SECTION 4 ADJUSTMENT

4.1
U OPTICAL ALIGNMENT ............................................................. 1
U
4.2
U Mechanical Adjustment ............................................................ 23
U

4.1.1
U Required Tools .................................................................... 1
U
4.2.1
U Piercer Position Adjustment .............................................. 23
U

4.1.2
U Optical Alignment Outline ................................................... 1
U
4.2.2
U Adjusting Rinse Cup Position ............................................ 23
U

4.1.3
U Replacing Flow Cell ............................................................ 2
U
4.2.3
U Adjusting Aspiration Position ............................................. 24
U

4.1.4
U Removing Photo Detector No.30 ........................................ 4 U
4.2.4
U Adjusting RBC Chamber ................................................... 24
U

4.1.5
U Removing XS_FCM Detector Block Cover B ...................... 4 U
4.2.5
U Adjusting Reaction Chamber............................................. 25
U

4.1.6
U Removing Flow Cell Unit .................................................... 5
U
4.2.6
U Sampler Adjustment .......................................................... 26
U

4.1.7
U Adjusting Laser Power ........................................................ 7
U
4.3
U Hydraulic Adjustment ............................................................... 30
U

4.1.8
U Adjusting Laser Diode Angle (when replacing Laser Diode)9 U
4.3.1
U Adjusting RBC Clog Level ................................................. 30
U

4.1.9
U Installing Flow Cell Unit No.11 .......................................... 10
U
4.3.2
U Adjusting HGB Blank ......................................................... 30
U

4.1.10 Adjusting Flow Cell Unit No.11 Angle ................................ 10


U U
4.3.3
U Adjusting Aspiration Sensor .............................................. 31
U

4.1.11 Installing XS_FCM Detector Block Cover B ........................11


U U
4.3.4
U Adjusting Sensitivity .......................................................... 32
U

4.1.12 Confirmation/Adjustment of Laser Diode Initial Position.....11


U U
4.4
U Electric Adjustment .................................................................. 33
U

4.1.13 Adjusting Laser Diode Position .......................................... 13


U U
4.4.1
U PCB No.6365T2 ................................................................ 33
U

4.1.14 Installing Photo Detector No.30 ......................................... 14


U U
4.4.2
U PCB No.2168 .................................................................... 33
U

4.1.15 Adjusting Beam Stopper Position ...................................... 14


U U

4.1.16 Confirmation/Adjustment of Photo Detector No. 28 (1) ..... 15


U U

4.1.17 Confirmation/Adjustment of Photo Detector No. 28 (2) ..... 15


U U

4.1.18 Connecting Oscilloscope ................................................... 16


U U

4.1.19 Optical Alignment (Rough) ................................................ 16


U U

4.1.20 Optical Alignment (fine) ..................................................... 19


U U

XS Series S/M December 13, 2007


May 23, 2008

SECTION 4 ADJUSTMENT
4.1 OPTICAL ALIGNMENT

4.1.1 Required Tools

- Hex wrench (1.5 mm and 2.5 mm)


- Screw Driver (Philips and slot headed)
- Laser power meter
- Latex particles (LATEX SS-071-P (10ML/VIAL)
- Latex particles (LATEX particles 4207A (10ML/VIAL)
and LATEX A-7312)
- Oscilloscope (4) Adjust LD unit X-direction roughly so that laser beam can position
at the center of the Flow Cell image behind the FSC Detector
4.1.2 Optical Alignment Outline location (now unit removed). (4.1.13 Adjusting Laser Diode
Position)
When replacing Laser Diode
Follow from below step 1 (5) Install FSC Detector at the specified position from the Flow Cell.
When replacing Flow Cell (4.1.14 Installing Photo Detector No.30)
Follow from below step 2
(6) Adjust X and Y direction of the LD unit so that laser beam can be
(1) Adjust the angle of the LD unit so that incident and reflect beam positioned at the center of the beam stopper on the FSC Detector.
are aligned using a compact mirror after removing Flow Cell and (4.1.15 Adjusting Beam Stopper Position)
FSC Detector unit. (4.1.8 Adjusting Laser Diode Angle (when
replacing Laser Diode) (7) Adjust the initial position of the SFL Detector (X,Y and Z direction).
(2) Install Flow Cell unit, and then adjust the angle of Flow Cell so that (4.1.16 Confirmation/Adjustment of Photo Detector No. 28 (1)) In
incident and reflect beam are aligned. (4.1.9 Installing Flow Cell case of Flow Cell replacement only (no LD unit replacement), skip
Unit No.11 and 4.1.10 Adjusting Flow Cell Unit No.11 Angle) this step.

(3) Adjust the initial position of the LD unit. (4.1.12 (8) Check FSC signal using an oscilloscope. (4.1.19 Optical
Confirmation/Adjustment of Laser Diode Initial Position) Alignment (Rough)) Adjust Z-direction of LD unit so that FSC
1) Z-direction rough adjustment by aligning the corner of the laser signal width will become narrower than specified limit.
diode cylinder and laser diode base.
2) X-direction rough adjustment by aligning optical base slotted
portion and LD base.
XS Series S/M 4-1
(9) Adjust following units by observing SFL signal using an 4.1.3 Replacing Flow Cell
oscilloscope. Verify the results in Fine mode menu, and then fix
the units. (4.1.20 Optical Alignment (fine)) (1) Disconnect tubing from nipple on SV 11.

- Z-direction of LD unit: to make signal width narrower


- X-direction of LD unit: to make uniform signal
- X-direction of SFL Lens: to make higher signal
- Y and Z-direction of SFL Detector: to make uniform signal

Cautions)
In case LD unit or Flow Cell angle adjustment is not well done,
uniform signal will be seen for a short time only (unstable).
If it is difficult to get SFL signal, check if Flow Cell image at the slit
of the SSC Detector is clear and sharp when shining a flash light
toward flow cell. If NOT, adjust SFL Lens X-direction to get sharp
and clear image on the slit.

(10) Adjust the SSC Detector left or right so that Flow Cell image can
be aligned at the slit of SSC Detector. (Step (9) in 4.1.20 Optical
Alignment (fine)) 11
(11) Adjust X-direction of the FSC Detector so that FSC signal
becomes the lowest. (This means direct laser beam is
interrupted at the beam stopper.) And then, fix the FSC Detector.
(Step (14) in 4.1.20 Optical Alignment (fine))

Caution)
Adjustment for FSC(W) data is not required as long as FSC(W) Valve Unit
data is within the specified limit.

6B

XS Series S/M 4-2 December 13,2007


(2) Connect a syringe and push CELLPACK in the flow cell to waste (3) Carefully remove thermistor No.48 upward so that a strong
chamber. stress is not applied onto the flow cell.

XS Series S/M 4-3 December 13,2007


4.1.4
7B Removing Photo Detector No.30 4.1.5 Removing XS_FCM Detector Block Cover B

(1) Remove Photo Detector No.30 by loosening fixing screws (1) Remove XS_FCM Detector Block Cover B by loosening fixing
(Screw Binding M3X6). Cut a tie wrap tying wiring cord. screws (Screw Binding M3X6).
Disconnect the connector from the PCB No.2168.

Fixing Screw
Fixing Screw Photo Detector No.30

PCB No.2168

Disconnect
8B

XS Series S/M 4-4 December 13,2007


(2) Lift and move FCM Detector Block toward you by loosening 4.1.6 Removing Flow Cell Unit
fixing screws.
(1) Remove fixing screw (Screw Hex-Socket Bolt M3X8) for flow cell
unit and carefully pull the flow cell unit downward so that the
flow cell is not damaged.

Fixing Screws

9B

XS Series S/M 4-5 December 13,2007


(2) Disconnect tubing from SV 34 and Reaction Chamber.

Disconnect

XS Series S/M 4-6 December 13,2007


4.1.7
10B Adjusting Laser Power (7) When below screen appears, double-click Service.

(1) Turn on IPU computer.

(2) Enter “sysmex” in the Logon Name.

(3) Enter “c”, “9”, “.” and “0” in the password.

(4) Turn on XS main unit.

(5) When below screen appears, select “Ignore all Errors except
Sampler Errors” and click OK.

(6) Select MENU – Controller.

XS Series S/M 4-7 December 13,2007


(8) When below screen appears, double-click Laser Power to (9) Set the wave length of the laser power meter to 633 nm and
activate laser. (Wear protection goggle when the LD is place the probe in front of the laser at a little inclined angle as
activated.) shown to prevent the reflect beam from going back to the laser
diode. (Laser power controlling device in the laser diode may
malfunction when a beam is incoming into the laser diode.)

Close

XS Series S/M 4-8 December 13,2007


(10) Adjust VR1 so that the laser power is 3.4 +/- 0.1 mW with a 4.1.8 Adjusting Laser Diode Angle (when replacing Laser
ceramic screw driver. (A ceramic screw driver is recommended Diode)
to prevent occurring static accident.)
(1) Remove Flow Cell by Step 4.1.3 Replacing Flow Cell.

(2) Place XS Laser Diode Angle Adjustment Tool or equivalent


(hand mirror). (Place the tool or a hand mirror by pressing to
arrow directions to reflect the laser beam to the laser diode.)

1B
VR1
(3) Place a hex wrench between flow cell and laser diode, and
adjust the laser diode angle so that the reflected light and laser
light are one line.

(4) Click OK to turn off the laser.


12B

XS Series S/M 4-9 December 13,2007


4.1.9 Installing Flow Cell Unit No.11 4.1.10 Adjusting Flow Cell Unit No.11 Angle

(1) Tighten temporary the Flow Cell Unit No.11 with Screw (1) Double-click Laser Power on the service screen to activate
Hex-Socket Bolt M3X8. laser.

(2) Place a hex wrench between flow cell and laser diode, and
adjust the flow cell angle so that the laser beam and reflected
beam from the flow cell on the wrench are one line.

(3) Tighten fixing screws so that the flow cell angle doe not move.

(4) Confirm that the laser beam and reflected beam on the wrench
are one line after tightening the fixing screws.
13B

(5) Click OK on the Laser Power screen to turn off the laser.
14B

XS Series S/M 4-10 December 13,2007


4.1.11 Installing XS_FCM Detector Block Cover B 4.1.12 Confirmation/Adjustment of Laser Diode Initial Position

(1) Tighten fixing screws so that tubing of the bottom of the flow cell (1) Confirm/Adjust the Laser Diode initial position (forward/backward)
is parallel to the bottom surface of the FCM Detector Block as follows.
Cover B.
Loosen these
screws when
moving cylinder
F/B

F/B Adjustment
Screw

15B

Chamfered corner of Align block surface


Laser Diode
and corner of laser
Aligned diode.

Block
Surface

XS Series S/M 4-11 December 13,2007


(2) Confirm/Adjust the Laser Diode initial position (right/left) as (3) Do not loosen following fixing screws.
follows.

Laser Diode
R/L Fixing
Screws

R/L Adjustment Screw

Aligned
Do Not Loosen Fixing Screws

Align these lines.

XS Series S/M 4-12 December 13,2007


4.1.13 Adjusting Laser Diode Position
16B (3) When step (2) is unclear to observe the two lines of the flow cell,
adjust the Laser Diode R/L position by observing laser beam in
(1) Double-click Laser Power on the service screen to activate flow cell as shown below. When laser beam hits the sample
laser. path in the flow cell, reflected beam can be seen.

(2) Place a white paper as shown and adjust Laser Diode R/L
position so that two lines of flow cell are both sides of the beam
by turning adjustment screw.

Laser beam not in center Laser beam in center

XS Series S/M 4-13 December 13,2007


4.1.14 Installing Photo Detector No.30
17B 4.1.15 Adjusting Beam Stopper Position

(1) Click OK on the Laser Power screen to turn off the laser. (1) Double-click Laser Power on the service screen to activate
laser.
(2) Fix the Photo Detector Block No.30 so that the distance
between flow cell and Photo Detector Bloc No.30 is 7 mm by (2) Adjust beam stopper position so that the beam image is on the
tightening screw binding M3X6. center of the beam stopper as shown.

(3) Click OK on the Laser Power screen to turn off the laser.

18B

XS Series S/M 4-14 December 13,2007


4.1.16 Confirmation/Adjustment of Photo Detector No. 28 (1)
19B 4.1.17 Confirmation/Adjustment of Photo Detector No. 28 (2)
20B

(1) Confirm/Adjust the adjustment plates position as follows. (Loose (1) Confirm/Adjust the Photo Detector No.28 initial position as
fixing screw when moving adjustment plates.) follows. (Loosen fixing screws (Screw Hex-Socket Bolt M3X6)
when moving Photo Detector No.28.)

Fixing Screws

XS Series S/M 4-15 December 13,2007


May 23, 2008

4.1.18 Connecting Oscilloscope 4.1.19 Optical Alignment (Rough)

(1) Remove rear panel. (1) Select Menu Æ Controller ( Maintenance ( Auto Rinse.

(2) Connect probes of the oscilloscope as shown. (Both FSC and SFL (2) Confirm that no background error is generated after performing
should be connected because SFL signal is triggered by FSC Auto Rinse.
signal.)
(3) Click Close to close the screen.

(4) Prepare latex material for optical alignment as follows.


FSC Latex: CELLPACK 30mL and SS-071-P two drops
FSC Latex: CELLPACK 30mL and 4207A six drops
SFL Latex: CELLPACK 50mL and A7312 two drops

Warning
Since undiluted latex is easy to coagulate, shake it enough and add
FSC(1) GND SCC SFL to CELLPACK.
Use the latex materials within 3 hours after making it. Do not add
GND latex to made latex materials after a time. (Use the latex materials
within an hour after making for confirmation of Optical Alignment
(fine))
Since the fluorescence intensity becomes weak when A-7312 (latex
for fluorescence) receives light, store A-7312 in dark place. Use
diluted latex in a bottle.
Store latex in room temperature and do not freeze it.

(5) Double-click Optical Axis Adjustment (rough) in service.

(3) Set the oscilloscope as follows.


Gain: 10mV
Time: 10 µ sec.

XS Series S/M 4-16


(6) Pour 1 mL of FSC latex to reaction chamber and press start
switch.

(7) Adjust laser diode position so that the FSC wave on the
oscilloscope is highest and narrowest by turning adjustment screw
A and B. (9) Adjust FSC(W) approximately 0.3 or lower. If the value does not
become 0.3, repeat the steps (6) to (8).
(8) Adjust the beam stopper R/L position so that FSC wave is lowest
by turning adjustment screw C.

XS Series S/M 4-17 December 13,2007


(10) Pour 1 mL of SFL latex to reaction chamber and press start
switch.

(11) Adjust Photo Detector position so that the SFL wave on the
oscilloscope is highest by turning adjustment screw D.

(12) Adjust Photo Detector Position so that the SFL wave on the
oscilloscope is highest and minimum variation by turning
adjustment screw E and F.

(13) Adjust the Photo Diode Position so that the SFL(W) is


approximately 0.3 ore lower by turning adjustment screw D, E
and F.
(14) Click Close in the screen.

F E

XS Series S/M 4-18 December 13,2007


4.1.20 Optical Alignment (fine)
23B

(9) If the value is out of the range, repeat steps (5) – (8).
(1) Double-click Optical Axis Adjustment (fine) in service.
X W
(2) Pour 1 mL of FSC latex to reaction chamber and press start FSC 60 ± 30 ≤ 0.35
switch. SFL 130 ± 50 ≤ 0.30
SSC 50 ± 40 ≤ 0.35
(3) Enter appropriate value into Gain and click Send so that the
FSC(X) value is within the below range.
FSC(X): 60 +/- 30ch

(4) Run FSC latex three times after adjusting Gain and confirm that
the FSC(W) is within the below range.
FSC(W)≤0.35

(5) Pour 1 mL of SFL latex to reaction chamber and press start


switch.

(6) Enter appropriate value into Gain and click Send so that the
SFL(X) value is within the below range. Leaf spring for Laser Diode Laser Diode base
SFL(X): 130+/-50ch

(7) Run SFL latex three times after adjusting Gain and confirm that
the SFL(W) is within the below range.
SFL(W)≤0.30

(8) Pour 1 mL of SFL latex to reaction chamber and press start


switch. Tighten following fixing screws while checking SFL
related values remain unchanged.1) Fixing screws for laser
diode leaf spring: four screws
2) Fixing screws for Laser diode base: two screws
3) Fixing screws for Photo Diode No.28: four screws
Photo Diode No.28

XS Series S/M 4-19 December 13,2007


(10) Shine the flow cell with a flashlight perpendicularly to flow cell
and observe the image of the Photo Detector No.29 on the
mirror of the Photo Detector No.28. Adjust the Photo Detector
No.29 position so that the slit is in the center of the flow cell
shadows.

Photo Detector No.28 Photo Detector No.29

Mirror

Shine with a flashlight Place the slit in center Fix the Photo Detector No.29
of the flow cell shadows while pressing to the base

An Example of FSC Alignment

XS Series S/M 4-20 December 13,2007


(11) Change the probe of the oscilloscope from FSC to SSC.

(12) Pour 1 mL of SFL latex to reaction chamber and press start


switch. Adjust Photo Detector No.29 position so that the SSC
wave is highest and minimum variation by turning adjustment
screw D.

(13) Pour 1 mL of SFL latex to reaction chamber and press start


switch. Confirm that below values are within the below range.
SSC(X):50+/- 40 ch
SSC(W):≤0.35

(14) After completing adjustment, fix two fixing screws.

An Example of SFL/SSC Alignment

XS Series S/M 4-21 December 13,2007


(15) Pour 1 mL of FSC latex to reaction chamber and press start (17) Apply lock tight following 20 fixing screws.
switch. Adjust beam stopper position so that the FSC wave is
lowest by turning adjustment screws C.

(16) After completing adjustment, tighten four fixing screws.

XS Series S/M 4-22 December 13,2007


4.2
1B Mechanical Adjustment 4.2.2 Adjusting Rinse Cup Position

4.2.1
24B Piercer Position Adjustment (1) Click Init. Position and adjust the piercer height so that the
piercer tip and Rinse Cup bottom are same height by clicking Up
(1) Replace adapter with XS Adapter (13 mm diameter) (P/N: or Down key.
442-3086-2.

Piercer tip and Rinse Cup bottom


are same height.

(2) Select Menu Æ Controller Æ Service ( Pierce/Pipette adjustment.

(3) Select Normal for Sample Tube Type.


25B

XS Series S/M 4-23 December 13,2007


4.2.3
26B Adjusting Aspiration Position 4.2.4 Adjusting RBC Chamber

(1) Click Aspiration Position and adjust the piercer position so that (1) Click RBC/HGB Chamber and adjust the piercer position so that
the piercer tip is as shown position by clicking Up, Down, Front the piercer tip is in the center of the chamber by clicking Up, Down,
and Back key. Front and Back key.

Same
height
with the
top
surface

Center
Adjust the center at Aspiration Position Same height with the top surface

Center
27B

XS Series S/M 4-24 December 13,2007


4.2.5
28B Adjusting Reaction Chamber

(1) Click WBC Chamber and adjust the piercer position so that the
piercer tip is in the center of the chamber by clicking Up, Down,
Front and Back key.

Center

Same height with the top surface (2) After adjusting, click Save and Close.

XS Series S/M 4-25 December 13,2007


4.2.6
29B Sampler Adjustment (2) Adjust Sample Set Unit Cover position so that the cover and
chassis are flat by clicking Back or Front.
(1) Select Controller Æ Service Æ Tube Holder Adjustment and click
Set Position.

Make them flat

XS Series S/M 4-26 December 13,2007


(3) Click ID Reader Position. (5) Pull Sample Set Unit manually and set the test tube with the
barcode facing to the barcode reader. Select Tub Holder
Adjustment and click ID Reader Position. Find the BR Unit No. 1
position so that the barcode reader can read barcode at front most
position and back most position. Move the BR Unit No. 1 position
so that the BR Unit No. 1 is placed the center of the previous two
positions.
Green: Reading barcode
Orange: Not readging barcode

(4) When movement completes, click Start in the Tube Rotation. Adust
the Sample Set Unit position so that the test tube rotates by
clicking Back and Front key. Adust the Sample Set Unit position so
that the test tube rotates discontinuously by clicking Front key.
Move Sample Set Unit backward – 15 pulses by clickin Back key.
Confirm that the test tube rotates continously. Click Save for
saving the position. Click Stop in the Tube Rotation and Close key.

BR Unit No. 1

(6) Tighten left fixing screw of the BR Unit No. 1. Click Close to close
the window. Tighten right fixing screw of the BR Unit No.1

XS Series S/M 4-27 December 13,2007


(9) Click Set Position Æ Catch. Place a test tube in the hand and click
(7) Set two racks to the sampler, select Controller Æ Service Æ Hand Down. Adjust the test tube position so that the test tube is center of
Adjustment and click Set Position and Down. the Sample Set Unit by clicking Back, Front, Left or Right.

(8) Adjust Hand height position so that the gap between rack and hand
is 1 to 2 mm by adjusting rod height position.

(10) After completing adjustment, click Save Æ Up and remove the test
tube from the Hand.

XS Series S/M 4-28 December 13,2007


(11) Click Rack-01 Æ Catch. Place a test tube in the hand and click (13) Click Rack-02 Æ Catch. Place a test tube in the hand and click
Down. Adjust the test tube position so that the test tube is center of Down. Adjust the test tube position so that the test tube is center of
the Rack-01 by clicking Back, Front, Left or Right. the Rack-01 by clicking Back, Front, Left or Right.

RACK 01
RACK-02

(12) After completing adjustment, click Save Æ Up and remove the test (14) After completing adjustment, click Save Æ Up and remove the test
tube from the Hand. tube from the Hand.

XS Series S/M 4-29 December 13,2007


4.3
2B Hydraulic Adjustment 4.3.2 Adjusting HGB Blank

4.3.1
30B Adjusting RBC Clog Level Note: Digital VR values inversely adjust blank values, i.e., increasing
digital VR value, decreases HGB blank value.
(1) Select Menu Æ Controller Æ Service Æ RBC Clogs Level.
(1) Confirm that 30 minutes past after turning on the main unit.
(2) When “RBC Clog Error” occurs, click Close to close error
screen. (2) Select Menu Æ Controller Æ Service Æ Sensitivity.

(3) RBC Clog Level appears in real time. (3) Enter appropriate values into HGB blank and ASP. Sensor blank,
and click Send. Repeat this step until value comes within
(4) Enter “99” into Target box and click “>>>” ( “Send” (”Continue”. Adjustment Value range as shown in table below.

(5) Repeat step (4) until RBC Clogs is 99 +/- 1. Click Close when RBC
Clogs becomes 99 +/- 1.

Temperature Adjustment Value


Up to 19 °C 2500 +/- 300
20.0 – 24.9 °C 3000 +/- 300
25.0 °C or 3500 +/- 300
higher
31B

(4) Click Close after values come within Adjustment Value range as
shown in table above.

XS Series S/M 4-30 December 13,2007


(2) Select Aspiration Sensor.
4.3.3
32B Adjusting Aspiration Sensor
(3) When Aspiration Sensor appears, run e-CHECK (level 2).
Note: Before performing this adjustment make sure the HGB blank
and HGB Asp. sensor blank values are in range (step 4.3.2). (4) After analysis, enter “HGB assay value for e-CHECK Level 2 X
105” into Target box and click “>>>”Æ “Send” and press start
(1) Select Menu Æ Controller Æ Service Æ Mode. Select Control 3 in switch.
Analytical Mode and Off in Continuous, and click OK.
(5) Confirm that the value is HGB assay value for e-CHECK Level 2
X 105 +/- 50 when analysis is over.

(6) Repeat until sample – blank value is in the range.

(7) Click Close when value becomes HGB assay value for e-CHECK
Level 2 X 105 +/- 50.

CTRL3 appears

XS Series S/M 4-31 December 13,2007


June 5, 2008

4.3.4 Adjusting Sensitivity (7) When the analysis is over select Sensitivity and confirm that
values are within the following range.
(1) Confirm that Aspiration Sensor adjustment is completed. (Select
Menu Æ Controller Æ Service Æ Mode. Select Control 3 in
Analytical Mode.) Target range
DIFF-X: assay value +/- 3ch
(2) Run e-CHECK (level 2). DIFF-Y: assay value +/- 2ch
Note: For American Markets, the X-CAL product is used rather FSC-X: assay value +/- 2ch
than e-CHECK level 2 when performing Sensitivity MCV: assay value +/- 1 fL
adjustments. MPV: assay value +/- 1fL
HGB: assay value +/- 0.2 g/dL
(3) Select Sensitivity in Service screen. (e-CHECK cannot be run in
the Sensitivity screen.)

(4) Enter each assay value in the Target box and click >>> and Make sure that the DIFF scattergram and FSC histogram are
Send. discrete and displayed normally.
Enter assay value Click >>>
(8) Click Close when the values are within the range.

(9) Repeat step (6) and (7) until the values are within the range.

Click Send For adjusting DIFF-Y using human blood, aspirate ten normal blood
samples and Calculate mean value. Enter target value in DIFF-Y and
click ‘Send’ so that the result value will be in the following target value
range.
Target Value: 43.5 +/- 3.0

Pick up the closest sample to the mean value.

(5) Click Close.

(6) Run e-CHECK (level 2) again.

XS Series S/M 4-32


4.4
3B Electric Adjustment 4.4.2 PCB No.2168

4.4.1
34B PCB No.6365T2 VR1: VR for LD power adjustment Increases LD power by turning to
CW.
(1) Dip Switch VR2: VR for high frequency overlap adjustment Increases high
frequency overlap.
bit ON OFF Default
8 Not used Not used OFF TP1: Ground for LD drive
7 Not used Not used OFF
TP2: Power output for LD Shows proportional voltage for LD power
6 Not used Not used OFF
5 Not used Not used OFF between TP2 and TP1. Proportional factor varies to laser
4 Initialize setting data in Not Initialize setting data in OFF diodes.
EEPROM at power ON EEPROM at power ON TP3: LD current monitor Shows following voltage.
3 Not used Not used OFF Voltage between TP3 and TP1=25 X LD current (example:
2 Not used Not used OFF
1 Initialize counters in Not Initialize counter in EEPROM OFF
2.5 V when LD current at 100mA)
EEPROM at power ON at power ON

35B

XS Series S/M 4-33 December 13,2007