Professional Documents
Culture Documents
1 - KMC-950 Service Manual (Eng) (V1.6)
1 - KMC-950 Service Manual (Eng) (V1.6)
Ver. 1.6
1
Service Manual
This manual may not be reproduced, in whole or in part, without the written permissions of
COMED Medical System, Inc.
The material in this manual is provided for informational purposes only and is subject to
change without notice.
KMC-950 is a trademark of COMED Medical Systems, Inc. In the USA and other countries. Other
product and company names mentioned herein are the property of their respective owners.
This manual contains descriptions, instructions, and procedures which apply only this product.
2
Service Manual
1. SPECIFICATION
C-ARM ----------------------------------------------------------------------------------------- 11
TV CAMERA --------------------------------------------------------------------- 12
3
Service Manual
2. SERVICE DATA
X-RAY generator 27
SETUP MENU 39
4
Service Manual
3. WORKSTATION
HARDWARE OVERVIEW -------------------------------------------------- 129
Location ----------------------------------------------------------------------------------- 129
Parts location ----------------------------------------------------------------
130
-
Connecting the Cables ---------------------------------------------------- 131
HOW TO change WorkStation & CXView image, data base --- 156
5
Service Manual
1. SPECIFICATION
SYSTEM POWER REQUIREMENTS
LINE VOLTAGE: 100 - 120 VAC 50/60 Hz : for UL
220 - 230 VAC 50/60 Hz
LINE PHASE: SINGLE
INPUT POWER: 3600 VA @ 120 VAC : for UL
12500 VA @ 230 VAC
LINE CODE: 26.3 Feet
LINE PLUG : 250 VAC, 63A LEGRAND NO. LV-63A-58700
ENVIRONMENT
OPERATIONAL: 10°C~40°C, 30~75% rH, 70~106kPa
TRANSPORT & STORGE: -40°C~70°C, 10~85% Rh, 50~106kPa
Extended Storage and Transportation > 2 days: 5 ~ 40°C, 5 ~ 90%RH
Short-term Storage and Transportation < 2 days: 0 ~ 45°C, 5 ~ 90%RH
X-RAY TUBE(INSERT)
MANUFACTURER: VARIAN
MODEL: RAD-99
TARGET MATERIALS: TUNGSTEN-RHENIUM
MOLYBDENUM TARGET
NOMINAL X-RAY TUBE VOLTAGE: 40~125kV
INHERENT FILTRATION: 0.5 mmAl @ 75kV ( IEC 60522 )
NOMINAL FOCAL SPOT VALUE: 0.3/0.6 mm ( IEC 336 )
ANODE HEAT STORAGE: 300kHU
ANODE HEAT COOLING: 6kHU/min
FILAMENT CURRENT: LARGE FOCUS : 4.3~5.0A
SMALL FOCUS : 3.5~4.1A
FILAMENT VOLTAGE: LARGE FOCUS : 4~12V
SMALL FOCUS : 4~6V
SERIAL NO. :
6
Service Manual
X-RAY TUBE ASSEMBLY (HOUSING)
ASSEMBLER: VARIAN
TYPE OF X-RAY TUBE: ROTATING ANODE
MODEL: OR-Ⅲ
TARGET ANGLE: 10
MAXIMUM HOUSING COOLING RATE: 15,000 HU
HOUSING HEAT STORAGE CAPACITY: 1,600 kHU
WEIGHT: 43 LBS. WITHOUT CABLES OR
ACCESSORIES.
POLARITY OF THE CABLE RECEPTACLES: ALDEN TYPE, P702-1(CATHODE)
AND P702(ANODE), HIGH VOLTAGE
RECEPTACLES. THE RECEPTACLES
MATE WITH ALDEN TYPE(P7010S)
HIGH VOLTAGE CONNECTOR AND
CABLE.
PERMANENT FILTRATION: 0.7 mmAl @75kV ( IEC 60522 )
7
Service Manual
X-RAY SOURCE ASSEMBLY ( Conforms to IEC-601-2-28 : 1993 )
ASSEMBLER: COMED MEDICAL SYSTEMS
POSITION OF THE FOCAL SPOT: 50.8mm from sticking surface of the Collimator
SERIAL NO. :
8
Service Manual
WARNING
In the normal fluoroscopy mode, it is desirable that shooting X-ray for 5minutes continuously, and
5minutes break, and then 5minutes shooting, and 5minutes break again. In the radiography mode,
it is desirable that shooting X-ray 1 time, and 40seconds break, and then shooting 1 time, and
40seconds break again. In the boost mode, you don’t have to push the foot switch over 3seconds
per 1 shooting, because the X-ray exposes just for 3seconds per 1time shooting in the boost
mode. As above, it is desirable that shooting X-ray for 3seconds per 1time, and then 30seconds
break, and then 3seconds shooting, and then 30seconds break again in the boost mode.
WARNING
If the tube is overheated, thermal sensor attached to the exterior of it warns the system and ‘Tube
Limit’ sign will be displayed on the TFT screen. Then the machine can work again by the time the
tube gets cooled.
9
Service Manual
C-ARM AND STAND MECHANICAL
OVERALL SIZE AND WEIGHT
HEIGHT : 1737mm
WIDTH : 800mm
LENGTH : 1974mm
WEIGHT : 300kg
FREE SPACE : 799mm
DEPTH : 692mm
10
Service Manual
C-ARM
ORBIT.ROTATION : 135 DEGREES
HORIZ.TRAVEL : 200mm
VERT.TRAVEL : 500mm"
SID Fluoro : 1000mm RAD : 970mm
☞ The size of X-Ray field does not differ from markings by more than 2% from the FOCAL SPOT
PANNING : ± 12.5 DEGREES
ROTATION : ± 180 DEGREES
11
Service Manual
TV CAMERA
NAME (type): LUNA1K
CCD SCANNING: InterlineTransfer CCD; progressive scan with microlenses
NUMBER OF ACTIVE PIXELS :1024(H) x 1024 (V) pixels
CCD Type 1/2” Interline Anti-blooming CCD sensor
Integrated optical lens High resolution with adjustable focus and diaphragms & iris
A/D converter 14 Bit
SIGNAL PROCESSING 16 Bit Digital signal processing
LIH Image rotating
Image reverse
Circular blanking
Edge enhancement
Noise Reduction
INTERFACES Video output: Camera link
Camera control interface: RS-232
LENS MOUNT C-MOUNT
POWER SUPPLY: 24Vdc ± 20%
OPERATION CONDITIONS
Working temperature -10°C to + 45°C
Store temperature -40°C to + 55°C
Humidity Operating to 95% relative humidity (non-condensing)
XTV MONITOR
MODEL MM-190
MANUFACTURER MEDITECH
TYPE B/W TFT Active Matrix LCD
POWER SUPPLY 230VAC±15%, 115VAC±15%
MONITOR SIZE: 19”
ASPECT: 5:4 (1280 x 1024)
BRIGHTNESS: 1,000cd/ (Center 1-Point, typ)
VIEWABLE ANGLE: R/L/U/D 85/85/85/85 (Min)
12
Service Manual
Another Type
XTV MONITOR( Conforms to EN60601-1/EN60950, EN60601-1-2 )
MODEL NF17F6AAD
MANUFACTURER NICAL SPA
POWER SUPPLY 230VAC±15%, 115VAC±15%
MONITOR SIZE: 17”
READOUT: FRAME
SCANNING LINES: 525LINE/60Hz
SCANNING SYSTEM: 2 : 1 INTERLACED
PICKUP DEVICE: 1/2 INCH CCD CAMERA
13
Service Manual
IMAGE INTENSIFER
( Conforms to EN60601-1:1993, EN60601-1-2:, IEC1262-1 to 7 : 1994 )
NAME : THALES TH 9428 HP2
INPUT SCREEN DIAMETER : 230/160/120 mm
OUT SCREEN DIAMETER : 20±0.5 mm
DQE : 65%
(Detective Quantum Efficiency)
RESOLUTION : 56 lp / Cm
CONTRAST : 25 / 1
CONVERSION FACTOR : 120 (cd / m2)
ACCESSORIES
The accessories delivered with the basic configuration of the KMC-950 consist of :
14
Service Manual
2. SERVICE DATA
System Circuit Diagram
15
Service Manual
16
Service Manual
Error Code
If there occurs a specific error in the machine, an appropriate error message will be displayed
on the graphic TOUCH SCREEN for easy identifiability:
17
Service Manual
BOARD LAYOUT
MAIN B/D LAYOUT
CAMERA COM
CN19 JP4,5 SELECTOR CN15 CN16
SW3
touch
CN4
CN11
CN18
CN3
CN2
CN9
TEMP_JP3
NOT USE
CN20 CN14
CN21 CN5
CN6
CN7
CN22
CN8
18
Service Manual
Main Board Connector
All connector
PCB point
No. 1 1
#1 CN 1 DC POWER
#2 CN 2 HFG INTERFACE
#4 CN 4 INTERFACE BOARD
#5 CN 5 EMERGENCY S/W
# 10 CN 10 COLLIMATOR
# 11 CN 11 COLLIMATOR SENSOR
# 12 - - -
# 13 - - - -
# 14 CN 14 OP PANEL
# 15 CN 15 RS232 PC
# 16 CN 16 RS232 TOUCH
# 17 CN 17 TUBE OVERHEAT
#18 CN 18 ELEVATOR
#19 CN 19 LASER
19
Service Manual
[ DIP switch 3 ]
DIP Switch 3
Switch 1 OFF
Switch 2 ON
Switch 3 ON
Switch 4 ON
20
Service Manual
F5 F4 F3 F2 F1
1. Fuse board
2. Generator
21
Service Manual
3. SMPS(POWER SUPPLY)
PSU BOARD
Fuse # 1 FUSE 1 3.15 A/250V
MAIN POWER
WARNING
- When you change fuse
1. You should check the rating, power, part no., and marking according with the manual.
2. It should be performed by an expert or an engineer who is well trained.
22
Service Manual
23
Service Manual
X-Ray CHECK
MAIN BOARD REFERENCE SIGNAL
This procedure is the adjustment and check job within Main Board. (KM9EEXT005)
24
Service Manual
25
Service Manual
MAIN BOARD SCOPE SIGNAL
26
Service Manual
27
Service Manual
28
Service Manual
X-Ray GENERATOR
This procedure is the adjustment and check job within HFG Control Inverter Unit (KM9EEXT005)
and High Voltage Tank Unit (KM9MEEXT005-4) for X-RAY generation.
There will be few cases in which the following checking and tuning of HFG unit is needed in the field. If you want to
check the details of HFG, you should uncover the back and left side cover of KMC-950 and then uncover the HFG
Inverter unit in the left side. Appropriate measurement devices, such as oscilloscope, DMM (Digital Multi Meter) etc.
are required to check the signals.
WARNING
- To avoid the electrical shock, you should be very careful in dealing the high voltage
connectors (Anode, Cathode) on tank and tube. Even after the main power is down, high
voltage charge still remains in tank and tube for about a few minutes.
- High voltage connectors should always be tightly fastened before the main power is
turned on.
29
Service Manual
Functional Blocks for X-RAY generator of KMC-950
X-RAY
Main Control Board X-RAY Tube
(KM9EPCB001) (Varian Rad99)
Anode Cathode
EP_DATA
X-RAY_ON
ROTOR ON
IP_DATA
L/S
IP_FB
EP FB
C(-KV/2)
A(+KV/2)
Small Vf
Large Vf
HV Calbe
Up to 62.5KV
30
Service Manual
X-RAY generator Feedback Stage in X-RAY generator of KMC-950
31
Service Manual
32
Service Manual
33
Service Manual
X-RAY generator Feedback inverter
EP_ kV feedback (Inverter TP 6 ) for 40KV IP_ mA feedback (Inverter TP 9 ) for 1mA
In Fluoroscopy(Small) In Fluoroscopy(Small)
EP_ kV feedback (Inverter TP 6 ) for 125KV IP_ mA feedback (Inverter TP 9 ) for 5mA
In Fluoroscopy(Small) In Fluoroscopy(Small)
34
Service Manual
EP_ kV feedback (Inverter TP 6 ) for 40KV IP_ mA feedback (Inverter TP 9 ) for 8mA
In Fluoroscopy(Small) In Fluoroscopy(Small)
EP_ kV feedback (Inverter TP 6 ) for 40KV IP_ mA feedback (Inverter TP 9 ) for 20mA
In Fluoroscopy(Large) In Fluoroscopy(Large)
35
Service Manual
36
Service Manual
WARNING
The measured data could be changed a little depend on each system(KMC-950)
37
Service Manual
X-Ray Beam Adjustment
Move the RAD Mask on the x-ray tube to focus X-Ray to center after x-ray calibration.
Method:
1. Set to Fluoro mode of 60KV, 3mA (or RAD mode, 40KV, 30mAs).
2. Stick the fluorescent plate on the I.I.
3. Make environment dark to watch the fluorescent.
Use the protector or black curtain. - Dark room is also useful.
4. Move RAD Mask so that the circle A on the fluorescent plate is
placed to center of X-Ray Beam.
(Fluorescent plate)
형광판
38
Service Manual
X-RAY Filed Size of KMC-950
The following figure is device for radiography mode. For more detailed information, refer to the
chap. 15 in OPERATION MANUAL(Doc.No. : KMM-I-0408-17)
39
Service Manual
SETUP(X-Ray ADJUSTMENT)
1. SETUP MENU
For the open Setup mode is selected function mode right side bottom area.
select
area
For the open Radiography X-ray Calibration mode is selected CALIBRATION button in
Setup mode.
40
Service Manual
For the open Fluoroscopy X-ray Calibration mode is selected FOCUS CHANGE button in
Calibration mode
90
80
70
60
mA 50
Change mA Point setting
data
40
30
20
40 50 60 70 80 90 100 110 120
kV point
41
Service Manual
2. Press the “offset”, “gain” on the touch screen, input the DATA(0-255) by using TEN-KEY, and
then adjust the left & right arrow button till the kVp displays “40”.
42
Service Manual
4. If the early data of feedback is high, re-adjust the offset data, and if the early data of feedback
is low, raise the offset data by using right-sided arrow button on the touch screen. If the latter
data of feedback is lower than early one, raise gain by using the right-sided arrow button, if
the latter data of feedback is higher than the early one, lower the gain by using the left-sided
arrow button on the touch screen. By the way, if the kVp measurement is different each other,
press the Arabic-button, and re-input the data by changing each 5
5. After re-adjusting, generate X-ray again, monitor a feedback type and compare it with
established data. If the difference of them is large, repeat the process of NO.5
6. When the installed data and real data is similar each other, raise the kVp data as 5 kVp by
using right-sided arrow button.
8. When these processes are finished, press the “SELECTOR” button on the left-lower sided,
change a number, and then repeat the adjustment process of NO. 3-6.
43
Service Manual
Radiography select range chart
44
Service Manual
Fluoroscopy X-ray calibration
45
Service Manual
4. Generate X-ray with pressing foot s/w, and compare the feedback data with installed one. If
the measured data is lower than the installed one, change the data as same method of NO. 2.
5. After adjusting, generate X-ray again, monitor the feedback type, and compare it with
installed one. If the difference of them is large, repeat the process of NO.5
46
Service Manual
8. When these processes are finished, adjust the kVp to “50-120” with right-sided arrow button,
and repeat the adjustment process of NO. 4-7.
9. When these processes are finished, press the “SELECTOR” button on the left-lower sided,
change a number “2-5” and then repeat the adjustment process of NO. 4-8.
47
Service Manual
RADIOGRAPHY
FLUOROSCOPY
48
Service Manual
90
80
70
60
50
40
30
20
40 50 60 70 80 90 100 110 120
49
Service Manual
50
Service Manual
APPENDIX I - MECHANICAL ASSEMBLIES
1. C-ARM COVER ASSEMBLY
2. C-ARM AND I.I. ASSEMBLY
3. C-ARM SYAY STOPPER ASSEMBLY
4. MOTOR, CCU AND HFG PART ASSEMBLY
5. IMAGE INTENSIFIER AND CAMERA HEAD ASSEMBLY
6. IMAGE INTENSIFIER POWER SUPPLY ASSEMBLY
7. C-ARM ROTATION LOCK ASSEMBLY
8. C-ARM ORBIT REVOLUTION LOCK ASSEMBLY
9. C-ARM SLIDING BEARING ASSEM BLY
10. TUBE COVER ASSEMBLY
11. TUBE AND COLLIMATOR ASSEMBLY
12. CART ACCESSORIES ASSEMBLY
13. CART MONITOR & CYLINDER ASSEMBLY
14. CART COVER ASSEMBLY
51
KMC-950 CONSOLIDATED MANUAL V1.5
53
KMC-950 CONSOLIDATED MANUAL V1.5
54
KMC-950 CONSOLIDATED MANUAL V1.5
2. C-ARM AND I.I. ASSEMBLY
55
KMC-950 CONSOLIDATED MANUAL V1.5
56
KMC-950 CONSOLIDATED MANUAL V1.5
3. C-ARM SYAY STOPPER ASSEMBLY
6
5
4
S1
3
8
S3
2
7
S2
57
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 FR-ARM T-ARM Ass'y 1
2 KM9MSTA001 STAY 1
3 KM9MTAR010-1 STAY-SLIP 1
4 KM9EEXT037 U FINE NUT 1
5 KM9MSTA002 S-BR-CLAMP1 1
6 KM9MSTA003 S-BR-GUIDE 1
7 KM9MSTA003-1 S-BR-SPRING 4
8 KM9MSTA004 S-BR-CLAMP2 1
9 KM9MSTA004-1 S-BR-LEVEL 1
58
KMC-950 CONSOLIDATED MANUAL V1.5
4. MOTOR, CCU AND HFG PART ASSEMBLY
59
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9MBOD002-1 BODY MC BK 1
2 KM9MBOD011 FIXING ANGLE 2
3 KM9MBOD022-5 CON BK ASSY 1
4 KM9EEXT005-4 HIGH VOLTAGE TANK 1
5 KM9EEXT016 MAIN TRANSFORMER 1
6 KM9EEXT005 INVERT HFG 1
7 KM6EBED000-1 CONTROLLER 1
60
KMC-950 CONSOLIDATED MANUAL V1.5
5. IMAGE INTENSIFIER AND CAMERA HEAD ASSEMBLY
61
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 Fig.5 Thales I.I Ass'y 1
2 KM9EPCB010 comed FC 1K CAMERA 1
3 KM9MCAM020 LENS I.I BRACKET 1
4 KM9MCAM021 CAM1K PWR BRACKET 1
5 KM6MCAR012-6 650 THALES 9" WG BK 1
6 KM6MCAR012-4 650 9" THALES SUB WG 1 1
7 KM9MCAR030 1K II COVER 1
8 KM9EEXT026 GRID 1
9 KM9MCAR011-2 II-COVER-SPACER 4
10 KM9MCAR011 II-COVER 1
11 KM9MCAR011-1 II-COVER BOLT 4
62
KMC-950 CONSOLIDATED MANUAL V1.5
63
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9EEXT001 Thales I.I 1
2 KM9MCAR012 I.I Bracket1 1
3 KM9MCAR012-1 I.I Bracket2 1
4 KM9MCAR003 TUBE I.I BK 1
64
KMC-950 CONSOLIDATED MANUAL V1.5
65
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 Fig. 7 Stay Ass'y 1
2 KM9MSTA005-2 SS-BR-SLIP 2
3 KM9MSTA015 HANDLE (L/R) 1
4 KM9MSTA001-5 C-Arm Stay Cover 2
5 KM9MSTA001-6 STAY-W-COVER 1
6 KM9MSTA016 S-CABLE HOLDER Ass'y 1
7 KM9MSTA001-2 S-COVER1 1
8 KM9MSTA016 CABLE FIXING SCREW 1
66
KMC-950 CONSOLIDATED MANUAL V1.5
8. C-ARM ORBIT REVOLUTION LOCK ASSEMBLY
67
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 Fig. 8 Stay Ass'y 1
2 KM9MSTA011 C-ARM STOPPER Ass'y 1
3 KM9MSTA010-1 C-BR-SHAFT2 1
4 KM9MSTA010-2 C-BR-RING 2
5 KM9MSTA010 Lock Arm 1
6 KM9MSTA005-2 S-BR-SLIP 2
7 KM9MSTA012 C-BR-SHAFT1 1
8 KM9MSTA013 STOPPER Ass'y 1
9 KM9MSTA014 CAM RING 1
10 KM9MSTA014-1 STOPPER RING 1
68
KMC-950 CONSOLIDATED MANUAL V1.5
9. C-ARM SLIDING BEARING ASSEMBLY
69
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9MSTA001 Stay 1
2 KM9MSTA008 C-Arm BR Shaft (2) 2
3 KM9MSTA007 C-Arm BR Shaft (1) 2
4 KM9MSTA009-14 UPPER BR 3 Ass'y 4
5 KM9MSTA009-13 LOWER BH BASE Ass'y 4
6 KM9MSTA009-11 LR BR 4 HOLDER Ass'y 4
7 KM9MSTA009-18 UPPER BR HOLDER FIX 4
8 KM9MSTA009-17 C-ARM TRAVEL STOPPER 2
9 KM9MSTA007-2 BR SHAFT STOPPER 16
10 KM9MSTA009-16 LR BR HOLDER SHAFT 4
70
KMC-950 CONSOLIDATED MANUAL V1.5
10. TUBE COVER ASSEMBLY
71
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 Fig.10 TUBE ASS’Y 1
2 KM9MCAR007 TUBE COVER 1
3 KM9MCAR007-4 CORN 1
4 KM9MCAR007-3 CORN COVER 1
72
KMC-950 CONSOLIDATED MANUAL V1.5
11. TUBE AND COLLIMATOR ASSEMBLY
73
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9MCAR004 TUBE BK 1
2 KM9MCAR004-1 Block 2
3 KM9MCAR007-1 Tube COVER2 1
4 - Tube FRONT COVER 1
5 KM9EEXT003 Tube 1
6 KM9MCAR004-4 Tube Al Filter 1
7 KM9MCAR004-5 Tube CIRCULAR PLATE 1
8 KM9MCOL002 Collimater-9” 1
9 KM9MCAR004-2 Collimator Fix Washer 3
74
KMC-950 CONSOLIDATED MANUAL V1.5
12. CART ACCESSORIES ASSEMBLY
75
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9MCRT040-2 SIDE COVER 2
2 KM9MCRT043-12 PC FRONT BRKT 2
3 KM9EEXT124 SONO PRINT (A6 SIZE) 1
4 KM9EDIS001 COMPUTER 1
76
KMC-950 CONSOLIDATED MANUAL V1.5
13. CART MONITOR & CYLINDER ASSEMBLY
77
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9MCRT044-1 CYLINDER BOTTOM HOUSING 1
2 KM9MCRT044-6 CYLINDER UNDER COVER 1
3 KM9MCRT044-7 CYLINDER UPPER COVER 1
4 KM9MCRT044-2 CYLINDER TOP HOUSING 1
5 KM9MCRT044-5 MONITOR UP DN HANDLE 1
6 KM9MCRT044-3 MONITOR BLOCK 1
7 KM9MCRT044-8 SPACE RING 1
8 KM9MCRT044-4 MONITOR ROATE SHAFT 1
9 KM9EEXT126 MONITOR ARM 1
10 KM9EDIS015-4 MONITOR 19” 2
78
KMC-950 CONSOLIDATED MANUAL V1.5
14. CART COVER ASSEMBLY
79
KMC-950 CONSOLIDATED MANUAL V1.5
ITEM NO. PART NO. UNIT NAME QTY. DESCRIPTION
1 KM9MCRT040-5 DRAWER 1
2 KM9MCRT040-1 REAR COVER 1
3 KM9MCRT040-3 PRINTER COVER 1
80
KMC-950 CONSOLIDATED MANUAL V1.5
CCD camera
Counterweight
Image intensifier
Service Manual
81
KMC-950 CONSOLIDATED MANUAL V1.5
TECHNICAL TS 61 417 355
SPECIFICATION TH 9428 HP2 H211 VR13
X-RAY IMAGE INTENSIFIER 2003-11-27 Page 1
SUMMARY
This specification contains:
-Introduction, particular features, regulatory requirements page 1
-Electro-optical characteristics pages 2, 3 and 4
-Appendix:
• outline drawing : 8 324 699 E
• wiring and connection : 8 324 663 E
INTRODUCTION
TH 9428 HP2 H211 VR13 is a triple field X-Ray Image Intensifier for medical applications
with a nominal entrance field size of 230 mm and an output image diameter of 20 mm.
Its main components are:
-X-Ray Image Intensifier tube
-mu metal magnetic shield
-lead radiation shield
-painted housing: Thales color reference: T23
-power supply: HV type TH 7195
Technical Specification ref. TS 4989-161 and
User’s Manual TIV/1M.1166/96
PARTICULAR FEATURES
Environmental data
With the maximum allowed static load on the output flange (see outline drawing) the mechanical
shocks should not exceed 1 g (6 ms duration).
REGULATORY REQUIREMENTS
DHHS 21 CFR 1020-30 IEC 60601-1; IEC 60601-1-2 UL 2601 CSA 22-2 n° 60601-1
Marking: CE, UL
82
KMC-950 CONSOLIDATED MANUAL V1.5
83
KMC-950 CONSOLIDATED MANUAL V1.5
ELECTRO-OPTICAL TS 61 417 355
CHARACTERISTICS TH 9428 HP2 H211 VR13
X RAY IMAGE INTENSIFIER 2003-11-27 Page 3
Average background luminance
The average background luminance shall be less than 0.02 cd/m² in all modes.
Just after switching on, or after switching modes, the XRII may exhibit some local luminance
fluctuations.These momentary fluctuations should not exceed 0.1 cd/m² and normally disappear after a
few minutes.
Image shift
The center of the entrance field in the magnifying modes should not shift more than 5 mm (typically 2
mm) with respect to the center of the entrance field in the normal mode of operation.
Output screen spectral response
The output screen contains a P20 type phosphor with a peak wavelength between 520 and 540 nm and a
full width at half maximum of less than 100 nm.
Optical
The thickness of the output window is 3.6 ± 0.2 mm.
The index of refraction of the output window is 1.50 ± 0.02 at 587 nm
The distance of the image plane to the pilot plane will be everywhere within ± 30 µm from the value at
the
center of the image.
Rise and decay time
At an exposure rate of 5 mR/s in the normal mode the luminance of the output screen will be more than
50 % and more than 80 % of its final value respectively 1 ms and 10 ms after starting the exposure. After
a continuous exposure rate of 5 mR/s in the normal mode, the luminance of the output screen will be less
than 30 % and less than 10 % of its initial value respectively 1 ms and 10 ms after stopping the exposure.
Minimum pulse width (valid when used with THALES power supplies only) The minimum pulse width
per DSA pulse is typically 20 ms. The minimum pulse width per cine pulse is typically 2 ms.
Blanking mode
The output luminance of the image intensifier can be blanked to a level less than 5 % (typically 1 %)
with respect to the unblanked mode in all fields, by applying a blanking signal to the power supply unit.
The blanking time constant is less than 1 ms.
84
KMC-950 CONSOLIDATED MANUAL V1.5
Getter anode (G4) U4 2.5 to 3.0 kV
Photocathode (blanking) UPC 1000 V
Operating voltage
* for optimum performances U1 and U2 should be adjusted for each mode.
Maximum electrical ratings
XRII maximum input voltages UPC = 1200 V (blanking)
U1 = 350 V
U2 = 1.5 kV
U3 = 13 kV
U4 = 3.5 kV
U5 = 31 kV
Photocathode maximum continuous current IPC = 5 µA
Photocathode maximum pulsed current IPC = 50 µA
Environmental data
The detailed description of the environmental testing conditions is given in TII/4.100/90.
The main characteristics are summarized hereunder :
- climatic
-operating temperature range +5°C to + 40°C
-storage temperature range -10°C to + 55°C
85
KMC-950 CONSOLIDATED MANUAL V1.5
86
KMC-950 CONSOLIDATED MANUAL V1.5
87
KMC-950 CONSOLIDATED MANUAL V1.5
88
KMC-950 CONSOLIDATED MANUAL V1.5
1. CCD CAMERA
1.1 Camera power supplay
89
KMC-950 CONSOLIDATED MANUAL V1.5
Power board
Remove 3 screws
Camera removal
Camera connections
90
KMC-950 CONSOLIDATED MANUAL V1.5
The camera is put onto the output window and is fixed by the mounting ring using 3 screws.
91
KMC-950 CONSOLIDATED MANUAL V1.5
1.2.4 Adjustment of the optical focus
• Switch off the unit.
• To focus the camera lens, the I.I cover has to be removed.
• For that, loosen the screws at the camera housing.
• Switch on.
• Center a line pair phantom to the input window.
The line pairs should be 45° to the monitor line structure and 90° to the scattered radiation grid
• Loosen the locking screw. Set to magnifier 2 format and adjust to maximum resolution.
And then magnifier 1,N focus adjustment.
• Switch to full field size and check edge focus. Tighten locking screw hand screwed.
• Assemble I.I cover
Locking screw(3)
for focusing ring
focusing ring
Locking screw(3)
for zoom
92
KMC-950 CONSOLIDATED MANUAL V1.5
93
Service Manual
- PARTⅠ. SCHEMATIC -
95
Service Manual
1) KMC-950 OP board
96
Service Manual
2) KMC-650 Main CPU board
AD[0..15] AD[0..15] AD[0..15]
U31A
1 2 +24V
SA[0..7] SA[0..7] 74LS07
U5 U31F
CN4
AD0 2 19 SA0 SD[0..7] SD[0..7] 13 12 CCU_FUNCTION 0 1
+12V AD1 3 D0 Q0 18 SA1 74LS07 14
+5V AD2 4 D1 Q1 17 SA2 CCU_FUNCTION 1 2
AD3 5 D2 Q2 16 SA3 U31B 15
D3 Q3
74HC573
ADIN7 AD4 6 15 SA4 3 4 CCU_FUNCTION 2 3
D4 Q4 +5V
+5V R6 +5V AD5 7 14 SA5 74LS07 FD_CCU 16
R2 AD6 8 D5 Q5 13 SA6 CCU_FUNCTION 3 4
AD7 D6 Q6 SA7 AR4 U31E VD_CCU
4.7K 1K 9 12 17
U21
2
R4 D7 Q7 1 XU3 U10 11 10 CCU_SUB 0 5
R
2 6 3 VR1 1 ALE 11 2 SA0 SA0 10 11 SD0 SD0 2 19 74LS07 18
4.7K RESIN RST 5 D1 1 LE 3 SA1 SA1 9 A0 O0 12 SD1 SD1 3 D1 Q1 18 CCU_SUB 1 6
3 RST LM336_5V 10K VAR OE 4 SA2 SA2 8 A1 O1 13 SD2 SD2 4 D2 Q2 17 U31C 19
1 CT +5V SA3 SA3 A2 O2 SD3 SD3 D3 Q3 CCU_SUB 2
+ C14 74ALS573 5 7 15 5 16 5 6 7
REF 8 0.1uF 0.1uF 6 SA4 SA4 6 A3 O3 16 SD4 SD4 6 D4 Q4 15 74LS07 20
VCC 10uF A4 O4 D5 Q5
+ C28 4 7 7 SA5 SA5 5 17 SD5 SD5 7 14 CCU_SUB 3 8
GND SEN +5V A5 O5 D6 Q6
10uF 8 SA6 SA6 4 18 SD6 SD6 8 13 U31D 21
0.1uF C27 U6 AR6 9 SA7 SA7 3 A6 O6 19 SD7 SD7 9 D7 Q7 12 9 8 9
TL7705A 1 AD8 25 A7 O7 D8 Q8 74LS07 22
R
0.1uF AD0 2 18 SD0 SD0 2 AD9 24 A8 IOCS1 11 ACK_CCU/FREEZE-B10
37
1
XU1 AD1 3 A1 B1 17 SD1 SD1 3 6.8K AD10 21 A9 CLK U32C 23
+5V AD2 4 A2 B2 16 SD2 SD2 4 AD11 23 A10 OUT_EN 1 5 6 11
VPP
VCC
R5 AD3 5 A3 B3 15 SD3 SD3 5 AD12 2 A11 OE 74LS07 ABC 24
74HC245
60 AD0 AD4 6 A4 B4 14 SD4 SD4 6 AD13 26 A12 74HC574 12
+5V P3.0/AD0 59 AD1 AD5 7 A5 B5 13 SD5 SD5 7 AD14 27 A13 U32D 25
P3.1/AD1 AD2 AD6 A6 B6 SD6 SD6 A14
80C196KC20
10K 58 8 12 8 9 8 13
C10 U20 C12 3 P3.2/AD2 57 AD3 AD7 9 A7 B7 11 SD7 SD7 9 SRD 22 74LS07
NMI P3.3/AD3 AD4 A8 B8 CS_ROM 20 OE
+
22uF 2 1 22uF 56
+V C1+ P3.4/AD4 55 AD5 RD- 1 CE U32A CCU_DB25_FEMALE
+
3 16 P3.5/AD5 54 AD6 19 DIR 6.8K M27C256B 1 2 CCU_STROBO
C11 6 C1- 4 RESET P3.6/AD6 53 AD7 OE 74LS07 1K_LP용 CUT
22uF -V C2+ 22uF P3.7/AD7 74HC245
+
+
5 C13 COL_LEAF_ROT_SV( ADIN0) 6 U32F
CN16 C2- ADIN1(mA F/B) 5 ACH0/P0.0 52 AD8 XU4 U11 13 12 CCU_FREEZE GATE
1 13 12 P2.1 COL_LEAF_SV(TEMP_SV) 7 ACH1/P0.1 P4.0/AD8 51 AD9 SA0 10 11 SD0 SD0 2 19 74LS07 1K_LP용 CUT
MWR
2 14 R1 IN R1 OUT 11 P2.0 ABC 4 ACH2/P0.2 P4.1/AD9 50 AD10 SA1 9 A0 D0 12 SD1 SD1 3 D1 Q1 18
3 T1 OUT T1 IN FV++ 11 ACH3/P0.3 P4.2/AD10 49 AD11 SA2 8 A1 D1 13 SD2 SD2 4 D2 Q2 17 U32B
XU7
8 9 FV-- 10 P0.4/ACH4 P4.3/AD11 48 AD12 AD9 1 12 SWR SA3 7 A2 D2 15 SD3 SD3 5 D3 Q3 16 3 4 CCU_ROT-CW
7 R2 IN R2 OUT 10 COL_IRIS_SV 8 P0.5/ACH5 P4.4/AD12 47 AD13 AD10 2 I0 F0 13 SRD SA4 6 A3 D3 16 SD4 SD4 6 D4 Q4 15 KEY _D0 74LS07
EASY VIEW T2 OUT T2 IN ADIN7 9 ACH6/P0.6 P4.5/AD13 46 AD14 +5V AD11 3 I1 F1 14 CS_RAM SA5 5 A4 D4 17 SD5 SD5 7 D5 Q5 14 KEY _D1
MAX232 ACH7/P0.7 P4.6/AD14 45 AD15 AD12 4 I2 F2 15 CS_ROM SA6 4 A5 D5 18 SD6 SD6 8 D6 Q6 13 KEY _D2 U32E
P2.0 18 P4.7/AD15 AD13 5 I3 F3 16 SA7 3 A6 D6 19 SD7 SD7 9 D7 Q7 12 KEY _D3 11 10 CCU_ROT-CCW
CN15 P2.1 17 P2.0/TXD(out) 43
AR7 AD14 6 I4 F4 17 AD8 25 A7 D7 D8 Q8 74LS07
1 VD 15 P2.1/RXD(in) READY 65 1 AD15 7 I5 F5 18 CS_0 AD9 24 A8 IOCS2 11
R
2 OVER_H 44 P2.2/EXINT(in) CLKOUT 41 WR- 2 RD- 8 I6 F6 19 IORW AD10 21 A9 CLK
3 42 P2.3/T2CLK(in) BHE 40 WR- 3 WR- 9 I7 F7 AD11 23 A10 OUT_EN 1
HE_P2.5 39 P2.4/T2RST(iN) WR 61 RD- RD- 4 11 I8 AD12 2 A11 OE
R3 +5V P2.6 33 P2.5/PWM0(out) RD 62 ALE ALE 5 I9 AD13 26 A12 74HC574
DIS PC P2.7 38 P2.6/T2Up-dOWM ALE 63 AD14 1 A13 SW3
16V8
P2.7/T2Capture INST A14
4.2K
4.7K ACK_HI0 24 19 P1.0 SRD 22
C? 25 HSI.0 P1.0 20 P1.1 SWR 27 OE
104 FD_HI2 26 HSI.1 P1.1 21 P1.2 CS_RAM 20 WE FV++
HI3 27 HSI.2/HSO.4 P1.2 22 P1.3 CS1
C4 HSI.3/HSO.5 PWM1/P1.3 23 P1.4 24257AK-15
PWM2/P1.4 30 P1.5
P1.5
11.0592
20 68
VCC 36 GND1 28 HO0 U8 1 SW DPDT/SM
R
27pF 1 14 GND0 HSO.0 29 SA5 1 15 IOCS0 IOCS0 2 IOCS0 IOCS0
+5V RST/VPP Vss HSO.1 A Y0
C3 2
EA HSO.2
34
+5V
SA6
SA7
2
B Y1
14 IOCS1
IOCS2
IOCS1 3
IOCS2 4
IOCS1
IOCS2
IOCS1
IOCS2
ADD Part's
4 12 35 3 13
XTAL2 ANGND HSO.3 C Y2
A682G
5 12 IOCS3 IOCS3 5 IOCS3 IOCS3
X1
X2
XTAL1 AR2 AD8 Y3 IOCS4 IOCS4 6 IOCS4 IOCS4
1 6 11
G1 Y4
R
67
66
R
3 P1.1 P1.1 13 P1.0/AIN0 2 80C196KC20 ACK_HI0 2 CS_0 5 G2A Y5 9 IOCS6 IOCS6 8 IOCS6 IOCS6 R89 +12V
P1.1/AIN1 P3.0/RXD G2B Y6
6.8K
1
HE_P2.5 ACK VD FD
74HC125 SD[0..7] SD[0..7] R87 100
VD FD_HI2 FD_CCU
C38
OVER_H +5V
+5V AR12
101 R86 100
AR1 VD VD_CCU
1 U14
R
1 2 RTN0 RTN0 2 18 SD0 C37
RTN0
RTN1
RTN2
RTN3
RTN4
RTN5
RTN6
RTN7
SCAN0
SCAN1
SCAN2
SCAN3
A1 Y1
R
10
11
12
13
14
1
2
3
4
5
6
7
8
9
8 A2 9 RTN7 RTN7 9 A7 Y7 11 SD7
9 OVER_H 7 A8 Y8
WP 1 IOCS6
1K G1 19
6.8K G2
AT24C64 CN14 74HC541
HIRO-14(OP PANEL)
SHUTTER
+24V IRIS_COLLIMATOR
PREVIEW 패 턴 수 정
CN11
1
2
3
4
+12V로 변 경 I.I SELECTOR3 TP1
IRIS_COLLIMATOR CUT RED RED CN3 I.I SELECTOR2 ?
1
+5V 연 결 D? SHUTTER6 R20 XU30A
COL_LEAF_ROT_SV( ADIN0)
3
5.6V
1 2 COL_IRIS_SV 부품 제거 +24V D2 LM324
5 3 4 1 5 U28 C? R99
1
2
3
4
C19
0.1u
C20
0.1u
-
+ C5 + 3 1 16 SHUTTER6 47K
1B 1C
3300uF/35V
3300uF/35V
100uF/35V 4 2 15 0.1u
AC14V GND 연 결 D15 D14 1 3 2B 2C 14 ROTOR_ON TP3
3B 3C
R16
200/2W
1
5B 5C
1
+12V IRIS_COLLIMATOR CUT R95 R96 PA1A-24V 1N4007 PREVIEW CUT R83 22K I.I SELECTOR3 U12_13
U12_12
6
6B 6C
11 I.I SELECTOR2
I.I SELECTOR3
R23 XU30B
1N49371N4937 3K 3K 7 10 1K - 6 R21
2 R84 22K I.I SELECTOR2 7B 7C ADIN1(mA F/B) 7 4.7K
3 U9 9 4.5 6 9 + 5
4 KSDA02 COM
C21
0.1u
C23
0.1u
5.6V
5
+ C6 IRIS COLLIMATOR용 패 턴 수 정 RLY 2 부품 제거 SD0
SD1
2
D1 Q1
19 2 1 0 0 D3 LM324
100uF/35V 5 3 18 ULN2003A 3 1 1 0 C? R100
6 3 1N4007 SD2 4 D2 Q2 17
7 4 SD3 5 D3 Q3 16 0.1u
47K CN2
1 D13 SD4 6 D4 Q4 15 U16 KVP_DATA 1
CON7 DIPSW2 D5 Q5
C22
0.1u
C24
0.1u
8
SDO
C26
0.1u
C25
0.1u
4
100uF/35V CN19 +5V 74HC595 3
+
XU30D R27 13
1
1
U23
R19 R37 20K R61 10K 1 12 1K Kv 14
+
1
2
3
4
5
6
7
8
9
2 6 OUT_EN 1 2 14 15
RESIN RST LASER
R -
4.7K 5 2 R38 20K R62 10K 13 16
SD[0..7]
RST -
3 IRIS_COLLIMATOR U17 LM358 LM324 R_F 17
4
1 CT +5V 패턴수정 AR8 14 15 R39 20K R63 10K 18
11
REF SDI QA
ELEVATOR 패 턴 수 정 +24V VCC
8 R19 LASER 6.8K 11
SRCLK QB
1 C46 19
+ C29 4 7 4.7K +12V/+5V로 변 경 +12V 12 2 R40 20K R64 10K 20
GND SEN +24V Q1 RCLK QC 3 R28 20K 100p
10uF QD
0.1uF +5V 4 R41 20K R65 10K
TL7705A +5V 1 3 +5V_7805 +5V_7805 QE 5
VIN VOUT QF HFG
AR9 A682G 10 6 R42 20K R66 10K C31 100p
D? RLY 3 D? RED 3K XU23 OUT_EN 13 CLR QG 7 TP3
8
R92 연결 16 6 R G QH U38B
1N4007
4
5 9 VBB VCC 2 7805 + C43 9 5
+
XU30C R29
1
2
3
4
5
6
7
8
9
1
3 VBB VREF SDO R43 20K R67 10K 7 10 1K mA
100uF/35V +
4 D3 1N4007 74HC595 6 8
+ = CLOSE LEAF -
1 10 1 U12 R44 20K R68 10K 9
OUTA BRAKE U12_19 SD0 -
2 15 8 19 2 LM358 LM324
4
OUTB EN 7 U12_18 18 Q1 D1 3 SD1 R45 20K R69 10K
CN18 + C?
11
PH U12_17 17 Q2 D2 4 SD2 C45
Q3 D3 ?uF
1 PA1A-24V 3 U12_16 16 5 SD3 R46 20K R70 10K
2 11 RC U12_15 15 Q4 D4 6 SD4 U18 R30 20K 100p
3 SENSE U12_14 14 Q5 D5 7 SD5 14 15 R47 20K R71 10K
4 C40 R10 R11 14 U12_13 13 Q6 D6 8 SD6 11 SDI QA 1
MODE U12_12 12 Q7 D7 9 SD7 12 SRCLK QB 2 R48 20K R72 10K C34 100p
A3952 Q8 D8 RCLK QC 3
HEADER 4 RLY 4 D? RED 3K 330p 3K 0.5/2W 11 IOCS4 QD 4 R49 20K R73 10K PREHEAT
R93 CLK +5V_7805 QE 5 XU47A
1
5 1 OUT_EN 10 QF 6 R50 20K R74 10K +12V LM358
OE CLR QG TLP521-4
3 IRIS COLLIMATOR용 패 턴 수 정 +24V
+12V로 변 경
+5V
OUT_EN 13
G QH
7 D23 RED 1.2K
ERROR_3
4 D4 1N4007 74HC574 R51 20K R75 10K 9 8 R98 4 8
1 9 Q4E 4C
A472G SDO
+
2 CN10 XU24 AR10 R52 20K R76 10K
+12V 16 6 74HC595 10 7 D22 RED 1.2K
SD[0..7]
3
9 VBB VCC 2 R R53 20K R77 10K 11 Q4C 4A 6 R97 CHARGE_OK_E2
PA1A-24V 3 1 VBB VREF 74HC541 Q3E 3C
9
8
7
6
5
4
3
2
1
10 2 IOCS3 19 +5V R54 20K R78 10K
R101 7 3 + = CLOSE IRIS 10 1 1 G2 12 5 D21 RED 1.2K + C44
OUTA BRAKE G1 Q3C 3A
OVER_H
6K 1% 6 4 15 8 13 4 R96 FIL_OK_E1 0.47uF
U34A 2 5 OUTB EN 7 SD7 11 9 Q2E 2C
9 6 PH SD6 12 Y8 A8 8
SD5 Y7 A7 D20
3 R108 R109 7 3 +12V 13 7 14 3 RED 1.2K
+ RC SD4 Y6 A6 Q2C 2A ROTOR_OK_E4
1 8 11 14 6 15 2 R95
C? SENSE SD3 Y5 A5 Q1E 1C
+ C? 2
-
+5V 15 5
R102 R104 10K 1% 50K 1% C41 R12 R13 14 SD2 16 Y4 A4 4 R98
100uF/16V
1
3
SD5 7 D5 Q5 14 6 IN5 OUT5 13 MWR Q3E 3C
D6 Q6 IN6 OUT6
D? R103 U34B R106 + = CLOSE LEAF ROT. 10
OUTA BRAKE
1 SD6
SD7
8
D7 Q7
13 7
IN7 OUT7
12 ELEVATOR UP
ELEVATOR DOWN
ADD Part's U35A
6K 1% 5K 1% 15 8 R94 9 12 8 11 12 5 D17 RED 1.2K CN7
OUTB EN D8 Q8 IN8 OUT8 Q3C 3A
74HC08
CN9 1N4007 6 - R107 7 CN17 1K 13 4 R91 2
7 PH IOCS7 11 10 Q2E 2C 1 FLUORO
1 5 + 3 1 CLK COM
RC OUT_EN D16
2 10K 1% 11 2 1 ULN2803 14 3 RED 1.2K
C? SENSE OE Q2C 2A
+ C? R111 15 2 R90 CN5
H_ENABLE 2
1
D? OP490 5K 1% C42 R14 R15 14 74HC574 Q1E 1C EMG. 2
100uF/16V MODE
TEMP S 1N4007 0.1u OVER HEATING 1 EMERGENCY
HE_P2.5
A3952 16 1
330p 3K 0.5/2W Q1C $PIN0
U26
97
Service Manual
3) HTC-120 Block diagram
99
Service Manual
L2
Raector
3
RB
TB1
102K/630V
102K/630V
D68 C110
BRIDGE Q101 C1 C2 Q103 105/400V
AC L101
IGBT1 R8 R7 R6 R5 IGBT2
2W22
2W22
2W22
2W22
AC500V 63A C111
HV FUSE C102 C104 505/700V*2
HV TANK
50W10K
222/AC250V 222/AC250V + + R2
C109
505/700V
102K/630V
102K/630V
C106 CN1
C101 4700/400V 4700/400V C3 C4 CN2 CN1 CN1 8
104/AC250V C103 C105 104/AC250V C107 C108 15
222/AC250V 222/AC250V Q102 Q104 1 RY 102 RY 103 1 1 7
2 2 2
AC
R99
R1 R2 R3 R4 BR-950-02P 3
4
3
4
14
6 +HV
2W22
2W22
2W22
2W22
IGBT1 IGBT2 L9 CT101 13
50W10 BR-950-04P 5
RY 101 BR-950-04P 12
L104 4
CR101 4
11
3
10
1
MON(+,-) CT101 2
BR-950-02P
3
2
1
4
3
2
1
3
2
1
3
2
1
3
2
1
3
2
1
CN4 D C B A CN1
35312-0410 35312-0310 35312-0310 35312-0310 35312-0310 35312-0310
IGBT
DRIVER
HIF3BA-14PA-2.54DSA
35312-0310
CN3 CN2
10
11
12
13
14
1
2
3
1
2
3
4
5
6
7
8
9
KV DR1
KV DR2
KV DR3
KV DR4
OV CYRRENT
CHARGE MON
+12V
-12V
GND
GND
GND
GND
X-RAY TUBE
CN2
CN8
1
2
3
4
1
2
3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
1
2 1
3
CN1 CN5
2 FILAMENT
53258-04 35312-0310 CN6 HIF3BA-14PA-2.54DSA
4 3
5 4
6 5
7
CONTROL SYGNAL 8
53259-05
9 CN7
53259-09
1
CN3
2 FEED BACK
1 3
2 4
MONITOR SYGNAL 3 5
4
TO SYSTEM 5
53259-05
6 TB102
53259-06
1 ROTOR1
CN4
2 ROTOR2
+12V 1 3 ROTOR3
-12V 2
GND 3
BR950C-03P
GND 4 CN10
53259-04
1
2 C112
CN9
AC 220V 1
53258-02P 30uF/AC370V
AC RTN 2
POSKOM CO.,LTD
Title
작 검 승
53258-02P
성 토 인 PK-HTC-120
Size Document Number Rev
A3 00
MAIN
Date: Friday , September 10, 2004 Sheet 1 of 1
100
Service Manual
5) HT Control board
C5
BSE AC250V TB102
ROTOR3
1 ROTOR2
L2 2 ROTOR1
F1 RTR CONTROL Q1 600V16A BTA16 600B T1
3
2:300 CT1 TP1 BR950ELW-03
6.3A 660uH 5A(OB630 1PIE) Q2 R5
1
RTR OK
S 100 OHM 1mA 0.1V STN1A60
470 CN10
10mA/1V max10mA U1 1 C112
R4 2
+ C6 ZD3
R1
L 10 OHM 100mA 1V 6 4 100/16V
1N4733
53258-02
D15 1W680 30uF/370VAC
100mA/1V MAX150mA 1W100
R2 1/4W120
R3 W02
2 1 +12V
1K
TLP560
F-VIN
L1 F2 TP2
SP503L192 CER472/AC250V D3 S3L60U
T2
AC220V C4 50T 3.15 D1
1
470/400V
C1 C2 RS405 R6 R10 C13
RTN 2 B104/AC275V C10 L3 R13 C14 R14
CN9 B104/AC275V + C8 2W24K 2w4.7K 222/1KV 222/1KV 2W24K 222/1KV 2W24K
53258-02 C3 103/1KV TRANSFORMER
CER222/AC250V C C7 PQ3230 C11
TH1A 2004.5변경 D2
10D11 R7 Q3 330/200V D12 D13 RY 1
2W 1 R11 1N4937 1N4937 53258-06
UF5406
6 COM
TP7 R81 2K K904 1%330K
5 L1
53258-04P +12V
TP6 4 L2
CN1
3 S1
4 +12V 1N4744 R8 S S2
4 3 C30 ZD1 F-G1 F-G2 2
3 2 +12V 10 C12 TP4 TP5 L 1
2 2SA1020
INRUSH RY 1 105/250 CH2-MD12 CN8
1 D11 104
D4 Q7 Q8
1N4003 C9
+ C26 Q4 1N4937
IRF840 IRF840
U2A
8
4.7/16V R31 102
R33 3
Q5 1 + R9 1K
C2655 10K
2 D5 R15 R16 R17 R18
10K - 1N4937 R12 47 10K 47 10K
LM358 +12V 1%18K
D9 C15
4
3~5 SEC DELAY R32 T3 105/63 GND
1N4148 C17 TP3
TRANSFORMER
820K 50KHz TP8
ZD2 D6
104
+12V
VR4 R22
R34 D10 C23 3266 5k R19 1%2W 1
Q6 105 1N4747 104 R20
PN2907 14 11 12
DIODE C19 1%4.7K 1%1K
10K
C40 1N4148
CAPACITOR NON-POL R23 U3B - 6
13 U4 TL494 1 7 D7 D8
R30 + 5
4 1N4148
4
LM358 1K 1K LM358
2 - 10 9 7 5 6 2 3 15 16 C18 1N4148
1 MN104
3 + C22 R25 R21
U3A R27 R26 R24
C27 VR1 VR2 CW1-M12S 102 1M 10K 10K 1M
3266 2K 3266 2K L TP9 R28 10K
8
1%390 1%2.7K
R54
10K
TP10 R52 10K - 6 CN7 TO HV TANK
R38 Q14 7 EP FB
C1815 U2B + 5 1 IP FB
1K 2 GND
HTR-ON 25
OCP 28
X-RAY ON SIG 32
21
26
22
23
R55 +12V -12V
3 GND
FIL-I SENSING
2.2K LM358
4 INT LOCK
FIL-GATE 5
16 +12V EP FB 34
53258-09 18 GND IP FB 36 53258-05
X-RAY CONT
9 17 -12V GND 35
PRE HEAT
8
EP CONT
7 INT LOCK 30
TO SYSTEM IP CONT
6 +12V CN6
FOCUS(LOW=L)
ALM RST 5 R43 TO DRV BOARD
ALARM 4
19 R42 1 KV DR1
3 2 KV DR2
RTR ON
2 PWM OUT 1k Q9 R44
3 KV DR3
GND
1 7 EP REF 100
4 KV DR4
9 RTN 1k PN2907
5 OV CYRRENT
IP REF
DRV-CONTROL
CN2 8 R45 100
6 CHARGE MON
53258-06 7 GND
EP-MONITOR SUB1 R47 TP14 8 GND
IP-MONITOR 6
1 + X RAY CONTROL 20 R46 9 GND
5 10
GND
4
+12V 2 - PWM OUT 1k Q10
11
TO SYSTEM FIN ON
GND 3
2
R39
3 + PRE HEAT 1k PN2907
R48 100
12
13
10 ARLARM OUT
11 EP/IP OVER
12 EP MONITOR
13 IP MONITOR
RTR OK +12V 4.7K 4 - R49 100
1 RTR OK 14
TP11 CH2-MD12 GND 24
CN3 GND U5 5 + ALM RESET
TP12 6 - TP15
HIF3BA-14PA-2.54DSA
-12V RY 1
RTN
+12V -12V TP13 PC123F D14
53258-04 1N4937 R50
4 GND 1%2W10(ABW2N10J)
+12V RY 2
3
TO SYSTEM -12V
14
27
29
31
33
Q11 CW1-M12S
GND 2 R40 BU806
GND 1
EC470/35V
10K R41
CN4 10K Q12 R51
IRF530
1%2W90(ABW2N90J)
+ + EC470/35V
C29
C28
-12V 3
GND 2
+12V 1
35312-03P
CN5
Title
TO DRV BOARD HTC-120 HT CONTROL
101
Service Manual
6) HT Tank
102
Service Manual
7) IGBT Drive board
103
Service Manual
8) SMPS(Power supply board)
104
Service Manual
- PARTⅡ. DRAWING(PCB) -
105
Service Manual
1) OP board(Main)
106
Service Manual
107
Service Manual
108
Service Manual
109
Service Manual
2) Main board
110
Service Manual
111
Service Manual
112
Service Manual
3) Fuse board
113
Service Manual
114
Service Manual
115
Service Manual
116
Service Manual
4) OP board(Sub)
117
Service Manual
118
Service Manual
119
Service Manual
120
Service Manual
5) HT control
121
Service Manual
122
Service Manual
123
Service Manual
6) IGBT drive
124
Service Manual
125
Service Manual
126
Service Manual
7) Power supply board(smps)
127
Service Manual
128
Service Manual
3. WORKSTATION
Hardware Overview
Location
Workstation
Standard Option
CART CART
129
Service Manual
Parts location
Remove 4 screws
1 2
4
7
2
The Processing Unit is Composed of:
Number Description
1 Power Supply connector: 220VAC
2 Monitor signal : HDMI + DVI
3 ABS signal to Interface board page:1-133, CN5
4 Camera Link connector page:1-133, CN2
5 RS-232C COM1: Main board connector page:1-133, CN4
6 RS-232C COM2: Camera connector page:1-133, CN3
7 RS-232C COM3: D.A.P Module connector (option)
130
Service Manual
Connecting the Cables
131
Service Manual
132
Service Manual
133
Service Manual
Patch & INSTALL GUIDE
How to Recovery CD
[Selector CDROM]
134
Service Manual
135
Service Manual
First, confirm EWF current states.
1. File in C driver and the patch in order description below first a current state, confirms.
1.1 My Document Batch file.
137
Service Manual
2) HOW TO EWF DISABLE C: for patch
My Documents folder
EWF-Disable Batch [ ] double click.
138
Service Manual
EWF ENABLE functions are NOT executed from live
Condition.
Only at the time of PC Booting, possible !!!
139
Service Manual
1.2 Also with different method to execute a direct command from
COMMAND windows, confirmation
From COMMAND windows which appear EWFMGR C: inputs and confirms a current status.
Type RAM Protected Type.
State ENABLE Current protecting.
Boot Command NO_CMD (current no command)
140
Service Manual
Changed Boot Command is DISABLE.
(Booting at the time: Disalbenon protection)
PC restart.
START RESTART.
From COMMAND windows which appear EWFMGR C: inputs and confirms a current status.
CHECK DISABLE
141
Service Manual
142
Service Manual
In future
Embedd xp EWF function HIBERNATE Becoming connection.
The file patch first after, COMMAND window, inputs EWFMGR C: -COMMIT
143
Service Manual
PC restart.
START RESTART.
144
Service Manual
HOW TO HIBERNATION Function OFF and ON
XP Embedded one technical HIBERNATION functions is
FAST START realized.
145
Service Manual
- HIBERNATION Function OFF and ON
Factory setting
1 pcie_frm11.sys patch
C:\Windows\System32\drivers
Hibernate-on.bat
2 EWF ENABLE
3 P.C Re-start
After Booting
4 CXView & StoSCU
(setting first screen)
execution
7 MY DOCUMENTS folder
CLOSE
실행
8 About 5sec. Wait
PC auto re-start.
3. Makes CXView widow
실행 With make top-level.
9 COMPLETE FINISH
실행
File modify patch C: HDD drive.
a Hibernate-off.bat MY DOCUMENTS
b File patch
Network Connection [(TCP/IP) Setting]
c Hibernate-on.bat
d EWF ENABLE
e P.C Re-start
146
Service Manual
■Network Connection
1. Internet Protocol(TCP/IP) Properties Setting.
147
Service Manual
2. DICOM & Machine setting
( )
configuration click.
PACS
148
Service Manual
A. Modality setting
Via Change Operator, input the ID and Password enable Machine TAP.
149
Service Manual
150
Service Manual
■CXVIEW IMAGE SETTING
caution :
1) Sharpness adjust
151
Service Manual
152
Service Manual
ABS X-ray setting:
Mouse right button click[ Feature Setting] selector.
153
Service Manual
2) LUT setting [ ]
- Selector LUT .
EDIT
154
Service Manual
LUT graph application effects.
3) LEVEL setting
LEVEL Selective
155
Service Manual
4) Profile KEY setting
KEYBOARD KEY(F1~F10) SET
Mouse right button click [Profile] selector.
156
Service Manual
5) Profile setting
157
Service Manual
It is not possible to restore all images from new and old PC.
Only available to use image and patient DB from one of PC
□ Restoration of setting
158
Service Manual
159
Service Manual