Professional Documents
Culture Documents
Atomlab - Plus - 100 - Operation and Service Manual
Atomlab - Plus - 100 - Operation and Service Manual
DOSE CALIBRATOR
OPERATION AND SERVICE MANUAL
VERSION 2.0
086-265
086-269
BIODEX
Biodex Medical Systems, Inc.
20 Ramsay Road, Shirley, New York, 11967-4704, Tel: 800-224-6339 (In NY and Int’l, call 631-924-9000), Fax: 631-924-9241, Email: sales@biodex.com, www.biodex.com
B. This warranty does not apply if the product, as determined by BIODEX MEDICAL
SYSTEMS, INC., is defective due to abuse, misuse, modification or service performed by
other than a BIODEX MEDICAL SYSTEMS, INC. authorized repair and calibration facility.
Misuse and abuse include, but are not limited to, subjecting limits and allowing the
equipment to become contaminated by radioactive materials.
C. In order to obtain warranty repair service, the equipment must be returned freight pre-
paid to one of our facilities. The Return Authorization number (R.A. #) should be included,
along with a statement of the problem. Equipment will be returned transportation prepaid.
2. Calibration
A. Instruments are warranted to be within their specified accuracy at the time of shipment.
If a question arises and BIODEX MEDICAL SYSTEMS, INC. determines that the initial cali-
bration is in error, the instrument will be recalibrated at no charge.
B. Mechanical products are warranted to meet written specifications and tolerances at the
time of shipment.
4. Non-Warranty Service
A. Repairs and/or replacements not covered by this warranty may be performed by
BIODEX MEDICAL SYSTEMS, INC. at a factory authorized service location. Estimates of
repair charges may be requested, however, a charge for estimate preparation may apply if
the repair is later not authorized by the customer.
B. The cost of transportation into and out of the service location will be the responsibility of
of the customer.
BIODEX
Biodex Medical Systems, Inc.
20 Ramsay Road, Shirley, New York, 11967-4704, Tel: 800-224-6339 (In NY and Int’l, call 631-924-9000), Fax: 631-924-9241, Email: sales@biodex.com, www.biodex.com
Service Procedure
If you think you have a service problem, take the following action.
If you still think you have a service problem, call BIODEX MEDICAL SYSTEMS, INC.,
Service Department at (800) 224-6339.
2. After taking the information, we will advise on the action we will take.
3. Sometimes service personnel must consult with engineering and it may take time to get
back to you. Be sure to let the service representative know your schedule so that we
can call at a convenient time.
4. The return call may be from a person other than whom you first reported the problem to.
5. After analyzing the problem, we will decide if the unit must be returned to us for repair,
or replacement parts will be sent.
6. If unit must be returned, it will be given a return authorization (R.A. #) number by us.
Pack the system in the carton that it was originally shipped in, or pack it safely and
securely to avoid shipping damage. It is the customer's responsibility for any damage
that occurs during shipping.
BIODEX
Biodex Medical Systems, Inc.
20 Ramsay Road, Shirley, New York, 11967-4704, Tel: 800-224-6339 (In NY and Int’l, call 631-924-9000), Fax: 631-924-9241, Email: sales@biodex.com, www.biodex.com
ATOMLAB Dose Calibrator Operation Manual Table of Contents
TABLE OF CONTENTS
Page
Chapter One: Introduction to the Atomlab 100 Plus Dose Calibrator.........1-1
Introduction to the Atomlab 100 Plus Dose Calibrator and this manual.
Chapter Two: Installing Your Dose Calibrator ...............................................2-1
Describes what you need to properly use your Atomlab Dose Calibrator,
powering up, installation and cleaning instructions.
Chapter Three: Getting Started .......................................................................3-1
This chapter allows you to gain a quick familiarization of the operation of
your Dose Calibrator.
Chapter Four: Description ...............................................................................4-1
Describes features and functions of your Dose Calibrator, including the
Detector and Display units.
Chapter Five: Operation...................................................................................5-1
Describes the full operation of your Atomlab 100 Plus system.
Glossary .........................................................................................................GL-1
Appendices
Appendix A .......................................................................................................A-1
Atomlab Dose Calibrator Troubleshooting Procedures
Appendix B .......................................................................................................B-1
Atomlab 100 Plus Dose Calibrator Specifications
Authorized European Community Representative
Appendix C .......................................................................................................C-1
Decay Calculations
Appendix D .......................................................................................................D-1
Pre-Set Calibration Values
Appendix E........................................................................................................E-1
Calibration Values for Other Isotope Selections
Appendix F ........................................................................................................F-1
Atomlab Dose Calibrator Calibration and Traceability
Appendix G.......................................................................................................G-1
Atomlab Computer Interconnection
Appendix H .......................................................................................................H-1
Determining the Firmware Level of Your Dose Calibrator
Appendix I ..........................................................................................................I-1
Electromagnetic Compatibility
Appendix J ........................................................................................................J-1
Quality Assurance Testing of Atomlab Calibrators
(Manufacturing Instructions)
Appendix K .......................................................................................................K-1
Specifications
Quick Reference Chart
Trouble Shooting Chart
Schematics
CHAPTER ONE:
INTRODUCTION TO THE ATOMLAB 100 PLUS
DOSE CALIBRATOR
OVERVIEW
CHAPTER TWO:
INSTALLING YOUR ATOMLAB 100 PLUS
DOSE CALIBRATOR
OVERVIEW
OPERATING REQUIREMENTS
Work Surface The surface area of the table top should provide ample area
to accommodate the Detector Unit, Display Unit, the Printer,
the L-Block shield, the shielded radionuclide preparation
area, and room for writing. See Figure 2-1 for a typical dose
preparation layout.
• 30 in. deep
• 48 in. wide
• 36 in. high
Physical Location Frequently, dose calibrators are located in areas called hot
and Environment labs in which there is a wet sink with plumbing. Do not locate
your dose calibrator on a table or other counter which can
become wet or is subject to splashing or spraying.
In general, make sure your work area is clean, dry, and dust
free.
POWER REQUIREMENTS
Line Voltage 100 to 120 VAC 1/2 A, or 200 to 240 VAC 1/4 A, selectable
with Fuse Holder (see Appendix A).
POSITIONING
CABLE CONNECTIONS
Detector Unit Refer to Figure 2-2 for an illustration of the Detector Unit's
Connection cable connector and rear panel. The Detector Cable is gray,
ten feet long, and has white identical male connectors on
both ends. The connectors are of a self-locking type.
Figure 2-2: The Rear Panel of the Detector Unit, end of the
Detector Cable, and the Operational Light.
Display Unit NOTE: For the next procedure ensure your dose calibrator
Connection is turned OFF.
Figure 2-3: The rear panel of the Display Unit with Chamber
Connector, the end of the Detector Cable, and the Display
end of the Power Cord.
Power Cord 1) Rotate the Display Unit so that you are looking at the
Connection rear panel. Directly to the right of the connector labeled
CHAMBER, locate the Fuse Holder. This is labeled USE
ONLY WITH 250V FUSES, with 100-120 read upright,
and 200-240 read upside down.
2) Make sure the Power Switch for the Printer is OFF (see
Figure 2-4).
3) Rotate both the Display Unit and Printer so that you are
looking at their rear panels.
NOTE: The ends of the Printer Power Cable are not inter -
changeable.
POWERING UP
PACKING MATERIAL
Now that you have set up your Atomlab Dose Calibrator, and
everything is intact and functioning properly, take a few
moments to repack the packing material. This material was
especially designed for safe shipment of the Atomlab Dose
Calibrator and should be considered part of the instrument.
Place the shipping material in the appropriate boxes and put
the boxes back into the shipping carton. Label the carton and
store it in a safe, out-of-the-way location. If you ever need to
return your dose calibrator for updates, calibration, or repair,
use the original shipping carton, boxes, and shipping material.
PRINTER TYPE
DIP SWITCHES
HANDLING NOTES
5) Keep your hands out of the Printer while the power is ON.
UNPACKING
Installing the Install the Paper Holders in the outermost holes in the rear of the
Paper Holders Printer (see Figure 2-7). Insert the holders, positioned as shown,
and Re-Roll and push down slightly so that they lock into place. (The guides
Prevention Guard on the holders fit into the indentations beneath the outermost
holes in the rear of the Printer.) Do not force the holders!
1) Turn the power OFF, lift the Printer Cover up, and
remove it (see Figure 2-8). There is an indentation to
the right and at the rear of the Printer Cover.
NOTE: For the next two steps, hold the ribbon taut as
you install it. Refer to Figure 2-10.
Wrap one side around the black Ribbon Guide on the end
of the Platen and drop one Spool on the Spool Shaft.
Make sure the Spool Driving Pins engage with the Spool
Drive Holes. As the spool drops into place there will be a
click.
Removing the Ink Hold the spool and lift gently, rotating it until the ribbon sags.
Ribbon Push the ribbon Detecting Lever out, lift the spool until it
comes off the shaft. Remove the second spool by following
the same procedure.
PAPER INSERTION
4) Insert the label paper evenly into the Paper Insertion Slot
until it hits the pin feed hubs.
PAPER REMOVAL
Cut the paper close to the slot and use the FEED button
until the paper has passed completely through the Printer.
When you are confident you have made the proper cable
connections and installed the required accessories correctly,
turn the Power Switch ON. If everything is right, the green
POWER and ON LINE LEDs will come ON, and the red
ALARM LED will not come ON.
CHAPTER THREE:
GETTING STARTED WITH YOUR
ATOMLAB 100 PLUS DOSE CALIBRATOR
OVERVIEW
METHODOLOGY
Front Panel Refer to the Front Panel Display in Figure 3-1. On the
Display Atomlab 100 Plus the four-digit display associated with the
Electronic Thumbwheel is referred to as the Electronic
Thumbwheel Display. This display is used to set the date,
time and dial values for the OTHER key.
GETTING STARTED
Read the Display NOTE: The top right four-digit LED Activity Display always
indicates activity in the Detector's Chamber Well, no matter
which keys are pressed. (There are momentary exceptions:
when you press the TEST key, for example.)
the LED on the key come on (the Tc-99m key may already
be selected). If there is no activity in the Detector, the dis-
played value should be about 1.00 µCi or less. This value
may get lower after "warm-up".
NOTE: The Mo-99 key has a special function for the Moly
Breakthrough test, which is described in Chapters 4 and 5.
Cs-137 Check After the above background correction has been completed,
remove your Cs-137 calibration Check Source from its
shield, place it in the Sample Holder and lower it into the
Chamber Well. Press the Cs-137 Isotope Selection key on
the front panel and observe that the LED on the key comes
on. The Activity Display will now indicate the activity of your
Cs-137 Check Source. Compare it to the decayed calibrated
AVERAGING PERIOD
RANGE SELECTION
ISOTOPE SELECTION
DISPLAY UNITS
(To access the isotope dial values, press the SETUP key.
The month/day setting will appear. Continue pressing the
SETUP key three more times to advance past the
Month/Date, Hour/Minute and Century/Year settings. The
unit should now display the dial value for the isotopes listed
on the key pad.)
The Atomlab 100 Plus printout will look similar to Figure 3-2.
CHAPTER FOUR:
ATOMLAB 100 PLUS DOSE CALIBRATOR DESCRIPTION
OVERVIEW
GENERAL DESCRIPTION
THE DETECTOR
THE CHAMBER
CURRENT MEASUREMENT
REAR PANEL
RESPONSE
The Atomlab 100 Plus Dose Calibrator Display Unit consists of con-
trol keys and displays that allow you to make activity measurements.
A built-in microprocessor executes commands input via the front
panel keys and computes activity values from Detector data. This
unit can also print dose/syringe shield labels and store and print
two months worth of Constancy for Co-57, Cs-137 and Ba-133.
The Display Unit, with a molded plastic case housing the elec-
tronics, has been specifically designed to perform Activity
Measurements in a laboratory setting. To allow easy fingertip
control of the keys, the front panel slopes gradually. The Activity
Display slopes as well, providing an optimum viewing angle. On
the rear panel of the unit are the power and communication con-
nectors, and the power switch, which remain out of the way as
they are infrequently needed.
ZERO CALIBRATION
DISPLAY UNITS
For example, 12.03 µCi = .445 MBq and 63.9 mCi = 2.36
GBq. If the numeric display reads 0.445 and MBq is indicat-
ed, the activity value is 445,000 Becquerels; if 2.36 is dis-
played and GBq indicated, the activity is 2,360,000,000
Becquerels.
Activity Display The four-digit LEDs which make up the Activity Display
always display measured activity, except during the Self-
Test.
To the right of the Activity Display are the four display unit indi-
cators: µCi, mCi, MBq, and GBq. If you select Ci as the dis-
play units, the LED behind either µCi or mCi will come ON,
depending on the range automatically selected by the Atomlab
Dose Calibrator; if you select Bq, MBq or GBq will be indicat-
ed.
Isotope Selection These keys allow you to select which radioisotope you will
Keys be measuring. When, for example, you press the key
marked Tc-99m, the dose calibrator will be calibrated to
accurately measure the activity of Tc-99m. The LED on the
key you press will come ON so that you will always know for
which radioisotope the system is currently calibrated.
Standard The standard radioisotope (Isotope Selection) keys for the
Isotopes Atomlab 100 Plus include: Co-57, Cs-137, I-131, In-111,
Ga-67, Xe-133, I-123, Tl-201, Ba-133, Sr-89, Mo-99 and
Tc-99m.
After 4 seconds, this display will turn OFF and the next
test given in 2) will occur.
8.8.8.8.
Background When you press the BKGND key to store the background
Correction activity, this value remains in memory until you press the
(BKGND Key) key again. Pressing the BKGND key a second time will abort
the process.
ELECTRONIC THUMBWHEEL
CHAPTER FIVE:
OPERATION OF YOUR ATOMLAB 100 PLUS
DOSE CALIBRATOR
DAILY CHECKS
Background Ensure the Chamber Well Liner and Sample Holder have
Correction been placed in the Detector. Press BKGND.
Time and Date values are programmed at the first level of the
Setup mode. Once established, these values are used for all
Atomlab 100 Plus functions for which Time and Date are
recorded including: test report printouts, activity record printouts
with the time the isotope was counted, and system constancy
functions. The clock is also needed to properly decay the
stored radionuclide standards that are used for constancy.
Procedure At the first level of the Setup mode, the month and day are
displayed. To set month and day, use the following procedure:
PRINTER OPERATION
The test method has been derived from the fact when Mo-99
decays, high energy gamma photons (~75O keV) are produced
which can “breakthrough” a lead wall with a thickness sufficient
to totally absorb the lower energy Tc-99m gamma. Therefore,
when a Tc-99m source is placed in the properly shielded lead
container and inserted into the Chamber Well of the Atomlab
lOO Plus, any measured radiation will be due to Mo-99
gamma transmission.
1) Press the SETUP key four times. The SETUP LED comes
ON. The OTHER key LED begins to flash indicating you
may select the other display units if desired.
1) Press OTHER.
2) Press SETUP four times. The SETUP LED will come ON.
CONSTANCY CHECK
CAUTION SYMBOLS
◆ Indicates decayed activity is less than 0.050 mCi and
which may fall below your license requirement
■ The measured activity error exceeds 5% but is less than
10%. Consult with your RSO or designated person from
your license
■ The measured activity error exceeds 10% which is
beyond your license requirements. Consult your RSO for
recomended action.
MISEN EN GARDE:
◆Le degré d'activité théorique indiqué est moins de 0.050 mC i
ce qui peut êlre sous les normes requises de votre pemis.
■La mesure d'erreur d'activité excréde 5% mais est moins de
10%. Consultez votre officier de radi protection (physicien)
ou la personne désignée sur votre permis.
■La mesure d'erreur d'activité excéde 10% ce qui dépasse
les normes de votre permis. Consultez votre officier de
radioprotection (physicien) pour ses recommentations.
In Case of Er. 7 If the Atomlab 100 Plus detects more than a 5% discrepancy
in its Daily Check it will beep, the TEST LED will flash and
Electronic Thumbwheel (#1) will display an Er. 7 message
after you have pressed the ENTER key for the second time.
To escape the error loop, press the TEST key, followed by the
TEST key. This performs the test loop and generates a printed
Form with the Evaluation Message noted in the caption for
Figure 5-3. Press an isotope key to return to normal operation.
Monthly Report The Atomlab Dose Calibrator begins storing Daily Checks
the day you first enter the Check Source Activity. T h e
Atomlab 100 Plus stores two months worth of Constancy
Check data for all three constancy isotopes. Two constancy
report keys are labeled “This Month” and “Last Month’:
Pressing “This Month” will print a report from the first of this
month up to and including today’s date. “Last Month” prints
all of the previous months data. This report remains available
for a full month. This allows ample opportunity for the user to
print their report. As the month rolls over to the next month,
the “Last Month” data is replaced by the “This Month” data.
Reports See end of chapter for sample pages that the reports could
be place on and kept in the files.
CONTAMINATION CHECKS
Detector Liner Repeat the above procedure, removing and replacing the
Check Well Liner.
SAMPLE REPORTS
Figure 5-4: A sample Test Report Page with Test Label put into place.
Figure 5-6: A sample Daily Constancy Page with a month end report in place.
GLOSSARY
OVERVIEW
Accuracy Tests The test of a dose calibrator to see how close the displayed
activity values are to the true values.
Averaging Period The time the Atomlab takes to display a stable activity
reading.
Calibration Value The value set on the Electronic Thumbwheel (#1) Display to
adjust the Atomlab to correctly respond to a particular
radioisotope. See Response, Ionization Chamber.
Decay Chart A chart projecting how the activity of a source will decay
over a period of time at specified time intervals.
Molybdenum-99 The parent radionuclide which decays into the daughter Tc-
(Mo-99) 99m, used to generate radiopharmaceuticals.
APPENDIX A:
ATOMLAB DOSE CALIBRATOR
TROUBLESHOOTING PROCEDURES
OVERVIEW
PROBLEM DESCRIPTIONS
2) Turn the power OFF again and remove the Power Cord
from the electrical outlet. Remove the other end of the
Power Cord from the receptacle on the rear of the
Display Unit.
4) Turn the power OFF, remove the Power Cord from the
outlet and verify that the voltage of the outlet is correct.
If not, repair the outlet or find another outlet that
provides the required voltage. Re-insert the Power Cord
and turn the power back ON.
5) Turn the power OFF, remove the Power Cord from the
outlet and remove the Fuse Holder, as described later in
this appendix (see Figure A-1). Verify that both fuses
are intact. If one or both are not, replace. Re-insert the
Fuse Holder into the rear panel of the Display Unit. Re-
insert the Power Cord into the outlet and turn the power
back ON.
If your system still will not power up, call the Biodex
Customer Service Department.
After you insert an isotope into the Detector and select the
properly calibrated key, the display may vary at low activity
values as the system acquires the new data. After a short
Display Fluctua- For some isotopes which require a high Calibration Value,
tions For High you may observe fluctuations of ±0.5 µCi or more. Keep in
Calibration mind that the Calibration Value for an Isotope is related to
Values gain of the electrometer, or by what factor the electrometer
has to amplify the signal from the detector. The Calibration
for Tc-99m is 33.6, and a reading of 10 µCi for Tc-99m is
equivalent to 10-12 Amps measured by the electrometer.
High Calibration Values impact gains so that the electrome-
ter amplifies inherent noise to a greater degree. Thus,
greater fluctuations are observed in the Activity Display.
UNEXPECTED READINGS
PRINTER PROBLEMS
NOTE: The form printed out after power up, and after
pressing the TEST key, also shows the software versions of
the Display and Chamber Units.
Procedure 1) Rotate the Display Unit so that you are looking at the
rear panel. Directly to the right of the connector labeled
CHAMBER (between the power plug and ON/OFF
switch), locate the Fuse Holder. This is labeled in the
center with USE ONLY WITH 250V FUSES, on the
bottom with100V-120V read upright and on the top with
200V-240V read upside down.
2) The orientation of the Fuse Holder determines the
Orientation power the Atomlab unit requires. If the label
100V-120V can be read upright (the label 200V-240V is
upside down), and the corresponding arrow directly
below the label points to the arrow on the rear panel
directly below the Fuse Holder, then you must plug your
dose calibrator into an 100 VAC power source.
Fuse Inspection Once the Fuse Holder is removed, inspect the filament at
the center of the two fuses. If it is burned or broken in any
way the fuse must be replaced. If the filament does not
appear to be broken or burned, test the fuse for continuity
with an ohm meter.
Fuse Holder To replace the Fuse Holder, position it into the indentation
Replacement from which it was removed and gently push it back into
place. It is critical to first determine the particular power
requirement (100 VAC or 240 VAC) and ensure the Fuse
Holder is properly oriented before turning the system back
ON. See step 2) above.
NOTE: For certain error faults you can press the TEST key
and continue to use your dose calibrator. The LED on the
TEST key will continue to flash, reminding you that action to
correct the problem must eventually be taken. Some of
these faults can be easily corrected over the phone by the
Biodex Customer Service Department. Other faults, however,
require that you return the dose calibrator to Biodex Medical
Systems for servicing.
APPENDIX B:
ATOMLAB 100 PLUS DOSE CALIBRATOR
SPECIFICATIONS
OVERVIEW
MEASUREMENT RANGE
DISPLAY RANGES
Ci Bq Ci Equivalent
RESPONSE TIME
Auto selection:
one second above 2 mCi;
3 seconds between 200 µCi and 2 mCi;
3-30 seconds below 200 µCi.
ACCURACY
Stability Short term (24 hours): within 0.3% above one mCi,
exclusive of background.
Long term (one year): within 1%.
NOISE
ENERGY RANGE
25 keV to 3 MeV
ISOTOPE SELECTION
DETECTOR TYPE
Detector Shield 1/4 in. (6.4 mm) lead surrounding ionization chamber, with
top well opening.
Temperature: 0 to 40°C
Humidity: 0 to 95% rh, non-condensing
PHYSICAL DATA
POWER
DETECTOR CABLE
Conductors Six total: two for power, one for chassis ground, three for
serial data.
FRONT PANEL
Activity Display Four digit, seven segment red LED, 0.57 in. (14.5 mm)
high, four floating decimal point indicators between digits.
Activity Units Four discrete inscriptions back lit with large red LEDs: mCi,
µC i, MBq or GBq (right adjacent to digits).
Electronic Four digit, seven segment red LED, 0.27 in. (6.9 mm) high,
Thumbwheel with colon between second and third digits, and three float-
Display ing LED decimal point indicators between digits. Range is
from 000.1 to 999.
CLOCK
APPENDIX C:
DECAY CALCULATIONS
OVERVIEW
This appendix provides methods of performing radioisotope
decay calculations.
4) Calculate A = A0 x 2(-t/T1/2)
1) A0 = 43 mCi.
4) A0 x 2(-t/T1/2)
A = 43 mCi x 2(-0.8833/6.007)
A = 43 mCi x 2(-0.147045)
A = 43 mCi x 0.903
A = 38.83 mCi
1) Ai = 43 mCi.
3) t = 32.5 h.
6) From the Decay Factor Chart, the .41 row and the .001
column, Xr (.410) = .75262.
N Xn (N) = 2 -N
0 1
1 0.5
2 0.25
3 0.125
4 0.0625
5 0.03125
6 0.01563
7 0.007813
8 0.003906
9 0.001953
10 0.000977
Table 1
Af = Ai e-λt
where
where
Af = X Ai
where
X = Xn (N) x Xr (R)
e(A + B) = eA x eB ;
eln(A) = A ; and
eAB = (eA)B .
Now t / T1/2 = N + R
= Ai x Xn (N) x Xr (R)
Xn (5) = 0.03125
Xr (0.410) = 0.75262
Af = X Ai = [Xn (N) x Xr (R)] Ai = [0.03125 x 0.75262] 43 mCi
Af = 1.01 mCi
APPENDIX D
ATOMLAB DOSE CALIBRATOR
PRE-SET CALIBRATION VALUES
OVERVIEW
Chart A
(For Chamber Units with level six firmware only.)
Chart B
(For Chamber Units with firmware level five or less.)
APPENDIX E:
CALIBRATION VALUES
FOR VERSION SIX FIRMWARE OTHER ISOTOPE
SELECTIONS
OVERVIEW
NOTE: For all Dose Calibrator Models, the well liner must
be installed before making measurements. The well liner
provides attenuation for the Y-90 betas that would otherwise
enter the chamber and cause high readings.
APPENDIX F
ATOMLAB DOSE CALIBRATOR CALIBRATION
AND TRACEABILITY
OVERVIEW
APPENDIX G:
ATOMLAB 100 PLUS COMPUTER INTERCONNECTION
PC PORT COMMUNICATIONS
Command
Dose Calibrator The first field starts with the first character after the title and
Records continues until a space is encountered. It may be from two
till six characters in length. There are thirteen isotope
names. Typical sample records are shown below (version 5
firmware):
The second field is the dial value that is being used to cal-
culate activity. The third field is the numerical value of activi-
ty. This number may be positive or negative. The fourth and
last field is the units of activity. They will be either µCi, mCi,
MBq or GBq. The final character is the end of record char-
acter as described above.
ADDITIONAL CONTROLS
CODE BUTTON
A Tc-99m
B Mo-99
C Ti-201
D I-123
E Xe-133
F Ga-67
G In-111
H I-131
I Sr-89
J OTHER
K Cs-137
L Ba-133
M Co-57
N SETUP
O TEST
P BKGND
Q Label
R This Month
S Last Month
T ENTER
U N/A
V N/A
W Up Arrow ETW#1
X Up Arrow ETW#1
Y N/A
Z N/A
[ N/A
APPENDIX H
DETERMINING THE FIRMWARE LEVEL OF YOUR
DOSE CALIBRATOR
APPENDIX I:
ELECTROMAGNETIC COMPATIBILITY
APPENDIX J:
QUALITY ASSURANCE TESTING OF ATOMLAB CALIBRATORS
(MANUFACTURER'S* INSTRUCTIONS)
The date on the Report of Calibration does not expire since the report is not
useful, in the typical sense, as a record of calibration in a periodic maintenance
schedule.
For example, when a survey meter is purchased, it normally comes with a calibra-
tion certificate, dated within the past 6 or 12 months. This meter is then placed on
a re-calibration schedule, as determined by local regulations. The re-calibration
date may be scheduled from the original calibration date, or simply placed on the
institution’s normal schedule interval for such meters.Re-calibration requires the
meter be shipped to a laboratory that is licensed to perform such a task. There is
no need to check the accuracy of the survey meter upon installation at the facility;
the facility generally has no means to check the accuracy of a survey meter.
The situation and requirement is different for Dose Calibrators. Although modern
manufacturing capabilities have improved the robust nature of dose calibrator
systems, these products tend to be more fragile than survey meters. Therefore,
the NRC determined at an early date that the dose calibrator accuracy must be
tested on site, prior to its introduction into clinical use. This was prudent since, a)
the dose calibrator was being used to measure radionuclides prior to administra-
tion to a patient and, b) the availability of sealed long lived isotopes made this
accuracy test very reasonable at a clinical facility.
The NRC now requires the facility using a dose calibrator to implement the
Manufacturer’s Instructions for Quality Assurance Testing of Dose Calibrators.
Such Instructions accompany all Atomlab Dose Calibrators, and requires the
Accuracy test "at installation and at least annually thereafter. After repair of the
dose calibrator, repeat the above tests as a new installation."
Therefore, the date on the Report of Calibration cannot be used as a substitute for
the installation accuracy test. The only significance of the date is when the dose
calibrator was calibrated at the manufacturer. It never receives another calibration
unless it is returned to the manufacturer. The accuracy test at the facility is not a
calibration; it is simply an accuracy test stating that the dose calibrator measures a
specific radionuclide within a stated error. The date on the Report of Calibration
does not expire until the dose calibrator is returned to the manufacturer.
APPENDIX K:
SPECIFICATIONS
Detector Unit:
Dimensions: 7.5" w x 7.5" depth x 16.25" h (19 x 19 x 41.3 cm)
Well I.D.: 2.75" dia x 10.5" h (7 x 26.7 cm)
Well I.D. with Liner: 2.5" dia x 10.25" h (6.35 x 26 cm)
Lead Shielding: .25" lead (.63 cm)
Weight: 35 lb (16 kg)
Approvals: ETL to UL 3101-1 and cETL to CAN/CSA C22.2 No. 1010.1-M92
Related:
086-301 Shielding Rings, Interlocking, 2" lead
For additional protection from high energy activity
086-423 Moly Shield, Vial, .3" lead
086-435 Moly Shield, Syringe, .3" lead
086-509 Lineator
Replacement:
086-242 Vial/Syringe Dipper
086-241 Well Insert
086-268 Labels, “peel & stick”, 200/roll
086-278 Ribbon, Printer, 4/pkg
BIODEX
Biodex Medical Systems, Inc.
20 Ramsay Road, Shirley, New York, 11967-4704, Tel: 800-224-6339 (In NY and Int’l, call 631-924-9000), Fax: 631-924-9241, Email: sales@biodex.com, www.biodex.com