Professional Documents
Culture Documents
Operating Manual
Operating Manual
Operating Manual
Hemodialysis Machine
Operating Instructions
0123
Part No.: M32 327 1 Fresenius Medical Care 4008 S 11/02.01 (OP) 0-1
0-2 Fresenius Medical Care 4008 S 11/02.01 (OP)
OPERATING INSTRUCTIONS
HEMODIALYSIS MACHINE 4008 S
These Operating Instructions are part of the machine documentation and thus an integral part of the 4008 S
hemodialysis machine. They contain any information necessary for the use of the 4008 S hemodialysis
machine. Before operating the machine, the operator must have been instructed by the manufacturer on
how to use the machine and must be thoroughly familiar with the contents of the Operating Instructions.
The machine may only be operated by persons who have been instructed on the proper operation and
handling of the unit.
The Operating Instructions must be carefully studied before attempting to operate the machine.
The Brief Operating Instructions are intended as a reference only and do not replace the Operating
Instructions manual with which the operator must be fully familiar.
☞ Note
Informs the operator that in case of a
failure to follow the steps as de-
scribed, a specific function will be
executed incorrectly or will not be
executed at all, or will not produce
the desired effect.
Caution
Advises the operator against certain
procedures or actions that could
cause damage to the equipment or
may have adverse effects on opera-
tors and patients.
The parameters entered must be verified by the operator, i.e. the operator must check, if the values entered
are correct. Should the desired value deviate from the parameters displayed, the setting must be corrected
before activating the function.
The actual values displayed must be compared with the desired values specified!
The 4008 S hemodialysis machine reflects the latest state of technology and complies with the regulations
of EN 60601-1 (IEC 601-1).
Assembly, extensions, adjustments, modifications or repairs may only be carried out by the manufacturer
or persons authorized by him.
Local Service:
Chapter Page
2 Preparation .................................................................................................................... 2-
2.1 Dialysis Machine ............................................................................................................. 2-3
2.2 Functional Check ............................................................................................................ 2-12
2.3 Blood Pump .................................................................................................................... 2-26
2.4 Heparin Pump ................................................................................................................. 2-28
2.5 Air Detector ..................................................................................................................... 2-32
2.6 Extracorporeal Blood Circuit ........................................................................................... 2-34
2.7 Priming the Extracorporeal Blood Circuit With a Saline Bag ......................................... 2-36
2.8 Dialysate Side ................................................................................................................. 2-41
3 Treatment ....................................................................................................................... 3-
3.1 Ultrafiltration .................................................................................................................... 3-3
3.2 Double Needle Dialysis ................................................................................................... 3-11
3.3 Single-Needle Dialysis .................................................................................................... 3-18
3.4 Single-Needle Click-Clack .............................................................................................. 3-23
3.5 Isolated Ultrafiltration ISO UF ......................................................................................... 3-25
3.6 Sodium and UF Profiles .................................................................................................. 3-30
3.7 DIASAFE ®plus (option) ..................................................................................................... 3-39
5 Disconnection ............................................................................................................... 5-
5.1 End of Treatment ............................................................................................................ 5-3
5.2 Reinfusion ....................................................................................................................... 5-6
5.3 Stopping Treatment ........................................................................................................ 5-11
5.4 Emptying the bibag® (option) .......................................................................................... 5-18
5.5 Emptying the Dialyzer ..................................................................................................... 5-20
6 Cleaning ......................................................................................................................... 6-
6.1 General Overview ........................................................................................................... 6-3
6.2 External Cleaning ........................................................................................................... 6-5
6.3 Basic Conditions for Cleaning Programs ........................................................................ 6-5
6.4 Disinfection, Decalcification ............................................................................................ 6-6
6.5 Cleaning / Degreasing Program ..................................................................................... 6-11
6.6 Hot Rinse ........................................................................................................................ 6-15
6.7 Rinsing ............................................................................................................................ 6-18
6.8 Priming During Cleaning Programs ................................................................................ 6-22
6.9 Precirculation During the Cleaning Programs ................................................................. 6-27
Chapter Page
The 4008 H dialysis machine is designed for performing chronic and acute hemodialysis. It can be
used in home dialysis, limited care centers and clinical hemodialysis.
Dialysis therapy occasionally causes hypotension, nausea, vomiting and cramping in some
patients. Please take note of the package inserts enclosed with the hemodialysis concentrates,
dialyzers, etc.
1.1.3 Contraindications
The hemodialysis machine allows dialysis treatments without any additional equipment. The
machine operates and monitors the dialysate circuit and the extracorporeal blood circuit.
The required treatment parameters can be entered via various menus provided for programming
and displayed on a high-resolution LC display. The current treatment data are shown on the
display.
In the dialysate circuit, purified water is mixed with hemodialysis concentrate, heated, degassed
and delivered to the dialyzer. Inflowing and outflowing quantities are balanced volumetrically. The
pressure at the dialyzer is adjusted depending on the ultrafiltration rate selected and the type of
dialyzer used.
The DIASAFE®plus option permits hemodialysis treatment with high-purity dialysate. The
DIASAFE®plus option basically consists of one filter stage. For the entire time of its use the filter is
an integral part of the hemodialysis machine and is rinsed and disinfected in conjunction with the
machine. The integrity of the filter is tested by means of a pressure holding test which is
integrated in the functional test of the hemodialysis machine. The filter can be used for a
maximum of 100 treatments or for a period of 12 weeks, whichever comes first. Should the filter
fail to pass the filter test, it must be replaced at an earlier date. For the use of this option, the
quality of the water and the dialysate must comply with the applicable standards. (See Chapter
1.6 Initial Start-Up.)
In the extracorporeal blood circuit, the blood is continuously heparinized and passes through the
dialyzer. An air detector prevents infusion of air . Any dangerous loss of blood is prevented by a
blood leak detector and a device monitoring the venous return pressure. The arterial pressure
monitoring unit detects any intake of the needle in the vessel.
The hemodialysis machine is designed for both acetate dialysis and bicarbonate dialysis.
The mixing ratio of concentrate to purified water can be programmed (default setting 1+34 –
selection of other mixing ratios is possible).
Depending on the concentrate used, the Na+ concentration can be readjusted within a range of 125
to 150 mmol.
With the bibag® option the bicarbonate solution is produced from a bag (sodium hydrogen
carbonate powder, NaHCO3). The powder is diluted in the machine to obtain ready-to-use
bicarbonate.
Na+ and UF profiles can be programmed by using the built-in variation function.
The basic version of the hemodialysis machine is designed for double needle dialysis.
Freely selectable programs are available for cleaning and disinfecting the machine.
The hemodialysis machine is equipped with all safety systems required for proper function and
patient safety EN 60601-1 (IEC 601-1).
The automatic test of the safety systems (functional test, also called T1 test) must be performed
prior to each dialysis treatment. This test also checks the integrity of the closed system (pressure
holding test PHT).
In the following cases, the hemodialysis machine will request the user to perform the functional test:
– after a start with external power supply (not power failure),
– after one of the cleaning programs.
For additional safety, cyclic pressure holding tests (CPHT) are also performed during the dialysis
treatment to detect possible leakages in the closed system.
The CPHT is performed automatically every 12.5 minutes. For the CPHT the dialysis machine is
switched for a few seconds into bypass and ultrafiltration is stopped. Balancing is not affected by
these CPHTs, but the effective dialysis time will be reduced by approx. 2 minutes per hour of
treatment.
Caution
Any additional equipment connected to the analog and digital interfaces of the
machine must comply with the applicable EN specifications (e.g. EN 60950 for
data processing equipment, and EN 60601 (IEC 601) for electro-medical equip-
ment). Apart from this requirement, all configurations must comply with the
system standard EN 60601-1-1 (IEC 601-1-1). The connection of additional
equipment to the signal input or output component affects the system configura-
tion and anyone connecting additional equipment is therefore responsible for
compliance with the system standard EN 60601-1-1 (IEC 601-1-1).
These accessories have been approved within the scope of an EC prototype test.
1.2.1 Transportation
● Inside buildings
● Outside buildings
– Never push the hemodialysis machine across an uneven paving (e.g. cobblestone pavement).
Always lift it.
– When transporting the hemodialysis machine in vehicles, accommodate it vertically or hori-
zontally and protect it with an appropriate padding.
– When transporting the hemodialysis machine outside of buildings for a prolonged period of
time, the range of storage temperature must be observed (if necessary, fill the machine with
an antifreezing solution).
The hemodialysis machine must be stored vertically in a well-ventilated room with low variations
in temperature.
Storage temperature:
– without antifreeze +5 °C up to +60 °C
– with antifreeze –20 °C up to +60 °C
When storing the hemodialysis machine with antifreeze, make sure to use antifreeze of the
following composition:
– 48.75 % water
– 48.75 % glycerine
– 2.5 % Teta-plus
Arterieller Venöser
Druck Druck TMP
Reinigen mmHg mmHg mmHg
280 500 500
200
400 400
SN 100
300 300
0
9
200 200
–100
100 100
–200
0 0
–300
Test
Leitfähigk.
mS/cm
Vorbereiten 15,5
Fluß UF
15 I/O I/O
Start/Reset 14,5
7 8 9 0 Esc
14
Dial. 4 5 6
Ton aus 13,5 UF
Flüssigk.
13 Menü Menü 1 2 3 C Best
Pven. 8
Start Start
Stop Stop
2
Rate: ml/h
(Bolus: ml)
( :h.min)
Bolus
Rate
Start
Stop
2a
7
3 6
Legend
1 Monitor 3 Hydraulics
2 Modules – Level 1 4 Brake
(from left to right) 5 Bicarbonate suction tube (blue)
Blood pump (arterial) 6 Concentrate suction tube (red)
Blood pump (SN) 7 Rinse chamber
Air detector 8 Interlock shunt for dialyzer connecting tubes
2a Modules – Level 2 9 IV pole
(from left to right)
Heparin pump
Rinse chamber, bibag®-Connector (option)
not used
3 14
4
5
15
13
12
6 11
10
9
8
Legend
1 2 3 4 5 6 7 8 9 10 11
15
16
17
14 13 12
Caution
Apart from the options mentioned on page 1-8, additional equipment should only
be connected to the ports on the rear of the monitor if the resulting overall system
complies with the requirements of EN 60601-1-1 (IEC 601-1-1), or if their
applicability with regard to technical safety has been proven by a certificate
issued by a testing agency authorized to test the ready-for-use machine.
1 Fan
2 Service switch
3 Audible alarm volume control
4 Connection socket for status indicator/traffic light (optional)
5 RS232 interface (galvanically decoupled by optocoupler) (optional)
6 Input/output port (for external equipment)
7 Screen socket
8 Time meter
9 ALARM IN port (for connection of an external pushbutton switch, which can be used to
initiate a machine stop).
10 ALARM OUT port (for connection of an external alarm indicator, nurse call)
11 Symbols. Caution: consult accompanying documents.
12 Power supply unit (voltage supply for the hemodialysis machine)
13 Power switch (main switch)
14 Power cord (supplies the hemodialysis machine with line voltage, 230 V AC; 50 – 60 Hz)
15 Loudspeaker for audible alarm
16 Port for connection of the remote control (optional)
17 Key interface (patient identification in conjunction with FINESSE) (optional)
1 Rate: ml/min
(Ø: mm)
2
11
3
10
4 Start
Stop 9
1 Display (indicates the delivery rate, the line diameter or an error code)
2 Alarm LED (red)
3 Operation LED (green)
4 Pressure connector (Luer-lock connector of the arterial pressure measuring line)
5 Clamping piece (holds the installed line segment securely in place)
6 Blood pump cover
7 Rotor (delivers the patient’s blood by means of the rollers and the present line set; the rotor
can be removed for cleaning purposes)
8 Sensor (blood pump cover open or closed)
9 Start/Stop key (to turn the blood pump on and off)
10 ▼ key (to reduce the delivery rate or the value of the line diameter)
11 ▲ key (to increase the delivery rate or the value of the line diameter)
12
Rate: ml/h
(Bolus: ml)
1
11
( :h.min)
10
2 Bolus 9
3 Rate 8
Start
4 Stop 7
Pven. 1
1 Keys (to raise (▲) and lower (▼) the fluid level in the venous bubble catcher)
2 Venous pressure connector (Luer-lock connector of the venous pressure measuring line)
3 Holder for the venous bubble catcher with ultrasonic sensors
4 Venous line clamp
5 Optical detector
The float switch (5) located in the heater block (66) controls the inflow of fresh water via the water
inlet valve (41).
The upstream pressure reducing valve (61) ensures a constant water inlet pressure.
In the chamber (88c) of the multifunction block (88), the concentrate pump (23) adds acetate-
containing concentrate. During bicarbonate dialysis, it adds acid concentrate.
When peforming bicarbonate dialysis the bicarbonate is pumped by the membrane pump (25)
from the bicarbonate suction tube (95) via the air separation chamber (131a) into the chamber
(88c) of the multifunction block (88) adding exact quantities of bicarbonate. The conductivity
measuring cell (132) and the temperature sensor (133) measure and monitor the bicarbonate
conductivity.
The heat exchanger (77) transmits part of the heat of the outflowing dialysate to the cold inflowing
water.
The dialysate is heated in the heater rod chamber (66b). The degassing pump (29) and the orifice
(89) generate a negative pressure for deaeration of the dialysate. The air removed is collected in
the air separating chamber from where it passes to the heater block (66) via the loading pressure
valve (65). The air is then discharged into the atmosphere via the float chamber (66c).
The dialysate flows from the chamber (88c) via the mixing chamber (131b) to the balancing
chamber (68) from where it passes through the conductivity measuring cell (7) and the dialyzer
valve (24) to the dialyzer.
In the conductivity cell (7), both the conductivity and the temperature (temperature sensor 3) are
measured and indicated on the monitor.
If temperature and conductivity are inside the alarm limits, the two dialyzer valves (24 and 24b)
are open and the bypass valve (26) is closed.
Should one of the two values be outside the alarm limits (temperature or CD alarm), the bypass
valve (26) opens and the dialyzer valve (24) closes.
The used dialysate coming from the dialyzer is monitored by a blood leak detector (8), from where
it flows into the secondary air separator (88b) of the multifunction block (88). The fluid coming
from this chamber is forced through the flow pump (21) into the balancing chamber (68). This
ensures that the spent dialysate discharged by the dialyzer is always replaced by the same
amount of fresh dialysate.
The pressure transducer (9) is located in the secondary air separator (88b) of the multifunction
block (88). The pressure measured here is taken as a quantity in the computation of the TMP.
A membrane pump (UF pump 22) with a volume of 1 ml per stroke withdraws fluid from the
system at a specified rate. Since the system is closed, the same volume of ultrafiltrate reflows
from the blood via the dialyzer.
After the balancing chamber (68), the used dialysate is directed to the drain via the valve (30), the
heat exchanger (77) and the drain valve (87).
With the dialysate filter (114), the dialysate passes through a capillary filter with a polysulfone
membrane and then flows to the dialyzer. The filter holds back pyrogenic substances and
microorganisms.
The pressure measuring points A, B, C, and D allow direct measurement of various pressures,
without the need to separate lines and tubing. Measuring point A measures the water inlet
pressure; B the loading pressure of the balancing chamber; C the pressure of the flow pump; and
D the negative degassing pump pressure.
During the hot rinse and disinfection programs, the machine is operated in recirculation mode. In
this case, the drain valve (87) is closed and the recirculation valve (86) open. This method of
recirculation serves to save energy (hot rinse) and disinfectant. The disinfectant is introduced into
the circuit by the UF pump (22) via the disinfectant valve (84) and the sensor (115). The sensor
(115) detects whether disinfectant is present or not.
The float switch (5) provided in the heater block (66) regulates the inflow of fresh water via the
water inlet valve (41).
The pressure reducing valve (61) ensures a constant water inlet pressure.
The fresh water is heated in the heater rod chamber (66b). The degassing pump (29) and the
degassing orifice (89) are used to generate a negative pressure, to deaerate the water. The air is
collected in the primary air separator (88c). From there, it is directed into the heater rod chamber
(66b) via the loading pressure valve (65) and then discharged into the atmosphere via the float
chamber (66c).
The preheated and deaerated water is delivered from the primary air separator (88c) to the
concentrate / bicarbonate mixing point (205).
Concentrate containing acetate (or in bicarbonate-based dialysis acid concentrate) flows from
the concentrate suction tube (94) through the concentrate air separator (201) and the concen-
trate pump (23) to the concentrate/bicarbonate mixing point (205). In bicarbonate-based dialysis,
bicarbonate is transported via the bicarbonate suction tube (95), the air separator (203) and the
bicarbonate pump (25) to the concentrate / bicarbonate mixing point (205).
The evacuation valve (188) and the bibag® drain valve (130) are provided for deaeration.
The conductivity measuring cell (132) und the temperature sensor (133) measure and monitor the
bicarbonate conductivity.
At the concentrate/bicarbonate mixing point (205) the appropriate concentrates are mixed with
water. The dialysate then flows across the balancing chamber (68), the dialysate filter (114) and
the conductivity cell (7) to the dialyzer.
The dialysate filter (114) holds back pyrogenic substances and microorganisms.
In the conductivity cell (7), both the conductivity and the temperature (temperature sensor 3) are
measured and indicated on the monitor. The temperature sensor (109) controls the temperature.
If temperature and conductivity are within the alarm limits, the two dialyzer valves (24 and 24b)
are open and the bypass valve (26) is closed.
Should one of the two values be outside the alarm limits (temperature or CD alarm), the bypass
valve (26) opens and the dialyzer valve (24) closes.
The integrity of the DIASAFE®plus (114) membrane is checked by ventilating the internal capillary
space via the vent valve.
The retentate valve (189) is used to rinse retentate from the internal capillary space of the
DIASAFE®plus (114).
The pressure transducer (9) is provided in the secondary air separator (88b) of the multifunction
block (88). The pressure measured there is used as a quantity in the computation of the TMP.
A membrane pump (UF pump 22) with a volume of 1 ml per stroke withdraws fluid from the
system at a specified rate. Since the system is closed, the same volume is removed as
ultrafilatrate from the blood across the dialyzer.
After the balancing chamber (68), the used dialysate is directed to the drain via the valve (30), the
heat exchanger (77) and the drain valve (87).
Part of the heat of the dialysate flowing to the drain is dissipated by the heat exchanger (77) into
the cold, inflowing water.
Any penetration of a larger volume of air into the system, caused by strong degassing or during
priming, is detected by the level sensor (6). This air is discharged into the atmosphere via the air
separation pump (97).
The pressure measurement points A, B, C, and D allow direct measurement of various pressures,
without having to separate lines and tubings. Measurement point A measures the water inlet
pressure; B the balancing chamber loading pressure, C the flow pump pressure, and D the
negative pressure created by the deaeration pump.
In the hot rinse and disinfection programs, the machine is operated in the recirculation mode.
Here the drain valve (87) is closed and the recirculation valve (86) opens. The recirculation mode
saves energy (hot rinse) and disinfectant. The circuit is supplied with disinfectant via the
disinfectant valve (84) and the sensor (115) of the UF pump (22). The sensor (115) detects
presence or absence of disinfectant.
When using the CDS option (central concentrate delivery), concentrate containing acetate, or in
bicarbonate-based dialysis acid concentrate, flows from the concentrate connector (121) through
the valve (102) to the rinse chamber (90a). Bicarbonate is delivered in the same way from the
bicarbonate connector (122) through the valve (104) to the rinse chamber (90b).
The desired type of concentrate supply must be selected in the „SETUP menu“ (see Technical
Manual). The following 5 options are available:
1. No central delivery system (factory setting)
2. Central bicarbonate supply
3. Central acid supply
4. Central bicarbonate and central acid supply
5. Central acetate supply
When selecting a type of central delivery, the user is not restricted to the centrally supplied
concentrate. Other concentrates can also be connected by disconnection or connection of the
concentrate suction tubes. The valves V102 and V104 are controlled automatically. The hemodialysis
machine also recognizes the type of dialysis performed (acetate or bicarbonate dialysis) and uses
this information to control the concentrate or the bicarbonate pump respectively.
☞ Note
Aspiration of concentrates from the containers has priority (see illustration on the
following pages).
If, with central delivery, the concentrate suction tube is repeatedly pulled off, the
rinse chamber will flow over. The rinse chamber will not be drained before the
next cleaning program.
The user has the option to switch off the central delivery system. (See Chapter 2.1.4 Concentrate
Supply.)
☞ Note
The operator is responsible for the proper installation and function of the CDS.
The following illustrations show the possible combinations of connection for CDS.
(The bibag ® option can be used as an alternative to bicarbonate from containers.)
● Mode of operation
The hemodialysis machine permits the optional performance of acetate or bicarbonate dialysis.
The type of treatment is determined by:
– the concentrate(s),
– the positioning of the concentrate suction tubes.
The concentrates are delivered by two pumps and are mixed with RO (reverse osmosis) water in
the hemodialysis machine. The concentrate suction tubes are connected to the appropriate
containers.
With CDS, the appropriate concentrates are delivered by the central delivery system.
The mixing process is volumetric and proportional, i.e. the concentrate pumps add an exact
quantity of concentrate to a fixed volume of water.
The mixing ratio can be varied. The desired ratio can be selected in the dialysate menu. To change
the ratio, the operator must press the Dialysate Menu key.
During bicarbonate dialysis, precipitation of calcium and magnesium salts can occur. The machine
should therefore be decalcified and disinfected after each bicarbonate dialysis, using a suitable
agent (e.g. Puristeril® 340 or Citrosteril®).
During a cleaning program, the concentrate suction tubes must be inserted in the rinse ports
located on the hemodialysis machine.
The correct ion concentration in the dialysate can only be achieved with concentrates that
match the mixing system.
Acid concentrate
SK-F 203 (PGS 21)/SK-F 003 (PGS 01), 35-fold, 6 l in 10-l container.
Bicarbonate concentrate
8.4% sodium hydrogen carbonate solution, 8 l in 10-l container.
☞ Note
The contents of opened bicarbonate concentrate containers cannot be stored.
After mixing, the acetic acid from the acid concentrate reacts with the equivalent quantity of
bicarbonate, producing carbonic acid and sodium acetate.
With the pumps at standard setting, this produces dialysate, which has the following ion
composition:
Procedure
The dialysate is continuously prepared from bicarbonate concentrate, acid concentrate, and
reverse-osmosis water.
Two membrane pumps, which are independent of each other, deliver the acid concentrate solution
and the bicarbonate solution in a volumetrically proportional mixing system.
Caution
Use only specially keyed containers and mating suction tubes.
This system does, however, not exclude the possibility of incorrect use of
uncoded containers.
Only the dedicated bibag® may be connected to the bibag® connector.
to dialyzer
Degassing
via balancing chamber
pump
Concentrate Bicarbonate
pump pump
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
○ ○ ○ ○ ○ ○ ○
RO Concentrate A Concentrate B
water (acetate or acid) (Bicarbonate) or (bibag ® Option)
to dialyzer
via balancing chamber
Concentrate Bicarbonate
pump pump
Degassing
pump
○ ○ ○ ○ ○ ○ ○ ○ ○ ○
○ ○ ○ ○ ○
Measurement of the conductivity and the bypass function prevent that dialysate of the wrong
concentration will enter the dialyzer.
Monitoring of the rated conductivity to within ±5% by setting alarm limits provides a safeguard
against errors in the mixing system.
In the event of dialysate of the wrong composition, the conductivity will be outside the alarm limits.
The CD alarm, which is then generated, will cause the machine to go into bypass, accompanied
by an audible and a visual alarm.
An initial error in the mixing system will be detected before the deviation of the dialysate values from
the rated value becomes dangerous.
B* Belgian bicarbonate
1
for detailed information refer to the Technical Manual
☞ Note
For treatment with the bibag ® (option) the mix ratio must be set to 1+34 or 1+44.
The total conductivity of the ready-to-use dialysate (in mS/cm) is determined by the acidic and
basic part of the bicarbonate hemodialysis concentrates.
The total conductivity of the ready-to-use dialysate with the appropriate mixing ratio is indicated
on each package of the acidic bicarbonate hemodialysis concentrate from Fresenius Medical
Care.
In the event of a ready-to-use dialysate with 32 mmol/l bicarbonate and 3 mmol/l acid (e.g.
acetate content), the contribution of the basic component of the bicarbonate concentrate to the
total conductivity is 2.2 mS/cm. The contribution of the acidic component of the bicarbonate
concentrate to the conductivity depends on the different electrolytic contents.
● Change in conductivity
Based on a basic sodium value of e.g. 140 mmol, the sodium value changes by 5%, if the desired
sodium value is increased to 147 mmol. The total conductivity of the ready-to-use dialysate then
also increases by 5 %.
If the bicarbonate proportion is changed, the acid proportion is changed in the opposite direction,
in order to maintain the desired sodium at a constant value. This may result in a minor CD
alteration.
Example:
A rated sodium value of 140 mmol/l corresponds to 105 mmol/l from the acid concentrate and
35 mmol/l from the bicarbonate concentrate. If the bicarbonate is readjusted by + 3 mmol/l
(= 38 mmol/l on the bicarbonate pump), the acid concentrate pump is readjusted by –3 mmol/l to
102 mmol/l. The rated sodium value of 102 + 38 = 140 mmol/l is preserved.
In order to improve the individual tolerance to dialysis treatments, ultrafiltration profiles and sodium
profiles can be combined either mutually dependent (e.g. high ultrafiltration with high sodium
concentration) or independent.
Separate documentation is available from the manufacturer on request.
Sodium
1 2 3 4 5 6
1
Ultrafiltration
● System Components
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer Venous
valve 1 bubble catcher
Air detector
V24
UF Flow
pump pump Dialyzer
V24b
V26 Bypass Venous
valve line
clamp
Dialyzer
valve 2
Arterial
blood
pump
The filter test (DIASAFE®plus test) is incorporated in the functional test provided it has been
preselected by the DIP switch setting on LP 632.
When the filter test is started, the internal capillary spaces of the DIASAFE®plus must be ventilated
by the vent valve. A special balancing chamber switching cycle is employed to remove fluid from
the interal capillary space across the membrane. The hydrophylic membrane rejects air so that a
negative pressure builds up in the system. Balancing chamber switching stops at a negative
pressure of about – 300 mmHg (peak value). The UF pump increases the negative pressure to
about – 450 mmHg. A differential pressure measurement is started. The result computed by the
microprocessor must not exceed a pressure change of ± 16.7. The maximum duration of the test
is 5 minutes. In the event of a more significant change in pressure, the UF pump will adjust the
negative pressure to the initial pressure. Differential measurement is then restarted.
Another test criterion is the fluid volume removed. A deviation of ± 30 ml from the expected value
of 145 ml will cause the test to stop.
An error message will be displayed if the filter fails to pass the test.
On successful completion of the functional test the hemodialysis machine switches into bypass
mode for a defined period of time. During this period the internal capillary space of the
DIASAFE®plus is automatically refilled with fluid.
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer
valve 1
V24
UF Flow
pump pump Interlock shunt
V24b
V26 Bypass
valve
Dialyzer
valve 2
The fresh dialysate flows from the balancing chamber via the conductivity and temperature
monitoring system to the DIASAFE ®plus .
This filter is integrated in the hydraulic system to ensure that in a no alarm state (no dialysate
alarm: V26 closed, V24 and V24b open) the "fresh" dialysate flows across the filter membrane.
In bypass mode (dialysate alarm: V26 open, V24 and V24b closed) fluid flows through the inner
capillary space of the DIASAFE ®plus . Possible retentate is rinsed from the capillaries and
transported to the drain thus overcoming the disadvantages of a dead end filtration.
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer Venous
valve 1 bubble catcher
Air detector
V24
UF Flow
pump pump Dialyzer
V24b
V26 Bypass Venous
valve line
clamp
Dialyzer
valve 2
Arterial
blood
pump
During the cleaning programs the dialysate lines (red/blue) are connected to the interlock shunt.
The cleaning agent flows through the balancing chamber to the DIASAFE®plus . Valves V24, V24b
and V26 are clocked so that the fluid alternately flows through the internal capillary space
(bypass) and across the filter membrane.
This ensures that in the bypass mode possible retentate is rinsed from the capillaries and is
transported to the drain. The filter is thus thoroughly cleaned, disinfected or rinsed.
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer
valve 1
V24
UF Flow
pump pump Interlock shunt
V24b
V26 Bypass
valve
Dialyzer
valve 2
● System components
DIASAFE ®plus
The DIASAFE ®plus is a hollow-fibre filter. Its excellent filtering characteristics can be attributed to
the Fresenius Polysulfon ® membrane.
Pressure transducer 2
This transducer is a sensor mounted as a standard component in the hydraulic unit. It is used to
control and evaluate the membrane integrity test.
Compressor
The compressor is required for testing the integrity of the membrane of the DIASAFE®plus. Within
a minimum of time it ventilates one side of the filter membrane until the test pressure is achieved.
Test valve
The test valve is only open as long as the compressor ventilates the filter membranes and the test
pressure has built up.
The integrity of the membrane of the DIASAFE®plus is tested by means of a pressure holding test.
This test is integrated in the functional test. For this test, the outer capillaries of the DIASAFE ®plus
are ventilated. This is done by the compressor which pumps air via the test valve into the
hydraulic system. The inflowing air displaces the fluid in the system via the hydrophilic filter
membrane and the retentate valve into the drain. This process continues until the entire outer
capillary space is filled with air. As air cannot pass across an intact membrane, a positive
pressure will build up on the ventilation side. The pressure transducer 2 controls the compressor
and evaluates the integrity test. When the test pressure has been achieved, the compressor is
switched off and the test valve closes. The pressure drop per unit of time is used as a measure for
the evaluation of the integrity of the membrane.
Drain Pressure
valve DIASAFE® P transducer 2
plus Cond./Temp. (182)
(87) monitoring
(114)
(3/7/109)
Retentate Bypass
valve valve
(189) (26)
Pressure
transducer Dialyzer
(9) valve 2
P (24b)
The fresh dialysate flows from the balancing chamber to the DIASAFE®plus.
This filter is integrated in the hydraulic system to ensure that in a no alarm state (no dialysate
alarm: V26 closed, V24 and V24b open) the "fresh" dialysate flows across the filter membrane.
In bypass mode (dialysate alarm: V26 open, V24 and V24b closed) fluid flows through the inner
capillary space of the DIASAFE ®plus . Possible retentate is rinsed from the capillaries and
transported to the drain thus overcoming the disadvantages of a dead end filtration.
In all cleaning programs the dialyzer lines (red / blue) must be connected to the interlock shunt.
The cleaning agent, water or disinfectant flow via the balancing chamber to the DIASAFE®plus. As
the retentate valve, V26 and V24/24b open alternatingly, fluid flows alternatingly through the
inner capillary space and the membrane of the filter.
Drain Pressure
valve DIASAFE® P transducer 2
plus Cond./Temp. (182)
(87) monitoring
(114)
(3/7/109)
Retentate Bypass
valve valve
(189) (26)
Pressure
transducer Dialyzer
(9) valve 2
P (24b)
Legend
The arterial blood pump pumps the blood from the patient’s vascular access to the dialyzer. The
pressure on the suction side of the pump is measured and monitored by the arterial pressure
monitor, which is separated by a hydrophobic filter. Should the pressure exceed or fall below the
set limits, a blood alarm is released: the arterial blood pump stops, the venous line clamp is closed,
a visual and audible alarm is emitted. Ultrafiltration is stopped.
The heparin pump is used to administer heparin doses to the blood at an adjustable rate.
After the dialyzer, the blood enters the venous bubble catcher. There, the venous return pressure
is measured and monitored within tight limits. The monitor for the venous return pressure is
separated by two hydrophobic filters. The venous bubble catcher is located in the air detector as
a safeguard against infusion of air. Should the level drop or should there be frothy blood in the
venous bubble catcher, a blood alarm is emitted: the arterial blood pump stops, the venous line
clamp is closed, a visual and audible alarm is emitted. Ultrafiltration is stopped.
After the bubble catcher, the blood passes the optical detector (OD), which differentiates as
follows:
– OD light (saline or air in the line set) or
– OD dark (blood in the line set).
From the optical detector, the blood is then returned to the patient (venous needle).
☞ Note
This procedure should be used in exceptional cases only, since the stroke
volumes and, thus, the corresponding recirculation shares can be very unfavor-
able.
Should blood flow problems arise in one of the vascular accesses during double needle dialysis,
it is still possible to continue the treatment using the installed line set, by selecting SN click-clack.
After the switching pressure values have been entered, the venous pressure transducer controls
the blood pump and the venous line clamp. Both the arterial and the venous blood line are
connected to the remaining vascular access by means of a Y-piece.
The venous line clamp is closed, as long as the arterial blood pump is delivering blood. As soon
as the upper limit at the venous pressure transducer has been reached in the blood line set, the
blood pump stops and the venous line clamp opens. The pressure, which has developed in the line
set, returns the stored blood to the patient. As soon as the set lower limit has been reached, the
venous line clamp closes, the arterial blood pump starts running, and the same procedure is
repeated periodically.
Should the cycle time be exceeded by 15 s, an alarm will be emitted. The venous line clamp is
closed. The blood pump and the UF pump stop running.
Legend
During SN dialysis, blood is alternately collected from and returned to the patient through one
needle. This is achieved by means of an additional blood pump module, which includes the single
needle control electronics (2-pump procedure with pressure-pressure control).
With SN (QSN) in the twin-pump SN system, the mean blood flow is computed as follows:
QSN = (BPRart x BPRSN) / (BPRart + BPRSN) [theoretical value]
The following approximation formula is sufficient for practical use:
QSN = (BPRart + BPRSN) / 4
The arterial blood pump pumps blood into the external compliance chamber. The blood volume is
dependent upon the set stroke volume (10 to 50 ml). The SN blood pump is not running. The venous
line clamp is closed. The external compliance chamber is connected to a pressure transducer in
the SN blood pump. The compliance chamber, which is filled with air, is located inside the SN blood
pump. As soon as the pressure exceeds the set upper limit, the arterial blood pump stops. The SN
blood pump starts and pumps the blood through the dialyzer, the venous bubble catcher and the
venous line clamp into the patient. As soon as the pressure in the internal compliance chamber
drops below the set lower limit, the described procedure is restarted.
The lower changeover point is fixed to 75 mmHg. The upper changeover point depends on the
stroke volume entered.
The position of the SN blood pump before the dialyzer prevents level fluctuations in the venous
bubble catcher and improves stability of the TMP.
The therapeutical result (clearance) depends on the effective blood flow and the stroke volume.
The higher the stroke volume, the lower the proportionate recirculation.
For this reason, the largest possible blood flow and the largest possible stroke volume should be
set. These settings differ from patient to patient and, owing to the individual vascular accesses,
have different limits.
All other components and their functions correspond to those already described for double needle
dialysis.
The stop detection of both blood pumps is activated. Should one of the pumps stop for longer than
the set time (30 seconds), a blood pump stop alarm is emitted.
☞ Note
The design (closed system) of 4008 hemodialysis machines ensures that pres-
sure fluctuations during SN operation will not affect the UF rate or the UF
volume.
The delivery rate display on the blood pump indicates the theoretical blood flow. This value is
computed from the rotor speed and the inside diameter of the blood line used.
The actual blood flow (effective blood flow) may deviate slightly from this value, because it
depends on the different input pressures (arterial pressure) on the pump.
The computer of the monitor contains an algorithm which corrects the theoretical blood flow,
dependent on the arterial pressure. The value is displayed in the current dialysis menu (Eff. Bld.
Flow).
Using the actual blood flow and the elapsed treatment time, the monitor computes the accumu-
lated blood volume. This information is displayed in the Cum.bld.vol. field in the Dialysis menu.
1 Rate: ml/min
(Ø: mm)
2
11
3
10
4 Start
Stop 9
1 Display (indicates the delivery rate, the line diameter, the SN stroke volume, or an error
code)
2 Alarm LED (red)
3 Operation LED (green)
4 Pressure connector (Luer-lock connector of the SN pressure measuring line)
5 Clamping piece (holds the installed line segment securely in place)
6 Blood pump cover
7 Rotor (delivers the patient’s blood by means of the rollers and the present line set; the rotor
can be removed for cleaning purposes)
8 Sensor (blood pump cover open or closed)
9 Start/Stop key (to turn the blood pump on and off)
10 ▼ key (to reduce the delivery rate, the value of the line diameter, or the SN stroke volume)
11 ▲ key (to increase the delivery rate, the value of the line diameter, or the SN stroke volume)
The following points must be strictly observed to assure safe operation of the hemodia-
lysis machine:
● Electric installation
– The national standards and regulations (e.g. in Germany VDE 0107) are to be observed when
connecting the machine to the local power supply system.
– The use of machines of safety classification I, the quality grade of the protective conductor of
the electric installation represents an essential factor. In many countries, pertinent regulations
have been specified by national authorities, which must be adhered to.
– If the power cord needs to be replaced, use only the original power cord listed in the spare
parts catalog.
– Connect the equipotential bonding to the rear of the machine, if this is required by the legal
regulations of the place of installation (e.g. according to VDE 0107 in rooms of application
group 2) or special applications (e.g. according to VDE 0753 relating to intracardiac surgery).
Caution
When using central venous catheters it is imperative that the potential equillisa-
tion be connected.
With this special type of application all other applied parts (other medical
products) must be of the CF type (isolated applied parts that are approved for
use on the heart).
Apart from this, applied parts of the CF type (e.g. defibrilators) where the patient
currents exceed the maximum allowable CF patient leakage currents may not be
used.
– If additional devices, which are not included in the accessories, are connected to the
hemodialysis machine, there will be the danger of exceeding the permissible leakage
currents.
● Batteries
– Only water, which is suitable for hemodialysis, should be used for operating the hemodialysis
machine. The quality of the water should fulfill the AAMI regulations for hemodialysis systems,
RD5, 1981, part 3.2.
– The ranges of the following parameters are specified in Chapter 1.11 Technical Data: water
inlet pressure, supply pressure of the dialysate concentrate, temperature, and flow rates,
which are required for operating the hemodialysis machine.
– The national rules and regulations with regard to the prevention of reverse flow into the water
supply network and the air gap between drain connection and sewage connection must be
observed when installing and operating the machine (e.g. in Germany VDE 0753, part 4). For
further information refer to Chapter 1.11 Technical Data.
When using the DIASAFE®plus attention should be given to the points listed below with
regard to the quality of product water, concentrate and dialysate:
Caution
Use only reverse osmosis water.
Always ensure that both product water and dialysate meet the applicable
standard (see below).
For microbiologic quality standards / recommendations refer to the table below.
General notes
The microbiologic purity of dialysate prepared in the dialysis center is of critical importance.
Concentrate quality
While acid concentrate may be considered bacteriostatic, bicarbonate concentrate is an ideal
nutrient medium for microorganisms. The use of sterile or microorganism-poor concentrates is
recommended. On-line dry concentrates (bibag®) aid the preparation of dialysate with high
microbiological quality. The use of residues from open canisters and the preparation of bicarbo-
nate concentrate in the dialysis centre are not recommended. Central delivery systems for
concentrate or dialysate must meet high hygienic standards and demand regular inspection.
– Prior to each treatment, the prescribed functional checks must be performed. (See Chapter
2.8.1 Functional Checks Prior to Each Dialysis Treatment.)
● Important documents
– On request, circuit diagrams, descriptions, spare parts lists, and other documents are made
available by the manufacturer. These are intended to support trained personnel in servicing
and repairing the machine.
☞ Note
The machine may only be used as specified in the machine documentation.
Only under these conditions will the manufacturer consider himself responsible
for any effects on the safety, the reliability, and the performance of the machine.
Initial start-up of the machine should be performed by the Fresenius Medical Care technical
customer service or a person authorized by them.
– When bringing the hemodialyis machine from a cooler into a warmer room, allow approx. 2
hours for the machine to adjust to the ambient temperature before turning the unit on.
– Prior to first use of the machine, check the position of the DIP switches, and adjust them, if
necessary. Also verify the setup settings.
(Refer to the Technical Manual)
The current DIP switch settings must then be entered in the Medical Device Register.
Caution
When connecting the hemodialysis
machine for the first time, make sure
to observe the technical specifica-
tions in Chapter 1.11.
Arterial Venous
The On/Off LED is illuminated.
Pressure Pressure TMP
☛ Cleaning mmHg mmHg mmHg
280 500 500
200
☞ Note
To avoid problems during the man-
datory rinse (to remove the anti-
freeze) the service switch must be
set to ON (up).
On completion of the mandatory
rinse set the service switch back to
OFF (down) again.
Mandatory rinse end At the end of the rinse program, the pro-
gram is automatically terminated .
The Cleaning LED stops lighting.
☞ Note
Installation of the DIASAFE®plus must
be entered in the Medical Device Re-
gister (date, batch number).
Verify, and if necessary adjust, the
DIP switch settings prior to initial
start-up. Record the settings in the
Medical Device Register.
● Basic requirements
● Operator steps
Arterial Venous
The On/Off LED is illuminated.
Pressure Pressure TMP
☛
–300
Test The Test indicator is flashing.
Conductivity If the Test indicator is not flashing,
mS/cm one of the conditions for the test to
Prime 15,5 start is not fulfilled.
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Test key.
Conductivity
The Test indicator is illuminated.
☛
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
If the machines fails to pass the
functional test, repeat the test by
pressing the Test key.
If a DIASAFE®plus failure (FXX) is
displayed after the functional test,
this test step has to be skipped.
Then run a heat disinfection program
or a disinfection program followed by
a hot rinse.
Caution
Disinfect prior to the first treatment.
☞ Note
Installation of the DIASAFE®plus must
be entered in the Medical Device Re-
gister (date, batch number).
Verify, and if necessary adjust, the
DIP switch settings prior to initial
start-up. Record the settings in the
Medical Device Register.
● Basic requirements
● Operator steps
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
☛
Cleaning Menu
C Conf
☛
Info message An Info message displays.
Empty filter
C Conf
☛
Info message An Info message displays.
Filling filter
Cleaning menu
Disinfect as specified in the Operating
Instructions.
Caution
Disinfect prior to the first treatment.
7 8 9 0 1 5 Esc
4 5 6
3 4
2
1 2 3 C 6 Conf
1 2
The currently selected menu and the text are always displayed as follows:
white characters on blue background
The following colors are used in the menu field and the menu bar:
Background Characters Data (numerical values), function or menu
blue white can be selected
green black can be changed
grey white cannot be selected or changed
Dialysis menu
Press the key to select the
Alarm limits menu.
7 8 9 0 Esc
4 5 6
1 2 3 C ☛ Conf
C Conf
☛
Press the key
to move to the Position field of the
venous pressure display.
7 8 9 0 Esc
4 5 6
1 2 3 C ☛ Conf
1 2 3 ☛
C
–300
Test Press the Dialysate Menu key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
The Dialysate menu displays.
7 8 9 0 Esc
4 5 6
1 2 3 C ☛ Conf
1 2 3 ☛ C
● Power-up screen
● T1 Test
The Dialysis menu shows the current status of the machine during dialysis.
The Dialysis menu also provides the option to graphically display parameters of a treatment in
progress or of recorded data (see also Dialysis representation).
If other menus are selected and no key is pressed for approx. 30 seconds, the screen automat-
ically returns to the Dialysis menu (exception: UF menu).
● Ultrafiltration menu
● Ultrafiltration menu
Use the 4
1
keys or the keys to
5
2
6
Use the 4
1
keys or the keys to
5
2
6
Use the 4
1
5
2
keys or the keys to se-
6
● Alarm
☞ Note
To enter values while messages are displayed, the message can be temporarily
cleared from the screen by pressing the keys.
● Warning
☞ Note
To enter values while messages are displayed, the message can be temporarily
cleared from the screen by pressing the keys.
The length of the individual cleaning programs can be freely selected (see Technical Manual,
adjustment instructions in the SETUP menu).
☞ Note
Flow alarms during cleaning programs will increase the length of the cleaning
program by the time of the alarm. Should the hemodialysis machine remain in
the no-program state for more than 10 minutes after completion of a cleaning
program, the machine turns off automatically.
Exception: The cleaning program was started with the Auto-On function.
The cleaning program was stopped (e.g. by turning the suction tube).
The following components and functions are monitored during the cleaning programs:
– Dialyzer valve
– Bypass valve
– Balancing chamber switching pulse (flow)
– Float switch
– UF pump
– Temperature
– 24 V and 12 V supply voltages
– Software watchdog
– Shunt door
– Dialyzer lines connected
– Concentrate and bicarbonate suction tubes connected
– CDS
– V84
– Retentate valve
Caution
Disinfect the machine after each dialysis treatment.
☞ Note
If the machine has been idle for more than 72 hours, it is recommended to run a
disinfection program before initiating the next treatment.
The agents for disinfection (Puristeril® 340) and for heat disinfection (Citrosteril® or Diasteril®) are
connected on the rear of the machine (disinfection programs 1 through 4).
Irrespecitve of the container connected to the rear of the machine, the special program 5 can be
used , if necessary, to draw in a hypochlorite-containing disinfectant (e.g. Sporotal® 100) from the
front via the concentrate suction tube.
☞ Note
The heat disinfection program may only be performed with Citrosteril® or Dia-
steril®. If a different type of disinfectant is connected to the rear of the machine
when changing the program (e.g. Puristeril® 340 instead of Citrosteril® or Dia-
steril®), the previously used disinfectant must be thorougly rinsed from the
suction tube.
When the disinfectant is to be drawn in from the front (PGM 5), ensure that a
container with a sufficient amount of disinfectant is connected. Avoid aspiration
of air.
–HR– –HR–
Temp.: 84 °C Temp.: 84 °C
Flow: 450 ml/min Flow: 450 ml/min
Time: adjustable from Time: adjustable from
15 to 30 min in the 15 to 30 min in the
SETUP menu SETUP menu
Timing: Timing:
from > 80 °C from > 80 °C
Perform the maintenance procedures specified in the Technical Manual. The technical safety
checks (TSC) are included in the maintenance intervals.
The items listed in this chapter are an extract from our extensive product range. Further
products on request.
1.10.1 Disposables
Caution
The dialyzers, blood line sets and syringes listed in this chapter are disposables,
identified by the symbol: 2 = do not re-use.
The 4008 H hemodialysis system has been approved for use with the disposables listed below
within the scope of the EC design certification.
● Dialyzers
Caution
These line sets match the pressure limiting system of the blood pumps.
In order to avoid cross infections at the pressure connectors, the use of
hydrophobic filters is required.
Optimum single needle operation is achieved by using the original line sets only.
During continuous operation, the blood line systems must be exchanged after 24
hours at the latest.
● Disposable syringes
● Hemodialysis concentrates
● Calibration concentrate
● Saline solution
Solutions for disinfecting and removing calcium deposits from the dialysate circuit
Caution
Strictly observe the following information:
1. The "use before" date printed on the container.
2. The storage conditions for the disinfectant concentrates.
3. The manufacturer’s instructions for use.
Caution
Prior to the use of disinfectants other than those listed below, their efficacy and
compatibility with the materials affected in the hemodialysis machine and the
DIASAFE ®plus (option) must be ensured.
Improper use of disinfectants (concentration, temperature range, dwell time)
may result in damage to the dialysis machine and the DIASAFE ®plus (option).
Disinfection
☞ Note
After installation, the DIASAFE ®plus can be cleaned with Sporotal® 100 for a
maximum of 11 times.
In machines with basic hydraulics, the cleaning procedures with Sporotal® 100
must be documented.
In machines with advanced hydraulics the cleaning procedures with Sporotal ®
100 will be monitored by the hemodialysis machine.
4. Fresenius Diasteril®:
(Water temperature in the cleaning program: 84 °C)
Active substance base: hydroxyacetic acid
Active substance concentration: approx. 0.8% (diluted)
2. Fresenius Citrosteril®:
(Water temperature in the cleaning program: 37 °C)
Active substance base: citric acid hydrate
Use concentration: approx. 0.8% (diluted)
3. Fresenius Diasteril®:
(Water temperature in the cleaning program: 37 °C)
Active substance base: hydroxyacetic acid
Use concentration: approx. 0.8% (diluted)
Potassium-iodide-starch paper:
Caution
Always use fresh (white) indicator paper.
Caution
Applicable for Diasafe®plus (option) only:
After a disinfection with peracetic acid the following must be observed:
After termination of the rinse program (mandatory rinse) the hemodialysis
machine must be checked for residual disinfectant using the peracetic acid test.
If the test shows a residual concentration of peracetic acid the rinse program
must be restarted. (In this case it is not necessary to run the entire rinse
program.)
Determination of the pH
Caution
Test on the dialysis machine drain fluid for approx. 10 to 15 sec.
If the pH is too high (yellow color pH ≤ 4.1), run a rinse program for a minimum of
5 min., then re-determine the pH.
Fresenius Freka®-NOL:
Type of disinfection: spraying
Active substance base: alcohol (42%)
Use concentration: as concentrate
Caution
Use only the original DIASAFE ®plus specified herein.
Dialyzers
Fresenius Cuprophan ® E1 500 196 1
hollow-fibre dialyzers E2 500 197 1
E3 500 198 1
E4 500 199 1
Disposable syringes
Type Fresenius 30-ml heparin syringe 503 032 1
Type Fresenius 50-ml heparin syringe 900 071 1
biBag® (option)
Sodium bicarbonate 700 g 508 507 1 GB, D, F, NL, E, P, I
Sodium bicarbonate 650 g 508 992 1 GB, D, F, NL, E, P, I
Sodium bicarbonate 650 g 508 727 1 S, SF, DK, N
Sodium bicarbonate 950 g 508 991 1 GB, D, F, NL, E, P, I
Sodium bicarbonate 950 g 508 732 1 S, SF, DK, N
Calibration concentrate
Fresenius UHB calibration concentrate 193 615 1 1000-ml bottle
Line voltage 230 V AC, +6%, –10%, 50/60 Hz Symbol: (alternating current)
240 V AC, ±10%, 50/60 Hz Symbol: (alternating current)
110 V AC, ±10%, 50/60 Hz Symbol: (alternating current)
Operating current:
– Dialysis: Approx. 6 A, at a water inlet temperature of 17 °C
dialysate temperature: 37 °C
dialysate flow: 500 ml/min
230 V~ 50/60 Hz
2 11 A
3 7
4 6
5
Fresenius Medical Care AG
D-61346 Bad Homburg
Tel.: ++49 (0) 61 72 / 609-0
Legend
Water drain: 0 up to 100 cm above the ground, no less than 5 cm free fall. The
water drain must be located at a lower level than that of the dialyzer
position.
Concentrate supply: 0 to –100 mbar; max. suction height: 1 m, max. height of fall: 0 m
(with central delivery system – optional: 0 up to 500 mbar)
Heat dissipation: Dialysis: approx. 400 watts (at an ambient temperature of 20 °C)
Hot rinsing: approx. 520 watts (at an ambient temperature of 20 °C)
Range of operating 15 °C up to 35 °C
temperature:
Temporary downtime: It is recommended to run a hot rinse program without cooling rinse
(PGM 2) before putting a machine temporarily out of use and to run
a disinfection program prior to reuse.
Stability: 10°
Caution
Any additional equipment connected to the analog and digital interfaces of the
machine must comply with the applicable EN specifications (e.g. EN 60950 for
data processing equipment, and EN 60601 (IEC 601) for electro-medical equip-
ment). Apart from this requirement, all configurations must comply with the
system standard EN 60601-1-1 (IEC 601-1-1). The connection of additional
equipment to the signal input or output component affects the system configura-
tion and anyone connecting additional equipment is therefore responsible for
compliance with the system standard EN 60601-1-1 (IEC 601-1-1).
Suppression of Mute alarm time: adjustable in the SETUP menu from 1 to 2 minutes
audible alarm: (factory setting: 1 minute).
Any new alarm reactivates the silenced audible alarm.
Preparation: Defined by the optical detector below the venous bubble catcher.
Preparation is terminated as soon as the optical detector senses
dark in the blood line set.
Priming the blood line set: Automatic priming of the blood line set upon pressing a key (Prime).
Priming is automatically terminated as soon as the venous bubble
catcher is filled, however, not later than after 1 to 5 minutes (time
adjustable in the SETUP menu).
Single needle With arterial blood pump and venous line clamp; pressure-pressure-
click-clack dialysis: controlled with adjustable pressure reverse values.
Exceptional procedure in case of problems with the vascular access
during double needle dialysis.
Single needle dialysis Two-pump procedure with adjustable upper pressure reverse value
(optional): (adjustable via stroke volume).
Blood leak detector: Threshold of response less than or equal to 0.5 ml blood loss per
minute to the dialysate at a hematocrit of 0.25.
(Flow rate 300 ml/min to 800 ml/min)
With the dialysate flow turned off, a blood leak alarm response is delayed. The
delayed response is dependent upon the fluid volume in the hydraulic compart-
ment between dialyzer and blood leak detector (approx. 110 ml total fluid volume)
as well as upon the size of the membrane rupture in the dialyzer. The response of
a blood leak alarm is, in addition, dependent upon the set ultrafiltration rate.
Degassing: Residual gas volume in the dialysate at the inlet to the dialyzer
approx. 11 ml air per 1 liter of water (at 1000 hPa and 0 °C)
When operating with reverse osmosis water of 22 °C and an air
content of approx. 21 ml air per 1 liter of water (at 1000 hPa and
0 °C)
Method: negative pressure
– Acetate: Standard setting 1+34 (other mixing ratios can be selected in the
SETUP menu). Setting range : 125 up to 150 mmol/l, depending
upon the concentrate used.
Volumetric proportional mixing system. The rated concentration is achieved not
later than after 10 minutes.
Arterial range:
– Arterial pressure Display range: –300 to +280 mmHg
measurement: Resolution: 20 mmHg
Accuracy: 10 mmHg
The OD in the LD senses light:
Alarm window width: –300 to +280 mmHg
The OD in the LD senses dark:
Alarm window width: 120 mmHg, adjustable over the entire
display range
With single needle dialysis as well as with single needle click-
clack dialysis: the OD in the LD senses either dark or light; upper
limit fixed to 280 mmHg; lower limit adjustable from 0 up to
–300 mmHg.
Monitoring of the arterial pressure.
Venous range:
– Venous pressure Display range: –60 to +520 mmHg
measurement: Resolution: 20 mmHg
Accuracy: 10 mmHg
The OD in the LD senses light:
Alarm window width: –60 to +520 mmHg
The OD in the LD senses dark:
Alarm window width: 40 to 200 mmHg, adjustable within a
range of 20 to 520 mmHg
With single needle click-clack dialysis:
Alarm window width: 80 to 460 mmHg, adjustable
from 20 up to 480 mmHg
Monitoring of the venous back pressure.
– Air detector: Threshold of response: drop in the fluid level, air bubbles, froth or
microfroth (blood/air mixture) over the entire blood flow range from
0 up to 600 ml/min
Procedure: ultrasonic transmission
Audible alarm:
– Setting range of the Adjustable with the volume control on the rear of the monitor,
loudness of the minimum setting: ≥ 65 dBA
audible alarm:
Only environmentally compatible and recyclable materials are used for the manufacture of
hemodialysis machines.
Prior to the disposal of hemodialysis machines, any possible risk of infection must be eliminated
by appropriate disinfection.
Electronic boards can be disposed of as set forth in the regulation on the disposal of electronic
scrap.
For the disposal of lead-acid batteries, the respective regulations in force at the time of disposal
apply.
Further information on disposal can be obtained on request.
Materials used:
(Materials marked with an asterisk come into contact with water, dialysate and dialysate concen-
trate.)
Plastics
Metals
Aluminum 1.4401*
Sheet metal, Cu55, 1.1203 1.4404*
Sheet metal, DIN 1541, MK75, unhardened Spring steel 1.4568 – DIN 17224
Sheet metal, DIN 1783, AlMg Titanium – TiAl 4 V6*
Sheet metal, DIN 1798, AlMgSi o.5 Spring steel 1.4310 – DIN 59381
DIN 174 – St37K Steel, 1.4300 (for systems)
1.4301 Steel, 9S MnPb, 28K
1.4305 – DIN 178/668 Galvanized steel
Motors Cables
Copper Copper
Cast steel PVC
Cast zinc Teflon
Miscellaneous
Adhesives
Loctite
Scotch-Weld DP 499
Scotch-Weld V 23
Cyanolit
Elastosil
Blood alarm Group of alarms which cause the blood systems to stop:
– Arterial pressure
– Venous back pressure
– Transmembrane pressure
– Blood leak
– Air
– BP stop alarm
Blood systems Systems which maintain and control the function of the extracor-
poreal circuit.
Bubble catcher Device integrated into the blood line set for separating gases
which are not dissolved in the blood.
Bypass Process during which the dialysate is diverted away from the
dialyzer.
Compliance chamber Device for levelling variations in pressure and volume, e.g. with
use being made of the single needle method.
Dialysate pressure The pressure present in a defined section of the system carrying
the dialysate, e.g. at the dialyzer outlet.
Dialyzer blood connector Arterial or venous blood connector provided at the dialyzer.
Extracorporeal circuit A section of the blood circulation which takes place outside of the
body.
Line clamp Device for automatically clamping the blood line set, e.g. in the
event of alarms or during single needle operation.
Pressure measurement Connector for the pressure measuring line, e.g. Luer-Lock.
connector
Pressure measuring line Line connecting the line set to the pressure monitoring unit.
Pump head The pump head comprises pump rotor and pump stator.
Pump line holder Device for fixing the blood pump line segment.
Pump stator Stationary support for the pressure pulleys of the pump rotor.
Single needle method Technique in which the vascular access for the treatment proce-
dure is established by means of one puncture only.
Transmembrane pressure Difference between the pressures acting upon the dialyzer mem-
brane (blood pressure, pressure in the dialysate).
Venous return pressure Pressure in the extracorporeal circuit before the venous cannula
(e.g. in the venous bubble catcher).
Water alarms Group of alarms which do not cause the blood systems to be
turned off: conductivity (bypass operation), temperature (bypass
operation), flow.
1.13.1 Abbreviations
AC = alternating current
BC = balancing chamber
BIC = bicarbonate
BP = blood pump
CD = conductivity
DC = direct current
HD = hemodialysis
HDF = hemodiafiltration
IS = interlock shunt
LD = air detector
OD = optical detector
SN = single needle
SV = solenoid valve
UF = ultrafiltration
Alternating current
Ground/system ground
I/O ON / OFF
1 Visual inspections
2 General inspections
4 Ultrafiltration system
6 Mixing system
8 Modules
8.1.1 Set the line diameter to 8 mm; 18 rpm or gauging the capacity □
check the speed at 200 ml/min. of the delivery volume of 200
ml/min in liters
8.1.2 Open the blood pump cover. The blood pump stops; BP stop □
alarm after 30 seconds
9 Electrical tests
(Values according to
DIN VDE 0751-1)
10 DIASAFE®plus (option)
Publications, investigating report on polysulfone membranes, with the enclosed literature refer-
ences.
– “No evidence for endotoxin transfer across high flux polysulfone membranes”, among others
by Professor Dr. med. Jürgen Bommer from the Ludolf-Krehl-Klinik in Heidelberg.
– “Filtration of dialysate using an ON-LINE dialysate filter”, dated June 5, 1990, among others by
Francise Dumler MD, Division of Nephrology & Hypertension, Henry Ford Hospital, 2799 West
Grand Blvd., Detroit, MI, 48202, USA as well as Fresenius brochures on DIASAFE and
polysulfone high-flux with literature.
– “Hemodiafiltration with ON-LINE Production of Bicarbonate Infusate: a new standard for high
efficiency, low-cost dialysis in elderly and uncompliant patients” among other by Dr. B.
Canaud, Division of Nephrology, Lapeyronie University Hospital, 555, Route de Ganges,
34059 Montpellier, France
Chapter Page
2.7 Priming the Extracorporeal Blood Circuit With a Saline Bag .................................. 2-36
2.7.1 Priming ............................................................................................................................ 2-36
2.7.2 Setting the Delivery Rate of the Blood Pump ................................................................. 2-37
2.7.3 Stopping Delivery ............................................................................................................ 2-38
2.7.4 Interrupting the Priming Program .................................................................................... 2-40
2.7.5 Stopping the Priming Program ........................................................................................ 2-40
☛
On/Off Air detector Blood leak
Arterial Venous
The On/Off LED is illuminated.
Pressure Pressure TMP
Rate: ml/min
The LED segments, the Alarm LED (red)
(Ø: mm)
and the Operation LED (green) light for
1 second.
Start
Stop
Rate: ml/min
The line size displays for approx. 2 se-
(Ø: mm)
conds.
Start
Stop
Rate: ml/min
The delivery rate displays.
(Ø: mm)
Start
Stop
Rate: ml/h
(Bolus: ml)
The LED segments, the alarm (red)
LED, the operation (green) LED and the
( :h.min)
stop-watch time (green) LED light for
Bolus
1 second.
Rate
Start
Stop
Start
Stop
( :h.min)
Bolus
Rate
Start
Stop
automatic program will be started
(lasting approx. 35 seconds).
This program will first remove all air from
the bibag® and will then fill the bag with
water.
Acid
concentrate
Rate: ml/h
( :h.min)
Bolus
Rate
Start
Stop
tube for acid and acetate in the acid con-
tainer.
Acid
concentrate
( :h.min)
Bolus
Rate
Start
Stop
automatic program will be started
(lasting approx. 35 seconds).
This program will first remove all air from
the bibag® and will then fill the bag with
water.
Acid
concentrate
General notes
☞ Note
When using CDS, connect the con-
centrate suction tubes to match the
selected type of concentrate delivery
(SETUP) (See Chapter 1.3.5, Central
Delivery System).
Default setting in SETUP: CDS off
☞ Note
With CDS-supplied bicarbonate and
acid , CDS must be deactivated in
the user menu when performing ace-
tate dialysis with containers.
–300
Test Press the Dialysate Menu key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
C Conf
☛
Fresenius Medical Care 4008 S 11/02.01 (OP) 2-11
2.2 Functional Check
– The functional check can be performed with or without previously installing the line set.
If the line set was filled during the cleaning program, the dialyzer must be deaerated before
starting the functional test. This is done by briefly opening and then closing the upper screw
cap of the dialyzer.
– Depending on the setting defined in the SETUP the functional test will be started automatically
when turning power on or must be started manually by pressing the Test key.
– Start conditions:
– The dialysate connectors (red/blue) are in the interlock shunt.
– The shunt door is closed.
– The optical detector in the air detector module senses clear fluid.
– The concentrate suction tubes are in the appropriate container or, for use with CDS, are
properly connected depending on the type of supply.
● Automatic Start
–300
Test The Test LED is flashing.
Conductivity Internal tests.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
Prime 15,5
Flow UF
The Test LED is illuminated.
15
If one of the starting conditions is not
On/Off On/Off
fulfilled, an audible alarm will be
Start/Reset 14,5 sounded.
7 8 9
14 Use the Test or Tone Mute key to ack-
nowledge the audible alarm.
Make sure all starting conditions are
fulfilled.
Then press the Test key to start the
test.
☞ Note
To prevent an automatic start of the
test proceed as follows:
– Open the shunt door.
– Turn the hemodialysis machine
on.
– Start the functional test as usual.
– Close the shunt door.
–300
Test The Test LED is flashing.
Conductivity If the Test LED is not flashing, one of
mS/cm the conditions for the test to start is not
Prime 15,5 fulfilled.
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
–300
☛
Test Press the Test key.
Conductivity
mS/cm The Test LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
● Basic condition
● Start
–300
Test Press the Prime key.
Conductivity
mS/cm The Test LED is illuminated.
☛
Prime 15,5
Flow UF The Prime LED is illuminated.
15 On/Off On/Off
14,5
For precirculation press the Start/Reset
Start/Reset
7 8 9 key.
14
Precirculation can only be started after
4 5 6 completion of the display test.
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
T1 Test Prime
Temporary display message
or
T1 Test Prime / Circulation
Rate: ml/min
The Operation LED (green) is illumin-
(Ø: mm)
ated.
The blood pump is delivering.
Start
☛ The arterial blood pump fills the line set
up to the venous bubble catcher.
Stop A delivery rate > 180 ml/min will auto-
matically be returned to 180 ml/min.
The delivery rate can be changed by
the operator as desired
● Delivery stops
–300
Test The Test LED is illuminated.
Conductivity
mS/cm
The Tone Mute LED is flashing.
Prime 15,5
Flow UF Audible signal
15 On/Off On/Off cleared automatically
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Start/Reset key.
Conductivity The blood pump starts running and the
mS/cm venous line clamp opens.
Prime 15,5 Fluid is recirculated.
Flow UF
15 On/Off On/Off The Test LED is illuminated.
☛
Start/Reset 14,5
7 8 9 The Prime LED is illuminated.
14
13,5 Dialysate
4 5 6 The Tone Mute LED is dark.
Tone Mute UF
13 Menu Menu 1 2 3
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
–300
Test Press the Start/Reset or the Prime key.
Conductivity
Prime
mS/cm
Continue priming by pressing the Prime
15,5
Flow UF key again or press the Start/Reset key
☛
15 On/Off On/Off to continue precirculation.
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
● Parallel test
● Serial test
T1 Test passed
Temporary display message
–300
Test The Test LED is dark.
Conductivity
mS/cm The Prime LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Preparation
Display message
–300
Test The Start/Reset LED is illuminated.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test The Test LED is flashing.
Conductivity
mS/cm The Tone Mute LED is flashing.
Prime 15,5
Flow UF Audible signal
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
T1 Test failed
Display message
–300
☛
Test Press the Test key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Test
Conductivity
mS/cm The Test LED is illuminated.
Prime 15,5
Flow UF The Tone Mute LED is dark.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
The following error messages can be
acknowledged with the Test key:
Skip battery?
Skip Diasafe?
–300
Press the Test key.
☛
Test
Conductivity
mS/cm
The Test LED is flashing.
Prime 15,5
Flow UF Audible signal
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Press the Test key.
☛
Test
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
– Failure to fulfill one of the conditions to start the test listed in Chapter 2.2.2 Start of the Test.
– The test will be continued after the problem has been corrected.
☛
On/Off Air detector Blood leak
Arterial Venous
The On/Off LED stops lighting.
Pressure Pressure TMP
The test can only be stopped by turning
the machine off.
Rate: ml/min
(Ø: mm)
➀ Open the cover.
➁ Press the Start/Stop key and keep it
pressed until the rotor stops.
Start
Stop
☛
➁
Rate: ml/min
(Ø: mm)
➀ Install the line segment on the left.
➁ Press the Start/Stop key and keep it
pressed until the rotor stops.
Start
Stop ➂ Install the line segment on the right.
➄ ☛ ➃ Close the cover.
➁
➄ Connect the pressure measuring line
to the pressure port.
➀ ➂
➃
Rate: ml/min
Should the blood pump rotor be running:
(Ø: mm)
Start
☛ Selectable range:
2 to 10 mm in 0.2 mm increments
Stop
Rate: ml/min
Press the Start/Stop key to accept the
(Ø: mm)
value.
(The value is displayed for approx. 2
seconds)
The delivery rate of the blood pump is
reset to 0.
Start
Stop
☛
Rate: ml/min
Use the ▲ key to set the delivery rate.
(Ø: mm)
☛
Start
Stop
Rate: ml/h
(Bolus: ml)
Press the ▲ key.
The slide carriage will move to the fully
open position.
( :h.min)
Bolus
Rate
☛
☞ Note
When the slide carriage moves to the
fully open positon, the bolus data are
reset to 0.
Rate: ml/h
(Bolus: ml)
➀ Connect a syringe filled with heparin
to the heparin line of the arterial line
( :h.min)
set.
Bolus ➁ Insert the syringe.
:h.min)
Note
The plate of the syringe plunger must
engage in the slide carriage.
Bolus
Rate
Start
Stop
☞ Note
The syringe wings must be in the
syringe wing slot.
Rate: ml/h
(Bolus: ml)
Press the Rate key.
The display is flashing.
( :h.min)
Bolus
☛ Rate
Start
Stop
Rate: ml/h
(Bolus: ml)
Use the ▲ and ▼ key to set the desired
rate.
Pressing the key for more than 3 sec-
( :h.min)
onds will increase the change rate.
Bolus
Selectable range: 0.1 to 10 ml/h
in 0.1 ml/h increments
Rate
Start
☛
Stop
Rate: ml/h
(Bolus: ml)
Press the Rate key to accept the value
displayed.
The display will continue flashing until
( :h.min)
the set rate has been accepted by
Bolus
pressing the Rate key.
During this time the heparin pump de-
Rate livers at the previous rate.
☛
Start
Stop
☞ Note
Administer the heparin dose accord-
ing to the physician's instructions!
☞ Note
For the injection of an initial heparin
dose use the Bolus key.
Rate: ml/h
(Bolus: ml)
Press the key.
The display is flashing.
( :h.min)
Bolus
Rate
☛ Start
Stop
Rate: ml/h
(Bolus: ml)
Use the ▲ and ▼ key to set the desired
time
(Pressing the key for more than 3 sec-
( :h.min)
onds will cause the change rate to
Bolus switch from 1-minute to 10-minute
steps.)
Rate ☛ Selectable range: 1 minute to 9 hours
59 minutes in steps of 1 minute.
Start
Stop
Rate: ml/h
(Bolus: ml)
Press the key to accept the selected
time.
The display will continue flashing until
( :h.min)
the set time has been accepted by
Bolus
pressing the key.
During this time, the delivery time pre-
Rate viously set remains active.
☞ Note
Place the clot catcher approx. 1 cm
below the lower edge of the bubble
catcher holder.
Caution
Make sure that no ultrasound-con-
ducting media is used on the air de-
tector.
➁
➀
➂
Caution
Do not remove the blood line from
the optical detector during dialysis
treatment.
Regarding the regulations for the application of hemodialysis machines the following has
to be observed:
Caution
Use aseptic technique for all blood
side connections.
Use hydrophobic filters in the pres-
sure return lines to prevent contami-
nation. A wetted hydrophobic filter
must be replaced!
Caution
When inserting the line sets make
sure that the blood lines are not
kinked.
Use the line holders provided.
Caution
During the treatment verify at appro-
priate intervals:
– the condition of the patient
– the function of the hemodialysis
machine and the extracorporeal
blood circuit. Pay particular atten-
tion to the venous insertion site,
as a possible dislocation of the
venous cannula may not always
be detected by the pressure
monitoring system.
Caution
It is important to set the lower
venous alarm limit to be as close as
possible to the actual value of the
venous pressure.
Pven.
Start Start
Stop Stop
Rate: ml/h
(Bolus: ml)
( :h.min)
Bolus
Rate
Start
Stop
2.7.1 Priming
☛
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Prime
Display message
Rate: ml/min
The Operation LED (green) is illumin-
(Ø: mm)
ated.
The blood pump is delivering.
Start
☛ The arterial blood pump fills the line set
up to the venous bubble catcher.
Stop A delivery rate > 180 ml/min will auto-
matically be returned to 180 ml/min.
The operator may select any delivery
rate desired.
or
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Start/Reset key.
Conductivity The blood pump starts running and the
mS/cm venous line clamp opens.
Prime 15,5 Allow fluid to recirculate until all air has
Flow UF
been removed from the line sets and
15 On/Off On/Off
the dialyzer.
Start/Reset 14,5
☛
7 8 9
14 The Start/Reset LED is illuminated.
4 5 6
Tone Mute 13,5 Dialysate UF The Tone Mute LED is dark.
13 Menu Menu 1 2 3
–300
Test Should the venous bubble catcher not be
Conductivity filled within the time defined in the SET-
mS/cm
UP:
☛
Prime 15,5
Flow UF
15
Press the Prime key a second time.
On/Off On/Off
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
Start
Stop
☛
2.7.5 Stopping the Priming Program
–300
Test Press the Start/Reset or the Prime key.
Conductivity
mS/cm Press the Prime key a second time to
continue priming.
☛
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Caution
The functional checks must be performed by the operator prior to each dialysis
treatment.
1. Concentrate container:
Use only the dedicated keyed containers for bicarbonate dialysis.
2. Verification of the dialysate composition:
Check the conductivity displayed on the monitor with the specifications on the
acid / concentrate container. (Observe the settings for sodium and bicarbon-
ate in the Dialysate menu!)
Caution
If the mixing ratio was set in the SETUP to be individually adjustable, always
make sure:
1. that the mixing ratio set by the technician complies with the mixing specifica-
tions for the concentrates used.
2. to check the dialysate for correct physiological composition before initiating
dialysis.
3. that the individually adjustable mixing ratio and the bibag® option are not used
simultaneously.
–300
Test Press the Dialysate Menu key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
☞ Note
For acetate dialysis (mixing ratio
1+34), with the dilution set to a diffe-
rent ratio, verify the basic conditions:
Acetate connected, bicarbonate suc-
tion tube in the rinse port, central bi-
carbonate delivery off, bibag ® not
connected.
Info message
Check Concentrate! Press the Conf key to confirm the safety
(Select with [Confirm] key) prompt.
Caution
Verify proper acetate connection and
make sure that the corresponding
mixing ratio has been selected.
Inadvertant use of an acid concen-
trate intended for a 1+45 mix ratio in-
stead of the acetate and with the mix
ratio set to 1+34 may result in a con-
ductivity that is within the display
range, although the composition of
the dialysate is not physiological.
Use of this dialysate presents a
hazard to the patient.
–300
Test To set the following parameters press
Conductivity the Dialysate Menu key to display the
Prime
mS/cm
Dialysate menu.
15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Use the 4
1
5
2
keys or the
6
3
keys
to enter the Base Na +.
Use the 4
1
5
2
keys or the keys
6
☞ Note
The respective sodium and bicarbo-
nate ion concentrations are specified
in the product information.
Caution
The alarm limits of the conductivity
display must be set around the ex-
pected value.
The actual value in the conductivity
display must have attained the ex-
pected target value after a maximum
of 10 minutes.
Should this not be the case, the ac-
tual value must first be checked in
the laboratory. If the result turns out
to be negative, the machine and/or
the concentrates should be checked.
–300
Test Press the Dialysate Menu key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Use the 4
1
5
2
keys or the
6
3
keys
to enter the flow rate.
Selectable rates:
300, 500, 800 ml/min
Use the 4
1
5
2
keys or the
6
3
keys
to enter the temperature.
Selectable range: 35.0 to 39.0 °C
in 0.5 °C increments
C Conf
☛
–300
Test If the dialyzer is to be prepared with a
Conductivity UF rate:
mS/cm
Prime 15,5
Flow UF
Press the UF Menu key.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 ☛ Menu Menu 1 2 3
Use the 4
1
5
2
keys or the
6
3
keys
to enter the desired rate.
☞ Note
When the optical detector senses
clear fluid and the dialysate pressure
is positive during preparation, the UF
rate is automatically set to 500 ml/h.
When entering and starting UF para-
meters, only the values entered are
relevant.
–300
Test Press the UF On/Off key.
Conductivity
mS/cm The UF LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Chapter Page
☞ Note
If the dialyzer has been prepared
with a UF rate, the UF volume must
be erased.
☞ Note
This function erases all treatment
parameters stored in the machine.
● Maximum values:
● Starting Position
with the 4
1
5
2
6
3
keys or the keys.
with the 4
1
5
2
6
3
keys or the keys.
with the 4
1
5
2
6
3
keys or the keys.
with the 4
1
5
2
6
3
keys or the keys.
with the 4
1
5
2
6
3
keys or the keys.
with the 4
1
5
2
6
3
keys or the keys.
● Rate function
with the 4
1
5
2
6
3
keys or the keys.
☞ Note
If only the UF rate is entered for ul-
trafiltration, check the UF rate dis-
played in the dialysis menu for plaus-
ibility after saving the data.
with the 4
1
5
2
6
3
keys or the keys.
–300
Test Press the UF On/Off key.
Conductivity
mS/cm The UF LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the UF On/Off key.
Conductivity
mS/cm
The UF LED is dark.
Prime 15,5
Flow UF The UF pump stops.
15 On/Off On/Off
Counting of the UF time left stops.
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the UF On/Off key a second time.
Conductivity
mS/cm
The UF LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the UF On/Off key to interrupt
Conductivity ultrafiltration.
mS/cm
Prime 15,5
Flow UF
The UF LED is dark.
15 On/Off On/Off The UF pump stops.
☛
Start/Reset 14,5
7 8 9 Counting of the UF time left stops.
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the UF Menu key to display the
Conductivity Ultrafiltration menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
with the 4
1
5
2
6
3
keys or the keys.
C Conf
☛
Prime 15,5
Flow UF
The UF LED is illuminated.
15 On/Off On/Off
14,5
☛
Start/Reset
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Rate: ml/min
Press the Start/Stop on the blood pump
(Ø: mm)
module.
The Operation LED (green) is dark .
The arterial blood pump has stopped.
Start
Stop
☛
Connect the arterial line set to the pa-
tient.
If infusion of the saline contained in the
line is desired, connect now the venous
line set.
Rate: ml/min
Set the delivery rate of the arterial blood
(Ø: mm)
pump to the desired value.
Start
☛
Stop
Caution
When using high-flux dialyzers make
sure that the alarm limits for monito-
ring the TMP are as narrow as possi-
ble.
Start/Reset 14,5
7 8 9 Connect the venous connector of the line
14
set to the patient (provided this has not
4 5 6
Tone Mute 13,5 Dialysate UF been done yet).
13 Menu Menu 1 2 3
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Display message
The green (operation) status indicator
is illuminated.
Dialysis
Caution
When using high-flux dialyzers and
selecting low UF rates there is a pos-
sibility of local backfiltration.
Backfiltration depends on:
– the type of high-flux dialyzer
– the flow resistance on the dia-
lysate and the blood side
– the viscosity of the blood
Rate: ml/h
(Bolus: ml)
Press the Start/Stop key on the heparin
pump module.
( :h.min)
The Operation LED (green) is illumin-
Bolus ated.
The heparin pump is delivering.
Rate
Start
Stop
☛
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
● Heparin
Rate: ml/h
(Bolus: ml)
Check the bolus volume:
Press the Bolus key on the heparin
( :h.min)
pump module.
☛ Bolus
Rate
The bolus volume already adminis-
tered will be indicated.
Start
Stop
Rate: ml/h
Administering a bolus:
(Bolus: ml)
☛ Bolus
Rate
Display: 0.0
The heparin pump will administer the
bolus in 0.1 ml increments.
After releasing the key, the display will
Start
Stop show for 5 seconds the bolus volume
injected.
(The bolus volume administered will be
added to the bolus volume stored.)
Then the delivery rate will be displayed
– Max. bolus administration: 5 ml
– Max. bolus volume in total:
1x syringe capacity
When adding physiological saline solution (saline) in the venous bubble catcher, the blood may
be diluted to a color which the optical detector no longer senses as blood.
3.3.1 Preparation
Start
Stop
☞ Note
Do not install the single-needle pump
segment!
Start
Rate Stop
☛
Rate: ml/min
Press the Start/Stop key on the blood
(Ø: mm)
pump module
The Operation LED (green) is dark.
The arterial blood pump has been
stopped.
Start
Stop
☛
Connect the arterial line set to the pa-
tient.
If infusion of the saline contained in the
line is desired, connect now the venous
line set.
Rate: ml/min
Use the ▼ or ▲ key to set the delivery
(Ø: mm)
rate for the arterial blood pump.
Start
☛
Stop
Start
Rate Stop
Start
Stop
Start
☛ higher than the delivery rate selected
for the arterial blood pump
Rate Stop
200 200
Rate: ml/min
Press the Start/Stop key on the arterial
(Ø: mm)
blood pump module.
The Operation LED (green) of the blood
pump is illuminated.
The blood pumps are delivering (pro-
Start
Stop vided there is no blood alarm)
☛
The optical detector senses dark fluid.
The yellow status indicator (warning/
info) is illuminated.
Preparation end
Display message
–300 The alarm limits (arterial, venous, TMP
Test
and conductivity) are set automatically.
Conductivity
mS/cm
The blood pump stops.
Prime 15,5
Flow UF
15 On/Off On/Off The venous line clamp closes.
Start/Reset 14,5
7 8 9 Audible signal
14
13,5 Dialysate
4 5 6 The Start/Reset LED is flashing.
Tone Mute UF
13 Menu Menu 1 2 3 The Tone Mute LED is flashing.
Connect the venous line set to the pati-
ent (unless this has already been done).
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF The blood pumps are started automat-
15 On/Off On/Off ically.
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
SN-Dialysis
Display message
☞ Note
When performing single-needle dia-
lysis, it is important to dialyze with
maximum blood flow.
This procedure should be used in exceptional cases only, as the stroke volumes and thus the
respective recirculation may vary widely.
The arterial and the venous blood line are connected with a Y-piece to the vascular access.
C Conf
☛
C Conf
☛
SN-Click-Clack
Display message
C Conf
☛
Use the keys
to move to the fields
– Upper Limit
– Lower Limit
Enter the desired values with the keys.
To stop:
Use the keys
to move to the SN Click -Clack field.
Set SN Click-Clack to Off
with the keys.
C Conf
☛
Isolated ultrafiltration can be started at any time and can be repeated as often as necessary.
The parameteres entered at the beginning of the treament (UF Goal and UF Time) must be taken
into consideration.
The total volume to be removed (UF Goal), the total treatment time (UF Time Left) or the UF Goal
and the UF Rate must always be programmed.
–300
Test ➀ Press the UF On/Off key to turn Ul-
Conductivity trafiltration off.
mS/cm
Start/Reset 14,5
➀ 7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
with the 4
1
5
2
6
3
keys or the keys.
C Conf
☛
ISO-UF start?
Display message
Prime 15,5
Flow UF
The UF LED is illuminated.
15 On/Off On/Off The dialysate flow is automatically set to
14,5 Off.
☛
Start/Reset
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
On completion of the isolated ultrafil-
tration, dialysis will automatically be
continued with dialysate flow and the
calculated UF parameters.
Conductivity alarm and temperature
alarm are suppressed for 3 minutes.
–300
Test Press the UF Menu key to display the
Conductivity Ultrafiltration menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Use the keys
to move to the UF Profile field.
Enter the desired profile
7 8 9 0
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
–300
Test Press the Dialysate Menu key to display
Conductivity the Dialysate menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Flow off
Display message
Audible signal after 30 minutes.
(adjustable in SETUP: 30, 45, 60
minutes)
–300
Test Press the UF On/Off key to turn the ul-
Conductivity trafiltration on.
mS/cm
Prime 15,5
Flow UF
The UF LED is illuminated.
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
The following values should have been entered before selecting profiles:
– Base Na +, concentrate and bicarbonate.
– UF Goal (minimum UF Goal: 200 ml)
– UF Time (minimum UF profile time: 2 hrs)
– UF Rate:
– minimum continuous UF rate:
Profiles 1 to 6 100 ml/h
– maximum continuous UF rate:
Profile Rate (ml/h)
1 3000
2 3000
3 2660
4 2050
5 2050
6 1250
Sodium
1 2 3 4 5 6
1
Ultrafiltration
The machine will prompt the operator in the event of missing or incorrectly entered data.
–300
Test Press the UF Menu key to display the
Conductivity Ultrafiltration menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Use the keys
to move to the UF Profile field.
Select the desired profile
7 8 9 0
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
–300
Test Press the UF On/Off key to start the pro-
Conductivity file.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the UF Menu key to display the
Conductivity Ultrafiltration menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Use the keys
to move to the UF Profile field.
Enter the desired profile
7 8 9 0
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
–300
Test Press the Dialysate Menu key to display
Conductivity the Dialysate menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
☞ Note
In this mode the machine automat-
ically adjusts the preset UF profile
number to match the selected Na+
profile and informs the operator.
C Conf
☛
–300
Test Press the UF On/Off key to start the pro-
Conductivity file.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
Missing or incorrectly entered
starting parameters are inquired
while entering the profiles.
Profiles, UF and dilution parameters
cannot be reprogrammed while the
program is in progress.
The profile must be stopped and re-
started with the altered parameters.
–300
Test Press the Dialysate Menu key to display
Conductivity the Dialysate menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
C Conf
☛
–300
Test Press the UF On/Off key to start the pro-
Conductivity file.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Dialysate Menu key to display
Conductivity the Dialysate menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
C Conf
☛
–300
Test Press the UF Menu key to display the
Conductivity UF menu.
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
with the 4
1
5
2
6
3
keys or the keys.
The machine will prompt the operator
in the event of missing or incorrectly
entered parameters.
UF-Goal reached
Display message
The yellow status indicator (warning/
info) is illuminated.
The Ultrafiltration LED is dark.
The UF pump has been stopped.
Audible signal
–300
Test The Start/Reset LED is flashing.
Conductivity
mS/cm The Tone Mute LED is flashing.
Prime 15,5
Flow UF The TMP alarm limits are spread fully
15 On/Off On/Off open.
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF The Tone Mute is dark.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Reinfusion?
–300
Test Press the UF On/Off key.
Conductivity
mS/cm The UF LED is flashing.
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
C Conf
☛
Rinsing Diasafe
Temporary display message
Retentate removed by rinsing (approx.
every 60 minutes for 5 balancing cham-
ber switching cycles)
Machine in bypass
Chapter Page
4.9 Error Messages When Using the Download Function ............................................. 4-45
Air detector Blood leak The arterial and/or venous alarm limits
Arterial Venous are not centered.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
New Limit?
Display message
–300
Test Press the Start/Reset key a second
Conductivity time.
mS/cm
If the key is not pressed within approx.
Prime 15,5
Flow UF 8 seconds, the original alarm limits are
15 retained.
On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The arterial and the venous alarm limits
Arterial Venous are centered around the actual value.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Air detector Blood leak The TMP alarm limits are not centered.
Arterial Venous
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
–300
Test Press the UF On/Off
Conductivity key to turn Ultrafiltration off and on
Prime
mS/cm
again.
15,5
Flow UF (This function is not available when
15 On/Off On/Off running a profile)
☛
Start/Reset 14,5
14
2x 7 8 9
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The TMP alarm limits are spread fully
Arterial Venous open.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Air detector Blood leak The TMP alarm limits are centered
Arterial Venous around the actual value.
Pressure Pressure TMP
(time delay depending on the UF
mmHg mmHg mmHg
280 coefficient)
500 500
200
400 400 The position of the TMP alarm window
100
300 300
may automatically track up during the
0 treatment.
200 200
–100
100 100
–200
0 0
–300
Caution
It is important that the lower venous
alarm limit is set to be as close as
possible to the actual value of the
venous pressure.
Caution
When the TMP alarm limits are ad-
justed by the operator, care must be
taken to select a window size as
small as possible to ensure appro-
priate protection.
☞ Note
Automatic adaption of the window
size is now deactivated.
Dialysis
Use the keys
to select the Alarm limits menu.
Press the Conf key.
This function permits correction of the needle without false alarms. The arterial and venous alarm
limits are deactivated for 2 minutes. The full scale values and the lower venous alarm limit of
20 mmHg are still active.
4.2.1 Selection
Dialysis
Use the keys
to select the Alarm limits menu.
Press the Conf key.
Needle-Adj-Pgm
Display message
4.2.2 Deselection
Caution
When overriding a safety system the
responsibility for the patient’s safety
rests with the operator of the ma-
chine.
–300
Test The Start/Reset LED is flashing.
Conductivity
mS/cm The Tone Mute LED is flashing.
Prime 15,5
Flow UF The blood pump(s) stop(s).
15 On/Off On/Off
The venous line clamp closes.
Start/Reset 14,5
7 8 9
14 The UF pump(s) stop(s).
4 5 6
Tone Mute 13,5 Dialysate UF The nurse call is active.
13 Menu Menu 1 2 3
Audible signal
!
e.g. Upper ven. alarm
Upper ven. alarm
!
e.g.: Lower art. alarm
Lower art. alarm
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The alarm limits are flashing for 8
Arterial Venous seconds.
Pressure Pressure TMP
The alarm limits are inactive
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Air detector Blood leak The arterial alarm limits are temporarily
Arterial Venous spread.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Air detector Blood leak The arterial alarm limits are centered
Arterial Venous around the actual value.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Caution
If the filter in the pressure measuring
line is wetted, disconnect the line
and use a syringe to purge the filter
or replace the filter.
!
e.g. Upper ven. alarm
Upper ven. alarm
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF The Tone Mute LED is illuminated.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The alarm limits are flashing for 8
Arterial Venous seconds.
Pressure Pressure TMP
The alarm limits are inactive
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The venous alarm limits are temporarily
Arterial Venous spread.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Air detector Blood leak The venous alarm limits are centered
Arterial Venous around the actual value.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
!
e.g. Upper TMP alarm
Upper TMP alarm
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF The Tone Mute LED is illuminated.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The alarm limits are flashing for some
Arterial Venous time depending on the calculated UFC.
Pressure Pressure TMP
The alarm limits are inactive
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Air detector Blood leak The TMP alarm limits are spread for
Arterial Venous some time, depending on the calculated
Pressure Pressure TMP
mmHg mmHg mmHg
UFC.
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Air detector Blood leak The TMP alarm limits are centered
Arterial Venous around the actual value.
Pressure Pressure TMP
mmHg mmHg mmHg
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Caution
Sudden changes of the TMP are in-
dicative of a defect.
–300
Test Press the Start/Reset key.
Conductivity See also Chapter 4.5 Warnings,
mS/cm Message Fillprogram
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
! Blood leak
[Override with [Confirm] key.
C Conf
☛
! Bloodpump-stop
Rate: ml/min
The Alarm LED (red) on the blood pump
(Ø: mm)
module is illuminated.
E.XX
Error message E.XX on the blood pump
module.
Start
Stop
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
E.XX
☛
Start
Stop
–300
Test Press the Dialysis Start key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
If the problem persists, call the
service.
Rate: ml/h
(Bolus: ml)
The Alarm LED (red) on the heparin
Err
pump is illuminated.
( :h.min)
Message Err displays on the heparin
Bolus pump.
Start
Stop
Rate: ml/h
(Bolus: ml)
Error message E.XX on the heparin
E.XX
pump.
( :h.min)
Bolus
Rate
Start
Stop
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
If the problem persists, call the
service.
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
! Bloodpump-stop
Rate: ml/min
The Alarm LED (red) on the blood pump
(Ø: mm)
is illuminated.
Cause:
The rotor stopped for more than 30
Start
Stop
(15) seconds.
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
! BP-stop
Rate: ml/min
The Alarm LED (red) on the arterial
(Ø: mm)
blood pump or the SN blood pump mod-
ule lights.
Cause:
Start
One of the blood pump rotors has
Stop stopped.
The response time is determined by
the speed.
Start
Stop
Start
Stop
☛
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
The blood pump starts running.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Rate: ml/h
(Bolus: ml)
Rate: ml/min
Use the syringe to drain the fluid from
(Ø: mm)
the compliance chamber while the arter-
( :h.min)
ial blood pump is running.
Bolus
Start
Rate Stop
Start
Stop
Rate: ml/min
Press the Start/Stop key to stop the ar-
(Ø: mm)
terial blood pump.
The Operation LED is dark.
Start
Stop
☛
4-22 Fresenius Medical Care 4008 S 11/02.01 (OP)
Rate: ml/h
(Bolus: ml)
Rate: ml/min
➀ Clamp the SN pressure measuring
(Ø: mm)
line before the hydrophobic filter.
( :h.min)
Bolus
➁ Replace the hydrophobic filter.
Start
Rate
➁ Stop
Start
Stop
➀
Start
Rate ➀ Stop
Start
Stop
Rate: ml/min
Use the ▲ and ▼ key to set the delivery
(Ø: min)
rate for the blood pump.
Press the Start/Stop key to start the ar-
terial blood pump.
Start
Stop
The Operation LED is illuminated.
☛
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
The Start/Reset LED is illuminated.
Prime 15,5
Flow UF The Tone Mute LED is dark.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
! Cycle alarm
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF The Tone Mute LED is dark.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test The Tone Mute LED is flashing.
Conductivity
mS/cm Nurse call active.
Prime 15,5
Flow UF Audible signal
15 On/Off On/Off
Flow indicator (external): the cone does
Start/Reset 14,5
7 8 9 not move.
14
13,5 Dialysate
4 5 6 The alarm message and the effect of the
Tone Mute UF
alarm are automatically reset when the
13 Menu Menu 1 2 3
problem causing the alarm has been cor-
rected.
The audible alarm can be suppressed for
1 minute by pressing the Tone Mute key
(factory setting).
Conductivity
mS/cm
The conductivity alarm indicator is
15,5 illuminated.
Flow
15 On/Off The machine is in bypass mode.
14,5
Cause:
14 The actual conductivity value is outside
13,5 Dialysate the alarm limits.
13 Menu
Remedy:
Verify the concentrate side.
Correct the problem causing the alarm.
or Remedy:
Wait for the temperature to reach the
Warning desired value or call the service.
or
Messages will be displayed in the text field or in a pop-up window. The messages are either
warnings or information messages.
Example:
Cyclic PHT F04 Blood alarm present Press the Approx. Approx.
for more than Start/Reset key 15 min 15 min
3 minutes. to acknowledge.
or or
Failure to perform Acknowledge by
the cyclic pressure pressing the
holding test (system Start/Reset key,
error). considering a
possible balancing
error.
Flow off warning The dialysate flow Press the Start/ 30 min 30 min
has been switched Reset key (unless 45 min 45 min
off. flow is desired at the 60 min 60 min
moment), or switch (SETUP) (SETUP)
the dialysate flow on.
Heparin pump not on The heparin pump Press the Start/ Approx. –
module is not on. Reset key (no 11 min
heparinization), or
switch the heparin
pump on.
Shunt Cover open The shunt cover has Close the shunt 1 min –
been opened. cover.
Emergency Operation
Display message
–300
Test The Start/Reset LED is illuminated.
Conductivity
mS/cm
The Tone Mute LED is flashing.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
13,5 Dialysate
4 5 6 No audible signal
Tone Mute UF
13 Menu Menu 1 2 3 Emergency operation is limited to
approx. 15 minutes.
☛
Caution
In the event of a prolonged power
failure, use the pump handle to ma-
nually return the blood to the patient.
For manual reinfusion, remove the
venous patient line from the venous
line clamp. Visually check the line for
air (emergency operation)!
In Single Needle mode, also remove
the BP segment for the BP (SN) from
the blood pump!
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Arterial pressure Arterial pressure outside the Check the line system.
(only while priming/ alarm limits. Connect the line system.
precirculation is in pro- Purge the hydrophobic filter.
gress during the clean- Increase the blood pump
ing programs) speed (>100 ml/min).
Call the service.
Bic line not conn The bicarbonate connector is Insert the bicarbonate connec-
not in the rinse port. tor in the rinse port.
Blood Sensed The air detector detects blood Check the optical detector.
by OD at the start of a cleaning pro- Call the service.
gram.
Conc line not conn The concentrate connector is Insert the concentrate connec-
not in the rinse port. tor in the rinse port.
Disinfectant empty ? Float switch does not detect Check the disinfectant con-
fluid. tainer.
Disinfection valve sensor 115 Can be acknowledged by
does not sense any conducti- pressing the Disinfection key.
vity. Call the service.
Concentrate float switch fails
to detect fluid.
Internal Error XXX Run time error in the program. Press the On/Off switch to turn
the machine off and on again.
Call the service.
Level Detect. failed Air detector test failed. Problem cannot be corrected
(only while priming/ in the cleaning program
precirculation is in pro- (precirculation impossible).
gress during the clean- After cleaning: perform a func-
ing programs) tional test.
Call the service.
opt. Detect. failed Optical detector test failed. Problem cannot be corrected
(only while priming/ in the cleaning program
precirculation is in pro- (precirculation impossible).
gress during the clean- After cleaning: perform a func-
ing programs) tional test.
Call the service.
Power Failure Power failure during operation. After power has returned, the
resp. program can be contin-
ued as usual.
Rinse Failure F02 V84 error. Press the On/Off switch to turn
In disinfection program the machine off and on again.
(PGM 5): Call the service.
Concentrate level sensor fails In disinfection program
to detect fluid. (PGM 8): Connect disinfectant
Rinse Failure F04 The float switch does not Press the On/Off switch to turn
detect fluid. the machine off and on again.
Call the service.
Rinse Failure F05 PSW 124 does not open. Call the service.
Observe the pressure exerted
on the distribution piping (no
more than 500 mbar).
Rinse Failure F06 PSW 123 does not open. Call the service.
Observe the pressure exerted
on the distribution piping (no
more than 500 mbar).
Rinse Failure F07* Pressure drop on PSW 124. Call the service.
Rinse Failure F08* Pressure drop on PSW 123. Call the service.
* This error message may be indicated alternately with the message DO NOT SWITCH OFF !!
Wait for 1 min. before turning the machine off!
Rinse Failure F09 PSW 123/124 does not open Call the service.
after pressure reduction.
Rinse Failure F11 3 min before end of mandatory Call the service.
rinse PSW 123 has not opened
after pressure reduction.
Rinse Failure F12 Failure to open PSW 123/124. Call the service.
Rinse Failure F14 Flow problems in the rinsing- Call the service.
area test.
Rinse Failure F21 Maximum number of strokes Press the I/O switch to toggle
for drawing in disinfectant the machine off and on again.
exceeded. Replace the disinfectant con-
tainer.
Call the service.
Rinse required! The mandatory rinse was Select the rinse program.
stopped.
Shunt cover defect Dialyzer lines connected to Change the blood lines.
alternating with dialyzer although the hemo- Check the dialyzer couplings.
Change blood lines! dialysis machine is in the Check the interlock shunt.
(only with priming/ cleaning mode. (If cleaning can be started al-
precirculation is in though the two dialyzer coup-
progress during the lings are not in the interlock
cleaning programs) shunt: call the service.)
Upper Flow Alarm The cleaning flow rises above Can be acknowledged by press-
1000 ml/min. ing the respective cleaning pro-
gram key.
Call the service.
V91/V100 Failure The valve does not open. Call the service.
V99 Failure The valve does not open. Call the service.
Venous pressure? Venous pressure outside the Check the line set.
(only while priming/ alarm limits. Connect the line set.
precirculation is in Purge the hydrophobic filter.
progress during the Increase the blood pump
cleaning programs) speed (>100 ml/min).
Call the service.
Voltage Failure One of the voltages is outside Acknowledge with the respec-
the limits (+12 V, +24 V). tive cleaning program key.
Press the On/Off switch to turn
the machine off and on again.
Call the service.
Water alarm Float switch for t>30 s high. Check the water supply.
Float switch for t>10 s low. Cleared automatically.
Call the service.
Audible Signal
The red status indicator is flashing.
Cause:
Bios Test passed Dual flashing rate: EPROM error
ROM Test passed Triple flashing rate: RAM error
RAM Test passed Quadruple flashing rate: NOVRAM error
Possible solution:
Call the service.
Keyboard Error Key on the monitor pressed Press the ON/Off key to turn
upon power up. the machine off and on.
Short-circuit on the keyboard. Call the service.
Watchdog Error This error message will only Press the ON/Off key to turn
be displayed immediately the machine off and on.
after turning power on. Call the service.
Download Error F001 Optical detector does not sense Ensure basic requirements are
clear fluid. fulfilled. Restart data transfer.
Download Error F002 Data transfer disabled or im- Call the service.
possible.
(DIP switch 1 LP763 set to OFF)
Download Error F003 Patient key not connected. Connect the patient key.
Restart data transfer, if
necessary.
Download Error F004 CRC check for communication Call the service.
inactive.
Download Error F007 Not in Preparation mode. Ensure basic requirements are
fulfilled. Restart data transfer.
Download Error F008 Illegal patient name Enter the patient name.
(19 blanks) Restart data transfer.
Download Error F011 Illegal UF or Na+ profile number. Check external data.
Restart data transfer.
Download Error F013 Value range for Base Na+ or Check external data.
concentrate exceeded or Restart data transfer.
plausibility ±13 mmol.
Download Error F014 Value range for Start Na+ Check external data.
exceeded. Restart data transfer.
Download Error F015 Value range for bicarbonate Check external data.
exceeded. Restart data transfer.
Download Error F019 Start Na+ plausibility check. Check external data.
Restart data transfer.
Download Error F020 Block check UF rate, UF goal, Check external data.
UF time, UF profile. Restart data transfer.
Na+ profile, Base Na+, concen-,
trate, start Na+.
Download Error F022 Na+ profile data plausibility. Check external data.
Restart data transfer.
Download Error F023 Invalid CRC of the patient key. Check patient key.
Restart data transfer.
Chapter Page
UF Goal reached
Display message, e.g.: UF goal reached
Fresenius Medical Care 4008 S The TMP alarm limits are spread fully
open.
On/Off Air detector Blood leak
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF Turn the heparin pump off, if necessary.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Bolus
Start
Rate Stop
☛
Rate: ml/h
(Bolus: ml)
Rate: ml/min
(Ø: mm)
➀ Open the cover.
( :h.min)
➁ Press the Start/Stop key and keep it
Bolus
pressed until the rotor stops.
Start
Rate Stop
Start
Stop
☛
➁
Rate: ml/h
(Bolus: ml)
Rate: ml/min
(Ø: mm)
➀ Remove the line segment on the left.
( :h.min)
➁ Press the Start/Stop key and keep it
Bolus
pressed until the line set has been
Start
Rate Stop completely removed.
Start
Stop
☛
➁
➂ Remove the line segment on the
right.
➀ ➂
C Conf
☛
Reinfusion
Display message
Air detector Blood leak The arterial, venous and TMP alarm
Arterial Venous limits are spread fully open for 5
Pressure Pressure TMP
mmHg mmHg mmHg
minutes.
280 500 500
200
400 400
100
300 300
0
200 200
–100
100 100
–200
0 0
–300
Start
Stop
☛
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
Reinfusion starts (max. 5 minutes).
Prime 15,5
Flow UF The Start/Reset LED is illuminated.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Dialysis end
Display message
The optical detector senses clear fluid.
The blood pump has stopped.
The venous line clamp is closed.
Audible signal
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
The Start/Reset LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Preparation
Display message
The blood pump starts.
The blood still present in the line set is
reinfused to the patient until an air detec-
tor alarm occurs.
(Audible alarms suppressed.)
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
C Conf
☛
Reinfusion?
Display message
–300
Test Press the Start/Reset to restart
Conductivity Reinfusion.
mS/cm
Prime 15,5
Flow UF
The Start/Reset LED is illuminated.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the Start/Reset to skip
Conductivity Reinfusion.
mS/cm
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Dialysis
Display message
C Conf
☛
Dialysis
Display message
–300
Test Press the UF On/Off key to turn the ul-
Conductivity trafiltration off.
mS/cm
Start/Reset 14,5
☛ 7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Rate: ml/min
Turn the blood pump off.
(Ø: mm)
Rate: ml/h
(Bolus: ml)
Turn the heparin pump off.
The Operation LED is dark.
( :h.min)
Bolus
Rate
Start
Stop
☛
Rate: ml/min
Turn on the blood pump with reduced
(Ø: mm)
blood flow.
The Operation LED is illuminated.
Start
Stop
☛
–300
Test The Start/Reset LED is flashing.
Conductivity
mS/cm
The Tone Mute LED is flashing.
Prime 15,5
Flow UF Use the ▲ and ▼ keys to select
15 On/Off On/Off Cancel treatment.
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
C Conf
☛
or
–300
Test Leave the cursor bar on the highlighted
Conductivity Continue treatment option (green).
mS/cm
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
Test Press the UF On/Off key to turn the ul-
Conductivity trafiltration off.
mS/cm
Prime 15,5
Flow UF
The UF LED stops lighting.
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
C Conf
☛
Reinfusion
Display message
The blood pump stops.
The venous line clamp closes.
–300
Test The Start/Reset LED is illuminated.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Rate: ml/min
Press the ▼ key to reduce the delivery
(Ø: mm)
rate of the blood pump.
Start
Stop
☛
( :h.min)
The Operation LED is dark.
Bolus
Rate
Start
Stop
☛
–300
Test Press the Start/Reset key.
Conductivity
mS/cm
The Start/Reset key is illuminated.
Prime 15,5
Flow UF The blood is returned to the patient.
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Dialysis end
The optical detector senses clear fluid.
Display message
The blood pump has stopped.
The venous line clamp is closed.
Audible signal
Empty the bibag® (option).
See Chapter 5.4 Emptying the bibag®
(option).
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Dialysis End
Display message
–300
Test Press the Start/Reset key.
Conductivity
mS/cm The Start/Reset LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
C Conf
☛
Emptying BIBAG
Display message
Drain program in progress
–300
Test The bibag ® drain program can be
Conductivity skipped by pressing the Start/Reset key
Prime
mS/cm
again.
15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
☛
☞ Note
Always clean the sealing area of the
connector after dialysis to remove
bicarbonate precipitate.
Emptying program
Display message
Caution
The machine should be disinfected
after each dialysis treatment.
Caution
Applicable for DIASAFE®plus (option)
only:
It is imperative to disinfect the ma-
chine after each dialysis treatment.
☞ Note
Applicable for DIASAFE®plus (option)
only:
If the machine has been idle for more
than 72 hours, it is recommended to
run a disinfection program before ini-
tiating the next treatment.
Chapter Page
Display
(during cleaning
program)
–R–*
–F–HR–C–
–F–HR–
–IHR–
–F–D–M–
–F–HDIS–M–
–F–D–M–HR–
–F–HDIS–M–HR–
–F–D(F)–M–
Indication of the disinfections
still available
Filter change program
(applicable for advanced hydraulics only)
Indication of
date, time and program
of the last disinfection
that was fully completed (arrow)
The part of the cleaning program currently in progress is displayed as follows: > < (e.g. >F<HDIS–
M–, free rinsing in progress).
☞ Note
The count-down times shown during
the cleaning programs are calculated
values. This time may vary depend-
ing on the ambient conditions.
☞ Note
Applicable for machines with
DIASAFE®plus (option) and advanced
hydraulics only:
When pressing the Cleaning key it
will be displayed in the Cleaning
menu when the filter needs to be
changed. (See Chapter 7.2.1 Indica-
tion of a Filter Change.)
After treatment, the exterior of the hemodialysis machine should be cleaned with a cleaning and
disinfecting agent with an active substance concentration <5 % for amphoteric surfactants and
<48 % for alcohol. (See also Chapter 1.10 Disposables and Consumables.)
Any failure to comply with the basic conditions will be indicated by a message on the display.
Caution
Applicable for DIASAFE®plus (option)
only:
Contact of the filter with organic sol-
vents can affect the properties of the
material of housing, seals and capil-
laries. Safe function can then no lon-
ger be ensured and the manufacturer
shall no longer have any liability.
(See DIASAFE®plus package insert.)
Factory setting: Citrosteril® and Diasteril® respectively. For use of a different disinfectant, this
setting must be changed by a technician.
Caution
During the heat disinfection cycle,
hot water or steam may be emitted
via the vent tubing.
Caution
Applicable for DIASAFE®plus (option)
only:
High surface temperatures on the
DIASAFE®plus during heat disinfec-
tion.
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
400 400
☛
SN 100
300 300
0
200 200
Caution
After disinfection with Puristeril®
check the machine for residual disin-
fectant after completion of the man-
datory rinse cycle (e.g. with potassi-
um iodide paper).
Caution
After a disinfection with Diasteril®,
the following must be observed:
After termination of the rinse pro-
gram, (mandatory rinse) the hemo-
dialysis machine must be checked
for residual disinfectant using the
pH test.
If the pH is too high (yellow color
pH ≤ 4.1), run a rinse program for a
minimum of 5 min., then re-deter-
mine the pH.
Caution
Applicable for DIASAFE®plus (option)
only:
Do not stop a disinfection program
before its completion. Premature ter-
mination will compromise the efficacy
of the disinfection.
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
400 400
☛
SN 100
300 300
0
200 200
C Conf
☛
☞ Note
Do not remove the concentrate and
bicarbonate suction tubes from their
ports to prevent leakage of fluid
which might then enter the interior
portions of the machine.
In this cleaning program the cleaning agent (Sporotal®) is drawn in from the front via the
concentrate suction tube.
☞ Note
Applicable for DIASAFE®plus (option) only:
After installation, the DIASAFE®plus may be cleaned with Sporotal® 100 for a
maximum of 11 times.
In machines with basic hydraulics the cleaning procedures with Sporotal® 100
must be documented.
In machines with advanced hydraulics the cleaning procedures with Sporotal®
100 are monitored by the hemodialysis machine.
(See Chapter 7.2.1 Indication of a Filter Change.)
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
400 400
☛
SN 100
300 300
0
200 200
Rate: ml/h
( :h.min)
Bolus
Rate
Start
Stop
the container with the cleaning agent
(Sporotal® 100).
Press the Conf key.
Cleaning
agent
Caution
Assure that the container used con-
tains sufficient cleaning agent to
complete the rinse program.
Avoid aspiration of air.
Caution
Check the machine for residual disin-
fectant (e.g. with potassium iodide
paper)
Caution
Applicable for DIASAFE®plus (option)
only:
If the test shows a residual concen-
tration, the respective rinse program
must be restarted.
(In this case it is not necessary to run
the entire rinse program.)
Caution
During the heat disinfection cycle,
hot water or steam may be emitted
via the vent tubing.
Caution
Applicable for DIASAFE®plus (option)
only:
High surface temperatures on the
DIASAFE®plus during hot rinse pro-
grams!
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
☛
400 400
SN 100
300 300
0
200 200
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
☛
400 400
SN 100
300 300
0
200 200
C Conf
☛
☞ Note
Do not remove the concentrate and
bicarbonate suction tubes from their
ports to prevent leakage of fluid
which might then enter the interior
portions of the machine.
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
400 400
☛
SN 100
300 300
0
200 200
Arterial Venous
Pressure Pressure TMP
Cleaning mmHg mmHg mmHg
280 500 500
200
☛
400 400
SN 100
300 300
0
200 200
C Conf
☛
Please wait
☞ Note
Do not remove the concentrate and
bicarbonate suction tubes from their
ports to prevent leakage of fluid
which might then enter the interior
portions of the machine.
–300
Test Press the Prime key for 1 second.
Conductivity
mS/cm
The Prime LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
Prime
Temporary display message
Rinse-volume: xxxx ml
Temporary display message if a rinse
volume has been defined in the SETUP.
>R<
Display message, e.g.: >R<
Rate: ml/min
The Operation LED (green) is
(Ø: mm)
illuminated.
The blood pump is delivering.
Start
☛ The arterial blood pump fills the line set
up to the venous bubble catcher.
Stop A delivery rate > 180 ml/min will auto-
matically be reset to 180 ml/min. The
delivery rate can be changed by the
operator as desired.
>R<
Display message
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
–300
Test Press the Prime key for 1 second.
Conductivity
mS/cm
Press the Prime key again to continue
Prime 15,5
Flow UF
priming.
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Prime 15,5
Flow UF
15 On/Off On/Off
☛
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
The dialysate couplings must be in
the interlock shunt.
6.9.2 Preselection
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Prime/Circulation
Temporary display message
Precirculation will be started automat-
ically after priming has been completed.
>R<
Display message e.g.: >R<
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Circulation
Temporary display message
>R<
Display message e.g.: >R<
Rate: ml/min
The operation (green) LED is illumin-
(Ø: mm)
ated.
The blood pump is delivering.
☛ Use the ▲ or ▼ key to set the desired
Start
delivery rate.
Stop (Pressing the key for more than 3 sec-
onds will accelerate the change rate.)
>R<
Display message
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
☛
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
Chapter Page
7.1 Changing the DIASAFE ®plus (option) on Machines with Basic Hydraulics ............. 7-3
7.1.1 General Notes ................................................................................................................. 7-3
7.1.2 Filter Change Cirteria ...................................................................................................... 7-3
7.1.3 Changing the Filter ......................................................................................................... 7-3
7.1.4 Priming a Filter ................................................................................................................ 7-5
7.2 Changing the DIASAFE ®plus (option) on Machines with Advanced Hydraulics ...... 7-7
7.2.1 Indication of a Filter Change ........................................................................................... 7-7
7.2.2 Overriding a Filter Change .............................................................................................. 7-11
7.2.3 Changing the Filter ......................................................................................................... 7-12
☞ Note
Replacement of the DIASAFE®plus
must be entered in the Medical De-
vice Register (date, batch number).
☛
On/Off Air detector Blood leak
The On/Off LED stops lighting.
Arterial Venous
Pressure Pressure TMP
☛
On/Off Air detector Blood leak
Arterial Venous
The On/Off LED is illuminated.
Pressure Pressure TMP
–300
Test The Test LED is flashing.
Conductivity If the Test is not flashing, one of the
mS/cm conditions for the test to start is not ful-
Prime 15,5 filled.
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
–300
☛
Test Press the Test key.
Conductivity
mS/cm The Test LED is illuminated.
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☞ Note
If the machine fails to pass the
functional test, repeat the test by
pressing the Test key.
If a DIASAFE®plus failure (FXX) is
displayed after the functional test,
this test step has to be skipped.
Then run a heat disinfection program
or a disinfection program followed by
a hot rinse.
● General Notes
☞ Note
Replacement of the DIASAFE® plus
must be entered in the Medical De-
vice Register (date, batch number).
● Basic conditions
☞ Note
A filter change may never be prema-
turely terminated, but must always
be properly completed.
400 400
SN 100
300 300
0
200 200
C Conf
☛
C Conf
☛
Filling filter
Cleaning menu
Disinfect as specified in the Operating
Instructions.
Caution
Run a disinfection program after
each filter change.
–300
Test Press the Dialysate Menu key.
Conductivity
mS/cm
Prime 15,5
Flow UF
15 On/Off On/Off
Start/Reset 14,5
7 8 9
14
4 5 6
Tone Mute 13,5 Dialysate UF
13 Menu Menu 1 2 3
☛
Dialysis
Use the keys
to move to the Empty BIBAG? field.
Use the keys to select Yes.
C Conf
☛
☞ Note
Do not remove the bicarbonate suc-
tion tube from its port to prevent lea-
kage of fluid which might then enter
the interior portions of the machine.
☛
☛
Caution
Wait until the cooling rinse program
following a hot rinse or a heat desin-
fection has been completed before
connecting the bibag®.
7.4.1 Description
The function to transfer (download) treatment parameters allows transfer of patient-related treat-
ment parameters to the dialysis machine.
The following treatment parameters are eligible for transfer:
An external system (data source) can be used to transfer the treatment parameters.
The treatment parameters downloaded to the dialysis machine must be accepted in total. It is not
possible to accept individual parameters only.
☞ Note
After the treatment parameters have been transferred, the downloaded data can
only be accepted on the dialysis machine, provided:
the UF unit has been turned off.
the UF volume has been set to 0000.
Caution
The data received must be verified by the user, i.e. the user must check the data
received to ensure they are correct. Only then may the data be accepted with the
Conf key.
C Conf
☛
C Conf
☛
C Conf
☛
Harold
Sample Preparation
Display message
The patient's name is displayed on the
left.
Displayed when all downloaded data
have been saved.
Menu
e.g. Ultrafiltration menu
Esc
Press the Esc key.
☛
C Conf
C Conf
☛