Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Wandong Products Specification

China Resources Wandong Medical Equipment Co., Ltd.

High Frequency 50kW, 150kV Radiography System


- HF50-R

Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China
E-mail: international@wandong.com.cn Tel/Fax: +86-10-84575792/3/5, 84575794

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 1 of 7
Wandong Products Specification

WDM: A successful journey of 60 years


China Resources Wandong Medical Equipment Co., Ltd. (WANDONG, originally Beijing Wandong
Medical Equipment Co., Ltd.) was established in 1955, and has been listed on the Shanghai Stock
Exchange since 1997. Headquartering in High-Tech industrial
park in Beijing, Wandong occupies an area of more than
100,000 square meters as its modern production base and
R&D center, and has 30+ branches in China and a wide sales
and service network in about 70 countries all over the world.
With the 60 years dedication to the medical imaging,
Wandong has made brand name of “WDM”, a well-known
brand in the world.

As one of the largest radiology imaging equipment


manufacturers in the world, Wandong has wide range of
product lines including General Radiography, Mobile X-ray &
C-arm, Digital Radiography (DR), Digital Fluoroscopy (DRF),
Digital Mammography, Cath-Lab, Computed Tomography (CT)
and Magnetic Resonance Imaging (MRI) machine, and the
annual production capability is more than 6,000 units.

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 2 of 7
Wandong Products Specification

Performance feature:
1) Excellent High-frequency Generator

 High-frequency generator output is stable and precise. System effectively reduced radiation to

patient by much lower scattering dose than traditional machine

 Transient exposure eliminates motion artifacts.

 APR (Anatomical Program Radiography) of more than 600 set exposure programs can meet all

clinical requirements. (5 kinds of body size, 30 anatomical positions, 4 projection directions)

2) Humanization Design Console

 Fast and reliable operation guaranteed by soft touch buttons and LCD display of parameters.

 Categorized buttons in control panel match with doctor’s operation procedure.

 Variety of console installation including wall-hanging, pedestal table, desktop placed fits for

different site requirements.

 Operation fatigue will be reduced by ergonomic design.

3) Integrated Radiography table

 The wide range of movement of Stand and radiography table to ensure patient to be covered

the whole body by X-ray tube, To fully meet the lateral position、oblique position and the head

of table and requirement of special position.

 Stand rotation makes stretcher patients to do diagnostic without moving.

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 3 of 7
Wandong Products Specification

1. Configurations:
 Standard Configurations:
No. Component Part Number Brand Quantity

1 High Frequency Generator GFS501-1 WDM 1

2 X-ray Tube Assembly E7843X Toshiba 1

3 HV cable DL7512-1B WDM 1

4 Collimator XS1-2 WDM 1

5 Radiography Table SC4-4 WDM 1

6 Vertical Bucky Stand LS-3 WDM 1

 Optional Configurations:

No. Component Part Number Brand Quantity

1 Ionization Chamber SSMC610 3 fields Claymount 1

2 Radiography Table SC4-1 WDM 1

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 4 of 7
Wandong Products Specification

2. Technical Specification:
Standard configuration

Component Specifications

Part Number: GFS501-1

Power: 50kW

Constant Frequency of Inverter: 25kHz

KV Range: 40~150kV

mA Range: 25~630mA

High Frequency Time Range: 2ms~5s

Generator mAs:0.5mAs~500mAs

AEC interface, APR

Power Supply: 380VAC, three-phase

Fault Auto Detection and Diagnosis

Console: LCD Displays

APR:600 positions

Part Number: Toshiba E7843X

Focal Spot: 0.6mm/ 1.2mm


X-ray Tube
Power: 22 / 50kW
Assembly
Anode Heat Capacity: 150kHU

Target Angle: 12°

Part Number: XS1-2

External Dimension: 250mm×224mm×250mm


Collimator
Projection Field(SID=65cm) Max: 35cm×35cm; Min: 0cm×0cm

Operation Mode: Manual

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 5 of 7
Wandong Products Specification

Part Number: SC4-4

External Dimension(Length* Width*Height): 2000mm*740mm*658mm

Tabletop Movement(Longitudinal):1200mm

Tabletop Movement(Transverse): 200mm

Longitudinal Travel of Tube Stand: 1200mm

Vertical Travel: 810mm~1710mm (Distance from X-ray Tube Focus to The Floor)

SID: 1100mm Max. (to Table Bucky)

Rotation Range of X-ray Tube Around Horizontal Arm: +120°~-120°

Tube Stand Self-Rotation Range: +180° ~ -180°, and the interval of 90°could be located.
Radiography
Tabletop Locking Mode: Electromagnetic lock, braking force≥100N
Table
Tabletop Locking Control: foot switch

Bucky Mode: Fixed

Grid Size: 356mm×432mm(14 ″×17″)

Bucky specification: f0=100cm, r=10, N=40L/cm

Cassette Size: Max 14″×17″

Distance between Tabletop and Bucky Film ≤65mm

Longitudinal Travel of Bucky(SID): 640mm

Weight Capacity: Max. 135 kg

Part Number: LS-3

Vertical Travel: ≥1000mm


Vertical Bucky
Maximum Dimension: 8″ ×10″-14″ ×17″
Stand
Grid (According to User’s Requirement):

N=40L/cm; r=8, 10 or 12; f0=150cm, 180cm or 200cm

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 6 of 7
Wandong Products Specification

Optional Components

Component Specifications

Part Number: Claymount SSMC610

Ionization AEC function, 3 fields

Chamber Exposure dose range:1~100uGy

Exposure time range: 1ms~5s

Part Number: SC4-1

External Dimension (Length* Width*Height): 2000mm*730mm*690mm

Tabletop Movement(Longitudinal): 1200mm

Tabletop movement(Transverse): 200mm

X-ray tube Stand Travel Range (Longitudinal): 580mm

Rotation Range of X-ray Tube Around Horizontal Arm: +90°~-90°

X-ray Tube Stand Self-Rotation Range: +180°~-180°, interval of 90°could be located.


Radiography
Distance from X-ray tube focus to the floor ≥ 1030mm~1740mm
table
Tabletop Locking Mode: Electromagnetic lock, footswitch control, braking force≥100N

Gird: 388mmх464mm, f0: 100cm, r10, N40 L/cm

Cassette Size: Max. 14″×17″

Distance between Tabletop and Bucky Film ≤ 80mm

Bucky Longitudinal Travel: 580mm

Bucky locking: Electromagnetic, braking force≥100N

This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 7 of 7

You might also like