Professional Documents
Culture Documents
System Owner Manual - Magnetom Aera Sapedm Som Aera Ve11e m7-01002g.629.16.02.02
System Owner Manual - Magnetom Aera Sapedm Som Aera Ve11e m7-01002g.629.16.02.02
MAGNETOM Aera
Manufacturer’s note:
This device bears a CE mark in accordance with the provisions of Council
Directive 93/42/EEC of June 14, 1993 concerning medical devices and
the Council Directive 2011/65/EU of June 08, 2011 on the restriction of
the use of certain hazardous substances in electrical and electronic
equipment.
The CE marking applies only to medical devices which have been put on
the market according to the above-mentioned EC Directives.
Unauthorized changes to this product are not covered by the CE mark
and the related Declaration of Conformity.
Contents
Introduction 1
Safety Information 2
Technical data 5
Location of labels 6
Maintenance Plan 7
Disposal 8
Register A - J 11
Dear customer,
On the occasion of inspections and controls by the authorities you must always have many different
documents at hand. The present System Owner Manual is intended to make archiving these documents
simpler for you so that you always have them completely at hand if necessary.
We have already filed the most important documents in the System Owner Manual on delivery of your
system.
The further updating of the Manual now lies in your hands.
We therefore recommend that you designate an employee right at the start-up of your new system who
will be responsible for continuously updating the System Owner Manual and who immediately has it at
hand as required.
Upon request, Siemens Service will provide technical documents for the MR system (e.g. circuit dia-
grams, spare parts lists, descriptions, calibration instructions, component part lists). However, this does
not constitute authorization for repairs.
! All labels displayed in the System Owner Manuals are examples only and may differ from the labels
attached to the system and the components.
Siemens Healthcare
representative
System Owner
Manufacturer
Siemens
■
■
■
P.S.: The table on the following page shows which documents are filed in the System Owner Manual
and who contributes the documents.
Introduction – Page 1 of 4
Who supplies what? The documents in the System Owner Manual.
representative
System Owner
Manufacturer
Siemens
Document
1 Introduction
■ Notes on handling the System Owner Manual
■ Operator and location of the product
2 Safety information
■ Important safety aspects for the owner of a MR system
5 Technical data
■ Technical data
■ Upgrades and additions
6 Location of labels
■ Location of labels
7 Preventive maintenance
■ Maintenance plan
■ Maintenance contract
■ Maintenance certificates
8 Disposal
■ Instructions for disposal of problematic substances
12 Certificates
■ Installation certificate
■ System handover certificate
■ Customer instruction certificate
■ Constancy test certificate
Introduction – Page 2 of 4
Who supplies what? The documents in the System Owner Manual.
representative
System Owner
Manufacturer
Siemens
Document
14 Updates
■ Performed updates
15 Room layout
■ Practice and room drawings
17 Declaration of conformity
Your Siemens service technician will find the appropriate conformity declaration under the following
address:
https://intranet.medical.siemens.com/Sales+Intranet+International/Divisions/Magnetic+Resonance/
CrossProductInformation/?languagecode=de
Select the “Certificates & Declarations” link under the “Regulatory” section of “Cross Product Informa-
tion”.
Introduction – Page 3 of 4
Information about branding
Siemens Healthineers
Siemens has recently reorganized its organizational group structure. The healthcare business now is a
separate legal entity. This is manifested in our new brand name:
While we have adopted this new brand name, the legal name of our company and the name of the
Legal Manufacturer do not change. This change also does not impact any legal obligations. This new
brand name is reflected on newly created product labeling (for example, user and technical manuals).
Already existing material will not bear the new brand name.
Introduction – Page 4 of 4
www.siemens.com/healthcare
MAGNETOM Family
System Owner Manual – Safety Information
syngo MR E11
Table of contents
Index 19
1.1 Preamble
This section of the system owner manual contains the most important safety
aspects for which you, as the owner of the MR system, will be responsible.
These include legal requirements, emergency plans, employee information and
qualifications, as well as requirements that must be met in the examination
room.
1.1.2 Responsibility
Siemens accepts no responsibility for the safety, reliability, and performance of
the MR system, if the MR system is not used in accordance with the instructions
for use (Operator Manual, System Owner Manual). Siemens is also not
responsible for any direct or indirect damages caused by incorrect operation.
This includes, but is not limited to, accidents with ferromagnetic objects. This
applies even if the consequences only become obvious at a later point in time.
1.2.6 Laser
The laser of the laser light localizer is classified as Class 2M according to IEC
60825-1 (Class II according to US CDRH).
WARNING
1.2.9 Installation
WARNING
WARNING
WARNING
Especially endangered patients and As special precautionary measures, a program for medical supervision as well as
MR workers a plan for using emergency equipment outside the magnet’s influence must be
in place for patients with a higher than normal risk factor, such as:
◾ Patients susceptible to cardiovascular collapse
◾ Patients who are likely to develop seizures
◾ Patients who are at an increased risk of heart attacks or other cardiac
problems
◾ Patients with limited thermoregulation
◾ Claustrophobic patients
◾ Patients who are unconscious, anesthetized or confused or who are not able
to communicate normally for other reasons
◾ Children
under two years of age. The responsible physician must evaluate the
benefits of the MR examination compared to those of other imaging
procedures.
WARNING
Quench During a quench, the super-conductivity of the magnet is lost. The energy of the
magnetic field is converted into heat and the magnet field strength falls off. The
liquid helium (coolant) boils off rapidly during this process and is released to the
outside via the exhaust vent line. The escape of gaseous helium via the exhaust
line is very noisy due to the high gas flow and you can hear a loud roaring noise
for several minutes.
A quench may occur as follows:
◾ Start-up of the MR system (ramping up or filling the magnet)
◾ An accident (earthquake, fire, etc.)
◾ Spontaneously without any obvious external reason (highly unusual)
Defective vent line If the vent line fails in part or fully, gaseous helium will enter the examination
room. In this case the air conditioning unit will not be able to ensure sufficient
air exchange and the following hazardous situations may arise:
◾ Poor visibility due to strong fog in the upper area of the room
◾ Rise in pressure in the examination room
◾ Hypothermia and risk of suffocation (e.g. in case of large leaks)
Mandatory reporting in case of fire 1 Prior to initial start-up of the MR system, ensure that the fire department is
informed about the MR system as well as structural on-site conditions.
2 Inform the fire department about the contents of the measurement
phantoms and the health risks caused by nickelous aerosol formations.
WARNING
MR workers and all personnel who have access to the MR system are not
sufficiently informed!
Personal injury, property damage
◆ Ensure that all personnel (incl. cleaning crews, rescue personnel, etc.)
are regularly informed about the potential risks inherent in MR systems
as well as the relevant safety information (for example, regarding
magnetic forces).
◆ The exclusion zone and corresponding safety measures must be
observed even when the system is switched off.
Training MR workers and physicians Personnel and physicians must be trained in the safe and effective use of MR
systems. The training must include the following topics:
◾ Emergency medical care
◾ Controlled access area
◾ Emergency switches
◾ Measures preventing fires
◾ Quench emergency plan
◾ Prevention of hazards related to magnet forces
◾ Combinations with other devices
The physician must complete a special training course on interpreting images.
WARNING
WARNING
WARNING
The air conditioning is installed on-site by the system owner. It is not part of
The functions of the air conditioning or the temperature and relative humidity
of the examination room must be checked at regular intervals.
( Page 16 Maintenance at regular intervals)
WARNING
1.8 Maintenance/repair
1.8.1 Responsibility
As a supplier, Siemens will not be held responsible for the safety, reliability, and
performance of the system in the following cases:
◾ Installations, additions, adjustments, modifications, and repairs to the MR
system, or changes to the software that are not performed by Siemens
Service.
◾ Assemblies are not replaced with original spare parts.
◾ The electrical wiring in the room does not meet the requirements of VDE
regulation 0100-710 or applicable national regulations.
Siemens is not responsible for potential damage in the event non-authorized
personnel refill the magnet with helium.
Modifications to the system or sub-systems during the service life require
evaluation to the requirements of IEC 60601-1.
WARNING
Daily checks In the course of system operation, technical and constructional changes may
have been made to the MR system and its environment. It must be ensured that
all components function satisfactorily and that the changes do not present
hazardous conditions.
After system acceptance by the customer, a daily visual inspection should be
performed for the MR system with respect to the following constructional
changes:
◾ Changes in the environment at the output of the exhaust line (e.g. window
installed retrospectively, inlet and outlet of air-conditioning units, new
buildings or temporarily installed containers)
◾ Changes to the air conditioning unit or venting system (e.g. by adding air
inlets and outlets in neighboring rooms)
◾ Installation of additional MR systems (e.g. prohibited use of the same exhaust
line for several MR systems)
◾ Constructional changes inside and outside the examination room
The examination room must be checked for (newly introduced) magnetic parts.
Annual checks The annual technical safety inspections are listed in this system owner manual
and may only be performed by Siemens Service.
Upon request, Siemens Service will provide technical documents for the MR
system (e.g. circuit diagrams, spare parts lists, descriptions, calibration
instructions). However, this does not constitute authorization for repairs.
Helium-related risks Liquid helium presents the following properties that, among other things, may
result in hazardous conditions when not handled professionally:
◾ Extremely cold: causes frostbites when it comes in touch with skin
◾ Oxygen in ambient air is displaced during boil-off: risk of asphyxiation
If the helium fill level is too low, the alarm box or the syngo Acquisition
Workplace will signal this accordingly.
◆ In case of alarm, notify Siemens Service and/or ensure refilling only through
trained and experienced personnel.
Storage
WARNING
A Responsibility
Safety instructions 16
Accidents
Fire fighting 10
S
Air conditioning
Ambient conditions 14 Safety information
System owner 5
Ambient conditions
Air conditioning 14 Safety instructions
Quench 10
C Signs and symbols
Warning and prohibition signs 14
Coolant
Safety information 17
W
E Warning labels
Warning and prohibition signs 14
Electromagnetic fields
Magnetic fringe field 14
Emergency 9
Emergency plan
Establishing 9
Emergency plans
Procedural instructions 8
Rescue scenarios 8
F
Fire
Requirement of information 10
Safety instructions 10
Fire|
Fire fighting 10
H
Hazards
Coolant 17
Laser 7
Helium
Hazards 18
Refilling 17
L
Laser light localizer
Safety instructions 7
M
Maintenance
Safety information 16
Maintenance/repair 16
P
Product safety
Safety instructions 16
Q
Quench
Safety instructions 10
R
Repair
Safety instructions 17
www.siemens.com/healthcare
MAGNETOM
MR
Aera
2
MR compatibility
The data and protocols included in this manual provide the operator with the means for
evaluating the MR compatibility of non-Siemens MR products with the MR system
described in the System Owner Manual.
The MR compatibility of non-Siemens MR products addresses products of both manufactur-
ers, the company producing the non-Siemens MR product and the company producing the
MR system. The operator is ensured the safe operation of the non-Siemens MR product
together with the MR system only when both manufacturers provide a test certificate
addressing the “testing the effect of non-Siemens MR products on MR systems” as well as
“testing the effect of MR systems on non-Siemens MR products”. In all other cases, the
operator has to ensure that use of the non-Siemens MR product does not conflict with the
functions of the MR system and vice versa.
According to IEC 60601-2-33, the manufacturer of the MR system is obligated to provide a
data sheet covering technical information of the MR system to enable items to be assessed
for MR compatibility.
A display of the contours of equal field strength of the MR system begins on page 5 of this
manual.
MR compatibility test
Manufacturers of a non-Siemens MR product, who would like to obtain a test certificate
(“testing the effect of non-Siemens MR products on MR systems”) for their device from
Siemens, should get in touch with their Siemens sales engineer.
Parameters
All system-specific parameters are included in the “Technical Data” register of the System
Owner Manual.
MR compatibility protocols
According to IEC 60601-2-33, protocols need to be proposed for testing the functionality of
non-Siemens MR products. The protocols listed below are routinely used in the system.
! The tests cannot be used to evaluate the effects of the non-Siemens MR product on
the image quality of the MR system.
! None of the service sequences listed below can be used to confirm MR compatibility
or to define MR Conditional parameters of devices in the MR environment.
Spatial distributions
This section includes the following spatial distributions diagrams:
Static magnetic field B0
Spatial gradient of B0
Product of the static magnetic field B0 and the spatial gradient of B0
1
Vertical (m)
−1
40 mT
20 mT
10 mT
5 mT
3 mT 1 mT
−2 0.5 mT
−3
−4
−5
−5 −4 −3 −2 −1 0 1 2 3 4 5
Horizontal (m)
0.5
Vertical Axis (m)
mT
1m
1
3m
5m T
10 T
T
20 mT
40
20
T
T
m
m
0m
T
0
−1
−2
−3
−4
−5
−5 −4 −3 −2 −1 0 1 2 3 4 5
Z Axis (m)
0.
5
1
Horizontal Axis (m)
m
m
T
T
20
1
3 T
5
10
m
m
40
20
m
T
T
m
0m
T
T
T
0
−1
−2
−3
−4
−5
−5 −4 −3 −2 −1 0 1 2 3 4 5
Z Axis (m)
Top view
Magnet
X Axis (m)
Z Axis (m)
Side view
Patient table
Y Axis (m)
Z Axis (m)
A small overview graphic is displayed on the following pages to indicate the area mapped
by the respective graph.
1.8
1.6
1.4
1.2
Vertical (m)
Side view
1
0.8
0.6
1.9 T
0.4
0.2 1.5 T
0.5 T
1T
0 Patient table
1.5 T
−0.2
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Z Axis (m)
! The iso-magnetic contour lines are accurate to a value better than 1%. There is very
little influence from the environment to the scanner. The graphic shows the cover in
nominal position. On a variety of scanners, the position can vary ±5 mm in the axial
direction. The maximum value therefore has a tolerance of ±10%, as small geometric
deviations can cause a significant change in the value.
1T
0.5 T
1.5 T
−0.2
1.5 T
−0.4 1.9 T
−0.6
−0.8
Horizontal (m)
−1
Top view
−1.2
−1.4
−1.6
−1.8
−2
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Z Axis (m)
1.8
1.6
1.4
1.2
Vertical (m)
Side view
1
0.8 0.
1T
/m
0.6
0.3
0.5
T/m
11 T/m
1T
T/m
/m
0.4 10 T/m
7 T/m
5 T/m
0.2
3 T/m
0.1 T/m
0.3 T/m
0.5 T/m
0
1 T/m
Patient table
3 T/m
−0.2
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Z Axis (m)
! The iso-magnetic contour lines are accurate to a value better than 1%. There is very
little influence from the environment to the scanner. The graphic shows the cover in
nominal position. On a variety of scanners, the position can vary ±5 mm in the axial
direction. The maximum value therefore has a tolerance of ±10%, as small geometric
deviations can cause a significant change in the value.
0.1 T/m
0.3 T/m
0.5 T/m
1 T/m
1 T/m
0.5 T/m
0.3 T/m
0.1 T/m
3 T/m
−0.2
5 T/m
7 T/m
−0.4
11 T/m
−0.6
−0.8
Horizontal (m)
−1
Top view
−1.2
−1.4
−1.6
−1.8
−2
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Z Axis (m)
1.8
1.6
1.4
1.2
Vertical (m)
Side view
1
0.8
0.
1T
0.3 T2 /m
0.6
2 /m
0.5 2 m
T/
1T
2 m
/
17 T2/m
0.4
10 T2/m
13 T2/m
5 T2/m
0.2
3 T2/m
0.1 T2/m
0.3 T/2m
0.5 T/m
1 T/2m
0 Patient table
3 T2/m
−0.2
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Z Axis (m)
0.1 T2/m
0.3 T2/m
0.5 T2/m
1 T2/m
1 T2/m
0.5 T2/m
0.3 T2/m
0.1 T2/m
3 T2/m
−0.2 5 T2/m
10 T2/m
−0.4 13 T2/m
17 T2/m
−0.6
−0.8
Horizontal (m)
−1
Top view
−1.2
−1.4
−1.6
−1.8
−2
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Z Axis (m)
|B| [T]
0.07
0.06
0.05
End of magnet
0.04
0.03
0.02
0.01
0.00
0.00 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 z [m]
Magnet isocenter
XQ gradient system
|B| [T]
0.06
0.05
0.03
0.02
0.01
0.00
0.00 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 z [m]
Magnet isocenter
XJ gradient system
The areas shaded in grey indicate the length of the magnet.
|B| [T]
0.04
0.02
0.60 m
0.40 m
0.01
0.20 m
0.00
0.0 0.5 1.0 1.5 2.0 z [m]
Magnet isocenter
XQ gradient system
|B| [T]
0.03
End of gradient coil
0.02
0.60 m
0.01 0.40 m
0.20 m
0.00
0.0 0.5 1.0 1.5 2.0 z [m]
Magnet isocenter
XJ gradient system
The areas shaded in grey indicate the length of the gradient coil.
1
End of magnet
0.1
relative B1(z)2/B1(0)2
0.01
0.001
0.0001
0.00001
0.000001
0.00 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 z [m]
Magnet isocenter
The RF field is calculated along the center line starting in the magnet isocenter. The area
shaded in grey indicates the length of the magnet.
The ratio B1(z)2/B1(0)2 provides a worst case estimation of the SAR contribution to a per-
son who is positioned at a distance z from the isocenter. The SAR contribution is relative to
the SAR that is applied to a person in the center of the patient bore.
For example a person standing in front of the system aperture absorbs a maximum of 0.2%
of the RF power which is applied to a patient scanned in the center of the bore.
Magnet data
Operating field strength 1.5 Tesla
Magnet type Superconductor
Field stability over time <0.1 ppm/h
Weight (with cryogens) 3118 kg
Magnet length 137 cm
System length cover to cover 145 cm
Open Bore design1 70 cm
Refill interval (typical)2 Not applicable
Boil-off rate (typical)2 0.0 l/year
Max. helium capacity approx. 1280 liters
Minimum helium level 35%
Cryostat Stainless steel
RF data
Peak power of transmitter amplifier 26.1 kW
Transmitter bandwidth 800 kHz
Receiver bandwidth 500 Hz–1 MHz (for each channel)
RF transmit coils
Body coil Max. applied RF field B1+ 27 µT
Max. specified B1+ rms 6.2 µT
Distance to isocenter, where RF transmit field is reduced by
3 dB 0.18 m
10 dB 0.38 m
TxRx CP Extremity Max. applied RF field B1+ 47 µT
Max. specified B1+ rms 17 µT
TxRx 15 Ch Knee Max. applied RF field B1+ 47 µT
Max. specified B1+ rms 13 µT
TxRx Knee 15 Max. applied RF field B1+ 47 µT
Max. specified B1+ rms 13 µT
TxRx CP Head Max. applied RF field B1+ 47 µT
Max. specified B1+ rms 12.5 µT
Tim table
Max. patient weight for vertical and horizontal table 250 kg (550 lbs)
movement
Max. scan range 140 cm; optional 205 cm1
Vertical table movement Range 52–102 cm2
+13 mm3
Speed 6 cm/s
Horizontal table move- Max. range 261 cm
ment
Max. speed 20 cm/s
Horizontal accuracy for ±0.5 mm
repositioning from one
direction
Continuous table movement during scan capable
Patient comfort
Open bore design1 70 cm
In-bore lighting can be set at 6 different levels
In-bore ventilation can be set at 6 different levels
In-bore intercom including loudspeaker, microphone and earphones
Gradient data
XQ gradients
Performance for each axis
Max. amplitude 45 mT/m
Min. rise time 225 µs
Max. slew rate 200 T/m/s
Vector gradient performance (vector addition of all 3 gradient axes)
Max. eff. amplitude 78 mT/m
Max. eff. slew rate 346 T/m/s
Gradient duty cycle 100%
XJ gradients
Performance for each axis
Max. amplitude 33 mT/m
Min. rise time 264 µs
Max. slew rate 125 T/m/s
Vector gradient performance (vector addition of all 3 gradient axes)
Max. eff. amplitude 57 mT/m
Max. eff. slew rate 216 T/m/s
Gradient duty cycle 100%
Aera
2
Data sheet
The following document applies to all products provided by the Magnetic Resonance (MR)
product group of Siemens Healthcare.
Deviations and additions to this document are provided in accompanying product-specific
documents. This information has to be followed respectively applied.
Medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the accompa-
nying documents.
Electromagnetic compatibility is the ability of an equipment or system to function satisfac-
torily in its electromagnetic environment without introducing intolerable electromagnetic
disturbance to anything in that environment.
Portable and mobile RF communications equipment can affect medical electrical equipment.
! Fixed equipment or system cabling, which can not be removed by the user, is not
listed. This cabling is part of the system and was regarded at all EMC-considerations.
Without this cabling there is no complete functionality of the system.
! The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the equipment or
system as replacement parts for internal components, may result in increased emis-
sion or decreased immunity of the equipment or system.
RF shielding
! Failure to use MR system in the specified type of shield location could result in degra-
dation of the performance of this equipment, interference with other equipment or
interference with radio services.
! Recommended test methods for measurement of RF shielding effectiveness and
RF filter attenuation are given in the “Planing Guide System”.
! The use of RF emitting equipment, not released or certified by Siemens, inside the
shielded location of the MR system is not allowed, unless the equipment is switched
to an RF silent mode before entering the shielded location. In case of doubt please
check the operator manual or the guidelines of the respective 3rd party equipment.
Electromagnetic emissions
The MR system is intended for use in the electromagnetic environment specified below. The
customer or the user of the MR system should assure that it is used in such an environment.
! It is essential that the actual shielding effectiveness and filter attenuation of the
shielded location be verified to assure that they meet the minimum specification.
! The emissions characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
! The MR system or equipment should not be used adjacent to other equipment, if adja-
cent or beside use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
Electromagnetic immunity
The MR system is intended for use in the electromagnetic environment specified below.
The customer or the user of the MR system should assure that it is used in such an environ-
ment. Normally, the MR system is used in a Professional Healthcare Environment.
Electromagnetic environ-
Immunity tests IEC 60601 test level Compliance level
ment — guidance
Electrostatic ±8 kV contact ±8 kV contact Floors should be wood, concrete
discharge (ESD) ±2/4/8/15 kV air ±2/4/8/15 kV air or ceramic tile. If floors are
IEC 61000-4-2 covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast ±2 kV for power supply ±2 kV for power supply Mains power quality should be
transient/burst lines lines that of a typical commercial or
IEC 61000-4-4 ±1 kV for signal lines ±1 kV for signal lines hospital environment.
Surge ±0.5/1/2 kV line to ±0.5/1/2 kV line to Mains power quality should be
IEC 61000-4-5 ground ground that of a typical commercial or
±0.5/1 kV line to line ±0.5/1 kV line to line hospital environment.
Voltage dips, short <5% of a.c. mains Voltage interruptions; Mains power quality should be
interruptions and volt- voltage 0% UT; that of a typical commercial or
age variations on (>95% dip in a.c. 250/300 cycle hospital environment. If the
power supply input mains voltage) user of the equipment or sys-
lines IEC 61000-4-11 for 5 s tem requires continued opera-
tion during power mains
interruptions, it is recom-
mended that the equipment or
system be powered from an
uninterruptible power supply.
The MR system has a rated
input current of more than
16 A per phase.
Power frequency 30 A/m 30 A/m Power frequency magnetic
(50/60 Hz) magnetic fields should be at levels char-
field IEC 61000-4-8 acteristic of a typical location in
a typical commercial or hospital
environment.
Radiated RF inter- 80 MHz to 2.7 GHz: 3 V/m frequency range: 40–500 MHz
ference 3 V/m minimum RF shielding effectiveness: >65 dB
IEC 61000-4-3
RF attenuation of the RF cabin itself (without
Proximity field 385 MHz: 27 V/m any feedthroughs or filters): >90 dB
from RF wireless 450 MHz: 28 V/m 1.5 T and 3 T, including Multi-Nuclear Option:
communications
equipment 710/745/780 MHz: frequency range: 15–125 MHz
IEC 61000-4-3 9 V/m minimum RF shielding effectiveness: >65 dB
810/870/930 MHz: RF attenuation of the RF cabin itself (without
28 V/m any feedthroughs or filters): >90 dB
1720/1845/ Verification of the shielding requirements is
1970 MHz: 28 V/m essential prior hand over of the system. The
2450 MHz: 28 V/m integrity of the RF shielding is in the responsi-
bility of the customer.
5240/5500/
5785 MHz: 9 V/m Field strengths outside the shielded location
from fixed RF transmitters, as determined by
an electromagnetic site survey, should be less
than 3 V/m1.
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
1. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength outside the shielded location in which the MR system is used exceeds 3 V/m, the MR system should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
relocating the MR system or using a shielded location with a higher RF shielding effectiveness and filter attenuation.
Test Immunity
Maximum
frequency Band (MHz) Service Modulation Distance (m) test level
power (W)
(MHz) (V/m)
385 380–390 TETRA 400 Pulse modulation 1.8 0.3 27
18 Hz
450 430–470 GMRS 460; FM 2 0.3 28
FRS 460 ±5 kHz deviation
1 kHz sine
710 704–787 LTE Band 13, Pulse modulation 0.2 0.3 9
17 217 Hz
745
780
810 800–960 GSM 800/900; Pulse modulation 2 0.3 28
TETRA 800; 18 Hz
870
iDEN 820;
930 CDMA 850;
LTE Band 5
1720 1700–1990 GSM 1800; Pulse modulation 2 0.3 28
CDMA 1900; 217 Hz
1845
GSM 1900;
1970 DECT;
LTE Band 1, 3,
4, 25;
UMTS
Test Immunity
Maximum
frequency Band (MHz) Service Modulation Distance (m) test level
power (W)
(MHz) (V/m)
2450 2400–2570 Bluetooth; Pulse modulation 2 0.3 28
WLAN 217 Hz
802.11 b/g/n;
RFID 2450;
LTE Band 7
5240 5100–5800 WLAN 802.11 Pulse modulation 0.2 0.3 9
a/n 217 Hz
5500
5785
! The MR system may be interfered with by other equipment, even if that other equip-
ment complies with CISPR emission requirements.
The MR system applies IQ-modulated (without encoding any data) repeated RF pulses. This
results in a total bandwidth of the modulated signal of less than ±500 kHz around applied
carrier (<125 MHz).
Technical data
Aera
2
General remark
The following pages describe common information and information specific for the system
configuration with XJ gradients. Please find specific information for MAGNETOM Aera sys-
tems with XQ gradients beginning on page 31.
Magnet system
Short-bore, patient-friendly design, high homogeneity
Easy siting due to AS (Active Shielding) and E.I.S. (External Interference Shielding)
magnet technology
Operating mode: Standard operating mode according to IEC 60601-2-33.
Magnet parameters
Operating field strength 1.5 Tesla
Magnet type Superconductor
Field stability over time <0.1 ppm/h
Weight (with cryogens) 3118 kg
Magnet length 137 cm
System length cover to cover 145 cm
Open Bore design1 70 cm
Type of installation Fixed
Decay characteristics from full field to approx. 20 seconds
20 mT
Technical data
3
Shimming
Both: passive and active shimming. Passive shimming during installation.
Standard active shim with 3 linear channels (1st order). Additional 5 non linear chan-
nels (2nd order) as option (XQ gradients).
3D Shim Patient-specific automated shim
Time to shim = approx. 20 s
Technical data
4
Shielding
Active Shielding (AS) 5th generation active shielding (AS) tech-
nology with counter coils
Fringe field (axial×radial) 0.5 mT1 4.0 m×2.5 m
0.1 mT 5.5 m×3.1 m
External Interference Shield (E.I.S.) Patented shielding system integrated into
the magnet
Continuous compensation and automatic
suppression of external magnetic field
interferences during measurement (caused
by moving ferromagnetic objects or nearby
power lines)
1. For typical clinical use, depending on sequences and operating time with running helium compressor.
The system needs to be serviced at regular interval. Undisturbed magnet cooling for 24 hours and
7 days a week.
Technical data
5
Technical data
6
Patient handling
The patient table is available in two different configurations:
Tim table
Max. patient weight for vertical and horizontal table 250 kg (550 lbs)
movement
Max. scan range 140 cm
optional 205 cm1
Vertical table movement Range 52–102 cm2
+13 mm3
Speed 6 cm/s
Horizontal table movement Max. range 261 cm
Max. speed 20 cm/s
Horizontal accuracy for ±0.5 mm
repositioning from one
direction
Continuous table movement during scan capable
Technical data
7
Technical data
8
Patient noise
Patients require hearing protection with an SNR1 = 18 dB or more
Technical data
9
Gradient system
General features
Actively shielded (AS) whole-body gradient coil system
Extremely low eddy currents
Water-cooled coil and amplifier for maximum performance
All axes force compensated
Gradient amplifier
Water-cooled, highly compact, modular design
Ultra-fast solid-state technology with very low switching losses
Max. output voltage1 2000 V
Max. output current1 625 A
Performance
Performance for each axis
Max. amplitude 33 mT/m
Min. rise time 264 µs
Max. slew rate 125 T/m/s
Vector gradient performance (vector addition of all 3 gradient axes)
Max. eff. amplitude 57 mT/m
Max. eff. slew rate 216 T/m/s
Gradient duty cycle 100%
Technical data
10
DirectRF™ technology
Technical data
11
RF receiver technology
Maximum number of channels1 204
Number of independent receiver 24, 48, 642
channels that can be used simulta-
neously in one single scan in one FOV,
each generating an independent partial
image
Quadrature demodulation and filtering Digital
Receiver bandwidth 500 Hz–1 MHz (for each channel)
Receiver signal resolution 32 bit
ADC sampling rate 80 MHz
Preamplifier noise figure <0.5 dB
Dynamic range at coil connector 164 dB instantaneous at receiver
(referred to 1 Hz resolution bandwidth) 169 dB with automatic gain control at local
coil connector
Technical data
12
RF coils
Intended use
Coil name Coil type Coil design Resonant
Applications
nuclei
Body 18 Receive 18 Channel 1H Examinations of the upper body and the
phased array abdomen
Examinations of the pelvis
Head/Neck 20 Receive 20 Channel 1H Examinations of the head and the neck
phased array (e.g. nape of the neck and vessels of the
neck)
Combined measurements with other
coils
Spine 32 Receive 32 Channel 1
H Examinations of the spine
phased array Combined measurements with other
coils
Flex Large 4 Receive 4 Channel 1
H Imaging of large regions such as medium
phased array to large shoulder, hip, and knee
Flex Small 4 Receive 4 Channel 1
H Imaging of small regions such as small to
phased array medium shoulder, wrist, elbow, and
ankle
Peripheral Angio 36 Receive 36 Channel 1H Vascular examinations of the lower
phased array extremities
Technical data
13
Intended use
Coil name Coil type Coil design Resonant
Applications
nuclei
Loop 11 Receive Linear 1H Examinations of the skin, for example,
melanoma, birth marks, and other skin
anomalies
Examinations of fingers, wrist bones, or
toes
Loop 7 Receive Linear 1
H Examinations of the skin, for example,
melanoma, birth marks, and other skin
anomalies
Examinations of fingers, wrist bones, or
toes
Loop 4 Receive Linear 1H Examinations of the skin, for example,
melanoma, birth marks, and other skin
anomalies
Examinations of fingers, wrist bones, or
toes
Hand/Wrist 16 Receive 16 Channel 1H Examinations of wrists, hands, and
phased array fingers
Foot/Ankle 16 Receive 16 Channel 1H Examinations of feet and ankles
phased array
Shoulder Large 16 Receive 16 Channel 1H Examinations of the shoulder
phased array
Shoulder Small 16 Receive 16 Channel 1H Examinations of the shoulder
phased array
Technical data
14
Intended use
Coil name Coil type Coil design Resonant
Applications
nuclei
TxRx CP Extremity Transmit/ Quadrature 1H Examinations of the left or right knee
Receive birdcage
TxRx 15 Ch Knee Transmit/ Transmit: 1H Examinations of the left or right knee
Receive quadrature
birdcage
Receive:
15 Channel
phased array
TxRx Knee 15 Transmit/ Transmit: 1H Examinations of the left or right knee
Receive quadrature
birdcage
Receive:
15 Channel
phased array
TxRx Knee 15 Flare Transmit/ Transmit: 1H Examinations of the left or right knee
Receive quadrature
birdcage
Receive:
15 Channel
phased array
4 Ch BI Breast Receive 4 Channel 1H Biopsy examinations
phased array High-resolution breast imaging
16 Ch AI Breast Receive 16 Channel 1
H Biopsy examinations
phased array High-resolution breast imaging
Breast 18 Receive 18 Channel 1
H Examinations of female patients
phased array Simultaneousl display of both sides of the
breast for comparative evaluations
Technical data
15
Intended use
Coil name Coil type Coil design Resonant
Applications
nuclei
2-/4-/8-Channel Receive Optional 2, 4, 1H Biopsy examinations
Sentinelle Breast or 8 Channel Conventional breast imaging
Coil phased array
2-/10-/16-Channel Receive Optional 2, 1H Biopsy examinations
Sentinelle Breast 10, or 16 Conventional breast imaging
Coil Channel
phased array
Body 6 Receive 6 Channel 1H Examinations of the upper body and the
phased array abdomen
Examinations of the pelvis
Body 6 Long Receive 6 Channel 1H Body 6 with extended coil cable
phased array Examinations of the upper body and the
abdomen
Examinations of the pelvis
Body 18 Long Receive 18 Channel 1H Body 18 with extended coil cable
phased array Examinations of the upper body and the
abdomen
Examinations of the pelvis
Body 301 Receive 30 Channel 1H Examinations of the upper body,
phased array abdomen, and liver
Examinations of the pelvis
Examinations of the heart
Body 602 Receive 60 Channel 1H Examinations of the upper body,
phased array abdomen, and liver
Examinations of the pelvis
Examinations of the heart
Technical data
16
Intended use
Coil name Coil type Coil design Resonant
Applications
nuclei
Head/Neck 163 Receive 16 Channel 1H Examinations of the head and the neck
phased array (e.g. nape of the neck and vessels of the
neck)
Combined measurements with other
coils
Spine 243 Receive 24 Channel 1
H Examinations of the spine
phased array Combined measurements with other
coils
Endorectal Receive Linear 1
H Examinations of the prostate, colon, and
Cervix Uteri
Prostate 2 Receive 2 Channel 1
H Examinations of the prostate, colon, and
phased array the surrounding tissue in the pelvis
TxRx CP Head Transmit/ Quadrature 1
H Vascular examinations in the head
Receive birdcage
Special Purpose 4 Receive 4 Channel 1
H Examinations of small body regions near
phased array the skin surface
Pediatric 16 Receive 16 Channel 1
H Imaging of the head and neck of
phased array neonates and infants4
Technical data
17
Sequences
Basic set of sequences provided on the scanner.
Technical data
18
Computer system
1. A standard monitor without calibration is not suitable for diagnostic purposes. Please consider the ini-
tial acceptance testing for image display devices and the follow-up service for constancy testing on a
regular base, as offered by Siemens service.
Technical data
19
Technical data
20
Network/data coupling
The MR system provides the level of safety according IEC 60950-1 outside patient environ-
ment. All equipment connected to the system’s network/data couplings must also provide
minimum level of safety according IEC 60950-1.
Technical data
21
Windowing
Freely selectable window width and center
Windowing on succeeding images
Auto-windowing for optimized contrast
Saves and sends window values
Evaluation
Parallel evaluation of up to 40 regions of interest
Circle
Rectangle
Freehand ROI
Pixel lens with position marker
Statistical evaluation
Area
Standard deviation
Mean value
Min/max values
Image scrolling
Magnification
Distance
Angle
Technical data
22
2D Post-processing
Image manipulations
Reversal of gray-scale values
Image rotation by 90° or by user-defined angle
Flip horizontally/vertically
Image zoom and pan
Shutter
Annotation
Position display
Displays measured slice positions on localizer image and selected series
Mean Curve
Time-intensity analysis
Creates and edits DICOM structured reports
Filming
Connection via DICOM Basic Print
Interactive filming
Filming parallel to other activities
Independent scanning and documentation – no wait time due to camera delays
Freely selectable positioning of images onto virtual film sheet
Selectable various film layouts
Mother-in-Child display
Windowing, image zoom and pan on film sheet
Configurable image text
Simultaneous handling of multiple film jobs
Up to 100 virtual film sheets
Technical data
23
Argus Viewer
Viewing software for cardiac MR studies and large data sets
Efficient cine review of cardiac and other dynamic data sets
Multiple sorting options
Single movie as well as 2, 4, or 8 simultaneous slices together in movie mode
Rapid avi creation of 1 to 8 slices simultaneously
Creates and edits DICOM structured reports
Dynamic Analysis
Arithmetic operations on images and series
Addition, subtraction, multiplication, division of single images and whole series
Arithmetic mean and standard deviation across a range of selected images
Calculation of T1 and T2, and logarithmic images
Differentiation/integration of selected images
Calculation of a mean slope image from a range of selected images
Calculation of z-score (t-test) images for evaluation of BOLD imaging data (Blood
Oxygenation Level Dependent)
Time-to-peak evaluation (TTP)
ADC maps
Several evaluation functions may be started consecutively in the background
Printing on paper
Interface and software for printing images on paper (laser printer not included)
Grey levels and color printing supported
Data format Postscript Level 2
Technical data
24
3D Post-processing
MPR – Multi-Planar Reconstruction
Real-time multi-planar reformatting of secondary views
Viewing perspectives: sagittal; coronal; axial; oblique; double oblique; curved (freehand)
Reconstruction along polygon and/or curved (freehand) cut lines
Reconstruction based on reconstructed planes possible
Reconstruction of user-defined ranges of parallel, radial or freehand cuts
Selectable slice thickness and slice increment of reconstructed images
Storing of post-processing protocols
Annotations and 2D evaluations such as distance and ROI
Technical data
25
Ambient conditions
Control room
Temperature 15–30 °C (59–86 °F)
Relative air humidity 40 to 60%
Absolute air humidity <11.0 g/kg
Air pressure 1060–700 hPa1
Electronics room
Temperature 15–30 °C (59–86 °F)
Relative air humidity 40 to 80%
Absolute air humidity <11.0 g/kg
Air pressure 1060–700 hPa1
Examination room
Temperature 18–22 °C (64–71 °F)
Relative air humidity 40 to 60%
Absolute air humidity <11.0 g/kg
Air pressure 1060–700 hPa1
Technical data
26
Cooling system
Two different customer specific cooling alternatives (Separator or Eco Chiller) are avail-
able.
Technical data
27
Values Tolerance
Voltage 380 V, 400 V, 420 V, 440 V, ±10%
460 V, 480 V
Frequency 50/60 Hz ±1 Hz
Connection value 63 kVA for 480 V/400 V,
62 kVA for 380 V
For all products, line power has to be supplied via an on-site system contact or via another
multipole shut-down mechanism. Room installation has to be in compliance with
VDE 0100-7101.
1. In all countries, compliance with local and national legal regulations is required.
However, we strongly recommend compliance with the regulations described herein – to the extent permit-
ted by relevant local and national laws – in order to ensure the safety of operating personnel, patients, and
third parties.
380/400/420/440/460/480 V ~
3PHASE 50/60 Hz
SHORT – TIME kVA: 75
LONG – TIME kVA: 20
Technical data
28
Power consumption
Off1 6.1 kW
System ready to measure1 8.7 kW
Scan1 18.9 kW
Remarks:
All values are typical values, applicable for 400 V/50 Hz.
Consumption for optional separator pump not included.
1. The power consumption described herein is based on results that were achieved in a setting according
to the COCIR methodology MRI - Measurement of the energy consumption (http://www.cocir.org/site/
index.php?id=46). Since many variables impact power consumption (e.g. sequences used for scan-
ning and sequence parameters, scan time), there can be no guarantee that each customer will achieve
the same values.
Technical data
29
General classifications
Protection class I
Components used
Patient table Type B applied part
Local coils Type B or BF applied part
Body coil Type B applied part
ECG/Pulse module Type BF applied part
IP protection class according to IP X0
IEC 60529
Explosion protection The MR system is not intended for opera-
tion in areas prone to explosion (e.g.,
highly flammable mixtures of anaesthesia
gases with air or oxygen or nitrous oxide)
Operating mode Continuous operation with short-term load
Disinfection receptors/system compo- Disinfectants without alcohol, ether
nents
Degree of safety in the presence of a No AP or APG category equipment
flammable anesthetics mixture with air
or with oxygen or with nitrous oxide
Mains operated equipment with addi- None
tional power sources
Technical data
30
Dimensions
Heat
Width Depth Height Weight dissipa-
Component
[cm] [cm] [cm] [kg] tion
[kW]
Examination Room
Magnet 1.5 Tesla AS 205 137 215 3118
(incl. Helium)
Magnet in operation, 231 405 219 4798
incl. gradient coil, 4331
body coil, Tim table,
and covers
Tim table 76 249 52–1022
+13 mm3
Required min. room 2404
height clearance
Min. transport dimen- 231 155 214
sions
Control Room
syngo Acquisition 120 80 117
Workplace (72+45)
(table+monitor)
Host computer 22 46 47
syngo MR Workplace 120 80 117
(optional) (72+45)
(table+monitor)
Equipment Room
Electronics cabinet, 160 65 1985 1500 56
incl. system control,
RF system, gradient
power system, image
processor
Cooling system 65 65 189 500
Technical data
31
Shimming
Both: passive and active shimming. Passive shimming during installation.
Standard active shim with 3 linear channels (1st order). Additional 5 non linear chan-
nels (2nd order).
3D Shim Patient-specific automated shim
Time to shim = approx. 20 s
Technical data
32
Hearing protection data
The A-evaluated, effective sound pressure level was measured according to NEMA
MS 4-2006 (National Electrical Manufacturers Association) using the maximum gradient
acoustic noise (MGAN) method.
Patient noise
Patients require hearing protection with an SNR1 = 22 dB or more
Gradient system
Gradient amplifier
Water-cooled, highly compact, modular design
Ultra-fast solid-state technology with very low switching losses
Max. output voltage1 2250 V
Max. output current1 900 A
Performance
Performance for each axis
Max. amplitude 45 mT/m
Min. rise time 225 µs
Max. slew rate 200 T/m/s
Vector gradient performance (vector addition of all 3 gradient axes)
Max. eff. amplitude 78 mT/m
Max. eff. slew rate 346 T/m/s
Gradient duty cycle 100%
Technical data
33
Cooling system
Two different customer specific cooling alternatives (Separator or Eco Chiller) are avail-
able.
Values Tolerance
Voltage 380 V, 400 V, 420 V, 440 V, ±10%
460 V, 480 V
Frequency 50/60 Hz ±1 Hz
Connection value 88 kVA for 480 V/400 V,
90 kVA for 380 V
For all products, line power has to be supplied via an on-site system contact or via another
multipole shut-down mechanism. Room installation has to be in compliance with
VDE 0100-7101.
1. In all countries compliance with local and national legal regulations is required.
However, we strongly recommend compliance with the regulations described herein – to the extent permit-
ted by relevant local and national laws – in order to ensure the safety of operating personnel, patients, and
third parties.
Technical data
34
Power rating plate
380/400/420/440/460/480 V ~
3PHASE 50/60 Hz
SHORT – TIME kVA: 102
LONG – TIME kVA: 20
Power consumption
Off1 6.1 kW
System ready to measure1 9.3 kW
Scan1 22.5 kW
Remarks:
All values are typical values, applicable for 400 V/50 Hz.
Consumption for optional separator pump not included.
1. The power consumption described herein is based on results that were achieved in a setting according
to the COCIR methodology MRI - Measurement of the energy consumption (http://www.cocir.org/site/
index.php?id=46). Since many variables impact power consumption (e.g. sequences used for scan-
ning and sequence parameters, scan time), there can be no guarantee that each customer will achieve
the same values.
Technical data
MAGNETOM
MR
Location of labels
Aera
Avantofit
Prisma/Prismafit
Skyra/Skyrafit
2
MAGNETOM XXXX
(01) 00000123456789 GTIN
(240) 01234567 MODEL
(422) 012 ORIGIN DE
(21) 01234 SERIAL 2016 0123
SYSTEM IVK
Siemens Healthcare GmbH, Henkestr. 127
91052 Erlangen, Germany
Made in Germany
Within the EU, products identified with this symbol are subject to guidelines 2002/96/EC
for old electrical or electronics system and 2012/19/EU respectively. Please contact
Siemens Service in case of questions about returning and disposing the MR system and/or
its components and accessories.
Revision label
Location of labels
3
MAGNETOM XXXX
(01) 00000123456789 GTIN
(240) 01234567 MODEL
(422) 012 ORIGIN DE
(21) 01234 SERIAL 2016 0123
SYSTEM IVK
Siemens Healthcare GmbH, Henkestr. 127
91052 Erlangen, Germany
Made in Germany
Location of labels
4
Location of labels
5
Location of labels
6
Location of labels
The Maintenance Plan has been removed
and is only available in the printed version
of the System Owner Manual.
MAGNETOM
MR
Disposal
Aera
Avantofit
Prisma/Prismafit
Skyra/Skyrafit
2
Equipment disposal
On disposing of the system or parts thereof, currently valid environmental legislation must
be observed.
Examples of environmentally relevant components are:
Accumulators and batteries
Transformers
Capacitors
Monitor picture tubes
Phantoms
For details contact your local customer service representative or your Siemens regional
office.
NOTICE: System components hazardous to persons or the environment must be dis-
posed of with care and in compliance with legally binding ordinances.
Disposal
MAGNETOM
MR
Aera
Avantofit
Prisma/Prismafit
Skyra/Skyrafit
Manufacturer Disclosure Statement for Medical Device Security – MDS2
HN 1-2013
The MRI Scanner is used to create diagnostic images which can be sent via the local network to DICOM nodes.
Note: External network connection (e.g. hospital network) is not mandatory to operate MAGNETOM systems.
Purpose of integrating the Device into an IT-Network (see Annex Z for details):
DICOM communication node for image transfer; SRS - Siemens Remote Service support
MANAGEMENT OF PRIVATE DATA
Yes, No,
Note #
Refer to Section 2.3.2 of HIMSS/NEMA HN 1-2013 standard for the proper interpretation of information requested
N/A, or
in this form.
See Note
A Can this device display, transmit, or maintain private data (including electronic Protected Health
Information [ePHI])? ........................................................................................................................................ Yes
B Types of private data elements that can be maintained by the device:
B.1 Demographic (e.g., name, address, location, unique identification number)? ........................................... Yes
B.2 Medical record (e.g., medical record #, account #, test or treatment date, device identification number)? Yes
B.3 Diagnostic/therapeutic (e.g., photo/radiograph, test results, or physiologic data with identifying
characteristics)? ........................................................................................................................................ Yes
B.4 Open, unstructured text entered by device user/operator? .................................................................... Yes
B.5 Biometric data? ....................................................................................................................................... No 1
B.6 Personal financial information? ................................................................................................................. No
C Maintaining private data - Can the device:
C.1 Maintain private data temporarily in volatile memory (i.e., until cleared by power-off or reset)? ............. Yes
C.2 Store private data persistently on local media? ....................................................................................... Yes
C.3 Import/export private data with other systems?........................................................................................ Yes
C.4 Maintain private data during power service interruptions? ...................................................................... Yes
D Mechanisms used for the transmitting, importing/exporting of private data – Can the device:
D.1 Display private data (e.g., video display, etc.)? ....................................................................................... Yes
D.2 Generate hardcopy reports or images containing private data? ............................................................... Yes 2
D.3 Retrieve private data from or record private data to removable media (e.g., disk, DVD, CD-ROM, tape,
CF/SD card, memory stick, etc.)? ............................................................................................................. Yes
D.4 Transmit/receive or import/export private data via dedicated cable connection (e.g., IEEE 1073, serial
port, USB, FireWire, etc.)? ........................................................................................................................ Yes 3
D.5 Transmit/receive private data via a wired network connection (e.g., LAN, WAN, VPN, intranet, Internet,
etc.)? ......................................................................................................................................................... Yes
D.6 Transmit/receive private data via an integrated wireless network connection (e.g., WiFi, Bluetooth,
infrared, etc.)? ........................................................................................................................................... No
D.7 Import private data via scanning? ........................................................................................................... No
D.8 Other? ....................................................................................................................................................... No
Private data are handled and stored in DICOM Format. You can refer to the DICOM Conformance Statement and
the link to the according internet page of Siemens Healthcare. http://www.healthcare.siemens.com/services/itstandards/
Management
dicom
of private
1) Biometric data can be created out of the MR DICOM data, e.g. facial recognition would be possible.
data notes:
2) Printer must be provided by customer.
3) Not required for intended use but possible.
3
© Copyright 2013 by the National Electrical Manufacturers Association and
the Healthcare Information and Management Systems Society.
HN 1-2013
SECURITY CAPABILITIES
Yes, No,
Note #
Refer to Section 2.3.2 of HIMSS/NEMA HN 1-2013 standard for the proper interpretation of information requested in
N/A, or
this form.
See Note
1-1 Can the device be configured to force reauthorization of logged-in user(s) after a predetermined length of
inactivity (e.g., auto-logoff, session lock, password protected screen saver)? ................................................. Yes 1
1-1.1 Is the length of inactivity time before auto-logoff/screen lock user or administrator configurable?
(Indicate time [fixed or configurable range] in notes.) ........................................................................... Yes
1-1.2 Can auto-logoff/screen lock be manually invoked (e.g., via a shortcut key or proximity sensor, etc.) by
the user? .............................................................................................................................................. Yes
ALOF
1) Depends on the configuration settings.
notes:
2-1 Can the medical device create an audit trail? .................................................................................................. Yes 1
2-2 Indicate which of the following events are recorded in the audit log:
2-2.1 Login/logout ............................................................................................................................................. Yes
2-2.2 Display/presentation of data .................................................................................................................... Yes
2-2.3 Creation/modification/deletion of data ..................................................................................................... Yes
2-2.4 Import/export of data from removable media ......................................................................................... Yes
2-2.5 Receipt/transmission of data from/to external (e.g., network) connection ............................................... Yes
2-2.5.1 Remote service activity .......................................................................................................................... Yes
2-2.6 Other events? (describe in the notes section) ......................................................................................... Yes 2
2-3 Indicate what information is used to identify individual events recorded in the audit log:
2-3.1 User ID .................................................................................................................................................... Yes
2-3.2 Date/time ................................................................................................................................................. Yes
AUDT
1) Audit trails are created in the means of Windows.
notes:
2) With the syngo Security package additional events (e.g. all events with regard to PHI) will be logged.
3 AUTHORIZATION (AUTH)
The ability of the device to determine the authorization of users.
3-1 Can the device prevent access to unauthorized users through user login requirements or other mechanism? Yes
3-2 Can users be assigned different privilege levels within an application based on 'roles' (e.g., guests, regular
users, power users, administrators, etc.)? ......................................................................................................... Yes 1
3-3 Can the device owner/operator obtain unrestricted administrative privileges (e.g., access operating system or
application via local root or admin account)? ....................................................................................................... Yes 2
AUTH 1) Yes, if the syngo Security package is activated.
notes: 2) No access to safety relevant functionality.
4
© Copyright 2013 by the National Electrical Manufacturers Association and
the Healthcare Information and Management Systems Society.
HN 1-2013
Yes, No,
Note #
Refer to Section 2.3. of HIMSS/NEMA HN 1-2013 standard for the proper interpretation of information requested in
N/A, or
this form.
See Note
4-1 Can the device owner/operator reconfigure product security capabilities? .................................................... Yes 1
CNFS
1) Some configurations need to be performed by SIEMENS Service.
notes:
6-1 Does the device provide an integral capability to de-identify private data? ....................................................... Yes 1
DIDT 1) Focus of the anonymization is on the DICOM patient quadruple. Additional country specific legal requirements are
notes: not fulfilled yet.
7-1 Does the device have an integral data backup capability (i.e., backup to remote storage or removable media
such as tape, disk)? ............................................................................................................................................. Yes 1
DTBK 1) Service data and configuration data can be backed up on HDD, CD-R or DVD-R by the service technician. Patient
notes: (image) data can be auto-transferred immediately after the MR exam (PACS).
9-1 Does the device ensure the integrity of stored data with implicit or explicit error detection/correction
technology? ......................................................................................................................................................... No 1
IGAU 1) The medical device is not intended to be used for patient data storage. Patient data should be transferred
notes: immediately to an appropriate archive.
5
© Copyright 2013 by the National Electrical Manufacturers Association and
the Healthcare Information and Management Systems Society.
HN 1-2013
Yes, No,
Note #
Refer to Section 2.3.2 of HIMSS/NEMA HN 1-2013 standard for the proper interpretation of information requested in
N/A, or
this form.
See Note
10-1 Does the device support the use of anti-malware software (or other anti-malware mechanism)? ................. Yes 1
10-1.1 Can the user independently re-configure anti-malware settings? ....................................................... No
10-1.2 Does notification of malware detection occur in the device user interface? ........................................ No
10-1.3 Can only manufacturer-authorized persons repair systems when malware has been detected? ........ Yes
10-2 Can the device owner install or update anti-virus software? .......................................................................... No
10-3 Can the device owner/operator (technically/physically) update virus definitions on manufacturer-installed anti-
virus software? ................................................................................................................................................ No 2
1) McAfee® Solidifier and Microsoft® EMET is used to protect the system.
MLDP
2) The protection of the system is based on Whitelisting and therefore no additional Anti Virus Software is part of the
notes:
System.
11 NODE AUTHENTICATION (NAUT)
The ability of the device to authenticate communication partners/nodes.
11-1 Does the device provide/support any means of node authentication that assures both the sender and the
recipient of data are known to each other and are authorized to receive transferred information? .................. No 1
NAUT
1) For remote service access (e.g. Siemens Remote Service) technical measures for node authentication are in place.
notes:
12-1 Does the device support user/operator-specific username(s) and password(s) for at least one user? ........... Yes
12-1.1 Does the device support unique user/operator-specific IDs and passwords for multiple users? ................... Yes 1
12-2 Can the device be configured to authenticate users through an external authentication service (e.g., MS Active
Directory, NDS, LDAP, etc.)? .............................................................................................................................. No
12-3 Can the device be configured to lock out a user after a certain number of unsuccessful logon attempts? ....... Yes
12-4 Can default passwords be changed at/prior to installation? ................................................................................ Yes
12-5 Are any shared user IDs used in this system? .................................................................................................... Yes
12-6 Can the device be configured to enforce creation of user account passwords that meet established complexity
rules? ................................................................................................................................................................... Yes
12-7 Can the device be configured so that account passwords expire periodically? .................................................. Yes
PAUT
1) Optional syngo Security package needs to be activated.
notes:
13-1 Are all device components maintaining private data (other than removable media) physically secure (i.e.,
cannot remove without tools)? ............................................................................................................................. No 1
PLOK
1) Customer can physically secure the device with very little effort
notes:
6
© Copyright 2013 by the National Electrical Manufacturers Association and
the Healthcare Information and Management Systems Society.
HN 1-2013
Yes, No,
Note #
Refer to Section 2.3.2 of HIMSS/NEMA HN 1-2013 standard for the proper interpretation of information requested
N/A, or
in this form.
See Note
14-1 In the notes section, list the provided or required (separately purchased and/or delivered) operating system(s)
- including version number(s). .......................................................................................................................... Yes 1
14-2 Is a list of other third party applications provided by the manufacturer available? ............................................ Yes 2
1) Windows 7 (64 bit) Embedded, Customized Linux (not visible for the customer network)
RDMP
2) Shortlist is part of the Security Whitepaper. Complete list can be provided on customer request. (on base of a non-
notes:
disclosure agreement)
15 SYSTEM AND APPLICATION HARDENING (SAHD)
The device's resistance to cyber attacks and malware.
15-1 Does the device employ any hardening measures? Please indicate in the notes the level of conformance to any
industry-recognized hardening standards. ........................................................................................................... Yes 1
15-2 Does the device employ any mechanism (e.g., release-specific hash key, checksums, etc.) to ensure the
installed program/update is the manufacturer-authorized program or software update? .................................... Yes
15-3 Does the device have external communication capability (e.g., network, modem, etc.)? ................................... Yes
15-4 Does the file system allow the implementation of file-level access controls (e.g., New Technology File System
(NTFS) for MS Windows platforms)? ................................................................................................................... Yes
15-5 Are all accounts which are not required for the intended use of the device disabled or deleted, for both users
and applications? ................................................................................................................................................. Yes
15-6 Are all shared resources (e.g., file shares) which are not required for the intended use of the device, disabled?
Yes
15-7 Are all communication ports which are not required for the intended use of the device closed/disabled? ....... Yes
15-8 Are all services (e.g., telnet, file transfer protocol [FTP], internet information server [IIS], etc.), which are not
required for the intended use of the device deleted/disabled? .......................................................................... Yes
15-9 Are all applications (COTS applications as well as OS-included applications, e.g., MS Internet Explorer, etc.)
which are not required for the intended use of the device deleted/disabled? ................................................... Yes
15-10 Can the device boot from uncontrolled or removable media (i.e., a source other than an internal drive or
memory component)? .......................................................................................................................................... Yes 2
15-11 Can software or hardware not authorized by the device manufacturer be installed on the device without the use
of tools? ............................................................................................................................................................... No
SAHD 1) DISA Secure Technical Implementation Guidelines (STIG)
notes: 2) Can be disabled on request.
16-1 Are security-related features documented for the device user? ........................................................................... Yes
16-2 Are instructions available for device/media sanitization (i.e., instructions for how to achieve the permanent
deletion of personal or other sensitive data)? ....................................................................................................... Yes 1
SGUD
1) Siemens Service has instructions in place to support the customer during the device/media sanitization.
notes:
7
© Copyright 2013 by the National Electrical Manufacturers Association and
the Healthcare Information and Management Systems Society.
HN 1-2013
Yes, No,
Note #
Refer to Section 2.3.2 of HIMSS/NEMA HN 1-2013 standard for the proper interpretation of information
N/A, or
requested in this form.
See Note
18-1 Can private data be transmitted only via a point-to-point dedicated cable? .................................................. No 1
18-2 Is private data encrypted prior to transmission via a network or removable media? (If yes, indicate in the
notes which encryption standard is implemented.) ......................................................................................... No 2
18-3 Is private data transmission restricted to a fixed list of network destinations? .............................................. Yes 3
1) Depends on the customer network design. In principle possible but not common.
TXCF 2) DICOM Encryption is supported and can be configured. Software based removable media encryption is not
notes: supported.
3) DICOM nodes are defined and identified using their IP address.
19 TRANSMISSION INTEGRITY (TXIG)
The ability of the device to ensure the integrity of transmitted private data.
19-1 Does the device support any mechanism intended to ensure data is not modified during transmission? (If
yes, describe in the notes section how this is achieved.) ............................................................................... Yes 1
TXIG
1) In the means of TCP/IP.
notes:
OTHR
n.a.
notes:
8
© Copyright 2013 by the National Electrical Manufacturers Association and
the Healthcare Information and Management Systems Society.
MAGNETOM
MR
Aera
Avantofit
Prisma/Prismafit
Skyra/Skyrafit
Product & Solution Security Whitepaper
Foreword
The Siemens Healthineers product and solution Performing static code analysis of medical
security program device software
At Siemens Healthineers, we are committed to Conducting security testing of medical devices
working with you to address cybersecurity and under development as well as medical
privacy requirements. Our Product and Solution devices already in the field
Security Office is responsible for our global Tailoring patch management to the medical
program that focuses on addressing device and depth of coverage chosen by you
cybersecurity throughout the product lifecycle Monitoring security vulnerability to track
of our medical devices. reported third party components issues in our
medical devices
Our program targets incorporating state of the Working with suppliers to address security
art cybersecurity in our current and future throughout the supply chain
products. We seek to protect the security of your Training of employees to provide knowledge
data while, at the same time, providing consistent with their level of responsibilities
measures to strengthen the resiliency of our regarding your data and device integrity.
products from external cybersecurity attackers.
Contacting Siemens Healthineers about
We comply with applicable security and privacy product and solution security
regulations from the US Department of Health Siemens Healthineers requests that any
and Human Services (HHS), including the Food cybersecurity or privacy incidents are reported
and Drug Administration (FDA) and Office for by email to:
Civil Rights (OCR), to help you meet your IT productsecurity@siemens-healthineers.com
security and privacy obligations.
For all other communication with Siemens
Vulnerability and incident management Healthineers about product and solution
Siemens Healthineers cooperates with security:
government agencies and cybersecurity ProductTechnologyAssurance.dl@siemens-
researchers concerning reported potential healthineers.com
vulnerabilities.
3
Network Information
The following image shows a high-level overview of the MR system and its integration into the hospital environment:
Equipment
Magnet Room
Room
Connection to
Siemens Remote machines physically not
MaRS accessible
Service
(OPTIONAL) located in a control cabinet
(technic room)
no USB
no PHI
pixels only
optional screen
CAN24 hidden network
in acquisition room no routing to Customer
network
Control Room
video
connection Acquisition Application
Shared Data Postprocessing Apps
Base Common syngo platform
PHI only on MRAW
syngo.via
MRWP (OPTIONAL) MRAW client syngo.via
Satellite Console Main Console server
DICOM DICOM
Customer Network
DICOM
The basic MR system is located in three different but adjacent rooms: The examination room, where the patient is
examined, the operator room, where the medical operators work, and the equipment room, where most of the control
hardware is installed.
Communication lines shown in red are Ethernet connections (TCP and UDP). Communication lines shown in black use
other internal protocols (CAN, PCIe, or proprietary protocols). The internal Ethernet connections between the MRAWP
and the MARS, and between the CAN24 and the MARS are not accessible from the hospital network.
It is assumed that the machine room is locked and not accessible to patients or non-authorized personnel. For
installations with elevated security requirements, it is recommended to make the operator room inaccessible to patients
and non-authorized persons as well.
4
Optional components:
The hospital network and all components outside the MR systems are assumed to be under control of the
customer. Siemens makes no prescriptions about how to structure the network.
The MRWP is an optional workstation for image inspection and post-processing. It could even be located in a
separate room. Information in the patient and image database located on the MRAWP is shared between the
MRAWP and the MRWP, i.e. sensitive information is transferred over the hospital network. For systems with
elevated security requirements it is recommended to either not using the MRWP, or to make sure that the
network communication between the MRAWP and the MRWP is secured.
Siemens remote service is an optional feature. Without remote service, the router to the Siemens service center is
not needed.
The CAN24 box is used to monitor the magnet status by remote service, even if the other system parts are
switched off. Without remote service, this cannot be done. The CAN24 is not optional; however, the connection
to the hospital network can be omitted.
Note: A network connection (e.g. hospital network) is not mandatory to operate MAGNETOM systems.
The MRWP requires one static IP addresses, which must be provided by the customer. The following ports can be used by
the system (depending on the configuration and installed options).
5
Security Controls
Network controls
The system is designed to make limited use of network
ports and protocols. Only ports and protocols which are
required for operation are open and utilized
Microsoft® Windows firewall is configured to block
unwanted inbound network traffic except for above
mentioned ports
Siemens Healthineers recommends operating the
system in a secured network environment, e.g. a
separate network segmented or a VLAN.
Connection to the Internet or private networks for
patients/guests is not recommended
In case of a denial of service (DoS) or malware attack,
the system can be taken off the network and operated
standalone
6
Software Bill of Materials
The following table lists comprises the most relevant third party technologies used.
7
Manufacturer Disclosure Statement
according to IEC60601-1
1-1. The device is connected via Ethernet cable to the hospital using a TCP/IP network with 1Gb/s.
• If the network is down the network services (see below) is not available which can lead to the risks
stated below.
• If the network is unavailable, medical images can not be transferred for remote consultation.
• If the recommended network performance (1Gbit/s) is not provided, the transfer of images is extended
and availability of images at destinations (e.g. for consulting) is delayed.
• Only the protocols shown in the table of used ports are needed for communication.
1-2. PACS system for archiving Images/Results
• If the PACS is not available, images can not be archived after the examination. In case of a system
hardware failure all not archived images can be lost.
• If the PACS is not available, images can not be archived after the examination. Examinations may be no
longer possible because the hard disk is full (because non-archived images can not be automatically
removed).
• If the PACS is not available, images can not be archived after the examination. In case of manual
deletion of images, not archived images can be lost.
• If the PACS is not available, images are not available for remote consultation via PACS consoles.
• If the PACS is not available, prior images are not available.
• If the recommended network performance (1Gbit/s) is not provided, the transfer time to PACS is
extended and the time to wait before switching off the System concecutive to last transfer operations is
prolonged.
1-3. DICOM Printer
• If the DICOM printer is not available, film is not available for diagnose/archive.
1-4. RIS System
• If the RIS system is not available, modality worklist is not available. This can lead to data
inconsistencies.
• If the RIS system is not available, modality worklist is not available. This can lead to images sent to the
PACS are possibly not available until manually coerced with the RIS data in the PACS.
• In case a reduced network performance shall be compensated by a long Worklist Query time-out, this
results in the possibility that non-actual RIS data are used when registering a patient from the list of
schedules on the System.
1-5. Network Connection to Siemens Remote Service Server
• If the connection to the Siemens Remote Service Server is not available, Virus pattern and SW patches
can not be distributed.
• If the connection to the Siemens Remote Service Server is not available, SIEMENS Support is restricted.
1-6. Common Medical protocol properties.
• Protocols used in medical environment are typically unsecure.
• Exceptions are Secure DICOM and Siemens Remote Service (using https).
8
2) Instructions for the responsible organization
2-1. Connection of the system to a NETWORK / DATA COUPLING that includes other equipment could result in previously
unidentified risks to patients operators or third parties. The RESPONSIBLE ORGANIZATION should identify, evaluate
and control these risks
2-2. Subsequent changes to the NETWORK / DATA COUPLING could introduce new RISKS and require additional analysis.
2-3. Changes to the network include:
• changes in NETWORK / DATA COUPLING configuration;
• connection to additional items to the NETWORK / DATA COUPLING;
• disconnecting items from the NETWORK / DATA COUPLING;
• update of equipment connected to the NETWORK / DATA COUPLING;
• upgrade of equipment connected to the NETWORK / DATA COUPLING;
2-4. The RESPONSIBLE ORGANIZATION is fully responsible for the security of the network the device is connected to.
2-5. The RESPONSIBLE ORGANIZATION is fully responsible to ensure staff with access to the device do not have the
possibility to provide any harm to the system.
2-6. The RESPONSIBLE ORGANIZATION has to ensure that the internal network can not be accessed phsysically by non
authorized persons.
2-7. Stuff of the RESPONSIBLE ORGANIZATION has to be trained in security. The RESPONSIBLE ORGANIZATION is
responsible to do this.
2-8. The RESPONSIBLE ORGANIZATION is fully responsible that only authorized medical/administrative staff shall have
access to the device.
2-9. The RESPONSIBLE ORGANIZATION is fully responsible that visitors/patients do not have unsupervised physical
access to the system.
2-10. The RESPONSIBLE ORGANIZATION shall provide access to the system for device administrators and device service
engineers
2-11. The RESPONSIBLE ORGANIZATION shall ensure that neither access from the public internet nor the public intranet to
the device is possible.
2-12. The RESPONSIBLE ORGANIZATION is responsible to ensure physical security for the device.
2-13. The RESPONSIBLE ORGANIZATION shall ensure that access to services of the device from other equipment is possible
only on a need to do basis. An adequate network topology with apropriat firewall settings shall be used.
2-14. The RESPONSIBLE ORGANIZATION is responsible for a secure infrastructure that makes it impossible to change,
prevent, tamper data in transit in any way.
2-15. RECOMMENDATION: It is higly recommende that the RESPONSIBLE ORGANIZATION monitors the network for
unusual traffic
3) Risks and hazardous situations
9
Disclaimer according to IEC 80001-1
1-1 The Device has the capability to be connected to a medical IT-network which is managed under full responsibility of
the operating responsible organization. It is assumed that the responsible organization assigns a Medical IT-Network Risk
Manager to perform IT-Risk Management (see IEC 80001- 1:2010/EN 80001-1:2011) for IT-networks incorporating medical
devices.
1-2 This statement describes Device-specific IT-networking safety and security capabilities. It is not a responsibility
agreement according to IEC 80001-1:2010/EN 80001-1:2011.
1-3 Any modification of the platform, the software or the interfaces of the Device - unless authorized and approved by
Siemens Healthcare GmbH Healthcare - voids all warranties, liabilities, assertions and contracts.
1-4 The responsible organization acknowledges that the Device’s underlying standard computer with operating system is
to some extent vulnerable to typical attacks like e.g. malware or denial-of-service.
1-5 Unintended consequences (like e.g. misuse/loss/corruption) of data not under control of the Device e.g. after electronic
communication from the Device to some IT-network or to some storage, are under the responsibility of the responsible
organization.
1-6 Unauthorized use of the external connections or storage media of the Device can cause hazards regarding the
availability and information security of all components of the medical IT-network. The responsible organization must
ensure – through technical and/or organizational measures - that only authorized use of the external connections and
storage media is permitted.
10
Statement on FDA Cybersecurity
Guidance
Siemens Healthineers will follow cybersecurity guidance issued by the FDA as appropriate. Siemens Healthineers
recognizes the principle described in FDA cybersecurity guidance that an effective cybersecurity framework is a shared
responsibility among multiple stakeholders (e.g., medical device manufacturers, health care facilities, patients and
providers), and is committed to drawing on its innovation, engineering and pioneering skills in collective efforts designed
to prevent, detect and respond to new and emerging cybersecurity threats. While FDA cybersecurity guidance is
informative as to adopting a risk-based approach to addressing potential patient harm, it is not binding and alternative
approaches may be used to satisfy FDA regulatory requirements.
The representations contained in this whitepaper are designed to describe Siemens Healthineers’ approach to
cybersecurity of its medical devices and to disclose the security capabilities of the devices/systems described herein.
Neither Siemens Healthineers nor any medical device manufacturer can warrant that its systems will be invulnerable to
cyberattack. Siemens Healthineers makes no representation or warranty that its cybersecurity efforts will ensure that its
medical devices/systems will be error-free or secure against cyberattack.
11
12
Register
To help you organize your own documentation, the following
registers are provided:
Certificates B
Upgrades D
Room layout E
Declaration of conformity G
... H
... I
... J
Register