7.5 Connecting to the Equipment (Starting PC- Enter the following username and password, select the language and
Utility) click [Login].
Login: dryprinter
Connect the communication of the PC for servicing (or settings-changeable models)
to the equipment.
<REMARKS>
Operation of the PC-Utility:
{MU:2.2_Flow of PC-Utility Operations}
Enter the following address in the address bar, and then press the
<ENTER> key on the key board.
http://172.16.1.30 (default IP address of this equipment):20051/USER/Login.htm
<NOTE>
When entering the address, discriminate between upper and lower cases.
equipment.
<REMARKS>
Input range: 0 to 65535
<DEFAULT>
<REMARKS>
<DEFAULT> To connect CR-IR 391CL/CR-IR 391V CL, CR-VW 674/CR-IR 355V CL or CR-IR
346CL/CR-IR 348CL, there is a need to register the Mac address, IP address, and
“0.0.0.0” hostname using the FF-WakeOnLAN Utility of CR-IR 391CL/CR-IR 391V CL, CR-VW
<REMARKS> 674/CR-IR 355V CL or CR-IR 346CL/CR-IR 348CL.
{ CR-IR 391CL/CR-IR 391V CL, CR-VW 674/CR-IR 355V CL, CR-IR 346CL/CR-IR
348CL Service Manual (MC Appendix)}
setting.
In [Establish Name], enter the establish name entered in “7.1 Creating <DEFAULT>
the Data Storage Folder”.
All [ON]
<DEFAULT> Enable QC
“Hospital Name”
Set whether to enable/disable the image QC function. In order to use the image QC
{IN:7.1_Creating the Data Storage Folder} function using U-Utility, this setting must be set to “Enable”.
<NOTE>
<Parameter>
Use one-byte alphanumeric when entering establish name.
[ON]/ [OFF]
Printing Dmax
[Execute]
Set whether to enable/disable the “density 4” step wedge for density measurement in
the QC test pattern of U-Utility.
equipment.
<Parameter>
[ON]/ [OFF]
<DEFAULT>
[OFF]
Set Date
Set the system date and time.
Set Alarm
Set whether to enable/disable the operation panel beep.
<Parameter>
[ON]/ [OFF]
<DEFAULT>
[ON]
Command for setting the power of this equipment to go OFF at the same time the
power of CR-IR 391CL/CR-IR 391V CL or CR-VW 674/CR-IR 355V CL is turned OFF.
<REMARKS>
This is a function of the DRYPIX PRIMA Monitor Tool. When connecting CR-IR 346CL/
CR-IR 348CL, this setting is not required as the DRYPIX PRIMA Monitor Tool is not
installed.
<Parameter>
[ON]/ [OFF]
<DEFAULT>
[ON]
[OK]
screen appears.
<REMARKS> Remove the modality connection software CD-ROM.
If it does not start automatically, double-click “DRYPRICopyTool.exe” in the CD-
ROM. Execute [Upgrading] of the PC-Utility.
*DRYPIX PRIMA
+System information and setup
+Transfer Indv. Data
+File Transfer
+Upgrading
Full Install
Upgrading
<REMARKS>
There is a need to reboot the equipment after installing the modality connection
software.
*DRYPIX PRIMA
+System information and setup
Setting1
Setting2
Setting3
Set Date
Ele Save Mode
Set Alarm
Reboot
Film Characteristic ID
Power Supply Voltage
<REMARKS>
If the PC-Utility was ended by clicking , the logout screen will not be
displayed unless the pop-up blocked setting of the browser is canceled.
{MU:2.1.1_Environment of PC for Servicing}
DRYPIX PRIMA
+System information and setup 0: 0116H[W]
+Transfer Indv. Data whether to send details of error
Details of errors are sent to the 272: 0110H[F]
Add Client
Details of errors (unique values 21317: 5345H[F]
Delete Client for FUJIFILM systems) are sent to
Change Client Name the client,without changing the
+Clients
default Use System Timeout Set timeout of reception from cli- 0: No
FCR-CSL ent or not 1: Yes
WatchDRYPIXLink Presentation LUT Use LUT information sent from cli- 0: No
ent or not 1: Yes
QC-TEST
Precede Picking Up Use precedent removal function 0: No
Common
or not 1: Yes
********
Use Max/Min Density Which minimum density/maximum 0: Printer
density value is to be used. 1: Modality
Set the various parameters according to the client setting list shown Extension Format ID Select ID for differentiating when CR (XG-1/CR-Console)
below.For details of each feature, refer to “Service Mode (MU) 4.4 PC- extension format is used CR (Synapse)
EV (Philips Easy Vision)
HM (Hitachi DR2000/Clavis
MU:2.4 [3] } view)
SM (Siemens)
Use Default Annotation Use Annotation or not 0: No
1: Yes
Change Base Color Change base color or not 0: disable
(Disable:110H[F]) 1: enable
0: 27.9x35.6
description as 25.7x36.4 cm or as 1: 25.7x36.4
27.9 x 35.6 cm
-
rently)