Technical Specification of

Process and Product Development Research Centre


Equipment Specification Quantity Piece Total
Cost Estimate
d Cost
1 Cutting mill Feed material = soft, medium-hard, elastic, fibrous
Feed size max.= 60 x 80 mm
Final fineness (sieve) = 0.25 - 20 mm 1 40,000 40,000
Collecting receptacle = 5 liter (standard), 30 liter
(option)
Throughput = 0.2 - 50 kg/h
parallel section rotor, 3 blades; 5 liter collecting
receptacle with filter bag; bottom sieves
with perforations from 0.25 mm - 20 mm;
motor rating = 1500 W; rotor speed of 1390 min-1
at
2 Heavy- Duty For large pieces of tough = 695 min-1 version at 50
Cutting Mill Hz 1 50,000 50,000
(835 min-1 at 60 Hz)
For thermally sensitive materials = 1390 min-1
version at 50 Hz (1690 min-1 at 60 Hz)
3 Disk Mill 6-disc rotor,
Feed size 1.5 kW single phase or 3-phaser
<6mm
Discharging size (mesh) = 8-180
Main shaft speed (rpm) = 1050
Throughput (Kg/hr) = 10-35 1 8000 8000
Dimensions (mm) = 550*490*365
Weight (Kg) = 130
Power (Kw) = 1.1

4 Readability= 0,0001 g
Mass Capacity = 210 g
Balance Pan size = (dual range) 1 17,550 17,550
(Semi-Micro
Analytical
balance)
5 Readability = 0,01 g
Mass balance Capacity = 610 g
Pan size = 142 x 130 mm 1 10,000 10,000
6 Brass frame with stainless steel mesh or stainless
Sieve frame and mesh 1 30,000 30,000
Diameter = 20 cm
analyzer
Width of Aperture = 4.75 mm-850 Microns

7 Hot plate Stirring Volume = 0~2000Ml


with mixing Stirring Speed = 0~1250rpm 1 16,000 16,000
Heating Temperature = 300°C
Top Plate Size = 135×135mm
Panel Material = Stainless Steel
Power Requirements = AC 220V/50Hz or
110V/60Hz
Weight = 2.2kg

8 Microwave Up to 1,000 watts of microwave power


Oven 5 different power levels
Bottom feed microwave using mode stirrer technology
for even heating 1 48,000 48,000
Cavity = 330mm wide x 205mm tall x 330mm deep
(13” x 8 1/16” x 13”)
Stainless steel door and front
Input power = 120 Volts / 60 Hz. 15-amp
Frequency = 2.45 GHz.
Cavity venting Powered
Microwave leakage = < 5 mW/cm2 at 5 cm meeting
requirements

9 Thermo  Balance
gravimetric Sensitivity = 0.1 µg
Analyzer Capacity = 1300 mg
Accuracy Better than = 0.02%
Precision = 0.01%
 Temperature 1 140,00 140,000
Furnace = Low mass (<10 g); Platinum heating 0
element with ceramic protective coating; resistant
to inert and oxidative gas over the full temperature
range
Range = -20 °C to 1200 °C
Scan rates 0.1 °C/min to 500 °C/min
Precision = ±1 °C
 Cooling
Method = Forced air cooled with an external fan
Cycle time = 1100 °C to 50 °C
Sample Pans = Platinum or ceramic (60 µL)
Dimensions (H x W x D) = 44 x 43 x 43 cm
Weight = 26 Kg (57 lbs.)
Power Requirements = 100 to 240 Volt, 50/60 Hz

10 Differential Dual, digital mass flow-controller


Scanning Cooling accessory upgrades
Calorimeter Automated DSC cover
Semi-automated sample loading accessory 1 120,00 120,000
Dynamic range = ±1300 mW 0
Accuracy = <±0.2%
Precision = <±0.03%
Indium height/width = 18.4 mW/˚C
Indium melting time = 2.3 sec

11 Oxygen Bomb calorimeter with embedded computer control,


Bomb us, LCD graphics display, high accuracy/resolution
Calorimeter / PRTs, 2 × RS232 interfaces, parallel port for printer 1 92,000 92,000
Bomb Oxygen bomb and bucket (calorimeter vessel)
Calorimeters Thermostatically controlled bath, circulator, cooler,
pipette (2L)
EN ISO 1716 sample preparation device, firing wire
and cotton, cigarette paper
Automatic Firing
Calibration

12 Kjeldahl Kjeldahl Connecting Bulb, Kjeldahl Digestion Flask,


assembly, Condenser, Connecting Tube, Erlenmeyer flask Pure
both as Gum Rubber Tubing. 1 600,00 600,000
digestion and Capacity to hold 6 round-bottomed flask of 300 ml 0
distillation capacity each complete with condenser and
set, connecting tubes Heating capacity of 500 W of each
heater
13 Fumehood - To hold Kjeldahl assembly of 6 sets and stand alone
digestion 1 60,000 60,000
chamber

14 Digital Viscosity = 20-6000000 mpa.s


Viscometer Maximum Viscosity Range = 6000000 mpa.s
Features = Selectable speed, 1 30,000 30,000
LCD Display
15 Freeze dryer Power = 1070 W
Max Freeze Dry Area = 0.12 m2 1 80,960 80,960
Plate load material = 1.2L
Cold Trap Temp = -55 /-80 deg C
Vacuum Pressure = < 10 Pa
Length 565 mm Width 460 mm
Performance (drying rate) = 4 Kg/24h

Capacity = 4.3 lit


16 Rotary Automation = Grade Automatic
Vacuum Frequency = 50 Hz 1 15,000 15,000
Evaporator Voltage = 220 V
Frequency = 50 Hz
Power = 400 W
Temperature Control = +10-200o C
Rotating Flask = 1L
Collecting Bottle = 1L
pH
17 Ion-Selective Range -2.00 to 16.00 Resolution 0.01
electrode Accuracy ± 0.02
Auto buffer recognition 4, 7 and 10.05 (with manual 1 10,000 10,000
override)
Calibration User selectable 1 or 2 point
mV
Range ±1999
Resolution 1mV
Accuracy ±1mV
Temperature
Range -10 to 105°C
Resolution 0.1°C
Accuracy ± 0.5°C
ATC range 0 to 100°C
Ion: Multipoint calibration 1 to 5 points
Activity :1x10-9 to 9.99x109 in integer or exponential
format
Concentration units: activity, %, ppm, mg/l, moles,
none
Dedicated ions: Ammonium, Calcium, Chloride,
Fluoride
and Nitrate
Clock: 24 hour, hours/min/sec or day of month, month
and year. Leap year corrected
Outputs: RS232 via interface cradle
Connector: Waterproof
Power: 3 AA batteries
Size (l x w x d): 200 x 80 x 60mm
Weight: 370g
Operation = Electrically Operated
18 Cream Capacity = 1000 LPH (250 Gallons)
Separator Supply of Milk = Online Supply
Technology = Centrifuge Process to Separate Cream 1 36,900 36,900
Power Source = 220-230 Volts 50 Hz / 110 Volts 60
Hz
Separation Bowl & Disc = Food Grade Stainless Steel.
Milk Spout & Cream Spouts = Food Grade Stainless
Steel

Soxhlet Measuring sample quantity: 6 simultaneously


19 Extraction Heating time: within 10 min
Fat Analyzer Recovery system: automatic 1 160,00 160,000
Recovery rate: ≥80% 0
Temperature controlling range: room temperature -
100°C
Power: AC220V/50Hz, 1000W
Shape size: 760×230×680mm
Weight : 25kg
Difference Even heating and automatic recycling

20 Nitrogen/Pro Analysis Time = 6 mins, depending on sample


tein Analyzer combustion conditions
Sample Loader: Stackable auto loader = up to 140
samples by 4 layers 1 40,000 40,000
Repeatability = Nitrogen≤0.05%
Sample Mass = Up to1000mg,
Temp. Resolution = 1℃ depending on sample matrix
Gas Required = Helium, 99.99%, 0.25± 0.01 Mpa
Oxygen, 99.5%, 0.25± 0.01 Mpa Nitrogen or
compressed air, 0.25± 0.01 Mpa
Consumption = Helium 200ml/min
Measurement Rang = Nitrogen: 0.04mg-50mg
Furnace Type = Resistance furnace, Max temp 1050℃
Power Supply = Single phase, AC220V ±10%,
50/60Hz, 5.5kW
Net Weight = 110kg
Dimensions(L×W×H) = 690mm×750mm×720mm

21 Chemical Temperature of Reaction = 105 oC or 150o C


oxygen Temperature Stability = +-0.5oC
demand Temperature range = -10oC to 160 oC
(COD) Temperature Accuracy = +-2oC (@ 25oC) 1 48,000 48,000
measurement Capacity = 25 vials (dia 16 x 100 mm), 1 receptacle
for a stainless-steel reference thermometer
Warm up time = 10 to 15 minutes, depending on
selected temperature
Operating Mode = Timed (0 to 180Min.) or infinity
mode

22 Biological Measurement range = 0~1000 mg/L


oxygen Application = laboratory use Automated 1 30,000 30,000
demand
(BOD)
Measurement
23 Soxhlet Material = Glass
Extraction Capacity = 500 ml 2 18,480 36,960
Temperature = 600 o C
Color = Transparent
Usage =Laboratory

24 Laboratory Distillation Column Diameter = 21 mm


Fractional Distillation Column Length = 90 cm Long
distillation Distillation Column Jacket = Vacuum Jacket and
Silvered “Mirror” 1 36,000 36,000
Condenser = Inner Coil and Outer Jacket High Surface
Area
Boiler Heater = Electrical
Boiler Sizes = 12-liter sizes Boiler Stirring Standard
Spinning Band Types = Teflon-Operates up to 225 °C
Stainless Steel
Operates Above 225 °C
Boiler Stirring Standard
Spinning Band Types = Teflon-Operates up to 225 °C
Stainless Steel - Operates Above 225 °C
Maximum Theoretical Plates
Teflon Band - 50
Stainless Steel - 20
Condensing Bath = Optional Circulating Bath

 Bath33 0-40 °C
 Bath140 -20 to 100 °C

Fraction Collector = Standard 2 Place Optional 4


Place or 8 Place
Vacuum Operation

25 Ion The system must be able to analyze most of the anions


chromatogra like fluoride, chloride, nitrate, bromide, phosphate,
phy sulphate, nitrite, nitrate, arsenate, chromate, carbonate; 599,41 599,412.9
cations like lithium, sodium, potassium, calcium, 1 2.94 4
potassium, amines, barium, strontium, amines and
transition metals by conductivity detection and using
diode array UV VIS detector.
• PC based system = with software-based control and
data acquisition which can perform both suppressed
and non-suppressed conductivity applications. The
software should be able to identify pumps, columns,
detection automatically.
Software should efficiently acquire & process the data
and it must have a complete system control. The
necessary software should be fully Windows based.
• PUMP: A high pressure pump of serial dual piston
type must be provided which must have flow rate of
0.001 to 10 mL/min with a resolution of 0.001 mL and
flow reproducibility of ±0.1%. Pulsation must be
lower than 1% with a pressure range of 0 - 5000 psi.
CONDUCTIVITY DETECTOR: A microprocessor-
based conductivity detector for analysis of anion and
cation, with a thermostat based micro-flow cell
conductivity block with an accuracy in the range of ≤
0.0010 C. The user should be able to set temperature of
the conductivity block between 20 - 50 ºC.
Conductivity measurement range must be between 0 -
15000 µS/ cm with electronic noise < 0.1nS/cm at
1uS/cm level and Temperature coefficient in the range
of 0-5%.
DIODE ARRAY UV-VISIBLE DETECTOR:
Detector must be of UV-Visible range must have
minimum 6 data channels and one reference channel
for speciation of Iron, chromium, organic acids using
Post Column Derivatization Technique. The detector
should be of variable wavelength type and the range
should be between 190 to 900 nm. It should have a
compact Post Column Derivatization Unit with a
complete set of reagent delivery pump, mixing T joint
and reaction coils.
IC COLUMNS AND COLUMN HOUSING:
Columns should be coated with respective guards and
must possess an electronic chip for storing data and
complete usage history including number of injections
and working hours. Housing should be able to identify
the columns and set the optimal operating conditions
for column operations.
INJECTOR: One injector port with number of Dual
position 6-Port injector valve with fast response time
and controlled through software. CHEMICAL
SUPPRESSOR: Chemical suppressor should have
high loading and high backpressure capacity (at least
350psi) with continuous regeneration and must be able
to take high flow rate up to 10mL/min. The
regeneration of the suppressor should be by external
mode with an option of controlling the regenerant flow
rate. Suppressor should have at least 8 years or more
warranty with warranty certificate from manufacturer.
AUTO SAMPLER: An autosampler of at least >30
positions with 0.1 - >10mL sample capacity vials
should be quoted along with the system. The entire set-
up provided should be automatic and controlled
through the software.
26 Crucible Capacity = 30 mL 10 700 7000
Gooch Height = 60 mm:
Diameter = 30mm:

27 Flame Purpose
Photometer • For analysis of sodium, potassium calcium and
lithium in (water) samples.
Conditions & Requirements 1 24,750 24,750
• The instrument shall be fully installed and
demonstrated by the supplier.
• The photometer shall generally comply with IS8651-
1977
Specifications
Unit
Microprocessor controlled flame photometry
Scale range
0 to 100% sensitivity for full scale deflection:
Response-sodium
2 mg/l or less
Potassium
1 mg/l or less
Calcium 5mg/l
Lithium 5 mg/l
Accuracy 2% of reading
28 Horizontal Max temp (°C) = 1200
Tube Number of heated zones = Three zone
Furnace Heated length (mm) = 450
Heat-up time (mins) = 55 1 200,00 200,000
Max outer ø accessory tube (mm) = 60 0
Tube length for use in air (mm) = 600
Tube length for use with modified atmosphere (mm)=
900
Dimensions = External H x W x D (mm) 560 x 615 x
390
Max power (W) = 2000
Thermocouple type = N
Weight (kg) = 20
Horizontal configuration

29 Bio line Chamber Volume = 250 Liter


Laboratory Number of Shelves = 3 Shelves 1 16,000 16,000
Incubator Temperature Range = 5 to 60 deg C
Accuracy = +/-0.5 deg C
Voltage = 230V AC
Frequency = 50 Hz
Temperature Uniformity = +/-1 deg C
30 Biosafety
Cabinet Cabinet Material = Stainless Steel
Main Filter = HEPA Filter 1 20,000 20,000
Sterilization = UV Germicidal Tube
Electric Supply = 230
Noise Level = Less than 65 db

31 Automatic Voltage = 220/110V


Bioreactor Stirring Power(W) = 60
Stirring Speed(rpm) = 0-600 (max 1300) 1 80,000 80,000
Motor Torque(g/cm) = 4800
Reaction Kettle Capacity(L) = 5
Stirring Shaft Diameter = 12mm
Discharge Method = Discharge from bottom,
glass discharge valve

32 Filtering long-stem funnel; rubber spatula; beaker covering and


funnel watch glass; 10 200 2000
Funnel stand: To hold eleven 5-7 cm diameter glass
funnels, to correspond to flask stand
33 Ph/Mv meter Range 0-14 Ph with accuracy of ±0.05 Ph, complete
with combination electrode, mains operated; to work
on 220 V, 50 cycles;Ph 1 15000 15000
• Range: -2.000 to +20.000
• Resolution: 0.001 / 0.01 / 0.1
• Accuracy: ±0.003
• Calibration: User selectable 1, 2 or 3 point
• Automatic buffer recognition: Jenway (2.00, 4.00,
7.00, 9.20 and 10.00), DIN, NIST, JIS Mv
• Range: ±1999.9Mv
• Resolution: 0.1/1Mv
• Accuracy: ±0.2Mv
• Temperature: Range -10 to 105°C
• Resolution: 0.1°C
• Accuracy: ±0.5°C
• ATC and manual temperature Correction: 0 to 100°C
• Alarm points: -2.000 to 19.999Ph
• GLP: Calibration reminder interval (1 to 999hrs)
Alarm outputs (open collector and audible). Security
code protected data
• Outputs Analogue: RS232 and IrDA
• Connector: BNC
• Power: 9V AC ±10% @ 50/60Hz
• Size: (l x w x h) 210 x 250 x 55mm
34 CHNS • Smooth wall tine capsules for elemental analysis ;
elemental pans: pressed weighing pans; aluminum pans for
analysis: moisture determinations; 1 750,00 750,000
• Operating Mode: CHN, CHNS and OXYGEN 0
• Accuracy: 0.3% abs
• Analysis Time: 6 to 8 minutes per sample
• Automatic Weight Transfer
• Sample Size: 1 to 200 mg
• Multiple Sample Mode if required: 60 samples
• Sample Combustion Temperature: Around 2000 º c.
35 Lab Portable Voltage; 220/110V
Plug Flow Stirring Power(W); 120
Jacket Glass Stirring Speed(rpm); 0-600 1 56,000 56,000
Chemical Motor Torque(g/cm); 1000
Reactor Reaction Kettle Capacity(L); 10
Stirring Shaft Diameter; 120mm
Discharge Method; Discharge from bottom, glass
discharge valve
36 Laboratory Vacuum degree (Mpa); -0.098
Glass Lined Reaction flask(L); 10
Reactor Rotation speed (Rpm); 0-500 1 60,000 60,000
Batch Stirring shaft diameter(mm); 12
Reactor Motor power(W); 90
Voltage(V); 220/50
Size (CM); 65x50x200

37 Thermostat Working room size; 100L


Incubator Temperature. Range; RT+5-65ºC
with Shaker Temperature fluctuation; ±0.5ºC 1 28,800 28,800
Vibration frequency (single layer); 20-260 rpm
Vibration frequency (double layer); 40-200 rpm
Load; 3kg
Amplitude of vibration; 20mm
Ambient temperature; 10-30ºC
Pallet dimension; 400*340mm
Timer range; 1-9999min.

38 Laboratory Blower; Built in centrifugal blower, speed adjustable


Chemical Internal Size(mm); 1380*750*745mm 3 17000 51,000
Laminar Air External Size(mm); 1500*800*2200
Flow Fume Base Cabinet;
Hood Height is 650 mm, the exterior size includes the cabin
et height
MAX Opening; 450mm
Work Surface Height; 850mm
Protection Class; Class I
Protection Type; IP 20
Air velocity; 0.3 to 0.8 m/s
Noise; ≤ 60Db
Exhaust Duct; Φ300,4meters, PVC
Fluorescent Lamp; 36w*1
UV Lamp; 30w*2
Exhaust Duct; PVC, standard length: 4 meters
Power Supply; 110-220V/50-60Hz(optional)
Power Consumption; 235W
Waterproof Socket; Two, totally 1000W
Standard Accessory;
Water tap: 1; Water sink: 1; Gas tap: 2; Water proof so
ckets2
;Base Stand; Active carbon filter, foot switch
Material; Exterior body: Cold-
roll steel sheets, coated with anti-bacteria powder
Side Panel and front window:5mm thickness toughene
d glass, anti-ultraviolet radiation
Work table: Using Solid Chemical-
Resistant physicochemical board
Gross Weight; 370kg

Laboratory Material; Borosilicate Glass


39 Color; transparent 1 4000 4000
Desiccator Size; 210mm
Shape; round
Back type; box
Temperature; 50-60 oC
Atomic Working spectral range; 190~900nm
40 Wavelength accuracy; ≤±0.5nm 1 280,00 280,000
absorption Wavelength repeatability; ≤0.3nm (single direction) 0
spectrophoto Spectrum bandwidth;
meter 0.2nm, 0.4nm, 0.7nm, 1.4nm, 2.4nm, 5.0nm
Resolution; <40%
Base line stability; ±0.004Abs/30min
Lamp stand; 1
Gases detected CH4, CO2, H2S, O2
41 Portable Measuring range; CH4: (0 ~ 100) %vol, CO2: (0~100) 1 60,000 60,000
biogas %vol H2S: (0~1000) ppm, O2: (0-30) %vol,
analyzer CH4: 1%vol, CO2: 1%vol, H2S: 1ppm, O2: 0.1%
Sensor; Infrared sensor and electrochemical senses
≥5years (CH4, CO2) ≥2years (H2S, O2)
CH4: ≤70S (T90), CO2: ≤70s (T90), H2S: ≤45s (T90),
O2: ≤45s (T90)
Alarms; Audible—buzzer with ≥80dB audible
Alarm point; Visible—flashing red LED, ≥20m visible
Data logging (optional);
Continuous Data logging (1900 data at
Continuous working time; configurable 0.1-
10 minutes intervals)
Display; ≥13hours
Battery model; Liquid Crystal Display (LCD)
Charging time; PL123450, 3.7V/1500mA
Dust and water resistance; 4-5 hours
Size & Weight; IP54
Ex sign; 130*66*30mm, 500g
42 Moisture Sample maximum weight ;10g
Analyzer Division value; 0-1g 1 11,600 11,600
Temperature range; 60 -170OC
Timer; 1-30min
Accuracy; +/-2%
Power supply; 220 V
43 Muffle Max. Temperature = 1450°C
furnace Working Temperature = 1350°C
Temperature Accuracy = +/- 1°C (+/- 1.8°F) 1 80,000 80,000
Temperature Controller = PID controller
Display = LED / LCD
External Chamber = Construction MS w/ Powder
Coating / 304 Grade Stainless Steel (Optional)
Internal Chamber Construction = Ceramic Board &
Grooved Refractory Chamber as per Temp.
Requirement
Insulation Ceramic; wool insulation
Alarm = Audible & Visual type
Power Supply = 220 / 440 Volts
Dimension (Inches) = 12 x 12 x 12
Volume (Liter) = 28
Muffle Furnace Accessories
 Chart Recorder
 Tongs
 Heating Element
 Gloves

44 K type Temperature Range with Thermocouple Grade Wire 5 200 1000


Thermocoupl Thermocouple grade wire = -454 to 2,300F (-270 to
e 1260C)
Extension wire = 32 to 392F (0 to 200C)
Accuracy (whichever is greater): Standard = +/- 2.2C
or +/- .75%
Special Limits of Error =+/- 1.1C or 0.4
45 Ultra-low Upright Freezer with capacity at least 400- 450 Liters
Temperature System should have Programmable operating 1 65,000 65,000
Freezer temperature from -50°C up to -86°C with 1°C
increment.
It should be an energy efficient system with power
consumption ≤ 11 kWh per day under standard test
conditions & noise level

Lab Furniture

1 Central 1 30,000 30,000


Bench

2 Wooden 2 15,000 32,000


Computer
Table
3 Adjustable 15 2,500 37,500
laboratory
chair
4 Desktop 5 900 45,000
computer
Total Grand 4,131432.