Silicate (Silicic Acid) Test: Spectroquant
Silicate (Silicic Acid) Test: Spectroquant
December 2018
1.00857.0001 7. Procedure
Measuring range 1.1 - 107.0 mg/l SiO2 (0.5 - 50.0 mg/l Si):
5 - 500 mg/l Si Reagent Si-2 2.0 ml Add with pipette and mix.
For programming data for selected photometers / spectrophotometers see Leave to stand for 2 min (reaction time A).
www.service-test-kits.com.
Reagent Si-3 4 drops 2) Add and mix.
3. Applications Leave to stand for 2 min (reaction time B), then fill the measurement
Sample material: sample into a 10-mm cell, and measure in the photometer.
All kinds of water
It is recommended to use water for on-line analysis, Cat. No. 101051.
1)
Boiler water and boiler feed water
This test is not suited for seawater. Hold the bottle vertically while adding the reagent!
2)
6. Preparation
• Analyze immediately after sampling.
• The pH must be within the range 2 - 10.
Adjust, if necessary, with sodium hydroxide solution or sulfuric acid.
• Filter turbid samples.