Professional Documents
Culture Documents
Evaluation and Routine Testing in Medical Imaging Departments
Evaluation and Routine Testing in Medical Imaging Departments
Evaluation and Routine Testing in Medical Imaging Departments
1:
1994
IEC 1223-2-1:
1993
The preparation of this British Standard was entrusted by the Health Care
Standards Policy Committee (HCC/-) to Technical Committee HCC/73, upon
which the following bodies were represented:
Contents
Page
Committees responsible Inside front cover
National foreword ii
Introduction 1
1 Scope and object 1
2 References to other publications 1
3 Terminology 1
4 General aspects of constancy tests 2
5 Test procedure 3
6 Action to be taken 4
7 Statement of compliance 5
Annex A Terminology — Index of terms 8
Annex B Example of a form for the standardized test report 9
Annex C Guidance on action to be taken 14
Annex D Rationale 14
Annex E Procedures for initial setting up of the film processor 16
Figure 1 — Variation in film base plus fog density, speed index and
contrast index with developer temperature (optimization procedure) 6
Figure 2 — Example of a control chart displaying the film base plus
fog density, speed index and contrast index as a function of time 7
List of references Inside back cover
© BSI 11-1999 i
BS 7725-2.1:1994
National foreword
This Section of BS 7725 has been prepared under the direction of the Health Care
Standards Policy Committee. It is identical with IEC 1223-2-1:1993 Evaluation
and routine testing in medical imaging departments — Part 2-1: Constancy
tests — Film processors, prepared by the Technical Committee 62 Electrical
equipment in medical practice and published by the International
Electrotechnical Commission (IEC).
Cross-references
International Standard Corresponding British Standard
IEC 601-1:1988 BS 5724 Medical electrical equipment
(without amendment Part 1:1989 General requirements for safety
No. 1, 1991) (Identical)
IEC 788:1984 BS 6641:1985 Glossary of medical radiological terms
(Identical)
IEC 1223-1:1993 BS 7725 Evaluation and routine testing in medical
imaging departments
Part 1:1994 General
(Identical)
IEC 1223-2-3:1993 Section 2.3:1994 Method for darkroom conditions
(Identical)
NOTE ISO 4090:1991, ISO 8374:1986 and ISO 8402:1986 referenced in clause 2 are not quoted in
this Technical Report. This error has been reported to the relevant body for action.
A British Standard does not purport to include all the necessary provisions of a
contract. Users of British Standards are responsible for their correct application.
Compliance with a British Standard does not of itself confer immunity
from legal obligations.
Summary of pages
This document comprises a front cover, an inside front cover, pages i and ii,
pages 1 to 18, an inside back cover and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on the
inside front cover.
ii © BSI 11-1999
BS 7725-2.1:1994
© BSI 11-1999 1
BS 7725-2.1:1994
2 © BSI 11-1999
BS 7725-2.1:1994
© BSI 11-1999 3
BS 7725-2.1:1994
The test results and conditions should be recorded 5.3 Hypo retention test*
in a uniform manner, examples of which are given in In order to ensure that radiographic materials are
Figure 2 and Annex B. fixed and washed adequately, a hypo retention test
5.1.3 ESTABLISHED CRITERIA should be performed at least annually, as follows.
The performance of the FILM PROCESSOR is of a) An unexposed control film is processed, and
adequate constancy if variations in the constancy then a few drops of hypo test solution are placed
parameters are within the following limits: on the RADIOGRAM. This normally produces a
a) the FILM BASE PLUS FOG DENSITY shall be stain on the RADIOGRAM ranging in colour from
within ± 0,05 of the BASELINE VALUE; pale yellow to dark brown.
b) the SPEED INDEX shall be within ± 0,15 of the b) The hue of the coloured stain is compared
BASELINE VALUE, and should be within ± 0,10 of against a standard estimator. This shall be
this value. within the limit specified in the ACCOMPANYING
DOCUMENTS for the film used, otherwise
5.1.4 Further action corrective action is necessary.
If any of the constancy parameters are outside the *NOTE The standard estimator may be obtained from the film
ESTABLISHED CRITERIA described in 5.1.3 the manufacturer. The hypo test solution may also be obtained from
CONSTANCY TEST should be repeated. the film manufacturer or can be made up as described in
Annex C.
If any of the constancy parameters still remain
outside these limits, or if the results of the last three 6 Action to be taken
CONSTANCY TESTS indicate a tendency for any index
The control charts according to 5.1.2, and records of
to consistently decrease or increase in value, then
test results according to Annex B, will clearly show
appropriate action should be taken as described in
any trend in the values of the constancy parameters
clause 6.
with time, and will indicate any tendency to exceed
5.1.5 Frequency of CONSTANCY TEST the ESTABLISHED CRITERIA described in 5.1.3.
a) Once per day at the same time, if use of the Should such a trend occur, or should the value of any
FILM PROCESSOR is intense; constancy parameter fall outside the limits
b) prior to use, if the FILM PROCESSOR has not described in 5.1.3, the following checks should be
been used for at least 24 h; carried out in order to determine what corrective
action should be taken to halt the deterioration in
c) each time the processing solutions are changed;
processing conditions:
d) after maintenance or repair of the FILM
PROCESSOR.
a) The temperature of the development solution
is measured.
5.2 Consistency of control films
b) The replenishment rates are measured.
When the current batch of control films is about to c) The radiographic materials currently being
become exhausted, a replacement batch of control processed are checked for changes in batch
films should be introduced in accordance with the numbers, types and sizes.
following procedure.
d) The replenishment solutions are checked for
a) At least three control films from the current correct preparation.
batch, together with at least three control films
from the replacement batch, are exposed as e) The expiry dates of control film and the
described in 5.1.1. radiographic materials currently in use in the
X-ray department are checked.
b) The control films are processed and evaluated
as described in 5.1.1. f) The wash water flow rate is measured.
c) The average values of the constancy g) The wash water temperature is measured.
parameters for the two batches of control films h) The operation of the recirculatory system is
are compared. The differences in these average checked.
values should be used to establish the new i) The developer filters are inspected.
BASELINE VALUES for the constancy parameters.
j) The expiry dates of the chemicals in use are
d) If the difference in any single constancy checked, if applicable.
parameter is outside the limits described in 5.1.3,
an investigation should be carried out into the
reason for the change in film characteristics. This
investigation is beyond the scope of this technical
report.
4 © BSI 11-1999
BS 7725-2.1:1994
k) The maintenance log-book of the FILM A record should be kept in the maintenance log-book
PROCESSOR is checked to confirm that the of all the faults discovered, the actions taken, and
established routine maintenance schedule has the effect of this action on the constancy parameters
been correctly implemented. of the FILM PROCESSOr.
l) The maintenance log-book is checked to identify See also Annex E.
the maintenance or checking operations applied
to the FILM PROCESSOR immediately prior to the 7 Statement of compliance
departure from constancy. This may provide
The test report should be headed:
some indication of the possible cause.
Test report
m) The duration of the processing cycle is on constancy test of equipment for film
checked. processing according to IEC 1223-2-1:1993
If any of these checks indicate the possible cause of
the deviation in processor performance, appropriate
remedial action should be taken, and a further two
control films should be used to evaluate the
effectiveness of the remedial action. If this action
fails to restore the performance of the FILM
PROCESSOR to within the limits described in 5.1.3, a
replacement of all the processing and replenishment
solution may be undertaken in a new initial set-up
of the FILM PROCESSOR, following the procedures
described in Annex E.
© BSI 11-1999 5
BS 7725-2.1:1994
Figure 1 — Variation in FILM BASE PLUS FOG DENSITY, SPEED INDEX and CONTRAST INDEX with
developer temperature (optimization procedure)
6 © BSI 11-1999
BS 7725-2.1:1994
Figure 2 — Example of a control chart displaying the FILM BASE PLUS FOG DENSITY,
SPEED INDEX and CONTRAST INDEX as a function of time
© BSI 11-1999 7
BS 7725-2.1:1994
8 © BSI 11-1999
BS 7725-2.1:1994
Identification
History of tests
STATUS TEST Date:
latest test on darkroom conditions Date:
latest test on film-processing equipment Date:
latest initial CONSTANCY TEST Date:
previous CONSTANCY TEST Date:
Frequency of testing
Identity of instrumentation
* Thermometer
* Sensitometer
* Densitometer
© BSI 11-1999 9
BS 7725-2.1:1994
— Concentration
w Temperature at point 1 °C °C °C °C
w Temperature at point 2 °C °C °C °C
* Fixing solution
— Type
— Rate of regeneration
w Temperature at point 1 °C °C °C °C
w Temperature at point 2 °C °C °C °C
w Difference in temperature °C °C °C °C
— Concentration
— Rate of regeneration
10 © BSI 11-1999
BS 7725-2.1:1994
w Temperature at point 1 °C °C °C °C
w Temperature at point 2 °C °C °C °C
* Starter solution
— Type
— Quantity
* Tanning agent
— Type
— Quantity
* Wash water
— Flow rate see 6 f)
— Identification of point 1 for
measuring
w Temperature at point 1 °C °C °C °C
w Temperature at point 2 °C °C °C °C
w Difference in temperature °C °C °C °C
Radiographic/photographic
material
* Identification of a list of materials
taken into account see 6 c) and 4.2.1
* Control film 1
w Material(s) chosen as control film see 4.2.1
w Type
w Speed
w Emulsion number
© BSI 11-1999 11
BS 7725-2.1:1994
w Identity of batch
* Control film 2
w Material(s) chosen as control film see 4.2.1
w Type
w Speed
w Emulsion number
w Identity of batch
w Expiry date
w Sheet 1
w Sheet 2
w Sheet 3
w Sheet 4
w Sheet 5
w Sheet 6
12 © BSI 11-1999
BS 7725-2.1:1994
— SPEED INDEX
w Sheet 1
w Sheet 2
w Sheet 3
w Sheet 4
w Sheet 5
w Sheet 6
— CONTRAST INDEX
w Sheet 1
w Sheet 2
w Sheet 3
w Sheet 4
w Sheet 5
w Sheet 6
— SPEED INDEX
— CONTRAST INDEX
* Measured duration s s s s
© BSI 11-1999 13
BS 7725-2.1:1994
Final decision:
— Performance of FILM PROCESSOR
satisfactory see 5.1.3 YES/NO YES/NO YES/NO YES/NO
14 © BSI 11-1999
BS 7725-2.1:1994
© BSI 11-1999 15
BS 7725-2.1:1994
Harmonization is usually achieved by adjusting the m) The time taken for the leading edge of a film
temperature of the development solution in each to enter and leave the FILM PROCESSOR is
FILM PROCESSOR to correspond to a common SPEED measured. The time of the processing cycle is
INDEX, although this may result in one or more of checked to be within the limits specified in the
the FILM PROCESSORS operating under suboptimum ACCOMPANYING DOCUMENTS.
conditions. n) *A series of “clean-up” films is passed through
the FILM PROCESSOR to check that no detrimental
Annex E Procedures for initial setting scratches and roller or pressure marks are being
up of the FILM PROCESSOR produced.
E.1 Initial setting up NOTE “Clean-up” films are exposed unprocessed films
which are processed daily and immediately after the
a) The FILM PROCESSOR is cleaned according to processing solutions have been changed.
the instructions given in the ACCOMPANYING p) Before any further measurements of the
DOCUMENTS. constancy parameters of the FILM PROCESSOR are
b) The developer and fixer tanks are drained. undertaken, the FILM PROCESSOR should have
The tanks are flushed. Racks are removed and been in use for the development of the equivalent
washed with fresh water. of at least 25 conventional RADIOGRAMS of
c) The developer recirculation filter, if fitted, is size 35 cm × 43 cm.
replaced with a new filter. E.2 Determination of optimum temperature of
d) The replenisher tanks and hoses are drained the development solution
and flushed with fresh water. E.2.1 Method
e) Fresh replenishment solutions are prepared in a) The temperature of the development solution
the replenisher tanks in accordance with the is set to a value 5 °C below the mid-point value of
manufacturer’s instructions. the range of temperature recommended in the
ACCOMPANYING DOCUMENTS. The temperature is
f) The replenishment systems are operated to
waste for a short time to ensure that all the checked by means of the thermometer and
water is flushed out of the replenisher lines. The adjusted if necessary. The temperature of the
correct functioning of the replenisher pumps is development solution is allowed adequate time to
checked. stabilize.
g) The fixer tank is filled to the correct level with b) Using the sensitometer, a series of six control
fresh fixer and the fixer rack inserted. films is exposed. It is sufficient to expose only one
side of a DOUBLE EMULSION FILM.
h) The developer tank is filled to the correct level
with fresh developer and.the correct amount of c) *After at least 30 min but no longer than 4 h,
the starter solution is added as specified in the the control films are processed.
ACCOMPANYING DOCUMENTS. The control films are processed so that the
j) The temperature of the water supplied leading edge is the least exposed end of the
continuously to the FILM PROCESSOR is set to the stepwedge pattern.
correct value. The correct wash water flow rates The sheets of control film are passed into the
are checked, as is the correct functioning of the FILM PROCESSOR always at the same side of the
recirculation system. The FILM PROCESSOR is feeding slit.
switched on for approximately 30 min and d) The optical densities are measured in the three
allowed to attain its working temperature. areas of each control film; see 4.2.3. This enables
k) The temperatures of the developer, fixer, and the constancy parameters (FILM BASE PLUS FOG
wash water are measured at each end of each DENSITY, SPEED INDEX, and CONTRAST INDEX) to
bath. If necessary, these are adjusted to be determined.
within ± 0,2 °C of the values recommended in the e) The procedure described in items a) to d)
ACCOMPANYING DOCUMENTS. The measured of E.2.1 is repeated a further 10 times, but with
temperatures of the processing solutions are the temperature of the development solution
checked against those indicated by the increased by 1 °C on each occasion.
thermometers of the FILM PROCESSOR, and any
differences recorded.
l) The replenishment rates are measured and, if
necessary, adjusted in accordance with the data
given in the ACCOMPANYING DOCUMENTS.
16 © BSI 11-1999
BS 7725-2.1:1994
© BSI 11-1999 17
18 blank
BS 7725-2.1:1994
List of references
© BSI 11-1999
BS 7725-2.1:
1994
IEC 1223-2-1: BSI — British Standards Institution
1993
BSI is the independent national body responsible for preparing
British Standards. It presents the UK view on standards in Europe and at the
international level. It is incorporated by Royal Charter.
Revisions
It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: 020 8996 9000. Fax: 020 8996 7400.
BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.
Buying standards
Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.
Information on standards
Copyright
Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written
permission from BSI.
This does not preclude the free use, in the course of implementing the standard,
of necessary details such as symbols, and size, type or grade designations. If these
details are to be used for any other purpose than implementation then the prior
written permission of BSI must be obtained.