Professional Documents
Culture Documents
Calibration of Micropipettes: Test Methods and Uncertainty Analysis
Calibration of Micropipettes: Test Methods and Uncertainty Analysis
www.elsevier.com/locate/measurement
a
Instituto Português da Qualidade (IPQ), Rua António Gião, 2, Caparica, Portugal
b
Nederlands Meetinstituut (NMi), Schoemakerstraat 97, 2628 VK Delft, The Netherlands
Abstract
Micropipettes or piston pipettes are used to make the most of volume measurements in fields like health, chemistry,
biology, pharmacy and genetics. Laboratories must ensure that obtained results using these instruments are reliable, for
that is necessary to calibrate micropipettes using the correct method and estimate uncertainties. The objective of this work
is to present the best method to calibrate micropipettes according to the describe measurement conditions. A model for the
calculation of the measurement uncertainty is also described.
Ó 2006 Elsevier Ltd. All rights reserved.
0263-2241/$ - see front matter Ó 2006 Elsevier Ltd. All rights reserved.
doi:10.1016/j.measurement.2006.05.012
E. Batista et al. / Measurement 40 (2007) 338–342 339
piston aspirates the liquid. To expel the liquid the all, the beaker must be filled with 5 mm film of
piston is slid between the volume defining marks. oil. Due to the fact that the oil does not evaporate
The volume delivered by the micropipette is deter- because it has a very high vapour pressure the bea-
mined using a gravimetric method and water as the ker can be weighted any time before the start of the
calibration liquid at a temperature of (20 ± 0.5) °C. measurements. As in the others methods, the micro-
The calibration of micropipettes was carried out pipette is wetted five times and the tips are replaced
using four different gravimetric methods based on as described before. A new tip is wetted and filled
different ISO standards. with the testing volume and delivered into the bea-
In all four methods a suitable mass comparator is ker, with caution in order to ensure that the water
needed for registering the mass delivered for a given goes beneath the oil. It is important not to deliver
pipetted volume. The water and room temperature near the beaker walls. After delivering, the beaker
must be carefully controlled during the measure- is weighted again and the mass obtained by differ-
ments. ence is a measure for the volume of the micropi-
The first method, the ‘‘normal’’ method, is based pette. The procedure is repeated ten times. Within
on ISO 4787 [1] and ISO 8655 [2]. The micropi- each measurement the mass of the beaker must be
pette is wetted five times to reach equilibrium with registered.
respect to humidity. Then the tip is changed and Five tests were performed to each micropipette of
wetted again. The weight of the dry beaker is deter- 5 ll, 100 ll, 250 ll and 1000 ll in all four methods.
mined, and so is the volume pipetted. The beaker is
weighted again and the mass obtained by difference 3. Evaluation of measurement uncertainty
is a measure for the volume of the micropipette. The
beaker is cleaned after each volume additions and The evaluation of the uncertainty of all four
this procedure is repeated ten times. methods follows what is commonly referred as
The second method is the ‘‘evaporation’’ method ‘‘mainstream GUM’’ [3] and is similar to the one
based only in ISO 8655. The micropipette is wetted described in the work of Batista et al. [4].
five times and then the tip is changed and wetted To determine the volume of a micropipette the
one time. The weight of the beaker with a 3 mm following formula (1) is used:
water film is determined (m0). A stopwatch is started
1 q
and the testing volume is pipetted. The beaker is V 20 ¼ ðI L I E Þ 1 A
qW q A qB
weighted again (m1), then the test volume is pipetted
again and added to the beaker (m2). The mass ½1 cðt 20Þ ð1Þ
obtained by difference (m1 m0, m2 m1, . . . , where
m10 m9) is a measure for the volume of the micro- V20 volume, at temperature of 20 °C, in ll;
pipette. The stopwatch is stopped after the tenth IL weighting result of the recipient after the
measurement and the weighing vessel is left on the pipetting in mg;
balance for the time used for the ten measurements IE weighting result of the recipient before the
and its mass is recorded (m11). The value determined pipetting, in mg;
due to evaporation loss is added to each measure- qW water density, at calibration temperature t,
ment. To calculate the evaporation loss the follow- in mg/ll;
ing formula is used: (m10 m11)/10. qA air density, in mg/ll (0.0012 mg/ll);
The third method is the ‘‘without evaporation’’ qB mass pieces density (8.0 mg/ll);
method, is also based on the ISO 8655 but the evap- c material cubic thermal expansion coeffi-
oration loss is not taken in count. The micropipette is cient of the micropipette, in 1/°C;
wetted five times, and then the tip is changed and wet- t water temperature, in °C.
ted once more. The weight of the beaker with a 3 mm
water film is determined and the testing volume is Defining 1/(qW qA) as A, 1 qA/qB as B,
pipetted. The beaker is weighted again and the mass 1 c(t 20) as C, and m as IL IE, Eq. (1) can
obtained by difference is a measure for the volume of be rewritten in a simpler way as
the micropipette. This cycle is repeated ten times.
V 20 ¼ m A B C
The last method is the ‘‘without evaporation
using oil’’ method. This method is very similar to Some of the variables in (1) depend on others. A
the third and is also based in ISO 8655. First of complete overview is given in Fig. 1.
340 E. Batista et al. / Measurement 40 (2007) 338–342
Table 1
Evaluation of uncertainty of a micropipette volume
Variable Xi Estimate xi Distribution Standard uncertainty u(xi) Sensitivity coefficient ci Contribution ui(y) (ll)
M (lg) 5.033 Normal 0.0088 1.00 0.0088
dmres (lg) 0 Rectangular 0.001 1.00 0.001
dmcal (lg) 0 Normal (k = 2) 0.005 1.00 0.005
t (°C) 19.5 Normal 0 0.00121 0
dtcal (°C) 0 Normal (k = 2) 0.05 0.00121 0.0000606
qW (lg/ll) 0.9983 Rectangular 0.0000116 5.06 0.0000588
qA (lg/ll) 0.0012 Rectangular 0.000000289 4.43 0.00000128
qB (lg/ll) 7.96 Rectangular 0.0346 0.000096 0.00000332
c (1/°C) 0.000240 Rectangular 0.00000289 2.52 0.00000729
V20 (ll) 5.05 U(V20), k = 2 0.02
5.5 255.0
5 μl micropipette 250 μl micropipette
5.4 254.0
5.3 253.0
5.2 252.0
Volume (μl)
Volume (μl)
5.1
251.0
5.0
250.0
4.9
249.0
4.8
248.0 With evaporation Without evaporation
4.7 Normal With Oil
With evaporation Without evaporation 247.0
4.6 Normal With oil 1 2 3 4 5
4.5 Test
1 2 3 4 5
Test
Fig. 4. Calibration of a 250 ll micropipette.
Fig. 2. Calibration of a 5 ll micropipette.
100.5 999.0
100.0 996.0
99.5 993.0
990.0
99.0
987.0 With evaporation Without evaporation
98.5 With evaporation Without evaportion Normal With oil
Normal With Oil
984.0
98.0 1 2 3 4 5
1 2 3 4 5 Test
Test
Fig. 5. Calibration of a 1000 ll micropipette.
Fig. 3. Calibration of a 100 ll micropipette.
surements. This can decrease the reliability of the
the major part of the evaporation occurs between method.
additions, and not between the additions and the The use of oil was thought to prevent evapora-
balance stability, so when a correction is made the tion of water while the calibration is undergoing,
measured volume is increased in a incorrect way in order to have better results but after evaluating
when compared with the real value. The correction the results, it is clear that this method does not have
value is a mean value added to each of the ten mea- any benefit over the ‘‘without evaporation’’ method.
342 E. Batista et al. / Measurement 40 (2007) 338–342
In the case of the 5 ll micropipette, the results are The presented uncertainty analysis can also be
even worse. There is a great difficulty in delivering used for other piston operating volumetric equip-
a small quantity of liquid inside the oil without ment and laboratory volumetric glassware if the
touching the oil or the wall of the beaker, which same gravimetric method is applied.
leads to evaporation loss and a reduction in the
volume. References
5. Concluding remarks [1] ISO 4787, 1984, Laboratory glassware – Volumetric glass-
ware – Methods for use and testing of capacity.
[2] ISO 8655-1/2/6, 2002, Piston-operated volumetric apparatus.
Several tests were preformed in order to deter- [3] BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML, Guide to
mine the best method for calibration of micropi- the expression of uncertainty in measurement (GUM),
pettes based on the existing ISO standards. After Geneva, 1995.
the evaluation of results and uncertainty it can be [4] E. Batista, A. Sabrosa, M. Ferreira, I. Castanheira, A.M.H.
concluded that the ‘‘without evaporation’’ method Van der Veen, ‘‘Uncertainty calculation in the calibration of
volumetric laboratory glassware’’, in Metrologia 2000, Brasil,
is the most efficient. 2000.
The described model for the calculation of the [5] F. Spieweck, H. Bettin, Review: Solid Liquid density deter-
measurement uncertainty was described and it mination, Tm – Technisches Messen 59 (1992) 237–292.
proved to be completely suitable for this type of [6] ASTM E542, Standard Practice for Calibration of Laboratory
Volumetric Apparatus, 2000.
gravimetric method.