Professional Documents
Culture Documents
AAPM TG198 ImplementationOfTG142-Compressed
AAPM TG198 ImplementationOfTG142-Compressed
AAPM Task Group 198 Report: An Implementation Guide for TG 142 Quality Assurance of Medical
Accelerators
Joseph Hanley, PhD1, Sean Dresser, MS2, William Simon, MS3, Ryan Flynn, PhD4, Eric E. Klein,
PhD5, Daniel Letourneau, PhD6, Chihray Liu, PhD7, Fang-Fang Yin, PhD8, Bijan Arjomandy, PhD9,
Lijun Ma, PhD10, Francisco Aguirre, MS11, Jimmy Jones, MS12, and John Bayouth, PhD13
1Princeton Radiation Oncology
2Emory University
3Sun Nuclear Corp
4University of Iowa
5Rhode Island Hospital/Warren Alpert Medical
6Princess Margaret Cancer Centre
7University of Florida
8Duke University
9Karmanos Cancer Institute at McLaren-Flint McLaren
10University of California, San Franciscosample
11MD Anderson Cancer Center
12The University of Colorado Health—Poudre Valley
13University of Wisconsin—Madison
Consultants: Todd Holmes (Varian Medical Systems), Carlos Sandin (Elekta)
TG 198 was constituted by the AAPM, Science Council, Therapy Physics Committee, and Quality
Assurance and Outcome Improvement Subcommittee.
This article has been accepted for publication and undergone full peer review but has not been
through the copyediting, typesetting, pagination and proofreading process, which may lead to
differences between this version and the Version of Record. Please cite this article as doi:
10.1002/MP.14992
This article is protected by copyright. All rights reserved
Corresponding Author: Joseph Hanley, Princeton Radiation Oncology, 9 Centre Drive, Monroe, NJ
Accepted Article
08831, USA; phone: 609-655-5755; email: joseph.hanley@asterahealthcare.org
Disclosure Statement
The Chair of the AAPM TG 198 has reviewed the required conflict of interest statement on file for
each member of AAPM TG 198 and determined that disclosure of potential conflicts of interest is an
adequate management plan. Disclosures of potential conflicts of interest for each member of AAPM
TG 198 are found at the close of this document.
Contents
1 Introduction
1.1 Purpose
1.2 Background
2 QA of Medical Accelerators
2.1 General
2.1.1 Clarifications and Deviations From TG 142
2.1.1.1 Test Frequency
2.1.1.2 Daily or Weekly QA
2.1.1.3 Monthly QA
2.1.1.4 Annual QA
2.1.2 Documentation
2.2 Daily
2.2.1 Dosimetry
2.2.1.1 Photon and Electron Output Constancy
2.2.2 Mechanical
2.2.2.1 Laser Localization
2.2.2.2 Optical Distance Indicator (ODI)
2.2.2.3 Collimator Size Indicator
2.2.3 Safety
The recommendations of this TG are not intended to be used as regulations. These recommendations are
guidelines for qualified medical physicists (QMPs) to use and appropriately interpret for their individual
institution and clinical setting. Each institution may have site-specific needs and requirements which may
modify their usage of these recommendations.
1.2 Background
The underlying principle behind TG 40 and TG 142 was the International Commission on Radiation
Units and Measurements’ (ICRU)11 recommendation that the dose delivered to the patient be within
±5% of the prescribed dose.12 Taking into consideration the many steps involved in delivering dose to
a target volume in a patient, each step must be performed with accuracy significantly better than 5%
to achieve this recommendation.
The report of the TG 142 considerably expanded the tests of TG 40. In addition, the report
recommended tolerances that accommodated differences in the intended use of the machine
functionality: non-IMRT, IMRT, and stereotactic delivery. It should be noted that this TG replaced all
references to “x-ray” with “photon” and updated the tolerances for a few of the tests recommended in
TG 142. Furthermore, in the gating sections, all references to amplitude have been replaced with
displacement or displacement magnitude to align with TG 76,13 the management of respiratory motion
in radiation oncology. Volumetric modulated arc therapy (VMAT) QA has also been included in this
report.
(1) Ideally, the solid phantom and ionization chamber should be stored in the treatment room with the
linac at all times to stay in a state of equilibrium with the ambient temperature in the room. If this
is not possible, the phantom and chamber should be placed in the treatment room for a sufficient
period to allow the equipment to reach room temperature. The temperature of the chamber cavity
in the phantom and the barometric pressure should be monitored periodically throughout the
measurements, and appropriate changes should be made to correct the temperature or pressure
factor if necessary. If the electrometer requires a long warmup period—that should be confirmed
from the user manual but is typically 10 to 15 min37—it should be turned on with the appropriate
voltage before any measurements.
(2) The phantom should be placed on the treatment couch or other supportive device at conditions
like those at the time of annual calibration (see Figure 2) per TG 5138 calibration (SSD, SAD,
gantry settings, etc.). If cross-calibrations are performed from water to a solid phantom,
measurement depths should be close to those depths used for TG 51 calibration.
(3) For statistical purposes, three measurements should be obtained for each energy to allow for any
drift that could occur with the linac or electrometer or ionization chamber. Variation between each
reading should not exceed ±0.5%.39
(4) Perform all appropriate corrections to the measurements following the protocol used for the initial
calibration (e.g., TG 51) or alternately using the previously established cross-calibration.
Calculate the result in cGy/MU at the depth indicated by the protocol (e.g., dmax, d = 10 cm, etc.).
(5) Adjustments should be made if the measured output exceeds ±2%. At the discretion of the facility,
adjustments could be made for smaller deviations from expected values. Repeated measurements
should be made after any adjustments to ensure the correct output adjustment was made.
(6) If adjustments are made to the outputs, any devices used for daily QA should be checked. Any
deviation from the new monthly readings should be investigated to determine the cause, and
2.4.3 Safety
2.4.3.1 Follow Manufacturer’s Test Procedures
2.4.3.1.1 Objective
The objective is to follow the manufacturer’s recommended safety procedures.
2.4.3.1.2 Tolerance
The tolerance is as per manufacturer’s tolerances.
2.7.4.4 Contrast
2.7.4.4.1 Objective
If kV imager is not functioning properly or if an inappropriate filter is applied to the acquired images,
the images may have poor contrast, which can compromise patient setup effectiveness due especially
to inability to differentiate between tissues.
2.7.4.4.2 Tolerance
Contrast should be better than or equal to baseline measured during acceptance.
2.7.4.4.3 Methods and Equipment
A QA device such as the Leeds phantom can be imaged to verify that the kV imaging system’s
contrast resolution has not fallen below baseline (see Figure 11). The same image acquired in section
2.7.4.3.3 can be used for this test. Inspect the images visually to determine the lowest contrast disk
visible in the images. This value is compared with baseline. It should be noted that there are several
commercially available software packages that will perform the analysis of these imaging tests
automatically.
2.7.4.5 Uniformity and Noise
2.7.4.5.1 Objective
The objective is to ensure that the performance of the kV imager with respect to uniformity and noise
has not degraded with time.
2.7.4.5.2 Tolerance
Uniformity and noise should be better than or equal to baseline measured during acceptance.
2.7.4.5.3 Methods and Equipment
Disclosure Statement
1. The members of AAPM TG 198 listed below attest that they have no potential conflicts of interest
related to the subject matter or materials presented in this document: Joseph Hanley, Sean Dresser,
William Simon, Ryan Flynn, Eric Klein, Daniel Létourneau, Chihray Liu, Fang-Fang Yin, Bijan
Arjomandy, Lijun Ma, Francisco Aguirre, Jimmy Jones.
2. The members of AAPM TG 198 listed below disclose the following potential conflict(s) of interest
related to subject matter or materials presented in this document: John Bayouth has ownership interest
in MR Guidance, LLC, which provides commissioning consulting services for MRI guided radiation
therapy. Todd Holmes is an employee and shareholder of Varian Medical Systems. Carlos Sandin is
an employee of Elekta LTD.
REFERENCES
1. Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: quality assurance of medical
accelerators. Med Phys. 2009;36:4197.
2. Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for radiation oncology: report of
AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21:581–618.
3. Fought AM, Trager M, Yin FF, Kirkpatrick J, Adamson J. Re-examining TG-142
recommendations in light of modern techniques for linear accelerator based radiosurgery. Med Phys.
2016;43:5437.
4. Al-Hallaq HA, Chmura S, Salama JK, et al. Rationale of technical requirements for NRG-
FIGURE LEGENDS
Figure 1. Beam-on indicator and x-ray-in-use lights outside a treatment vault.
Figure 2. Solid water setup for monthly output measurements.
Figure 3. Winston-Lutz setup with cone and film holder.
Figure 4. LaserGuard collision test setup. Object is set up on couch such that it will enter the
protection zone of the LaserGuard system. Once in protection zone, all motion should be inhibited.
Figure 5. Reciprocating phantom with marker block for respiratory gating tests for the Varian RPM
system.
Figure 6. Stereotactic motion lockouts on a Varian linac.
Figure 7. Planar kV images of a cube phantom used for positioning or repositioning tests. A central
target BB, along with 2 setup BBs are visible. The blue crosshairs are the digital graticule.
Figure 8. Planar MV image of an image quality phantom for monthly QA. The bar patterns or line
pairs along the central axis can be used to determine spatial resolution.
Figure 9. Las Vegas phantom for MV planar image QA. The number of visible holes in the image
should be consistent with the number visible during acceptance testing.
Dosimetry
Photon & Electron ±3% Ionization chamber, 10-20 min RTT
Output Constancy Diode/MOSFET system
Mechanical
Laser Location ±2mm/±1.5mm/±1mm Front pointer, reference marks 2 min RTT
Distance Indicator ±2mm Front pointer, jig 2 min RTT
Collimator Size Indicator ±2mm/±2mm/±1mm Graph paper, ruler, jig 3 min RTT
Safety
Door Interlock Functional NA 1 min RTT
Door Closing Safety Functional NA 1 min RTT
Audiovisual Monitors Functional NA 1 min RTT
Stereotactic interlocks NA/NA/ Functional NA 1 min RTT
Radiation Area Monitors Functional NA 1 min RTT
Beam-on Indicator Functional NA 1 min RTT
NA = not applicable.
C= central axis, T = target point, G = gun point, L = left point, R = right point.
Mechanical
Light / Radiation Field ±2.0 mm per Jaw Film or EPID 30 min. (Film) QMP or
Coincidence (Symmetric & 15 min. (EPID) Designee
Asymmetric)
±2mm/ ±1 mm/ ≤ 1mm Front pointer/Winston-Lutz test equipment 5 min. QMP or
Lasers
deviation Designee
Gantry / Collimator Angle ±1° Level 5 min. QMP or
Indicators (@ cardinal angles, Designee
digital only)
Accessory Trays (i.e., port film ±2 mm NA 1 min. QMP or
graticule tray) Designee
Jaw Position Indicators ±2 mm per Jaw Graph paper, template 15 min. QMP or
(Symmetric) Designee
±2.0mm or ±1 mm per Graph paper, template 15 min. (can be QMP or
Jaw Position Indicators jaw when used for done simultaneously Designee
(Asymmetric) field matching at the with Symmetric)
central axis
Cross-Hair Centering ≤1 mm Graph paper, template 15 min. QMP or
(Walkout) Designee
Treatment Couch Position ±2 mm, ±1°/ ±2 mm, Graph paper, ruler 15 min. QMP or
Safety
Laser guard-interlock test Functional NA 1 min. QMP or
Designee
Respiratory Gating
Beam output constancy ± 2% ADCL Calibrated Ionization 15 min. (using same QMP or
Chamber/Electrometer or setup as output Designee
Chamber/Electrometer cross-calibrated with constancy)
ADCL Chamber/Electrometer, solid phantom
or water phantom
Phase, displacement Functional NA 1 min. QMP or
magnitude beam control Designee
In-room respiratory monitoring Functional NA 1 min QMP or
system Designee
Gating interlock Functional NA 1-2 min. QMP or
Designee
VMAT
Patient Specific VMAT QA Institutional VMAT QA VMAT QA phantom 20-30 min QMP or
passing criteria based Designee
on TG-218 guidelines
NA = not applicable.
Mechanical
Collimator Rotation Isocenter ± 1 mm from baseline Front pointer set. 10-15 min. QMP
Gantry Rotation Isocenter ± 1 mm from baseline Front pointer set. 15-30 min. QMP
Couch Rotation Isocenter ± 1 mm from baseline Front pointer set. 10-15 min. QMP
Electron Applicator Interlocks Functional NA 10-15 min. QMP
Coincidence of Radiation and ±2/ ±2/ ±1 mm Front pointer set. Film/ GafChromic 30-60 min. QMP
Mechanical Isocenter film/EPID.
Couch Top Sag ± 2 mm from baseline Front pointer set. Lasers and ruler. 10-15 min QMP
Couch Angle ±1°/ ±1°/ ±0.5° Graph paper. 10-15 min QMP
Couch Travel Maximum Range Graph paper. Ruler. 10-15 min QMP
± 2 mm
Movement in All Directions
Stereotactic Accessories, Lockouts, NA/ NA/ Functional NA 10-15 min QMP
etc.
Safety
Follow Manufacturer's Test Functional NA 30-45 min. QMP
Procedures
Respiratory Gating
Beam Energy Constancy ± 2% of non-gated Ionization Chamber/Electrometer, 15-30 min QMP
beam solid phantom or water phantom
Temporal accuracy of ±100 ms of expected Motion phantom. Film. GafChromic 60-120 min QMP
phase/displacement magnitude gate film. EPID
on
Calibration of surrogate for ±100 ms of expected Motion phantom. 60-120 min QMP
phase/displacement magnitude
Interlock testing Functional NA 5-10 min QMP
VMAT
Test Patient Specific VMAT QA Plan passes QA at a VMAT QA phantom 20-30 min QMP
constancy pass rate ≥ baseline
pass rate minus 2%
AND Plan passes
institutional VMAT QA
criteria
Interruption test ± 1% deviation from VMAT QA phantom 10-15 min QMP
uninterrupted passing
NA = not applicable
Backup diaphragm settings ±2mm Film or EPID and radio- Film: 30-35 min QMP or
(Elekta only) opaque markers if the light EPID: 15-20 min Designee
field crosshair is used as a
surrogate for the radiation
isocenter.
Leaf Travel speed (IMRT) leaf speed within Log-file analysis software or Log file method: QMP or
±0.5 cm/s EPID and corresponding 10-12 min Designee
analysis software EPID method:
15-20 min
Leaf position accuracy (IMRT) ±1mm Films or EPID and software Film: 25-40min QMP or
analysis application EPID: 15-25 min Designee
Graph paper, template Light Field: 10-15
min
Annual
MLC transmission (average of ±0.5% from Ionization chamber and Obj. 1: 20-30 QMP
leaf and interleaf baseline water equivalent scattering min for one
transmission), all energies material. energy. 10 min
per additional
energies
Leaf position repeatability ±1mm EPID or graph paper for the 30-40 min with QMP
light field-based approach EPID. 15 min
with the light-field
approach
Coincidence of light field and ±2mm per MLC Film or EPID and software Film: 25-25 min QMP
Photon field (all energies) bank analysis application EPID: 10-15 min
(for one energy)
Moving window IMRT RMS of leaf Delivery log file analysis 10-20 min QMP
positioning error <
software tool
0.35 cm for all leaves
NA = not applicable.
system
Imaging dose ± 20% of the baseline Radiographic dosimetry 15-30 min. QMP
dose system
Cone-beam CT (kV & MV)
Imaging dose Baseline ± 20% for kV- Radiographic dosimetry 15-30 min. QMP
mp_14992_f1.jpg
mp_14992_f2.jpg
mp_14992_f3.jpg
mp_14992_f4.jpg
mp_14992_f5.jpg
mp_14992_f6.jpg
mp_14992_f7.jpg
mp_14992_f8.jpg
mp_14992_f9.jpg
mp_14992_f10.jpg
mp_14992_f11.jpg
mp_14992_f12.jpg
mp_14992_f13.jpg
mp_14992_f14.jpg
mp_14992_f16.jpg
mp_14992_f17.jpg
mp_14992_f18.jpg