Operating Manual
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL
Premise
This manual contains important information on the
functions of the AUTION MICRO PU-4210.
e
1 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05
Introduction
This manual contains outlines, instructions for the
operation, maintenance and troubleshooting for the
AUTION MICRO PU-4210. Read this manual carefully
before use the Analyzer in order to operate it correctly.
After you read the manual, keep it near the Analyzer to
refer to it when necessary.
e
AUTION MICRO 2
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL
Caution Marks
Pay careful attention to these caution marks in this Operating Manual.
e
3 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
●Do NOT place the Analyzer where it may ●Select a room in which to set up the
be affected by chemicals, corrosive gases Analyzer where the temperature can be
and electronic noise. controlled between 10℃ and 30℃, and
humidity in a range of 20 to 80 %.
●Do NOT place the Analyzer where it may ●Do NOT place the Analyzer where it may
be affected by direct sunlight or wind. be affected by magnetic materials including
magnets or speakers.
AUTION MICRO 4
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
●Should the Analyzer stop functioning, ●If an abnormal sound, odor or smoke is
contact your distributor for repairs. Do NOT found, turn the power off immediately and
attempt to repair it your self. Doing so may consult your distributor after making sure
result in damaging the Analyzer or injury. that the abnormal situation is not continuing.
●Do NOT place the Analyzer where it may ●Do NOT drop or hit the Analyzer.
be affected by water. Samples may cause
damage to the Analyzer.
5 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
Table of Contents
AUTION MICRO(PU-4210)Operating Manual
Premise …………………………………………………………1
Introduction ……………………………………………………2
Caution Marks …………………………………………………3
Precautions for Installation …………………………………4
Table of Contents ……………………………………………6
AUTION MICRO 6
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
Characteristics
Minimum Maintenance
The only necessary maintenance item is cleaning the Strip Holder.
Power Source
Use a dry battery or a AC adapter when using the Analyzer alone, and use the
AC adapter when using it with a Printer.
Regent Selection
AUTION Sticks 10EA and Uriflet 9UB are available to use.
100 memories
Maximum 100 measurement results can be stored.
When more than 100 samples have been measured, older data will be automatically
deleted in order.
7 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
Analyzer Printer
Operating Manual
Optional Optional
Accessories Consumables
Strip Holder
AUTION MICRO 8
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
Front
8
1 2
3
9
9 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
LCD
5. Date
1. Memory 4. Delete
2. ID 6. Time
3. No.
7. Test Strip
9. Item Name
AUTION MICRO 10
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL
Start Measurement
Dip the Test Strip in the sample for 2 seconds in accordance with the alarm.
The Strip Holder moves inside of the Analyzer and measurement starts.
Measurement Results are displayed and the Strip Holder comes out from the Analyzer.
Continue measurement
Dispose of Test Strip.
Finish measurement
Clean the Strip Holder, and turn off the main power.
e
11 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
Cautions Measurement
Handling Samples
● Use a fresh urine sample. If the sample urine is not measured within an
hour after collection, cap it tightly and keep it refrigerated. Be sure to
allow it to reach to room temperature before measurement.
AUTION MICRO 12
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
Cautions Measurement
13 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05
Start Up Measurement
1. Preparation.
① When using the Printer, make sure that the
Printer is connected to the Analyzer correctly
and paper is loaded. If there is a red strip on
the both sides of the paper, replace the roll.
15 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05
1. Prepare sample.
① Prepare sample in a sample cup.
Make sure that sample is
Important
enough to cover the whole Reagent Pads
reagent pads.
e
AUTION MICRO 16
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
17 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
7. Measurement.
The Strip Holder is pulled into the Analyzer
and the measurement starts.
8. Finish Measurement.
When measurement is completed,the
measurement result is displayed and the
Strip Holder comes out from the Analyzer.
Discard the used Test Strip, and when
finishing the whole measurement process,
remove the Strip Holder from the Analyzer
and clean it.
AUTION MICRO 18
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL
Check Strip(Gray)
e
19 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
●If the second Check Measurement results are still out of the range:
Contact your distributor.
AUTION MICRO 20
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
21 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
3. Redisplay of Measurement
Results.
① Press [MEMORY▲] or [MEMORY▼] from the
Ready screen, the latest data is redisplayed
([MEMORY] is displayed at the top left of the
screen).
AUTION MICRO 22
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
4. Delete Data.
A: Delete Single Data.
① Press [DELETE] on the Redisplay screen of
measurement results,“DELETE”blinks.
(The measurement results are also displayed.)
5. Retransmit.
A: Retransmit Single Data.
① Press [SEND] on the Redisplay screen of the
measurement results, the Transmission Icon blinks.
(The measurement Results are also displayed.)
23 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
6. Reprint.
(when the Printer is connected.)
A: Reprint Single Data.
① Press [PRINT] on the Redisplay screen of the
measurement results, the Printer Icon blinks.
(The measurement results are also displayed.)
AUTION MICRO 24
DOC. No. OM-28373 REV.01 01/01 AUTION MICRO PU-4210 OPERATING MANUAL
7. Battery Icon.
A battery icon at the lower right of the analyzer
display tells you the timing of battery exchange.
Follow the instructions below, and replace the
battery to a new one if necessary.
The supplied AC adapter can be used as well.
25 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
The following special parameters are available to set with the PU-4210.
(The parameters are stored until the settings are changed.)
●Orientation of the Test Strip set.
Setting of Test Strip from the right side or from the left side of the Analyzer
(default is set to the right side).
To change the Test Strip orientation, remove the Strip
Note
Holder from the Analyzer beforehand.
●Date Order.
Choose Date order among Y-M-D, M-D-Y, and D-M-Y.
(The default setting is Y-M-D).
Setting is reflected on the screen display and printing. (Fixed for external output.)
●Setting of measurement units and symbols for semi-quantitative ranks.
Set the measurement units for the measurement results (SI or usual) and format
of semi-quantitative ranks ( +++ or +n ) ( default is set to usual, +++).
Setting is reflected on the printing and external output. (Fixed for the screen display.)
1. Operation.
previous setting
① Turn on the power while pressing [STOP].
(Keep pressing [STOP] during initialization.)
② After initialization, the entry screen for 3 digits
numbers is displayed and the parameter setting
mode appears.
③ The numeral of each digit corresponds to the
following. Input the numerals for the desired setting.
In this case, if numerals other than the designated
ones are input, they are not accepted.
Hundreds 1:Right setting 2:Left setting
Tens 1:Y-M-D 2:M-D-Y 3:D-M-Y
Ones 1:mg/dl,+++ 2:mg/dl,+n 3:mmol/l,+++ 4:mmol/l,+n
④ Press [ENT] after the entry to change the settings.
The Ready Screen is restored.
e.g.) Change from the default (Right setting, Y-M-D, mg/dl,+++)
●When“232”is entered,(Left setting, D-M-Y, mg/dl, +n) is set. Then the Ready Screen is restored.
●When“063”is entered,(Right setting, Y-M-D, mmol/l, +++) is set. Then the Ready Screen is restored.
AUTION MICRO 26
DOC. No. OM-28373 REV.0 04/01 AUTION MICRO PU-4210 OPERATING MANUAL
Printer Printing
1. Attachment or Removal of
Printer.
When connecting the Printer to the Analyzer,
insert the hook of the Printer in the hole at the
bottom of the Analyzer, and then slide it firmly
enough to achieve a secure fit.
When removing the Printer, slide the Analyzer lock release button
to the front while pressing the lock release button of the Printer.
Before you attach or remove the Printer, make sure to turn
Note
off the power and pull out the AC adapter of the Printer.
2. Paper Jam.
When paper is jammed, press [FEED] of the
FEED
Printer to remove the paper.
3. Loading of Paper.
After confirming the face and the back of the
paper, put the thermal printer paper into the
insertion position, and then the paper will be
fed automatically.
When inserting a new
Note
thermal Printer paper,
cut the edge with scissors
to make it even, and set it in
the position.
4. Error Comments.
When an error occurs, error codes and error comments are printed.
(Only error codes are displayed on the screen.)
※See“Troubleshooting”.
27 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
Measurement No.
Abnormal Mark
Patient ID
Date・Time・
Type of Test Strip・
Temperature in Instrument
Marks for
abnormal values
(Normal:Blank) Measurement Results
Measurement Units
Abnormal Mark When the measurement results is abnormal (Rank is more than“+1”.),
sample abnormal mark‘●’is displayed.
AUTION MICRO 28
DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL
Clean the Strip Holder after finishing measurement at the end of the day.
1. Slide the Strip Holder to remove. 2. Wash the Strip Holder with
detergent or alcohol and then rinse
it with tap water. Especially, clean
the groove with special care.
3. Dry the Strip Holder completely 4. Insert the Strip Holder completely
with a soft cloth. by using grooves of the backside of
the Strip Holder.
e
29 AUTION MICRO
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
Error Code
Error Comment Cause Remedy
E001 Optical System does not Make sure that the Strip Holder
Optical function correctly. and Test Strip are set properly
Movement
and turn on the power.
E002 Lamp intensity of LED is Clean the White Plate of the Strip
Lamp Intensity
High
too strong. Holder with a soft cloth
moistened with water or alcohol
and turn on the power.
E003 Lamp intensity of LED is Clean the White Plate of the Strip
Lamp Intensity too weak. Holder with a soft cloth
Low
moistened with water or alcohol
and turn on the power.
E005 Printer does not function Make sure that the Analyzer and
correctly. the printer are connected
correctly. (Press [STOP] to restore
the Ready Screen.)
E006 The Test Strip is not Insert the Test Strip and Strip
Strip Error inserted correctly. Holder correctly and start
measurement again.
(Press [STOP] to restore the
Ready Screen.)
Important
Test Strip type is not judged.
AUTION MICRO 30
Specifications
Analyzer
No. Items Contents
1 Sample Urine
2 Test Strip AUTION Sticks 10EA, Uriflet 9UB
3 Measurement Item GLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G.
4 Measurement Method Test strip
Dual or Single (BLD only) Wavelength Reflectance Method
5 Measurement Wavelength Triple Wavelength, Individual LED (565, 632, 755nm)
6 Sample Supply Manual Dipping
7 Reaction Time Approx. 45 seconds/test
8 Calibration Function Automatic Calibration by CAL pad
9 Display LCD
10 Operator Panel 12 keys
11 Memory Capacity 100 tests
12 Thermal Correction Function Automatic Correction Function (10-30℃)
13 SG Correction Function Automatic Correction according to pH measurement value
14 External Output Conformed to RS-232C
15 Measurement Environment Condition Temperature:10∼30℃, Humidity:20∼80%RH
16 Dimensions 125(W)×87(D)×31(H) mm
17 Weight Approximately 300g
18 Power Supply Size AA alkali cells (×2) (Or AC adapter)
Printer
No. Items Contents
1 Printer 36 digits of half-sized characters, Line Thermal Printer
2 Printer Paper Thermal Printer Paper 58mm (Width)×φ26mm
3 Dimensions 125(W)×163(D)×38(H) mm
4 Weight Approximately 300g
5 Power Supply AC adapter (Approximately 20W)