Professional Documents
Culture Documents
Evaluation and Routine Testing in Medical Imaging Departments
Evaluation and Routine Testing in Medical Imaging Departments
2:
1994
IEC 1223-2-2:
1993
The preparation of this British Standard was entrusted by the Health Care
Standards Policy Committee (HCC/-) to Technical Committee HCC/73, upon
which the following bodies were represented:
© BSI 10-1999
Amd. No. Date Comments
Contents
Page
Committees responsible Inside front cover
National foreword ii
Introduction 1
1 Scope and object 1
2 References to other publications 1
3 Terminology 1
4 General condition of radiographic cassettes, intensifying
screens and intensifying screens in film changers 2
5 Testing the constancy of adequate contact in film-screen
combinations 2
6 Testing the relative sensitivity of intensifying screen(s) and
associated radiographic cassettes 3
7 Test report 5
Annex A Terminology — Index of terms 6
Annex B Example of a form for the standardized test report 8
Annex C Guidance on action to be taken 10
Annex D Rationale 11
List of references Inside back cover
© BSI 10-1999 i
BS 7725-2.2:1994
National foreword
This Section of BS 7725 has been prepared under the direction of the Health Care
Standards Policy Committee. It is identical with IEC Publication 1223-2-2:1993
Evaluation and routine testing in medical imaging departments —
Part 2-2: Constancy tests — Radiographic cassettes and film changers —
Film-screen contact and relative sensitivity of the screen-cassette assembly,
prepared by the Technical Committee 62 Electrical equipment in medical practice
and published by the International Electrotechnical Commission (IEC).
Cross-references
International Standard Corresponding British Standard
IEC 788:1984 BS 6641:1985 Glossary of medical radiology terms
(Identical)
BS 7725 Evaluation and routine testing in medical
imaging departments
IEC 1223-1:1993 Part 1:1994 General
(Identical)
IEC 1223-2-1:1993 Section 2.1:1994 Film processors
(Identical)
IEC 1223-2-3:1994 Section 2.3:1994 Method for darkroom safelight
conditions
(Identical)
A British Standard does not purport to include all the necessary provisions of a
contract. Users of British Standards are responsible for their correct application.
Compliance with a British Standard does not of itself confer immunity
from legal obligations.
Summary of pages
This document comprises a front cover, an inside front cover, pages i and ii,
pages 1 to 12, an inside back cover and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on the
inside front cover.
ii © BSI 10-1999
BS 7725-2.2:1994
IEC 788:1984, Medical radiology — Terminology. luminous surface in combination with additional
provisions for the observation of transparent objects
IEC 1223-1:1993, Evaluation and routine testing in
such as RADIOGRAMS on RADIOGRAPHIC FILM
medical imaging departments — Part 1: General
aspects.
IEC 1223-2-1:1993, Evaluation and routine testing
in medical imaging departments —
Part 2-1: Constancy tests — Film processors.
© BSI 10-1999 1
BS 7725-2.2:1994
4 General condition of RADIOGRAPHIC • regions where the protective coat has been
CASSETTES, INTENSIFYING SCREENS and worn away.
INTENSIFYING SCREENS in FILM
5 Testing the constancy of adequate
CHANGERS*
contact in film-screen combinations
Any relevant instructions given in the
ACCOMPANYING DOCUMENTS of the RADIOGRAPHIC
5.1 Summary
CASSETTES, FILM CHANGER, and the INTENSIFYING The adequacy of the contact between the
SCREEN(S) should be observed when examining the RADIOGRAPHIC FILM and INTENSIFYING SCREENS is
devices. determined by viewing the uniformity in blackening
The following measures should be taken: of the DIRECT RADIOGRAM, and the clarity of the
image of a standard TEST DEVICE placed in front of
— Each single INTENSIFYING SCREEN or each pair
the RADIOGRAPHIC CASSETTE or FILM CHANGER to be
of INTENSIFYING SCREENS is given an individual
tested.
identification such as a number;
5.2 Test equipment
• this identification is written with permanent
ink in an unobtrusive position on one 5.2.1 RADIOGRAPHIC FILM
INTENSIFYING SCREEN, so that it will be The RADIOGRAPHIC FILM used for the test should be
recorded on the film; typical of that used with the INTENSIFYING SCREENS
*NOTE Attention has to be paid to the fact that and associated RADIOGRAPHIC CASSETTES or FILM
permanent ink may affect or destroy the protective layer of CHANGERS to be tested.
INTENSIFYING SCREENS.
• this identification is indicated on the outside For successive tests, RADIOGRAPHIC FILM of the same
of the RADIOGRAPHIC CASSETTE. type should be used.
— Each RADIOGRAPHIC CASSETTE and FILM 5.2.2 Standard TEST DEVICE
CHANGER should be labelled with The TEST DEVICE required for this test is a metal
• name of manufacturer of INTENSIFYING wire mesh of linear dimensions not smaller than the
SCREEN(S) contained; IMAGE RECEPTION AREA of the RADIOGRAPHIC
• type(s) of INTENSIFYING SCREEN(S) contained; CASSETTE or FILM CHANGER to be tested.
• date of acquisition of the INTENSIFYING *The wire mesh should have a square pattern of
SCREEN(S). such a quality that its structure appears uniform
when viewed on a FILM ILLUMINATOR.
The general condition of all RADIOGRAPHIC
CASSETTES and the INTENSIFYING SCREENS should be To facilitate the measurement of the optical density
checked as follows, at least every six months: of the RADIOGRAM obtained during the test, the
standard TEST DEVICE should have an opening in the
1) The exterior and interior of each cassette is centre of the wire mesh of 1 cm square or 1 cm in
examined for
diameter.
• correct labelling;
*The wire should be made from metal composed
• cleanliness; predominantly of elements with an atomic number
• signs of warping and fatigue of the material of at least 26, for example copper or zinc.
provided to ensure the contact between The diameter of the wire should be
RADIOGRAPHIC FILM and INTENSIFYING SCREEN. approximately 0,5 mm and the separation between
2) The hinge assembly and closure mechanism of the centres of adjacent wires should be
the cassette are examined for wear and damage. approximately six diameters.
3) Each INTENSIFYING SCREEN is examined for 5.2.3 FILM PROCESSOR
• cleanliness; The FILM PROCESSOR used for the test should have
• worn and stained areas; and been tested and found satisfactory, according to
IEC 1223-2-1.
• *as appropriate, any screen or pair of screens
is cleaned or replaced, and the date recorded. 5.2.4 Densitometer
NOTE It is generally undesirable to replace only one of a Optical densities are measured with a densitometer
pair of INTENSIFYING SCREENS. which reads consistently within ± 0,02.
4) *Taking appropriate precautions to protect the
5.2.5 FILM ILLUMINATOR
eyes of the examiner, INTENSIFYING SCREENS are
examined under ultra-violet light to reveal The FILM ILLUMINATOR used for the test should have
• artefacts; been tested and found satisfactory, according to
IEC 1223-2-12.
2 © BSI 10-1999
BS 7725-2.2:1994
5.3 Criterion for satisfaction* Rapid type FILM CHANGERS should be tested in the
The contact between RADIOGRAPHIC FILM and static and dynamic mode, by making:
INTENSIFYING SCREEN(S) is satisfactory if the optical — two RADIOGRAMS in static mode, and
density of the image of the standard TEST DEVICE on — a series of at least six RADIOGRAMS in dynamic
the RADIOGRAM appears uniform. mode at the highest frequency normally used in
5.4 Test procedure the department.
1) Prior to starting any constancy check, the Other FILM CHANGERS are tested in the static mode.
identity is established of any device such as a Equipment identification, test conditions and other
RADIOGRAPHIC CASSETTE, INTENSIFYING SCREEN, pertinent information shall be included in the test
FILM CHANGER, or ACCESSORY used for the test, report (see clause 7).
and having properties that can influence the 5.5 Evaluation
result of the test.
The RADIOGRAMS are viewed on the FILM
2) The conditions of the RADIOGRAPHIC CASSETTE
ILLUMINATOR from a distance of 2 m to 3 m.
and INTENSIFYING SCREENS as described in
clause 4 are checked, and any action necessary If regions of the image of the TEST DEVICE on the
before commencing the test is taken. RADIOGRAM appear dark or non-uniform, or details
of the image are blurred, the film-screen contact is
3) *If shown worthwhile, any ANTI-SCATTER GRID poor, and will impair the quality of the recorded
present is removed from the equipment used for
diagnostic information.
the test and from a FILM CHANGER to be tested.
5.6 Further action
4) The RADIOGRAPHIC CASSETTE or FILM CHANGER
is loaded with the type of RADIOGRAPHIC FILM If the image of the TEST DEVICE shows the defects
selected according to 5.2.1. described in 5.5, consideration should be given to
5) *Unless otherwise chosen for the test, the test the repair or replacement of screen-cassette
is not continued until at least 15 min after assemblies showing poor film-screen contact.
loading the RADIOGRAPHIC CASSETTE. 5.7 Frequency of testing
6) The standard TEST DEVICE is placed flat on the Adequate, homogeneous contact between the
outside of, and in contact with, the incident face RADIOGRAPHIC FILM and INTENSIFYING SCREEN(S)
of the RADIOGRAPHIC CASSETTE, or when testing a should be tested:
FILM CHANGER, the TEST DEVICE is placed on the — initially on acquisition; and then
PATIENT SUPPORT or another suitable support
immediately in front of the FILM CHANGER to — at least annually; or
avoid contact with the surface of the FILM — when poor film-screen contact is suspected.
CHANGER.
7) The FOCAL SPOT TO IMAGE RECEPTOR DISTANCE
6 Testing the relative sensitivity of
is set to the distance used in normal clinical INTENSIFYING SCREEN(S) and associated
practice. RADIOGRAPHIC CASSETTES*
8) The dimensions of the X-RAY FIELD are adjusted 6.1 Summary
to the IMAGE RECEPTION AREA, and the The relative sensitivity of screen-cassette
RADIOGRAPHIC CASSETTE or FILM CHANGER is
assemblies is determined by comparing the optical
aligned with the X-RAY BEAM. density produced by simultaneously and equally
9) The individual RADIOGRAPHIC CASSETTE or irradiated identical films placed in each of a group
FILM CHANGER is identified with lead numbers. of three test and one reference RADIOGRAPHIC
10) The standard TEST DEVICE is irradiated, using CASSETTES. The process is repeated for all
the smallest available FOCAL SPOT and LOADING RADIOGRAPHIC CASSETTES to be tested, using the
FACTORS that will produce an optical density in same reference screen-cassette assembly for all the
the processed RADIOGRAM of about 2 above FILM groups.
BASE PLUS FOG DENSITY in the image of the *The screen-cassette assemblies are then sorted
opening in the wire mesh. into groups that produce similar optical densities,
11) In general, for testing RADIOGRAPHIC and therefore are of similar sensitivity.
CASSETTES and FILM CHANGERS, an X-RAY TUBE
VOLTAGE of not more than 70 kV is used.
© BSI 10-1999 3
BS 7725-2.2:1994
4 © BSI 10-1999
BS 7725-2.2:1994
© BSI 10-1999 5
BS 7725-2.2:1994
ABSORPTION rm-12-05–
ACCEPTANCE TEST AG-3.2.4
ACCESSORY rm-83-06
ACCOMPANYING DOCUMENTS rm-82-01
ANTI-SCATTER GRID rm-32-06
AUTOMATIC CONTROL SYSTEM rm-36-45
CONSTANCY TEST AG-3.2.6
CURRENT TIME PRODUCT rm-36-13
DENTAL PANORAMIC TOMOGRAPHY rm-41-12
DIRECT RADIOGRAM rm-32-03
DIRECT RADIOGRAPHY rm-41-07
ELECTRO-OPTICAL X-RAY IMAGE INTENSIFIER rm-32-40
ESTABLISHED CRITERIA AG-3.2.8
FILM BASE PLUS FOG DENSITY 1-3.2.2
FILM CHANGER rm-31-07
FILM ILLUMINATOR 2-3.2.1
FILM PROCESSOR 1-3.2.1
FILTRATION rm-12-11
FOCAL SPOT rm-20-13s
FOCAL SPOT TO IMAGE RECEPTOR DISTANCE rm-37-13
IMAGE RECEPTION AREA rm-37-16
INTENSIFYING SCREEN rm-32-38
IRRADIATION TIME rm-36-11
LOADING FACTOR rm-36-01
LOADING TIME rm-36-10
MEDICAL DIAGNOSTIC RADIOLOGY rm-40-04
OPERATOR rm-85-02
PATIENT rm-62-03
PATIENT SUPPORT rm-30-02
QUALITY ASSURANCE AG-3.2.1
6 © BSI 10-1999
BS 7725-2.2:1994
© BSI 10-1999 7
BS 7725-2.2:1994
8 © BSI 10-1999
BS 7725-2.2:1994
*Back screen:
— Cleanliness according to item 1) of clause 4 satisfactory YES/NO YES/NO YES/NO YES/NO
— Absence of defects according to item 3) of clause 4 YES/NO YES/NO YES/NO YES/NO
— Absence of defects according to item 4) of clause 4 YES/NO YES/NO YES/NO YES/NO
RADIOGRAPHIC FILM:
— Type of film according to 5.2.1
— Package or batch identification according to 5.2.1
Identification of screen-cassette assembly
TEST DEVICE:
— Identification of TEST DEVICE according to 5.2.2
Densitometer:
— Identification of densitometer according to 5.2.4
Identification of X-RAY EQUIPMENT:
— X-RAY GENERATOR
— X-RAY SOURCE ASSEMBLY
— FOCAL SPOT
— FILTRATION
— Other layers in the X-RAY BEAM
*LOADING FACTORS:
— X-RAY TUBE VOLTAGE
— X-RAY TUBE CURRENT, and
— LOADING TIME, or
CURRENT TIME PRODUCT, or
Selection of AUTOMATIC CONTROL SYSTEM
Geometrical conditions:
— FOCAL SPOT TO IMAGE RECEPTOR DISTANCE
— .............
— .............
Identification of FILM ILLUMINATOR
Test results:
— Value of density measured
*Decisions made:
— New association to a group of screen-cassette assemblies YES/NO YES/NO YES/NO YES/NO
— Designation of new group
— Mechanical state and visual inspection satisfactory YES/NO YES/NO YES/NO YES/NO
— Contact film-screen satisfactory YES/NO YES/NO YES/NO YES/NO
— Relative sensitivity satisfactory YES/NO YES/NO YES/NO YES/NO
— ............. YES/NO YES/NO YES/NO YES/NO
— ............. YES/NO YES/NO YES/NO YES/NO
— .............
— .............
— .............
*Final decision:
— Screen-cassette assembly passed the test YES/NO YES/NO YES/NO YES/NO
Labelling of date of next CONSTANCY TEST
© BSI 10-1999 9
BS 7725-2.2:1994
10 © BSI 10-1999
BS 7725-2.2:1994
Annex D Rationale
Concerning 4 General condition
The performance of RADIOGRAPHIC CASSETTES and INTENSIFYING SCREENS may change with handling and
age in normal use.
Worn or damaged screen-cassette assemblies may cause unsharp and fogged RADIOGRAMS, thus requiring
that cassettes or screens be repaired or replaced.
Inadequate film-screen contact, dirt, and other causes of impaired performance may also be present in
FILM CHANGERS and other systems in DIRECT RADIOGRAPHY, not using RADIOGRAPHIC CASSETTES.
Therefore, screen-cassette assemblies and other systems for DIRECT RADIOGRAPHY need to be inspected at
least annually.
In order to facilitate survey of the performance of the single components involved, they should be
individually identified, and their age and date of commissioning should be known.
In order to trace an individual INTENSIFYING SCREEN that might have caused a recurring film artefact, the
individual identification of the INTENSIFYING SCREENS or pairs of screens is extremely important.
If the identification with permanent ink presents the risk of damaging the INTENSIFYING SCREEN, other
ways of identification, transferable to the RADIOGRAM obtained, must be used, such as the individual
production number, if provided.
A careful inspection of INTENSIFYING SCREENS should pay attention to:
— worn areas;
— stained areas;
— dirty specks;
which may produce artefacts on RADIOGRAMS; and to yellowing due to age.
Within the diagnostic X-ray department an inventory of the screen-cassette assemblies and other systems
for DIRECT RADIOGRAPHY should be maintained, listing as appropriate:
— individual identification;
— type of the RADIOGRAPHIC CASSETTE or image reception system;
— date of acquisition of the RADIOGRAPHIC CASSETTE;
— individual identification of the INTENSIFYING SCREENS;
— types of INTENSIFYING SCREENS;
— sensitivity of a screen-cassette assembly, or its assignment to a group of similar sensitivity.
Concerning 4 General condition — item 3)
Cleaning should be done using only methods recommended in the ACCOMPANYING DOCUMENTS of the
INTENSIFYING SCREEN(S).
The necessary frequency of cleaning INTENSIFYING SCREENS will depend upon the extent and conditions of
use.
Concerning 4 General condition — item 4)
Special care should be taken in work with ultra-violet light.
Light sources for inspection of INTENSIFYING SCREENS under ultra-violet light are commercially available.
A description of their use is given in:
B.L. Diffey and W.E. Mallion
“An apparatus for the routine inspection of intensifying screens”,
Radiography 44, 223 (1978)
Concerning 5.2.2 Standard TEST DEVICE
As ferromagnetic materials may affect the function of certain electromagnetic equipment, for example,
ELECTRO-OPTICAL X-RAY IMAGE INTENSIFIERS, such materials should not be used for these TEST DEVICES.
© BSI 10-1999 11
BS 7725-2.2:1994
12 © BSI 10-1999
BS 7725-2.2:1994
List of references
© BSI 10-1999
BS 7725-2.2:
1994
IEC 1223-2-2: BSI — British Standards Institution
1993
BSI is the independent national body responsible for preparing
British Standards. It presents the UK view on standards in Europe and at the
international level. It is incorporated by Royal Charter.
Revisions
It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: 020 8996 9000. Fax: 020 8996 7400.
BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.
Buying standards
Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.
Information on standards
Copyright
Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written
permission from BSI.
This does not preclude the free use, in the course of implementing the standard,
of necessary details such as symbols, and size, type or grade designations. If these
details are to be used for any other purpose than implementation then the prior
written permission of BSI must be obtained.