IN-35

7.5 Connecting to the Equipment (Starting PC- Enter the following username and password, select the language and
Utility) click [Login].
Login: dryprinter
Connect the communication of the PC for servicing (or settings-changeable models)
to the equipment.
<REMARKS>
Operation of the PC-Utility:
{MU:2.2_Flow of PC-Utility Operations}

Start Internet Explorer on the PC for servicing (or settings-changeable


models).

Enter the following address in the address bar, and then press the
<ENTER> key on the key board.
http://172.16.1.30 (default IP address of this equipment):20051/USER/Login.htm

<NOTE>
When entering the address, discriminate between upper and lower cases.

equipment.

020-201-02E FM-DL 100 IN-35


IN-36
7.6 System Settings 7.6.1 Setting 1
Using the PC-Utility, perform the following system settings of the equipment. Skip <INSTRUCTIONS>
items which can be left at their default values when operating the equipment.
administrator of the installation site.
*DRYPIX PRIMA
+System information and setup
Setting1 AE Title
Check Version
Input Serial No.
Set DICOM <DEFAULT>
Set Network
Display Mac address “DRYPIX”
Registration service pc <REMARKS>
Setting2
Set Rem.Films
Set Tray remaining number of digits.
Enable QC
Set Target Den. the 1-byte code table.
Printing Dmax
Auto F.D.C.
Set Film infomation
Display indv. data
Setting3
Set SMPTE (V2.0 or later)
Set Date
Port No.
Ele Save Mode Set the port number (decimal/5 digits) used for the network.
Set Alarm
<DEFAULT>
Reboot “17238”
Film Characteristic ID
Power Supply Voltage <NOTE>
Do not set the following port numbers as they are reserved.

<REMARKS>
Input range: 0 to 65535

020-201-02E FM-DL 100 IN-36


IN-37
IP Address Host Name
Set the IP address (decimal) used for the network. Set the host name (maximum 16 digits) used for the network.
<DEFAULT> <NOTE>
“172.16.1.30” Set the host name entered at “7.1 Creating the Data Storage Folder” of the
Installation (IN) section.
<REMARKS>
{IN:7.1_Creating the Data Storage Folder}

<DEFAULT>

<REMARKS>

lower case), numerals, and hyphen “-”.


Subnet Mask
Set the subnet mask (decimal) used for the network.
<DEFAULT>
“255.255.0.0”
<REMARKS>
Display Mac address
The MAC address of the CPU board is displayed.
<DEFAULT>
Gateway “255.255.0.0”
Set the gateway address (decimal) used for the network. <REMARKS>

<DEFAULT> To connect CR-IR 391CL/CR-IR 391V CL, CR-VW 674/CR-IR 355V CL or CR-IR
346CL/CR-IR 348CL, there is a need to register the Mac address, IP address, and
“0.0.0.0” hostname using the FF-WakeOnLAN Utility of CR-IR 391CL/CR-IR 391V CL, CR-VW
<REMARKS> 674/CR-IR 355V CL or CR-IR 346CL/CR-IR 348CL.
{ CR-IR 391CL/CR-IR 391V CL, CR-VW 674/CR-IR 355V CL, CR-IR 346CL/CR-IR
348CL Service Manual (MC Appendix)}

setting.

020-201-03E FM-DL 100 IN-37


IN-38
Registering the PC for Servicing (or Settings-Changeable 7.6.2 Setting 2
Models)
Set Tray
Any unregistered PC for servicing (or settings-changeable models) is limited in
function and does not support “Transfer Indv. Data” and “File Transfer” of PC-Utility. <REMARKS>
<REMARKS>
Only one PC for servicing (or settings-changeable models) can be registered. Only the
<Parameter>

Select [Registation service pc] in the maintenance tree.

In [Establish Name], enter the establish name entered in “7.1 Creating <DEFAULT>
the Data Storage Folder”.
All [ON]
<DEFAULT> Enable QC
“Hospital Name”
Set whether to enable/disable the image QC function. In order to use the image QC
{IN:7.1_Creating the Data Storage Folder} function using U-Utility, this setting must be set to “Enable”.
<NOTE>
<Parameter>
Use one-byte alphanumeric when entering establish name.
[ON]/ [OFF]

In [Service PC IP Address], enter the IP address of the current PC for


<DEFAULT>
servicing (or settings-changeable models). [ON]

<DEFAULT> Set Target Den.


“172.16.1.20” Set the density of each step wedge used when recording the image QC pattern of
U-Utility.
<REMARKS>
Default IP address of CR-IR 391CL/CR-IR 391V CL/CR-VW 674/CR-IR 355V CL/ <DEFAULT>
CR-IR 346CL/CR-IR 348CL: “172.16.1.20” [Density 1: 48, Density 2: 120, Density 3: 192]

Printing Dmax
[Execute]
Set whether to enable/disable the “density 4” step wedge for density measurement in
the QC test pattern of U-Utility.
equipment.
<Parameter>
[ON]/ [OFF]
<DEFAULT>
[OFF]

020-201-03E FM-DL 100 IN-38


IN-39
Auto F.D.C. 7.6.3 Setting 3
Select the timing for executing automatic density correction.
<Parameter>
Command for setting parameters for outputting SMPTE test patterns using U-Utility.
[off]/ [every pack]
<Parameter>
<DEFAULT>
No. : [1] to [5]
[every pack] AE Title
Image Matrix : [1]/[2]/[4]/[6]/[8]/[12]
Set Film information LUT Number : [LUT 1] to [LUT 8]
Interpolation type : [SHARP]/[MEDIUM]/[SMOOTH]
Interpolation algorithm/method : [NONE]/[CUBIC]
<Parameter>
<DEFAULT>
[Upper Left]/ [Upper Right]/ [Lower Left]/ [Lower Right]
1 default 1 LUT 1 SHARP CUBIC
<DEFAULT> 2 default 1 LUT 1 SHARP CUBIC
3 default 1 LUT 1 SHARP CUBIC
[Upper Left]
4 default 1 LUT 1 SHARP CUBIC
5 default 1 LUT 1 SHARP CUBIC

Set Date
Set the system date and time.

Ele Save Mode


Set the timer value when this equipment shifts into the sleep mode.(unit: minute)
<Parameter>
[0]/ [15]/ [30]/ [45]
<DEFAULT>
[15]

Set Alarm
Set whether to enable/disable the operation panel beep.
<Parameter>
[ON]/ [OFF]
<DEFAULT>
[ON]

020-201-02E FM-DL 100 IN-39


IN-39.1

Command for setting the power of this equipment to go OFF at the same time the
power of CR-IR 391CL/CR-IR 391V CL or CR-VW 674/CR-IR 355V CL is turned OFF.
<REMARKS>
This is a function of the DRYPIX PRIMA Monitor Tool. When connecting CR-IR 346CL/
CR-IR 348CL, this setting is not required as the DRYPIX PRIMA Monitor Tool is not
installed.

<Parameter>
[ON]/ [OFF]
<DEFAULT>
[ON]

Power Supply Voltage


Set the voltage of the power to which the equipment is connected.
<Parameter>
[ON]/ [110V]/ [120V]/ [200V]/ [220V]/ [230V]/ [240V]
<DEFAULT>
[100V]
<REMARKS>
The default value of the 200V model is “200V”.

020-201-03E FM-DL 100 IN-39.1


IN-39.2
7.7 Installing the Modality Connection Software Click the [Copy] button.
(to Support Multi Modalities)
<REMARKS>
The modality connection software needs to be installed in this equipment (with
software version V2.0 or later) so that it can support multi modalities. The [Input drive] and [Make or delete directory] of the [DPX PRIMA Copy Tool]
screen have the same functions as “Creating data storage folder”.
Load the modality connection software CD-ROM into the CD-ROM {IN:7.1_Creating the Data Storage Folder}
drive of the PC for servicing (or settings-changeable model).

[OK]
screen appears.
<REMARKS> Remove the modality connection software CD-ROM.
If it does not start automatically, double-click “DRYPRICopyTool.exe” in the CD-
ROM. Execute [Upgrading] of the PC-Utility.
*DRYPIX PRIMA
+System information and setup
+Transfer Indv. Data

+File Transfer
+Upgrading
Full Install
Upgrading
<REMARKS>
There is a need to reboot the equipment after installing the modality connection
software.

020-201-03E FM-DL 100 IN-39.2


IN-39.3
7.8 Rebooting the Equipment
Rebooting the equipment enables various system settings changed.
Select [Reboot] on the PC-Utility.

*DRYPIX PRIMA
+System information and setup
Setting1
Setting2
Setting3
Set Date
Ele Save Mode
Set Alarm
Reboot
Film Characteristic ID
Power Supply Voltage

close the window.

The equipment will reboot after about 75 seconds.

020-201-02E FM-DL 100 IN-39.3


IN-39.4
7.9 Disconnecting the Equipment Close the logout window after the logout screen shows “It logged out
from the printer. Please close this page.”
After rebooting the equipment, click the [Logout] button of the
maintenance tree.

<REMARKS>
If the PC-Utility was ended by clicking , the logout screen will not be
displayed unless the pop-up blocked setting of the browser is canceled.
{MU:2.1.1_Environment of PC for Servicing}

Reboot the PC-Utility.

Check that “DRYPIX PRIMA maintenance tree with Multi-Modality” is


displayed on the menu tree of the PC-Utility if the modality connection
software has been installed.

020-201-02E FM-DL 100 IN-39.4


IN-39.5
7.10 Client Settings (Multi Modality Equipment Protocol
Bold: Default
Only) Item Function Value
Set parameters for the connected client using PC-Utility. Attribute list error Send error message for unexpect- 0: not indicate
Set the items according to the corresponding client settings in “Modality Connection (0107H Warning) ed data received or ignore warn- 1: indicate
List”. ing
N Event Report RQ Send event report or not 0: disable
<NOTE> 1: enable
The default values for client parameters listed below are the values for a newly Attribute Value Out Of Send error message for out-of- 0: not indicate
registered client and differ from the default values set for “default”, “FCR-CSL”, Range (0116H Warning) range data received or ignore 1: indicate
“WatchDRYPIXLink”, and “QC-TEST”. warning
Notify that images have been 0: not indicate
imagebox 1: indicate
Newly register the client according to the installation environment. (B604H Warning) frame, or ignore warning

DRYPIX PRIMA
+System information and setup 0: 0116H[W]
+Transfer Indv. Data whether to send details of error
Details of errors are sent to the 272: 0110H[F]
Add Client
Details of errors (unique values 21317: 5345H[F]
Delete Client for FUJIFILM systems) are sent to
Change Client Name the client,without changing the
+Clients
default Use System Timeout Set timeout of reception from cli- 0: No
FCR-CSL ent or not 1: Yes
WatchDRYPIXLink Presentation LUT Use LUT information sent from cli- 0: No
ent or not 1: Yes
QC-TEST
Precede Picking Up Use precedent removal function 0: No
Common
or not 1: Yes
********
Use Max/Min Density Which minimum density/maximum 0: Printer
density value is to be used. 1: Modality
Set the various parameters according to the client setting list shown Extension Format ID Select ID for differentiating when CR (XG-1/CR-Console)
below.For details of each feature, refer to “Service Mode (MU) 4.4 PC- extension format is used CR (Synapse)
EV (Philips Easy Vision)
HM (Hitachi DR2000/Clavis
MU:2.4 [3] } view)
SM (Siemens)
Use Default Annotation Use Annotation or not 0: No
1: Yes
Change Base Color Change base color or not 0: disable
(Disable:110H[F]) 1: enable
0: 27.9x35.6
description as 25.7x36.4 cm or as 1: 25.7x36.4
27.9 x 35.6 cm
-
rently)

020-201-02E FM-DL 100 IN-39.5


IN-39.6
Magnify LUT Common
Bold: Default Bold: Default
Item Function Value Item Function Value
Smoothing Type Selection of smoothing type 1: Sharp Default LUT Selection of LUT number used as 1 - 8 (1)
2: Medium default
3: Smooth Illumination Lighting 0 - 65535 (2000) *1
Default Requested 1: DECIMATE
Ambient Light 0 - 65535 (10) *1
Decimate/Crop Behavior client does not send information 2: CROP
when images are reduced 3: FAIL Which LUT number is to be used 0: Printer
(Default LUT#) 1: Modality
Set the image enlargement/reduc- 0: NONE
tion processing type 1: CUBIC Default Illumination Which lighting setting is to be 0: Printer
used 1: Modality
Specify the length of line to be 0 - 7000 (0)
printed on the image for one Default Ambient Light 0: Printer
frame to be used 1: Modality
Which enlargement/reduction pro- 0: Printer Procedure Selection of tone correction pro- 0: SAR
cessing type is to be used 1: Modality cessing method 1: BAR
Default Smoothing Type Which smoothing type is to be 0: Printer
used 1: Modality
Default Requested Image 0: Printer LUT1 to LUT8
to be used 1: Modality
Bold: Default
Default Decimate/Crop Which Decimate/Crop Behavior is 0: Printer
Behavior to be used 1: Modality Item Function Value
Procedure Selection of enlargement/reduc- 0: A-VR Max Density Maximum density value (x100 for 0 - 300
tion processing method 1: SSM (BAR: 0 - 100) *2
Edge Detection Perform enlargement/reduction 0: OFF Min Density Minimum density value (x100 for 0 - 300 *2
processing for edge or not 1: ON SAR, No setting for BAR)
Detection Level Edge detection level setting 1 - 1023 (600/800/600) 11 - 99 (BAR: 1 - 10)
LUT *2, *3
Sharpness -70 - 300 (130/30/80)
setting Contrast Contrast setting (x100, No setting (BAR: 10 - 400) (100)
for SAR)
Character Color Character color setting 0: White
1: Black Number of Tuning Points 0 - 5 (0)
SAR
Density Setting of x100 of density value to 1 - 300 (120)

Shift Setting of x100 value of move- -100 - 100 (0)


ment density value
Contrast Setting of x 100 of contrast of 10 - 400 (100)

*2: The default value differs in LUT1 to LUT8.


*3: Up to 20 can be entered for BAR but enter between 1 and 10.

020-201-02E FM-DL 100 IN-39.6


IN-40
Output Format Item Function Value
Bold: Default Margin between Image Setting of image frame intervals 0 - 50 (20)
Item Function Value (Unit: pixel)
1: 8x10 Image Layout Setting of image layout method
2: 10x14 0: concentration
3: 14x14
4: 14x17 Center of remaining segment set 1: spread
5: 10x12 with frame interval value
Medium Type 1: Blue Processing Type Setting of density processing cal- 0: Type1
0: Clear culation method 1: Type2
Film Orientation 1: PORTRAIT Image Crop UPDWN Area - 0: Upper
2: LANDSCAPE for CR Image tom is to be trimmed from a CR 1: Lower
image that is larger than the im- 2: Both
Border Density Background density (x100 of ac- 0 - 300 (300) age box.
tual value)
Image Crop Side Area for 0: Left
Polarity Reverse the black/white of the im- 1: Normal CR Image is to be trimmed from a CR image 1: Right
age 2: Reverse that is larger than the image box. 2: Both
Trim Whether there is image trim 0: OFF
1: ON Annotation
Trim Width Width of the trim (pixels) 1 - 9 (1) Bold: Default
Trim Density Image trim density (x100 of actual 0 - 300 (0) Item Function Value
value)
Upper Left/Upper Center/Up- Setting of annotation position, and
Number of Copies Default number of print copies 1 - 99 (1) per Right/Lower Left/Lower entering
Print Priority Setting of print priority 1: HIGH Center/Lower Right
2: MEDIUM
3: LOW Common
Default Number of Copies 0: Printer Bold: Default
be used 1: Modality Item Function Value
Default Print Priority Which Print Priority is to be used 0: Printer Data Transfer Timeout Standby time setting when data is 30 - 3600 (300)
1: Modality suspended
Default Medium Type Film base color priority 0: Printer (Becomes error when setting time
1: Modality is exceeded)
Default Film Orientation Film orientation priority 0: Printer Association Count Maximum number of accesses to 1 - 10 (10)
1: Modality the equipment at the same time
by DICOM communication
0: Printer
1: Modality Logging DICOM DICOM communication log level 0: none
setting 1: normal
Default Border Density Which background density is to be 0: Printer 2: detail
used 1: Modality
Model Printer manufacture model name DRYPIX
Default Trim Which Trim setting is to be used 0: Printer
1: Modality Manufacturer Printer manufacturer name FUJIFILM Corporation
Defalut Polarity Which Polarity setting is to be 0: Printer Default Client Whether to use or not the [default] 0: Yes
used 1: Modality setting of the [Client:] list for un- 1: No
registered clients
Mirror 0: OFF (Normal)
1: ON (Mirror)

020-201-02E FM-DL 100 IN-40