Autoclaving disposable Petri dishes: How do you use a biohazard bag and can plastic
disposable Petri dishes be autoclaved in them?
We have just purchased an autoclave, and we have a large supply of plastic Petri dishes. I
noticed that the plastic Petri dishes melt when I put them into the autoclave. To solve the
problem, I am wondering whether it is ok practice to place inoculated agar Petri dishes into a
biohazard bag and place it in the autoclave to be sterilized, even if they melt? Also, what is
the procedure for using biohazard bags? For example, do you leave them open? Or do you
seal them before use? etc.
Voting:
No votes yet
Laboratory Technicians:
Laboratory Technicians
The best way to sterilise Petri dishes containing agar inoculated with microorganisms in an
autoclave or pressure cooker is by placing them into a bag that will withstand the sterilisation
conditions and contain the treated contents. Schools have the option of using two different
types of bags:
Autoclavable biohazard bags which are available from scientific suppliers, see the
Science ASSIST School science suppliers list. These are made from a heavy duty
plastic e.g. polypropylene marked with the international biohazard symbol and usually
have the word autoclavable written on them. They are available in a variety of sizes.
Oven bags which can be purchased from supermarkets. These are also available in
different sizes.
If you are unsure if your biohazard bags are suitable for autoclaving we suggest that you
contact the supplier for advice.
Procedure for using an autoclavable biohazard or oven bag for sterilising inoculated
agar in plastic Petri dishes:
Loosely pack agar plates into bags to no more than 2/3 full. This will ensure that the
steam during sterilisation will penetrate the entire load. Bags that are tightly filled to
capacity will not allow effective steam penetration and the contents will not be sterilised
even if all sterilisation parameters are met.
Make sure there are no sharp objects present that may puncture the bag.
Loosely tape shut the bag leaving an opening of about 5-6cm to allow good steam
penetration. This can be done with autoclave tape or a rubber band. Never tightly close
the bags as they are impervious to steam and therefore the temperature of the inside of
the bag will not be sufficient for sterilisation.
It is advisable to place the bag into a secondary container within the steriliser to
prevent any leakage into the steriliser should the bag rupture. The container must be
able to withstand the autoclaving conditions.
Do not overload the steriliser with too many bags as this may block steam circulation.
Use a sterility compliance strip to indicate if the correct time, temperature and
pressure have been reached during the run time. These are available from scientific
suppliers.
Sterilise at 15psi, 121oC for 15-20 minutes.
After sterilisation has been verified, the autoclave or oven bag containing waste items
should be disposed of by placing it into a sturdy garbage bag which is sealed for
immediate disposal in industrial bins.
Wear heat protective gloves when removing waste from the steriliser.
Sterilisation of these items ideally takes place as soon as possible after completion
of a practical activity and occurs within the science laboratory or prep room area.
Science ASSIST has developed the following resources with related information:
References:
1'Petri dishes, plastic', Marienfeld-Superior website. http://www.marienfeld-superior.com/petri-
dishes-plastic.html (Link updated: April 2018)