Troubleshooting
EVA-HF325/525 generator
EVA-HF325/525
Troubleshooting
ADVISORY SYMBOLS
1
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
CONTENTS
1. I N T R O D U C T I O N - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
2. E R R O R C O D E S - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
3. F L O W C H A R T - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6
4. E R R O R C H E C K - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9
5. D ATA TA B L E - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11
6. W AV E F O R M D ATA - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 3
7. J O B C O D E - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 6
8. X - R AY C H E C K - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4 0
1 0 . T E C H N I C A L I N F O R M AT I O N - - - - - - - - - - - - - - - - - - - - - - - - - - 4 6
11 . J O B C O D E I N F O R M AT I O N - - - - - - - - - - - - - - - - - - - - - - - - - - 5 6
2
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
1. INTRODUCTION
3
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
2. ERROR CODES
System error codes are displayed on the kVp display LCD(condition display LCD)
Each code is preceded by the letter “ E”(i.e. E16). Error codes are described in below
tables.
4
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
5
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
3. FLOW CHART
Power ON
Text 1
LED 9 ON Check 1
LED10,11,12 ON
Text 2
LED 7,8 ON
Check 2
Text 3
LED 1 OR 2 ON
Check 3
LED 3 OR 4 OR 5 ON
Text 4
Contactor 1
Check 4
Contactor 2
Text 5
LED 4 ON
Initialize OK
6
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
TEXT 1
Led 9 ON is meaning of system power input OK.
Led 10,11,12 ON is meaning of power supply +5V,+12V,-12V OK.
That Leds located on Relay power board.
TEXT 2
Led 7 ON is meaning of charge of capacitor ON.
Led 8 ON is meaning of
Filament power input(AC60) ON.
Bucky power input(AC110) ON.
Lock power input(DC30) ON.
Collimeter power input(AC20) ON.
And relay power(DC24V) in relay power board
That Leds located on Relay power board.
TEXT 3
Led 1 ON is meaning of selected large focus filament.
Led 2 ON is meaning of selected small focus filament.
Leds ON will select Led 1 or Led 2.
Led 3 ON is meaning of selected bucky 1.
Led 4 ON is meaning of selected bucky 2.
Led 5 ON is meaning of selected bucky 3.
Led 3,4,5 all OFF is meaning of selected NO bucky and That leds ON will selected
Led 3 or Led 4 or Led 5.
That Leds located on Relay Power Board.
TEXT 4
Contactor 1, 2 is meaning of Main contactor and charge contactor ON.
TEXT 5
That meaning is System initializing OK.
Led 4 located on HT contol cpu board.
7
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
4. ERROR CHECK
CHECK 1
- Line input power check(see JOB CODE 001)
- Communication cable check(see JOB CODE 002)
- Membrane key check(see JOB CODE 003)
- Bridge diode in Relay power board(see JOB CODE 004)
- AC12V input check(see JOB CODE 005)
CHECK 2
- Power(+5V,+12V,-12V) in Relay power board(see JOB CODE 006)
- Power ON signal check(see JOB CODE 007)
CHECK 3
- Check ERROR CODE TABLE
CHECK 4
- Check charge/discharge monitor board(see JOB CODE 008)
- Check charge contactor(see JOB CODE 009)
- Check charge signal in HT control cpu board(see JOB CODE 010)
8
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
5. DATA TABLE
The Calibration data and some technique data stored in the U7-UV_PROM
And APR data stored in the U23-EEPROM of the Console display Board.
5.1 CALIBRATION DATA TABLE IN UV_PROM
9
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
10
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
6. WAVEFORM DATA
11
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
12
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
13
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
13. TL594 output(at low kV data) 14. TL594 output(at high kV data)
14
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
15
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
7. JOB CODE
Check 1. input power of line power cable L1 and L2 voltage is 215 ~230VAC
Check 2. connection of AC connection wire are tight to each terminals
16
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
J1 J2 J2 J1
Pic1.view in mother board Pic2. wiew in console
J1 J2
+5V(H→C) +KV
GND(H→C) -KV
TXD DC24V-2(H→C)
HX-RAY TUBE2X-RAY
HTRANSFER POWER ON
CX-RAY
CTRANSFER
CINHIBIT C+5V
+12V(C→H) C+12V
+12V(H→C) C GND
17
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
18
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
SW49
SW50
misprint silkscreen
membrane key
19
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
R40
1.2K D16
RY9
1N4001 SW
LD9
J1-21,43(AC220V)
LED J1-19,41(AC220V)
J1-20,42(AC220V)
- + R34 R35 Q5
1K 1K 2N2222A
D9
DF06 + C12 R37 Q6
2N2222A R36
1K
220uF/35V 4.7K
R38
4.7K
to console control
J1-1(S/W ON))
J1-2,24(S/W COM)
J1-23(S/W OFF)
J1-22(AC12V)
J1-44(AC12V)
LED9 ON
SWITCH ON/COM/OFF LINE IS CONNECTED WITH COMMUNICATION AND CONSOLE.
20
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
AC20V
J5(11P)
J5
DC24V-2
GND
DC24V-1
AC220V(N)
J9(6P) AC220V(N)
IPM AC220V
J9 IPM AC220V
AC12V
AC220V AC12V (To mother board)
AC220V AC220V
AC12V (From sub-transformer) AC220V
AC12V
AC20V
AC20V
21
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
LED10 ON : +5V
LED11 ON : +12V
LED12 ON : -12V
R10
R33 R25
22
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
Relay 3 and D5
20A
RY3
D5
+24V POWER ON
1N4007
23
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
J2-13,35(DC24V-1)
J1-4(DC24V-2)
R39
2.2K
J1-12,34(DC24V-2 OUTPUT)
to 1pin of J11and 2pin of J2 in
D17 mother board and
LD8 RY8
LED 1N4001
J2-26(POWER ON)
To Relay Power
4. Check input power DC24V-2, DC24V-1(All power is from Main power board)
If stopped LCD display “ initialize ' “ always and No ERROR CODE display in
LCD dispay , Check this section
24
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
P2(charge sense)
P3(discharge signal)
R3- R9(33K) this resistor endure at 230VAC
Must check input power voltage(210 – 230VAC)
D1 Q1
IN4764 IRFP450
LED1
D3
D2 1N4733
IN4764 U1 P3
1 6 R4
1
3.3M 2 To 3,4pin of J10 in
DISCHARGE Mother Board
2 5
P2
1 To 5,6pin of J10 in
3 4 2 Mother Board
3
CHARGE SEN
TLP531
AC 380/3Phase -> 9K
25
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
main
charge
precharge resistor
If power not charge in storge capacitor will displayed “E28” in console display.
26
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
If need change Calibration data with new data saved in Serial-ROM. You will read Serial-ROM
with Rom_writer and re_save data in same address. if you save abnormal data, it will be
27
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
U7(M27C256B)
U25(25256)
28
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
KVp reference
data at 200mA
mA reference
data at 25mA
mA reference
data at 50mA
29
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
30
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
J10
SSR
CT
J7(To Interface Terminal)
SP(spark killer)
J10
1
2
CT1
0VAC
ROTOR DC24V-1 COM/MAIN/PHASE
SP1 SSR1
220VAC
D4
SSR 1N4007
J7-1
J7-2
J7-3
J7-4
J7-5
C4
24uF
ROTOR ON
Pic. Interface Terminal
To ROTOR of TUBE
31
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
TEMP
G(PHASE)
B(MAIN)
W(COM)
32
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
U19 R39-46 10
1 18
I1 Q1 J2-27(MAIN MAG)
2 17
I2 Q2 J2-5(POWER ON)
3 16
I3 Q3 J2-28(PRECHARGE)
4 15
I4 Q4 J2-6(CHARGE MAG)
5 14
I5 Q5 J2-7(TUBE SEL)
6 13
I6 Q6 J2-29(220/60)
7 12
I7 Q7 J2-30(HIGH SPEED RT ON)
8 11
I8 Q8 J2-8(ROTOR ON)
9 10
GND COM 1K
ULN2803
If did not output sensing signal of ROTOR sense, it will display “E26/27” in Console
display LCD.
33
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
+12V C49
TP11
104
0.6V
8
VR8 U12A D24
3
+
1 6 11
10k 2 U6
- J1_12(FIL LOW)
ULN2004
TL082A 1N4148
R63
4.7K
4
10k
C48
104 +12V
R64
4.7K
8
U12B D25
5
+
VR9 7 5 12
6 U6
- TL082A J1_34(FIL HIGH)
ULN2004
10k 1N4148
4
TP10
C50
3.3V-12V
104
If you want LOW limit change, use with VR8. and HIGH limit is VR9.
If filament sensing value is lower than TP11, will display “E29” in Console Display
LCD.
And higher than TP10, will display “E30” in Console Display LCD.
34
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
Temp(from tube)
If TUBE temperature signal is open, will display “E32” in Console display LCD.
J7 J7(10P)
BUCKY1
BUCKY1
BUCKY2
BUCKY2
BUCKY3
BUCKY3
TUBE1 TEMP
TUBE1 TEMP
TUBE2 TEMP
TUBE2 TEMP
35
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
From HV tank
36
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
HCPL4503
VPC(GND1)
J1-4
PC1
8 2 VP1(+15VOUTPUT1)
J1-1
C5
6 0.1
3 FPO(-FAULT1)
J1-5
5
HCPL4503
Table. Check Point(All power was setted originally in factory. Don’t need contrast)
Checking point Value Accuracy limit
J1 – 1,4 VDC 14.9 ~ 15.1VDC
J1 – 2,4 VDC 4.95 ~ 5.05VDC
J2 – 1,4 VDC 14.9 ~ 15.1VDC
J2 – 2,4 VDC 4.95 ~ 5.05VDC
Pic.IPM and IPM drive board
IPM
IPM drive board
J2
J1
J3
+5V
GND
J1(J2) KDR1
KDR2
FAULT FAULT
GND AC220V
-KV DR AC220V
+5V
+15V
37
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
+mA
+kV
GND
When measure “+kV” wih OSC, real kV is 10kV per 1V.
Also, real mA is 20mA per 1V.
Case 1
12 D0 VDD 17
1 16 5 12 11 D1 RS
RT
I1 O1 CT VCC 11 D2
2 15 C1 10
D3
3 I2 O2 14 C2
8 9
I3 O3 16 9 8 D4 3
4 13 15 +IN1 E1 10 D5 VREF
I4 O4 -IN1 E2 7 14
5 12 3 14 6 D6 CS 16
6 I5 O5 11 1 FB REF 13 D7 LDAC
+IN2 CTL 4
7 I6 O6 10 2 4 AGND 5
-IN2 DTC DGND
8 I7 O7 9
7
GND 1
GND COM 15 VSS 2
WR VOUT
U3
ULN2004 TL594 AD7224
38
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
2. Aanother factor.
1. If IPM driver board broken, always display “E00”(power on state ,not exposure state)
change IPM driver board
check IPM drive input power(220VAC) and signal connection.
2. If IPM was damaged, sound “ ting—“and display “E00” at high kVp exposing
change IPM
39
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
8. X-RAY CHECK
1. FILAMENT DRIVE BOARD CHECK
Tools & Equipment required : DVM,OSC
Pic. Filament drive board
① filament enable signal (high low)
② ③
DC0V - 100V
+12V
D12 1uF
1
VIN
RT CA
COMP
13 U3 ZD4 Q12
5K 4.7K CB 4.7 1/2W
3 1 8 IRFP250
C18 OSCOUT NC NC FR102 D13 D14 D15
7 11 2 7 1N4745 C13 + C30
CT EA 3 INA OTA 6 T4
472 GND VDD R42
14 4 5 224 470uF/200V
10 EB INB OTB ZD9 R38
SHTDWN FR102 FR102 FR102 1K
4 D8 R32 47K/3W
+SENSE TC4427
5 R31 Q9
VREF
-SENSE 1K 1N4745
8 R19 R18 FR102 1K 2N2907 C27
V+
GND
1K 1K
U2 Q10 473 C16 C19
C26 ZD10
16
2
LM3524
R33 102 104 104
Q4 1N4745
R23 VR3 R22 IRF630 1 1/2W 2N2907
ZD8
1K 5K 1K
1N4733 C28
R9
J2-30(MSY) D17 1uF Q11
1K R34 2P4M R35
+12V
4.7 1/2W 560
C32 R36 R39
R51 FR102
R49 R46 100 0.5/5W
FSENSE 104
D19
10K 20K 20K
1N4148
R47 R45
R50 C34
+
+12V
-12V
0.1 1N4148 J1-4,5,26,27(FIL OUTA)
C37
8
+ U5B U5A
6 2
- - J1-6,7,28,29(FIL OUTB)
100uF/35V 7 1
5 3
+ +
TL082 TL082
4
J2-19,20,41,42(-12V) C36
-12V
0.1
① ④
40
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
Q1 R1
J2-10,11,32,33(AC60V) BUX48A 0.5/3W
BD1
NF1 - +
Q2 R2
KBU10M
BUX48A 0.5/3W
J2-13,14,35,36(AC60V)
NOISE FILTER Q3 R3
BUX48A 0.5/3W
J2-15,37(EARTH) + C1 + C2
R11 Q4 R4 R10
47K/3W DC0 - 100V
470uF/200V 470uF/200V BUX48A 0.5/3W 0.2/5W
+12V
C7
Q5 R5 R12 R14
TIP42 1K/3W 104 12K 12K
U6B
8
D1 LM358 6 R16
R6 -
R24 7
47K
5.2K 5
2.2K/1/2W +
R13 R15
D3 1N4148 R17
Q6 390 820
4
2N3440 47K
+12V C4
1N4148
C5
J2-21,22,43,44(GND)
0.1 0.1
R7 R8 U6A
8
+ C17 2.4K 2.7K
J2-17,18,39,40(+12V) D2 LM358 R20
2
-
100uF/35V 1
47K
3
+
1N4148 R21
4
47K
C3
R54 0.1
10K
J2-31(FOUT) C8
+12V
8
R58
-12V
8
10K R56 6
104 -
7
100K
5 U4B
+
R26 C6
TL082 R25
4.7K 0.1
10K
4
C10
+12V
8
104
8
6
-
7 2
J1-13,24,35(FSEN) -
U1B 5 1
+
U1A 3
TL082 + FSENSE
C11 TL082
4
-12V
104
⑤ ⑥ VR1
41
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
42
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
R38 R37
R9
4.7K 4.7K U4B
10K
U6 15 14
7 10 16 D S
IN
U1A J2_26(R/F)
TL082A R8 ADG201A
8
KV
ZD3,ZD4을 1N4744을을 1N4746을 을을을 2001을 8을 29을 2 ULN2004 U4A
C8 - 1M
1 2 3
R19 J1_1(KSEN) D S
3 1
100P + IN
J1_27(-MASEN) R15 1M
J1_5(+MASEN) C4 ADG201A
4
154K
C21 R16 C9 C10 103
-KV
+12V -MA
104 10K 273 100P
8
R10 R12
8
6 U2B U1B
- R17 R18 R1 R3
7 6
10K 1M -
5 7
+ 34K 100K 1K 100K
C5 ZD3 D5 D6 5
TL082A +
C1 ZD1 D1 D2
TL082A R6
102 1N4744 1N4148 1N4148
4
C22
104 1N4752 1N4148 1N4148 10K MA
4
J1_26(-KVSEN) -12V
T2
J1_4(+KVSEN) +12V -12V 104
100uH T1
U4A +12V -12V
C7 VR2 100uH
C6 ZD4 D7 D8 3 2
S D 1 C2 ZD2 D3 D4 C3
100P 20K IN
102 1N4744 1N4148 1N4148 R7
103
R11 R13 R14 ADG201A 104 1N4752 1N4148 1N4148 10K
U4B R2 R4 R5 J1_23(MSEN)
10K 1M 154K VR1
14 15
S D 16 1K 100K 10K
+KV 10K IN
+MA
ADG201A R28
+12V
U5B
1K
14 15 TP2
S D 16 C23
IN C15
6
104 U6
C18 5 12 TP3 TP4
ADG201A
RT
CT VCC 11 1 16 R33-R36
C1 I1 O1 J1_7(KDR1)
8
U2A 683 8 2 15
J2_32(KOUT) 103 C2 I2 O2 J1_29(KDR2)
3 16 9 3 14
+ R22 +IN1 E1 I3 O3 J1_8(KDR3)
1 15 10 4 13
-IN1 E2 I4 O4 J1_30(KDR4)
2 3 14 5 12
- 24.9K FB REF I5 O5 10
D9 D10 1 13 6 11
TL082A R23 C12 2 +IN2 CTL 4 7 I6 O6 10
7 -IN2 DTC 8 I7 O7 9
1N4148 1N4148 52K
4
ADG201A 104
U5
U5D ADG201A
7 6 4 C27
8 D S
IN 104
ADG201A
① ② ③ ⑤ ⑥ ⑦
43
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
① kV output data voltage from 2 of U27 in Ht control cpu board 1(when exposured)
kV low (40kV see waveform 6)
kV high(70kV see waveform 7)
② exposure/kV enable signal
high low (see waveform 8)
③ TP 2 is control waveform(see waveform 9)
Checking point Value Accuracy limit
TP2 kHz 39 ~ 41kHz
Adjust point : VR1
④ test point : -kV, +kV feedback signal waveform
+kV feedback (see waveform 10)
-kV feedback(see waveform 11)
1V ≒ 10kV
⑤TP1(dead time control)
Adjust point : VR3 (see waveform 12, when exposure)
⑥TP3,TP4 : TL594 output point
kV data low(see waveform 13)
kV data high(see waveform 14)
E1 is 90 degree phase ahead of E2
⑦KDR1,KDR4 (see waveform 15)
KDR1,4 is 90 degree phase ahead of KDR2,3
KDR1 ~ KDR4 signal input to IPM DRIVE BOARD
⑧ test point : -mA, +mA feedback signal waveform
+mA feedback (see waveform 16)
-mA feedback(see waveform 17)
1V ≒ 20mA
44
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
J11(5P) J10(6P)
J14(10P) J1(15P) J2(15P) J15~J18(NOT USED) J19(6P) J13(7P) J12(7P)
J3 J7 J11 J14
TUBE2 AC220V BUCKY1 DC24V-2 -KV
BUCKY1 GND +KV
TUBE2 AC220V BUCKY2 PRECHARGE +MA
BUCKY2 CHARGE -MA
BUCKY3 MAIN GND
J4 BUCKY3 SMALL
TUBE2 READY TUBE1 TEMP COM
TUBE2 X-RAY TUBE1 TEMP J12 LARGE
FLUORO MODE TUBE2 TEMP GND TUBE2 AC220V
GND TUBE2 TEMP +5V TUBE2 AC220V
KDR4
J5 J8 KDR3 J19
AC220V AC220V(N) IPM FAULT AC220V
AC220V AC220V(N) AC220V AC220V
AC12V AC60V AC220V +12V GND
AC12V AC60V +12V
IPM AC220V +5V GND
IPM AC220V J9 J13 +5V
AC220V(N) AC220V GND
AC220V(N) AC220V +5V J1,J2
DC24V-1 AC15V KDR2
GND AC15V KDR1 : REFER JOB CODE002
DC 24V-2 AC20V IPM FAULT
AC20V RTN AC220V
AC20V AC220V
J6
ROTOR ON
POWER ON J10
220/40 CHANGE DISCHARGE4
TUBE SELECT DISCHARGE3
ROTOR SENSOR DISCHARGE2
GND DISCHARGE1
AC110V CHARGE SENSE
AC110V GND
45
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
CONSOLE
DISPLAY AND
MEMBRANE
CONTROLLER
CONSOLE CPU
COMMUNICATION CABLE HT CONTROL
BOARD
CPU BOARD KOUT
Feedback
HT CONTROL IPM
IPM
CONTROL BOARD driver
RELAY POWER
BOARD
HV transformer
MAIN POWER
+
Filament power BOARD
- terminal
T1 and T2
ROTOR power
46
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
kV mA Time Bucky
Console display CPU give kV data,mA data,time data,bucky select data to HT control CPU.
And HT control CPU re-give same data to Console display CPU.
So, Console display CPU check data error or good. If error detected, display EROOR code.
Ex) KV data error(E16,E17), mA data error(E18,E19), time data(E20) I
TP1(Txd), TP2(Rxd)
47
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
kV mA
mA , time, bucky select signal not transfer. Only kV data data transfer
8
cf . kV data 8bit = 2 P
P
= 256 information
48
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
kV data(138+1 = 139)
KV-SET data “139” in Console display cpu board give to HT control cpu board
And, HT control cpu give data to AD7224
When communicate HT control cpu board with console display cpu
board, sometimes error detected by electrical noise.
1. Check ground state.
2. Check shield state of communication cable.
3. Check noise of Power line.
]
49
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
50
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
51
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
AD7224
If you want to see waveform TP5,TP7(when push x-ray s/w) see Chap.6 6 and7 waveform.
BECAREFUL WHEN YOU CHECKING “FOUT SIGNAL”. IF ELECTRICAL
SHORT HAPPEN, U25(AD7224) IN HT CONTROL CPU BOARD WILL BE
BROCKEN.
1 2 3 4
1. KOUT signal 2. Feedback signal
52
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
TL594(494) receive two signal (KOUT(-ini) and feedback signal(+ini))and give output
signal(E1,E2) to IPM drive.
If KOUT(analog data) is much, width of TL594 output signal would be widen.
53
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
54
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
55
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting
56
COMED Medical Systems Co.,Ltd.