Professional Documents
Culture Documents
Ziehm C-Arm - Center Beam Adjustment
Ziehm C-Arm - Center Beam Adjustment
After the generator has been mounted to the system and is placed into operation, the center beam
needs to be adjusted, before the collimator is installed.
In order to perform this adjustment, a center beam test device is required (91116).
Aditionally a pre collimator should be used for radiation protection during adjustment.(65699)
Adjust_Center_Beam.fm -1
17. April 2012, 12:08 DRAFT
1.1.2 Setup for center beam adjustment
In order to do the center beam adjustment, the center beam test device has to be put on the gen-
erators output window. Insert the pre collimator first, in order to provide radiation protection (beam
will be limited)
Note
Horizontal and vertical image reversal (mirroring) and Image rotation must be inac-
tive in order to provide a correct readout of the image on the screen.
• First: find the correct position of the center beam test device
• Second: adjust the generator position to get the image of the center beam test device into the
center of the screen
-2 Adjust_Center_Beam.fm
17. April 2012, 12:08 DRAFT
1.1.3.1 Correct position of the center beam test device.
By shifting the position of the center beam test device, a concentric shape of the center beam test
device has to appear on the screen. This concentric shape may appear uncentered to the screen.
Warning
Center beam adjustment is done while screening.
Take care for radiation protection.
wrong correct
Fig. 1-4 wrong and correct position of the center beam test device
Adjust_Center_Beam.fm -3
17. April 2012, 12:08 DRAFT
1.1.3.2 Adjusting the generator position
-4 Adjust_Center_Beam.fm
17. April 2012, 12:08 DRAFT
Warning
Center beam adjustment is done while screening.
Take care for radiation protection.
Adjust_Center_Beam.fm -5
17. April 2012, 12:08 DRAFT
1.1.3.2.2 Adjusting Generator attachment
If there is a top down deviation, this is adjusted by turning the generator attachment.
In order to to this, the fixation bolts have to be losened. (8mm allen key required)
Then the generator can be adjusted in a small range.
Fix the bolts tightly when the correct position is found (46Nm) .
Warning
Center beam adjustment is done while screening.
Take care for radiation protection.
fixation bolts
In the image above generator is removed in order to make the attachment visible.
For adjustment, the generator is not removed.
Center beam adjustment is finished if the concentric shape appears centered to the screen.
-6 Adjust_Center_Beam.fm
17. April 2012, 12:08 DRAFT