Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

AK_96_Flow_path_HCEN12302_01_200Page 1 6/13/2008 1:08:43 PM

AK 96 Fluid unit - flow path


CHVA

C P BiCart

BiCart
CBVA holder*

Mixing chamber Mixing chamber


B
A pump

RIVA
Cond. T FLVA Cond. T
A B pump cell A cell B

Flow Expansion Degassing


switch chamber chamber
Heater pH Cell DRVA* P Flow
(upgrade kit) M pump Cond.
C Cell P
M AIVA Degassing T
P
Y

PR 1 INPS PR 2 pressure
CM

Pure water INVA Flow restrictor


T
inlet
MY

Blood leak
CY

ZEVA
P HPG
Outlet restrictor detector
CMY

Drain
K

T
EVVA Cond.
Draining
cell C*
Heat restrictor
exchangers REVA Suction UF-measuring
pump Deairating 2 1 cell
PD P
chamber

Dialyzer

TAVA BYVA

SAGS FIVA* DIVA

Pure water
A concentrate Fluid connector
B concentrate detection
Dialysis fluid Dialysis fluid
Dialysate
‘filter (U9000S)*
*=optional

HCEN12302 © 2007 Gambro Lundia AB, Rev. 01.2008