Troubleshooting
EVA-HF325/525 generator
EVA-HF325/525
Troubleshooting

ADVISORY SYMBOLS

The following advisory symbols will be used throughout this manual.


Their application and meaning are described below.

DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT IF


NOT HEED OR AVOIDED WILL CAUSE SERIOUS PERSONAL
INJURY OR DEATH.

ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT


HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL
INJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT OR
DATA.

ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT


HEEDED OR AVOIDED COULD CAUSE PERSONAL INJURY
OR DAMAGE TO EQUIPMENT OR DATA.

1
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

CONTENTS

1. I N T R O D U C T I O N - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3

2. E R R O R C O D E S - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4

3. F L O W C H A R T - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6

4. E R R O R C H E C K - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9

5. D ATA TA B L E - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11

6. W AV E F O R M D ATA - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 3

7. J O B C O D E - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 6

8. X - R AY C H E C K - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4 0

9. MAIN CARD BOARD(MOTHER BOARD)DRAWING---------- 45

1 0 . T E C H N I C A L I N F O R M AT I O N - - - - - - - - - - - - - - - - - - - - - - - - - - 4 6

11 . J O B C O D E I N F O R M AT I O N - - - - - - - - - - - - - - - - - - - - - - - - - - 5 6

2
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

1. INTRODUCTION

The generator contains many self-diagnostic routines which greatly facilitates


troubleshooting. The self-diagnostic functions require that all microprocessors function
properly. Each microprocessor contains LEDs, LCD that indicates their proper operation.

THE MAIN STORAGE CAPACITORS RETAIN A LARGE PORTION OF


THEIR FOR APPROXIMATELY 5MIN. AFTER THE UNIT IS TURNED
OFF.

ALWAYS HAVE “IPM DRIVER PCB” CONNECTED WHEN MAIN


POWER IS ACTIVATED IN THE GENERATOR. WHITOUT “IPM
DRIVER PCB” CONNECTED, PERMANENT DAMAGE WILL OCCUR
TO IPM’S.

DO NOT USE THE GENERATOR POWER SUPPLIES TO POWER


EQUIPMENT EXTERNAL TO THE GENERATOR. ALL GENERATOR
INPUTS AND OUTPUTS MUST BE TTL LOGIC LEVELS
(TYPICALLY ACTIVE LOW) AND ISOLATED FROM OTHER
SUBSYSTEMS OPTICALLY OR WITH DRY CONTACT RELAYS.

IF CHANGE EEPROM ON CONSOLE DISPLAY BOARD, WILL


LOSE CALIBRATION DATA, APR DATA, AND SOME DATA. SO
MUST CHECK EEPROM DATA.
AND IF CHANGE UV-PROM ON HT CONTROL CPU BOARD AND
CONSOLE DISPLAY BOARD, WILL CHECK THE VIRSION OF
PROGRAM.

3
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

2. ERROR CODES

System error codes are displayed on the kVp display LCD(condition display LCD)
Each code is preceded by the letter “ E”(i.e. E16). Error codes are described in below
tables.

ERROR CODE DESCRIPTION

A system failure has been detected. The system must be turned


Display OFF and then ON to clear the error.
E xx Besides display “E xx”, another error code could be shown on the
Console display LCD.
KV DATA is more than “FD”(see JOB CODE 020)
E16
Check KV DATA located APR data in EEPROM
KV DATA is less than “20”(see JOB CODE 020)
E17
Check KV DATA located APR data in EEPROM
mA DATA is more than “FD”(see JOB CODE 020)
E18
Check mA DATA located APR data in EEPROM
mA DATA is less than “20”(see JOB CODE 020)
E19
Check mA DATA located APR data in EEPROM
TIME DATA is abnormal(see JOB CODE 021)
E20
Check TIME DATA located APR data in EEPROM

COMP DATA is abnormal(see JOB CODE 020)


E21/E22
Check COMP DATA located Calibration data in UV-PROM
ROTOR is abnormal(see JOB CODE 022)
E26/E27 Check ROTOR sensing circuit section or
ROTOR DATA located in UV-PROM address
CHARGE is abnormal(see JOB CODE 008 ~010)
E28 Charge capacitor(4700uF) is not charge
Check charge sensing circuit section and control cpu board
Filament sensing “LOW” (see JOB CODE 023)
E29 Low current input to filament
Check filament sensing circuit section and connection of HV cable

4
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Filament sensing “HIGH”(see JOB CODE 023)


E30 High current input to filament
Check filament drive board section
TEMP is abnormal(see JOB CODE 024)
E32 Tube temperature is overheated
Check Tube temperature section and signal connection.
IPM fault(see JOB CODE 025)
E00
Check IPM fault section

5
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

3. FLOW CHART

Power ON

Text 1

LED 9 ON Check 1
LED10,11,12 ON

Text 2

LED 7,8 ON
Check 2

Text 3

LED 1 OR 2 ON
Check 3
LED 3 OR 4 OR 5 ON

Text 4

Contactor 1
Check 4
Contactor 2

Text 5

LED 4 ON

Initialize OK

6
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

TEXT 1
Led 9 ON is meaning of system power input OK.
Led 10,11,12 ON is meaning of power supply +5V,+12V,-12V OK.
That Leds located on Relay power board.
TEXT 2
Led 7 ON is meaning of charge of capacitor ON.
Led 8 ON is meaning of
Filament power input(AC60) ON.
Bucky power input(AC110) ON.
Lock power input(DC30) ON.
Collimeter power input(AC20) ON.
And relay power(DC24V) in relay power board
That Leds located on Relay power board.
TEXT 3
Led 1 ON is meaning of selected large focus filament.
Led 2 ON is meaning of selected small focus filament.
Leds ON will select Led 1 or Led 2.
Led 3 ON is meaning of selected bucky 1.
Led 4 ON is meaning of selected bucky 2.
Led 5 ON is meaning of selected bucky 3.
Led 3,4,5 all OFF is meaning of selected NO bucky and That leds ON will selected
Led 3 or Led 4 or Led 5.
That Leds located on Relay Power Board.
TEXT 4
Contactor 1, 2 is meaning of Main contactor and charge contactor ON.
TEXT 5
That meaning is System initializing OK.
Led 4 located on HT contol cpu board.

7
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

4. ERROR CHECK

CHECK 1
- Line input power check(see JOB CODE 001)
- Communication cable check(see JOB CODE 002)
- Membrane key check(see JOB CODE 003)
- Bridge diode in Relay power board(see JOB CODE 004)
- AC12V input check(see JOB CODE 005)

CHECK 2
- Power(+5V,+12V,-12V) in Relay power board(see JOB CODE 006)
- Power ON signal check(see JOB CODE 007)

CHECK 3
- Check ERROR CODE TABLE

CHECK 4
- Check charge/discharge monitor board(see JOB CODE 008)
- Check charge contactor(see JOB CODE 009)
- Check charge signal in HT control cpu board(see JOB CODE 010)

Additional CHECK point


- If stopped LCD display “ initialize I “ always. And No ERROR CODE display in
LCD dispay(see JOB CODE 007)

8
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

5. DATA TABLE
The Calibration data and some technique data stored in the U7-UV_PROM
And APR data stored in the U23-EEPROM of the Console display Board.
5.1 CALIBRATION DATA TABLE IN UV_PROM

9
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

5.2 APR DATA TABLE IN EEPROM

10
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

6. WAVEFORM DATA

1. control waveform 2. output in gate of Q3,Q4

3. preheating 4. ready preheating(at high mA data)

11
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

5. ready preheating(at low mA data) 6. kV data output(at low kV data)

7. kV data output(at high kV data) 8. exposure enable

12
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

9. control waveform 10.+kV feedbak

11. –kV feedback 12. deadtime control

13
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

13. TL594 output(at low kV data) 14. TL594 output(at high kV data)

15. KDR1~KDR4 output 16. +mA feedback

14
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

17. –mA feedback 18. rotor signal

15
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7. JOB CODE

7.1 JOB CODE 001


Tools & Equipment required : DVM
THIS GENERATOR IS PERMANENTLY CONNECTED TO THE POWER LINE.
INTERNAL PARTS OF THE GENERATOR(LINE FUSE, ETC.) ARE
PERMANENTLY POWERED ON THROUGH POWER LINE
ALTHOUGH THE CONSOLE IS OFF. BE SURE THAT THE SAFETY
SWITCH IS OFF BEFORE HANDLING ANY INTERNAL PART OF
THE EQUIPMENT.

THIS 325/525 GENERATOR LINE POWER IS SINGLE PHASE


220VAC. FOR SINGLE PHASE GENERATOR, CONNECT THE
POWER CABLE L1 AND L2(N : NEUTRAL) AND GND TO THE
TERMINAL BLOCK.

system power terminal

x-ray power terminal line power cable AC connection wire

Check 1. input power of line power cable L1 and L2 voltage is 215 ~230VAC
Check 2. connection of AC connection wire are tight to each terminals

16
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.2 JOB CODE 002


Tools & Equipment required : DVM
CHECK POWER ON/OFF/COM OF J2 IN COMMUNICATION CABLE SIGNAL

J1 J2 J2 J1
Pic1.view in mother board Pic2. wiew in console
J1 J2
+5V(H→C) +KV

GND(H→C) -KV

TXD DC24V-2(H→C)

RXD TUBE2 READY

HX-RAY TUBE2X-RAY

HREADY FLUORO MODE

HINHIBIT POWER OFF

HTRANSFER POWER ON

CREADY POWER COM

CX-RAY

CTRANSFER

CINHIBIT C+5V

+12V(C→H) C+12V

+12V(H→C) C GND

J1 AND J2 IS CONNECTED CONTROLLER AND CONSOLE WITH RS-232


COMMUNICATION CABLE SO, FIX TIGHTLY WITH SCREW WHEN CONNECT J1,J2. AND
THEN CHECK IT.

17
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

ONESIDE COMMUNICATION CABLE LOCATED MOTHER BOARD


INSIDE CONTROLLER INTERCONNECTION ARE FACTORY MADE.
ONLY OPPOSITE CONNECTION SHOULD BE CONNECT TO
CONSOLE.
AND COMMUNICATION CABLE LENGTH IS 10M. FOR CABLE
LENGTH REFER TO STANDARD RS-232 COMMUNICATION WIRE
LENGTH. IF ROOM REQUIRE MORE THAN LENGTH MUST NOT OVER THAN 15M. IF WIRE
LENGTH MORE THAN 15M WILL BE OCCUR COMMUNICATION FAILURE.

18
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.3 JOB CODE 003


Tools & Equipment required : DVM
Check SW49(power off), SW50(power on) connection.
PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7
PC0 SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
PC1
PC2
PC3
PC4
PC5 SW9 SW10 SW11 SW12 SW13 SW14
PC6
PC7
NC
SWOFF
SWON S17 SW18 SW19 SW20 SW21
SWCOM

SW25 SW26 SW27 SW28 SW29 SW30

SW33 SW34 SW35 SW36 SW37 SW38 SW39

SW41 SW42 SW43 SW44

SW49

SW50

power OFF power ON

misprint silkscreen
membrane key

19
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.4 JOB CODE 004


Tools & Equipment required : DVM
Check D9(bridge diode) of Relay power board

bridge diode COM/ON/OFF LED9


Check diode input AC12V and output DC20V

R40
1.2K D16
RY9
1N4001 SW

LD9
J1-21,43(AC220V)
LED J1-19,41(AC220V)

J1-20,42(AC220V)
- + R34 R35 Q5
1K 1K 2N2222A

D9
DF06 + C12 R37 Q6
2N2222A R36
1K
220uF/35V 4.7K
R38
4.7K

to console control

J1-1(S/W ON))

J1-2,24(S/W COM)

J1-23(S/W OFF)

J1-22(AC12V)
J1-44(AC12V)

LED9 ON
SWITCH ON/COM/OFF LINE IS CONNECTED WITH COMMUNICATION AND CONSOLE.

20
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.5 JOB CODE 005


Tools & Equipment required : DVM

Check AC12V in Pictures


Pic 1. SUB TRANSFORMER

AC12V(to main power board)

AC20V

SYSTEM POWER TERMINAL

Pic 2. MAIN POWER BOARD

J5(11P)

J5
DC24V-2
GND
DC24V-1
AC220V(N)
J9(6P) AC220V(N)
IPM AC220V
J9 IPM AC220V
AC12V
AC220V AC12V (To mother board)
AC220V AC220V
AC12V (From sub-transformer) AC220V
AC12V
AC20V
AC20V

mother board drawing refer

21
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.6 JOB CODE 006


Tools & Equipment required : DVM
Check TEST POINT +5V,+12V,-12V in Relay Power Board

Check +5V(4.95V ~5.10V)

Check +12V(11.95V ~12.10V)

Check –12V(11.95 ~12.10V)

LED10 ON : +5V

LED11 ON : +12V

LED12 ON : -12V

R10 : +5V ADJUST


R25 : +12V ADJUST
R33 : -12V ADJUST

R10

R33 R25

22
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.7 JOB CODE 007


Tools & Equipment required : DVM
Check POWER ON signal.
1. Check output signal 17pin of U19 in HT Control Cpu Board

17pin of U19(power on signal) is High→Low

2. Check input signal 2pin of J3 in Main Power Board


J3
ROTOR ON
POWER ON
220/40 CHANGE
TUBE SELECT
ROTOR SENSE
ROTOR SENSE
BUCKY AC110V
BUCKY AC110V

Relay 3 and D5

TO MAIN TRANSFORMER SP4


J2-4

FROM SUB TRANSFORMER F7


J2-6

20A
RY3

D5
+24V POWER ON
1N4007

Pic. Main power board drawing


One wire of sub transformer (Input power 220VAC) and main transformer (220VAC) is
interconnected by Relay 3. So, operating by Relay 3 will give Input power to system.

23
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

3. Check input signal to Relay8 in Relay Power Board

J2-13,35(DC24V-1)

J1-4(DC24V-2)

R39
2.2K
J1-12,34(DC24V-2 OUTPUT)
to 1pin of J11and 2pin of J2 in
D17 mother board and
LD8 RY8
LED 1N4001

J2-26(POWER ON)

To Relay Power

4. Check input power DC24V-2, DC24V-1(All power is from Main power board)

If stopped LCD display “ initialize ' “ always and No ERROR CODE display in
LCD dispay , Check this section

24
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.8 JOB CODE 008


Tools & Equipment required : DVM
Check Led1 light ON and Charge Capacitor storge voltage(about 310VDC)

P2(charge sense)
P3(discharge signal)
R3- R9(33K) this resistor endure at 230VAC
Must check input power voltage(210 – 230VAC)

Discharge signal refer to Relay Power Board drawing


J1 LED1
+ 1
- 2
NOTE1
R2 R4 R5 R6 R7 R8 R9
R1 27k/3w R3 15K/3W 15K/3W 15K/3W 15K/3W 15K/3W 15K/3W
68K/3W 100K/3W

D1 Q1
IN4764 IRFP450
LED1

D3
D2 1N4733
IN4764 U1 P3
1 6 R4
1
3.3M 2 To 3,4pin of J10 in
DISCHARGE Mother Board
2 5
P2

1 To 5,6pin of J10 in
3 4 2 Mother Board
3

CHARGE SEN
TLP531

NOTE1 : AC220/1PHASE --> 27K/3W


AC380/3PHASE --> 18K/3W

2005.10.07 Note2 : AC 220/1Phase -> 33K

AC 380/3Phase -> 9K

Pic. Charge/discharge Monitor Board


THIS GENERATOR IS PERMANENTLY CONNECTED TO THE POWER LINE.
INTERNAL PARTS OF THE GENERATOR(LINE FUSE, ETC.) ARE
PERMANENTLY POWERED ON THROUGH POWER LINE ALTHOUGH THE
CONSOLE IS OFF. BE SURE THAT THE SAFETY SWITCH IS OFF BEFORE

HANDLING ANY INTERNAL PART OF THE EQUIPMENT.

THE MAIN STORGE CAPACITORS RETAIN A LARGE PORTION OF THEIR


CHARGE FOR APPROXIMATELY 3 MIN. AFTER THE UNIT IS TURNED OFF.

25
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.9 JOB CODE 009


Tools & Equipment required : DVM
Check main and charge contactor action.

main

charge

precharge resistor

Sub Relay Board


J2 J3
DC24V-2 MAIN MAG
GND MAIN MAG
MAIN RELAY CHARGE
CHARGE RELAY CHARGE
PRECHARGE

7.10 JOB CODE 010


Tools & Equipment required : DVM
Check each signal actions in HT control cpu board.
U19
U19 R39-46 10
1 18
I1 Q1 J2-27(MAIN MAG)
2 17
I2 Q2 J2-5(POWER ON)
3 16
I3 Q3 J2-28(PRECHARGE)
4 15
I4 Q4 J2-6(CHARGE MAG)
5 14
I5 Q5 J2-7(TUBE SEL)
6 13
I6 Q6 J2-29(220/60)
7 12
I7 Q7 J2-30(HIGH SPEED RT ON)
8 11
I8 Q8 J2-8(ROTOR ON)
9 10
GND COM 1K
ULN2803

POWER OFF POWER ON MAIN MAG CHARGE MAG

18PIN low high low low

17PIN high low low low

16PIN low high low high

15PIN low high high low

If power not charge in storge capacitor will displayed “E28” in console display.

26
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.20 JOB CODE 020


Tools & Equipment required : ROM-WRITER, CALCULATOR
Check kV, mA, comp data in Calibration Mode or read and re_write Serial-ROM with ROM-
WRITER.
See Serial-ROM TABLE(chapter.6)
address description address description
0007600 ~ 0007610 KVp data at 25mA 0007820 ~ 0007830 mA data at 50mA
0007620 ~ 0007630 kVp data at 50mA 0007840 ~ 0007850 mAp data at 75mA
0007640 ~ 0007650 KVp data at 75mA 0007860 ~ 0007870 mA data at 100mA
0007660 ~ 0007670 KVp data at 100mA 0007880 ~ 0007890 mA data at 150mA
0007680 ~ 0007690 KVp data at 150mA 00078A0 ~ 00078B0 mA data at 200mA
00076A0 ~ 00076B0 KVp data at 200mA 00078C0~ 00078D0 mA data at 250mA
00076C0~ 00076D0 KVp data at 250mA 00078E0 ~ 00078F0 mA data at 300mA
00076E0 ~ 00076F0 KVp data at 300mA 0007900 ~ 0007910 mA data at 400mA
0007700 ~ 0007710 KVp data at 400mA 0007920 ~ 0007930 mA data at 500mA
0007720 ~ 0007730 kVp data at 500mA 0007A00 ~ 0007A00 Comp data
0007800 ~ 0007810 mA data at 25mA 0007A20 ~ 0007A20 Technical data
See EEPROM TABLE(chapter.6)
address description
0003000 ~ 00030D0 KVp, mA, time,bucky selecting data at very big human
0003100 ~ 00031D0 KVp, mA, time,bucky selecting data at big human
0003200 ~ 00032D0 KVp, mA, time,bucky selecting data at small human
0003300 ~ 00033D0 KVp, mA, time,bucky selecting data at very small human

Pic. Calibration Mode LCD display

KVp data, mA data, comp data

If need change Calibration data with new data saved in Serial-ROM. You will read Serial-ROM
with Rom_writer and re_save data in same address. if you save abnormal data, it will be

27
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

displayed ERROR message in Console Display LCD.


If APR data brocken by noise. You can re_save with pushing set switch(save switch) in APR
MODE.
Pic. Serial-ROM and EEPROM location(Console display board(operating_panel_console board)

U7(M27C256B)

U25(25256)

28
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.21 JOB CODE 021


Tools & Equipment required : ROM-WRITER, CALCULATOR
Check reference of calibration data in Serial-ROM with ROM-WRITER.

Reference of Calibration data

All calibration data of KV(MA)SET is sum of reference data and


calibration data. Must not change value of reference data at Serial-ROM.
Only changeable calibration data.
KV-SET = reference data + calibration data
MA-SET = reference data + calibration data
MA-COMP = calibration data
Pic. Reference address and data in Serial-ROM

KVp reference

data at 200mA

mA reference

data at 25mA

mA reference

data at 50mA

29
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

mA reference data at 75mA

Pic. Reference data address table in Serial-ROM


address description
0006E00 ~ 0006E55 KVp reference data at 200mA
0006E80 ~ 0006ED5 mA reference data at 25mA
0006F00 ~ 0006F55 mA reference data at 50mA
0006F80 ~ 0006FD5 mA reference data at 75mA
0007000 ~ 0007055 mA reference data at 100mA
0007080 ~ 00070D5 mA reference data at 150mA
0007100~ 0007155 mA reference data at 200mA
0007180 ~ 00071D5 mA reference data at 250mA
0007200 ~ 0007255 mA reference data at 300mA
0007280 ~ 00072D5 mA reference data at 400mA
0007300 ~ 0007355 mA reference data at 500mA

If you want detail calibration information see CALIBRATION MANUAL

30
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.22 JOB CODE 022


Tools & Equipment required : DVM,OSC
1. Check ROTOR connecton
Pic. Main power board and drawing

J10

SSR
CT
J7(To Interface Terminal)

SP(spark killer)

J10
1
2

CT1
0VAC
ROTOR DC24V-1 COM/MAIN/PHASE

SP1 SSR1
220VAC
D4
SSR 1N4007

J7-1
J7-2
J7-3
J7-4
J7-5
C4

24uF

ROTOR ON
Pic. Interface Terminal

To ROTOR of TUBE

31
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Pic. Rotor(anode of tube)

TEMP

G(PHASE)
B(MAIN)
W(COM)

Check connection of ROTOR wire.(Checking point is interface terminal)

Must power OFF when check the ROTOR connection.


Will accuse electrical shock.

Checking Point Value Accuracy limit


COM - PHASE Ω 58 ~ 62Ω
COM - MAIN Ω 29 ~ 32Ω
MAIN - PHASE Ω 85 ~ 95Ω

Check ROTOR input signal when pushed ready switch.


Checking Point Value Accuracy limit
COM - PHASE VAC 280 ~ 310VAC
COM - MAIN VAC 210 ~ 230VAC
MAIN - PHASE VAC 280 ~ 310VAC

Check ROTOR feedback signal and waveform(use OSC equipment)


Checking Point Value Accuracy limit
TP9 VDC 0.8 ~ 0.95VDC

Rotor feed back signal must check with Osciloscope.


Rotor rotating signal is too short time to exam.
ON OFF ON OFF

0.2sec 3sec 0.2sec

32
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Pic.Rotor feedback signal in HT control control board

Waveform of ROTOR feedback signal refer chapter 6 18.ROTOR signal

U19 R39-46 10
1 18
I1 Q1 J2-27(MAIN MAG)
2 17
I2 Q2 J2-5(POWER ON)
3 16
I3 Q3 J2-28(PRECHARGE)
4 15
I4 Q4 J2-6(CHARGE MAG)
5 14
I5 Q5 J2-7(TUBE SEL)
6 13
I6 Q6 J2-29(220/60)
7 12
I7 Q7 J2-30(HIGH SPEED RT ON)
8 11
I8 Q8 J2-8(ROTOR ON)
9 10
GND COM 1K
ULN2803

If did not output sensing signal of ROTOR sense, it will display “E26/27” in Console
display LCD.

33
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.23 JOB CODE 023


Tools & Equipment required : DVM
1. Check the feedback signal.(HT control control board)
filament feedback signal(from filament drive board)
+12V

+12V C49

TP11
104
0.6V

8
VR8 U12A D24
3
+
1 6 11
10k 2 U6
- J1_12(FIL LOW)
ULN2004
TL082A 1N4148
R63
4.7K
4

J1_13,35(FIL SEN) R65

10k
C48
104 +12V
R64
4.7K
8

U12B D25
5
+
VR9 7 5 12
6 U6
- TL082A J1_34(FIL HIGH)
ULN2004
10k 1N4148
4

TP10
C50

3.3V-12V
104

Check time is not ready switch pushed.


Check point Value Accuracy limit
Filament feedback sensing VDC 0.6 ~ 0.9VDC
Sensing LOW limit(TP11) VDC 0.1 ~ 0.5VDC
Sensing HIGH limit(TP10) VDC 3.0 ~ 3.5VDC

If you want LOW limit change, use with VR8. and HIGH limit is VR9.
If filament sensing value is lower than TP11, will display “E29” in Console Display
LCD.
And higher than TP10, will display “E30” in Console Display LCD.

34
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.24 JOB CODE 024


Tools & Equipment required : DVM
Pic. TUBE(anode)

TUBE temperature signal is


mormal close but temperature about
over than 70° will open.

Pic. Interface terminal

Temp(from tube)

If TUBE temperature signal is open, will display “E32” in Console display LCD.
J7 J7(10P)
BUCKY1
BUCKY1
BUCKY2
BUCKY2
BUCKY3
BUCKY3
TUBE1 TEMP
TUBE1 TEMP
TUBE2 TEMP
TUBE2 TEMP

Tube 2 temp not used


325/525 generator use two tube(option), so,tube2 temperature signal must come to ground

35
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.25 JOB CODE 025


Tools & Equipment required : DVM
1. Check that the wires maked “+kV, -kV, +mA, -mA” in the HV Tank are well placed and
tighten.
Pic. HV tank(High tension transformer)
1.-kV(black)
2.+kV(red)
3.+mA(green)
4.-mA(white)
5.GND(yellow/shield)
6.small
7.com
8.large

Pic. Mother board(Main card board)

From HV tank

Connection check point


Check Point Value Accuracy limit
-kV – GND kΩ 113 ~ 114kΩ
+kV – GND kΩ 112 ~ 113kΩ
+mA – GND kΩ 2 ~ 2.5kΩ
-mA - GND kΩ 2 ~ 2.5kΩ

Must power OFF When check the connection

36
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

2. Check the IPM driver


Pic. IPM Drive Board drawing.
D3
PC2
FR107
2 8 SPR(+5V INPUT1)
J1-2
R1 C7
6 6.8K MAYER CAPACITOR
3 0.1
5 CP(-KV DR1)
J1-3

HCPL4503
VPC(GND1)
J1-4

PC1
8 2 VP1(+15VOUTPUT1)
J1-1
C5
6 0.1
3 FPO(-FAULT1)
J1-5
5

HCPL4503

Table. Check Point(All power was setted originally in factory. Don’t need contrast)
Checking point Value Accuracy limit
J1 – 1,4 VDC 14.9 ~ 15.1VDC
J1 – 2,4 VDC 4.95 ~ 5.05VDC
J2 – 1,4 VDC 14.9 ~ 15.1VDC
J2 – 2,4 VDC 4.95 ~ 5.05VDC
Pic.IPM and IPM drive board
IPM
IPM drive board

J2
J1

J3
+5V
GND
J1(J2) KDR1
KDR2
FAULT FAULT
GND AC220V
-KV DR AC220V
+5V
+15V

ALWAYS HAVE “IPM DRIVER PCB” CONNECTED WHEN MAIN POWER IS


ACTIVATED IN THE GENERATOR. WHITOUT “IPM DRIVER PCB” CONNECTED,
PERMANENT DAMAGE WILL OCCUR TO IPM’S.

37
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

7.26 JOB CODE 026


If “E00” display at console display LCD when x-ray shooting.
Check point.
1. feedback waveform
Pic. Testing point kV, mA

+mA

+kV

GND
When measure “+kV” wih OSC, real kV is 10kV per 1V.
Also, real mA is 20mA per 1V.
Case 1

IPM FAULT “E00”


1. HV cable bad(over than 80kVp)  change HV cable.
2. Calibration data abnormal  check calibration data
3. IPM damage(sound “ting”)  change IPM
4. check IC TL594,ULN2004 in HT control control board and AD7224 in HT control cpu board 
KOUT, TL594 output signal waveform check.
U27
U6 13 18
6

12 D0 VDD 17
1 16 5 12 11 D1 RS
RT

I1 O1 CT VCC 11 D2
2 15 C1 10
D3
3 I2 O2 14 C2
8 9
I3 O3 16 9 8 D4 3
4 13 15 +IN1 E1 10 D5 VREF
I4 O4 -IN1 E2 7 14
5 12 3 14 6 D6 CS 16
6 I5 O5 11 1 FB REF 13 D7 LDAC
+IN2 CTL 4
7 I6 O6 10 2 4 AGND 5
-IN2 DTC DGND
8 I7 O7 9
7
GND 1
GND COM 15 VSS 2
WR VOUT
U3
ULN2004 TL594 AD7224

38
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

5. HV transformer bad  change HV transformer.


Case 2

Sometimes IPM FAULT “E00”


1. Input power(kW) “LOW”  Change exposure condition
100kVp 300mA 0.3sec  100kVp 100(150)mA 0.9(0.6)sec
This is input current low, so change mA condition.
2. IC SG3524 in filament drive board is damage.  change new IC.

2. Aanother factor.
1. If IPM driver board broken, always display “E00”(power on state ,not exposure state)
 change IPM driver board
 check IPM drive input power(220VAC) and signal connection.
2. If IPM was damaged, sound “ ting—“and display “E00” at high kVp exposing
 change IPM

39
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

8. X-RAY CHECK
1. FILAMENT DRIVE BOARD CHECK
Tools & Equipment required : DVM,OSC
Pic. Filament drive board
① filament enable signal (high  low)

② control waveform (see waveform 1)

③ output in gate of Q3,Q4(see waveform 2)

④ preheating wave(see waveform 3)

ready wave with high mA data(see waveform 4)

ready wave with low mA data(see waveform 5)

② ③

DC0V - 100V

+12V

+ C22 ZD3 Q11


IRFP250 + C29
D9 D10 D11 C12
100uF/35V 1N4745 T3
470uF/200V R37
224 47K/3W
ZD6
FR102 FR102 FR102
D7
R27 R28 Q7 1N4745
FR102 1K 1K 2N2907 C24
R40
473
Q8 C14 C15
100/3W
CLK C23 ZD7
R29 102 104 104
Q3 1N4745 R41
IRF630 1 1/2W 2N2907
+12V C31 0.5/3W
ZD5
1N4733 C25
C21
15

D12 1uF
1

VR2 R52 0.1 R30 222/1KV


6 12 T5
V-

VIN

RT CA
COMP

13 U3 ZD4 Q12
5K 4.7K CB 4.7 1/2W
3 1 8 IRFP250
C18 OSCOUT NC NC FR102 D13 D14 D15
7 11 2 7 1N4745 C13 + C30
CT EA 3 INA OTA 6 T4
472 GND VDD R42
14 4 5 224 470uF/200V
10 EB INB OTB ZD9 R38
SHTDWN FR102 FR102 FR102 1K
4 D8 R32 47K/3W
+SENSE TC4427
5 R31 Q9
VREF

-SENSE 1K 1N4745
8 R19 R18 FR102 1K 2N2907 C27
V+

GND
1K 1K
U2 Q10 473 C16 C19
C26 ZD10
16
2

LM3524
R33 102 104 104
Q4 1N4745
R23 VR3 R22 IRF630 1 1/2W 2N2907
ZD8
1K 5K 1K
1N4733 C28
R9
J2-30(MSY) D17 1uF Q11
1K R34 2P4M R35
+12V
4.7 1/2W 560
C32 R36 R39
R51 FR102
R49 R46 100 0.5/5W
FSENSE 104
D19
10K 20K 20K
1N4148
R47 R45
R50 C34
+

10K 20K R43


10K 10uF
R48 1K
D22 D18
20K
R44 1N4148
1N4148
D21 1K C33
C35 1N4148 D20 102

+12V
-12V
0.1 1N4148 J1-4,5,26,27(FIL OUTA)
C37
8

+ U5B U5A
6 2
- - J1-6,7,28,29(FIL OUTB)
100uF/35V 7 1
5 3
+ +
TL082 TL082
4

J2-19,20,41,42(-12V) C36

-12V

0.1

① ④

40
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Filament control frequency check


Checking point Value Accuracy limit
CLK in U2(SG3524) ㎲ec 65 ~ 75㎲ec
Adjust point : VR2

Q1 R1
J2-10,11,32,33(AC60V) BUX48A 0.5/3W
BD1
NF1 - +
Q2 R2
KBU10M
BUX48A 0.5/3W
J2-13,14,35,36(AC60V)
NOISE FILTER Q3 R3

BUX48A 0.5/3W

J2-15,37(EARTH) + C1 + C2
R11 Q4 R4 R10
47K/3W DC0 - 100V
470uF/200V 470uF/200V BUX48A 0.5/3W 0.2/5W
+12V
C7
Q5 R5 R12 R14
TIP42 1K/3W 104 12K 12K
U6B

8
D1 LM358 6 R16
R6 -
R24 7
47K
5.2K 5
2.2K/1/2W +
R13 R15
D3 1N4148 R17
Q6 390 820

4
2N3440 47K
+12V C4
1N4148
C5
J2-21,22,43,44(GND)
0.1 0.1
R7 R8 U6A

8
+ C17 2.4K 2.7K
J2-17,18,39,40(+12V) D2 LM358 R20
2
-
100uF/35V 1
47K
3
+
1N4148 R21

4
47K
C3
R54 0.1

10K
J2-31(FOUT) C8

+12V
8

VR1 R57 104


2
-
1
2K 10K
3 U4A
+ R55
TL082 R53
C9 120K
1K
4

R58
-12V

8
10K R56 6
104 -
7
100K
5 U4B
+
R26 C6
TL082 R25
4.7K 0.1
10K
4

C10
+12V
8

104
8

6
-
7 2
J1-13,24,35(FSEN) -
U1B 5 1
+
U1A 3
TL082 + FSENSE
C11 TL082
4

-12V

104

⑤ ⑥ VR1

⑤ Filament feedback signal (to HT control control board)


⑥ FOUT signal (from HT control cpu board)
All mA step condition will change uniformly with VR1 contrasting FOUT
signal.
And, If you want each mA step condition change, use Calibration mode

BECAREFUL WHEN YOU CHECKING “FOUT SIGNAL”. IF ELECTRICAL


SHORT HAPPEN, U25(AD7224) IN HT CONTROL CPU BOARD WILL BE
BROCKEN.

41
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Pic.Filament drive board

SomeTimes, U1(TL082) is broken by filament noise.


This IC operate feedback sensing regulator.
If U1 broken, E30 display in Console LCD. Because, filament
feedback signal is pull-off(12VDC) by broken IC.
So filament feedback signal feedback “HIGH” by filament sensing
circuit in HT control control board.(see JOB CODE 023)

42
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

2. HT CONTROL CONTROL BOARD CHECK


Tools & Equipment required : DVM,OSC
Pic. HT control control board

Test exposure condition : 40kV/70kV, 200mA ,0.025sec)


④ ⑧
+12V

R38 R37
R9
4.7K 4.7K U4B
10K
U6 15 14
7 10 16 D S
IN
U1A J2_26(R/F)
TL082A R8 ADG201A
8

KV
ZD3,ZD4을 1N4744을을 1N4746을 을을을 2001을 8을 29을 2 ULN2004 U4A
C8 - 1M
1 2 3
R19 J1_1(KSEN) D S
3 1
100P + IN
J1_27(-MASEN) R15 1M
J1_5(+MASEN) C4 ADG201A
4

154K
C21 R16 C9 C10 103
-KV
+12V -MA
104 10K 273 100P
8

R10 R12

8
6 U2B U1B
- R17 R18 R1 R3
7 6
10K 1M -
5 7
+ 34K 100K 1K 100K
C5 ZD3 D5 D6 5
TL082A +
C1 ZD1 D1 D2
TL082A R6
102 1N4744 1N4148 1N4148
4

C22
104 1N4752 1N4148 1N4148 10K MA

4
J1_26(-KVSEN) -12V
T2
J1_4(+KVSEN) +12V -12V 104
100uH T1
U4A +12V -12V
C7 VR2 100uH
C6 ZD4 D7 D8 3 2
S D 1 C2 ZD2 D3 D4 C3
100P 20K IN
102 1N4744 1N4148 1N4148 R7
103
R11 R13 R14 ADG201A 104 1N4752 1N4148 1N4148 10K
U4B R2 R4 R5 J1_23(MSEN)
10K 1M 154K VR1
14 15
S D 16 1K 100K 10K
+KV 10K IN
+MA
ADG201A R28
+12V
U5B
1K
14 15 TP2
S D 16 C23
IN C15
6

104 U6
C18 5 12 TP3 TP4
ADG201A
RT

CT VCC 11 1 16 R33-R36
C1 I1 O1 J1_7(KDR1)
8

U2A 683 8 2 15
J2_32(KOUT) 103 C2 I2 O2 J1_29(KDR2)
3 16 9 3 14
+ R22 +IN1 E1 I3 O3 J1_8(KDR3)
1 15 10 4 13
-IN1 E2 I4 O4 J1_30(KDR4)
2 3 14 5 12
- 24.9K FB REF I5 O5 10
D9 D10 1 13 6 11
TL082A R23 C12 2 +IN2 CTL 4 7 I6 O6 10
7 -IN2 DTC 8 I7 O7 9
1N4148 1N4148 52K
4

C11 GND R29 R30 GND COM


273
R24 C14 TP1 1K 1K
+12V -12V U3 ULN2004
100P 100K R25 R26 C17
TL594 C16 R27 VR3
104 1K 1K
D11 D12 C13 10K 104
0.22 10K

1N4148 1N4148 103


R20 R21
10K 4.99K

D14 U5A 13 C24


J2_9(EXP)
2 3
J2_31(KSY) 1N4148 D S 104
1 U4
IN ADG202A
R31 R32 D13 4 C25
ADG201A
1K 1K 1N4148 U5C 104
10 11
9 D S
+5V IN 13 C26

ADG201A 104
U5
U5D ADG201A
7 6 4 C27
8 D S
IN 104

ADG201A

① ② ③ ⑤ ⑥ ⑦

43
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

① kV output data voltage from 2 of U27 in Ht control cpu board 1(when exposured)
kV low (40kV see waveform 6)
kV high(70kV see waveform 7)
② exposure/kV enable signal
high  low (see waveform 8)
③ TP 2 is control waveform(see waveform 9)
Checking point Value Accuracy limit
TP2 kHz 39 ~ 41kHz
Adjust point : VR1
④ test point : -kV, +kV feedback signal waveform
+kV feedback (see waveform 10)
-kV feedback(see waveform 11)
1V ≒ 10kV
⑤TP1(dead time control)
Adjust point : VR3 (see waveform 12, when exposure)
⑥TP3,TP4 : TL594 output point
kV data low(see waveform 13)
kV data high(see waveform 14)
E1 is 90 degree phase ahead of E2
⑦KDR1,KDR4 (see waveform 15)
KDR1,4 is 90 degree phase ahead of KDR2,3
KDR1 ~ KDR4 signal input to IPM DRIVE BOARD
⑧ test point : -mA, +mA feedback signal waveform
+mA feedback (see waveform 16)
-mA feedback(see waveform 17)
1V ≒ 20mA

BECAREFUL WHEN YOU CHECKING “KOUT SIGNAL”. IF ELECTRICAL


SHORT HAPPEN, U27(AD7224) IN HT CONTROL CPU BOARD WILL BE
BROCKEN.

44
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

9. MAIN CARD BOARD DRAWING


J3(3P) J4(4P) J5(11P) J6(8P) J7(10P) J8(4P) J9(7P)

J11(5P) J10(6P)
J14(10P) J1(15P) J2(15P) J15~J18(NOT USED) J19(6P) J13(7P) J12(7P)
J3 J7 J11 J14
TUBE2 AC220V BUCKY1 DC24V-2 -KV
BUCKY1 GND +KV
TUBE2 AC220V BUCKY2 PRECHARGE +MA
BUCKY2 CHARGE -MA
BUCKY3 MAIN GND
J4 BUCKY3 SMALL
TUBE2 READY TUBE1 TEMP COM
TUBE2 X-RAY TUBE1 TEMP J12 LARGE
FLUORO MODE TUBE2 TEMP GND TUBE2 AC220V
GND TUBE2 TEMP +5V TUBE2 AC220V
KDR4
J5 J8 KDR3 J19
AC220V AC220V(N) IPM FAULT AC220V
AC220V AC220V(N) AC220V AC220V
AC12V AC60V AC220V +12V GND
AC12V AC60V +12V
IPM AC220V +5V GND
IPM AC220V J9 J13 +5V
AC220V(N) AC220V GND
AC220V(N) AC220V +5V J1,J2
DC24V-1 AC15V KDR2
GND AC15V KDR1 : REFER JOB CODE002
DC 24V-2 AC20V IPM FAULT
AC20V RTN AC220V
AC20V AC220V
J6
ROTOR ON
POWER ON J10
220/40 CHANGE DISCHARGE4
TUBE SELECT DISCHARGE3
ROTOR SENSOR DISCHARGE2
GND DISCHARGE1
AC110V CHARGE SENSE
AC110V GND

45
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

10. TECHNICAL IMFORMATION


10.1 MAIN SIGNAL DRAWING

CONSOLE

DISPLAY AND

MEMBRANE
CONTROLLER

CONSOLE CPU
COMMUNICATION CABLE HT CONTROL
BOARD
CPU BOARD KOUT
Feedback

HT CONTROL IPM
IPM
CONTROL BOARD driver

FILAMENT DRIVE FOUT


+ TUBE ―
BOARD
HV cable

RELAY POWER
BOARD

HV transformer
MAIN POWER
+
Filament power BOARD
- terminal

T1 and T2
ROTOR power

46
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

10.2 MAIN SIGNAL EXPLANATION


10.2.1 communication

Console display CPU and HT Control CPU communicate with cable.


This communication included below like as.

kV mA Time Bucky

1ms 1ms 1ms

8bit 8bit 8bit 8bit

Console display CPU give kV data,mA data,time data,bucky select data to HT control CPU.
And HT control CPU re-give same data to Console display CPU.
So, Console display CPU check data error or good. If error detected, display EROOR code.
Ex) KV data error(E16,E17), mA data error(E18,E19), time data(E20) I

Pic. Console display CPU board

TP1(Txd), TP2(Rxd)

47
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Pic. HT control CPU board


TP1(Txd)
TP2(Rxd)
TP3(Txd)
TP4(Rxd)

TP3 TP4 TP1 TP2

If push KV up/down key in Console membrane.

kV mA

mA , time, bucky select signal not transfer. Only kV data data transfer
8
cf . kV data 8bit = 2 P
P
= 256 information

48
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

kV data(138+1 = 139)

KV-SET data “139” in Console display cpu board give to HT control cpu board
And, HT control cpu give data to AD7224
When communicate HT control cpu board with console display cpu
board, sometimes error detected by electrical noise.
1. Check ground state.
2. Check shield state of communication cable.
3. Check noise of Power line.
]

49
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

10.2.2 Exposure data


When exposed, kV data, mA data input to AD7224 with digital value( 0 – 256 information).
And then, analog value(Kout and Fout) output from AD7224.

Pic. HT control CPU board

AD7224(8-Bit DAC with Output Amplifiers)


ESD (electrostatic discharge) sensitive device. Electrostatic charges as high
as 4000 V readily accumulate on the human body and test equipment and
can discharge without detection. Although the AD7224 features proprietary
ESD protection circuitry, permanent damage may occur on devices
subjected to high energy electrostatic discharges. Therefore, proper ESD precautions are
recommended to avoid performance degradation or loss of functionality.

50
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Analog Kout data give to HT control control board

Analog Fout data give to Filament drive board

51
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

Pic. HT control cpu board

TP5 : Analog Kout data


TP6 : not used
TP7 : Analog Fout data

AD7224

If you want to see waveform TP5,TP7(when push x-ray s/w) see Chap.6 6 and7 waveform.
BECAREFUL WHEN YOU CHECKING “FOUT SIGNAL”. IF ELECTRICAL
SHORT HAPPEN, U25(AD7224) IN HT CONTROL CPU BOARD WILL BE
BROCKEN.

BECAREFUL WHEN YOU CHECKING “KOUT SIGNAL”. IF ELECTRICAL


SHORT HAPPEN, U27(AD7224) IN HT CONTROL CPU BOARD WILL BE
BROCKEN

10.2.3 Feedback signal


Pic. Ht control control board

1 2 3 4
1. KOUT signal 2. Feedback signal

52
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

3. TL594 output signal 4. IPM drive signal

TL594(494) receive two signal (KOUT(-ini) and feedback signal(+ini))and give output
signal(E1,E2) to IPM drive.
If KOUT(analog data) is much, width of TL594 output signal would be widen.

TL494(594) Technical data

53
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

54
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

55
COMED Medical Systems Co.,Ltd.
EVA-HF325/525
Troubleshooting

10. JOB CODE IMFORMATION

ERROR CODE OR DESCRIPTION JOB CODE


ERROR
JOB CODE
NOT POWER ON/OFF Power on/off circuit abnormal
001/002/003/004
E16/17/18/19/20/21/22 Calibration and APR data broken JOB CODE 020/021
E26/27 Rotor Rotating abnormal JOB CODE 022
JOB CODE
E28 Charge and discharge abnormal
008/009/010
E29/E30 Filament drive board abnormal JOB CODE 007/023
E32 Tube temperature abnormal JON CODE 024
E00 IPM fault JOB CODE 025/026
JOB CODE
Only display “Initializing.” Not communication control and console
002/004/005/006/007
NOT KEY UP/DOWN Membrane key abnormal JOB CODE 003

56