CV-200 Ventilator Technical Specifications

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

CV 

200 Ventilator System 

CV 2 0 0 V E N T I L A T O R S Y S T E M
 
  CV  200  Ventilator  System  consists  of  CV  200  Ventilator  Unit  and  PAVAN  Air  Medical  Compressor 
mounted  on  a  trolley.  CV  200  Ventilator  Unit  comes  with  a  built‐in  battery  for  2  hours  backup. 
 
Half an hour UPS backup is provisioned for PAVAN Medical Air Compressor.  

 
 
  TECHNICAL SPECIFICATIONS
 

A  CV 200 VENTILATOR UNIT 
Sl No.  Parameter  Specification 

Parameters 

 Assisted/Controlled Mandatory 
Ventilation‐Volume Control (A/CMV‐VC) 
 

 Volume Control‐Synchronized 
Intermittent Mandatory Ventilation (VC‐SIMV) 
 

1   Modes   Assisted/Controlled Mandatory 
Ventilation‐Pressure Control (A/CMV‐PC) 
 

 Pressure Control‐Synchronized 
Intermittent Mandatory Ventilation (PC‐SIMV) 
 

 Pressure Support Ventilation (PSV) 

Input gas supply should be oxygen and air at 2.75 to  
5 bar (40 to 70 psi). Input gas supply must be connected to 
CV 200 ventilator unit from direct line maintained at 
specified pressure and can be used from external panel 
2   Input Gas Supply 
outlets and/or cylinders with necessary attachments and 
adapters for both oxygen and air. Pressure difference 
between air and oxygen must be maintained within 20 psi for 
correct FiO2 delivery. 

4 to 100 b/min (measured value can be within  
3   Respiratory Rate (RR) 
±2 bpm of the set bpm) 

4:1 and 1:1 to 1:8 (measured value can be within ±0.25 of  
4   I:E ratio, I:E 
set value) 

50 to 1600 mL (measured value can be within ±10% of  
5   Tidal Volume (VT) 
set value) 

10 to 65 cmH2O (measured value can be within ± 5 cmH2O  
6   Peak Inspiratory Pressure, PIP 
of set value) (5 cmH2O less than Plimit) 
 
CV 200 Ventilator System 

 
Sl No.  Parameter  Specification 

Positive End Expiratory Pressure,  0 to 50 cmH2O (measured value can be within ± 2 cmH2O of 
7  
PEEP  set value) (increments of 1 cmH2O) 
8   Pressure Support, Psupport  5 to 40 cmH2O 
0.3 to 5 sec ((measured value can be within ± 0.05 sec of 
9   Inspiratory Time, INSP TIME 
 set value)  
10   Peak Flow Rate  135 L/min 
11   Pressure limit, Plimit  10 to 70 cmH2O (VC) (5 cmH2O above PIP in PC) 
12   Pressure Trigger, PTrig  OFF, 1 to 15 cmH2O 
13   Alarm Volume  1 to 12 (minimum to maximum) 
14   Standby/Start  Start or Pause Ventilation 
15   Inspiratory Pause   Available 
16   Volume Accuracy  2% to 3% of the full scale between 10 L/min to 80 L/min 
17   Patient type  Paediatric & Adult 
18   Intended use  ICU or Recovery room 
19   Breathing circuit type  Dual‐limb circuit, disposable  
20   Mechanical pressure relief  80 ± 5 cmH2O 
70 cmH2O (Breathe terminated when Paw reaches 
21   Maximum working pressure 
maximum working pressure) 
22   Inspiratory valve response time  < 10 ms 

23   Minimum expiratory time  20% of cycle time, 0.2 to 0.8 s 

24   Flow measurement  dP sensing proximal to patient 

25   Airway pressure measurement  Proximal sensing 

26   Time measurement  Microprocessor (inside ventilator) 

27   Oxygen monitoring  Galvanic cell 

28   Display   6.5 inch Colour TFT 
Touch screen for parameter selection encoder for value 
29   User interface 
increment/decrement 
30   Nebulizer  Ultrasonic nebulisation (Jet not available) 

31   Flow  2.5 to 65 L/min 

Waveforms and Loops 
Paw vs time, Flow vs time,  Volume vs time, P‐V loop and  
32   Graphics 
F‐V loop 
X : 0 to 10 s 
33   Airway pressure vs time 
Y : ‐5 to 80 cmH2O 
CV 200 Ventilator System 

 
Sl No.  Parameter  Specification 

X : 0 to 10 s 
34   Flow vs time 
Y : ‐150 to 150 L/min 
X : 0 to 10 s 
35   Volume vs time 
Y : 0 to 1.6 L 
X : 0 to 1600 mL 
36   Pressure‐Volume loop 
Y : ‐10 to 80 cmH2O 
X : 0 to 1600 mL 
37   Flow‐Volume loop 
Y : ‐150 to 150 L/min 
Alarms 

38   AC Mains OFF  Power supply unplugged 

39   Patient disconnected  PIP < 4 cmH2O in 3 consecutive breathes 

40   High Inspiratory Pressure  Airway pressure exceeds the set Plimit 

41   Negative Airway Pressure  Pressure falls below ‐10 cmH2O 

42   High Air/O2 Supply  System pressure (internally regulated) exceeds 45 psi 

43   Low Air/O2 Supply  System pressure (internally regulated) drops to 20 psi or less
Measured PEEP is more than the Set High PEEP limit for  
44   High PEEP 
3 consecutive breathes 
Measured PEEP is less than the Set Low PEEP limit for  
45   Low PEEP 
3 consecutive breathes 
Measured VTe is more than the Set High VTe limit for  
46   High VTe 
3 consecutive breathes 
Measured VTe is less than the Set Low VTe limit for  
47   Low VTe 
3 consecutive breathes 
Measured MVe is more than the Set High MVe limit for  
48   High MVe 
3 consecutive breathes 
Measured MVe is less than the Set Low MVe limit for  
49   Low MVe 
3 consecutive breathes 
Measured RR is more than the Set High RR limit for  
50   High Respiratory Rate 
3 consecutive breathes 
Measured RR is less than the Set Low RR limit for  
51   Low Respiratory Rate 
3 consecutive breathes 
52   High (FiO2)  Measured FiO2 is more than the Set High FiO2 limit  

53   Low (FiO2)  Measured (FiO2) is less than the Set Low FiO2 limit  

54   Check Sensor (FiO2)  Oxygen sensor is not connected 

55   Low Battery  Remaining battery level is below 14.5 V ± 0.5 V tolerance  
Remaining battery level is below 8 V ± 0.5 V tolerance or 
56   Check Battery 
Battery is disconnected 
CV 200 Ventilator System 

 
Sl No.  Parameter  Specification 

57   Check Flow Sensor connection  Flow Sensor is not connected 

58   Reverse Flow Sensor  Flow Sensor is reversely connected 

59   Alarm mute symbol  Displays status of audible alarm 

60   Volume symbol  Displays status of audible alarm volume 

61   Battery symbol  Displays charge status of battery 

62   Visual alarm  Flashing RED message display 

Power Supply Requirements* 

63   Power supply  220 V to 240 V AC, 50 Hz 

64   Internal battery  14.8 V/2.2 Ah for more than 2 hours backup 

65   Charging time  10 hours (charges automatically during mains operation) 

66   Power consumption  16 W (typical) 

Mechanical Dimensions 

67   Dimension  421 x 328 x 195 mm (W x D x H) 

68   Weight  11 kg (approximately) 
 

B  PAVAN MEDICAL AIR COMPRESSOR 
Sl No.  Parameter  Specification 

Operating Parameters 

1   Output Pressure  44 to 65 psi (3 to 4.5 bar) internally adjustable 

2   Output Flow (continuous)  >50 lpm at 50 psi (3.5 bar) 

3   Internal Tank Size  2 litres 

4   Oil and Particulate Filtration  5 micron 

5   Air Drier System  Electronic drier 

6   Dew Point Depression  < 5 °C below ambient 
(a) Normal Mode 
7   Modes of Operation 
(b) Service Mode 
8   Display  2 Line LCD display (16x2 character) 

9   Stand‐by Mode  Compressor stand‐by option available 

10   Cut‐in Pressure  40 ± 3 psi (2.8 ± 0.2 bar) 

11   Cut‐out Pressure  48 ± 3 psi (3.3 ± 0.2 bar) 
 
CV 200 Ventilator System 

 
Sl No.  Parameter  Specification 

Equipment Classification 

12   Noise Level  < 60 dBA at 1 m distance 

13   Protection against Electric Shock  Class I 

14   Mode of Operation  Long Term and Continuous 

15   Degree of Mobility  Mobile (with Trolley) 

Parameters Displayed 

16   Outlet Pressure  0 to 999 psi (Compressor air outlet pressure) 

Alarms Displayed 

Alarms when the pressure is low for more than 1 sec in 
17   Low Pressure  normal running. 
Low Pressure : < 30 ± 3 psi (2 ± 0.2 bar) 

Alarms when the temperature of the pump is more than 
18   High Pump Zone Temperature  the specified temperature for more than 30 sec. 
Pump Zone Temperature : > 90 °C ± 2 °C 

Service Mode 

Indicates the temperature of the pump chamber. Display 
19   Pump Zone Temperature 
range : 0 °C to 99 °C 

Indicates the temperature of the ambient air. 
20   Ambient Temperature 
Display range : 0 °C to 99 °C 

Indicates temperature of compressed air at the cooler 
21   Dew Point Temperature  outlet, i.e., < 5 °C below ambient. 
Display range : 0 °C to 99 °C 

Indicates the output pressure of the compressor. 50 psi 
22   Compressor Outlet Pressure  (Factory set) 
Display range : 0 ‐ 99 psi 

Displays compressor usage in hours 
23   Hour Meter 
Display range : 0 to 99999 hours 

Mechanical Dimensions 

24   Dimension  425 x 475 x 448 ± 2 mm (L x B x H) 

25   Weight  42.2 ± 0.5 kg (approximately) 

Power Supply Requirements 

26   Voltage  220 V to 240 V AC, 50 Hz 

27   Fuse  5 A, 250 V fuse in‐line and neutral 

28   Maximum current  2.5 A at 220 V nominal 
 
CV 200 Ventilator System 

C  UNINTERRUPTED POWER SUPPLY (UPS) 
Sl No.  Parameter  Specification 

1   Type of UPS  Online 

2   Input Voltage  230 V AC ± 10%, 1 ph, 50 Hz 

3   Output Voltage  230 V AC, 1 ph, 50 Hz pure sine wave 

4   Output Power Rating  1000 VA 

5   Output Voltage Regulation  ± 1% both mains and battery 

6   Battery  Integrated battery required as single UPS unit 

7   Battery Type  Sealed Maintenance Free 

8   Battery Backup Time @750 W  30 minutes (minimum) 

9   Display  LCD/LED 

10   Dimension (including battery)  200 x 420 x 420 mm (W x D x H) (maximum) 

11   Weight  < 40 kg (including battery)  

12   Noise Level  52 dB maximum at 1 meter from UPS surface 

13   Input connector & cable length  Indian plug, minimum 3 m length 

14   Output Connectors  Minimum 4 Nos. (5 A or 6 A) Indian Type 

15   Operating Temperature  10 °C to 40 °C 

16   Certification  CE/BIS 
 
 
 
 
 
 