Compact Urine Analyzer

Operating Manual
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL

Premise
This manual contains important information on the
functions of the AUTION MICRO PU-4210.

Supplied by : ARKRAY, Inc.


distributed by : A. Menarini I.F.R. srl
Diagnostic Division

This manual has been issued by: ARKRAY, Inc.


Read carefully prior to starting up the instrument.
It is recommended to retain this manual for future
use.

This product conforms to European Directive


98/79/EC.

e
1 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05

Introduction
This manual contains outlines, instructions for the
operation, maintenance and troubleshooting for the
AUTION MICRO PU-4210. Read this manual carefully
before use the Analyzer in order to operate it correctly.
After you read the manual, keep it near the Analyzer to
refer to it when necessary.

● This instrument uses urine as a sample. Urine may be


contaminated by pathogenic microbes that cause
infectious diseases. Always be careful when handling
urine samples. If operated incorrectly, there is a danger
of infection to you and those around you by pathogenic
microbes.

● Never touch places where samples may be contacted


with bare hands. When you perform maintenance,
always wear protective gloves to prevent exposure to
pathogenic microbes.

● Separate used Test Strips, etc. from general waste


and dispose of them according to local regulations on
biohazardous waste.

● This instrument may become

infectious in the course of use.


Discard the product in accordance
with local regulations for biohazardous
waste.

e
AUTION MICRO 2
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL

Caution Marks
Pay careful attention to these caution marks in this Operating Manual.

Regarding accidents resulting injury.

To prevent infection of operator or others from pathogenic


microbes, follow the instructions described.

To prevent injuries to operator or others, or material


damages, follow the instructions described.

Regarding damage and performance of Analyzer.

Failure to follow the instructions may lead to incorrect


Important
results.
To prevent damage to the unit and assist you in
Note
making best use of the capabilities of the Analyzer,
additional explanations and notes are provided herein.

Reference information on operation, additional


Reference
explanations and related functions are provided herein.

e
3 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Precautions for Installation


■For installation, read the
following precautions carefully
and always follow safety
measures.

● Place the Analyzer on a stable and level


surface, free of vibration.

●Do NOT place the Analyzer where it may ●Select a room in which to set up the
be affected by chemicals, corrosive gases Analyzer where the temperature can be
and electronic noise. controlled between 10℃ and 30℃, and
humidity in a range of 20 to 80 %.

●Do NOT place the Analyzer where it may ●Do NOT place the Analyzer where it may
be affected by direct sunlight or wind. be affected by magnetic materials including
magnets or speakers.

AUTION MICRO 4
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Precautions for Installation

●Plug in an exclusive AC adapter. ●When the instrument is not in use, unplug


the AC adapter.

●Should the Analyzer stop functioning, ●If an abnormal sound, odor or smoke is
contact your distributor for repairs. Do NOT found, turn the power off immediately and
attempt to repair it your self. Doing so may consult your distributor after making sure
result in damaging the Analyzer or injury. that the abnormal situation is not continuing.

●Do NOT place the Analyzer where it may ●Do NOT drop or hit the Analyzer.
be affected by water. Samples may cause
damage to the Analyzer.

5 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Table of Contents
AUTION MICRO(PU-4210)Operating Manual

Premise …………………………………………………………1
Introduction ……………………………………………………2
Caution Marks …………………………………………………3
Precautions for Installation …………………………………4
Table of Contents ……………………………………………6

Introduction Outline of AUTION MICRO …………………………………7


Shipping Carton ………………………………………………8
Parts Description and Functions ……………………………9
Display Screen and Key Arrangement ……………………10

Measurement Flow of Measurement ………………………………………11


Cautions ………………………………………………………12
Start Up ………………………………………………………14
Setting Up Parameters ……………………………………15
Normal Measurement ………………………………………16
Check Measurement ………………………………………19
Auxiliary Function ……………………………………………21
Special Parameters …………………………………………26

Printing Printer …………………………………………………………27


Printing Item …………………………………………………28

Maintenance and Maintenance …………………………………………………29


Troubleshooting Troubleshooting ………………………………………………30

AUTION MICRO 6
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Outline of PocketChem UA Introduction

The AUTION MICRO (PU-4210) is a small portable urine-analyzing device


which anyone can use anytime, anywhere. Since the Printer is removed
from the Analyzer, you can operate it by a minimal key operation.
The device is equipped with transmission capability to adapt to
systematization.

Characteristics
Minimum Maintenance
The only necessary maintenance item is cleaning the Strip Holder.

Long Life Light Source


Use of LED and short-time lighting enables a longer light source life.

Temperature Correction Function


An automatic correction function, which is provided by the internal thermal
sensor, solves any errors caused by temperature changes in the environment.

Power Source
Use a dry battery or a AC adapter when using the Analyzer alone, and use the
AC adapter when using it with a Printer.

Regent Selection
AUTION Sticks 10EA and Uriflet 9UB are available to use.

100 memories
Maximum 100 measurement results can be stored.
When more than 100 samples have been measured, older data will be automatically
deleted in order.

7 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Shipping Carton Introduction

Shipping carton of AUTION MICRO (PU-4210) contains the following.


Make sure that all of the following are included:

Analyzer Printer

AC Adapter     Check Strip Set  Thermal Printer Paper  Operating Manual


(this booklet)

Compact Urine Analyzer

Operating Manual

Optional Optional
Accessories Consumables

RS-232C Cable Thermal Printer Paper    Check Strip Set

Strip Holder

AUTION MICRO 8
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Parts Description and Functions Introduction

Parts description in this Operating Manual is as follows:

Front
8

1 2

3
9

Bottom (Analyzer) Rear (Analyzer)


5 6
7
4

    Name          Function


① Display Screen Displays measurement results and error codes
② Operator Panel For setting conditions and starting measurement
③ Optical System Measures samples
④ Power Input Terminal For connection of exclusive AC adapter
⑤ Connection Terminal For electrical connection to the Printer
⑥ DATA OUT Terminal Terminal for connection to a PC or other devices
⑦ Battery Box Place batteries here
⑧ FEED key Feeds printer paper
⑨ Lock Release Button Unlocks connection between the Analyzer and Printer

9 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Display Screen and Key Arrangement Introduction

LCD
5. Date
1. Memory 4. Delete
2. ID 6. Time
3. No.
7. Test Strip

10. figure display


8. Semiquantitative 13 digits
Symbol

9. Item Name

15. Battery Icon


11. Test Strip Icon

14. Printer Icon


12. Waiting Icon 13. Transmission Icon

1. Memory Displayed when the data is redisplayed.


2. ID Displayed (Blinks during ID setting) when ID is displayed.
3. No. Displayed (Blinks during No. setting) when No. is displayed.
4. Delete Blinks when the memory is deleted. For deletion, enter [ENT].
5. Date Light is on (Blinks during Date setting) when the Date is displayed.
6. Time Light is on (Blinks during Time setting) when the Time is displayed.
7. Test Strip Displayed when the Test Strip type is displayed. Blinks during Test Strip type setting.
8. Semiquantitative Symbol Displayed according to measurement results when measurement results are displayed or redisplayed.
9. Item Name Displayed according to test items when measurement results are displayed or redisplayed.
10. Figure display 13 digits Measurement No., ID, Date, Time, Test Strip name, etc. are displayed. (13 digits)
11. Test Strip Icon Blinks when the Test Strip is dipped or the Test Strip Holder is moved.
12. Waiting Icon Displayed from the measurement start to finish.
13. Transmission Icon Displayed for the external output.
14. Printer Icon Displayed when printing is performed.
15. Battery Icon Displayed in three different signs depending on the degree of exhaustion.

Key Arrangement Specification

1. Function Key or Numeric Key

2. Start Key, Enter Key or Power Switch

3. Stop Key or Hyphen Key

AUTION MICRO 10
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL

Flow of Measurement Measurement

Wear protective gloves to prevent exposure to pathogenic


microbes.

Separate used test papers, etc. from general waste and


dispose of them according to local regulations on
biohazardous waste.

Flow chart of measurement


The following flow chart explains a measurement procedure.
Before measurement, understand the whole procedure.

Turn on the main power.

Enter Measurement No. and Patient ID.


Select Test Strip Type

Start Measurement

Dip the Test Strip in the sample for 2 seconds in accordance with the alarm.

Remove the excess sample.

Set the Test Strip onto the Strip Holder.

The Strip Holder moves inside of the Analyzer and measurement starts.

Measurement Results are displayed and the Strip Holder comes out from the Analyzer.

Continue measurement
Dispose of Test Strip.
Finish measurement

Clean the Strip Holder, and turn off the main power.

e
11 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Cautions Measurement

Handling Samples

● Wear protective gloves to prevent exposure to pathogenic microbes.

● Use a fresh urine sample. If the sample urine is not measured within an
hour after collection, cap it tightly and keep it refrigerated. Be sure to
allow it to reach to room temperature before measurement.

● Mix the sample thoroughly before measurement. Do NOT centrifuge them.

● Prepare an adequate amount of samples so that the whole pad portion of


the Test Strip can be immersed.

● Do NOT add any preservative, disinfectant, or detergent to the sample.

● Keep the sample out of direct sunlight.

● If samples containing ascorbic acid are measured, the measured values


of glucose and occult blood are occasionally lower than the real values.

● If drug-administered urine or visually observed hematuria are measured,


correct measurement results sometimes cannot be obtained.

AUTION MICRO 12
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Cautions Measurement

Handling Test Strips

● Do NOT use Test Strips of other manufacturers.


Use only our Test Strips,“AUTION Sticks, Uriflet”
.
Read the attached manuals of the Test Strips.

● Do NOT use expired Test Strips.


Do NOT use expired Test Strips and/or Test Strips if the reagent pads
show any signs of discoloration, deformation or deterioration.

● Prepare Test Strips just before the measurement.


Take out only the necessary number of Test Strips to use and close the
cap. If Test Strips are left out for a long time, they may absorb humidity
or collect dust, causing incorrect measurement results.

● Do NOT touch the reagent pad of the Test Strip.


Do NOT touch the reagent pad of the Test Strip, otherwise correct
measurement results may not be obtained.

● Set Test Strip type.


Set the Test Strip type before measurement. If different Test Strips are
used, correct measurement results may not be obtained.

● Keep the desiccant.


Keep the desiccant contained in the Test Strip container until all Test
Strips are used.

13 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05

Start Up Measurement

Do NOT touch the optical system.


Doing so may result in injury.

1. Preparation.
① When using the Printer, make sure that the
Printer is connected to the Analyzer correctly
and paper is loaded. If there is a red strip on
the both sides of the paper, replace the roll.

2. Turn on the power.


① Connect the exclusive AC adapter with the
power input terminal on the rear of the device
(while using the AC adapter, battery is not
consumed). When only the Analyzer is
operated with batteries, insert batteries into
the battery box at the bottom of the Analyzer.

② Press [START-ENT] on the operator panel to


turn on the power. After initialization, the
Ready Screen appears.

There are two screens for the measurement-standby stage.(Ready Screen).


In either case, press [START] to start measurement.

Ready Screen 1 Ready Screen 2


The latest measurement Parameters for the next
result is displayed on measurement are
the LCD. displayed.
(Measurement No., ID,
Test Strips, Measurement
Item for Test Strip)

③ Make sure that the Strip Holder is set up correctly.


e
AUTION MICRO 14
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Setting Up Parameters Measurement

1. Setting up the measurement number.


① Press [No.] from the Ready Screen.
Measurement No. setting mode appears and“No.”and
the cursor blinks.
② Enter any numbers using the numeric keys.(Max. 4 digits)
③ Press [ENT] to confirm the entry. Press [STOP] to cancel,
and the Ready Screen is restored.
Measurement No. can be set arbitrarily.
Reference
Once the Measurement No. is set, the number increases
with every measurement.

2. Setting up of Patient ID.


① Enter [ID] from the Ready Screen. The patient ID setting
mode appears and“ID”and the cursor blinks.
② Enter any numbers using the numeric keys.
(Max. 13 digits including hyphens)
③ Press [ENT] to confirm the entry. The Ready Screen is
restored.
1: [STOP] functions as a hyphen key.
Reference
To cancel the entry, press [ENT] once before re-entry.
2: When the patient ID is entered, the Ready or Measurement
screen displays Measurement No. and Patient ID
alternately.
3: Patient ID is included in the printing results and external
transmission. (When not entered, -----------is displayed.)

3. Setting of Test Strip Type.


① Press [STRIP] from the Ready Screen, Test Strip type
selection mode appears and“STRIP”blinks, and the
Test Strip type and Item are displayed.
② To change the Test Strip type and Item, press [STRIP].
(10EA→9Ub(9UB)→C(CHECK)→10EA)
③ Press [ENT] to confirm the entry.
Press [STOP] to cancel, and the Ready Screen is restored.

15 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05

Normal Measurement Measurement

Wear protective gloves to prevent exposure to pathogenic


microbes.

1. Prepare sample.
① Prepare sample in a sample cup.
Make sure that sample is
Important
enough to cover the whole Reagent Pads

reagent pads.

② Stir sample well.


Do NOT use a centrifuge for
Important
mixing.

③ Prepare tissue paper to wipe off excess urine.

2. Prepare the Test Strip.


① Take out only the necessary number of strips
tl y

from the Test Strip container. C lo


s e th e lid ti g
h

② Close the cap.


Read the attached manuals
Important
for the Test Strip.

e
AUTION MICRO 16
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Normal Measurement Measurement

3. Press [START] to start Normal Measurement.


Test Strip Icon blinks and the timer starts to count
down the timing of dipping of the Test Strip from [3
sec]. (A timing buzzer also sounds in accordance with
the count down.)
To stop the measurement,
Reference
keep pressing [STOP] for 2-3
seconds during measurement.
Do NOT move the Analyzer during
Important
measurement. If impact or shock is given to the Analyzer,
measurement cannot be done correctly. The Test Strip
may become jammed in the Analyzer.

4. Dip a Test Strip in the sample.


Following the indicator timing (--sec), a timing buzzer
sounds for 2 seconds.
Do NOT dip
Keep dipping the Test Strip while the buzzer is the black marker
sounding.
Dip the Test Strip properly for 2
Important
seconds following the sound of the
timing buzzer. Shorter dipping time
may result in insufficient coloring
and longer dipping time may cause
discharging of the reagent.

5. Remove the excess sample.


When the timing buzzer stops, pull out the Test
Strip and remove the excess sample with tissue
paper.

17 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Normal Measurement Measurement

6. Set the Test Strip.


While the Test Strip Icon is blinking, place the Test
Strip onto the Strip Holder.
A few seconds before the Strip
Reference
Holder moves inside of the
Analyzer, the Test Strip Icon
stops blinking. When the Strip
Holder is set inside the Analyzer,
the Test Strip Icon disappears. Make sure that the tip of Test Strip
If the Test Strip is not correctly reaches to the left end of the holder.
Important
set on the Strip Holder, correct
measurement results cannot be
obtained.

7. Measurement.
The Strip Holder is pulled into the Analyzer
and the measurement starts.

8. Finish Measurement.
When measurement is completed,the
measurement result is displayed and the
Strip Holder comes out from the Analyzer.
Discard the used Test Strip, and when
finishing the whole measurement process,
remove the Strip Holder from the Analyzer
and clean it.

9. Turn off the power.


Keep pressing [START] for 2-3 seconds to turn off the power (the Strip Holder is
pulled into the Analyzer automatically).
When no key entry is made for more than 3 minutes,
Reference
the power automatically turns off.

AUTION MICRO 18
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL

Check Measurement Measurement

For installation, read the following precautions carefully and always


follow safety measures.

Wear protective gloves to prevent exposure to pathogenic


microbes.

1. Clean the Strip Holder.


Wash the Strip Holder and dry it completely with
the soft cloth. See“Maintenance”.
Make sure the holder is clean,
Note
otherwise dirt may adhere to
the Check Strip and damage
it.

2. Set the Test Strip Type.


Set the Test Strip Type to C (Check).
See“Setting Up Parameters”.

3. Prepare a Check Strip.


Remove 1 Check Strip from the Check Strip
container.
Do NOT touch the gray-
Important
colored part of Check Strip.

Check Strip(Gray)

Semiquantitative Value Label

e
19 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Check Measurement Measurement

4. Press [START] to start Check


Measurement.
While the Test Strip Icon is blinking, place the
Check Strip onto the Strip Holder.
A few seconds before the Strip
Reference
Holder moves inside of the
Analyzer, the Test Strip Icon
stops blinking.
When the Strip Holder is set
inside the Analyzer, the Test
Strip Icon disappears.

5. Confirm the results.


Compare the Check Measurement results with the ranges of semiquantitative
values attached to the Check Strips container.

●When the Check Measurement results are within the range:


The Analyzer is performing correctly. End Check Measurement.

●When the Check Measurement results are out of the range:


Repeat the Check Measurement with the other Check Strip.

6. Repeat the Check Measurement with


the other Check Strip.
●If the second Check Measurement results are within the range:
The Check Strip used for the first measurement was defective.
Do NOT use it again.

●If the second Check Measurement results are still out of the range:
Contact your distributor.

AUTION MICRO 20
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Auxiliary Function Measurement

1. Calendar Display and Setting.


① Press [DATE] from the Ready Screen, the current
date is displayed. (The Ready Screen is restored
in 2-3 seconds automatically.)

② Press [DATE] again while the date is displayed,


the Date Setting Screen appears and“DATE”
and cursor blink.

③ Enter the current date.

④ Press [ENT] to confirm the entry. Press [STOP]


to cancel, and the Ready Screen is restored.

2. Time Display and Setting.


① Press [TIME] from the Ready Screen, current
time is displayed.
(The Ready Screen is restored in 2-3 seconds
automatically).

② Press [TIME] again while the time is displayed,


the Time Setting Screen appears and [TIME]
and cursor blink.

③ Enter the current time.

④ Press [ENT] to confirm the entry.


Press [STOP] to cancel, and the Ready Screen
is restored.

21 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Auxiliary Function Measurement

3. Redisplay of Measurement
Results.
① Press [MEMORY▲] or [MEMORY▼] from the
Ready screen, the latest data is redisplayed
([MEMORY] is displayed at the top left of the
screen).

② Press [MEMORY▼] again, one previous data is


displayed.
Repeat this procedure, if necessary.

③ Press [MEMORY▲], the next data is


displayed.

④ Press [STOP], the Ready Screen is restored.


If No., ID, DATE and TIME are
Reference
pressed during Redisplay of
data, Measurement No.,
Patient ID, Measurement Date
and Measurement Time on the
Redisplayed data are displayed.
(Measurement No. is displayed
as default. This setting is
reflected in all data during a
series of redisplay operation.)

AUTION MICRO 22
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Auxiliary Function Measurement

4. Delete Data.
A: Delete Single Data.
① Press [DELETE] on the Redisplay screen of
measurement results,“DELETE”blinks.
(The measurement results are also displayed.)

② Press [ENT] to delete the displayed measurement


data and the next data is displayed.
To cancel it, press [STOP] and the Redisplay
screen is restored.

B: Delete All Data.


① Keep pressing [DELETE] on the Redisplay screen
of the measurement results for 2-3 seconds,
“DELETE”blinks and“ALL”is displayed.
(Other displayed items disappear.)

② Press [ENT] to delete all measurement data in the


memory.
To cancel it, press [STOP] and the Redisplay
screen is restored.

5. Retransmit.
A: Retransmit Single Data.
① Press [SEND] on the Redisplay screen of the
measurement results, the Transmission Icon blinks.
(The measurement Results are also displayed.)

② Press [ENT] to transmit the displayed measurement


data. To cancel it, press [STOP] and the Redisplay
screen is restored.

23 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Auxiliary Function Measurement

B: Retransmit All Data.


① Keep pressing [SEND] on the Redisplay screen of
the Measurement Results for 2-3 seconds, the
Transmission Icon blinks and“ALL”is displayed.
(Other displayed items disappear.)

② Press [ENT] to transmit all measurement data in


the memory from the latest data.
To cancel it, press [STOP] and the Redisplay screen
is restored.
To stop transmission, press
Reference
[STOP] during transmission.

6. Reprint.
(when the Printer is connected.)
A: Reprint Single Data.
① Press [PRINT] on the Redisplay screen of the
measurement results, the Printer Icon blinks.
(The measurement results are also displayed.)

② Press [ENT] to print the displayed measurement


data. To cancel it, press [STOP] and the Redisplay
screen is restored.

B: Reprint All Data.


① Keep pressing [PRINT] on the Redisplay screen of
the measurement results for 2-3 seconds, the
Printer Icon blinks and [All] is displayed.
(Other displayed items disappear.)

AUTION MICRO 24
DOC. No. OM-28373 REV.01 01/01 AUTION MICRO PU-4210 OPERATING MANUAL

Auxiliary Function Measurement

② Press [ENT] to print all measurement data in


memory from the latest data. To cancel it, press
[STOP] and the Redisplay screen is restored.
To stop printing, press [STOP]
Reference
during printing.
To print the latest data,press
[PRINT] on the Ready Screen.

7. Battery Icon.
A battery icon at the lower right of the analyzer
display tells you the timing of battery exchange.
Follow the instructions below, and replace the
battery to a new one if necessary.
The supplied AC adapter can be used as well.

① The battery is full with power.


②③ Although measurement can be continued


with this battery, the power is getting ②
short. Prepare a new battery.

This icon sometimes


Note
changes to the next icon

(blinking) even if enough
power remains in.

④ The battery runs out. ④


To continue measurement, replace the
battery to a new one.
(Blinking)

25 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Special Parameters Measurement

The following special parameters are available to set with the PU-4210.
(The parameters are stored until the settings are changed.)
●Orientation of the Test Strip set.
Setting of Test Strip from the right side or from the left side of the Analyzer
(default is set to the right side).
To change the Test Strip orientation, remove the Strip
Note
Holder from the Analyzer beforehand.
●Date Order.
Choose Date order among Y-M-D, M-D-Y, and D-M-Y.
(The default setting is Y-M-D).
Setting is reflected on the screen display and printing. (Fixed for external output.)
●Setting of measurement units and symbols for semi-quantitative ranks.
Set the measurement units for the measurement results (SI or usual) and format
of semi-quantitative ranks ( +++ or +n ) ( default is set to usual, +++).
Setting is reflected on the printing and external output. (Fixed for the screen display.)

1. Operation.
previous setting
① Turn on the power while pressing [STOP].
(Keep pressing [STOP] during initialization.)
② After initialization, the entry screen for 3 digits
numbers is displayed and the parameter setting
mode appears.
③ The numeral of each digit corresponds to the
following. Input the numerals for the desired setting.
In this case, if numerals other than the designated
ones are input, they are not accepted.
Hundreds 1:Right setting 2:Left setting
Tens 1:Y-M-D 2:M-D-Y 3:D-M-Y
Ones 1:mg/dl,+++ 2:mg/dl,+n 3:mmol/l,+++ 4:mmol/l,+n
④ Press [ENT] after the entry to change the settings.
The Ready Screen is restored.
e.g.) Change from the default (Right setting, Y-M-D, mg/dl,+++)
●When“232”is entered,(Left setting, D-M-Y, mg/dl, +n) is set. Then the Ready Screen is restored.
●When“063”is entered,(Right setting, Y-M-D, mmol/l, +++) is set. Then the Ready Screen is restored.

AUTION MICRO 26
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL

Printer Printing

1. Attachment or Removal of
Printer.
When connecting the Printer to the Analyzer,
insert the hook of the Printer in the hole at the
bottom of the Analyzer, and then slide it firmly
enough to achieve a secure fit.
When removing the Printer, slide the Analyzer lock release button

to the front while pressing the lock release button of the Printer.
Before you attach or remove the Printer, make sure to turn
Note
off the power and pull out the AC adapter of the Printer.

2. Paper Jam.
When paper is jammed, press [FEED] of the
FEED
Printer to remove the paper.

3. Loading of Paper.
After confirming the face and the back of the
paper, put the thermal printer paper into the
insertion position, and then the paper will be
fed automatically.
When inserting a new
Note
thermal Printer paper,
cut the edge with scissors
to make it even, and set it in
the position.

4. Error Comments.
When an error occurs, error codes and error comments are printed.
(Only error codes are displayed on the screen.)
※See“Troubleshooting”.

27 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Printing Item Printing

Measurement No.

Abnormal Mark
Patient ID
Date・Time・
Type of Test Strip・
Temperature in Instrument
Marks for
abnormal values
(Normal:Blank) Measurement Results

Measurement Units

Measurement Item Ranks Concentration values


(Semi-quantitative symbols)

Abnormal Mark When the measurement results is abnormal (Rank is more than“+1”.),
sample abnormal mark‘●’is displayed.

Measurement No. 4 digits

Patient ID ID is entered using keys. When there is no ID, [------------------]is printed.

Date, Time There are 3 printing patterns of the date


YMD, MDY, DMY
Example 2001-07-05 09:53
      07-05-2001 09:53
     05-07-2001 09:53

Type of Strip The Test Strip Type is displayed.


Example 9UB

Temperature in Temperature in the instrument is printed in ℃


Instrument Example 29℃

Measurement Abnormal Mark /


Results When an abnormal measurement result (rank is over“+1”) is recognized,
‘*’is printed before the name of item.
Measurement Item /
GLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G.
Ranks (Semi-quantitative symbols), Concentration,
Measurement Units.

AUTION MICRO 28
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL

Maintenance Troubleshooting and Maintenance

Clean the Strip Holder after finishing measurement at the end of the day.

Wear protective gloves to prevent exposure to pathogenic


microbes.

Discard used cotton swabs and used gloves according to


local regulations for biohazardous waste.

1. Slide the Strip Holder to remove. 2. Wash the Strip Holder with
detergent or alcohol and then rinse
it with tap water. Especially, clean
the groove with special care.

3. Dry the Strip Holder completely 4. Insert the Strip Holder completely
with a soft cloth. by using grooves of the backside of
the Strip Holder.
e
29 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Troubleshooting Troubleshooting and Maintenance

When measurement and/or printing trouble occurs, the Analyzer displays


error codes. Perform the appropriate remedy described in the following
chart. If measurement is still impossible, contact your distributor.

Error Code
Error Comment Cause Remedy

E001 Optical System does not Make sure that the Strip Holder
Optical function correctly. and Test Strip are set properly
Movement
and turn on the power.

E002 Lamp intensity of LED is Clean the White Plate of the Strip
Lamp Intensity
High
too strong. Holder with a soft cloth
moistened with water or alcohol
and turn on the power.

E003 Lamp intensity of LED is Clean the White Plate of the Strip
Lamp Intensity too weak. Holder with a soft cloth
Low
moistened with water or alcohol
and turn on the power.

E004 No Printer Paper Load printer paper.


(Press [STOP] to restore the
Ready Screen.)

E005 Printer does not function Make sure that the Analyzer and
correctly. the printer are connected
correctly. (Press [STOP] to restore
the Ready Screen.)

E006 The Test Strip is not Insert the Test Strip and Strip
Strip Error inserted correctly. Holder correctly and start
measurement again.
(Press [STOP] to restore the
Ready Screen.)

Important
Test Strip type is not judged.

AUTION MICRO 30
Specifications
Analyzer
No. Items Contents
 1 Sample Urine
2 Test Strip AUTION Sticks 10EA, Uriflet 9UB
3 Measurement Item GLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G.
4 Measurement Method Test strip
Dual or Single (BLD only) Wavelength Reflectance Method
5 Measurement Wavelength Triple Wavelength, Individual LED (565, 632, 755nm)
6 Sample Supply Manual Dipping
7 Reaction Time Approx. 45 seconds/test
8 Calibration Function Automatic Calibration by CAL pad
9 Display LCD
10 Operator Panel 12 keys
11 Memory Capacity 100 tests
12 Thermal Correction Function Automatic Correction Function (10-30℃)
13 SG Correction Function Automatic Correction according to pH measurement value
14 External Output Conformed to RS-232C
15 Measurement Environment Condition Temperature:10∼30℃, Humidity:20∼80%RH
16 Dimensions 125(W)×87(D)×31(H) mm
17 Weight Approximately 300g
18 Power Supply Size AA alkali cells (×2) (Or AC adapter)

Printer
No. Items Contents
 1 Printer 36 digits of half-sized characters, Line Thermal Printer
2 Printer Paper Thermal Printer Paper 58mm (Width)×φ26mm
3 Dimensions 125(W)×163(D)×38(H) mm
4 Weight Approximately 300g
5 Power Supply AC adapter (Approximately 20W)

84-00259E