CERTIFICATE OF CALIBRATION

CITADEL ISO/IEC17025 ACCREDITION LABORATORY


Issued By: Citadel Company Certificate Number: CEC-2020/CAD/CP-0002
Date of Issue: 05 February 2020 CL011

No. 201, Italian Village, Erbil Kurdistan Region, Iraq Approved Signatory
Phone Number: +964 -7506489336/+964 –7717023040 Chia Burhan
Website: www.citadel-engineering.com
Email: info@citadel-engineering.com

Customer:
Name:Crescent Petroleum & Dana Gas Address: Office 301 & 302, World Trade Center, Erbil, Kurdistan Region,
Iraq
Date of Calibration 09 March 2020 Due date of Calibration: 09 March 2021
Unit Under Calibration (U.U.C):
Description: Dry Well Calibrator
Manufacturer: FLUKE
Model / Type: 9141
Serial/Tag No.: A8A888 / SPTI024
Environmental Condition:
Temperature: 24 °C +/- 1°C
Relative Humidity: 45% °C +/- 10%
Location of Calibration: CEC YARD.
Calibration Procedure:
• BIPM, IEC, IFCC, ISO, IUPAG, IUPAP, OIML: 1993; International/ Vocabulary of Basic and General Terms in
Metrology.
• EA-4/02: 2013; Evaluation of the Uncertainty of Measurement in Calibration
• EURAMET/cg-15/v Guidelines on the Calibration of Digital Multimeters
• Calibration Philosophy In practice - Second Edition, Fluke Corporation, 1994
Calibration Information:
Traceability:
The instrument was calibrated against laboratory standards whose values are traceable to recognized National Standards.
The uncertainty limits quoted refer to the measured values only, with no account being taken of the instruments ability
to maintain its calibration.
Uncertainty:
The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a
level of confidence of approximately 95%. The uncertainty evaluation has been carried out in accordance with
ISO/IEC GUIDE98-3:2008 requirements.
Calibration equipment used:
No. Description Model Manufacture Serial No. Cal. due date Certificate number
1 RTD Calibrator 712 Fluke 9728181 1 Mar 2021 CEC-2020/CAD/CEC-0029/CEC
2 Thermocouple Calibrator 714B Fluke 27030068 N/A N/A
5 Thermocouple TYPE K NAGMAN T11013 30 Mar 2020 3092/3/NAGMAN
9 RTD 4 Wire PT 100 NAGMAN T1102 30 Mar 2020 3092/2-NAGMAN

1. This certificate is issued in accordance with the laboratory accreditation requirements of the Iraqi Accreditation System (IQAS).
2. This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory.

Doc No: CEC[EN]-IND/F/10-01 Page 1 of 2 Revision Date: 10/05/2019


CERTIFICATE OF CALIBRATION
CITADEL ISO/IEC17025 ACCREDITION LABORATORY
Issued By: Citadel Company Certificate Number: CEC-2020/CAD/CP-0002
Date of Issue: 05 February 2020 CL011

CALIBRATION RESULT
Deviation
Temperature UUT Setting Ref. Reading
Step REF. - UUT
(°C) (°C) (°C)
(°C)
1 50 50 49.80 0.20
2 100 100 99.61 0.39
3 150 150 149.60 0.40
4 200 200 199.57 0.43
5 250 250 249.55 0.45
6 300 300 299.52 0.48
7 350 350 349.50 0.50
8 400 400 400.6 -0.60
9 450 450 451.0 -1.0
10 500 500 501.0 -1.0
11 550 550 550.7 -0.7
12 600 600 600.4 -0.4
13 650 650 649.7 0.3

Unit under calibration (U.U.c):


Range: 50 ~ 650 °C
Accuracy : ± 0.5 °C to 400 °C; ± 1.0 °C to 650 °C

Calibrated by: Farooq.S Reviewed by: Hojat Movafegh Nia


Signature: Signature:
Date: 09 March 2020 Date: 11 March 2020

End of Document