Operating Instructions Sirona C8
Operating Instructions Sirona C8
léÉê~íáåÖ=fåëíêìÅíáçåë=
båÖäáëÜ
Sirona Dental Systems GmbH Table of contents
Operating Instructions C8+, C8+ Turn
Table of contents
båÖäáëÜ
1 Notes on these operating instructions................................................................. 6
1.1 General information on the Operating Instructions .................................................................... 6
1.2 Explanation of the symbols used ............................................................................................... 7
3 Maintenance............................................................................................................ 20
3.1 Care and cleaning instructions for the practice team................................................................. 20
3.2 Inspection and maintenance...................................................................................................... 20
3.3 Safety checks ............................................................................................................................ 21
båÖäáëÜ
4 Unit description ...................................................................................................... 22
4.1 Standards/approvals.................................................................................................................. 22
4.2 Technical data ........................................................................................................................... 23
5 System versions..................................................................................................... 24
7 Patient chair............................................................................................................ 29
7.1 Patient positioning ..................................................................................................................... 29
7.2 Adjusting the armrest................................................................................................................. 30
cêÉåÅÜ
7.3 Flat headrest/articulated headrest ............................................................................................. 30
7.4 MultiMotion headrest ................................................................................................................. 31
7.5 Changing the patient position by turning (C8+ Turn only).......................................................... 34
7.6 Chair programs – factory settings.............................................................................................. 35
7.7 Chair programs – reprogramming.............................................................................................. 36
7.8 Foot switch – reprogramming .................................................................................................... 37
7.9 Emergency stop function for chair movement ........................................................................... 38
7.10 Patient chair, options ................................................................................................................. 39
8 Dentist element....................................................................................................... 40
8.1 Locking the dentist element ....................................................................................................... 40
8.2 Adjusting the instrument position............................................................................................... 41
8.3 Swivel arm ................................................................................................................................. 41
pé~åáëÜ
59 42 896 D3417
D3417.201.01.22.02 3
Table of contents Sirona Dental Systems GmbH
Operating Instructions C8+, C8+ Turn
10 Assistant element................................................................................................... 60
10.1 Suction handpiece/saliva ejector .............................................................................................. 60
12 Water unit................................................................................................................ 63
12.1 Cuspidor flushing time - factory settings ................................................................................... 63
12.2 Cuspidor flushing time - reprogramming ................................................................................... 64
12.3 Tumbler filling - factory settings ................................................................................................ 65
12.4 Tumbler filling - reprogramming ................................................................................................ 65
12.5 Combination of rinsing position S
with tumbler filling and flushing controlled by dentist element .................................................. 66
12.6 Assistant element control panel - cuspidor flushing .................................................................. 68
12.7 Assistant element control panel - cuspidor flushing - reprogramming ...................................... 68
12.8 Assistant element control panel - tumbler filling........................................................................ 69
12.9 Assistant element control panel -
combination of mouth rinsing position S with tumbler filling and cuspidor flushing................... 70
12.10 Fresh water bottle ..................................................................................................................... 71
12.11 Switching between tap water/fresh water C8+ .......................................................................... 72
12.12 Hydrocolloid connection ............................................................................................................ 73
12.13 Dürr wet suction system............................................................................................................ 74
12.14 Amalgam separator................................................................................................................... 75
12.15 Disinfection system ................................................................................................................... 76
12.16 Conversion of water unit right-handed left-handed version.................................................. 77
59 42 896 D3417
4 D3417.201.01.22.02
Sirona Dental Systems GmbH Table of contents
Operating Instructions C8+, C8+ Turn
båÖäáëÜ
14.3 Dentist element.......................................................................................................................... 88
14.4 Assistant element ...................................................................................................................... 91
14.5 Water unit .................................................................................................................................. 92
14.6 Sanitation of the treatment center.............................................................................................. 98
14.7 SIROLUX E operating light ........................................................................................................ 104
14.8 SIROLUX FANTASTIC operating light ...................................................................................... 106
14.9 LEDview..................................................................................................................................... 107
15 Options.................................................................................................................... 108
15.1 Working stool CARL .................................................................................................................. 108
15.2 PAUL working stool (not available in the Federal Republic of Germany) .................................. 108
15.3 HUGO working stool .................................................................................................................. 108
båÖäáëÜ
cêÉåÅÜ
pé~åáëÜ
59 42 896 D3417
D3417.201.01.22.02 5
1 Notes on these operating instructions Sirona Dental Systems GmbH
1.1 General information on the Operating Instructions Operating Instructions C8+, C8+ Turn
Observe the Operating Instructions Please familiarize yourself with the unit by reading through these Operat-
ing Instructions before putting it into operation. It is essential that you
comply with the specified warning and safety information.
Keep documents safe Always keep the Operating Instructions handy in case you or another
user require(s) information at a later point in time. Save the Operating
Instructions on the PC or print them out.
If you sell the unit, make sure that the Operating Instructions are included
with it either as a hard copy or on an electronic storage device so that the
new owner can familiarize himself with its functions and the specified
warning and safety information.
Online portal for technical documents We have set up an online portal for Technical Documents at
http://www.sirona.com/manuals. From here, you can download these
Operating Instructions along with other documents. Please complete the
online form if you would like a hard copy of a particular document. We will
then be happy to send you a printed copy free of charge.
Help If you reach an impasse despite having thoroughly studied the Operating
Instructions, please contact your dental depot.
59 42 896 D3417
6 D3417.201.01.22.02
Sirona Dental Systems GmbH 1 Notes on these operating instructions
Operating Instructions C8+, C8+ Turn 1.2 Explanation of the symbols used
båÖäáëÜ
ON Patient chair up
T 10A H
Individual programming,
lower jaw treatment 2
59 42 896 D3417
D3417.201.01.22.02 7
1 Notes on these operating instructions Sirona Dental Systems GmbH
1.2 Explanation of the symbols used Operating Instructions C8+, C8+ Turn
>10s
<10m
1
80
60
ENDO 40
20
1
80
60
ENDO 40
20
59 42 896 D3417
8 D3417.201.01.22.02
Sirona Dental Systems GmbH 1 Notes on these operating instructions
Operating Instructions C8+, C8+ Turn 1.2 Explanation of the symbols used
båÖäáëÜ
Heat dissipation or
Reprogramming
59 42 896 D3417
D3417.201.01.22.02 9
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety information Operating Instructions C8+, C8+ Turn
Intended use This dental treatment center is designed to be used in the field of dentistry
and must only be operated by qualified dental personnel.
Contraindications for the use of the treatment center, if any, are listed in the
individual sections, e.g. Treatment instruments.
This unit is not intended for operation in areas subject to explosion hazards.
This unit is permanently installed. Operation is not permitted in mobile vehi-
cles.
Intended use also includes compliance with these Operating Instructions and
the relevant maintenance instructions.
On-site installation The on-site installation must have been performed according to our require-
ments. The details are described in the document "Installation Require-
ments".
Installing the treatment center Installation must be carried out by authorized technicians in accordance with
the installation instructions.
Media quality The air and water supplies must meet the requirements specified in the
installation requirements.
As the owner of the treatment center, you are generally responsible for
the water quality.
The colony count must fulfill the national regulations for drinking water and
must not exceed 500 CFU/ml under any circumstances (CFU: colony forming
unit).
In the event of an increased colony count, the indoor installation must be
checked and the cause of contamination eliminated. Alternatively, a stand-
alone water supply can also be installed. In addition, the optional fresh water
vessel can be used.
Before installation of the treatment center, an acceptable microbiological
water quality for the domestic water supply must be ensured and docu-
mented in the form of a colony count. Sampling and the colony count must
be performed by a competent laboratory.
59 42 896 D3417
10 D3417.201.01.22.02
Sirona Dental Systems GmbH 2 Warning and safety information
Operating Instructions C8+, C8+ Turn 2.1 General safety information
Check the quality of the water from the treatment center at regular intervals
and after periods of disuse of over > 1 week, see "Microbiological water test"
on page 97. Please contact your specialized dealer or your relevant dental
association for the respective national requirements and measures.
båÖäáëÜ
Highly immunosuppressed patients should not come in contact with water
from the treatment center. The use of sterile solutions is recommended.
Connection to the public drinking water Treatment center with isolation from public drinking water supply
supply
Provided it is equipped with a disinfection system, the treatment center fulfills
the requirements of EN 1717 (free outlet with sections ≥ 20 mm and the Ger-
man Gas and Water Association (DVGW). It is intrinsically safe in accor-
dance with worksheet W540 and therefore also fulfills the W270
requirements as well as the requirements for plastics used in the transport of
drinking water. It can be connected directly to the public drinking water sup-
ply.
Provided it is equipped with a fresh water vessel without switchover to the
public drinking water supply then the treatment center fulfills the require-
ments of both EN 1717 and those of the German Gas and Water Association
(DVGW).
The treatment center therefore also bears the symbol for the German Gas
and Water Association (DVGW) along with the rating plate on the chair base.
The DVGW approval of the treatment center expires if a media block is
installed in the base of the chair, see page 39 and 73.
59 42 896 D3417
D3417.201.01.22.02 11
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety information Operating Instructions C8+, C8+ Turn
Trouble-free operation Use of this unit is permissible only if it works properly without malfunctions. If
trouble-free operation cannot be ensured, the unit must be taken out of ser-
vice, checked by authorized technicians for malfunctions and, if necessary,
repaired.
Vacuum system The suction removal of aluminum and other metal oxides from blasting
devices via the amalgam separator and the automatic separator installed in
the treatment center is prohibited! This would cause extreme wear and clog-
ging of the suction and water paths.
A separate vacuum system must be used in connection with metal oxide
blasting devices. Treatment centers equipped with a central wet suction sys-
tem are generally suitable for suction removal of the above material. How-
ever, make sure to observe the instructions provided by the manufacturer of
your vacuum system.
No restrictions apply when using salt blasting devices in connection with
Sirona treatment centers. However, in such cases, make sure that the system
is subsequently flushed with an adequate amount of water.
Intermittent operation The motors of the treatment center and of the treatment instruments are
designed for intermittent operation corresponding to the dental mode of
treatment.
Drive motors for patient chair and backrest: max. 6% duty cycle (max. 15 s
"ON"/3 min, 55 s "OFF")
Care, cleaning agents, and disinfec- Unsuitable care and cleaning agents or disinfectants may corrode the sur-
tants face of the unit or impair its functioning.
Therefore, use only care and cleaning agents and disinfectants which have
been approved by Sirona. For further information, see "Care, cleaning
agents, and disinfectants" on page 84.
59 42 896 D3417
12 D3417.201.01.22.02
Sirona Dental Systems GmbH 2 Warning and safety information
Operating Instructions C8+, C8+ Turn 2.1 General safety information
Ventilation slots Under no circumstances may the ventilation slots on the unit be covered,
since otherwise the air circulation will be obstructed. This can cause the unit
to overheat.
båÖäáëÜ
Do not spray liquids such as disinfectants into the ventilation slots. This may
lead to malfunctions. Use only wipe disinfection in these areas.
Modifications and extensions to the Modifications to this unit which might affect the safety of the system owner,
unit patients or other persons are prohibited by law.
For reasons of product safety, this product may be operated only with original
Sirona accessories or third-party accessories expressly approved by Sirona.
The user assumes the risk of using non-approved accessories.
If any units not approved by Sirona are connected, they must comply with the
applicable standards, e.g.:
• IEC 60950 for information technology equipment (e.g. PC) and
• IEC 60601-1 for medical electrical equipment.
The treatment center monitor must fulfill the requirements of the IEC 60950
standard.
The loudspeaker port of the monitor may be connected only to a unit that
complies with IEC 60950 (e.g. a PC) or IEC 60601-1. Under no circum-
stances may it be connected e.g. to a stereo system, etc..
If a system is created during the installation process, the requirements of IEC
60601-1, 3rd edition, must be fulfilled. The system author assumes responsi-
bility for conformity of the system with the
93/42/EEC Directive.
Electromagnetic compatibility (EMC) The C8+ and C8+ Turn comply with the requirements of IEC 60601-1-2.
Medical electrical equipment is subject to special precautionary measures
with regard to electromagnetic compatibility (EMC). They must be installed
and operated as specified in the document "Installation Requirements".
Portable and mobile RF communications equipment may interfere with medi-
cal electrical equipment. Therefore, the use of such devices (e.g. cellular
phones) in practice or hospital environments must be prohibited.
Operating an RF surgical device
Treatment with HF surgical devices creates strong electromagnetic fields,
which may impact other electronic devices. Do not place external RF surgical
devices on the shelf space of the treatment center and do not guide the RF
handpiece cable over this. Electromagnetic disturbances can often be
reduced, if the external RF surgical unit is operated with a neutral electrode.
Video system
Electromagnetic disturbances in the environment of the treatment center
may result in reduced image quality. In such cases, it is advisable to repeat
the exposure before saving the image.
When a line drop occurs, the image data stored in the video camera may be
lost. Deposit the camera and repeat the exposure.
Image artifacts may possibly occur as a result of radio-frequency interfer-
ence caused by the transmitters of e.g. radio services or radio amateurs.
Influencing of cardiac pacemakers The treatment center contains strong magnets. These magnets can affect
cardiac pacemakers implanted in patients, users and technical personnel.
Therefore, please observe the safety information on page 30.
59 42 896 D3417
D3417.201.01.22.02 13
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety information Operating Instructions C8+, C8+ Turn
Dismantling / Installation When dismantling and reassembling the unit, proceed according to the
installation instructions for new installation in order to guarantee its function-
ing and stability.
Disposal Your product is marked with the adjacent symbol. Within the European Eco-
nomic Area, this product is subject to Directive 2002/96/EC as well as the
corresponding national laws. This directive requires environmentally sound
recycling / disposal of the product. The product must not be disposed of as
domestic refuse!
Please observe the disposal regulations applicable in your country.
Disposal procedure
Please note that this product is subject to the stipulations in EC Directive
2002/96 governing waste electrical and electronic equipment and must be
disposed of in line with these special requirements within the European
Union (EU).
Prior to disassembly / disposal of the product, it must be fully prepared
(cleaned / disinfected / sterilized).
When disposing of equipment permanently, please proceed as follows:
In Germany:
To initiate return of the electrical device, please send a disposal order to
"enretec GmbH".
1. You can find a form for placing a disposal order on the company's homep-
age at www.enretec.de under the menu item "Entsorgung elektrischer
und elektronischer Geräte" (Disposal of electrical and electronic devices).
The form can either be downloaded or filled out online.
2. Fill out the form with the corresponding details and send it either as an on-
line order or fax it to enretec GmbH at +49(0)3304 3919 590.
You can also get in touch with the following contacts for disposal orders
and any questions relating to this you may have:
Phone: +49(0)3304 3919 500;
E-mail: pickup@eomRECYCLING.com
Post: enretec GmbH, Geschäftsbereich eomRECYCLING
Kanalstrasse 17, 16727 Velten
Your equipment that is not permanently installed will be collected in the prac-
tice, while your permanently installed equipment will be collected curbside at
your address at the agreed time and date.
All disassembly, transport and packaging costs are to be borne by the owner
/ operator of the equipment. The disposal itself is free of charge.
Worldwide (outside Germany):
Please contact your local dental equipment specialist for country-specific
information on disposal.
59 42 896 D3417
14 D3417.201.01.22.02
Sirona Dental Systems GmbH 2 Warning and safety information
Operating Instructions C8+, C8+ Turn 2.1 General safety information
båÖäáëÜ
Please observe the patient chair's maxi-
mum load capacity of 135 kg.
The weight distribution complies with
ISO 6875. The mechanical stability test is
performed with a multiple safety factor
according to IEC 60601-1.
The maximum permissible weight of
accessories mounted on the patient chair
is 5 kg.
STOP STOP
2.
CAUTION
Make sure that no obstacles (e.g. window
wings, drawers, devices, …) extend into
the movement range.
59 42 896 D3417
D3417.201.01.22.02 15
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety information Operating Instructions C8+, C8+ Turn
STOP
STOP
CAUTION
When moving the chair (via program or
manually), make sure that the patient
does not under any circumstances come
into contact with the support arm or the
dentist element.
59 42 896 D3417
16 D3417.201.01.22.02
Sirona Dental Systems GmbH 2 Warning and safety information
Operating Instructions C8+, C8+ Turn 2.1 General safety information
båÖäáëÜ
In accordance with IEC 62471, they are
classified in risk group 2 with the light bar
removed.
CAUTION
right! wrong! right! wrong! Do not stare into the beam path for longer
periods, as this can cause eye damage.
ATTENTION! Potentially dangerous
optical radiation! Do not under any circumstances stare into
the beam path when the light bar is re-
iKbKaK=ÄÉ~ã moved.
Éñáí=~éÉêíìêÉ
59 42 896 D3417
D3417.201.01.22.02 17
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety information Operating Instructions C8+, C8+ Turn
59 42 896 D3417
18 D3417.201.01.22.02
Sirona Dental Systems GmbH 2 Warning and safety information
Operating Instructions C8+, C8+ Turn 2.1 General safety information
SIROSONIC L
wrong ! right !
båÖäáëÜ
right SIROLUX E/SIROLUX FANTASTIC
wrong
SIROLUX E
CAUTION
Never operate without cover:
Watch out for overheating of lamp compo-
nents and let them cool down.
SIROLUX FANTASTIC
CAUTION
Pay attention to elevated temperature of
the lamp components; and let these cool
down.
CAUTION
Use only Sirona fresh water bot-
tles!
N
MO
LEMON
LE
wrong
59 42 896 D3417
D3417.201.01.22.02 19
3 Maintenance Sirona Dental Systems GmbH
3.1 Care and cleaning instructions for the practice team Operating Instructions C8+, C8+ Turn
3 Maintenance
59 42 896 D3417
20 D3417.201.01.22.02
Sirona Dental Systems GmbH 3 Maintenance
Operating Instructions C8+, C8+ Turn 3.3 Safety checks
båÖäáëÜ
way that the first occurrence of a fault
does not create a hazard to the safety of
patients, users or other persons. Hence it
is important to detect such faults before a
second fault occurs, which might then
lead to safety hazards.
For that reason it is essential to perform
safety tests every 2 years which aim par-
ticularly at detecting electrical faults (e.g.
isolation defects). This is done by an
authorized service engineer from your
dental depot, most practically together
with the work to be performed according
to section 3.2.
The check includes a visual inspection as
well as measurements of the protective
ground wire connections and the equiva-
lent leakage currents.
The inspections and measurements to be
performed are specified in the "Mainte-
nance Manual". The measured values
must be documented there by the service
engineer.
Safety tests must also be performed and
documented during initial startup, after
extensions/upgrades (conversion) of
your treatment center and after repair
work which might affect the electrical
safety of the system.
i NOTE
The treatment center must not be operat-
ed if it has failed to pass the safety tests!
59 42 896 D3417
D3417.201.01.22.02 21
4 Unit description Sirona Dental Systems GmbH
4.1 Standards/approvals Operating Instructions C8+, C8+ Turn
4 Unit description
4.1 Standards/approvals
The treatment center complies with the following standards:
• IEC 60601-1 (electrical, mechanical, and software safety)
• IEC 60601-1-2 (electromagnetic compatibility)
• IEC 60601-1-6 (serviceability)
• ISO 6875 (Patient chair)
• ISO 7494-1 (Dental treatment units)
• ISO 7492-2 (dental treatment units, water and air supply)
• ISO 9680 (Operating light)
• ISO 11143 (amalgam separator), see also below (provided amalgam
separator option is available)
• EN 1717 (connection to the drinking water system),
see also below (DVGW)
Original language of the present document: German
This product bears the CE mark in accordance with the provisions of the
Council Directive 93/42/EEC of June 14, 1993 concerning medical devices
(MDD).
0123
The treatment center fulfills the requirements of the directive on the restric-
tion of hazardous substances, 2011/65/EU.
SIRONA
Dental Systems The amalgam separator achieves a separation efficiency of > 95%. The
GmbH
64625 Bensheim
Fabrikstrasse 31
requirements of the ISO 11143 standard are thereby fulfilled.
Z-64.1-24 Separating procedure type 1: Centrifugal system
The amalgam separator is approved by the German Institute for Structural
Engineering (DIBT) and by the AFNOR (France).
This unit meets the requirements of BELGAQUA and may therefore be con-
nected to the public drinking water supply in Belgium.
59 42 896 D3417
22 D3417.201.01.22.02
Sirona Dental Systems GmbH 4 Unit description
Operating Instructions C8+, C8+ Turn 4.2 Technical data
båÖäáëÜ
Model designation:
Power connection: 100 V / 115 V / 127 V / 230 VAC 50 / 60 Hz
Rated current: 11.5 A at 100 V AC
9.5 A at 115 V AC
4.5 A at 230 V AC.
In addition, max. 6 A for external units see page 39
Main unit fuse: T 10 A H, 250 V AC , REF 10 77 460
Protection class: Class I device
Device class in accordance Class IIa device
with Directive 93/42/EEC:
Degree of protection against elec- Type B applied parts
trical shock:
59 42 896 D3417
D3417.201.01.22.02 23
5 System versions Sirona Dental Systems GmbH
Operating Instructions C8+, C8+ Turn
5 System versions
C8+ OTP
1. Patient chair/Patient chair Turn1
2. OTP dentist element
5. 3. Water unit
2. 4. Assistant element
3. 5. Operating light
4.
6. Instrument foot switch with program
1.
positions
0
S
P
C8 +
S
0
1
2
6.
P1
P2
7.
59 42 896 D3417
24 D3417.201.01.22.02
Sirona Dental Systems GmbH 5 System versions
Operating Instructions C8+, C8+ Turn
båÖäáëÜ
2. Dentist element with swivel arms
5. 3. Water unit
2. 4. Assistant element
3. 5. Operating light
4.
1. 6. Instrument foot switch with program
positions
2
0
1
S
P
C8 +
S
0
1
2
6.
P1
P2
7.
C8+ Cart
1. Patient chair/Patient chair Turn1
2. Dentist element
5. 3. Water unit
4. Assistant element
3. 5. Operating light
4. 2.
1. 6. Instrument foot switch with program
positions
7. Instrument foot switch
6.
7.
59 42 896 D3417
D3417.201.01.22.02 25
5 System versions Sirona Dental Systems GmbH
Operating Instructions C8+, C8+ Turn
C8+ Turn
1. Patient chair Turn
2. OTP dentist element/Dentist
5.
element with swivel arms1
2. Cart dentist element1
3. 3. Water unit
4.
1. 4. Assistant element
5. Operating light
C8 +
2
1
100
80
60
ENDO 40
20
1
P
positions
S
0
1
2
P1
P2
8.
7.
i NOTE
For operation of the SIVISION 3 video
system, please refer to Operating Instruc-
tions 59 67 893.
59 42 896 D3417
26 D3417.201.01.22.02
Sirona Dental Systems GmbH 5 System versions
Operating Instructions C8+, C8+ Turn
båÖäáëÜ
The BL ISO C option is not available in all
countries.
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
1 2 3 4 5
SPRAYVIT
SPRAYVIT Highspeed
Turbine Highspeed
Turbine Highspeed
Turbine SIROSONICL L
SIROCONIC
handpiece handpiece handpiece
oderor/ or or / or
oder / or
oderor or/ or
oder / or
oderor
Option: ou / o ou / o ou / o ou / o ou / o
Option
SL motor
Motor SL SIROCAM 3/C
SIROCAM 3/C
SL motor
Motor SL SL motor
Motor SL Mini LED curing
Polymerisations-
light
oderor/ or licht Mini L.E.D.
ou / o
oderor/ or or/ or
oder / or
oderor
Option: ou / o ou / o ou / o
Option
BL ISO C SIROCAM 3/C
Motor SIROCAM 3/C
motor
BL ISO
Mini LED curing
Polymerisations-
light
oderor/ or licht Mini L.E.D.
ou / o
59 42 896 D3417
D3417.201.01.22.02 27
6 Putting the system into operation Sirona Dental Systems GmbH
6.1 System ON/OFF switch Operating Instructions C8+, C8+ Turn
100
80
60
ENDO 40
20
1
P
S
0
1
2
59 42 896 D3417
28 D3417.201.01.22.02
Sirona Dental Systems GmbH 7 Patient chair
Operating Instructions C8+, C8+ Turn 7.1 Patient positioning
7 Patient chair
7.1
båÖäáëÜ
Patient positioning
C8 +
S 0
100
80
60
ENDO 40
P 1
20
P C8 +
0
S
P
Option
2 S 0
1 S
0 2
1
1
2
C8 +
1 2
0
1
S
100
2
80
60
ENDO 40
20
1
P
S
0
1
2
Option
S 0
S
0
0
Option
S
2
1 1 2 1
2
P P
59 42 896 D3417
D3417.201.01.22.02 29
7 Patient chair Sirona Dental Systems GmbH
7.2 Adjusting the armrest Operating Instructions C8+, C8+ Turn
B B Wide backrest
C
C Narrow backrest
CAUTION
Risk of collision!
The following versions must not be retrofit-
ted with an armrest:
• C8+ with fixed water unit in combination
with long cuspidor
A Flat headrest
WARNING
The head pad (A) contains a strong
B magnet (1) on its bottom side. This
magnet could affect any cardiac pace-
maker located nearby.
59 42 896 D3417
30 D3417.201.01.22.02
Sirona Dental Systems GmbH 7 Patient chair
Operating Instructions C8+, C8+ Turn 7.4 MultiMotion headrest
båÖäáëÜ
The headrest can be adjusted to the
patient's height by pulling out or pushing
in the headrest extension. This adjust-
ment is usually required at the beginning
of treatment!
i NOTE
Before positioning the patient on the
headrest, make sure that the head sup-
port has been optimally adjusted to the
patient's height. This simplifies all subse-
quent work with the MultiMotion consider-
ably, since readjustment to the patient's
height can thus be omitted when changing
over from mandibular to maxillary treat-
ment.
Adjustment of hyperextension
Mandibular position
A The mandibular position can be set by
pulling the operating handle (A). The ana-
tomical movement of the arched exten-
sion keeps the patient's head in the sup-
port.
Pull the headrest out of the guide by
pulling handle A.
i NOTE
You can reduce the adjustment noise by
pressing release A.
59 42 896 D3417
D3417.201.01.22.02 31
7 Patient chair Sirona Dental Systems GmbH
7.4 MultiMotion headrest Operating Instructions C8+, C8+ Turn
Maxillary position
A Take the load off the headrest by lift-
ing it slightly.
Press release A in the key recess
area on the handle.
Let the headrest slide down slowly.
Let go of the release when the head-
rest reaches the desired position.
59 42 896 D3417
32 D3417.201.01.22.02
Sirona Dental Systems GmbH 7 Patient chair
Operating Instructions C8+, C8+ Turn 7.4 MultiMotion headrest
båÖäáëÜ
tion completely in order to obtain better
access to the patient. The patient's head
will then be supported on the contact sur-
C face of the headrest extensionF.
To remove the MultiMotion, proceed as
follows:
1. Press release button C.
D 2. Pull the complete headrest out of the
E headrest extension.
3. Cover the opening of the headrest ex-
tension with cover cap E.
F 4. If a children's head pad D is available,
place it on the headrest extension.
The head pad is held magnetically.
WARNING
The children's head pad (D) contains a
strong magnet on its bottom side. This
magnet could affect any cardiac pace-
maker located nearby.
Therefore, do not allow patients, users
and technical personnel with a cardiac
pacemaker to be located near this mag-
net.
i NOTE
Store the removed MultiMotion in a safe
place where it cannot fall onto the floor.
59 42 896 D3417
D3417.201.01.22.02 33
7 Patient chair Sirona Dental Systems GmbH
7.5 Changing the patient position by turning (C8+ Turn only) Operating Instructions C8+, C8+ Turn
CAUTION
Risk of collision!
When turning the chair, make sure that it
A does not collide with any other compo-
nents of the treatment center, persons,
furnishings or equipment.
Turn the chair only if all treatment instru-
ments are in place.
S
1. Step on foot pedal A to release the
0 locking brake.
2. Turn the patient chair to the desired
position.
3. The chair locks in place as soon as
you release foot pedal A.
17° 17°
59 42 896 D3417
34 D3417.201.01.22.02
Sirona Dental Systems GmbH 7 Patient chair
Operating Instructions C8+, C8+ Turn 7.6 Chair programs – factory settings
båÖäáëÜ
Last Memory function
S 0 1 2
S
S
1.S
0
2.S
P S
P 0
<3s <3s
1. 2. 1. 2.
C8 + C8 +
100
P
Option
80 0
S
60 2
1
ENDO 40
P 1
20
P
0
S
S
0 1
S
0 Option
1
1 0
P C8 +
S
2 alternatively
1
0
P P
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
P 1
P 2
<3s <3s
1. 2. 1. 2.
1 2
P1
Option
P2
S
1
P1
P2
0 2
59 42 896 D3417
D3417.201.01.22.02 35
7 Patient chair Sirona Dental Systems GmbH
7.7 Chair programs – reprogramming Operating Instructions C8+, C8+ Turn
S 0 1 2
S 0 1 2
C8 +
C8 +
10
60
80
P
Option
ENDO 40
P 1
20 0
S
P
2
1
0
S
S
0
2
1
S
0
Option
1
1
0
P S
P
C8 +
1
2 alternatively
2
0
S P
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
S 0 1 2
100
80
60
ENDO 40 100
20
80
C8 + P 1
ENDO
20
60
40
1
P
0 S
0
S
2
P 1
2 >3s S
0
1
1
P 0
S
2
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
59 42 896 D3417
36 D3417.201.01.22.02
Sirona Dental Systems GmbH 7 Patient chair
Operating Instructions C8+, C8+ Turn 7.8 Foot switch – reprogramming
båÖäáëÜ
P1
Option
P2
>3s
59 42 896 D3417
D3417.201.01.22.02 37
7 Patient chair Sirona Dental Systems GmbH
7.9 Emergency stop function for chair movement Operating Instructions C8+, C8+ Turn
C8 +
P
Option
0
S
2
1
Option
0
Option
S
2
1
P
alternatively
or
59 42 896 D3417
38 D3417.201.01.22.02
Sirona Dental Systems GmbH 7 Patient chair
Operating Instructions C8+, C8+ Turn 7.10 Patient chair, options
båÖäáëÜ
Additional equipment which is approved
for dental use (e g by EMS) and the hydro-
colloid coolant supply (see page 73) may
be connected to a media block on the
base of the chair (observe manufacturers’
specifications for the additional equip-
ment).
DVGW approval:
Pressure Flow rate The DVGW approval of the treatment
center expires if a media block is installed.
min. 2.5 bar
max. 6.0 bar
Water max. 1000 ml/min
(corresponding to input pressure on
treatment center)
59 42 896 D3417
D3417.201.01.22.02 39
8 Dentist element Sirona Dental Systems GmbH
8.1 Locking the dentist element Operating Instructions C8+, C8+ Turn
8 Dentist element
Brake applied
Maximum load
Brake released
CAUTION
A = You must not place any objects on this
dentist element!
Brake applied
A
A
Maximum load
Brake released
Option
59 42 896 D3417
40 D3417.201.01.22.02
Sirona Dental Systems GmbH 8 Dentist element
Operating Instructions C8+, C8+ Turn 8.2 Adjusting the instrument position
båÖäáëÜ
8.3 Swivel arm
1.
OK
3.
2.
Click CAUTION
IMPORTANT: Swivel arms
that are locked in place must
not be pushed backward! To
reset, pull the
instrument hose out again.
NO
59 42 896 D3417
D3417.201.01.22.02 41
8 Dentist element Sirona Dental Systems GmbH
8.4 Dentist element C8+ Cart Operating Instructions C8+, C8+ Turn
CAUTION
Risk of toppling over! Move cart slowly
and only on even floors without any obsta-
cles!
CAUTION
Distribute the additional loads on the den-
tist element evenly!
59 42 896 D3417
42 D3417.201.01.22.02
Sirona Dental Systems GmbH 8 Dentist element
Operating Instructions C8+, C8+ Turn 8.5 Starting an instrument
båÖäáëÜ
A = Unused holders should be sealed
with a ball!
A
NO
CAUTION
Do not operate the instrument without
spray.
Always place the instrument back in the
OK correct holder.
A i NOTE
A Also observe the operating instructions for
OK
the treatment instruments and the video
system.
CAUTION
Use only treatment instruments approved
by Sirona!
NO
59 42 896 D3417
D3417.201.01.22.02 43
8 Dentist element Sirona Dental Systems GmbH
8.6 Adjusting the driving air pressure Operating Instructions C8+, C8+ Turn
! !
Only by a
service engineer
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
alternatively
or
1P
2P
+ _
59 42 896 D3417
44 D3417.201.01.22.02
Sirona Dental Systems GmbH 8 Dentist element
Operating Instructions C8+, C8+ Turn 8.7 Instrument spray ON/OFF – chip blower
båÖäáëÜ
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
A B C
A = Spray OFF
B = Spray ON
C = Chip blower
or alternatively
B A C
i NOTE
For the adjacent foot switch, "Spray ON"
(B) is indicated by a green signal in the up-
per left corner.
P1
P1
P2 P2
59 42 896 D3417
D3417.201.01.22.02 45
8 Dentist element Sirona Dental Systems GmbH
8.8 Instruments – adjusting the spray intensity Operating Instructions C8+, C8+ Turn
! ! 2.
!
1.
1P
2P
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
3.
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
alternatively
or
4.
-
+
5.
1P
2P
_ A +
A = Spray WATER
B = Spray AIR
C8+ Cart
Control choke can be
6. adjusted by a service engi-
neer only
B
59 42 896 D3417
46 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.1 Negatoscope
9.1
båÖäáëÜ
Negatoscope
1. x 2. x
59 42 896 D3417
D3417.201.01.22.02 47
9 Dentist element options Sirona Dental Systems GmbH
9.3 Silicone handles, right/left Operating Instructions C8+, C8+ Turn
9.4 Tray
A
A= Silicone mat, optional
B Maximum load
59 42 896 D3417
48 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.6 3-way SPRAYVIT syringe
båÖäáëÜ
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
3 min
134°C 274°F
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
Click
Check
Option
59 42 896 D3417
D3417.201.01.22.02 49
9 Dentist element options Sirona Dental Systems GmbH
9.7 3-way syringe Operating Instructions C8+, C8+ Turn
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
134°C 274°F
3 min
Check
59 42 896 D3417
50 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.8 Satelec Mini L.E.D. curing light
båÖäáëÜ
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
Operating modes
Mode 1,2,3
CAUTION
Always use the glare shield!
CAUTION
Do not stare into the beam path for longer
periods, as this can damage your eyes.
59 42 896 D3417
D3417.201.01.22.02 51
9 Dentist element options Sirona Dental Systems GmbH
9.8 Satelec Mini L.E.D. curing light Operating Instructions C8+, C8+ Turn
Technical data
General technical data
Model: Mini L.E.D.
Weight of hand- 105 g
piece without
hose:
Dimensions: dia. 23 x 240 mm
Current draw of 5 V DC / 0.65 A
handpiece:
Thermal safety: Overheat protection
Optical specification
Wavelength: 420nm – 480nma
Max. intensity: at 450nm
Light power 1100 mW/cm2
dia.7.5 mm
(standard ver-
sion):
Light power 2,000 mW/cm2
dia.5.5 mm
(optional,
available from
Satelec):
Light power: 450 mW – 500 mW
Classification Risk group 2 with
acc. to IEC light bar removed
62471:
59 42 896 D3417
52 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.9 SL or BL ISO C electric motor
båÖäáëÜ
i NOTE
The BL ISO C option is not available in all
countries.
CAUTION
Damage to BL ISO C electric motor:
The BL ISO C electric motor may be con-
nected only to a treatment center intended
for this motor. The suitability of the treat-
ment center for operation with the BL ISO
BLBL-ISO
ISO C motor
motor
C electric motor is indicated by a label on
C8+
C8+holder3
hanger 3 the dentist element.
BL ISO C motor
Brushless motors are designed as three-
phase motors (without carbon brushes).
BL ISO C They are characterized by precise con-
3
012
59 42 896 D3417
D3417.201.01.22.02 53
9 Dentist element options Sirona Dental Systems GmbH
9.9 SL or BL ISO C electric motor Operating Instructions C8+, C8+ Turn
SL motor
Collector motors are DC motors in which
the voltage supply is realized via carbon
brushes. The speed range lies between
200 and approx. 40,000 rpm (revolutions
per minute).
The green coded instrument hose (GN)
must be used for the SL motor. The SL
GN GN motor can be operated with two different
motor couplings For more information,
please refer to the "SL Motor Operating
Instructions".
59 42 896 D3417
54 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.9 SL or BL ISO C electric motor
båÖäáëÜ
Speed in % LED/display
1 20 40 60 80 100
1-4 X
5 - 14 X X
15 - 24 X
25 - 34 X X
35 - 44 X
45 - 54 X X
55 - 64 X
65 - 74 X X
75 - 84 X
85 - 94 X X
95 - 100 X
Speed +
Speed -
A
alternatively
or
59 42 896 D3417
D3417.201.01.22.02 55
9 Dentist element options Sirona Dental Systems GmbH
9.10 SIROSONIC L scaler Operating Instructions C8+, C8+ Turn
wrong !
Power in % LED/display
1 20 40 60 80 100
1-4 X
5 - 14 X X
100
80
60
ENDO 40
15 - 24 X
20
1
P
S
0
1
2
25 - 34 X X
35 - 44 X
right ! C8 +
P
ENDO
1
20
60
40
100
80
45 - 54
55 - 64
X X
X
1
2
S
0
65 - 74 X X
75 - 84 X
85 - 94 X X
+
95 - 100 X
A 60
100
80
A = Water +/-
ENDO 40
20
1
P
_ 1
2
S
0
B = Preselected maximum intensity
_ -
+ +
- Intensity +
A B
+
+ _ Intensity -
100
80
60 60
100
40 alternatively
1P
2P
20
80 or
1
60
40 60
20
1
100 100
80 80 CAUTION
60 60 The set value is saved when the instru-
40 = 60 40 = 70 ment is placed back in its holder.
20 20
1 1
59 42 896 D3417
56 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.11 SIROSONIC L endodontic scaler
båÖäáëÜ
In endodontics mode, the maximum inten-
sity of the ultrasonic handpiece is
restricted to ensure safe operation.
Power LED/display
1 20 40 60 80 100
Step 1 X
Step 2 X
Step 3 X
Step 4 X
Step 5 X
Step 6 X
A A = Water +/-
A B Intensity +
Intensity -
alternatively
or
CAUTION
The set value is saved when the instru-
ment is placed back in its holder.
59 42 896 D3417
D3417.201.01.22.02 57
9 Dentist element options Sirona Dental Systems GmbH
9.12 Rinsing the water lines Operating Instructions C8+, C8+ Turn
Purge function
To reduce the amount of germs, the water
lines of the water carrying instruments of
the dentist element and the SPRAYVIT M
multifunctional syringe of the assistant
1. element are purged with water.
i NOTE
Pressing and holding the footswitch B
5.
suspends the purge function.
4.
7. Place the rinsed instrument back in
holder 2.
8. Remove the next instrument (holder
3) and hold it above the cuspidor.
After 5 seconds have elapsed, the in-
strument hose water line is rinsed for
30 seconds.
B 9. Repeat the procedure for all instru-
ments.
10. After completing the purge proce-
dure, turn the switch back to the
"OFF" position.
6. 10.
59 42 896 D3417
58 D3417.201.01.22.02
Sirona Dental Systems GmbH 9 Dentist element options
Operating Instructions C8+, C8+ Turn 9.12 Rinsing the water lines
i NOTE
It is possible to rinse all instruments at
once. The purge procedure is then execut-
båÖäáëÜ
ed automatically.
59 42 896 D3417
D3417.201.01.22.02 59
10 Assistant element Sirona Dental Systems GmbH
10.1 Suction handpiece/saliva ejector Operating Instructions C8+, C8+ Turn
10 Assistant element
A
A = Suction handpiece
B
B = Saliva ejector
wrong
59 42 896 D3417
60 D3417.201.01.22.02
Sirona Dental Systems GmbH 11 Assistant element - options
Operating Instructions C8+, C8+ Turn 11.1 3-way syringes
11.1
båÖäáëÜ
3-way syringes
59 42 896 D3417
D3417.201.01.22.02 61
11 Assistant element - options Sirona Dental Systems GmbH
11.4 Satelec Mini L.E.D. curing light Operating Instructions C8+, C8+ Turn
59 42 896 D3417
62 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.1 Cuspidor flushing time - factory settings
12 Water unit
12.1
båÖäáëÜ
Cuspidor flushing time - factory settings
30s
C8 +
C8 +
100
80 P
Option
60 0
ENDO 40
S
P
2
1
20
max.
P 1
0
S 1 kg
2 S
1
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
59 42 896 D3417
D3417.201.01.22.02 63
12 Water unit Sirona Dental Systems GmbH
12.2 Cuspidor flushing time - reprogramming Operating Instructions C8+, C8+ Turn
10s 10 m
1. 2.
C8 +
P
Option
0
S
2
1
2s
alternatively
10s 10 m
4.
3.
5.
C8+
5.
C8+ Turn
59 42 896 D3417
64 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.3 Tumbler filling - factory settings
båÖäáëÜ
4s
C8 + C8 +
P
Option
0
S
100 2
1
80
60
ENDO 40
Pmax. 1
20
0
1 kg P alternatively
S
2 S
1
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
>3s <15s
8 +
1.
C8 +
100
60
80 2s P
Option
ENDO 40
20 0
S
1
P 2
1
alternatively
2.
>3s <15s
8 +
3.
100
80
60
ENDO 40
20
1
P
59 42 896 D3417
D3417.201.01.22.02 65
12 Water unit Sirona Dental Systems GmbH
12.5 Combination of rinsing position S with tumbler filling and flushing controlled by dentist element Operating Instructions C8+, C8+ Turn
+ +
P S
<3s
C8 +
100
80
60
ENDO 40
P 1
20
P
0
S
2 S
1
0 C8 +
1
2 P
Option
0
S
2
1
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
P S
+
<3s
P S
+
<3s
P S
<3s
59 42 896 D3417
66 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.5 Combination of rinsing position S with tumbler filling and flushing controlled by dentist element
båÖäáëÜ
Only by a service engineer
S
+ +
P1
P2
max.
P1
Option
P2
1 kg
S
+
P1
P2
S
+
P1
P2
S
P1
P2
59 42 896 D3417
D3417.201.01.22.02 67
12 Water unit Sirona Dental Systems GmbH
12.6 Assistant element control panel - cuspidor flushing Operating Instructions C8+, C8+ Turn
30s
max.
1 kg
10s 10 m
1. 2.
2s
10s 10 m
3.
4.
5.
C8+
C8+ Turn
59 42 896 D3417
68 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.8 Assistant element control panel - tumbler filling
Operation
båÖäáëÜ
4s
0
S
2
1
P
Reprogramming
>3s <15s
1.
0 2s
S
2.
>3s <15s
3.
0
S
59 42 896 D3417
D3417.201.01.22.02 69
12 Water unit Sirona Dental Systems GmbH
12.9 Assistant element control panel - combination of mouth rinsing position S with tumbler filling and cuspidor flushing Operating
P
<3s
S
+ +
0
Option
S
2
1
P
0
S
2
1
P
P
<3s
S
+
P
<3s
S
+
P S
<3s
59 42 896 D3417
70 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.10 Fresh water bottle
båÖäáëÜ
CAUTION
Fill with
1. 2. Sirona: Dentosept P
N
MO
LE
CAUTION
Replace deformed, scratched or discol-
ored bottles immediately!
100 : 1
Operating temperature:
+10 °C – +45 °C.
H2 O
or
Do not continue using the bottle after its
! ! ! maximum period of use has expired!
!
4. 5.
LEMON
59 42 896 D3417
D3417.201.01.22.02 71
12 Water unit Sirona Dental Systems GmbH
12.11 Switching between tap water/fresh water C8+ Operating Instructions C8+, C8+ Turn
P1
P2
alternatively
or
1P
+
2P
59 42 896 D3417
72 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.12 Hydrocolloid connection
Extra equipment
båÖäáëÜ
OPEN
CLOSE
59 42 896 D3417
D3417.201.01.22.02 73
12 Water unit Sirona Dental Systems GmbH
12.13 Dürr wet suction system Operating Instructions C8+, C8+ Turn
H2 O
C8+ Turn
simultaneously
10s
59 42 896 D3417
74 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.14 Amalgam separator
båÖäáëÜ
!
AM
TEST
95 %
AMALG
DESINF
SAN
0%
A
AM
A= ! Please replace
TE
amalgam container !
ST
AM
ALG
DES
INF
SA
N
!!
AM
TEST
100 %
AMALG
DESINF
SAN
0%
B
AM
B= !! Replace
amalgam container !!
TE
ST
AM
ALG
DES
INF
SA
N
AM
TEST
AMALG
Error/fault
DESINF
SAN
AM
TE
ST
AM
ALG
DES
INF
SA
N
20042005 ....
20022003
1 2 3 4 5 6 7
1 2 3 4 8 5 9 610 711 12 13 14
1 2 3 4 8 5 9 610 71115121613171418 19 20 21
AM
1 2 3 4 8 5 9 610 71115121613171418 19 20 21
22 23 24 25 26 27 28
8 9 10 111512161317141822192320242125292630273128 TEST
15 16 17 1822192320242125292630273128
22 23 24 25292630273128
29 30 31 AM AMALGAM test
TEST at least 1x annually
AM
TEST
AMALG
DESINF
i NOTE
Information on replacing the amalgam
SAN separator can be found on page 96.
AM
TE
ST
!
AM
ALG
DES
INF
SA
N
OK
59 42 896 D3417
D3417.201.01.22.02 75
12 Water unit Sirona Dental Systems GmbH
12.15 Disinfection system Operating Instructions C8+, C8+ Turn
A
A = ! If the DESINF indicator flashes,
refill 1 liter of DENTOSEPT P!
Refill with
Sirona: Dentosept P
59 42 896 D3417
76 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.16 Conversion of water unit right-handed left-handed version
båÖäáëÜ
Conversion only possible on a C8+
Turn
CAUTION
Danger of falling objects! Prior to con-
version, remove all objects loosely depos-
ited on any components of the treatment
center.
i NOTE
Please observe the minimum clearances
specified in the document “Dimension
Drawing, Technical Data”.
59 42 896 D3417
D3417.201.01.22.02 77
12 Water unit Sirona Dental Systems GmbH
12.16 Conversion of water unit right-handed left-handed version Operating Instructions C8+, C8+ Turn
59 42 896 D3417
78 D3417.201.01.22.02
Sirona Dental Systems GmbH 12 Water unit
Operating Instructions C8+, C8+ Turn 12.16 Conversion of water unit right-handed left-handed version
båÖäáëÜ
59 42 896 D3417
D3417.201.01.22.02 79
13 Operating light - option Sirona Dental Systems GmbH
13.1 SIROLUX E Operating Instructions C8+, C8+ Turn
13.1 SIROLUX E
C8 +
P
Option
C8 +
2
0
1
S
100
80
60
ENDO 40
P 1
20
P alternatively
0
S
2 S
1
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
Composite setting
100
80
60
ENDO 40
P 1
20
P
0
S
Light intensity: 8000 lux
2 S
1
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
CAUTION
wrong right Do not use without transparent cover!
59 42 896 D3417
80 D3417.201.01.22.02
Sirona Dental Systems GmbH 13 Operating light - option
Operating Instructions C8+, C8+ Turn 13.1 SIROLUX E
Focusing
CAUTION
båÖäáëÜ
When focusing, make sure you turn the
cover cap only slightly and just up to its
stop. If the cover cap is turned counter-
clockwise past the stop, it will spring out of
its catch and the halogen lamp can fall
down.
Click
75 mm
3 ’’
m 700 mm
0m 28 ’’
18
’’
7
59 42 896 D3417
D3417.201.01.22.02 81
13 Operating light - option Sirona Dental Systems GmbH
13.2 SIROLUX FANTASTIC Operating Instructions C8+, C8+ Turn
C8 +
P
Option
C8 +
2
0
1
S
100
80
60
ENDO 40
P 1
20
P alternatively
0
S
2 S
1
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
Composite setting
100
80
60
ENDO 40
P 1
20
P
0
S
2
Light intensity: 8000 lux
S
1
0
1
2
C8 +
P
0
S
2
1
100
80
60
ENDO 40
20
1
P
S
0
1
2
59 42 896 D3417
82 D3417.201.01.22.02
Sirona Dental Systems GmbH 13 Operating light - option
Operating Instructions C8+, C8+ Turn 13.3 LEDview
Focusing
CAUTION
båÖäáëÜ
When focusing, make sure you turn the
cover cap only slightly and just up to its
stop. If the cover cap is turned counter-
clockwise past the stop, it will spring out of
its catch and the halogen lamp can fall
Click
down.
100 mm
4 ’’
m 700 mm
0m 28 ’’
20
’’
8
13.3 LEDview
Operation of the LEDview operating
light: see LEDview Operating Instruc-
tions:
REF 62 07 687
59 42 896 D3417
D3417.201.01.22.02 83
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.1 Care, cleaning agents, and disinfectants Operating Instructions C8+, C8+ Turn
i NOTE
The time intervals specified for care,
cleaning, and disinfection are reference
values. Please adapt the time intervals to
suit your personal method of working and
your national requirements
CAUTION
Unsuitable care and cleaning agents and
disinfectants can damage or affect the
surface of the unit or functions.
To this end, only use care and cleaning
agents and disinfectants that have been
approved by Sirona.
59 42 896 D3417
84 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.2 Patient chair
Upholstery/foot switch/chair
båÖäáëÜ
Care for and clean upholstery: The arti-
ficial leather upholstery must be cared for
and cleaned regularly (at least 1x/week),
especially light colored upholstery.
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
Foot control
Clean the bottom of the base plate using
a damp cloth moistened in water.
59 42 896 D3417
D3417.201.01.22.02 85
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.2 Patient chair Operating Instructions C8+, C8+ Turn
MultiMotion headrest
Disinfecting the surfaces
The surfaces can be spray
and wipe disinfected with
surface disinfectants.
i NOTE
Fast and frequent removal of the head pad
from the front or rear can damage the
snap-in noses.
CAUTION
Damaged pads, e.g. with scratched sur-
faces, must be replaced.
For reorders:
Contact surface: REF 61 82 047
59 42 896 D3417
86 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.2 Patient chair
båÖäáëÜ
check the water and air filters for permea-
bility. The filters must be changed if nec-
essary:
1. Switch the treatment center off at the
main switch.
The water and air supply is discon-
B
nected.
C
2. Remove the cover A.
A i NOTE
Any remaining water flows out of the water
filter when it is opened. You should there-
fore place an absorbent cloth underneath
the filter.
59 42 896 D3417
D3417.201.01.22.02 87
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.3 Dentist element Operating Instructions C8+, C8+ Turn
ENDO
1
20
60
40
Instrument hoses/instrument rack
P
1
S
0
CAUTION
2
7.
i NOTE
For care and cleaning of the motors, the
straight and contra-angle handpieces, the
highspeed handpiece and the
SIROSONIC L scaler, please refer to the
corresponding operating instructions.
6.
5x 5x
Wipe disinfection
8.
7.
A
6.
8. 5x 5x
59 42 896 D3417
88 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.3 Dentist element
båÖäáëÜ
1.
2.
3.
3-way syringe
H2O
134°C 274°F Spray disinfection/wipe
disinfection
3 min
59 42 896 D3417
D3417.201.01.22.02 89
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.3 Dentist element Operating Instructions C8+, C8+ Turn
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)
3 min
Always make sure that the Mini L.E.D.
light is switched off before using disinfec-
tants. Do not use any abrasive cleaners.
Prior to each use, check the light guides to
make sure that no composite residues
have adhered to them. If this is the case,
remove these residues immediately and
check to make sure that the surface of the
insert has not been damaged. If you find
any damage, replace the damaged light
guide.
Surfaces
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
Wipe disinfection
C8 +
C8 +
P P
Option
0 0
S S
2 2
1 1
100 100
80 80
60 60
ENDO 40 ENDO 40
20 20
1
P P 1
S S
0 0
1 1
2 2
Option
134°C 274°F
P1
P2
3 min
C8 + C8 +
P P
0
0
S
S
2 2
1 1
100 100
80 80
60 60
ENDO 40
ENDO 40
20 20
1
P 1
P
S S
0 0
1 1
2 2
59 42 896 D3417
90 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.4 Assistant element
båÖäáëÜ
ENDO
60
100
80 2.
40
20
P 1
CAUTION
S
0 Use only care and cleaning agents and
1
2
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
1.
CAUTION
Egress of air when changing hose:
No amalgam! If the hose is removed from the 3-way sy-
ringe when the treatment center on the as-
sistant element is switched on, air es-
capes from the connection flange. To
prevent this, switch the treatment center
off and release pressure from the 3-way
syringe before replacing the hose.
Amalgam recycling
3.
Wipe disinfection
According to manufacturer’s
3 min specifications
Every 2 hours:
2 glasses
Every day
after work
1 liter
59 42 896 D3417
D3417.201.01.22.02 91
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.5 Water unit Operating Instructions C8+, C8+ Turn
Amalgam recycling
Spray disinfection/wipe
disinfection
CAUTION
ap
ap
So
tergents or acidic agents!
0,
De 75 l
sin
f. Daily before the lunch break and after
consulting hours
The vacuum system is subjected to septic
secretions, saliva and blood daily. For
hygienic reasons the system must there-
fore be disinfected with a detergent/disin-
fectant approved by Sirona.
Daily
Sterilize the tumbler plate A.
CAUTION
Do not operate the unit without the tumbler
plate!
59 42 896 D3417
92 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.5 Water unit
båÖäáëÜ
8 9 10 11 12 14
15 16 17 18 19 21
22 23 24 25 26 28 60s 2. Open the flap on the base of the water
C8+ 29 30
unit.
8 +
2. 3. 4.
ENDO
60
40
100
80
CAUTION
The collecting tank may contain amalgam
20
1
P
C
C8+ Turn 3. Remove the siphon and empty it into
a special collecting tank.
4. Replace the siphon and close the flap
again.
5. Activate the cuspidor flushing of the
cuspidor bowl for at least one minute.
59 42 896 D3417
D3417.201.01.22.02 93
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.5 Water unit Operating Instructions C8+, C8+ Turn
i NOTE
Perform sanitation very
2.
4 weeks!
1.
2.
Sanitation (1)
Fill with
Sirona: Dentosept P
3. 4.
1P
2P
4.
20 s
P1
P2
alternatively
or
5.
5.1 5.2 5.3 5.4
100
80
60
ENDO 40
20
1
P
S
0
1
2
6.
59 42 896 D3417
94 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.5 Water unit
båÖäáëÜ
Sanitation (2)
H2 O
2.
1. Fill with water
2.
100
80
60
ENDO 40
20
1
P
1
6.3
S
0
3.
alternatively
or
1P
4. 2P
5.
P1
P2
20 s
6.
6.1 6.2 6.4
Tumbler filling 2x
59 42 896 D3417
D3417.201.01.22.02 95
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.5 Water unit Operating Instructions C8+, C8+ Turn
Amalgam separator
D
C
2.
C8+ Turn
G = Do not reuse!
3.
Amalgam recycling!
Replacement:
Amalgam rotor
REF 14 34 138
4. G
5.
6.
59 42 896 D3417
96 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.5 Water unit
båÖäáëÜ
use > 1 week. Start the checkups in inter-
vals of no more than two weeks and
adjust the time intervals depending on the
results.
In addition to running laboratory tests, you
can also use the "total count tester" as a
simple means of performing this test. You
can obtain the "total count tester" using
REF 58 53 775 via the dental specialist
30s dealer.
CAUTION
The “Total Count Tester” has a maximum
durability of 1 year after its date of delivery.
CAUTION
If the number of germs significantly ex-
ceeds 100, then sanitation is required
(see next page).
59 42 896 D3417
D3417.201.01.22.02 97
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.6 Sanitation of the treatment center Operating Instructions C8+, C8+ Turn
+ DESINF
SAN
= o.k.! • every 4 weeks
• after prolonged periods of disuse
(> 1 week)
• if the germination index exceeds 100
AM
TEST
germs per milliliter (see page 97).
AMALG
No sanitation possible,
+ =
DENTOSEPT P
DESINF
Refill with Dentosept P! A= 1% DENTOSEPT P
SAN
99% H2O
B= 100% DENTOSEPT P
!
A For reorders:
CAUTION
A
Sanitation can only be started if the DIS-
TEM
ST
4.
AM
DE
AL
SI
G
NF
A INF LED is not lit. If the DISINF LED is lit,
SA
!
5. audible signal sounds.
The following steps are performed auto-
matically:
• The mixing tank is emptied through the
tumbler outlet.
• The SAN LED begins to flash as soon
AM as the mixing tank is completely empty.
TEST
AMALG
DESINF
6. SAN
A
TEM
ST
AM
AL
G
DE
SI
NF
SA
N
59 42 896 D3417
98 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.6 Sanitation of the treatment center
båÖäáëÜ
B= 100% DENTOSEPT P
ap-
B
OK !
59 42 896 D3417
D3417.201.01.22.02 99
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.6 Sanitation of the treatment center Operating Instructions C8+, C8+ Turn
alternatively
or
2.
15s
3.
4.
59 42 896 D3417
100 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.6 Sanitation of the treatment center
båÖäáëÜ
effective time
B= 100% DENTOSEPT P
>3s
2.
B 3.
5.
59 42 896 D3417
D3417.201.01.22.02 101
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.6 Sanitation of the treatment center Operating Instructions C8+, C8+ Turn
1. A
5 - 10 s
A
2.
3.
3. 10s 10s
59 42 896 D3417
102 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.6 Sanitation of the treatment center
båÖäáëÜ
Rinse the water-carrying instruments with
a solution of 1% DENTOSEPT P and 99%
water (A).
1.
1. Remove the 3-way syringe and rinse
A
it for at least 20s.
20s
alternatively
or
2.
15s
3. + 5.
OK !
59 42 896 D3417
D3417.201.01.22.02 103
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.7 SIROLUX E operating light Operating Instructions C8+, C8+ Turn
CAUTION
Always turn the main switch of the lamp off
before replacing the lamp (see page 28)!
CAUTION
Pay attention to elevated temperature of
the lamp components; and let these cool
down.
CAUTION
Only:
REF 48 89 965
i Type OSRAM 64 440
CAUTION
Use only cleaning agents which are ap-
proved by Sirona (see page 84, section
14.1)!
Spray disinfection/
wipe disinfection
59 42 896 D3417
104 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.7 SIROLUX E operating light
båÖäáëÜ
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
Spray disinfection/
wipe disinfection
i NOTE
To change the handle position, proceed as
follows:
59 42 896 D3417
D3417.201.01.22.02 105
14 Care and cleaning instructions for the practice team Sirona Dental Systems GmbH
14.8 SIROLUX FANTASTIC operating light Operating Instructions C8+, C8+ Turn
CAUTION
Always turn the main switch of the lamp off
before replacing the lamp (see page 28)!
CAUTION
Pay attention to elevated temperature of
the lamp components; and let these cool
down.
i CAUTION
Only use halogen bulbs REF 62 03 629!
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
Spray disinfection/
wipe disinfection
59 42 896 D3417
106 D3417.201.01.22.02
Sirona Dental Systems GmbH 14 Care and cleaning instructions for the practice team
Operating Instructions C8+, C8+ Turn 14.9 LEDview
båÖäáëÜ
CAUTION
Use only care and cleaning agents and
disinfectants which have been approved
by Sirona (see page 84, section 14.1)!
Spray disinfection/
wipe disinfection
i NOTE
To change the handle position, proceed as
follows:
14.9 LEDview
• Care and cleaning of the LEDview
operating light: see LEDview Operat-
ing Instructions:
REF 62 07 687
59 42 896 D3417
D3417.201.01.22.02 107
15 Options Sirona Dental Systems GmbH
15.1 Working stool CARL Operating Instructions C8+, C8+ Turn
15 Options
15.2 PAUL working stool (not available in the Federal Republic of Germany)
Please observe the operating instructions
for the PAUL working stool:
Order No. 63 16 827
59 42 896 D3417
108 D3417.201.01.22.02