The document provides instructions for calibrating and testing an IB300 device using a calibration tool application. It outlines steps to install the app, calibrate the weight and body composition functions, run tests on weight and body composition, and then remove the app. Key steps include placing known weights on the device during weight calibration, waiting 7 minutes for body composition calibration to complete, and checking for PASS/FAIL results during body composition testing.
The document provides instructions for calibrating and testing an IB300 device using a calibration tool application. It outlines steps to install the app, calibrate the weight and body composition functions, run tests on weight and body composition, and then remove the app. Key steps include placing known weights on the device during weight calibration, waiting 7 minutes for body composition calibration to complete, and checking for PASS/FAIL results during body composition testing.
Original Description:
Guideline for calibrating inbalance 300 using calibration tool
The document provides instructions for calibrating and testing an IB300 device using a calibration tool application. It outlines steps to install the app, calibrate the weight and body composition functions, run tests on weight and body composition, and then remove the app. Key steps include placing known weights on the device during weight calibration, waiting 7 minutes for body composition calibration to complete, and checking for PASS/FAIL results during body composition testing.
The document provides instructions for calibrating and testing an IB300 device using a calibration tool application. It outlines steps to install the app, calibrate the weight and body composition functions, run tests on weight and body composition, and then remove the app. Key steps include placing known weights on the device during weight calibration, waiting 7 minutes for body composition calibration to complete, and checking for PASS/FAIL results during body composition testing.
2. Weight Calibration 3. Weight Test 4. Body Composition Calibration 5. Body Composition Test 6. Remove IB300 Caliration Tool Application 1. Install IB300 Tool app
• Install IB300 Tool App using USB
Storage. 2. Weight Calibration
• Click the weight calibration button
on the main screen. 2. Weight Calibration
• Place a 20 kg weight on the device.
2. Weight Calibration
• Press the start measurement button
and wait for the popup to appear. • Press the OK button and increase the weight for further measurements. (Minimum 60kg recommended) • The more additional measurements you have, the more accurate your scale measurement. 3. Weight Test
• On the Main screen, click Weight
Test. • Put the weight to be measured on the device. • Click the Start Measurement button. 3. Weight Test
• Check that the correct value is
measured on the screen. • If error occurs, recalibrate the scale. 4. Body Composition Calibration
• Click the Body Composition
Calibration button on the Main screen. 4. Body Composition Calibration
• Click the Calibration button.
• If it fails, restart the Tool device and click the Start button. 4. Body Composition Calibration
• Wait until the body composition
calibration measurement is completed. It takes about 7 minutes. 4. Body Composition Calibration
• After successful calibration of the
body composition, click the Apply Calibration button. 5. Body Composition Test
• Click Body Composition Test on the
Main screen. 5. Body Composition Test
• Click the Measure button and wait
for the measurement to complete. It takes about 2-5 minutes. 5. Body Composition Test
• After the measurement, check if
there is a Fail on each impedance value. • If it says PASS, then the body composition Calibraion is perfectly measured. 6. Remove IB300 Tool App
• Once all scale calibration and
body composition calibration are complete, delete the app. • On the Main screen, click the Delete App button at the bottom and press OK.