Professional Documents
Culture Documents
T20-A User Manual
T20-A User Manual
O P E R AT I O N TA B L E
111704 T-IM102d
Read these Instructions before use
Preliminary
Information
Keep these ‘Instructions for use’ in a safe convenient place for future reference. Read in
Safety notes conjunction with the Publications detailed in Section 3.2.9.
and cautions
All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :
UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
Tel: +44 (0) 1903 765040. Fax: +44 (0) 1903 762006.
Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.
Copyright © 2003
All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a
retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying,
recording or otherwise without written permission from Eschmann Holdings Limited.
The information in this publication was correct at the time of going to print. The Company, however,
reserves the right to modify or improve the equipment referred to.
If the CE mark is affixed to the product, it indicates compliance with Council Directive
93/42/EEC of 14 June 1993 concerning medical devices.
CONTENTS
1.0 Preliminary information 4.0 Operation (continued)
1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 4.7 Fuse replacement . . . . . . . . . . . . . . . . . . . . . . 24
1.2 About this manual . . . . . . . . . . . . . . . . . . . . . . . 5 4.8 Using accessories . . . . . . . . . . . . . . . . . . . . . . 25
1.3 Part identification . . . . . . . . . . . . . . . . . . . . . . . . 8 4.9 Patient positioning . . . . . . . . . . . . . . . . . . . . . . 27
1.4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 4.10 Treatment of the obese patient . . . . . . . . . . . . 27
1.5 Symbols and graphics . . . . . . . . . . . . . . . . . . . . 8 4.11 Radiographic procedures . . . . . . . . . . . . . . . . 27
Symbols general . . . . . . . . . . . . . . . . . . . . . . 8 4.12 After use procedures . . . . . . . . . . . . . . . . . . . . 27
Handset button symbols . . . . . . . . . . . . . . . . 9 4.13 Environmental considerations . . . . . . . . . . . . . 28
Handset graphics . . . . . . . . . . . . . . . . . . . . . 9 4.14 Table positions and safe loading . . . . . . . . . . . 28
Standby control panel button symbols . . . . 10
1.6 Interference considerations . . . . . . . . . . . . . . . 10 5.0 Cleaning, disinfection,
Interference from other equipment . . . . . . . 10 care & maintenance
Interference with other equipment . . . . . . . 10
Use of infrared . . . . . . . . . . . . . . . . . . . . . . 10 5.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
1.7 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Advice . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Visual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Operation table and accessories . . . . . . . . 37
Audible . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Mattresses and pads . . . . . . . . . . . . . . . . . 37
5.2 Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2.0 Safety notes & cautions Disinfection procedure . . . . . . . . . . . . . . . . 37
Table, accessories mattresses and pads . . 38
2.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 5.3 Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
2.2 Do’s and Don’ts . . . . . . . . . . . . . . . . . . . . . . . . 12 Operation table and sections . . . . . . . . . . . 38
2.3 Daily ‘Before use’ test . . . . . . . . . . . . . . . . . . . 12 Storing the operation table (long term) . . . . 38
2.4 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Lifting the operation table . . . . . . . . . . . . . . 40
5.4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.0 Description General . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Fault diagnosis . . . . . . . . . . . . . . . . . . . . . . 40
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Technical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.0 Technical data
6.1 Weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.0 Operation 6.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3 Movements . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.4 Table loading . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.2 Moving and braking table . . . . . . . . . . . . . . . . 14
6.5 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
General . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Antistatic requirements . . . . . . . . . . . . . . . . 43
‘Wheel’ orientation (from ‘braked’) . . . . . . . 15
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
‘Castor’ orientation (from ‘wheel’) . . . . . . . . 15
Internal battery charger . . . . . . . . . . . . . . . 43
‘Braked’ (from ‘wheel’ or ‘castor’) . . . . . . . . 16
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
‘Castor’ orientation (from ‘braked’) . . . . . . . 17
Duty cycle . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.3 Attaching/Removing table sections . . . . . . . . . 17
6.6 Classification and symbology . . . . . . . . . . . . . 45
General notes . . . . . . . . . . . . . . . . . . . . . . . 17
6.7 Use with other equipment . . . . . . . . . . . . . . . . 45
Attaching a tabletop section . . . . . . . . . . . . 17
Electrosurgical equipment (h.f.) . . . . . . . . . 45
Removing a tabletop section . . . . . . . . . . . 18
Defibrillation equipment . . . . . . . . . . . . . . . 45
Changing sections during a procedure . . . . 18
Other . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.4 Attaching/Removing mattresses . . . . . . . . . . . 18
6.8 Standards compliance . . . . . . . . . . . . . . . . . . . 45
4.5 Batteries - Charging/Changing/Standby . . . . . 19
6.9 Environmental conditions . . . . . . . . . . . . . . . . 45
Battery charging . . . . . . . . . . . . . . . . . . . . . 19
Operating environment . . . . . . . . . . . . . . . . 45
Battery changing . . . . . . . . . . . . . . . . . . . . 19
Transport & storage environment . . . . . . . . 45
Standby batteries . . . . . . . . . . . . . . . . . . . . 20
4.6 Tabletop control . . . . . . . . . . . . . . . . . . . . . . . . 20 (continued)
General . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Tabletop powered motions . . . . . . . . . . . . . 21
Corded handset . . . . . . . . . . . . . . . . . . . . . 21
Handset button functions . . . . . . . . . . . . . . 23
Standby control panel . . . . . . . . . . . . . . . . . 23
Head and leg section control . . . . . . . . . . . 24
T-IM102d P3/48
Illustrations
P4/48 T-IM102d
T20-a
OPERATION TABLE
T-IM102d P5/48
NOTE: Table illustrated has external locking catches for
the pedals. Not all tables have these, see section 4.2.
P6/48 T-IM102d
T20-a
OPERATION TABLE
1 Head section
2 Long trunk section
3 Short trunk section
4 Leg section
5 Leg section gas spring
6 Short trunk section release button
7 Long trunk section release button
8 Battery fuses
9 Wheel foot pedal
10 Head section release bar
11 Head section gas spring
12 Standby control panel
13 Base
14 Pedestal
15 Castor foot pedal
16 Leg section release bar
17 Mains fuses
18 Connection socket for mains (ONLY use Eschmann mains cord, Part No. 391177)
19 Main battery charging state LED:
Red = Mains ‘on’, batteries in ‘bulk’ charge
Amber = Mains ‘on’, batteries in ‘top-up’ charge (duration 2 hours)
Green = Mains ‘on’, batteries in ‘trickle’ charge (i.e. fully charged)
20 Standby battery charging state LED (colour code as main battery above)
21 Main table ‘On/Off’ switch ( = Off, = On)
22 Connection socket for Eschmann footswitch (ONLY)
23 Table ‘On’ LED (green)
Bright = Table in use, micro-controllers active
Dim = Table ready for instant use but in battery save (sleep) mode
24 Connection socket for Eschmann handset (ONLY)
25 Head and leg section end blocks (use to stow UK handset)
26 Accessory side rails (can also be used to stow handset*)
27 Standby battery switch
* Clip on handset must match side rail type, see accessory section 4.8
T-IM102d P7/48
1.3 Part identification also provided with an internal mains powered battery
charger, to charge both the main batteries and the standby
1.3.1 Fig. 1.1 shows the table in its normal batteries.
configuration and with the key provided it identifies the
major parts of the table and the manually operated controls CAUTION
(see section 4.6.1-4.6.5 for details of powered controls). It is most important that fuses of the correct
1.3.2 Fig. 1.1 also shows the location of the electrical type, size and rating are installed (see Technical
connections, fuses, switches, buttons and LEDs that the Data, section 6.5.4).
user needs to be familiar with (for more information on the
1.4.6 The T20-a operation table requires a mains
handset see sections 4.6.3 and 4.6.4, for the standby
electrical supply corresponding to the voltage shown on
control panel see section 4.6.5).
the Serial Number Plate (for location see inside front cover).
1.4 Installation Only use the Eschmann mains supply cord provided with
the table (part number 391177 Australian, or 714188 U.K.).
1.4.1 In the U.K. the table is delivered un-crated with If the plug supplied fitted to the cord is not suitable it should
the head and leg section fitted. The mattress set is boxed be replaced with a suitable plug wired as below. If the plug
individually and placed on the trunk sections. The handset, is a fused type, a 10A fuse must be fitted. The mains supply
mains cord and literature are supplied loose. Any cord must always be wired as follows:
accessories ordered with the table will be packed Brown internal cord to LIVE
individually. Some of the following unpacking and assembly Blue internal cord to NEUTRAL
information is not therefore applicable to U.K. customers. Green internal cord to Earth
For overseas markets the table is usually packed in a (Note: The T20-a is a Class II, Type BF, there
container with the head and leg section fitted, together with is no EARTH connection through to the table)
a boxed mattress set a mains cord and the literature. The
handset is packed within the container in an antistatic bag. 1.4.7 The table batteries should be recharged (see
Other accessories are usually packed separately, but some section 4.5.1) and the table operated through the cycle of
may be included in the main case and should be unpacked movements detailed below to check and ensure correct
and stored separately during the table installation. function, before the tables first use.
1.4.2 Carefully remove the T20-a operation table Full Trendelenburg / reverse Trendelenburg
(having first removed any accessories and packing Maximum to minimum height
restraints from within the container) from the packing case Maximum tilt, left and right
as follows:- Maximum traverse, caudally and cranially
Maximum to minimum break
i Remove the walls of the case leaving the table on
the base still in the braked position as packed. 1.4.8 The table has an antistatic pathway from the
Remove any chocks from the pallet base to enable tabletop, through an internal resistor, to the castors. To
later table movement (iv below). complete the antistatic pathway, the table must be used
ii Position the ramp provided (in the packing case) on an electrically conductive, or on an antistatic floor (also
adjacent to the base of the case. see the warning in section 4.9).
iii Follow the instruction provided in section 4.2.2 of 1.4.9 The table should be cleaned and disinfected prior
this manual and raise the table base into the ‘Wheel’ to its first use as detailed in section 5.1 and 5.2 and then
orientation. commissioned in accordance with any local procedures
iv With at least two people to support the table’s weight, applicable to new equipment, this should include staff
push it (do not pull it) off of the pallet base and down training. Eschmann can supply a range of wall charts and
the ramp. training aids and provide on-site training, contact
v Note that the table should not be wheeled over rough Eschmann for more information.
ground, always use a trolley until a smooth floor area
has been reached. 1.5 Symbols and graphics
vi Should it be necessary to lift the operation table refer
to section 5.3.3 where suitable lifting points and 1.5.1 To enable an easy reference to all the symbols
methods are detailed. and graphics used on the Operation Table (and within these
‘Instructions for use’) the following grouped sections show
1.4.3 Unpack the mattress set and fit a mattress to all the symbols and graphics used.
each tabletop section as detailed in section 4.4.
1.4.4 Any packaging materials should be recycled or
1.5.2 Symbols general
disposed of in accordance with current legislation (e.g. The The following symbols are shown on various parts of the
European Packaging Directive). table or Serial Number Plate.
1.4.5 The T20-a operation table is powered by internal This symbol is used to indicate the infrared
rechargeable batteries which are connected and charged code of the handset, see the infrared handset ‘User
before delivery (see disposal note in section 4.13). It is Handbook’.
P8/48 T-IM102d
T20-a
OPERATION TABLE
IPX 4 This symbol (splash proof) denotes that the This graphic, adjacent to the mains
equipment (the table) meets the requirements of IEC529 socket, indicates ‘mains on’ and
for protection from splashing water. battery charge state for both
batteries, see section 4.5.1.3.
IPX 6 This symbol (protection against heavy seas) denotes
that the equipment (the handset) meets the requirements
of IEC529 in that water from heavy seas or water projected 1.5.3 Handset button symbols
from powerful jets shall not enter in harmful quantities.
The following symbols are shown on the handset buttons
This symbol indicates that the equipment is for
to indicate their function. Use of the handset is fully detailed
use on alternating current.
in section 4.6.3 - 4.6.4 of this manual.
This symbol indicates that fuses adjacent to the
symbol have a rating and type as detailed. Trendelenburg - Press to rotate tabletop in the
This symbol warns the user to read the accom- Trendelenburg (head down) direction.
panying documents, the ‘Instructions for use’.
Reverse Trendelenburg - Press to rotate
Symbols and above the main table ‘on/off’ tabletop in the Reverse Trendelenburg (head
switch, indicate ‘OFF’ and ‘ON’ respectively. up) direction.
With the mains cord attached the equipment has
‘Class II’ protection against electric shock. Height down - Press to move tabletop down.
The patient leakage current, with mains voltage on
the applied parts, meets the requirement for type
BF medical electrical equipment and are defibrillator Height up - Press to move tabletop up.
proof.
This symbol is used to indicate the table’s duty cycle Tilt - Press to tilt tabletop down on the left (when
which is the ratio of the operating time to the sum of viewed from the long trunk end)
the operating time and the ensuing interval.
SN This symbol indicates the unit serial number is as Tilt - Press to tilt tabletop down on the right
indicated adjacent to the symbol. (when viewed from the long trunk end)
REF This symbol indicates the catalogue number is as
indicated adjacent to the symbol. Break down - Press to move the break down
This symbol indicates that the date of manufacture (i.e. short trunk moves down w.r.t. long trunk)
is as indicated adjacent to the symbol.
Break up - Press to move the break up
This symbol indicates the connection point for
(i.e. short trunk moves up w.r.t. long trunk)
a footswitch.
Traverse Cranially (towards the head) - Press
This symbol indicates the connection point for to move the tabletop Cranially (i.e. tabletop
the corded handset. moves in the long trunk direction).
This symbol indicates that the table section Traverse Caudally (towards the feet) - Press
to which it is applied (e.g. head section, under to move the tabletop Caudally (i.e. tabletop
the mattress) should not be used as a seat. moves in the short trunk direction).
This symbol indicates the ‘Safe working load’
of the section to which it is applied can safely Flexion - Press to move tabletop into Flexion.
support an evenly distributed load to the value
indicated, in this example 25kg.
Extension - Press to move table into Extension.
This symbol indicates the ‘Minimum breaking
load’ of the section to which it is applied. An
evenly distributed load (in this example 100kg
Return to Level - Press to return tabletop
or greater) may break the section.
to a preset level position.
This symbol on the table base indicates the 1.5.4 Handset graphics
table should not be moved with a load above
135kg, nor loaded above 300kg. This graphic is shown on the corded
hand control to indicate the battery
charge level. The indication provided
This symbol is used to identify the standby by the six coloured LEDs is explained
battery switch. in section 4.5.1.
T-IM102d P9/48
1.5.5 Standby control panel button symbols 1.6.1 Interference from other equipment
The following symbols are shown on the standby control The T20-a table has been designed to ensure that when
panel buttons, indicating the function they select. The using it in close proximity with ‘other correctly designed’
arrows (upper or lower) indicate the direction the selected electrical equipment, interference with the T20-a control
function will move, if the corresponding direction button systems does not occur. Eschmann cannot guarantee that
(i.e. upper or lower) is pressed, see section 4.6.5. other equipment used in an operating theatre is properly
constructed so as to avoid electrical interference with the
Button selects Trendelenburg function. T20-a. This could be a problem particularly with very old
equipment (i.e. it does not conform to the latest standards).
In the unlikely event that electromagnetic interference is
Button selects Break function. caused by ‘other’ noncompliant equipment, such ‘other’
equipment should not be activated at the same time the
T20-a is switched ‘on’ and its control systems are active.
Button selects Height function.
1.6.2 Interference with other equipment
The T20-a has been designed to ensure that when using it
Button selects Tilt function. in close proximity with ‘other correctly designed’ equipment
the T20-a table does not cause interference with such ‘other
correctly designed’ equipment. Eschmann cannot
Button selects Traverse function. guarantee that other equipment used in an operating
theatre is properly constructed to withstand electrical
interference with adjacent equipment. This could be a
Direction button - Press button to obtain movement problem particularly with very old equipment (i.e. it does
indicated by upper arrow of function button. not conform to the latest standards). In the unlikely event
that electromagnetic interference is caused by the T20-a
Direction button - Press button to obtain movement with such ‘other equipment’, this ‘other equipment’ should
indicated by lower arrow of function button. not be activated at the same time the T20-a is switched
‘on’ and its control systems are active.
1.6 Interference considerations 1.6.3 Use of infrared
The T20-a table has been designed and manufactured in The use of infrared remote control handsets for electronic
such a way as to remove, or minimise as far as possible, equipment is becoming more common in the operating
risks connected with reasonably foreseeable environmental theatre, and an infrared handset is available for the T20-a.
conditions such as magnetic fields and external electrical Before using an infrared handset (an optional accessory)
influences (i.e. electrical interference). The risks of electrical read the User Handbook supplied with it (T-UH16). The
interference from or to other devices normally used with table as supplied will not respond to an infrared handset
the T20-a have been taken into account. The steps taken until it has been configured to do so.
to achieve this are to ensure compliance with relevant inter-
national electromedical standards {i.e. IEC 60601-1-2:1993 1.7 Alarms
and IEC 60601-2-46:1998 (clause 36) regarding electro-
magnetic compatibility}. This compliance has been 1.7.1 Visual
confirmed by independent testing. However it is not
possible to simulate all the conditions that may be Visual alarm functions of this table are LED indicators for
encountered, the compliance testing therefore provides mains ‘on’ and battery charge state (see Fig. 1.1) situated
only a very good indication as to the susceptibility or on the table base and battery level indication (see Fig.
suppression of emissions to or from a device. The T20-a 4.8) provided on the corded handset. Ensure you are
table has been subjected to levels of electromagnetic familiar with these before using this operation table (see
interference up to ten times higher than the minimum sections 4.5.1.2 and 4.5.1.3). There is also an LED indicator
standard requirement and has not been affected by them to show that the table is switched ‘on’ (see item 22, Fig.
(i.e. no unwanted table movements were encountered). 1.1).
It is highly unlikely that any electrical interference problems
1.7.2 Audible
will be encountered with the T20-a, however, should such
interference be suspected Eschmann provide the following Audible alarm indication is provided (a double ‘beep’
guidance. repeating every eight seconds) to warn that the table has
be left ‘on’ for an extended period and should be switched
‘off’ (see section 4.12.4 and 4.12.5 for more information).
A single ‘beep’ also sounds when the table is switched
‘on’.
P10/48 T-IM102d
T20-a
OPERATION TABLE
T-IM102d P11/48
DO NOT:
CAUTIONS
♦ Lift the table by its tabletop.
Do not place either heavy accessories, or, long
♦ Push the table over rough surfaces, use a trolley.
accessories that could impose high torques,
to the side rails, as this may lead to damage of ♦ Drop the table (or individual sections).
the rails. ♦ Put heavy weights on the table sections, observe the
maximum advised loading.
Do not exceed the duty cycle for any table ♦ Put sharp objects on, or against, mattresses, pads, or
motor drive as detailed in the technical data the radiographic tabletop.
section 6.5.5.
♦ Place any objects on the base covers
2.2 Do’s and Dont’s ♦ Drop heavy objects onto the radiographic tabletop or
base covers.
Attention to the following points will prolong the life and
efficiency of the T20-a operation table and will help to avoid ♦ Spill oil, ether, or other fluids onto the mattresses or
the risk of accidents, or damage. Other safety notes and the pads.
warnings are also given within the text of this manual and ♦ Pull the table by any of the tabletop sections, or
these should be noted during use of the table. accessories, always push it.
♦ Service this equipment unless you have been trained
DO:
by Eschmann.
♦ Keep these ‘Instructions for use’ close-to-hand.
♦ Read these ‘Instructions for use’ carefully before 2.3 Daily ‘Before use’ test
adjusting, moving or using the table.
It is recommended that a ‘Daily test’ is carried out before
♦ Use the table on an antistatic floor to prevent using the table every day. First check that the batteries
inadvertent static buildup. are charged (see section 4.5.1.2) if they are not charge
♦ Use only the correct Eschmann mattresses and them (see section 4.5.1.3) and ensure that a suitable
accessories that are compatible with the T20-a as practice is put in place so they are charged at the end of
detailed in these instructions. every shift in future. Then check the table responds to the
♦ Check that handset cables and standby controls are following movements from either handset (each movement
not damaged before use. only needs to be brief, one or two seconds each):
♦ Check that the table and its accessories are not worn Trendelenburg / reverse Trendelenburg
or damaged, or are in any way not suitable for the Height Up / Down
intended purpose, before use. Tilt Left / Right
♦ Check that all the sections (e.g. head and leg) and Traverse Caudally / Cranially
accessories are secure, and put the table base in the Break Up /Down
‘braked’ position before use.
Then check at least two of the above plus Trendelenburg
♦ Ensure that all cables are not stretched leading to using the Standby control panel.
disconnection or damage during movement or
readjustment of the operation table or patient.
2.4 Accessories
♦ Remove table accessories and their clamps (in
particular rotary clamps) from sidebars, when they The accessories available from Eschmann for the T20-a
are not being used. operation table are listed in section 4.8. Use of other
♦ Read and follow the instructions for cleaning, and for equipment with the T20-a should only be considered after
the care of the table and mattresses. evaluating the safety of the patient and personnel.
Inadvertent use of incorrect accessories could damage the
♦ Switch ‘off’ and disconnect from the mains electrical table and lead to injury. Always ensure that the ‘User/
supply prior to cleaning and/or disinfecting the table Service Handbook’ supplied with T20-a accessories is
and when it is not being recharged. complied with and follow all the safety notes contained
♦ Ensure that the table and accessories are serviced at within them during use. For accessories that fit onto the
regular intervals (every six months is the side rails ensure that they are compatible with the side
recommended frequency) only by Eschmann trained rails fitted to this table. This table can be supplied with
personnel, or by accredited agents. standard UK or Denyer style side rails see section 1.1.2.
♦ Ensure that only the Eschmann mains cord supplied
with the table is used to connect the table to the mains.
♦ Ensure that only Eschmann supplied parts are used
during part replacement.
P12/48 T-IM102d
T20-a
OPERATION TABLE
T-IM102d P13/48
3.2.5 Only Eschmann accessories listed in this manual 4.0 OPERATION
should be used on the table and in accordance with the
‘User Handbook’ supplied with the accessory. Other
accessories, especially those that could compromise table 4.1 Introduction
stability, must not be used. 4.1.1 This section is split into various parts that are
3.2.6 Provision of a diagnostic port within the table sequenced to match those during use, starting with moving
enables access for reprogramming the tables software, the table into position and ending with after use procedures.
down-loading of fault information and service functions. By the nature of the product, this sequence can only be
Use of this port MUST be limited to trained service approximate and will not match all occasions. If the table
personnel only, and should only be used in accordance has not been used for some time it is recommended that
with the correct Eschmann manuals (see 3.2.9 below). the cycle of movements detailed in section 1.4.7 are run
through before using the table.
3.2.7 The table has four fuses that the user has access
to in the event of failure. The position of these are shown 4.2 Moving and braking the table
in Fig. 1.1 and are replaced as detailed in section 4.7.
3.2.8 Eschmann can provide customers with manuals WARNING
(see 3.2.9 below), for use by them in maintaining their own Always push the table (do not pull it) at a
equipment. These manuals contain schematic diagrams, suitable height ensuring that it is stable at
component part lists, descriptions and calibration all times, take care not to collide with
instructions which will assist the customers’s Eschmann
personnel or equipment. Maximum stability
trained personnel to service the equipment or replace parts
(which should only be obtained from Eschmann). will be at minimum height. Ensure that the
patient is adequately supported and
3.2.9 The following manuals are applicable to the
T20-a operation table and its accessories (the part number restrained (especially limbs) using cot
is in brackets following the manual reference) they are sides as appropriate. Ensure that all
available to order, see inside front cover for contact details:- connections via cord or tube, to the patient
T-SM47 (111707) - Service manual* or table, have either been disconnected, or
T-IPL35 (111706) - Illustrated parts list are only attached to equipment that will
T-IM94 (111012) - T20 Application software manual move with the table to avoid inadvertent
T-IM56 (698907) - General accessory leaflet disconnection.
Note: Some T20 accessories are provided with their
own ‘User/Service Handbooks’. When changing from ‘castor’ or ‘wheel’
* (see section 5.3.1.2) orientation, to ‘braked’, ensure that no
objects (e.g. cords, tubing etc.) can become
3.2.10 Appendix 1 provides a log that can be used to
record those people trained in the safe use of this operation
trapped beneath the table base and floor.
table. It is suggested that this is used to ensure that ALL Do not move the table with a patient
personnel using this table, are aware of all the warnings weighing more than 135kg and ensure
and cautions contained within these ‘Instructions for use’.
stability is maintained.
3.2.11 The T20-a table and its accessories, as listed in
these ‘Instructions for use’, do not contain ‘Latex’.
4.2.1 General
3.3 Accessories 4.2.1.1 The table can be moved easily on its own
built-in castors and wheels. Normally the table rests on
3.3.1 The accessories available from Eschmann for
brake pads at the long trunk end and wheels at the short
the T20-a operation table are listed in section 4.8. Use of
trunk end, these provide a secure and static location on
other equipment with the T20-a should only be considered
the operating theatre floor. When in the ‘braked’ position,
after evaluating the safety of the patient and personnel.
both pedals (items 9 and 15, Fig. 1.1) are in their raised
For accessories that fit onto the side rails ensure that they
positions. Lower the table to a suitable height to achieve a
are compatible with the side rails fitted to this table. This
stable position before moving the table.
table can be supplied with standard UK or Denyer style
side rails see section 1.1.2. 4.2.1.2 To move the table it can be placed onto its
wheels and castors in two ways, providing both a ‘castor’
orientation and a ‘wheel’ orientation. The latter enables
easy movement of the table in a straight line (down a
corridor for example).
P14/48 T-IM102d
T20-a
OPERATION TABLE
4.2.1.3 Moving the table into either of these
orientations does not require battery power, they are
manual operations achieved by using one or both of the
foot pedals on the table base (items 9 and 15 of Fig. 1.1).
4.2.1.4 When moving the table in ‘wheel’ orientation
always push it (do not pull it) from the long trunk end. It
can be pushed (not pulled) from either end when in the
‘castor’ orientation.
4.2.1.5 The table should always be left in the ‘braked’
orientation with both pedals (items 9 and 15, Fig. 1.1) in their
raised positions. Do not leave the table in a half braked Fig. 4.1b Pedal (external catch) locked down
position (pedal 9 of Fig. 1.1 up and pedal 15 of Fig. 1.1 down). (Castor foot pedal illustrated)
4.2.1.6 The operation of the foot pedal varies 4.2.2.2 When placed into the ‘Wheel’ orientation the
depending on the ‘design state’ of the table. This can be table base is supported on two wheels at the short trunk
checked by noting the Serial Number. Tables with Serial end and two castors at the long trunk end. To move the
number prefixed ‘T2AA...’ will have an external locking table push it from the long trunk end, moving the end nearest
catch, tables with Serial Number prefixed ‘T2AC...’ will have to you, left or right, to steer the table in the required direction.
an internal locking catch. The external pedal locking catch
is easily noticed and users should note if this is present
yyy
before operating the pedal as detailed below.
yyy
4.2.2.1 Pedal operation is easily achieved if the pedal
is pressed down with the right foot whilst steadying yourself
with hands on the tabletop. Stand on the side of the table
on which the footpad is located. Note if the pedal has an
internal or external locking catch and proceed as applicable.
With external catch To place the table into ‘Wheel’
orientation from the ‘braked’ orientation press the wheel
pedal (item 9, Fig. 1.1) down and towards the table until the
yyy
lever part of the pedal locks under the catch (see Fig. 4.1a
and 4.1b). Check that the lever has correctly engaged and
locked under the catch.
With internal catch To place the table into the
‘Wheel’ orientation from the ‘braked’ orientation, press the
wheel pedal (item 9, Fig. 1.1) down steadily until you hear
an audible ‘click’ (action indicated in Fig. 4.2). This ‘click’
indicates that the pedal has locked in the down position. Do Fig. 4.2 Lowering a foot pedal (internal catch)
not continue to press the pedal after the ‘click’ as this will (Wheel foot pedal illustrated)
release the internal catch and the pedal will not lock down.
yyy
If this does happen the pedal must be allowed to rise fully 4.2.2.3 The table is in ‘wheel’ orientation when the
(this resets the internal catch) before pressing it down again. wheel pedal (item 9, Fig. 1.1) is in the lowered position
and castor pedal (item 15, Fig. 1.1) is in the raised position.
yyy
4.2.3 ‘Castor’ orientation (from ‘wheel’)
4.2.3.1 Pedal operation is easily achieved if the pedal
is pressed with the left foot whilst steadying yourself with
hands on the tabletop. Stand on the side of the table on
which the footpad is located. Note if the pedal has an internal
or external locking catch and proceed as applicable.
With external catch To place the table into the
‘castor’ orientation from the ‘wheel’ orientation, press the
castor pedal (item 15, Fig. 1.1) down and towards the table
Fig. 4.1a Lowering a pedal (external catch) until the lever part of the pedal locks under the catch (see
(Wheel foot pedal illustrated) Fig. 4.1a). Check that the lever has correctly engaged and
locked under the catch, see Fig. 4.1b.
T-IM102d P15/48
With internal catch To place the table into the them both, in any sequence. Note if the pedal has an internal
‘castor’ orientation from the ‘wheel’ orientation, press the or external locking catch and proceed as applicable.
castor pedal (item 15, Fig. 1.1) down steadily until you hear With external catch Press the pedal down and
an audible ‘click’ (action indicated in Fig. 4.2). This ‘click’ away from the table base until it is clear of its locking catch,
indicates that the pedal has locked in the down position. Do see Fig. 4.3a (use the foot advised in 4.2.2.1 or 4.2.3.1 as
not continue to press the pedal after the ‘click’ as this will appropriate). Release pressure on the pedal and allow it
release the internal catch and the pedal will not lock down. to move into the raised position. The table will gently lower
If this does happen the pedal must be allowed to rise fully onto its brake pads and wheels, the motion is softened
(this resets the internal catch) before pressing it down again. and braked by an internal damper.
With internal catch Press the pedal ‘firmly down’
4.2.3.2 When placed into the ‘castor’ orientation the
see Fig. 4.3b (use the foot advised in 4.2.2.1 or 4.2.3.1 as
table base is supported on four castors, two at each end
appropriate) this will disengage the internal locking catch.
of the table. This orientation enables the table to be moved
Release pressure on the pedal and allow it to move into
in any direction including sideways and swivelling within
the raised position. The table will gently lower onto its brake
its own length. To move the table always push it in the
pads and wheels, the motion is softened and braked by an
required direction, never pull it.
internal damper.
4.2.3.3 The table is in ‘castor’ orientation when both
4.2.4.2 The table is in ‘braked’ orientation when both
pedals (items 9 and 15, Fig. 1.1) are in their lowered positions.
pedals (9 and 15, Fig. 1.1) are in their raised positions.
yyy yyy
yyy
Fig. 4.3a Raising a pedal (external catch)
(Castor foot pedal illustrated)
= BRAKED
TABLE WILL NOT MOVE
= WHEEL
PUSH IN A STRAIGHT LINE
= CASTOR
PUSH IN ANY DIRECTION
P16/48 T-IM102d
T20-a
OPERATION TABLE
4.2.5 ‘Castor’ orientation (from ‘braked’)
WARNING
4.2.5.1 To place the table into ‘castor’ orientation from Take care when realigning a sections pins,
the ‘braked’ orientation follow 4.2.2 to place the table into
‘wheel’ orientation and then follow 4.2.3 to complete the
ensure that fingers are clear of the gas springs
move into the ‘castor’ orientation. during actuation of the release handle.
Note: It is not critical that sections 4.2.2 and 4.2.3 are
carried out in this sequence the reverse is equally suitable
and correct.
WARNING
Ensure that nothing becomes trapped
(e.g. fingers, tubing, cords) when attaching
tabletop sections. When removing a
section be prepared to support the full
weight of the section when the guide pins
disengage. For users of small stature,
when handling larger sections (e.g. the leg
section) two people should work together
to avoid strain injuries. Also to minimise
weight for all users remove mattresses
before fitting or removing a section. Users
of earlier Eschmann tables (e.g. MR and RX
Series) should note that the T20-a guide
pins are shorter and disengage earlier.
Always ensure that the sections have been Fig. 4.5 Section pins in and out
correctly and securely fitted before use and
of alignment
only use the correct Eschmann sections.
4.3.1.4 To realign the pins of a table section place
4.3.1 General notes the section on its side (see Fig. 4.5) and support it by its
4.3.1.1 When removing a tabletop section (e.g. a upper side rail only keeping fingers etc. well clear of the
head or leg section) ensure that it is horizontal before gas springs (see 1, Fig. 4.5). Then actuate the release bar
removal as this will aid later replacement (pins will be (see 2, Fig. 4.5) until the gas springs have moved both
aligned with the section). If the section is not horizontal pins to the ends of their travel as shown (see 3, Fig. 4.5).
adjust it to horizontal before removal (see section 4.6.6). Insert the pins into the tabletop for 90% of their length,
actuate the release handle and adjust the section until
4.3.1.2 During long term storage of a section it should horizontal and then push the section fully home until the
be placed with the pins up, this ensures continued lubri- locking catches engage (also see section 4.3.2).
cation of the gas spring seals. Take care not to actuate the
release handle during storage, this will alter alignment of 4.3.2 Attaching a tabletop section
the pins, should this happen see 4.3.1.4. For short term
storage and to aid manual handling (e.g. during section 4.3.2.1 To attach the removable tabletop sections
reversal) use of the ‘T’ series Head/Leg accessory trolley (i.e. head or leg) hold the section firmly with two hands
(REF TA-040013) is recommended. aligning the pins of the section (item 1, Fig. 4.6) with the
location holes (item 2, Fig. 4.6) in the fixed tabletop section
4.3.1.3 When attaching a section first check to see if as shown in Fig. 4.6. Also refer to section 4.3.1.3.
the pins are aligned with each other. It is possible that during
storage the release handle may have been actuated and 4.3.2.2 Gently insert the section pins into the tabletop
the pins are no longer aligned as shown in Fig. 4.5. If they and slide the section into the tabletop evenly until the
are aligned attach the section as detailed in section 4.3.2. locking catches engage with an audible click. The section
If they are not, align them and attach the section as detailed will slide in easily if the weight of the section is gently
in section 4.3.1.4. supported with both hands. Take care not to trap anything
(especially fingers) between the section and the tabletop.
T-IM102d P17/48
4.3.3.3 Now using both hands (one on each side of the
section as shown in 4 of Fig. 4.7) continue to gently pull
the section out evenly from the trunk section until the guide
pins are free. Again gently lifting and supporting the weight
of the section will make this action easier. When employing
two people, one should hold the second button in whilst
the other pulls the section out evenly, both can then support
the weight when the pins are free.
Fig. 4.7 Removing a table section 4.4.4 The larger mattress sections can be bulky to handle
individually if of small stature, use two people to avoid strain.
P18/48 T-IM102d
T20-a
OPERATION TABLE
4.5 Batteries - charging/changing/standby Red LED ‘on’ = Critically low battery. Battery level is
critically low and only adequate for one Trendelenburg
4.5.1 Battery charging movement. They MUST be recharged as soon as
possible (i.e. during the current procedure or at least
WARNING before the next procedure).
Ensure that the mains supply used is Battery level
suitable (i.e. 100-240V, 50-60Hz) before indicators.
attaching the table to the mains supply.
Only use the mains cord supplied with the
table and charge the batteries daily.
CAUTION
Continued use of the table batteries when
‘critically low’ can damage the batteries. Charge Battery
batteries regularly to maintain peak performance. charging
Do not remove the table from charge until both LED
charging LED’s are ‘green’ to avoid false battery code
level indication on the handset.
4.5.1.1 Batteries within the table base are mains Fig. 4.8 Handset and charging indicators
rechargeable and should power the table continuously for
3 days normal use (from a full charge). However to maintain
When the main battery ‘critically low’ warning is indicated
peak battery performance the table batteries should be
the table cannot be operated by either handset or the
placed on charge (see section 4.5.1.3) at the end of each
footswitch. To continue to operate the table either connect
day or shift. Daily charging should be encouraged as good
it to a suitable mains supply and recharge the batteries
practice, do not wait for either ‘low’ battery indication
(see 4.5.1.3) or use the standby batteries (see 4.5.3).
before charging. Battery charge level is indicated by LEDs
Note: The standby control panel can be used even when
on the corded handset, see Fig. 4.8 and section 4.5.1.2.
the ‘critically low battery’ indication is given but note the
The table should be placed on charge as detailed in section
warning in section 4.6.5.
4.5.1.3. Also see battery management sheet Appendix 3.
Note: Over the first few charge cycles (from new) battery 4.5.1.3 To recharge the table batteries (normal and
capacity increases until they reach a stable maximum level. standby) connect the mains cord supplied with the table
4.5.1.2 Batteries should be checked for adequate (do not use any other mains cord) into the mains inlet
charge before using the table for any surgical procedure. (item 18, Fig. 1.1). Connect the mains cord to a suitable
The corded handset has two battery level indicators, one mains supply (check mains voltage is 100-240V, 50-60Hz)
for the main battery and one for the standby battery, see and switch the supply ‘on’ if controlled by a switch. The
Fig. 4.8. These indicators will only operate when the corded battery charging LEDs (see Fig. 1.1 and 4.8) will illuminate
handset is plugged into the table and the table is switched ‘red’ to indicate ‘bulk’ recharge, ‘amber’ to indicate that the
‘on’. Each indicator contains three coloured LEDs, these batteries are in a 2 hour ‘top-up’ charge and will illuminate
indicate the following information: ‘green’ to indicate a fully charged battery on ‘trickle’ charge.
The batteries should be fully charged within six hours (if
Main battery level indicator: the table is not in use) but it is safe to leave the mains
Green LED ‘on’ = Battery level satisfactory. supply ‘on’ when the batteries are fully recharged and both
Amber LED ‘on’ = Low battery. An indication that LEDs are ‘green’. See CAUTION above section 4.5.1.1.
the remaining battery charge is only adequate for Note: Always ensure the table is switched ‘off’ to minimise
one day’s average use. Batteries MUST be the battery charging time.
recharged at the end of the current shift.
Red LED ‘on’ = Critically low battery. An indication 4.5.1.4 It is recommended that the mains cord is
that the remaining battery charge is critically low. stored on the T20 accessory trolley (an accessory) to avoid
Batteries MUST be recharged as soon as possible loss and inadvertent use of an incorrect mains cord. It
(i.e. during the current procedure or at least before should be wrapped around the cable cleats.
the next procedure).
4.5.2 Battery changing
Standby battery level indicator:
Green LED ‘on’ = Battery level satisfactory. 4.5.2.1 Batteries in the table base should only be changed
Amber LED ‘on’ = Low battery. Battery level is low by service personnel. They should last for many years if
and only adequate for two or three Trendelenburg recharged regularly and should not require user intervention
movements, they MUST be recharged at the end of (also see section 4.13 for the caution concerning disposal).
the current shift.
T-IM102d P19/48
4.5.3 Standby batteries
CAUTION
WARNING Do not exceed the duty cycle for any motor
The standby batteries are for emergency drive as detailed in section 6.5.5.
use only and will only provide power for a
4.6.1 General
few movements.
4.6.1.1 The major tabletop adjustments and motions
4.5.3.1 The table is provided with standby batteries to are powered. These powered motions are controlled by a
cover the unlikely event of a main battery failure or fault. corded handset supplied with the table (see 4.6.3) and the
To use the standby batteries, the standby battery switch standby control panel (item 12, Fig. 1.1, see section 4.6.5).
(see 27, Fig. 1.1) must be pressed and held depressed An optional infrared handset or an optional footswitch (see
whilst the required motions are conducted using the normal accessory list section 4.8) can also be used to control the
hand and foot controls. tabletop. The use of these accessories is explained in the
4.5.3.2 The standby batteries are only used when the leaflet provided with them.
standby battery switch is depressed and they are automatically 4.6.1.2 All signals from controllers will be treated in a
recharged during the normal battery charging process. strict priority order to avoid multiple activation errors from
different controllers. Inputs will be responded to in the
4.6 Tabletop control sequence below:
1 Standby control panel
WARNING 2 Corded handset
Always ensure that the patient is secure 3 Infrared handset
4 Footswitch control.
or adequately supported during all tabletop Button presses from a controller lower on this list will be
adjustments and that such adjustments do ignored if any button is pressed on a controller higher on
not compromise table stability. the list and the input from the higher controller will be
actioned.
If a patient is on the table and the handset is
clipped onto the tabletop, take care that the 4.6.1.3 The table cannot be powered by footswitch,
handset buttons are not inadvertently or either handset, when the main battery ‘critically low’
indication is displayed (LED item 19, Fig. 1.1 ‘red’) unless
activated (e.g. by a patient Transporter or the standby battery switch (item 27, Fig. 1.1) is held
person) causing unwanted table movements. depressed. However the tabletop can be controlled by the
Similarly take care at all times to ensure that standby control panel (item 12, Fig. 1.1) situated on the
side of the column. This standby control panel has limited
the handset is not inadvertently activated. functionality in comparison to either handset and should
During all tabletop adjustments be aware ONLY be used in an emergency (e.g. handset or footswitch
of pinch points between the moving and failure, control system failure, or ‘critically low’ battery). Use
of this standby control panel in any situation other than an
static parts (e.g. the break hinge and head
emergency SHOULD be avoided as certain safety devices
and leg hinges). are overridden. The controls and buttons of the standby
Ensure that all theatre drapes, electrical control panel are detailed in section 4.6.5. In the unlikely
event of insufficient battery power when using the standby
cords and any medical tubing etc. are clear
control panel, press and hold the standby battery switch
from entrapment in pinch points and that (see 27, Fig. 1.1) whilst conducting powered motions.
adequate excess drape, cord, or tube, are
4.6.1.4 To operate the tabletop powered motions the
available to cater for the adjustment required.
table must be switched ‘on’ ( ) at the table on/off switch
Ensure that any motion and adjustment (item 21, Fig. 1.1). The table will emit a single ‘beep’ and
selected will not cause any part of the table the green LED, item 23 Fig. 1.1, will illuminate to show the
to hit or collide with any person or object. table is ‘on’. The LED will be bright during use but only dim
if the operating system has gone into ‘sleep’ mode (saving
Examples being:- battery power). The table will instantly respond to any
i) Leg section hitting floor when table is control input even when in ‘sleep’ mode. Also see sections
low and reverse Trendelenburg selected. 4.12.4 and 4.12.5.
ii) A fully lowered leg section hitting table 4.6.1.5 If two buttons are pressed on a handset or
base when tabletop is lowered. footswitch at the same time neither will have any effect.
iii) Head section hitting anaesthetist’s seat The function of the second button pressed is ignored and
when Trendelenburg selected. automatically cancels the function of the first button pressed
P20/48 T-IM102d
T20-a
OPERATION TABLE
and any table motion stops. Releasing either of the buttons 4.6.3 Corded handset
will enable the function of the remaining button (if still
pressed) to be actioned. This is to eliminate multiple button WARNING
activation errors.
Do not plug two handsets into the table at
4.6.1.6 In the unlikely event of failure of the main table the same time, this disables their function.
batteries, ‘press and hold’ the standby battery switch (see
27, Fig. 1.1) in the ‘on’ position and use the corded or infrared
4.6.3.1 The corded handset will operate any T20-a
handset to control the table. This enables the standby operation table (note that the infrared handset is configured
batteries to power the table, however the main battery fault
to match a specific table). Input from the corded handset
should be rectified as soon as possible. The standby control
will override any signals from an infrared handset. The
panel can also be used see section 4.6.5. corded handset simply plugs into either of the table handset
4.6.1.7 Control and adjustment of the standard head sockets (item 24 of Fig. 1.1 and item 2, Fig. 4.9). Never
and leg sections are manual operations, but assisted by plug handsets into both sockets at the same time. See
gas spring supports. Adjustment of these sections is section 4.6.4 for details of button function and section 4.5.1
detailed in section 4.6.6. for an explanation of the battery state LEDs.
4.6.2 Tabletop powered motions 4.6.3.2 When using the corded handset always ensure
that the cord is well clear of any moving parts, pinch points
and possible entrapment from table movements. Also ensure
WARNING that the cord will cater for any table movement and that
Ensure you have read and understood the such table movements do not stretch the cord excessively.
safety warnings listed in section 2.1 before 4.6.3.3 When not in use it is suggested the handset is
using any of the powered motions. clipped onto the accessory side rails (item 26, Fig 1.1). Note
that when clipped onto the table in this way it can be operated
4.6.2.1 Powered tabletop motions are, with one finger, without the need to hold it in the other hand.
Trendelenburg, height, tilt, brake, traverse, flexion and The handset must be compatible with the side rails fitted to
return to level. Each motion in either direction is controlled the table, see sections 1.1.2 and 4.8. The UK handset will
by pressing and holding the appropriate handset, footswitch also clip onto head/leg end blocks (see Fig. 1.1).
or standby control panel button. The table must be switched
‘on’ for powered motions to operate (see 4.6.1.4 above).
Section 4.6.4 details handset button functions, section 4.6.5
provides standby control panel information.
4.6.2.2 All powered tabletop movements are
programmed to provide a gentle transition from stationary
into the required motion and back again to stationary
(i.e. initially the motion starts slowly and then speeds up to
maximum speed and then slows again before coming to a
stop). This also enables accurate positioning at slow speed
by using repeated short button presses. Alternatively long Fig. 4.9 Controller sockets and
duration button presses allow large changes in tabletop release buttons
position at maximum speed.
4.6.2.3 Most powered tabletop motions will pause 4.6.3.4 To remove the handset plug from its socket
briefly when they pass through the level position (Note: the release button (item 1, Fig. 4.9) must be pressed and
not applicable to height and traverse). This is to enable held in during removal. When removing the corded
each motion to be returned to a level position individually.handset plug from its socket always grip and pull the plug,
do not pull the cord only as this may damage the cord or
4.6.2.4 Be aware that a lowered leg section could hit internal connections.
the table base when the top is lowered, or the head section
could collide with an anaesthetist’s chair during a 4.6.3.5 Handset buttons provide a tactile feedback
Trendelenburg movement. Also note the warnings at the to enable the user to detect when a button has been
beginning of section 4.6. pressed or released, this is in the form of a ‘pop’ or ‘click’.
Also, all buttons are ‘de-bounced’ to ensure that only
4.6.2.5 Always ensure that there is adequate space deliberate button presses are responded to (i.e. an
around the table for the movement selected and that the accidental quick activation is ignored).
movement required will not cause injury to patient or
medical personnel. Look for possible trap and or pinch
points between parts of the table and stationary objects.
T-IM102d P21/48
Fig. 4.10 Handset button functions and serial labels
(labels are examples only and text on actual table will be different to that shown above)
P22/48 T-IM102d
T20-a
OPERATION TABLE
4.6.4 Handset button functions 4.6.4.3 Tilt, Trendelenburg and Break motions will
pause momentarily when passing through a ‘level’ position.
4.6.4.1 The handset buttons have their function Note that ‘Break’ is level when the trunk sections are in-
indicated by graphics printed on the handset, see Fig. 4.10 line with each other, they will not be horizontal unless
and section 1.5.3. Trendelenburg is horizontal.
4.6.4.2 To activate the required tabletop motion press
4.6.5 Standby control panel
and hold the appropriate button until the required amount
of change in tabletop position has been achieved, then
release the button (also see section 4.6.1.5). Refer to WARNING
section 4.6.2.2 for details of the tabletop movement control. The standby control panel MUST be used
The actual motion for each button is fully detailed below, with extreme care for Trendelenburg move-
refer to Fig. 4.10 for button reference. Note: Buttons B1 ments. All programmed safety features are
and B2 provide the described movement for a table top in
the normal orientation (head section in long trunk) with the overridden in this mode (e.g. should the
patient’s head on the head section. If the patient orientation tabletop hit an object, motor protection is
is reversed button functions B1 and B2 are also reversed. inhibited and damage to them could occur).
B1 Trendelenburg (orange button) - Press to rotate the Also do not exceed 30° of Trendelenburg
tabletop in the Trendelenburg (head down) direction. (or reverse) from the standby panel.
B2 Reverse Trendelenburg (orange button) - Press to
4.6.5.1 The standby control panel is located on the
rotate the tabletop in the Reverse Trendelenburg
side of the column (item 12, Fig. 1.1). The panel has five
(head up) direction.
function buttons (OB1 to OB5, see Fig. 4.11) and two
B3 Height down - Press to lower the whole tabletop. direction buttons (OB6 and OB7, see Fig 4.11). The motion
of the function button is described by its graphical symbol
B4 Height up - Press to raise the whole tabletop. (symbols are shown and detailed in section 1.5.5).
B5 Tilt left - Press to tilt the whole tabletop down on the 4.6.5.2 When controlling the tabletop from this
left-hand side when viewed from the long trunk end. standby control panel certain inbuilt safety features are
overridden. Also there is no ‘soft start’ to the powered
B6 Tilt right - Press to tilt the whole tabletop down on Trendelenburg movements and the tabletop will not pause
the right-hand side when viewed from long trunk end. as it passes through the level (Trendelenburg) position.
Only five functions can be controlled from this panel
B7 Break down - Press to move the tabletop in the break
Trendelenburg, Height, Tilt, Break and Traverse.
down (Extension) direction.
4.6.5.3 To operate the tabletop from this panel
B8 Break up - Press to move the tabletop in the break involves selecting a required function by ‘pressing and
up (Flexion) direction. holding’ the appropriate function button (i.e. OB1 to OB5,
Fig. 4.11) and then selecting a direction for this function by
B9 Traverse Cranially - Press to move the tabletop
‘pressing and holding’ the appropriate direction button (OB6
Cranially (i.e. the tabletop will traverse and increase
or OB7, Fig. 4.11). The functions button graphic has two
the distance between the end of the long trunk section
arrows indicating two possible directions for each function.
and the column).
T-IM102d P23/48
Pressing the upper or lower direction button will select the 4.6.6 Head and leg section control
direction indicated by the corresponding arrow on the
function button. The table will move whilst both the function 4.6.6.1 Both of these sections (head and leg) are
button and the direction button are pressed, releasing either operated in the same way. The section’s weight is
will stop the motion. Motions for each button are fully supported during adjustment and held locked in position
detailed below:- by two gas springs (items 5 and 11, Fig. 1.1). Lowering the
section release handle during adjustment will automatically
OB1 Trendelenburg - Press and hold this button, then lock the section in place eliminating the possibility of the
press required direction button to rotate the tabletop in section dropping under its own, or the combined weight of
the Trendelenburg or reverse Trendelenburg direction the section and any other load placed upon it.
(button OB6, Fig. 4.11 is for Trendelenburg; button OB7,
Fig. 4.11 is for reverse Trendelenburg) 4.6.6.2 To adjust either section (having noted the
warnings at the beginning of section 4.6) grasp the end of
OB2 Height - Press and hold this button, then press
the section on both sides with thumbs up and the finger
required direction button to change the height of the
tips resting on the release handle (items 10 or 16, Fig. 1.1)
tabletop (button OB6, Fig. 4.11 is for Height up;
underneath the section. Whilst supporting the section in
button OB7, Fig. 4.11 is for Height down).
place (it may move when the release handle is raised),
OB3 Tilt - Press and hold this button, then press gently lift the release handle with the finger tips until the
the required direction button to tilt the tabletop section is free to move up or down (see 1, Fig. 4.12).
(button OB6, Fig. 4.11 is for Tilt down on the left;
button OB7, Fig. 4.11 is for Tilt down on the right, 4.6.6.3 When raising the section the gas springs will
when viewed from the long trunk end of the table). assist the movement, when lowering the section gentle
pressure may be required to overcome slight resistance
OB4 Break - Press and hold this button, then press of the gas springs. Adjust the section by rotating it up or
required direction button to move the tabletop in the down to the required position (see 2, Fig. 4.12).
Break up or down direction (button OB6, Fig. 4.11
is for Break up or Flexion; button OB7, Fig. 4.11 is 4.6.6.4 When the correct position has been obtained
for Break down or Extension). release the locking bar. The section is now locked in place
automatically by the gas springs (see 3, Fig. 4.12).
OB5 Traverse - Press and hold this button, then
press required direction button to traverse the
tabletop (button OB6, Fig. 4.11 is for cranially; button
4.7 Fuse replacement
OB7, Fig. 4.11 is for caudally). 4.7.1 The fuses are located as shown in Fig. 1.1 (items
8 and 17). Fuses adjacent to the mains inlet socket are
only applicable to the mains supply for the battery charger.
Fuses in the side of the table base are applicable to the
main batteries (left hand fuse) and standby batteries (right
hand fuse) adjacent to standby battery switch.
4.7.2 These fuses are replaced as follows:-
i Remove the mains cord from the table before
replacing any of the fuses and switch the table ‘off’
with switch item 21, Fig. 1.1.
ii Turn the fuse cover cap anticlockwise to remove
the cap with the fuse inside.
iii Ensure you have the correct fuse (consult the
markings adjacent to the fuse or the Technical Data,
section 6.0).
iv Place the new fuse into the cap and replace the cap
by screwing it in clockwise.
4.8 Using accessories
WARNING
When fitting any accessory ensure that it has
been securely attached and the maximum
loading has been noted before using it to
Fig. 4.12 Adjusting a head or leg section support part of the patient’s weight. Also see
guidance notes in section 2.4.
P24/48 T-IM102d
T20-a
OPERATION TABLE
Note: The use of the corded handset has been covered in Tables with Serial number T2AA-3A-2037 or later
section 4.6.3 and 4.6.4. Use of the standard table sections order the kit (T00-521-0001 or T00-551-0001,
have been covered in the sections listed below:- depending on tables side rail) for tables with an earlier
Head section - section 4.3 & 4.6.6 serial number order the handset only (T00-520-0001
Leg section - section 4.3 & 4.6.6 or T00-550-0001, depending on tables side rail).
Mattresses - section 4.4 4.8.2 The following accessories can be used with the
4.8.1 The catalogue (REF) numbers for the standard T20-a table. Their use is self explanatory but always ensure
T20-a accessories are shown in the following list, where all clamps are fully tightened before using to support the
applicable these are supplied with a ‘User/Service applicable part of the patient’s weight. Some of these require
Handbook’. This list may not be exhaustive as new additional clamps to secure them to the table this is indicated
accessories are added to the range . Please check with in the following list. Again this list may not be exhaustive as
Eschmann Equipment for the latest additions. new accessories are added to the range. Please check with
IMPORTANT NOTE: Please check with Eschmann Eschmann Equipment for the latest additions.
Equipment or their local agent before using access- REF Description
ories that fit onto the side rails of tables supplied with
non-standard Denyer side rails, see section 1.1.2. 81-201-10 Circular socket accessory clamp
81-213-11 Rectangular socket accessory clamp
REF Description 81-220-16 Circular socket rotary accessory clamp
T00-311-0001 Head section², UK side bars 81-250-15 Lithotomy supports **
T00-312-0001 Head section², Euro. side bars 81-264-10 Shoulder rests
T00-321-0001 Leg section², UK side bars 81-286-18 Anaesthetic screen, single *
T00-322-0001 Leg section², Euro. side bars 81-290-10 Arm support, Carter Braine #
T00-520-0001 Infrared handset only (fits UK side rail) ²² 81-300-19 Narrow arm table with pad #
T00-521-0001 Infrared handset only (fits Denyer side rail) ²² 81-314-14 Arm table, square without support ##
T00-540-0001 Corded handset (fits UK side rail) 81-344-13 Instrument table *
T00-541-0001 Corded handset (fits Denyer side rail) 81-367-85 Douche tray
T00-550-0001 Infrared handset kit (fits UK side rail) ²² 81-368-23 Perspex arm support, right-angle
T00-551-0001 Infrared handset kit (fits Denyer side rail) ²² 81-378-38 Foot stocks
T20-411-0001 Mattress, 50mm, complete set 81-404-13 Pelvis support #
T20-412-0001 Mattress, 75mm, complete set 81-412-15 Buttock support #
T20-421-0001 Mattress, 50mm, trunk section 81-436-17 Pelvis / chest support #
T20-422-0001 Mattress, 75mm, trunk section 81-446-13 Wristlet #
T20-431-0001 Mattress, 50mm, head/footrest 81-462-17 Patient restraint strap
T20-432-0001 Mattress, 75mm, head/footrest 81-462-33 Cot side, tall
T20-441-0001 Mattress, 50mm, leg section 81-463-06 Infusion pole
T20-442-0001 Mattress, 75mm, leg section 81-490-03 Kidney position supports ##
TA-0202001 Width extender², long, UK side bars 81-492-16 Lateral support, adult
TA-0202005 Width extender², long, Euro. side bars 81-504-19 Lateral support, child
TA-0202007 Width extender², short, UK side bars 81-569-13 Laminectomy support pad
TA-0202012 Width extender², short, Euro. side bars 81-569-80 Laminectomy support **
TA-0203001 Foot rest/extension² 81-812-17 X-ray cassette holder, lateral
TA-0301001 Lightweight leg 81-820-19 Arm table, square with support
TA-0302001 Ophthalmic head section 81-827-52 X-ray cassette tray, end loading
TA-0303004 Accessory attachment bar 81-828-41 X-ray cassette tray, side loading
TA-0401013 Accessory trolley * Requires one 81-220-16
TA-0404003 Footswitch ** Requires two 81-220-16
TA-0501002 Mattress, 50mm, width extender, long # Requires one 81-201-10
TA-0501006 Mattress, Ophthalmic head section ## Requires two 81-201-10
TA-0501008 Mattress, 50mm, width extender, short
TA-0501009 Mattress, 75mm, width extender, long Eschmann also supply a range of support pads, suitable
TA-0501010 Mattress, 75mm, width extender, short for use on the T20-a table for information contact Eschmann
TA-0601011 Leg abduction support Equipment at the address on the front cover.
81-342-00 Perineal instrument tray 4.8.3 Spare mains leads can be ordered under the
²
These accessories come without a mattress which following Spare Part Numbers:
should be ordered separately, see list above. 391177 - Australian mains lead.
²²
When ordering an infrared handset check tables 714188 - Standard U.K. mains lead.
Serial Number and order as follows:
T-IM102d P25/48
The traversing top allows full access for ‘C’ arm imaging. Additionally there
is a full length X-ray cassette tunnel (accessible from either end).
P26/48 T-IM102d
T20-a
OPERATION TABLE
4.9 Patient positioning
WARNING
Do not use any table sections or accessories
without the correct Eschmann mattress or pad
in place (except for accessories such as the
lightweight leg section which can be used for
short periods of use, without a mattress,
during patient positioning). Mattresses are an
important part of the antistatic pathway and
help prevent pressure sores developing.
P28/48 T-IM102d
T20-a
OPERATION TABLE
T-IM102d P29/48
Fig. 4.16 Maximum patient weight v table position graphs
(head and leg section reversed).
P30/48 T-IM102d
T20-a
OPERATION TABLE
(Note: For maximum patient weights see Fig. 4.15 and Fig. 4.16)
T-IM102d P31/48
Fig. 4.18 Patient in various ‘Prone’ positions
(Note: For maximum patient weights see Fig. 4.15 and Fig. 4.16)
P32/48 T-IM102d
T20-a
OPERATION TABLE
(Note: For maximum patient weights see Fig. 4.15 and Fig. 4.16)
T-IM102d P33/48
Fig. 4.20 Patient in various ‘Chair’ positions
(Note: For maximum patient weights see Fig. 4.15 and Fig. 4.16, but note that with
the table and patient positioned as above these are basically stable positions)
P34/48 T-IM102d
T20-a
OPERATION TABLE
(Note: For maximum patient weights see above and Fig. 4.15 and Fig. 4.16, note that table
stability in the top two graphics is the lowest and can only be used for the lighter patient)
T-IM102d P35/48
Fig. 4.22 Patient in ‘Lithotomy’, ‘Prone Laminectomy’ and ‘Lateral Nephrectomy’
(Note: For maximum patient weights see Fig. 4.15 and Fig. 4.16)
P36/48 T-IM102d
T20-a
OPERATION TABLE
T-IM102d P37/48
5.2.2 Table, accessories mattresses and pads ii Remove all mattresses and check them for any cuts,
scuffs or other damage and replace as required.
5.2.2.1 Disinfect the operation table, accessories,
mattresses and pads as follows: iii Check that the table covers are not cracked,
chipped, or otherwise damaged and arrange
i Disassemble the table and/or accessories as far replacement as required.
as possible without the use of tools, remove
mattresses and pads. iv Check the table for any signs of wear or damage
that requires attention and arrange remedial action
ii Scrub all surfaces and crevices with hot (55°C) if required.
neutral (pH7) detergent solution (diluted in
accordance with the manufacturers instructions) to v Ensure batteries have been regularly charged and
remove all visible contamination. Use a small brush charge them if required.
to clean areas of limited access repositioning the 5.3.1.2 Once every six months the table should receive
table as required to gain access to all surfaces. For a safety check and service as detailed in the service manual
stubborn mattress stains see section 5.1.3.2. (which provides full details of part replacement, safety
iii Wash down with hot (55°C) water. checks and routine maintenance). During this service the
tables calibration should be checked and if required it
iv Dry all surfaces with absorbent paper. should be recalibrated. Also at least once every year the
v Wash down all surfaces and crevices with one of electrical conductivity of the table should be checked. To
the solutions below: arrange these contact the Eschmann After Sales Service
Department, see inside front cover for contact details.
v a 70% solution of industrial methylated spirit in
water. 5.3.2 Storing the operation table (long term)
v a 1000 to 5000mg/litre solution of hypochlorite
5.3.2.1 The Operation table should be covered, and
in water (see Caution note above).
stored in a clean environment, with no extremes of
vi Wash down thoroughly with clean water. temperature, see the Technical Data section for details.
vi Dry all surfaces with absorbent paper, to avoid 5.3.2.2 The table should be stored:
damage do not leave mattresses wet in contact with
v With the base in the brake position.
another surface.
v With the head and leg sections fully lowered.
vii Dispose of all cleaning material and solutions in
accordance with authorized disposal procedures. v With the tabletop level in both planes.
v With ‘on/off’ switch (item 21, Fig. 1.1) ‘off’.
5.3 Care 5.3.2.3 Every two weeks:
5.3.1 Operation table and sections v Charge the table batteries (see section 4.5.1)
until both charging LEDs are green.
5.3.2.4 The storage maintenance inspection must be
WARNING conducted by a trained engineer. When the inspection is
The head and leg section gas-springs are finished, the ‘Storage Maintenance Record’ (see page 39)
filled with gas at high pressure, do not try should be completed and signed.
to open them. The gas-springs should be CAUTION
replaced immediately any signs of leaking Do not store mattresses and pads with
or deterioration in performance are noted objects (especially objects having sharp
(e.g. movement of section when locked). edges and protrusions) resting on their
CAUTION padded surface as this could damage them.
Always store them flat.
Do not lubricate the head or the
leg section, gas-springs. 5.3.2.5 Mattresses and pads should not be left in direct
or excessive heat and they must be stored flat. Do not
5.3.1.1 Once a week: leave them with other accessories resting on them or
i Remove fluff and debris from the head and leg leaning against another object.
section guide pin location sockets and if required
spray the pins and into the sockets with a suitable
light lubricant such as WD40.
P38/48 T-IM102d
T20-a
OPERATION TABLE
T-IM102d P39/48
5.3.3 Lifting the operation table in both directions (i.e. neutral tilt and Trendelenburg) and
traverse the top so the long trunk is central to the column
WARNING as shown in Fig 5.1. Lower the short trunk and then the
table top to their maximum limits.
The table is heavy and at least four strong
people are required to lift it. Ensure that 5.3.3.5 When lowering the table after the lift take care
not trap feet under the tables base. Inspect the table for
adequate precautions are taken (e.g. wear
any signs of damage and check all functions prior to placing
protective shoes, use the correct straps). the table back into service.
5.3.3.1 The T20-a table should only be lifted as a last 5.4 Maintenance
resort. Ideally it should be placed on a trolley directly from
the delivery vehicle, or moved on the base of the delivery
5.4.1 General
packing case (overseas only) and then rolled down the
ramp provided. 5.4.1.1 The Service Manual, which can be ordered
from the Eschmann After Sales Service Department,
5.3.3.2 If required the T20-a operation table should only
contains the routine service schedule and details of how
be lifted by placing suitable webbing straps underneath
to replace parts when required. Some parts of the table
the table base in the positions indicated in Fig. 5.1 (having
are not user serviceable and this is detailed in the Service
raised the table into the ‘castor’ position) and observing
Manual, also refer to section 5.3.1 and 1.1.5.
the notes that follow. The table should only be lifted the
minimum amount required and not carried. Lift the table
sufficiently high to allow a fully decked pallet to be slid 5.4.2 Fault diagnosis
underneath. The table should be placed into its ‘braked’ 5.4.2.1 The Table 1 lists possible causes for faults
orientation whilst on the pallet. Transport the table on the and conditions that may arise. Some of these may require
pallet using a forklift truck or similar equipment ensuring further investigation by trained engineers in conjunction
the table is strapped securely to the pallet. with the Service Manual and circuit diagrams. Where the
5.3.3.3 Extreme care should be taken to pad the straps remedy will require an engineer to rectify the fault or
where they pass the base covers and the table top sections condition this is indicated by the phrase “Engineer to...” in
to avoid damage. Take special care not to cause damage the remedy column. If any fault persists (e.g. blown fuse)
to the lower edge of the base covers. this should be investigated by an engineer.
yy
yy yy
yy
yy
yy yy
yy
yy yy
Fig. 5.1 Lifting the T20-a operation table
P40/48 T-IM102d
T20-a
OPERATION TABLE
Table not configured for infrared Arrange to have table configured for an
handset. infrared handset by a Service Engineer.
T-IM102d P41/48
TABLE 1 - FAULT DIAGNOSIS (continued)
Table will not move Table base in ‘braked’ position. Place table into the ‘castor’ or ‘wheel’
when pushed. orientation as detailed in section 4.2.
Foreign object trapped under Check for foreign object and remove
a wheel or castor. it if found.
Table difficult to manoeuvre. Table in ‘wheel’ orientation. Place table into ‘castor’ orientation.
Table difficult to move Table in ‘castor’ orientation. Place table in ‘wheel’ orientation.
in a straight line.
Table not stable on floor Table is not in ‘braked’ position. Place table into the ‘braked’ orientation
and moves when pushed as detailed in section 4.2.
Table not stable on floor Floor uneven or object under Move table to a flatter area
and wobbles when ‘braked’. a wheel or castor. or remove object.
No charge LED ‘on’ Mains supply faulty or not Check mains supply, or switch ‘on’
when table connected switched ‘on’. at mains supply.
to mains.
Mains cable not connected Reconnect mains cable correctly at
correctly at both ends. each end.
Charge LED colour does Table batteries at fault. Engineer to change batteries.
not change to green (even
after twelve hours).
Table section does not Weight of section not supported Support section weight when attaching
attach or release during attaching or removal. and removing section (i.e. gently lift end
easily from table. furthest from table).
Unexpected table movement. Another infrared controller Check for another infrared controller
being used locally. and discontinue its use, or engineer to
change table infrared code.
Table audible alarm Table left switched ‘on’ and no See section 4.12.5 and either stop the
sounding (‘beeping’) command input given for 6 hours. alarm (if table still in use) or switch the
table ‘off’ (if table not in use). If the table
is not in use, place it on charge.
P42/48 T-IM102d
T20-a
OPERATION TABLE
The following table movements are maximums and cannot 6.5.3 Internal battery charger
be assumed to be available in all combinations of table Input
positioning (see Fig. 6.1 for illustration, the number in
100-240V a.c. 50-60Hz 2.5A max
brackets after the item details which part). For example,
with the table in its normal configuration and at minimum Output
height the maximum Trendelenburg and leg section 29.2V d.c. 2A (max.) when charge state LED
movements cannot be achieved due to the proximity of red or orange, 27.6V d.c. when charge state
the floor. Similarly with a large amount of tilt set, maximum LED green
Trendelenburg cannot be achieved. 6.5.4 Fuses
Max. Trendelenburg (1) .. .. .. 35° 6.5.4.1 External fuses
Max. reverse Trendelenburg (1) .. .. 35° Mains input fuses (item 17, Fig. 1.1)
Max. extension (4) .. .. .. .. 230° 2 x T4A (5 x 20mm) 250V
Max. flexion (4) .. .. .. .. .. 90° Battery fuses (item 8, Fig. 1.1)
Head section (5) .. .. .. .. ±45° 2 x T20A (6.35 x 32mm) 500V
Leg section (3) .. .. .. .. -100° +55° Note: The left hand fuse is linked to the main batteries
Max. lateral tilt (7) .. .. .. .. ±18° and the right hand fuse (i.e. nearest to the standby
battery switch) is linked to the standby batteries.
T-IM102d P43/48
1 4
3
7
These are maximum movements for each aspect, they may not be available in certain combinations (e.g. maximum
tilt and maximum Trendelenburg). Movements that could cause damage cannot be catered for (e.g. position of the
leg section). Also see sections 6.2 and 6.3.
Fig. 6.1 Major dimensions and movements
P44/48 T-IM102d
T20-a
OPERATION TABLE
Mains plug fuse .. .. 10A (type is 6.7 Use with other equipment
dependent on plug which may have been changed
from that originally supplied) 6.7.1 Electrosurgical equipment (h.f.)
The T20-a operation table has been designed to minimise
6.5.4.2 Internal fuses (only accessible by engineer)
the possibility of accidental electrosurgery burns and can
Battery fuses (joining each battery pair) be used in conjunction with electrosurgical equipment.
2 x 30A blade type (1 per battery pair) However contact with any metal surfaces (e.g. table side
bars, or other equipment etc.) can cause burns during
electrosurgery and must be avoided.
6.5.5 Duty cycle
6.7.2 Defibrillation equipment
This symbol is used to indicate the table’s duty cycle With the mains cord attached the equipment has a
which, in the worst case, is ‘60s : 600s’, the ratio of the defibrillator proof applied part with type BF protection
operating time to the sum of the operating time and the against electric shock.
ensuing interval (see note below). Each motor drive has
its own duty cycle and this is dependent on loading and 6.7.3 Other
table position as detailed below. When the T20-a is used in conjunction with other infrared
Trendelenburg 1 : 4 (60s : 240s) at a maximum torque controlled devices, their controllers should be checked to
of 417Nm (e.g. 135kg load offset 31cm from the ensure that no interference between them and the T20-a
fulcrum, or a 300kg load offset 14cm from the is possible, prior to surgical intervention. Also see
fulcrum. Where this load is the patient and section 1.6.3.
accessories weight and the offset is how far the loads
centre of gravity is, from the centre of the column).
6.8 Standards compliance
Traverse 1 : 4 (60s : 240s) when the table is horizontal The table has been designed and built to comply with the
and at the maximum load of 300kg, or at 17.5° with following international standards:
a patient weight of 225kg, or at 35° with a patient BS EN 60601-1: 1990 (issue 2, October 1997)
weight of 150kg. BS EN 60601-1-2: 1993
BS EN 60601-2-46: 1998
Break 1 : 10 (60s : 600s) at maximum patient BS EN 60601-1-4: 1997
weight of 300kg.
6.9 Environmental conditions
Height 1 : 4 (60s : 240s) at maximum patient
weight of 300kg (not offset). 6.9.1 Operating environment
Tilt 1 : 4 (60s : 240s) at maximum patient The table has been designed to operate in the following
weight of 300kg. environment:
Note: The duty cycles above are all for the worst case (i.e. Temperature range .. .. 10°C to +40°C
maximum loads). For reduced loading the above duty Pressure range .. .. .. 69KPa to 106KPa.
cycles can be increased. Humidity range .. .. .. 30% to 75% RH
non-condensing.
6.6 Classification and symbology
6.9.2 Transport & storage environment
All the symbols used on this table are shown and explained
in section 1.5. These detail the safety category and class The table can be transported and stored safely, in the
of this table as marked on the table or section by the use following environment:
of these symbols. Temperature range .. .. -30°C to +50°C
Pressure range .. .. .. 69KPa to 106KPa
Humidity range .. .. .. 30% to 90% RH
non-condensing.
T-IM102d P45/48
APPENDIX 1
Ensure that ALL the personnel listed above have been fully trained in the safe use of this T20-a Operation
table. This should include a thorough understanding of all the safety notes and cautions contained within
this ‘Instructions for use’ manual.
Photocopy this page as required and ensure that ALL personnel have not passed the review date without
a refresher course. Append additional documents applicable to your unit/hospital (e.g. cross contamination
procedures) and list them below in the space provided, if they should be included in the training.
ADDITIONAL DOCUMENTS
P46/48 T-IM102d
T20-a
OPERATION TABLE
APPENDIX 2
DECONTAMINATION CERTIFICATE
This is to certify that the above equipment/item has been decontaminated in accordance with the
attached procedure and that this certificate has been issued by the authorized person detailed
below. The equipment/item has also been suitably marked to show it has been decontaminated in
accordance with the attached procedure and that this certificate has been issued.
T-IM102d P47/48
APPENDIX 3
* BAT’Y IND’N - Indicate battery state after charge. ‘G = Green or OK’, ‘A = Amber or Low’, or ‘R = Red or Critically Low’ as
indicated on the corded handset. Please photocopy this page and use to record battery charging data.
P48/48 T-IM102d
Eschmann Equipment, Peter Road, Lancing, West Sussex, BN15 8TJ, England.
Tel: +44 (0) 1903 753322. Fax: +44 (0) 1903 766793. www.eschmann.co.uk