1.1. PURPOSE OF THE SERVICE MANUAL
This manual includes servicing and maintenance information for MD series blood bank
refrigerators. It is prepared to be used by technicians who were formerly trained by
Nuve. This manual informs the technicians about the operating principles, diagnosing
and repairing methods and spare part replacing.
In the case that any problem which is not identified in this manual exists, please contact
Nuve.
Factory
Esenboga Yolu 22. Km.
Akyurt 06287 ANKARA TURKEY
Tel: 00 90 312 399 28 30 (3 lines)
Fax: 00 90 312 399 21 97
e-mail: service@nuve.com.tr
Control units are located at the top of the instrument. The air inside is sucked by the fan
that is located at the top of the chamber then passes through the evaporator that is
located at the top, between the chamber and the body and again enters to the chamber
from the openings at the back side of the chamber. The cooling function and the air
circulation are provided by this way. The MD series kit storage refrigerators are
microprocessor controlled units.
3
SECTION 2
OPERATING PRINCIPLES
The MD Series Kit Storage Refrigerators may be split into 4 main groups;
• Power supplies
• Control unit
• Cooling unit
• Lightening unit
Power
Glass Fuse Power inlet
Consumption
MD 72 10 A 330 W 230 V AC
The microprocessor controlled main PCB operates on the “On/Off” controlling system and all
excluding the circular chart recorder all the control elements are connected to the main PCB.
These are;
• Temperature sensor
• Lightening unit terminals
• Keyboard and display PCB
• Cooling group terminals
• Circulation motor terminals
• Remote alarm and NC contact inlets
4
2.2.3. Control Panel
Figure 1
1. Temperature Display
2. Cooling Led
3. Door Open Led
4. Value Increase and Decrease Keys
5. Alarm Led
6. Alarm Mute Button
7. Lighting Button
8. Set Button
9. Open/close Key Lock
5
2.2.4. Cooling Unit
The cooling function is obtained by circulating the refrigerant gas within a close circuit system.
The elements of this close circuit system are;
• Compressor
• Condenser Cooling Fan
• Condenser
• Dryer
• Capillary Pipe
• Evaporator
Figure 2
Samples can easily be observed by the lightening lamp inside the chamber. The lightening unit
elements are connected to the main PCB. They are;
• Lightening switch
• Ballast
• Starter
• Lightening lamp
In case of cut of electricity chamber temperature and the other displays are kept working by the
aid of a battery which is charged automatically. Battery charging elements are;
6
SECTION 3
SERVICING
Caution: Before servicing the instrument, please take the necessary precautions both
for your health and environmental safety. Please respect the warnings on the unit!!
The components on the main PCB must not be replaced even the failure is caused by one of
the components on the main PCB. In this case, please send the failed PCB to factory service
along with a note on which the failure explanations are written.
Before replacing the PCB or any control element, please make sure that the failure is not
caused by weak wire and terminal connections.
7
3.2. GENERAL FAILURE
1 The on / off switch is on but the Power supply inlet failure. Check that mains voltage is
power led does not lighten and supplied.
the display is blank. Check that the terminals and
power supply sockets are well
connected.
The glass fuse has blown. Replace the glass fuse and
check the short circuit that
causes the blowing of the fuse.
2 The on / off switch is on and the The main PCB / display PCB Remove this connection cable
power led lightens but the display connection cable is not and replace it carefully.
is blank or some segments are connected properly.
blank.
The main PCB / display Replace the connection cable.
connection cable is defective.
3 The unit blows fuse frequently. Short-circuit exists. Check the electrical terminals
and
cables for short circuit.
4 The circulation motor does not The motor cable connection Check these terminals if they
work. terminals are loose. are connected properly.
5 The lightening lamp does not turn Lightening lamp is defective. Replace the lamp.
on.
8
FAILURE PROBABLE CAUSES SOLUTION
7 The compressor does not work. The compressor cable Check the cable connection terminals
connection terminals are and ensure that they are connected
loose or broken. properly.
1 Temperature does not reach There is excessive heat Check that the chamber gasket is tight
the set value. loss. enough.
2 Actual temperature inside Temperature sensor reads Replace the temperature sensor.
differs from the displayed one. incorrect values.
9
3.4. COOLING SYSTEM FAILURES
CAUTION: Cooling system failures should only be handled by the trained and
experienced technicians.
• Connect the maximum pressure hose of the gas manifold to the vacuum pump, the
minimum pressure hose to the compressor service line and the gas charge hose to gas
charge cylinder.
• Open the minimum and maximum pressure valves.
• Operate the vacuum pump and vacuum the system for at least 60 minutes.
• Stop the vacuum pump after closing the maximum pressure valve.
10
Figure 3
11
SECTION 4
CALIBRATION PROCEDURE
4.1 HOW TO MEASURE THE ACTUAL TEMPERATURE
• Insert the calibrated external sensor in the chamber from the calibration probe enterance
at the left side of the unit.
• Start the program.
• Please wait approximately 4 hours (after the set temperature) before measuring the actual
temperature by the external temperature sensor.
• Be sure that calibrated external sensor does not touch any metal part or any of the blood
bags, it measures only the air temperature.
• If there is difference between the equipment’s display and the external sensor apply the
calibration procedure as below:
4.2 CALIBRATION
• Press “value increase & decrease buttons” together continuosly until you see “tEC” on the
display.
• Press “SET” and you see “tCP” on the display.
• Press “value increase button” to enter, “000” starts to flash.
• Make this value “211” by using “value increase button”.
• Press “SET” and you see “PrG / Con” flashing alternately.
• Press “SET” and you see “P00” on the display.
• Press “SET” with small intervals until you see “P03” on the display.
• Press “value increase button” to enter and you see a value on the display.
• Change the value that appears on the display according to the difference between the
equipment’s display and the external sensor.
For example, if the set temperature is 4°C and the external sensor measures 7°C, increase
the value, which appears on the display, by adding 3; if the set temperature is 4°C and the
external sensor measures 0°C, decrease the value, which appears on the display, by
subtracting 4 by using “value increase or decrease keys”.
• Press “SET” to validate and you see “P04” on the display.
• Press “SET” with small intervals until you see “PrG / Con” flashing alternately.
• Then press “alarm mute button” to return to the stand-by mode.
12
SECTION 5
SPARE PART REPLACING
CAUTION: Unplug the refrigerator before replacing any part!!
• Disconnect all cable terminals which are connected to the main PCB.
• Unplug the display PCB / main PCB connection cable ending from the main PCB.
• Remove the 4 connection clamps and then the PCB.
• Place a new PCB on the connection clamps and make the terminal connections according
to the electric diagram provided in Section 6.
• Plug the display PCB / main PCB connection cable.
• Disconnect the display PCB / main PCB connection cable from the display PCB.
• Remove the 4 connection screws and remove the display PCB.
• Assemble a new PCB.
• Check that the display keys and the led lamps meet their places on the panel.
13
5.7. REPLACING MAGNETIC GASKET
14
SECTION 6
15
6.1. ELECTRICAL CIRCUIT DIAGRAM
MD 72 - 120
16
MD 294
17
MD 504
18
SECTION 7
19
20
21
22
7.1 Spare Parts List