration probe, then press “RUN” for counting.
Pre-diluent Mode test: Drain Diluent into the sample tube through probe
and then inject 20μL peripheral blood into it for mixing. Place the tube
NOTE: When collect peripheral blood, avoid squeezing out the tissue fluid with
puncture is needed for peripheral blood collection. Wipe out the first drop of
Chapter 3 Circuit
The circuit system is composed of switched-mode power supply (SMPS), ARM
board, FPGA board,valves and motors drive board, analog amplifier board,
print server board, button board, AVDD board, thermal printer (recorder), LMS
Figure 3-1
Page 17
3.1 Introduction
3.1.1 Electrical System
Provide a stable DC power supply for analyzer, there is a triple power supply:
12V Power Supply: Provides power for valve drive, thermal printer and
AVDD Board.
5V Power Supply: Provides power for ARM board, FPGA board, print
3.1.3 ARM Board
ARM board is the control center of analyzer. The software is based on Linux,
which is responsible for data processing, print server and FPGA board control.
FPGA board is mainly used for software logic control and data collection,
provides parameters for ARM board and executes orders. See the Figure
bellow.
3.1.5 Analog Amplifier Board
Analog amplifier board is used for weak signal amplification and processing,
then adjusts it to be the right signal to the FPGA board for data conversion.
3.1.6 Valves and Motors Drive Board
Valves and motors drive board is the executive circuit unit for solenoid valves
AVDD board is the secondary power supply, which provides 12V power to
switched-mode power supply for conversion and lower noise and stable power
supply for analog board. Output voltage is +12V、-12V and 100V. The quality of
AVDD board is very important to cell signals. If it is damaged, analyzer can not
work normally.
Button board is for the shortcut keys on front panel. If one of the shortcut key
gets stuck, other shortcut keys cannot be used, and . the “RUN” Key is out of
order.
Print server board is for executing print order of recorder and external printer.
3.1.10 Reorder
Recorder uses thermal paper for printing. Thermal surface of the thermal paper
3.2 Electrical System Fault
Phenomenon:
Screen goes black after startup, indicators do not work and analyzer has no
reaction.
Troubleshooting:
1. This may result from the circuit protection of SMPS due to the voltage or
current overshoot during startup. Turn off the analyzer and restart it after
30 seconds
2. SMPS does not work, check whether the fuse has burned off. If so, replace
Phenomenon:
Screen goes black or white after startup. Analyzer has no action, but power
damaged.
Troubleshooting: Turn off the power supply. Replace FPGA board due to the
Troubleshooting: Open the sample shielding case, check whether the sample
no sample. At this moment, check whether there are problems with flow
Phenomenon:
Results of WBC, RBC and PLT are 0, but HGB value and counting time are
normal.
AVDD board 100V Power supply. Replace constant current source or AVDD
board.
Troubleshooting: AVDD board ±12V power supply has no output. Open the
shielding case of AVDD board and check whether the ±12V power supply
replace it.
Page 25
3.3 Parameters Detection Principle
Figure 3-12
Page 26
Chapter 4 Flow System
Figure 4-1 shows the whole framework of flow system behind right side door.
Figure 4-1
4.1 Dilutor
Dilutor is used for counting, rinse, prime and blood dilution at startup. It also
provides Diluent and power for cleaning flow system. Circuit board provides
DC24V to motor.
Dilutor consists of small syringe, Lyse syringe, Diluent syringe, motor, seal ring
Installation of dilutor motor differs from that of other motors. Motor is above the
Lyse syringe and Diluent syringe are controlled by the same motor. And the
4.2 Sample Cup
Sample cup is the sensor part of counting. It is the forefront detecting unit of
data collection.
Functionally, it consists of inner and outer electrode, front and back chamber,
sample cup through dilute conductive liquid. Loop resistance will change if
cells pass through ruby aperture. Cells in different size will produce different
calculated.
colorimetric analysis on blood sample with Lyse through lighting and receiving
4.3 Vacuum Pump
Vacuum pump is the power unit for creating negative pressure. Its working
voltage is DC12V. Fluid section connector has one air outlet and one air inlet.
When working, vacuum pump is driving the suction film by internal rotary motor
to make air in from inlet and out from outlet. (the influent liquid in the course of
4.4 Sampling Mechanism
Sampling mechanism controls the probe for sampling, dilution and rinse etc..
Figure 4-6
Page 31
4.5 Solenoid Valve
with tubes to form the whole flow system and controlling the liquid flow
Figure 4-7
4.6 Rolling Pump
Rolling pump with DC12V is for mixing blood sample in sample cup by bubbles
epoxy resin. It provides negative pressure for flow system under the force of
vacuum pump.
There are two connectors on the upper of vacuum chamber: one is connected
within this range, control board will provide drive voltage for vacuum pump to
start working until the pressure reaches the specified value; the other one is
There is an outlet at the end of vacuum chamber. It is for exhausting air and
4.8 JWLM Board
JWLM board is for detecting the liquid level of Diluent and Lyse in glass tube.
Connect electrical system and flow system before use. Switch on the power
and check whether the voltage of “VCC” is 5V without connecting with Diluent
and Lyse. If the voltage is normal, detect the voltage of UP” ,“DOWN”,”EZ” and
Page 33
“HEM”.
4.9 LMS Board
liquid is 4.8±0.2V, and for the one without liquid is 2.9±0.1V. Deviation of
optocoupler parameter points and contamination of glass tube inwall etc. will
WBC, RBC and PLT. LMS board has 2 channels: one is WBC channel, the
other one is RBC and PLT channel. Each channel is compose of 1 metering
tube and 2 optocouplers. Before counting, V15 and V17 open to let air go into
WBC, RBC metering tube in LMS board for emptying liquid. When analyzer
starts counting, V15 and V17 close, liquid passes metering tube through ruby
aperture. Meanwhile the liquid column in metering tube starts falling. When it
passes though the top optocoupler, comparator outputs start counting signal;
counting signal.
V5 Back flush Back flushing with air produced by rolling pump
V6 Discharge air and liquid Drain Waste of vacuum chamber and establish
vacuum
V12 Control Diluent addition Control Diluent addition of WBC, RBC cup
V16 Mixing (WBC) Mix sample with recoil air produced by rolling
pump
V18 Mixing (RBC) Mix sample with recoil air produced by rolling
pump
Page 36
4.11 Flow System Fault
Troubleshooting:
2. Position switch fault causes motor cannot stop working normally and
comes out abnormal sound. Check whether the switch is well connected
4. Motor is corroded by liquid and its internal units get stuck. Disconnect the
Troubleshooting:
1. Vacuum pump does not work, or the valve under the sample cup is
2. There are foreign bodies at the bottom of sample cup or the valve under
those of other valves, and then test them again. If there is no problem with
valve, there may be something wrong with the drive part of the circuit
board.
Troubleshooting:
1. Low vacuum or tube leakage, adjust vacuum and check flow system.
3. The set time of calibration is very different from the actual time. Correct it.
Page 37
4. Optocoupler voltage of LMS board is too low, which cause false alarm.
Troubleshooting:
1. Rinse and prime WBC/RBC cups with probe Detergent or remove the ruby
for cleaning, and check whether the voltage of the LMS board is normal.
2. If the liquid column of LMS board does not go down, it indicates the ruby
aperture has serious clog. At this moment, the test time may be 0.0 second.
3. If the test time is long, it indicates the ruby aperture has slight clog. Please
Troubleshooting:
2. The connector connected with sample aspiration probe has air leakage;
Troubleshooting:
1. The sample cup is dirty. Clean sample cup or prime it with probe Detergent
3. Keep analyzer away from the instruments which are with strong magnetic
wire.
heated up.
Page 38
Phenomenon: High PLT value
Troubleshooting:
1. There is blood left on sample aspiration probe inwall or the sample cup is
program.
2. Keep analyzer away from the instruments which are with strong
Troubleshooting:
1. Small motor of dilutor cannot work normally. Check whether the motor is
2. Check whether the small syringe has leakage or air leakage. Please
replace it if it is.
3. Small motor cannot work results from the problems of its circuit drive.
Phenomenon: All test results are too low in Whole Blood Mode.
Troubleshooting:
1. Sample in small syringe is not enough or the small syringes has air leakage.
2. Tube connector between sample aspiration probe and small syringe has air
Troubleshooting:
2. Lyse cannot dissolve blood cells well due to the low temperature. Please
4. Rolling pump cannot work and mix sample, or the intensity of rolling pump
Troubleshooting:
1. Valve on the LMS board for emptying air is not closed well. Nip the tube
between the valve and LMS board to check whether the liquid column still
dropping. If does, it indicates the valve is not closed well; if not, there is no
Troubleshooting:
2. Whether WBC cup indicator is dark or not on. Please replace indicator or
WBC cup.
Troubleshooting:
1. Diluent syringe does not work. Friction of leader is too large, please apply
grease to it.
Troubleshooting:
1. Diluent syringe does not work. Friction of leader is too large, please apply
grease to it.
5.1 System Calibration
Click ”Func”— “Sev”, input code ”1999” and click “OK” to enter System
Operator can test motors, set motor steps and WBC, RBC test time etc. in
System Calibration screen. Only the Items in Table 5-1 are those need to be
5.2 Gain Adjustment
Input “1999” and click “Gain Adjust” on top left corner to enter the screen
shown as below.
Page 43
5.2.1 WBC, RBC Gain Adjustment
Test with control and check the gain of WBC and RBC. Adjustment is not
necessary if it is within the acceptable range. If not, input current gain value in
Test with control again to check whether the gain has been corrected.
5.2.2 HGB Voltage Adjustment
HGB_BACK is about 4V. Adjust the voltage as following: correct the value in
“OK” on lower left corner for saving. The larger the channel gain is, the lower
NOTICE: PLT gain has been adjusted strictly before delivery. Generally it
5.2.4 Vacuum Adjustment
Input value in box and click . Vacuum pump
will start pumping. When finished, vacuum value will be shown in box.
NOTICE: Vacuum has been tested strictly before delivery. Generally it
5.3 System Check
5.3.1 Motor Check
At Motor Check screen, operator can check if motors are in normal condition.
Click the item wanted to check, then the result will be shown. Click “Back” to
Press “Func”—“sev.”, input the code “2006” to enter Motor Check screen.
Operator could test MA, MB, MC,MD motors and P1, P2. P1 is for vacuum and
P2 is for rolling pump. Click corresponding test item button to check its
condition. This operation is only for test, operator could not correct the motor
steps here.
5.3.2 Valve Check
Click “Valve Check” on top left corner of System Check screen to enter
valve with that of other valves and test the faulted valve through other drive
circuit. If the valve still does not work, itself is damaged; If not, the problem
5.3.3 System Status Check
Click “System Status Check” on top left corner of System Check screen to
Operator could not correct parameter value in this screen. Value displayed is
5.4 Internal Calibration
Click “Func”—“Sev” and input the code “2008” to enter Internal Calibration
2. Application program
6.1.2 Upgrade Process
1. Copy the application program into the root directory of U disk. Other
2. Plug the U disk into the 4th USB port. See Figure 6-1.
3. Click “Func”—“Sev”, input “9999” in prompt dialog box and then click “OK”.
4. Read-write process will last for about 5 seconds. A dialogue box will pop
complete upgrade.
step 2-4.
unsupported U disk. Analyzer may not read some U disks, please replace
U disk.
Page 49
6.2 Online
Analyzer is able to transmit data to computer by connected with computer.
Install online program in computer allows operator to process, save and print
data.
Because of the COM ports of rear panel and computer are both male
(DB9), please use the serial line whose terminals are both female (DB9).
And the internal connection of the serial line should be as following picture
The rest ports are not necessary to be connected. Serial line whose
terminals are both female and internal connection is just as above picture
2. Online program
Page 50
6.2.2 Online Procedures
1. Connect analyzer with computer through the COM port on rear panel of
3. Set “Auto Trans” on. Hexadecimal and HL7 are available transmission
modes for operator. Ensure that the transmission mode is in accord with
4. Computer is able to receive the data from analyzer after each test
Chapter 7 Common Faults
Fault Cause