Vision 2186

How to calibrate the


Oxygen Cells of Vision 2186
Infant Incubator
Vision 2186

Oxygen Cells
Calibration
Oxygen Cells
Oxygen Cells Compartment
REF: 090.158.600

Oxygen Cells
REF: 090.855.020
Oxygen Cells Calibration

Press “Menu” and “Inhibit Sound” keys at the same time;


Oxygen Cells Calibration

Wait until the screen changes and press “Menu” key one more time;
Oxygen Cells Calibration

Press “Home” key to go back to the main screen;


Oxygen Cells Calibration

At the main screen, press “Oxygen” key to go into Oxygen screen;


Oxygen Cells Calibration

Press “Cal” key to go to the Calibration screen;


Oxygen Cells Calibration

Since the system is operating in Calibration Mode, the calibration


screen will be different than the calibration screen at User Mode;
Oxygen Cells Calibration

Open the O2 compartment and wait 2 minutes


until the reading is stabilized by the Oxygen Cells;
Select the “21%” key;
Oxygen Cells Calibration

The values in sensors A and B will be updated to


21% (±1% tolerance) ;
Oxygen Cells Calibration

Apply 100% of O2 to the Oxygen Cells and wait 1 minute until the
reading is stabilized;
The 100% of O2 can be applied using a plastic bag over the cells and
connected to an O2 outlet, with no leakage;
Oxygen Cells Calibration

The values in sensors A and B will be updated to 100% (±2% tolerance) ;


Calibration is done.