Professional Documents
Culture Documents
Technical Bulletin: Alanine Aminotransferase Activity Assay Kit
Technical Bulletin: Alanine Aminotransferase Activity Assay Kit
Technical Bulletin: Alanine Aminotransferase Activity Assay Kit
TECHNICAL BULLETIN
homogenized with 200 L of ALT Assay Buffer. 5. Continue taking measurements until the value of
Centrifuge at 15,000 g for 10 minutes to remove the most active sample is greater than the value of
insoluble materials. the highest standard. At this time the most active
sample is near or exceeds the end of the linear
Serum samples can be directly added to wells. Add range of the standard curve.
1–20 L samples into wells of a 96 well plate.
6. The final measurement for calculating the enzyme
Bring samples to a final volume of 20 L with ALT activity would be the penultimate reading or the
Assay Buffer. For unknown samples, it is suggested to value before the most active sample is near or
test several sample volumes to make sure the readings exceeds the end of the linear range of the standard
are within the standard curve range. curve. The time of the penultimate reading is Tfinal.
For the positive control (optional), add 5 L of the ALT 7. Calculate the change in measurement from Tinitial to
Positive Control to wells. Adjust well volume to 20 L Tfinal for the samples and positive control.
with ALT Assay Buffer. A570 = (A570)final – (A570)initial
or
FLU = (FLUfinal) – (FLUinitial)
Results
Calculations The ALT activity of a sample may be determined by the
Correct for the background by subtracting the value following equation:
obtained for the 0 (blank) standard from all standard
readings. ALT Activity = B Sample Dilution Factor
(Tfinal – Tinitial) V
Plot the pyruvate standard curve using the Tfinal
readings. B = Amount (nmole) of pyruvate generated between
Tinitial and Tfinal
Compare the measurement value (A570 or FLU) of Tinitial = Time of first reading in minutes.
each sample to the standard curve to determine the Tfinal = Time of penultimate reading in minutes.
amount of pyruvate generated between Tinitial and Tfinal V = sample volume (mL) added to well.
(B).
Note: A new standard curve must be set up each time ALT activity reported as nmole/min/mL = milliunit/mL,
the assay is run. where one milliunit (mU) of ALT is defined as the
amount of enzyme that generates 1.0 nmole of
pyruvate per minute at 37 C.
4
Troubleshooting Guide
Problem Possible Cause Suggested Solution
Cold assay buffer Assay Buffer must be at room temperature
Omission of step in procedure Refer and follow Technical Bulletin precisely
Plate reader at incorrect wavelength Check filter settings of instrument
Assay Not Working
For fluorescence assays, use black plates
Type of 96 well plate used with clear bottoms. For colorimetric assays,
use clear plates
Use the Assay Buffer provided or refer to
Samples prepared in different buffer
Technical Bulletin for instructions
Repeat the sample homogenization,
Cell/Tissue culture samples were
increasing the length and extent of
incompletely homogenized
homogenization step.
Samples with erratic
Samples used after multiple freeze-thaw Aliquot and freeze samples if samples will be
readings
cycles used multiple times
Presence of interfering substance in the
If possible, dilute sample further
sample
Use of old or inappropriately stored Use fresh samples and store correctly until
samples use
Thaw all components completely and mix
Improperly thawed components
gently before use
Use of expired kit or improperly stored Check the expiration date and store the
Lower/higher reagents components appropriately
readings in samples Allowing the reagents to sit for extended
Prepare fresh reaction mix before each use
and standards times on ice
Refer to Technical Bulletin and verify correct
Incorrect incubation times or temperatures
incubation times and temperatures
Incorrect volumes used Use calibrated pipettes and aliquot correctly
Thaw and resuspend all components before
Use of partially thawed components
preparing the reaction mix
Pipetting errors in preparation of standards Avoid pipetting small volumes
Prepare a master Reaction Mix whenever
Pipetting errors in the Reaction Mix
possible
Non-linear standard Pipette gently against the wall of the plate
Air bubbles formed in well
curve well
Standard stock is at incorrect Refer to the standard dilution instructions in
concentration the Technical Bulletin
Recheck calculations after referring to
Calculation errors
Technical Bulletin
Substituting reagents from older kits/lots Use fresh components from the same kit
Samples measured at incorrect
Check the equipment and filter settings
wavelength
Unanticipated results Samples contain interfering substances If possible, dilute sample further
Sample readings above/below the linear Concentrate or dilute samples so readings
range are in the linear range
KVG,LS,MAM 05/17-1
2017 Sigma-Aldrich Co. LLC. All rights reserved. SIGMA-ALDRICH is a trademark of Sigma-Aldrich Co. LLC, registered in the US and other
countries. Sigma brand products are sold through Sigma-Aldrich, Inc. Purchaser must determine the suitability of the product(s) for their
particular use. Additional terms and conditions may apply. Please see product information on the Sigma-Aldrich website at
www.sigmaaldrich.com and/or on the reverse side of the invoice or packing slip.