Professional Documents
Culture Documents
Fridlund-Leberer1976 Article AnInexpensiveIntegratedCircuit
Fridlund-Leberer1976 Article AnInexpensiveIntegratedCircuit
An Intracranial stimulation unit which offers precise control of stimulus duration. frequency, and
current IS de~cribed. The. uni~ generates a square wave stimulus output utilizing very inexpensive, easily
constructed integrated circuitry.
Commercially available stimulators which allow the generates the square-wave pulse train that constitutes
experimenter to control pulse train frequency, the intracranial stimulus. This arrangement is fully
duration. and amplitude may cost several hundred capable of generating frequencies from 10 Hz to
dollars or more.' This report describes an integrated 10 kHz. depending on the selection of a suitable CF.
circuit stimulator that permits control of the above The square wave output from Pin 3 of IC2 varies
parameters. yet can be constructed for under $25.00. from ground to +V. Such a unidirectional square
The chief constraint ofthe proposed circuit is that the wave has been found to cause electrolytic tissue
stimulus waveform is not selectable; the pulse train damage at the intracranial electrode site (Mickle.
provided is a fixed 50% duty cycle biphasic square 1961). In the present circuit. the unidirectional dc
wave. More elaborate waveforms. such as the biphasic (direct current) square wave is transformed into an ac
pulse-pair advocated by Lilly (1955). require more (alternating current) waveform. This is achieved by
elaborate circuitry. However. the 50% duty cycle referencing the Pin 3 output not to ground. but to
square wave employed in the present design has two +V 12 via a voltage divider circuit composed of two
advantages when compared with shorter pulses: it 2.2-kQ resistors. The stimulus at the electrode is thus
allows the use of lower voltages to achieve required a true ac stimulus with voltage varying from +V /2 to
coulomb thresholds at the electrode site (Mickle. -V /2. Additionally. the voltage divider circuit serves
1961); and. it facilitates computation of peak current to current-limit the output of IC2 to under 5 m A.
reaching the electrode site during stimulation. This rendering the circuit impervious to damage due to
design can be used in student laboratories and general shorted stimulus outputs. The two voltage divider
experimental work where variation of the stimulus resistors should be closely matched to minimize dc
waveform is not necessary. offset at the electrode site.
A schematic of the stimulator is shown in Figure I. Stimulus current is adjusted via the .5-MQ
The circuit utilizes two LM 555 integrated circuits", potentiometer. Further. the current delivered to the
one in monostable (!C1) and one in astable OC2) electrode may be read off the 0-100 de micro-
multivibrator configuration. Upon shunting of the ammeter connected to the full-wave diode
trigger input to ground (by the closing of a bridge as shown. The diode bridge enables the use of
microswitch , for example), lCl is set to produce a an inexpensive dc meter movement; ac micro-
pulse of given duration. determined by RD. A 10-kQ ammeters are prohibitively expensive for this
resistor prevents spurious triggering by "pulling up" application. Since IC2 operates on a 50% duty cycle
the open trigger input of lCl Pin 2. to +V. (symmetrical) square wave, current read from the
The R-C circuit formed by the 2.2-MQ resistor and meter equals one-half the current being delivered to
.OI5-I-lF capacitor limits the triggering to only one the electrode site.
pulse from ICI regardless of the length of time that In intracranial stimulation research. it is important
the trigger inputs are shunted. The sustain switch that stimulus current be relatively constant despite the
bypasses the R-C circuit and allows continuous-duty impedance changes expected with chronic prep-
operation of ICI by grounding input Pin 2. arations. In this stimulator a large resistance. the
The pulse from lCl Pin 3 powers both the stimulus .S-MQ potentiometer. is placed in series with the
on light-emitting diode (LED 2) and astable IC2. IC2 electrode in order to adjust. and to regulate stimulus
current. Any change in the electrode/tissue
impedance. therefore. is a small proportion of the
This research was supported in part by the Committee on Facultv total load impedance presented to IC2. For example.
Research of the Graduate School of the University of Mississippi.
Requests for reprints should be addressed to Alan 1. Fridlund, P. assuming an electrode/tissue impedance of 15 kQ.
O. BOll 7457. Department of Psychology, University of Mississippi, and a setting of the potentiometer such that its
University. Mississippi 38677. resistance is 250 kQ. the total load impedance
21
22 FRIDLUND AND LEBERER
ruSE
IA OIOOES ARE IN'I\48
orr_~ .......... POWER F..;;J.lil...:a;.LloI..llKA-_,- .., OR EQUIV.
L...-.-_ _- ' SUPPLY
oN/Off
Rp tsEE TEXT)
TRIGGER
!~T 10m 'I 8
.
l.;--------.e...---4z
._~
555 6t---..
Itt
7t--~
:> 5 2.2KO 1%
10
2.2 KO 1%
r-----1r:;:i:1:;1-1,.5
.015
8
Kfl 2 7
Hon e
l
'I :> 6
2 7t-~-....
4 5
US,." IN
'/
LE02
STllAULUS ON
555
tC2 "t-'--.... 180KO
LM 555
TOP I/£!N
ON/JFF 3
presented to IC2 is 265 kQ. Even if the impedance of Stimulus duration and frequency are determined by
the preparation increases by one-third, to 20 kQ, total appropriate selection of duration resistor RO and
load impedance only increases to 270 kQ. Since in a frequency capacitor CF. These parameters have been
series circuit current in any part of the circuit is a shown to influence substantially responses induced at
function of the total circuit impedance. current at the specific electrode sites (Ervin & Kenney, 1971).
electrode site decreases by only 2% {(2S6 kQ- In our laboratory. the prototype stimulator
270 kQ)/26SQ]. Thus for practical intensities of employed six-position switches for RO and CF. such
stimulation. current regulation is quite satisfactory. that stimulus duration could be varied from
The operation oflC2 itself is relatively independent of .05-1.0 sec. and frequency from 60 Hz to I kHz.
wide variations in electrode impedance. CF should be determined first, Use a temporary
Power supply requirements for this circuit are not value of 1 MQ for RD. Close the sustain switch and
critical. Any 12-18 V de supply capable of producing shunt the trigger inputs; this turns IC2 on
300 rnA, and with some filtering. will work. The indefinitely. The use of a digital frequency counter is
circuitry is compact and. excluding power supply. can highly recommended. and is connected to the
easily be placed on one 4 x 6 in. printed circuit card. stimulus outputs for this measurement. If such a
although lead dress is not critical and point-to-point counting meter is not available, the stimulus outputs
wiring may be used effectively. If possible, the may be connected directly to 600-Q headphones. The
stimulator ground should "float" free from the stimulus frequencies can then be calibrated with an
conductive elements of the animal's cage." audio generator by "beating" the stimulus output
frequency against the generator frequency. The
frequencies match when the beat frequency is 0 Hz.
Table I Values of CF are selected for the frequencies desired;
Values for C F for Varying Stimulus Frequencies
approximate values for different frequencies are
CF (Microfarads) Frequency (Hz) shown in Table 1. Precise frequency determinations
will require addition of appropriate "gimmick"
.0033 1000
.005 750 capacitors either in series or parallel with CF. In
.0075 500 general. a close approximation for CF can be obtained
.01 350 from the formula:
.02 200
.027 120
.033 100
.075 60
INTEGRATED CIRCUIT INTRACRANIAL STIMULATOR 23
Table 2 Table 3
Values of R O (01 Varying Stimulus Durations Parts List fOI the Intracranial Stimulation Unit
Ro (Ohms) Duration (Sec) Quantity Component